MyCareLink™ 24950M (UDI 00763000367800)Recalls & reported problems
Medtronic, Inc. · MON 24950M MYCARELINK/SVC USA · Class 3
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This is a Medtronic MyCareLink remote monitor, the home unit that relays information from an implanted heart device. The 65 reports mostly describe the monitor overheating; a few mention a burning smell, brown residue oozing from the unit, moisture getting inside, or delivered stimulation not matching what was set. Reports also note four burns and one allergic reaction, and most describe no lasting patient effect. No deaths and no injuries are reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 65
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 65 adverse event reports that list this exact device (DI 00763000367800), received between Aug 25, 2020 and Jul 24, 2025 — classified as 65 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (56 reports).
10 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device56
- Device Emits Odor4
- Inappropriate/Inadequate Shock/Stimulation3
- Moisture Damage2
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions59
- No Consequences Or Impact To Patient5
- Burn(s)4
- Hypersensitivity/Allergic reaction1
Recent adverse event reports
The 25 most recent of 65 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2025Overheating of Device; Overheating of Device
THE REMOTE MONITOR WAS REPLACED AND RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22587373 ↗MalfunctionDec 5, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS OVERHEATING. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20853943 ↗MalfunctionNov 22, 2023Moisture Damage; Moisture Damage
IT WAS REPORTED THAT THERE WAS A "BROWN STUFF" OOZING OUT OF THE MONITOR. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE RE PLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18190353 ↗MalfunctionJun 14, 2023Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17126129 ↗MalfunctionApr 12, 2023Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16722755 ↗MalfunctionMar 9, 2023Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16513094 ↗MalfunctionMar 6, 2023Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16484780 ↗MalfunctionFeb 23, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16428127 ↗MalfunctionDec 28, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS HOT LIKE BURNING. NO TROUBLESHOOTING TAKEN AT THE TIME OF CALL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16060636 ↗MalfunctionDec 22, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE MONITOR GOT REALLY HOT AND THE READER WAS SO HOT THAT THE PATIENT COULD NOT HOLD ON TO IT. THE PATIENT STATED THAT EVERYTHING WAS CHECKED OUT AND THE MONITOR WAS STILL ACTING UP. THE MONITOR WAS TOTALLY UNPLUGGED FOR SAFETY'S SAKE. NO FURTHER TROUBLESHOOTING PERFORMED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16037873 ↗MalfunctionNov 21, 2022Moisture Damage; Moisture Damage
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15834721 ↗MalfunctionNov 18, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR BECAME HOT TO TOUCH THAT IT WAS REMOVED FROM THE PATIENT'S HOUSE FOR FEAR OF STARTING A FIRE. NO TROUBLESHOOTING WAS TAKEN AT THE TIME OF CALL. IT WAS STATED THAT THE MONITOR HAD BEEN DISCARDED. THE REMOTE MONITOR REPLACEMENT STATUS IS UNKNOWN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15824265 ↗MalfunctionOct 11, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. THE PATIENT UNPLUGGED THE MONITOR FROM THE POWER SUPPLY. NO FURTHER TR OUBLESHOOTING PERFORMED AT THE TIME OF CALL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15579330 ↗MalfunctionSep 19, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15445111 ↗MalfunctionSep 8, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. THE REMOTE MONITOR WAS UNPLUGGED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15378521 ↗MalfunctionAug 16, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE READER WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15237291 ↗MalfunctionAug 16, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING WAS TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15237153 ↗MalfunctionJul 7, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR STARTED TO BURN AFTER PLUGGING IT. IT WAS FURTHER REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING WAS TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14950003 ↗MalfunctionMay 23, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT. THE PATIENT WAS INFORMED TO UNPLUG THE MONITOR. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14472421 ↗MalfunctionMay 13, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS OVERHEATING. THE PATIENT WAS ADVISED TO UNPLUG THE MONITOR AND NO FURTHER TROUBLESHOOTING WAS INDICATED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14399406 ↗MalfunctionMay 7, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE THE READER WAS GETTING TOO HOT TO TOUCH. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14315521 ↗MalfunctionApr 24, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE PATIENT SAW SPARKS COMING FROM THE REMOTE MONITOR SCREEN. THE MONITOR WAS ALREADY UNPLUGGED. NO FURTHER TROUBLESHOOTING TAKEN. THE REMOTE MONITOR IS EXPECTED TO BE RETURNED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14191222 ↗MalfunctionApr 18, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR READER WAS HOT TO THE CHEST AND BURNED THE PATIENT. NO TROUBLESHOOTING INDICATED. THE MONITOR REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Burn(s)
MAUDE report 14137723 ↗MalfunctionApr 14, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING INDICATED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14114196 ↗MalfunctionApr 14, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT. NO TROUBLESHOOTING INDICATED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14114166 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.
- Z-2168-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
- Z-2169-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2505-2020 · Jun 4, 2020 · Open, ClassifiedCareLink SmartSync Device Manager, Model Number 24970A
- Z-2506-2020 · Jun 4, 2020 · Open, ClassifiedPatient Connector, Model Number 24967
- Z-0845-2019 · Jan 17, 2019 · TerminatedMedtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
- Z-0399-2018 · Jul 14, 2017 · TerminatedMyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with
- Z-0400-2018 · Jul 14, 2017 · TerminatedMyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
- Z-2209-2010 · Jun 11, 2010 · TerminatedMedtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page