Unify Assura™ CD3357-40QRecalls & reported problems

St. Jude Medical, Inc. · Cardiac resynchronization device, tiered-therapy cardioverter/defibrillator VVED DDDRV · Class 3 · Implant

All models of Unify Assura (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Unify Assura is an implanted defibrillator that paces the heart and delivers shocks. Of its 1,781 reports, many describe the device reading electrical noise as heartbeats, unexpected resets, and inappropriate or inadequate shocks. The reports include 55 deaths and 506 injuries, and some mention infection and pocket erosion. Reports fell from 140 in the prior 12 months to 79 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,784
Deaths reported
54
Injuries reported
506
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,784 adverse event reports that list this exact device (DI 05414734508216), received between Jun 11, 2018 and Sep 18, 2026 — classified as 54 deaths, 506 injuries, 1,224 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (956 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 10 reportsJan 2024: 18 reportsFeb 2024: 20 reportsMar 2024: 19 reportsApr 2024: 17 reportsMay 2024: 16 reportsJun 2024: 11 reportsJul 2024: 16 reportsAug 2024: 16 reportsSep 2024: 10 reportsOct 2024: 13 reportsNov 2024: 13 reportsDec 2024: 15 reportsJan 2025: 12 reportsFeb 2025: 18 reportsMar 2025: 15 reportsApr 2025: 11 reportsMay 2025: 9 reportsJun 2025: 14 reportsJul 2025: 11 reportsAug 2025: 13 reportsSep 2025: 2 reportsOct 2025: 7 reportsNov 2025: 11 reportsDec 2025: 10 reportsJan 2026: 6 reportsFeb 2026: 10 reportsMar 2026: 8 reportsApr 2026: 3 reportsMay 2026: 11 reportsJun 2026: 7 reportsJul 2026: 1 reportsAug 2026: 9 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing956
  • Adverse Event Without Identified Device or Use Problem284
  • Inappropriate or Unexpected Reset110
  • Inappropriate/Inadequate Shock/Stimulation85
  • Under-Sensing79
  • Battery Problem59
  • Signal Artifact/Noise46
  • Pacing Problem44

Patient problems

  • No Clinical Signs, Symptoms or Conditions784
  • No Known Impact Or Consequence To Patient515
  • Unspecified Infection182
  • No Consequences Or Impact To Patient68
  • Pocket Erosion58
  • Discomfort48
  • Arrhythmia42
  • Insufficient Information42

Recent adverse event reports

The 25 most recent of 1,784 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Over-Sensing; Pacing Problem; Environmental Compatibility Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50831355 ↗
  • InjurySep 18, 2026Inappropriate or Unexpected Reset

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829662 ↗
  • MalfunctionSep 17, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812667 ↗
  • MalfunctionSep 10, 2026Over-Sensing

    It was reported that the patient presented remotely. A remote transmission showed that the patient's Implantable Cardioverter Defibrillator (ICD) exhibited Post-Paced T-wave Oversensing. No intervention was performed. The patient had no adverse consequences due to the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50717720 ↗
  • MalfunctionSep 2, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50555483 ↗
  • MalfunctionSep 1, 2026Under-Sensing

    It was reported that the patient presented remotely via merlin.net. Review of transmission revealed that the Implantable Cardioverter Defibrillator (ICD) exhibited under-sensing. No intervention was performed. There were no patient consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50549255 ↗
  • MalfunctionAug 26, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50464564 ↗
  • MalfunctionAug 25, 2026Failure to Charge

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50442442 ↗
  • MalfunctionAug 21, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50382920 ↗
  • MalfunctionAug 20, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50359498 ↗
  • InjuryAug 20, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient presented in the emergency room due to pocket erosion. The physician explanted the entire implantable cardioverter-defibrillator system to resolve the issue. The patient's status was unknown.

    Patient problems: Pocket Erosion

    MAUDE report 50336113 ↗
  • MalfunctionAug 14, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50260278 ↗
  • MalfunctionAug 12, 2026Over-Sensing

    Patient problems: Insufficient Information

    MAUDE report 50227587 ↗
  • MalfunctionAug 6, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50161559 ↗
  • InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient presented with a bacteremia infection. The implantable cardioverter defibrillator (ICD), right ventricular lead, and left ventricular lead were explanted. The patient was reported to be recovering following the procedure.

    Patient problems: Unspecified Infection

    MAUDE report 50137445 ↗
  • MalfunctionJul 9, 2026Pacing Asynchronously; Sensing Intermittently

    IT WAS REPORTED DURING REMOTE FOLLOW UP THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED INTERMITTENT SENSING DUE TO PSEUDO-FUSION. THE PRODUCT WILL CONTINUE TO BE MONITORED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25772021 ↗
  • InjuryJun 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH AN INFECTION AT THE IMPLANTED DEVICE POCKET SITE. THE PHYSICIAN EXPLANTED THE ENTIRE SYSTEM DUE TO INFECTION. THE PATIENT CONDITION WAS UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 25555999 ↗
  • MalfunctionJun 18, 2026Difficult to Remove

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A STANDARD GENERATOR CHANGEOUT PROCEDURE. DURING THE PROCEDURE, THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)'S SET SCREW WAS DIFFICULT TO REMOVE. IT WAS EVENTUALLY LOOSENED, AND THE ICD WAS SUCCESSFULLY REPLACED. THE PATIENT REMAINED STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25555370 ↗
  • MalfunctionJun 18, 2026Over-Sensing

    IT WAS REPORTED DURING IN CLINIC FOLLOW UP THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED POST-PACED T-WAVE OVERSENSING (PPTWOS). PROGRAMMING CHANGES WERE SUCCESSFULLY IMPLEMENTED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25549542 ↗
  • MalfunctionJun 17, 2026Failure to Capture

    IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED A LOSS OF CAPTURE. NO INTERVENTION WAS PERFORMED, AND THE PATIENT'S STATUS WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25539958 ↗
  • MalfunctionJun 12, 2026Under-Sensing

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED ATRIAL UNDER-SENSING. PROGRAMMING CHANGES WERE PERFORMED TO RESOLVE THE ISSUE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25488090 ↗
  • DeathJun 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPIRED DUE TO SYSTOLIC HEART FAILURE, VENTRICULAR TACHYCARDIA, AND ACUTE HYPOXIC RESPIRATORY FAILURE. THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), RIGHT ATRIAL (RA) LEAD, AND RIGHT VENTRICULAR (RV) LEAD WERE EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446600 ↗
  • MalfunctionJun 5, 2026Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC DUE TO SYMPTOMS OF DIZZINESS. A DEVICE INTERROGATION SHOWED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED POST-PACED T-WAVE OVERSENSING. PROGRAMMING CHANGES WERE MADE. THE PATIENT WAS STABLE THROUGHOUT.

    Patient problems: Dizziness

    MAUDE report 25411829 ↗
  • InjuryMay 29, 2026Failure to Deliver Shock/Stimulation

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR FAILED TO DELIVER SHOCK. FURTHER INFORMATION WAS REQUESTED; HOWEVER, WAS NOT PROVIDED.

    Patient problems: Arrhythmia

    MAUDE report 25340380 ↗
  • MalfunctionMay 27, 2026Over-Sensing

    IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED REMOTELY VIA MERLIN.NET. IT WAS NOTED THAT POST PACED T WAVE OVERSENSING WAS OBSERVED ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25307908 ↗

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →