SYNCHROMED®II 8637-40 (UDI 00763000422615)Recalls & reported problems

Medtronic, Inc. · PUMP 8637-40 SM II 40ML EMAN SYMBL GRC · Class 3 · Implant

All models of SYNCHROMEDII (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic SynchroMed II is an implantable pump that delivers medicine into the body. The 1,471 reports most often describe the pump moving out of place, infusion or flow problems, too little medicine being delivered, and communication failures. The reports list 3 deaths and 981 injuries. One open recall names this device, started in November 2023, over the pump not resuming drug delivery after an MRI scan when it had been sounding an alarm.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,463
Deaths reported
3
Injuries reported
983
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1,463 adverse event reports that list this exact device (DI 00763000422615), received between Jul 15, 2021 and Sep 30, 2026 — classified as 3 deaths, 983 injuries, 477 malfunctions.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (235 reports).

1 FDA recall record name this device's identifier, 1 still open.

39 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 23 reportsJan 2024: 28 reportsFeb 2024: 30 reportsMar 2024: 16 reportsApr 2024: 16 reportsMay 2024: 16 reportsJun 2024: 31 reportsJul 2024: 23 reportsAug 2024: 24 reportsSep 2024: 21 reportsOct 2024: 19 reportsNov 2024: 20 reportsDec 2024: 10 reportsJan 2025: 30 reportsFeb 2025: 18 reportsMar 2025: 21 reportsApr 2025: 12 reportsMay 2025: 21 reportsJun 2025: 17 reportsJul 2025: 22 reportsAug 2025: 16 reportsSep 2025: 12 reportsOct 2025: 12 reportsNov 2025: 14 reportsDec 2025: 16 reportsJan 2026: 28 reportsFeb 2026: 49 reportsMar 2026: 39 reportsApr 2026: 13 reportsMay 2026: 11 reportsJun 2026: 10 reportsJul 2026: 5 reportsAug 2026: 3 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202649
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 13, 2023Class IIOpen, Classified

    Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.

    Z-0615-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem446
  • Migration or Expulsion of Device235
  • Infusion or Flow Problem213
  • Insufficient Device Problem Information177
  • Insufficient Flow or Under Infusion159
  • Communication or Transmission Problem128
  • Obstruction of Flow118
  • Material Integrity Problem105

Patient problems

  • No Clinical Signs, Symptoms or Conditions649
  • Pain253
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available217
  • Insufficient Information143
  • Unspecified Infection129
  • Muscular Rigidity83
  • Seroma59
  • Cerebrospinal Fluid Leakage46

Recent adverse event reports

The 25 most recent of 1,463 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Insufficient Flow or Under Infusion; Insufficient Device Problem Information

    Additional information was received from a manufacturer representative (rep) stating that the exact volumes were unknown regarding the discrepancy. They also added that they suspected a roller malfunction.

    Patient problems: Pain

    MAUDE report 50948157 ↗
  • InjurySep 22, 2026Insufficient Flow or Under Infusion; Material Integrity Problem; Insufficient Device Problem Information

    Information was received from a patient who was receiving Dilaudid (hydromorphone) via an implantable pump for non-malignant pain. It was reported that a couple days ago (b)(6) 2026 the patient's pump started alarming. The patient had been having problems with the pump not working properly. The patient was supposed to go on the 24th to have the pump replaced. The patient then got the message empty reservoir alarm occurred contact your clinician immediately for assisted service code 98. The issue was not resolved. The caller was redirected to their healthcare provider to further address the issue. Additional information was received from a healthcare provider (hcp). It was reported that the Hcp stated that patient being seen to silence a low reservoir alarm. The caller stated that the pump was going to be replaced due to underdosing and overdosing and it would not be filled in the meantime. The Hcp programmed a reservoir volume and silenced pump alarm.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50859774 ↗
  • InjurySep 16, 2026Pumping Stopped

    Information was received from multiple sources (Manufacturer Representative, Healthcare Provider, Clinical Study) regarding a patient who was receiving Baclofen (2000 mcg/ml at 1,543 mcg/day) via an implantable pump for unknown indications for use. It was reported that the patient had been having some episodes of concerning seizure activity. She had a side port tap on (b)(6) 2025, which showed a patent intrathecal catheter. On (b)(6) 2026 the patient was admitted with a complex medical history including a baclofen pump for hypertonia. The pump had started to experienced intermittent mechanical stalls. They self-recovered, but given the unexpected and uncertain nature of these stalls the Doctor felt the most appropriate step was to replace the pump as it is not guaranteed that this would restart after any given stall. Additional information received from the healthcare provider via a clinical study indicated that it was noted that the alarm had been occurring for several weeks every 1 to 2 days but would recover each time and was giving the appropriate daily dose of IT medication. A Catheter Access Port (CAP) contrast study was done and no issue was found. An Xray was performed and

    Patient problems: Convulsion/Seizure; Muscular Rigidity

    MAUDE report 50793191 ↗
  • InjuryAug 26, 2026Human-Device Interface Problem; Adverse Event Without Identified Device or Use Problem

    Information was received from a patient receiving Baclofen via an implantable infusion pump for non-malignant pain. It was reported that a few years ago, (b)(6) 2024, the hospital overdosed them by 'changing from concentration to compound and adjusting their dosing'. The patient called back and reiterated they were overdosed by the local hospital who switched their pump 'from 2000 to compound'. That hospital no longer checks the patient's pump.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50461926 ↗
  • InjuryAug 23, 2026Insufficient Device Problem Information

    Information was received from the healthcare provider (HCP) regarding a patient receiving intrathecal medication via an implantable pump. It was reported that the patient had a pump and catheter "revision"/explant. The reason for the pump and catheter removal were listed as "other".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50397257 ↗
  • InjuryAug 3, 2026Insufficient Device Problem Information

    Information was received from a patient with an implantable pump for non-malignant pain. It was reported that they fell in 2023 and had to have the pump replaced.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50123986 ↗
  • InjuryJul 29, 2026Infusion or Flow Problem

    Information was received from a healthcare provider via company representative regarding a patient with an implantable pump containing Compounded Baclofen (900 mcg/ml at 300 mcg/mo). It was reported that the patient experienced a loss of therapy. It was not known if there were any environmental/external/patient factors that may have led or contributed. It was reported that the 8780 catheter was explanted and replaced with a new 8780. Additional information was received from a healthcare provider via company representative stating that the cause of the patient's lack of therapy was not known. It was noted that during the replacement there was no flow from the catheter. There were no obvious cause for flow disruption discovered. After replacement of a new catheter that resolved the issue.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50069194 ↗
  • InjuryJul 29, 2026Insufficient Device Problem Information

    Information was received from a friend/family member regarding a patient who was receiving Dilaudid (hydromorphone) and Unknown Baclofen via an implantable pump for intractable spasticity. It was reported that patient was admitted to the hospital last night after the pain doctor adjusted the pump yesterday. Caller stated patient's blood pressure reached 200/100 and could not be lowered. Caller also stated patient had never received a bolus before, and the bolus had worked fine before the pump adjustment. Caller reported the doctor was unsure how to manage the pump and called Patient and Technical Services to ask whom to contact to interrogate the device in person, with concern that the pump might be malfunctioning. Agent redirected caller to healthcare provider (HCP).

    Patient problems: High Blood Pressure/ Hypertension

    MAUDE report 50067985 ↗
  • InjuryJul 29, 2026Adverse Event Without Identified Device or Use Problem

    Information was received from the patient via the company representative (Rep) regarding a patient receiving intrathecal medication via an implantable pump. The patient's pump system implanted in 2022 was explanted a month after it was implanted due to infection. Additional information provided by the company representative (Rep) reported that this issue had resolved.

    Patient problems: Unspecified Infection

    MAUDE report 50067367 ↗
  • MalfunctionJul 28, 2026Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50052928 ↗
  • InjuryJul 21, 2026Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM A COMPANY REPRESENTATIVE (REP) REGARDING A PATIENT RECEIVING INTRATHECAL BACLOFEN (UNKNOWN) 2000 MCG/ML VIA AN IMPLANTABLE PUMP. THE COMPANY REP REPORTED THAT PATIENT HAD AN INFECTION AT THE PUMP POCKET SITE AND THE HEALTHCARE PROVIDER (HCP) WANTED TO LOWER THE INFUSION RATE WITHOUT REFILLING THE PUMP. THE PUMP CURRENTLY HAD 2000 MCG/ ML AND THE COMPANY REP STATED THAT THE PUMP WAS RUNNING AT THE LOWEST FLOW RATE FOR THAT CONCENTRATION. THE COMPANY REP WANTED TO KNOW IF THEY COULD SOMEHOW PROGRAM THE PUMP TO GO SLOWER WITHOUT CHANGING THE CONCENTRATION OF THE PUMP.

    Patient problems: Unspecified Infection

    MAUDE report 25892087 ↗
  • InjuryJun 25, 2026Infusion or Flow Problem

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS RECEIV ING UNKNOWN DRUG VIA AN IMPLANTABLE PUMP. IT WAS REPORTED THAT THE PATIENT NTOICED INCREASE IN OLD PAIN. THE CATHETER WAS CHECKED DURING REVISION AND THEY WERE NOT ABLE TO ASPIRATE AT ANY POINT. NO VISIBILE KINKS WERE SEEN, THE PHYSICIAN ELECTED TO REPLACE THE CATHETER.

    Patient problems: Insufficient Information

    MAUDE report 25627893 ↗
  • MalfunctionJun 23, 2026Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A PATIENT WITH AN IMPLANTABLE PUMP FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT IT HAD BEEN QUITE A WHILE SINCE THE HANDSET KEPT MESSING UP EVERY TIME THEY TRIED TO CONNECT TO THE APP AND GET A BOLUS. THEY MENTIONED THE SCREEN FROZE UP TO THE LOCKOUT SCREEN EACH TIME THEY CONNECTED. THEY MENTIONED THEIR LOCKOUT DURATION WAS EVERY 2 HOURS. THE AGENT REVIEWED 90% LAG. THE AGENT HAD THE PATIENT CLOSE ALL APPS AND WALKED THEM THROUGH CLEARING OUT THE DATA. THEY CONNECTED TO THE PUMP WITHOUT ANY LAG OR ISSUES, AND THEY SAW THE TUTORIALS SCREEN AND MENTIONED THEY WERE CURRENTLY LOCKED OUT. THE AGENT REVIEWED INFORMATION AND BEST PRACTICES. THE PATIENT SAID THEY WOULD MONITOR THE ISSUE AND WOULD CALL BACK IF THE ISSUE PERSISTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25596961 ↗
  • MalfunctionJun 23, 2026Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A PATIENT (CON) REGARDING AN EXTERNAL DEVICE TO AN IMPLANTABLE PUMP INFUSING UNKNOWN DRUG FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE PATIENT MENTIONED THAT IT WAS TAKING A WHILE TO PROGRESS THRU THE PERCENTAGES SO THE AGENT ASSISTED THE PATIENT WITH CLEARING HANDSET DATA. THE TROUBLESHOOTING STEPS THAT WERE TAKEN ON THE CALL RESOLVED THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25595254 ↗
  • InjuryJun 17, 2026Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE (REP) REGARDING A PATIENT WHO WAS RECEIVING UNKNOWN BACLOFEN (2000 MCG/ML AT 102.1 MCG/DAY) VIA AN IMPLANTABLE PUMP FOR INTRACTABLE SPASTICITY. IT WAS REPORTED THAT THE PATIENT FELT THEY WERE HAVING AN INCREASE IN GROIN PAIN AFTER THEIR LAST PUMP REPLACEMENT. THE REPORTED AN INCREASE IN GROIN PAIN WHEN SITTING IN A CHAIR. THEY STATED THAT THEY FELT MORE PRESSURE ON THEIR GROIN WHEN SITTING FROM THE PUMP AND WOULD NEED TO MANIPULATE OR REPOSITION TO EASE THE PAIN PATTERN. THE PATIENT UNDERWENT A PUMP POCKET REVISION TO RELOCATE THE PUMP IN THE PATIENT'S ABDOMEN. THE ISSUE WAS RESOLVED AT THE TIME OF THE REPORT. THE HCP HAD NO ADDITIONAL INFORMATION FOR THE MANUFACTURER.

    Patient problems: Insufficient Information; Pain

    MAUDE report 25540832 ↗
  • MalfunctionJun 15, 2026Display or Visual Feedback Problem; Charging Problem; Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM THE PATIENT RECEIVING INTRATHECAL MEDICATION VIA AN IMPLANTABLE PUMP FOR NON-MALIGNANT PAIN. THE PATIENT REPORTED THAT THE COMMUNICATOR WAS NOT CHARGING AND THEY NOTICED THE ISSUE ABOUT A WEEK AGO. PATIENT MENTIONED THE CHARGING CORD HAD WORN OUT AND WHEN THEY TRY TO CONNECT TO THE COMMUNICATOR THEY NEEDED TO "WIGGLE AROUND" AND IF IT GETS CONNECTED THE LIGHT WOULD NOT GO GREEN, IT ONLY "STAYED RED ALL THE TIME". AGENT SENT REPAIR REQUEST TO REPLACE THE COMMUNICATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25504840 ↗
  • MalfunctionJun 12, 2026Display or Visual Feedback Problem; Failure to Power Up

    INFORMATION WAS RECEIVED FROM A PATIENT REPRESENTATIVE (REP) REGARDING A PATIENT RECEIVING FENTANYL VIA AN IMPLANTED PUMP. IT WAS REPORTED THAT THE REPLACEMENT COMMUNICATOR HAD BEEN PLUGGED IN FOR 2 DAYS AND WAS STILL SHOWING BATTERY INDICATOR LIGHT OF YELLOW AND WOULD NOT POWER ON. THE PATIENT REP TRIED DIFFERENT CORDS AND CHARGING BLOCKS AND DIFFERENT OUTLETS. AGENT SENT REQUEST TO REPAIR. THE PATIENT REP WOULD CALL BACK IF THEY NEED FURTHER ASSISTANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25486685 ↗
  • MalfunctionJun 11, 2026Charging Problem

    INFORMATION WAS RECEIVED FROM A CONSUMER REGARDING A PATIENT RECEIVING FENTANYL VIA AN IMPLANTED PUMP. THE INDICATION FOR PUMP USE WAS SPINAL PAIN. THE PATIENT'S REPRESENTATIVE REPORTED THAT AS OF A COUPLE DAYS AGO, THE PATIENT¿S COMMUNICATOR WENT OUT AND WAS DEAD. THEY TRIED PLUGGING THE CORD IN A DIFFERENT OUTLET, BUT IT DID NOT TURN ON OR RECHARGE, SO THEY COULD NOT DELIVER A BOLUS NOW. DURING THE CALL, THE CALLER TRIED USING A DIFFERENT CORD BUT VERIFIED IT DID NOT CHARGE. A REPLACEMENT COMMUNICATOR WAS REQUESTED FROM THE REPAIR DEPARTMENT BECAUSE THE COMMUNICATOR WOULD NOT RECHARGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25472950 ↗
  • MalfunctionJun 8, 2026Application Program Problem; Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A PATIENT REGARDING AN EXTERNAL DEVICE TO AN IMPLANTABLE PUMP INFUSING DILAUDID (HYDROMORPHONE) FOR SPINAL PAIN. IT WAS REPORTED THAT THE PATIENT REPORTED THEY WERE HAVING TROUBLE THE COMMUNICATOR SINCE LAST FRIDAY ((B)(6) 2026). THE PATIENT STATED THEY HAD A BAD COMMUNICATOR AND THEY RECEIVED A NEW COMMUNICATOR A COUPLE WEEKS AGO OR A WEEK OR SO AGO (2026-MAY-15). THE PATIENT MENTIONED THEY GOT THE EQUIPMENT BACK AND IT DID NOT WORK. THE PATIENT REPORTED GETTING A MESSAGE THAT THE APPLICATION HAD UNEXPECTEDLY STOPPED AND THEY HAD TO RESYNC ALL OVER AGAIN. THE PATIENT STATED THEY WERE GETTING SYSTEM ERROR 0X7500SA93 AND RESYNCED. THE AGENT ASSISTED PATIENT WITH CLOSING OUT ALL THE RUNNING APPLICATION, RESET THE HANDSET AND COMMUNICATOR AND CONNECT WITH THE DEVICE AND PATIENT SAID THEY COULD NOT REMEMBER WHERE THE PUMP WAS PLACED ON THE BODY. THE PATIENT STATED THEY WERE GETTING NO PUMP RESPONSE AND THE SCREEN GET STUCK AT 75% WHEN TRYING TO CONNECT. THE AGENT ASSISTED THE PATIENT WITH CLEARING DATA ON THE HANDSET. THE PATIENT SERVICE ASKED PATIENT TO CONNECT WITH THE PATIENT THERAPY MANAGER (MYPTM) APPLICATION AND PATIENT WAS GETTING NO PUMP RESPONSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25436419 ↗
  • MalfunctionJun 5, 2026Electromagnetic Interference

    INFORMATION WAS RECEIVED FROM A COMPANY REPRESENTATIVE (REP) REGARDING A PATIENT RECEIVING INTRATHECAL MEDICATION VIA AN IMPLANTABLE PUMP FOR INTRACTABLE SPASTICITY. THE COMPANY REP REPORTED THAT THE PATIENT WAS HAVING PUMP INTERROGATED AND THEY WERE SEEING CODE 100 [PUMP STALLED] . THE COMPANY REP STATED PATIENT DID HAVE A MAGNETIC RESONANCE IMAGING (MRI) YESTERDAY AND IT WAS NOT CHECKED POST MRI.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25411516 ↗
  • InjuryJun 4, 2026Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT RECEIVING UNKNOWN MORPHINE VIA AN IMPLANTABLE PUMP FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE HCP STATED THAT THE PUMP HAD POTENTIALLY MALFUNCTIONED BECAUSE THE PATIENT PRESENTED WITH SYMPTOMS CONSISTENT WITH ACUTE WITHDRAWAL. THE HCP STATED THAT THE PATIENT HAD TO BE INTUBATED. THE HCP STATED THEY HAD SPOKEN TO SOMEONE YESTERDAY AND WERE TOLD THAT THEY WOULD BE THERE TODAY TO CHECK THE PUMP, BUT NO ONE HAD COME. AN AGENT TRANSFERRED THE HCP TO THE NATIONAL ANSWERING SERVICE (NAS).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25392542 ↗
  • MalfunctionMay 29, 2026Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A PATIENT WITH AN IMPLANTABLE PUMP FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE HANDSET WAS LAGGING AT 90%. THE PATIENT MENTIONED THAT IT SEEMED LIKE THE EQUIPMENT WAS FINNICKY WHEN IT WAS PLACED OVER THE PUMP IN ORDER TO FIND THE PUMP. THE AGENT REVIEWED CONSIDERATIONS AND SUGGESTIONS WITH THE PATIENT THE AGENT REVIEWED AND GUIDED THE PATIENT THROUGH CLEARING THE DATA. THE PATIENT WAS THEN ABLE TO CONNECT TO THE PUMP WITHOUT ANY LAG. WHEN ASKED FOR THE EVENT DATE, THE PATIENT SAID IT WAS PROBABLY THREE YEARS AGO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25334713 ↗
  • MalfunctionMay 25, 2026Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A PATIENT REGARDING AN IMPLANTED INFUSION PUMP FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE PATIENT WAS HAVING ISSUES CONNECTING THE HANDSET TO THE COMMUNICATOR. A LOT OF THE TIME, WHEN THEY OPENED THE MYPTM APPLICATION AND PLACED THE COMMUNICATOR OVER THE PUMP, THE LOADING SCREEN WOULD KEEP ON SPINNING AND WOULD NOT CONNECT EVEN AFTER MOVING THE COMMUNICATOR AROUND. THE PATIENT ALSO MENTIONED THAT THE DEVICE TOOK A WHILE TO CONNECT AND THEY HAD TO RESYNC SEVERAL TIMES. PATIENT SERVICES REVIEWED INFORMATION. THE PATIENT STATED THAT THE NURSE HAD ALREADY HELPED THEM TO CLEAR THE DATA AND THE CACHE ON (B)(6) 2026 BUT THEY WERE STILL HAVING ISSUES CONNECTING. THE PATIENT STATED THAT TH ISSUE STARTED "PROBABLY COUPLE OF WEEKS AGO". THE PATIENT MENTIONED THAT THE NURSE ASKED THEM ABOUT THE LOADING SCREEN GETTING STUCK AT 90% BUT THE PATIENT WAS ABLE TO DO A BOLUS ON (B)(6) 2026 WITHOUT A LAG. PATIENT SERVICES REVIEWED INFORMATION. THE PATIENT EXPLAINED THE ISSUE VAGUELY. DURING THE CALL TO PATIENT SERVICES, THE PATIENT WAS ALREADY CONNECTED TO THE MYPTM APPLICATION AND IT SHOWED THAT THEY WERE LOCKED OUT FOR AN HOUR AND 55 MINUTES. PATIENT SERVICES HAD

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25270750 ↗
  • MalfunctionMay 20, 2026Inappropriate or Unexpected Reset

    INFORMATION WAS RECEIVED FROM A PATIENT AND A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE (REP) REGARDING A PATIENT RECEIVING BUPIVACAINE AND DILAUDID VIA AN IMPLANTABLE PUMP FOR SPINAL PAIN INDICATIONS. IT WAS REPORTED THAT THE REP GOT A CALL ON (B)(6) THAT THE PATIENT'S PUMP GOT AN ERROR CODE THAT IT HAD RESET. THE REP STATED THAT THE ERROR CODE WAS 101 (PUMP RESET OCCURRED) AND THEN NOTED CODE 87 FOR AN UNDERDOSE (PUMP IN SAFE MODE). THE REP STATED THAT THE PATIENT REPORTED HAVING MORE PAIN AND SOME WITHDRAWAL SYMPTOMS. THE REP ENDED UP MEETING WITH THE PATIENT AND LOOKED AT THE LOGS WHICH SHOWED A PUMP RESET, BUT THE PATIENT WANTED TO KNOW WHY THAT WOULD HAVE OCCURRED. AN AGENT ASKED THE REP IF THE PATIENT WAS AROUND ANY TYPE OF WELDING EQUIPMENT OR STRONGER ELECTROMAGNETIC INTERFERENCE (EMI) SOURCES, AND THE REP SAID THAT THE DAY BEFORE THE PATIENT WAS PREVIOUSLY WORKING OUT IN HIS GARAGE, BUT NOT THAT DAY, SO THE PUMP WENT OFF OUTSIDE OF THOSE PARAMETERS. THE REP STATED THAT THE PUMP WAS UPDATED TO THE DESIRED SET TINGS AND THE PATIENT REPORTED FEELING BETTER SOON AFTER. THIS INFORMATION WAS LATER RE-REPORTED BY ANOTHER REP. THIS REP STATED THAT THE PUMP LOGS

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain

    MAUDE report 25233731 ↗
  • InjuryMay 19, 2026Insufficient Device Problem Information

    INFORMATION WAS RECEIVED FROM A CONSUMER REGARDING A PATIENT RECEIVING BACLOFEN VIA AN IMPLANTABLE PUMP. THE INDICATION FOR USE WAS INTRACTABLE SPASTICITY. THE PATIENT REPORTED THAT THEIR PUMP ALARM GOES OFF EVERY HOUR. THE PATIENT ASKED IF THAT WOULD EVENTUALLY DECREASE LIKE 45MINUTES, 30MINUTES. THE PATIENT STATED THAT THEY HAD BEEN GETTING A SINGLE TONE ALARM SINCE THE 15TH OF THIS MONTH BUT MENTIONED THEY WERE TOLD IT WAS SUPPOSED TO BE ON THE 14TH. DURING THE CALL, THE PATIENT ALSO MENTIONED THE TWO-TONE ALARMS THAT THEY MIGHT GET. THE AGENT REVIEWED THE INFORMATION AND DEVICE FUNCTION. THE PATIENT DID NOT HAVE A HEALTHCARE PROVIDER AT THIS TIME AS THEY WERE LOOKING FOR A HEALTHCARE PROVIDER WHO ACCEPTS THEIR INSURANCE. THE PATIENT REQUESTED AND WAS SENT PHYSICIAN LISTINGS. ADDITIONAL INFORMATION WAS RECEIVED FROM THE PATIENT'S REPRESENTATIVE ON 2026-MAY-07. THE CALLER REPORTED THAT THE PUMP HAD BEEN BEEPING FOR 2 WEEKS SINCE LATE APRIL. THE CALLER STATED WHEN THEY FINALLY GOT AN APPOINTMENT WITH THE NEW HEALTHCARE PROVIDER (HCP), THE DOCTOR LOOKED AT THE PUMP INCISION AREA, AND IT WAS SWOLLEN. THE DOCTOR SAID THEY THOUGHT THE PUMP WAS INFECTED SO THEY DIDN'T WANT TO DO ANYTHI

    Patient problems: Cramp(s) /Muscle Spasm(s); Swelling/ Edema; Unspecified Infection

    MAUDE report 25217909 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P860004 on Mar 14, 1988. Submitted by Medtronic, Inc. as “Medtronic(R) Synchromed(Tm) Pump & Infusion System”. The FDA review took 770 days. The FDA has since approved 445 supplements (changes to the device or its labeling), most recently on Oct 30, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Recalls under the same FDA submission

These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.

  • Z-2201-2026 · Apr 8, 2026 · Open, Classified
    Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
  • Z-1506-2025 · Jan 9, 2025 · Open, Classified
    A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
  • Z-2352-2024 · May 10, 2024 · Open, Classified
    Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
  • Z-0614-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-20, Programmable pump
  • Z-0522-2024 · Jul 13, 2020 · Completed
    Medtronic SynchroMed, Model A10
  • Z-0508-2020 · Oct 11, 2019 · Terminated
    Synchromed II programmable pump, Model numbers 8637-20 and 8637-40
  • Z-0537-2018 · Sep 21, 2017 · Terminated
    Medtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
  • Z-0896-2018 · Aug 9, 2017 · Open, Classified
    Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
  • Z-0788-2017 · Oct 3, 2016 · Terminated
    Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
  • Z-1271-2016 · Feb 9, 2016 · Terminated
    Medtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
  • Z-1681-2015 · Apr 10, 2015 · Terminated
    Medtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
  • Z-1060-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow
  • Z-1062-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
  • Z-1061-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
  • Z-2276-2009 · Jul 1, 2009 · Terminated
    Medtronic SynchroMed II Programmable Pump, 8637, Rx Only, Sterilized by ethylene oxide gas, Medtronic inc., Minneapolis, MN 55432-5604. (The SynchroMed II Pump is supplied in 20 m

…and 24 more.

Common questions about the SYNCHROMED®II 8637-40

Has the SYNCHROMED®II 8637-40 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0615-2024 (Class II), began on Nov 13, 2023: Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the SYNCHROMED®II 8637-40?

The FDA's MAUDE database lists 1,463 adverse event reports for this device, received between Jul 15, 2021 and Sep 30, 2026. In those reports, the most common specific device problems are “Migration or Expulsion of Device” (235), “Infusion or Flow Problem” (213) and “Insufficient Flow or Under Infusion” (159); the most common patient problems are “Pain” (253), “Unspecified Infection” (129) and “Muscular Rigidity” (83). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SYNCHROMED®II 8637-40?

3 reports are classified as deaths, 983 as injuries and 477 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SYNCHROMED®II 8637-40 FDA approved?

It was approved through premarket approval (PMA) P860004 on Mar 14, 1988.

Is the SYNCHROMED®II 8637-40 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the SYNCHROMED®II 8637-40, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand SYNCHROMED®II, model 8637-40). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Pump, infusion, implanted, programmable devices

All Pump, infusion, implanted, programmable devices →