KYPHON Express II Inflatable Bone Tamp KE202 (UDI 00763000312053)Recalls & reported problems
Medtronic Sofamor Danek, Inc. · BONE TAMP KE202 20/2 EXPRESS II IBT · Class 2
All models of KYPHON Express II Inflatable Bone Tamp (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000312053), received between Oct 21, 2021 and May 2, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2025Material Rupture; Material Rupture
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING T12 KYPHOPLASTY FOR AN INDICATION OF COMPRESSION FRACTURE. IT WAS REPORTED THAT THE BALLOON WAS RUPTURED DURING INFLATION. THE BALLOON INFLATION WAS NOT VISIBLE ON X-RAY, CONTRAST IN INTELLEF LATOR WAS CHECKED. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21948978 ↗MalfunctionSep 21, 2022Material Rupture; Material Rupture
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING A SPINAL PRODUCT THAT WAS USED IN KYPHOPLASTY SURGERY FOR VERTEBRAL COMPRESSION FRACTURE OF L2. IT WAS REPORTED THAT THE BALLOON GOT STUCK IN THE VERTEBRAL BODY AND WHEN TRYING TO REMOVE THE BALLOON THROUGH THE INTRODUCER, THE END OF THE BALLOON BECAME DETACHED FROM THE TETHER AND WAS LEFT IN THE PATIENT. NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE BALLOON RUPTURED ON INFLATION. THERE CURRENTLY IS NO REVISION SURGERY SCHEDULED OR PLANNED FOR REMOVAL OF THE BALLOON FRAGMENTS.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 15460839 ↗MalfunctionOct 21, 2021Material Rupture; Material Rupture
INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH SYMPTOMS OF TWO COMPRESSION FRACTURES L1 AND L4 INVOLVED IN THE KYPHOPLASTY PROCEDURE. IT WAS REPORTED THAT INTRA-OPERATIVELY, WHILE INFLATING BALLOON, PRESSURE WAS 300 PSI AND THEN THE BALLOON BURST/RUPTURED. PATIENT WAS NOT ALLERGIC TO THE CONTRAST MEDIA. PRESSURE PRIOR TO RUPTURE WAS 305 PSI. VOLUME PRIOR TO RUPTURE WAS 2 CC. BALLOON WAS NOT IMPLANTED IN THE PATIENT'S BODY, IT WAS PLACED IN BODY THEN REMOVED. THERE WAS DELAY IN OVERALL PROCEDURE TIME FOR 4 MINUTES. THERE WERE NO RUPTURED FRAGMENTS FOUND IN THE PATIENT'S BODY, ONLY CONTRAST MEDIUM WAS FOUND. PROCEDURE WAS COMPLETED WITH NEW PRODUCT. PATIENT IS ALIVE AND NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12670474 ↗
Other Arthroscope devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page