JAZZ LOCKRecalls & reported problems
Implanet · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03607503507004), received between Feb 16, 2026 and Feb 16, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth3
- Migration or Expulsion of Device3
Patient problems
- Unspecified Infection2
- Inadequate Pain Relief1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2026Migration or Expulsion of Device; Loosening of Implant Not Related to Bone-Ingrowth
2X JAZZ BAND AND 2X JAZZ LOCK USED IN A POSTERIOR LUMBAR FUSION, (B)(6) 2024. BANDS PASSED AROUND THE UIV+1 (L1) SPINOUS PROCESS AND TENSIONED AROUND THE ROD BILATERALLY. REOPERATION ON (B)(6) 2024. SURGEON NOTED THAT IN ONE LOCK CONNECTOR BOTH PORTIONS OF THE BAND WERE OUT OF THE LOCK CONNECTOR AND IN FOR THE SECOND LOCK CONNECTOR ONE PORTION OF THE BAND REMAINED IN THE CONNECTOR. THE REASON FOR THE RE-OPERATION WAS PATIENT WAS STILL SYMPTOMATIC, SO THE REOPERATION WAS FOR FURTHER DECOMPRESSION, AND THEY FOUND SLIPPING OF THE BAND UPON INSPECTION. DUE TO THE PATIENT'S PATHOLOGY, THE SURGEON EXTENDED THE FIXATION UP ONE LEVEL WITH PEDICLE SCREWS AND RODS. SURGICAL INDICATION : LUMBAR DECOMPRESSION AND FUSION. TYPE OF CONSTRUCT: PLIF + PEDICLE SCREWS + JAZZ LOCK. THIS REPORT IS MADE AT THAT DATE FURTHER TO THE OBSERVATIONS WRITTEN IN A FORM FDA 483 BY (B)(6), FDA INVESTIGATOR ON JULY 4, 2025.
Patient problems: Inadequate Pain Relief
MAUDE report 24342935 ↗MalfunctionFeb 16, 2026Migration or Expulsion of Device; Loosening of Implant Not Related to Bone-Ingrowth
TWO JAZZ BAND IMPLANTS (LOT#: 210954 AND LOT#: 220386) AND TWO JAZZ LOCK IMPLANTS (LOT#: LXJA-LXJI AND LOT#: MQXZ-LXJM) WERE IMPLANTED IN A 58-YEAR-OLD PATIENT ON (B)(6) 2024 AT (B)(6) HOSPITAL (USA) DURING A POSTERIOR LUMBAR FUSION SURGERY. REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 DUE TO INFECTION AND THE SURGEON NOTED THAT THE BANDS WERE "LOOSE". THIS REPORT IS MADE AT THAT DATE FURTHER TO THE OBSERVATIONS WRITTEN IN A FORM FDA 483 BY MR (B)(6), FDA INVESTIGATOR ON JULY 4, 2025.
Patient problems: Unspecified Infection
MAUDE report 24342934 ↗MalfunctionFeb 16, 2026Migration or Expulsion of Device; Loosening of Implant Not Related to Bone-Ingrowth
TWO JAZZ BAND IMPLANTS (LOT # 210954 AND LOT # 220386) AND TWO JAZZ LOCK IMPLANTS (LOT # LXJA-LXJI AND LOT # MQXZ-LXJM) WERE IMPLANTED IN A 58-YEAR-OLD PATIENT ON (B)(6) 2024 AT (B)(6) HOSPITAL DURING A POSTERIOR LUMBAR FUSION SURGERY. REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 DUE TO INFECTION AND THE SURGEON NOTED THAT THE BANDS WERE "LOOSE". THIS REPORT IS MADE AT THAT DATE FURTHER TO THE OBSERVATIONS WRITTEN IN A FORM FDA 483 BY MR (B)(4), FDA INVESTIGATOR ON JULY 4, 2025.
Patient problems: Unspecified Infection
MAUDE report 24342933 ↗
FDA clearance and classification
Cleared through 510(k) K153348 on Mar 31, 2016 (decision: Substantially Equivalent). Submitted by Implanet, S.A. as “Jazz Lock”. The FDA review took 132 days. FDA record ↗
The FDA regulates product code OWI (bone fixation cerclage, sublaminar) under 21 CFR 888.3010 as a Class II device. 21 CFR 888.3010 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Acts as a bone anchor for temporary stabilization, or used in conjunction with other medical implants of similar metals when wiring is needed, during development of a spinal fusion.
Common questions about the JAZZ LOCK
Has the JAZZ LOCK been recalled?
No FDA recall record in our data names this exact device identifier (DI 03607503507004), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the JAZZ LOCK?
The FDA's MAUDE database lists 3 adverse event reports for this device, received on Feb 16, 2026. In those reports, the most common specific device problems are “Loosening of Implant Not Related to Bone-Ingrowth” (3) and “Migration or Expulsion of Device” (3); the most common patient problems are “Unspecified Infection” (2) and “Inadequate Pain Relief” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the JAZZ LOCK?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the JAZZ LOCK FDA approved?
It was cleared through 510(k) K153348 on Mar 31, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the JAZZ LOCK, and is it still sold?
The FDA's device database lists Implanet as the company responsible for it (brand JAZZ LOCK, model JAZZ LOCK). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Bone Fixation Cerclage, Sublaminar devices
- TRANSLACE™ Spinal Tethering System 8219500Medtronic Sofamor Danek, Inc. · Bone fixation cerclage, sublaminar11 reports
- VersaTie 86101245Nuvasive, Inc. · Bone fixation cerclage, sublaminar4 reports
- VersaTie 8510260Nuvasive, Inc. · Bone fixation cerclage, sublaminar3 reports
- LIGAPASS B08110005Medicrea International · Bone Fixation Cerclage, Sublaminar3 reports
- KARMA 75035-175Spinal Elements, Inc. · Bone Fixation Cerclage, Sublaminar3 reports
- VersaTie 8450001Nuvasive, Inc. · Bone fixation cerclage, sublaminar2 reports
- TRANSLACE™ Spinal Tethering System 8205300Medtronic Sofamor Danek, Inc. · Bone fixation cerclage, sublaminar2 reports
- NILE™ Alternative Fixation System 5416-F04730-SGVB Spine LLC · Bone fixation cerclage, sublaminar2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.