Cobalt™ XT HF Quad CRT-D MRI SureScan™ DTPA2QQRecalls & reported problems

Medtronic, Inc. · CRTD DTPA2QQ COBALT XT HF QUAD MRI DF4 · Class 3 · Implant

All models of Cobalt XT HF Quad CRT-D MRI SureScan (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is the Medtronic Cobalt XT HF Quad CRT-D, an implanted defibrillator with heart resynchronization pacing. The 1,618 reports often describe the device misreading heart signals, pacing problems, and inappropriate or inadequate shocks, and many mention infections. The records include 38 deaths, 1,098 injuries and 482 malfunctions. There are 4 open recalls, covering reduced shock energy, a telemetry error and a manufacturing weld crack.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,639
Deaths reported
40
Injuries reported
1,112
Recalls naming it
4

Summary

The FDA's MAUDE database contains 1,639 adverse event reports that list this exact device (DI 00763000178093), received between Mar 1, 2021 and Sep 30, 2026 — classified as 40 deaths, 1,112 injuries, 487 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (266 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 27 reportsDec 2023: 25 reportsJan 2024: 23 reportsFeb 2024: 27 reportsMar 2024: 28 reportsApr 2024: 42 reportsMay 2024: 27 reportsJun 2024: 29 reportsJul 2024: 31 reportsAug 2024: 37 reportsSep 2024: 34 reportsOct 2024: 54 reportsNov 2024: 44 reportsDec 2024: 31 reportsJan 2025: 52 reportsFeb 2025: 28 reportsMar 2025: 35 reportsApr 2025: 18 reportsMay 2025: 37 reportsJun 2025: 37 reportsJul 2025: 60 reportsAug 2025: 54 reportsSep 2025: 28 reportsOct 2025: 35 reportsNov 2025: 62 reportsDec 2025: 42 reportsJan 2026: 48 reportsFeb 2026: 60 reportsMar 2026: 38 reportsApr 2026: 44 reportsMay 2026: 38 reportsJun 2026: 25 reportsJul 2026: 13 reportsAug 2026: 35 reportsSep 2026: 41 reportsOct 2026: 0 reports20242025202662
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 16, 2024Class IIOpen, Classified

    Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

    Z-2194-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. May 10, 2023Class IOpen, Classified

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

    Z-1779-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Jun 22, 2022Class IOpen, Classified

    There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

    Z-1486-2022 · Root cause: Device Design · Recall details · FDA record ↗

  4. Apr 12, 2022Class IIOpen, Classified

    Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

    Z-1267-2022 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem660
  • Device Sensing Problem266
  • Pacing Problem232
  • Inappropriate/Inadequate Shock/Stimulation182
  • Appropriate Device Problem Term/Code Not Available133
  • Mechanical Problem104
  • Premature Discharge of Battery86
  • Electromagnetic Interference68

Patient problems

  • No Clinical Signs, Symptoms or Conditions561
  • Unspecified Infection446
  • Shock from Patient Lead(s)184
  • Erythema113
  • Purulent Discharge89
  • Discomfort75
  • Erosion70
  • Pain68

Recent adverse event reports

The 25 most recent of 1,639 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952738 ↗
  • InjurySep 30, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50951371 ↗
  • MalfunctionSep 29, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941286 ↗
  • MalfunctionSep 29, 2026Migration or Expulsion of Device

    Patient problems: Palpitations

    MAUDE report 50938823 ↗
  • MalfunctionSep 28, 2026Electromagnetic Interference

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50923904 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50923676 ↗
  • InjurySep 28, 2026Failure to Deliver Shock/Stimulation; Device Sensing Problem

    Patient problems: Tachycardia

    MAUDE report 50922786 ↗
  • InjurySep 28, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50922491 ↗
  • MalfunctionSep 25, 2026High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903728 ↗
  • InjurySep 24, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889025 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that three months post Cardiac Resynchronization Therapy Defibrillator (CRT-D) implant, the patient experienced a device pocket infection, discomfort, pain, purulent discharge, redness, and swelling. The CRT-D system was explanted and then replaced six days later. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Erythema; Medical device site infection; Pain; Purulent Discharge; Swelling/ Edema

    MAUDE report 50888990 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that approximately one month post replacement procedure, the patient developed a pocket infection. It was noted that the patient developed purulent discharge/pus. The Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted and replaced with a Leadless Implantable Pulse Generator (IPG). No further patient complications have been reported as a result of this event.

    Patient problems: Purulent Discharge; Unspecified Infection

    MAUDE report 50877036 ↗
  • InjurySep 23, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) may have provided inappropriate Anti-Tachycardia Pacing (ATP) and high voltage therapy for an Supra Ventricular Tachycardia (SVT) episode detected as Ventricular Fibrillation (VF). The CRT-D remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50871323 ↗
  • InjurySep 23, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted due to infection and septic shock. A leadless Implantable Pulse Generator (IPG) was implanted. No further patient complications have been reported as a result of this event.

    Patient problems: Septic Shock; Unspecified Infection

    MAUDE report 50870640 ↗
  • InjurySep 23, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Bacteremia; Bacterial Infection; Erosion; Hematoma; Muscle Weakness/Atrophy

    MAUDE report 50870057 ↗
  • MalfunctionSep 18, 2026Mechanical Problem; Low impedance; Unstable Impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829288 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50828186 ↗
  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50810932 ↗
  • InjurySep 17, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809256 ↗
  • MalfunctionSep 16, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: Confusion/ Disorientation; Low Blood Pressure/ Hypotension

    MAUDE report 50794119 ↗
  • MalfunctionSep 16, 2026Failure to Deliver Shock/Stimulation; Device Sensing Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) inappropriately withheld detection and therapy due to Supra Ventricular Tachycardia (SVT) that was identified as sustained Ventricular Tachycardia (VT).  The CRT-D remains in use. No patient complications have been reported as a result of this event.

    Patient problems: Tachycardia

    MAUDE report 50793021 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) device system was explanted due to a pocket infection with purulent discharge. No further patient complications have been reported as a result of this event.

    Patient problems: Medical device site infection; Purulent Discharge

    MAUDE report 50787145 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Bacterial Infection

    MAUDE report 50780920 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Erythema; Purulent Discharge; Unspecified Infection

    MAUDE report 50762489 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted due to pocket infection with erosion, redness and discomfort. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Erosion; Erythema; Medical device site infection

    MAUDE report 50757530 ↗

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →