Skip to content

NA BB8G55R15Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK BB8G55R15 A PACK · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000373955), received between Nov 1, 2022 and Dec 16, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1027-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Break1
  • Delivered as Unsterile Product1
  • Material Separation1
  • Misassembled During Installation1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE VENOUS LINE WAS CONTAINED WITHIN STERILE WRAPPING MAKING IT UNABLE FOR THE CUSTOMER TO OPEN THE WRAP WITHOUT CONTAMINATION. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23817428 ↗
  • MalfunctionNov 25, 2024Material Separation; Misassembled During Installation; Material Separation; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM PACK, IT WAS REPORTED THAT THE CARDIOPLEGIA LINE WAS NOT CONNECTED. THE CUSTOMER STATED THAT THE LOOSE/SEPARATED COMPONENT DID NOT FALL INTO THE PATIENT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20770300 ↗
  • MalfunctionNov 12, 2024Obstruction of Flow; Obstruction of Flow

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THERE WAS AN ISSUE WITH THE LUER ON TOP OF THE ARTERIAL LINE FILTER. THE CUSTOMER REPORTED THAT THERE APPEARED TO BE PLASTIC IN IT WHEN THEY TRIED TO FLUSH THE DEVICE WITH CO2. THE CUSTOMER NOTICED A WHISTLING NOISE AND THE CO2 WAS UNABLE TO FLUSH THROUGH THE PORT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE OBSERVED ON THE BOX AND STERILE BARRIER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20665617 ↗
  • MalfunctionFeb 16, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THERE WAS A LEAKAGE OBSERVED AT THE Y CONNECTOR. SEE ATTACHED VIDEO. THE DEVICE WAS USED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS. THERE WAS NO VISIBLE AIR IN THE SYSTEM/TUBING. PATIENT BLOOD LOSS WAS MINIMAL. A TRANSFUSION WAS NOT REQUIRED AS A RESULT OF THIS LEAK.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18722022 ↗
  • MalfunctionJun 14, 2023Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, DURING INITIAL SET UP THE CUSTOMER NOTICED DURING PRIMING THAT THE STOPCOCK WAS BROKEN INSIDE OF THE ARTERIAL LINE WHEN IT WAS SCREWED INTO PLACE.THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECT OCCURRED. ADDITIONAL INFORMATION: THERE WAS NO DAMAGE REPORTED TO THE OUTER BOX. THERE WAS NO OTHER DAMAGE REPORTED TO ANY OTHER COMPONENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17125405 ↗
  • MalfunctionNov 1, 2022Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED A LEAK WAS NOTICED FROM THE LINE CONNECTION. THE CUSTOMER STATED THAT THE BLOOD LOSS WAS MINIMAL, AND A BLOOD TRANSFUSION WAS NOT REQUIRED. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADDITIONAL PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO VISIBLE AIR IN SYSTEM/TUBING REPORTED. THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 15708007 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.

  • Z-1872-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8M63R4, Custom Pack HY8M63R4 Adult ECC, Custom Perfusion System, Sterile, Rx.
  • Z-1890-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10L28R3, Custom Pack HY10L28R3 1/4 Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1880-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1873-2021 · Apr 29, 2021 · Terminated
    Medtronic BB10R94R6, Custom Pack BB10R94R6 1/4 ECCPK, Custom Perfusion System, Sterile, Rx.
  • Z-1874-2021 · Apr 29, 2021 · Terminated
    Medtronic BB11J14R2, Custom Pack BB11J14R2 Bridge, Custom Perfusion System, Sterile, Rx.
  • Z-1879-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10W63R1, Custom Pack CB10W63R1 HL 20 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1895-2021 · Apr 29, 2021 · Terminated
    Medtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.
  • Z-1876-2021 · Apr 29, 2021 · Terminated
    Medtronic CB7J97R8, Custom Pack CB7J97R8 U14 E Sup, Custom Perfusion System, Sterile, Rx.
  • Z-1878-2021 · Apr 29, 2021 · Terminated
    Medtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
  • Z-1885-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10Y54R4, Custom Pack HY10Y54R4 ECC PED 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1877-2021 · Apr 29, 2021 · Terminated
    Medtronic HY11B40R1, Custom Pack HY11B40R1 1/4 Respirato, Custom Perfusion System, Sterile, Rx.
  • Z-1889-2021 · Apr 29, 2021 · Terminated
    Medtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1871-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10Z48R2, Ctr., Custom Pack HY10Z48R2 Adult ECC 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1894-2021 · Apr 29, 2021 · Terminated
    Medtronic BB9G78R8, Custom Pack BB9G78R8 ECC Supp, Custom Perfusion System, Sterile, Rx.
  • Z-1888-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10R55R5, Custom Pack CB10R55R5 3/8X3/8 Loop, Custom Perfusion System, Sterile, Rx.

…and 10 more.

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →

Sources and updates