KYPHON® Express™ Inflatable Bone Tamp K15B (UDI 00763000312008)Recalls & reported problems
Medtronic Sofamor Danek, Inc. · TAMP K15B IBT KYPHX EXPRESS SIZE 15/2 · Class 2
All models of KYPHON Express Inflatable Bone Tamp (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000312008), received between Mar 25, 2020 and Mar 6, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
5 other recalls cover devices cleared under the same FDA submission (K041454); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Mechanical Jam1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2024Material Rupture; Mechanical Jam; Material Rupture; Mechanical Jam
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING KYPHOPLASTY FOR VERTEBRAL COMPRESSION FRACTURE. IT WAS REPORTED THAT THE BALLOON POTENTIALLY MALFUNCTIONED DURING A PROCEDURE. THE BALLOON RADIOPAQUE MARKERS/ BALLOON BECAME STUCK IN THE PATIENT AND BROKE OFF IN THE VERTEBRAL BODY. THERE WERE NO PATIENT SYMPTOMS REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18843661 ↗MalfunctionJan 3, 2022Material Rupture; Material Rupture
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING INTRA-OP OF THE REPORTED PRODUCT. THE PRE-OP DIAGNOSIS WAS MENTIONED AS COMPRESSION FRACTURE. IT WAS REPORTED THAT, BALLOON RUPTURED DURING INFLATION. WHILE ATTEMPTING TO WITHDRAW THE DEVICE, PART OF THE POLYMER BALLOON SEPARATED FROM THE BODY OF THE BALLOON AND WAS NOT RETRIEVABLE. THERE WAS NO DELAY IN PROCEDURE REPORTED. THERE WERE NO SYMPTOMS TO PATIENT OR PHYSICIAN AND NO HOSPITALIZATION WAS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13140499 ↗MalfunctionMar 25, 2020Material Rupture; Material Rupture
PRESSURE PRIOR TO RUPTURE- 300 VOLUME PRIOR TO RUPTURE- 3 ML IT WAS REPORTED THAT THE PATIENT UNDERWENT KYPHOPLASTY UNILATERAL AT T8 DUE TO VERTEBRAL COMPRESSION FRACTURE. INTRA-OP, T8 OSTEOCOOL PROCEDURE WAS COMPLETED. THE PROBES WAS REMOVED AND WAITED FOR 30 SECONDS, THEN THE INFLATABLE BONE TAMPS (IBT) WAS INSERTED AND INFLATED. IT BURST AT LOW PRESSURE AND LOW VOLUME. THERE WAS A CONTRAST OBSTRUCTED VIEW OF CEMENT IMPLANTATION. THE OPERATION WAS COMPLETED USING NEW PRODUCT. THERE WAS A DELAY IN OVERALL PROCEDURE TIME AS A RESULT OF THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9876506 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K041454. One submission often covers many models, so check the product and lot details in each record.
- Z-2390-2020 · Apr 24, 2020 · TerminatedMedtronic Bone Void Filler, Small - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c
- Z-2391-2020 · Apr 24, 2020 · TerminatedMedtronic Bone Void Filler, Large - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c
- Z-2389-2020 · Apr 24, 2020 · TerminatedExtremities, Large Kit Assy, US, 15/3 - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Ki
- Z-3175-2011 · Mar 23, 2011 · TerminatedKyphon brand KyphoPak Express Tray First Fracture, 15/2 with KyphX Osteo Introducer System; Catalog Number: KPE1003; Product is manufactured and distributed by Medtronic Spine LLC
- Z-1209-2009 · Feb 4, 2009 · TerminatedKyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland. For use in spinal surgery.
Other Arthroscope devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page