NA BB8G55R15Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB8G55R15 A PACK · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000373955), received between Nov 1, 2022 and Dec 16, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1027-2023 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak2
- Break1
- Delivered as Unsterile Product1
- Material Separation1
- Misassembled During Installation1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Hemorrhage/Blood Loss/Bleeding2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 16, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE VENOUS LINE WAS CONTAINED WITHIN STERILE WRAPPING MAKING IT UNABLE FOR THE CUSTOMER TO OPEN THE WRAP WITHOUT CONTAMINATION. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23817428 ↗MalfunctionNov 25, 2024Material Separation; Misassembled During Installation; Material Separation; Misassembled During Installation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM PACK, IT WAS REPORTED THAT THE CARDIOPLEGIA LINE WAS NOT CONNECTED. THE CUSTOMER STATED THAT THE LOOSE/SEPARATED COMPONENT DID NOT FALL INTO THE PATIENT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20770300 ↗MalfunctionNov 12, 2024Obstruction of Flow; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THERE WAS AN ISSUE WITH THE LUER ON TOP OF THE ARTERIAL LINE FILTER. THE CUSTOMER REPORTED THAT THERE APPEARED TO BE PLASTIC IN IT WHEN THEY TRIED TO FLUSH THE DEVICE WITH CO2. THE CUSTOMER NOTICED A WHISTLING NOISE AND THE CO2 WAS UNABLE TO FLUSH THROUGH THE PORT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE OBSERVED ON THE BOX AND STERILE BARRIER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20665617 ↗MalfunctionFeb 16, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THERE WAS A LEAKAGE OBSERVED AT THE Y CONNECTOR. SEE ATTACHED VIDEO. THE DEVICE WAS USED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS. THERE WAS NO VISIBLE AIR IN THE SYSTEM/TUBING. PATIENT BLOOD LOSS WAS MINIMAL. A TRANSFUSION WAS NOT REQUIRED AS A RESULT OF THIS LEAK.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 18722022 ↗MalfunctionJun 14, 2023Break; Break
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, DURING INITIAL SET UP THE CUSTOMER NOTICED DURING PRIMING THAT THE STOPCOCK WAS BROKEN INSIDE OF THE ARTERIAL LINE WHEN IT WAS SCREWED INTO PLACE.THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECT OCCURRED. ADDITIONAL INFORMATION: THERE WAS NO DAMAGE REPORTED TO THE OUTER BOX. THERE WAS NO OTHER DAMAGE REPORTED TO ANY OTHER COMPONENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17125405 ↗MalfunctionNov 1, 2022Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED A LEAK WAS NOTICED FROM THE LINE CONNECTION. THE CUSTOMER STATED THAT THE BLOOD LOSS WAS MINIMAL, AND A BLOOD TRANSFUSION WAS NOT REQUIRED. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADDITIONAL PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO VISIBLE AIR IN SYSTEM/TUBING REPORTED. THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 15708007 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.
- Z-1872-2021 · Apr 29, 2021 · TerminatedMedtronic HY8M63R4, Custom Pack HY8M63R4 Adult ECC, Custom Perfusion System, Sterile, Rx.
- Z-1890-2021 · Apr 29, 2021 · TerminatedMedtronic HY10L28R3, Custom Pack HY10L28R3 1/4 Pack, Custom Perfusion System, Sterile, Rx.
- Z-1880-2021 · Apr 29, 2021 · TerminatedMedtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
- Z-1873-2021 · Apr 29, 2021 · TerminatedMedtronic BB10R94R6, Custom Pack BB10R94R6 1/4 ECCPK, Custom Perfusion System, Sterile, Rx.
- Z-1874-2021 · Apr 29, 2021 · TerminatedMedtronic BB11J14R2, Custom Pack BB11J14R2 Bridge, Custom Perfusion System, Sterile, Rx.
- Z-1879-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W63R1, Custom Pack CB10W63R1 HL 20 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1895-2021 · Apr 29, 2021 · TerminatedMedtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.
- Z-1876-2021 · Apr 29, 2021 · TerminatedMedtronic CB7J97R8, Custom Pack CB7J97R8 U14 E Sup, Custom Perfusion System, Sterile, Rx.
- Z-1878-2021 · Apr 29, 2021 · TerminatedMedtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
- Z-1885-2021 · Apr 29, 2021 · TerminatedMedtronic HY10Y54R4, Custom Pack HY10Y54R4 ECC PED 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1877-2021 · Apr 29, 2021 · TerminatedMedtronic HY11B40R1, Custom Pack HY11B40R1 1/4 Respirato, Custom Perfusion System, Sterile, Rx.
- Z-1889-2021 · Apr 29, 2021 · TerminatedMedtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
- Z-1871-2021 · Apr 29, 2021 · TerminatedMedtronic HY10Z48R2, Ctr., Custom Pack HY10Z48R2 Adult ECC 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1894-2021 · Apr 29, 2021 · TerminatedMedtronic BB9G78R8, Custom Pack BB9G78R8 ECC Supp, Custom Perfusion System, Sterile, Rx.
- Z-1888-2021 · Apr 29, 2021 · TerminatedMedtronic CB10R55R5, Custom Pack CB10R55R5 3/8X3/8 Loop, Custom Perfusion System, Sterile, Rx.
…and 10 more.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY11R00R6Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page