Midas Rex EM200-R (UDI 00763000174859)Recalls & reported problems
Medtronic Powered Surgical Solutions · MOTOR EM200-R LEGEND EHS STYLUS REFURB · Class 2
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00763000174859), received between Mar 4, 2019 and Mar 25, 2025 — classified as 1 injury, 13 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device13
- Device Displays Incorrect Message1
- Loss of Power1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Consequences Or Impact To Patient1
- Perforation1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2025Overheating of Device; Loss of Power; Device Displays Incorrect Message; Overheating of Device; Loss of Power; Device Displays Incorrect Message
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING AND ERROR MESSAGE GIVEN AND STALLING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21685049 ↗MalfunctionMar 5, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH REPORT OF OVERHEATING.. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21522819 ↗MalfunctionFeb 25, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21461257 ↗MalfunctionFeb 24, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF DEVICE OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21455287 ↗MalfunctionFeb 21, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431337 ↗MalfunctionFeb 12, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21365622 ↗MalfunctionFeb 12, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21365315 ↗MalfunctionFeb 11, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT DUE TO REASON FOR RETURN AND RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR. ON FOLLOW UP IT WAS REPORTED THAT THERE IS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21353692 ↗MalfunctionFeb 11, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR THE DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21352958 ↗MalfunctionFeb 10, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN AND DUE TO RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21341645 ↗MalfunctionFeb 10, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN AND DUE TO RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21341403 ↗MalfunctionFeb 7, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN AND DUE TO RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21331599 ↗MalfunctionNov 26, 2019Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING AND NOISE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN. ON FOLLOW UP IT WAS CONFIRMED THAT THERE WAS NO PATIENT OR STAFF IMPACT AS THE DEVICE WAS TESTED BEFORE THE PATIENT WAS BROUGHT INSIDE THE OPERATING ROOM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9375057 ↗InjuryMar 4, 2019Vibration; Vibration
IT WAS REPORTED THAT DURING SURGERY, THE FIRST MOTOR CHATTERED SO MUCH IT CAUSED A TEAR IN THE DURA. TWO MOTORS WERE USED AFTERWARDS, WHICH BOTH OVERHEATED, AND THE THIRD ONE ALSO CHATTERED. ON FOLLOW UP, IT WAS MENTIONED THAT THERE WAS A DELAY FOR SEVERAL MINUTES ON THE PROCEDURE TO OPEN 4 MIDAS REX TRAYS IN ORDER TO FIND A MOTOR THAT FUNCTIONED PROPERLY. IT WAS ALSO CONFIRMED THAT ACTIONS WERE UNDERTAKEN TO REPAIR THE TORN DURA DURING THE SAME SURGERY. ADDITIONAL INFORMATION ABOUT THE CURRENT STATUS OF THE PATIENT IS BEING FOLLOWED UP ON.
Patient problems: Perforation
MAUDE report 8387381 ↗
Other MOTOR, DRILL, ELECTRIC devices
- NA CRANI-A-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC1 recall358 reports
- NA QD8Synthes GmbH · MOTOR, DRILL, ELECTRIC386 reports
- NA QD11-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC312 reports
- Midas Rex EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC254 reports
- MIDAS REX MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC186 reports
- Midas Rex MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC158 reports
- NA SHORT-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC134 reports
- Midas Rex® EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC123 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.