Affinity NT™ 95217Recalls & reported problems
Medtronic, Inc. · FPA 95217 AFFINITY TRILLIUM GLBL · Class 2
All models of Affinity NT (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000232122), received between Mar 8, 2022 and Jul 27, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
- Crack1
- Fluid/Blood Leak1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2022Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR, IT WAS REPORTED THE DURING PREPARATION OF THE EXTRACORPOREAL CIRCULATION FOR NEW CASES THE CUSTOMER DETECTED A LEAK AT THE OXYGENATOR'S TEMPERATURE SENSOR. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE PACKAGING OR OTHER DEVICES WITHIN THE SAME PACKAGING AS THE OXYGENATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15114690 ↗MalfunctionMay 16, 2022Break; Break
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR AND CARDIOTOMY VENOUS RESERVOIR, THE CUSTOMER REPORTED THAT THE SAMPLING PORT OF THE DEVICE WAS BROKEN. THE CUSTOMER STATED THAT THE PORT WAS PARTIALLY BROKEN. THE DEVICE WAS REPLACED FOR THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECTS OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION STATING THAT THE DEVICE WAS USED 7 DAYS FROM PRODUCT RECEIPT AND THE STORAGE ROOM TEMPERATURE WAS 20~25 °C AND THE HUMIDITY WAS 20~30%.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14407538 ↗MalfunctionMar 8, 2022Obstruction of Flow; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THIS AFFINITY NT CARDIOTOMY VENOUS RESERVOIR (CVR), THE CUSTOMER NOTICED THAT THE BLOOD COMING INTO THE CARDIOTOMY SECTION AT THE REAR OF THE CVR SEEMED TO NOT BE COMING THROUGH THE SCREEN FILTER AS IT NORMALLY DOES. THE CUSTOMER SAID THAT THEIR HEPARIN DOSAGE WAS NORMAL AND THE ACTS WERE GREATER THAN 480 SECONDS THROUGHOUT THE CARDIOPULMONARY BYPASS PROCEDURE. USE OF THE CVR WAS CONTINUED AND THERE WAS NO PATIENT IMPACT AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13693572 ↗
FDA clearance and classification
Cleared through Special 510(k) K191029 on May 17, 2019 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Affinity NTTM Oxygenator (Model 511), Affinity® NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (PRF) with TrilliumTM Biosurface (Model 511T), Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541), Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541-R), Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface (Model 541T), Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurfac”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code DTZ (oxygenator, cardiopulmonary bypass) under 21 CFR 870.4350 as a Class II device. 21 CFR 870.4350 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Affinity NT™ 95217
Has the Affinity NT™ 95217 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000232122), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Affinity NT™ 95217?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 8, 2022 and Jul 27, 2022. In those reports, the most common specific device problems are “Break” (1), “Crack” (1) and “Fluid/Blood Leak” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Affinity NT™ 95217?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Affinity NT™ 95217 FDA approved?
It was cleared through Special 510(k) K191029 on May 17, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Affinity NT™ 95217, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Affinity NT™, model 95217). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other OXYGENATOR, CARDIOPULMONARY BYPASS devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths87 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page