Skip to content

VersaTie 8510260Recalls & reported problems

Nuvasive, Inc. · VersaTie Clamp, 5-6mm Side Loading · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517709875), received between Apr 25, 2019 and Aug 27, 2020 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Material Rupture2
  • Output above Specifications1

Patient problems

  • Bone Fracture(s)1
  • Failure of Implant1
  • Pain1
  • Spinal Column Injury1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2020Material Rupture; Material Rupture

    ON (B)(6) 2020, PATIENT UNDERWENT A REVISION PROCEDURE FOR AN UNKNOWN REASONS WHERE THE CONSTRUCT WAS EXTENDED TO THE L2 LEVEL AND SUBLAMINAR BONE FIXATION CERCLAGE WAS UTILIZED. AN ADDITIONAL REVISION PROCEDURE WAS REPORTED TO HAVE OCCURRED ON (B)(6) 2020 FOR UNKNOWN REASONS AND IT WAS IDENTIFIED THAT IMPLANTED BANDS HAD RUPTURED. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 10462716 ↗
  • MalfunctionAug 7, 2020Material Rupture; Material Rupture

    PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2019, WHERE THE CONSTRUCT WAS EXTENDED TO THE T10 LEVEL. THE PATIENT BEGAN TO EXPERIENCE PAIN AT THE TOP OF THE CONSTRUCT. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER REPORTER, IT WAS CONFIRMED THAT 4 OF THE 6 IMPLANTED BANDS HAD RUPTURED, WHICH IN TURN HAD LOOSENED THE CONSTRUCTS SCREWS. THE PATIENT REPORTED NO ISSUES POST THE REVISION PROCEDURE.

    Patient problems: Pain

    MAUDE report 10377330 ↗
  • InjuryApr 25, 2019Output above Specifications; Output above Specifications

    ON (B)(6) 2019 A PATIENT UNDERWENT A POSTERIOR FIXATION PROCEDURE UTILIZING THE VERSATIE SYSTEM. AS PER REPORTER WHILE TENSION WAS BEING PERFORMED THE BAND RIPPED THE SPINOUS PROCESS. THE SURGEON MOVED TO THE NEXT LEVEL AND THE SAME THING OCCURRED. ALTERNATIVE FIXATION DEVICE WAS UTILIZED. THE PATIENT WAS REPORTED TO HAVE RE-COOPERATED.

    Patient problems: Bone Fracture(s); Spinal Column Injury

    MAUDE report 8547866 ↗

FDA clearance and classification

Cleared through Special 510(k) K190418 on Mar 21, 2019 (decision: Substantially Equivalent). Submitted by Nu Vasive, Incorporated as “NuVasive® VersaTie® System”. The FDA review took 28 days. FDA record ↗

Cleared through 510(k) K173117 on Dec 7, 2017 (decision: Substantially Equivalent). Submitted by Nuvasive, Inc. as “NuVasive® VersaTie™ System”. The FDA review took 69 days. FDA record ↗

Cleared through 510(k) K161265 on Jul 15, 2016 (decision: Substantially Equivalent). Submitted by Nu Vasive, Incorporated as “NuVasive® VersaTie System”. The FDA review took 71 days. FDA record ↗

The FDA regulates product code OWI (bone fixation cerclage, sublaminar) under 21 CFR 888.3010 as a Class II device. 21 CFR 888.3010 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Acts as a bone anchor for temporary stabilization, or used in conjunction with other medical implants of similar metals when wiring is needed, during development of a spinal fusion.

Common questions about the VersaTie 8510260

Has the VersaTie 8510260 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00887517709875), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VersaTie 8510260?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Apr 25, 2019 and Aug 27, 2020. In those reports, the most common specific device problems are “Material Rupture” (2) and “Output above Specifications” (1); the most common patient problems are “Bone Fracture(s)” (1), “Failure of Implant” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VersaTie 8510260?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VersaTie 8510260 FDA approved?

It was cleared through 510(k) K161265 on Jul 15, 2016; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the VersaTie 8510260 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the VersaTie 8510260, and is it still sold?

The FDA's device database lists Nuvasive, Inc. as the company responsible for it (brand VersaTie, model 8510260). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Bone fixation cerclage, sublaminar devices

All Bone fixation cerclage, sublaminar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page