Skip to content

NA BB8B96R14Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK BB8B96R14 WO PLEG · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000300692), received between Apr 30, 2024 and Apr 30, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K171308); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1027-2023 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Crack1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THERE WAS A LEAK FROM THE NON-COLLARED SIDE OF THE VENT VALVE CONNECTOR. IT WAS STATED THAT THERE WAS MINIMAL PATIENT BLOOD LOSS AS A RESULT OF THE LEAK AND AN UNPLANNED BLOOD TRANSFUSION WAS NOT REQUIRED BECAUSE OF THE LEAK. THE LEAK ORIGINATED FROM THE VALVE. THE CUSTOMER STATED THAT A COMPLETE CRACK/BREAK DID NOT OCCUR. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 19212523 ↗
  • MalfunctionApr 30, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THERE WAS A LEAK FROM THE NON-COLLARED SIDE OF THE VENT VALVE CONNECTOR. THERE WAS NO PATIENT BLOOD LOSS AS A RESULT. AN UNPLANNED BLOOD TRANSFUSION WAS NOT REQUIRED BECAUSE OF THE LEAK. THE LEAK ORGINATED FROM THE RED VENT VALVE COMPONENT. THE SAME LEAK APPEARED IN MULTIPLE PACKS. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19211656 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171308. One submission often covers many models, so check the product and lot details in each record.

  • Z-0535-2018 · Jan 3, 2018 · Terminated
    Medtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.