Skip to content

N/A EMBSNV20Recalls & reported problems

Medtronic, Inc. · ADAPTOR EMBSNV20 20CM FOR BSX NVRO LEADS · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000278267), received between Apr 3, 2023 and Aug 20, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High impedance” (2 reports).

9 other recalls cover devices cleared under the same FDA submission (P840001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2024High impedance; High impedance

    THE MANUFACTURER REPRESENTATIVE (REP) REPORTED THAT ON JUNE 29, 2022 DURING AN ANIMAL STUDY, HIGH IMPEDANCES / OPEN (>40K OHMS) WAS SEEN ON CONTACT 2.  IT WAS NOTED THAT THE IMPLANTED NEUROSTIMULATOR (INS) USED IS A PRE-MARKET DEVICE, BUT THE ADAPTOR-EXTENSION AND THE COMPETITOR LEAD USED WERE POST-MARKET DEVICES.    TROUBLESHOOTING PERFORMED INCLUDED RESEATING THE ADAPTOR IN THE IMPLANTED NEUROSTIMULATOR (INS) WITHIN THE OR, BUT CONTACT 2 CONTINUED TO SHOW UP AS OPEN. AT EXPLANT, CONTACT 2 SHOWED AN OPEN WITH BOTH THE INS AND WITH THE HANDHELD OHMMETER. IT WAS NOTED THAT THE BOOT WAS SUTURED / GLUED WITH MEDICAL ADHESIVE TO THE ADAPTOR AND LEAD TO KEEP FLUID FROM GETTING IN AND AFFECTING THE ECAP MEASUREMENTS, SO THE INVESTIGATOR DID NOT DEFINITIVELY KNOW IF THE ADAPTOR WAS CAUSING THE HIGH IMPEDANCES. FDA CONFIRMED RECEIPT OF NON-MDT PRODUCT NOTIFICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20001870 ↗
  • MalfunctionApr 3, 2023High impedance; High impedance

    NO NEW INFORMATION SUBMITTING A CORRECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16667746 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P840001. One submission often covers many models, so check the product and lot details in each record.

  • Z-0372-2025 · Sep 18, 2024 · Open, Classified
    Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
  • Z-1691-2024 · Mar 28, 2024 · Open, Classified
    Intellis AdaptiveStim Product Number 97715, Spinal Cord Stimulator
  • Z-0780-2024 · Dec 19, 2023 · Open, Classified
    Model A710 Intellis Clinician Programmer Application, version 2.0.97
  • Z-2159-2021 · Jul 9, 2021 · Open, Classified
    Medtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b) 977A275 c) 977A290
  • Z-1926-2016 · May 3, 2016 · Terminated
    Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or li
  • Z-1147-2015 · Jan 9, 2015 · Terminated
    The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
  • Z-1524-2014 · Feb 27, 2014 · Terminated
    Medtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713), RestoreSENSOR (37714), RestoreSENSOR SureScan MRI (97714), PrimeAdvanced
  • Z-0278-2014 · Sep 30, 2013 · Terminated
    Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
  • Z-1536-06 · Aug 2, 2006 · Terminated
    Medtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis

Other Stimulator, spinal-cord, totally implanted for pain relief devices

All Stimulator, spinal-cord, totally implanted for pain relief devices →

Sources and updates