Medtronic SynchroMed II, Model 8637-40, Programmable pump recallZ-0615-2024 · Medtronic Neuromodulation · initiated Nov 13, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION notice to healthcare providers and medical facilities on 77/13/2023 via mail and email. The notice explained the problem with the device, hazards to patient, and recommended the following: Patient Management Recommendations: . Upon completion of an MRI scan, interrogate the pump with the Clinician Programmer (or Personal Therapy Manager) to end telemetry mode and resume drug delivery. . Consult the MRI Guidelines for additional information on MRI preparation and post-examination review, and motor stall recovery timing (see MRI Guidelines at www.manuals.medtronic.com). . Remind your patients about the importance of interrogating the SynchroMed II pump after an MRI to ensure continuation of therapy. . Educate patients, caregivers, and family members to recognize the signs and symptoms associated with intrathecal drug therapy underdose or withdrawal. Patients receiving intrathecal baclofen therapy (e.g., Lioresal Intrathecal) are at higher risk for adverse events, as baclofen withdrawal can lead to a life-threatening condition if not treated promptly and effectively. For questions or concerns associated with this correction, please contact your Medtronic Representative or Technical Services at 1-800-707-0933. On 07/10/2024, Medtronic sent an additional notification to customers to correct an error in the previous communication. The previous communication incorrectly included a reference to the use of the Patient Therapy Manager as part of the recommended mitigations. The updated communication has been revised to remove reference to the Patient Therapy Manager. All other aspects of the mitigations remain unchanged.
Recall details
- Recall number
- Z-0615-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 13, 2023
- Posted by the FDA
- Dec 22, 2023
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 78532 units
- Product code
- LKK
- FDA submissions
- P860004
- Distribution
- Worldwide
Product description
Medtronic SynchroMed II, Model 8637-40, Programmable pump
Code information (models, lots, serial numbers)
Show the full code information (946k characters)
GTIN 00613994483195, Serial Numbers: NGV502657H, NGV508302H, NGV612281H, NGV612282H, NGV717221H, NGV723564H, NGV727392H, NGV727397H, NGV727400H, NGV727906H, NGV730310H, NGV741180H; GTIN 00643169100824, Serial Numbers: NGV503518H, NGV503522H, NGV504577H, NGV504580H, NGV504581H, NGV507269H, NGV507271H, NGV507272H, NGV507273H, NGV507431H, NGV507433H, NGV507434H, NGV507436H, NGV508247H, NGV508248H, NGV508257H, NGV509489H, NGV509490H, NGV510082H, NGV512179H, NGV512180H, NGV502833H, NGV503294H, NGV502489H, NGV511809H, NGV510811H, NGV507369H, NGV511074H, NGV512393H, NGV502490H, NGV509057H, NGV506230H, NGV503295H, NGV507614H, NGV509098H, NGV509769H, NGV503297H, NGV502903H, NGV504121H, NGV511164H, NGV512019H, NGV502487H, NGV504523H, NGV503292H, NGV511885H, NGV511163H, NGV509373H, NGV502488H, NGV506232H, NGV510872H, NGV509770H, NGV510474H, NGV512471H, NGV508053H, NGV504241H, NGV507628H, NGV507635H, NGV503352H, NGV512033H, NGV512037H, NGV507723H, NGV508260H, NGV509172H, NGV509477H, NGV503596H, NGV504429H, NGV502450H, NGV507824H, NGV510692H, NGV505405H, NGV506725H, NGV511877H, NGV512666H, NGV505006H, NGV509513H, NGV500356H, NGV501018H, NGV503021H, NGV503099H, NGV503186H, NGV503187H, NGV503487H, NGV504710H, NGV504901H, NGV505165H, NGV508052H, NGV509450H, NGV509762H, NGV512254H, NGV513130H, NGV505328H, NGV506132H, NGV508199H, NGV504567H, NGV512052H, NGV501951H, NGV502809H, NGV502344H, NGV506202H, NGV505245H, NGV505772H, NGV508420H, NGV510444H, NGV510215H, NGV502774H, NGV504027H, NGV509803H, NGV511504H, NGV504240H, NGV504389H, NGV511052H, NGV504392H, NGV507692H, NGV509561H, NGV511591H, NGV500401H, NGV503155H, NGV503634H, NGV504300H, NGV504477H, NGV504499H, NGV504503H, NGV504866H, NGV505396H, NGV505728H, NGV506421H, NGV506888H, NGV507130H, NGV507553H, NGV510135H, NGV510995H, NGV511124H, NGV512684H, NGV501082H, NGV507786H, NGV500391H, NGV500398H, NGV507732H, NGV504276H, NGV500358H, NGV502146H, NGV510357H, NGV502422H, NGV502423H, NGV506876H, NGV509283H, NGV512496H, NGV513129H, NGV505345H, NGV509783H, NGV511611H, NGV511903H, NGV503233H, NGV502982H, NGV503659H, NGV505207H, NGV507109H, NGV511643H, NGV511698H, NGV511741H, NGV506755H, NGV502367H, NGV504370H, NGV504375H, NGV510066H, NGV505958H, NGV501978H, NGV502023H, NGV503397H, NGV503043H, NGV506006H, NGV508280H, NGV511149H, NGV511064H, NGV508139H, NGV504019H, NGV502608H, NGV505225H, NGV511757H, NGV506199H, NGV501014H, NGV501896H, NGV511111H, NGV511731H, NGV502132H, NGV506176H, NGV506204H, NGV506529H, NGV506554H, NGV507088H, NGV507165H, NGV507678H, NGV508306H, NGV510328H, NGV502994H, NGV505401H, NGV506284H, NGV507615H, NGV502156H, NGV505314H, NGV501090H, NGV505329H, NGV505457H, NGV505500H, NGV505648H, NGV509119H, NGV510330H, NGV511806H, NGV504679H, NGV507077H, NGV508938H, NGV513135H, NGV512645H, NGV504876H, NGV505939H, NGV507572H, NGV503898H, NGV512833H, NGV506826H, NGV508758H, NGV512338H, NGV502787H, NGV504247H, NGV507146H, NGV507573H, NGV510448H, NGV511060H, NGV511063H, NGV512164H, NGV505128H, NGV511725H, NGV501823H, NGV504275H, NGV504495H, NGV509382H, NGV511752H, NGV506188H, NGV509553H, NGV500349H, NGV509097H, NGV512565H, NGV507524H, NGV513125H, NGV510408H, NGV501983H, NGV505806H, NGV500384H, NGV501010H, NGV501011H, NGV507538H, NGV510807H, NGV501844H, NGV502044H, NGV502420H, NGV503287H, NGV503382H, NGV505310H, NGV505381H, NGV509055H, NGV502466H, NGV502738H, NGV502739H, NGV508421H, NGV508274H, NGV510338H, NGV507613H, NGV510040H, NGV510179H, NGV503750H, NGV511478H, NGV503534H, NGV505211H, NGV503488H, NGV512107H, NGV505139H, NGV506404H, NGV504012H, NGV504586H, NGV505301H, NGV505605H, NGV506027H, NGV506783H, NGV508317H, NGV508862H, NGV510792H, NGV510838H, NGV511535H, NGV512425H, NGV503938H, NGV504808H, NGV507173H, NGV508258H, NGV510565H, NGV504060H, NGV508220H, NGV510758H, NGV506221H, NGV509318H, NGV504017H, NGV504018H, NGV504955H, NGV507463H, NGV507647H, NGV507867H, NGV508956H, NGV510383H, NGV510386H, NGV512251H, NGV503114H, NGV505218H, NGV508259H, NGV510310H, NGV502571H, NGV512820H, NGV512838H, NGV505475H, NGV502634H, NGV504946H, NGV504337H, NGV504514H, NGV505921H, NGV506073H, NGV506438H, NGV506720H, NGV506827H, NGV507219H, NGV507557H, NGV509222H, NGV509335H, NGV511623H, NGV512306H, NGV502111H, NGV504827H, NGV509076H, NGV510027H, NGV505741H, NGV504954H, NGV510865H, NGV504202H, NGV505195H, NGV506000H, NGV506643H, NGV508261H, NGV503537H, NGV503841H, NGV504812H, NGV506550H, NGV508528H, NGV511104H, NGV502570H, NGV505693H, NGV507363H, NGV507404H, NGV512883H, NGV507039H, NGV506420H, NGV507100H, NGV504194H, NGV504312H, NGV504648H, NGV512661H, NGV506021H, NGV508483H, NGV508942H, NGV505731H, NGV505857H, NGV510252H, NGV511675H, NGV503906H, NGV505413H, NGV508711H, NGV512591H, NGV512161H, NGV512416H, NGV504738H, NGV506049H, NGV507730H, NGV506292H, NGV502266H, NGV502418H, NGV510337H, NGV512650H, NGV509050H, NGV513120H, NGV508221H, NGV511147H, NGV504914H, NGV507232H, NGV507234H, NGV512266H, NGV506907H, NGV501012H, NGV503772H, NGV506540H, NGV507155H, NGV507441H, NGV510068H, NGV511110H, NGV502852H, NGV511704H, NGV510186H, NGV506521H, NGV508496H, NGV511652H, NGV513222H, NGV506952H, NGV508868H, NGV507152H, NGV503542H, NGV501042H, NGV501076H, NGV510213H, NGV509284H, NGV505377H, NGV509370H, NGV503180H, NGV507327H, NGV509077H, NGV509539H, NGV510199H, NGV511708H, NGV505717H, NGV505320H, NGV502441H, NGV503401H, NGV506806H, NGV510587H, NGV511093H, NGV511777H, NGV511783H, NGV507662H, NGV504163H, NGV504751H, NGV504754H, NGV504756H, NGV504758H, NGV510020H, NGV511099H, NGV512401H, NGV510133H, NGV506608H, NGV508857H, NGV511726H, NGV505173H, NGV507211H, NGV512386H, NGV508909H, NGV503547H, NGV509039H, NGV502245H, NGV504262H, NGV506088H, NGV506090H, NGV506509H, NGV506510H, NGV507503H, NGV508515H, NGV509307H, NGV510460H, NGV511462H, NGV511463H, NGV506878H, NGV512739H, NGV505361H, NGV504609H, NGV506805H, NGV501703H, NGV507458H, NGV508542H, NGV508739H, NGV512818H, NGV510322H, NGV501686H, NGV505323H, NGV511030H, NGV502386H, NGV505988H, NGV509312H, NGV509364H, NGV509365H, NGV511143H, NGV503227H, NGV507746H, NGV508491H, NGV509728H, NGV510141H, NGV510534H, NGV506139H, NGV509473H, NGV505858H, NGV506710H, NGV508455H, NGV508465H, NGV509258H, NGV510230H, NGV510829H, NGV510002H, NGV504491H, NGV504626H, NGV508872H, NGV500983H, NGV505869H, NGV505744H, NGV505305H, NGV508931H, NGV510315H, NGV510679H, NGV503441H, NGV504308H, NGV506213H, NGV506722H, NGV510480H, NGV501864H, NGV503523H, NGV508937H, NGV512297H, NGV501828H, NGV506711H, NGV507294H, NGV510074H, NGV510945H, NGV504110H, NGV511994H, NGV503748H, NGV500933H, NGV512679H, NGV508042H, NGV508044H, NGV509263H, NGV504899H, NGV509778H, NGV511068H, NGV503343H, NGV507001H, NGV504310H, NGV504342H, NGV504343H, NGV504344H, NGV504347H, NGV511711H, NGV512078H, NGV504052H, NGV507494H, NGV502475H, NGV502750H, NGV502985H, NGV502986H, NGV504592H, NGV504680H, NGV507048H, NGV507373H, NGV508984H, NGV509343H, NGV510124H, NGV510491H, NGV501855H, NGV504078H, NGV504169H, NGV509919H, NGV500994H, NGV510693H, NGV511603H, NGV502911H, NGV507462H, NGV502169H, NGV502717H, NGV503109H, NGV505818H, NGV506046H, NGV507589H, NGV507648H, NGV508418H, NGV509768H, NGV510075H, NGV510708H, NGV511833H, NGV502230H, NGV509872H, NGV506886H, NGV510362H, NGV502253H, NGV502425H, NGV505872H, NGV506901H, NGV510345H, NGV512211H, NGV512965H, NGV507525H, NGV506115H, NGV507820H, NGV502435H, NGV504554H, NGV504757H, NGV505415H, NGV510821H, NGV502258H, NGV502973H, NGV504860H, NGV506459H, NGV508756H, NGV504486H, NGV512092H, NGV502080H, NGV502371H, NGV504039H, NGV502364H, NGV507624H, NGV510549H, NGV504435H, NGV504443H, NGV504450H, NGV507216H, NGV507536H, NGV509328H, NGV509957H, NGV510045H, NGV504711H, NGV506062H, NGV504735H, NGV507178H, NGV511800H, NGV502440H, NGV509595H, NGV508019H, NGV505908H, NGV506693H, NGV507682H, NGV509601H, NGV510723H, NGV503363H, NGV506018H, NGV509021H, NGV502613H, NGV504287H, NGV507739H, NGV510690H, NGV511536H, NGV511544H, NGV505012H, NGV505026H, NGV508946H, NGV507023H, NGV508057H, NGV507982H, NGV503743H, NGV505535H, NGV505636H, NGV508357H, NGV509068H, NGV510981H, NGV510982H, NGV510985H, NGV503178H, NGV502138H, NGV502361H, NGV502506H, NGV503146H, NGV503163H, NGV504650H, NGV505228H, NGV505229H, NGV505727H, NGV506097H, NGV506313H, NGV506767H, NGV507242H, NGV507245H, NGV507247H, NGV508235H, NGV508747H, NGV508760H, NGV509202H, NGV509468H, NGV509469H, NGV510168H, NGV511737H, NGV511882H, NGV512312H, NGV504155H, NGV505424H, NGV503148H, NGV504987H, NGV504252H, NGV505735H, NGV505767H, NGV508181H, NGV509310H, NGV504651H, NGV510212H, NGV504156H, NGV509831H, NGV501088H, NGV510746H, NGV509344H, NGV501336H, NGV501338H, NGV502130H, NGV502645H, NGV502651H, NGV502724H, NGV502907H, NGV503385H, NGV504349H, NGV504968H, NGV505311H, NGV505745H, NGV506225H, NGV507115H, NGV507117H, NGV507290H, NGV507827H, NGV508202H, NGV508224H, NGV508523H, NGV509303H, NGV509314H, NGV509445H, NGV510087H, NGV510927H, NGV511071H, NGV511795H, NGV512142H, NGV512514H, NGV506879H, NGV507313H, NGV512075H, NGV501949H, NGV512215H, NGV504539H, NGV505514H, NGV507736H, NGV507837H, NGV503461H, NGV507767H, NGV510629H, NGV501912H, NGV501913H, NGV503037H, NGV503288H, NGV504895H, NGV504974H, NGV506064H, NGV506872H, NGV507834H, NGV502296H, NGV510380H, NGV510925H, NGV511968H, NGV512091H, NGV512102H, NGV512377H, NGV512665H, NGV512672H, NGV502456H, NGV504885H, NGV506585H, NGV506777H, NGV508093H, NGV508708H, NGV511043H, NGV508951H, NGV501970H, NGV502658H, NGV503046H, NGV507763H, NGV510579H, NGV512174H, NGV502614H, NGV507231H, NGV506092H, NGV512522H, NGV503074H, NGV503821H, NGV505770H, NGV512060H, NGV502430H, NGV505397H, NGV511880H, NGV504143H, NGV509218H, NGV505800H, NGV503306H, NGV503579H, NGV506776H, NGV507698H, NGV507125H, NGV507715H, NGV512351H, NGV506719H, NGV502271H, NGV503044H, NGV506966H, NGV504139H, NGV506189H, NGV504857H, NGV506489H, NGV507059H, NGV508557H, NGV507633H, NGV510922H, NGV502204H, NGV511161H, NGV507674H, NGV507675H, NGV507680H, NGV511007H, NGV501057H, NGV504068H, NGV504720H, NGV505839H, NGV509032H, NGV511685H, NGV502493H, NGV504911H, NGV505365H, NGV503483H, NGV511094H, NGV505759H, NGV510467H, NGV512612H, NGV501965H, NGV505797H, NGV505807H, NGV506552H, NGV511662H, NGV506948H, NGV507103H, NGV507323H, NGV504773H, NGV507086H, NGV509446H, NGV511506H, NGV505986H, NGV506786H, NGV511490H, NGV509582H, NGV512419H, NGV504998H, NGV502173H, NGV504625H, NGV505208H, NGV506834H, NGV507638H, NGV504030H, NGV508736H, NGV510545H, NGV506303H, NGV509934H, NGV505436H, NGV501528H, NGV507591H, NGV509319H, NGV511162H, NGV512453H, NGV503849H, NGV505742H, NGV508473H, NGV512603H, NGV510185H, NGV510453H, NGV512693H, NGV507669H, NGV509392H, NGV501954H, NGV502293H, NGV502360H, NGV504868H, NGV505801H, NGV506853H, NGV507133H, NGV508190H, NGV508890H, NGV509281H, NGV510284H, NGV504614H, NGV511813H, NGV509852H, NGV512216H, NGV510128H, NGV504686H, NGV506054H, NGV509094H, NGV510508H, NGV510664H, NGV512369H, NGV513250H, NGV502469H, NGV506454H, NGV507774H, NGV509499H, NGV510306H, NGV512668H, NGV505376H, NGV506916H, NGV508238H, NGV510582H, NGV503200H, NGV506544H, NGV509101H, NGV511929H, NGV504367H, NGV508443H, NGV502947H, NGV505181H, NGV502203H, NGV507160H, NGV509128H, NGV511887H, NGV504206H, NGV509287H, NGV501971H, NGV502476H, NGV502639H, NGV504725H, NGV507170H, NGV507180H, NGV507343H, NGV509152H, NGV509185H, NGV509187H, NGV511556H, NGV503574H, NGV512242H, NGV510732H, NGV503347H, NGV505354H, NGV500410H, NGV502636H, NGV504099H, NGV509482H, NGV512000H, NGV512055H, NGV502442H, NGV502865H, NGV506954H, NGV512633H, NGV504616H, NGV504617H, NGV502233H, NGV504405H, NGV507080H, NGV512758H, NGV512772H, NGV506758H, NGV502758H, NGV504897H, NGV505098H, NGV510260H, NGV512148H, NGV503739H, NGV510803H, NGV511493H, NGV503033H, NGV503589H, NGV505494H, NGV510616H, NGV511788H, NGV504191H, NGV510714H, NGV502003H, NGV502899H, NGV506629H, NGV509550H, NGV509590H, NGV502741H, NGV504925H, NGV503223H, NGV507411H, NGV510102H, NGV510859H, NGV510861H, NGV512103H, NGV512924H, NGV502424H, NGV502974H, NGV504032H, NGV510648H, NGV505842H, NGV509073H, NGV501668H, NGV502030H, NGV503318H, NGV506197H, NGV509309H, NGV509602H, NGV501926H, NGV505700H, NGV512578H, NGV507683H, NGV506133H, NGV509871H, NGV511841H, NGV510023H, NGV503612H, NGV510564H, NGV511669H, NGV510070H, NGV506017H, NGV506030H, NGV512623H, NGV508015H, NGV501865H, NGV502496H, NGV502504H, NGV506918H, NGV511069H, NGV510862H, NGV510863H, NGV511650H, NGV505392H, NGV507009H, NGV503022H, NGV503555H, NGV504434H, NGV507162H, NGV502897H, NGV504127H, NGV504128H, NGV503188H, NGV507364H, NGV507580H, NGV507612H, NGV508690H, NGV509292H, NGV504833H, NGV506929H, NGV507348H, NGV505454H, NGV507154H, NGV506959H, NGV512398H, NGV511883H, NGV503583H, NGV505282H, NGV508961H, NGV505512H, NGV507998H, NGV501820H, NGV502656H, NGV508229H, NGV510949H, NGV510681H, NGV502628H, NGV506140H, NGV511470H, NGV504187H, NGV509195H, NGV506849H, NGV505289H, NGV505736H, NGV507583H, NGV511854H, NGV504772H, NGV501650H, NGV502968H, NGV503121H, NGV504212H, NGV504685H, NGV506191H, NGV507581H, NGV512170H, NGV513279H, NGV508774H, NGV508859H, NGV509531H, NGV509532H, NGV511832H, NGV510620H, NGV511113H, NGV510435H, NGV505773H, NGV510930H, NGV512257H, NGV503712H, NGV508194H, NGV503911H, NGV505391H, NGV510083H, NGV506748H, NGV509944H, NGV507425H, NGV501049H, NGV505017H, NGV508681H, NGV506100H, NGV507036H, NGV510812H, NGV511913H, NGV512721H, NGV501646H, NGV507576H, NGV504122H, NGV511153H, NGV504374H, NGV503668H, NGV508777H, NGV509045H, NGV501098H, NGV502491H, NGV504599H, NGV507268H, NGV508779H, NGV509031H, NGV509133H, NGV510606H, NGV512549H, NGV513271H, NGV505752H, NGV505843H, NGV501039H, NGV506822H, NGV511634H, NGV506530H, NGV508501H, NGV509130H, NGV510523H, NGV511040H, NGV500374H, NGV502976H, NGV506672H, NGV507096H, NGV504528H, NGV506201H, NGV509960H, NGV504884H, NGV510242H, NGV512600H, NGV500922H, NGV510856H, NGV504028H, NGV504632H, NGV505203H, NGV505952H, NGV509394H, NGV512308H, NGV503591H, NGV504915H, NGV510538H, NGV502726H, NGV504496H, NGV506938H, NGV506940H, NGV508915H, NGV508916H, NGV508920H, NGV510219H, NGV510239H, NGV510248H, NGV502585H, NGV505217H, NGV508891H, NGV501974H, NGV507623H, NGV510642H, NGV502394H, NGV502449H, NGV502451H, NGV503649H, NGV503905H, NGV504029H, NGV504057H, NGV504074H, NGV504132H, NGV504160H, NGV504164H, NGV504165H, NGV504172H, NGV504475H, NGV504647H, NGV504649H, NGV504988H, NGV505931H, NGV506512H, NGV506802H, NGV506811H, NGV506920H, NGV507376H, NGV507676H, NGV508270H, NGV508283H, NGV508429H, NGV508721H, NGV509466H, NGV509876H, NGV510188H, NGV510194H, NGV510321H, NGV510327H, NGV511563H, NGV511893H, NGV512357H, NGV512540H, NGV512741H, NGV503725H, NGV504589H, NGV504917H, NGV506813H, NGV512011H, NGV503859H, NGV503229H, NGV504823H, NGV505863H, NGV509148H, NGV509162H, NGV503372H, NGV504263H, NGV505485H, NGV507394H, NGV508543H, NGV512436H, NGV507535H, NGV508328H, NGV508878H, NGV502288H, NGV509766H, NGV506136H, NGV509467H, NGV511710H, NGV512448H, NGV502588H, NGV509530H, NGV511531H, NGV511537H, NGV511895H, NGV512197H, NGV506902H, NGV510341H, NGV511605H, NGV502243H, NGV503608H, NGV507631H, NGV510354H, NGV502123H, NGV504791H, NGV512664H, NGV504761H, NGV506925H, NGV507750H, NGV509771H, NGV510572H, NGV507179H, NGV508079H, NGV512917H, NGV509578H, NGV509371H, NGV504798H, NGV504770H, NGV503111H, NGV504841H, NGV504905H, NGV504967H, NGV508103H, NGV504898H, NGV509826H, NGV501046H, NGV503760H, NGV506684H, NGV507356H, NGV510595H, NGV503588H, NGV504553H, NGV504811H, NGV502251H, NGV502839H, NGV504334H, NGV504336H, NGV508368H, NGV509452H, NGV509623H, NGV510729H, NGV512528H, NGV505597H, NGV506310H, NGV503071H, NGV504430H, NGV508738H, NGV503365H, NGV504306H, NGV506298H, NGV507872H, NGV512195H, NGV502794H, NGV503047H, NGV504005H, NGV506972H, NGV504674H, NGV506283H, NGV501100H, NGV502594H, NGV502208H, NGV503065H, NGV510524H, NGV502109H, NGV505699H, NGV509377H, NGV512341H, NGV502990H, NGV500348H, NGV501817H, NGV502073H, NGV507899H, NGV504681H, NGV512246H, NGV504568H, NGV511694H, NGV503966H, NGV506485H, NGV508163H, NGV509511H, NGV510485H, NGV512624H, NGV503718H, NGV507337H, NGV505327H, NGV510452H, NGV504969H, NGV505318H, NGV501052H, NGV504853H, NGV507240H, NGV504665H, NGV507832H, NGV511822H, NGV510393H, NGV507426H, NGV508555H, NGV508447H, NGV507274H, NGV503395H, NGV503400H, NGV503833H, NGV504038H, NGV504794H, NGV505679H, NGV507813H, NGV504355H, NGV505823H, NGV505932H, NGV502326H, NGV503765H, NGV503317H, NGV506309H, NGV510858H, NGV504008H, NGV500939H, NGV503293H, NGV503904H, NGV505264H, NGV503346H, NGV504222H, NGV504655H, NGV504656H, NGV506699H, NGV506809H, NGV509568H, NGV502085H, NGV502625H, NGV503392H, NGV507067H, NGV507861H, NGV509221H, NGV509224H, NGV509245H, NGV509666H, NGV509670H, NGV509949H, NGV505783H, NGV509375H, NGV503118H, NGV502747H, NGV505257H, NGV509652H, NGV510177H, NGV505194H, NGV502569H, NGV511943H, NGV508955H, NGV507906H, NGV505978H, NGV507574H, NGV505626H, NGV509272H, NGV502055H, NGV505126H, NGV507687H, NGV502728H, NGV504837H, NGV509816H, NGV510742H, NGV501979H, NGV502223H, NGV502595H, NGV503042H, NGV504111H, NGV504113H, NGV504152H, NGV511759H, NGV511900H, NGV504597H, NGV504818H, NGV507093H, NGV510109H, NGV501035H, NGV507504H, NGV508433H, NGV506756H, NGV506910H, NGV507655H, NGV508945H, NGV502938H, NGV504706H, NGV504708H, NGV509838H, NGV502225H, NGV502574H, NGV506919H, NGV509440H, NGV510155H, NGV511860H, NGV500945H, NGV501150H, NGV505743H, NGV507357H, NGV505119H, NGV502996H, NGV510607H, NGV501013H, NGV509458H, NGV505285H, NGV502937H, NGV508240H, NGV508481H, NGV509333H, NGV510541H, NGV512561H, NGV507410H, NGV511600H, NGV507708H, NGV512816H, NGV503103H, NGV508697H, NGV512168H, NGV509543H, NGV509544H, NGV510464H, NGV511578H, NGV506516H, NGV508313H, NGV506858H, NGV505254H, NGV512287H, NGV503566H, NGV504737H, NGV507331H, NGV506212H, NGV510469H, NGV510696H, NGV502934H, NGV508707H, NGV501871H, NGV502776H, NGV503125H, NGV506087H, NGV507842H, NGV507864H, NGV511709H, NGV509011H, NGV501060H, NGV507748H, NGV506157H, NGV512601H, NGV506109H, NGV510443H, NGV510804H, NGV504380H, NGV505785H, NGV510170H, NGV510409H, NGV512136H, NGV512247H, NGV505956H, NGV509792H, NGV503300H, NGV505035H, NGV510761H, NGV511165H, NGV504795H, NGV510623H, NGV502609H, NGV510131H, NGV502983H, NGV507738H, NGV509428H, NGV509522H, NGV511152H, NGV502114H, NGV502248H, NGV503340H, NGV511461H, NGV501026H, NGV503628H, NGV503648H, NGV504119H, NGV504525H, NGV507743H, NGV509002H, NGV509093H, NGV504727H, NGV507468H, NGV511097H, NGV501320H, NGV503301H, NGV505224H, NGV507038H, NGV510316H, NGV505864H, NGV507389H, NGV511673H, NGV512584H, NGV512698H, NGV512770H, NGV506741H, NGV503738H, NGV504070H, NGV506737H, NGV510517H, NGV511670H, NGV511859H, NGV503936H, NGV509080H, NGV509507H, NGV509779H, NGV505132H, NGV502633H, NGV502663H, NGV505495H, NGV509865H, NGV511575H, NGV512692H, NGV504250H, NGV504253H, NGV505334H, NGV505621H, NGV507439H, NGV510559H, NGV512087H, NGV513172H, NGV501991H, NGV501992H, NGV504299H, NGV508215H, NGV508335H, NGV510377H, NGV511787H, NGV510946H, NGV502136H, NGV504148H, NGV510470H, NGV511821H, NGV502743H, NGV504971H, NGV512526H, NGV512385H, NGV503857H, NGV504329H, NGV508291H, NGV508750H, NGV508766H, NGV511484H, NGV508370H, NGV501027H, NGV501054H, NGV506066H, NGV507253H, NGV511141H, NGV512184H, NGV510266H, NGV512962H, NGV512607H, NGV509509H, NGV503088H, NGV506166H, NGV506957H, NGV510593H, NGV507275H, NGV503599H, NGV505738H, NGV502434H, NGV504940H, NGV510343H, NGV502294H, NGV503307H, NGV503309H, NGV506116H, NGV507157H, NGV510868H, NGV503030H, NGV503204H, NGV510445H, NGV502607H, NGV506962H, NGV507339H, NGV507345H, NGV501933H, NGV506112H, NGV507291H, NGV508273H, NGV509464H, NGV510011H, NGV504044H, NGV503447H, NGV502349H, NGV504061H, NGV507315H, NGV508870H, NGV502828H, NGV506001H, NGV502396H, NGV507761H, NGV506627H, NGV505989H, NGV504293H, NGV504014H, NGV506851H, NGV509846H, NGV504712H, NGV505172H, NGV506763H, NGV507171H, NGV509351H, NGV503482H, NGV507986H, NGV511750H, NGV510203H, NGV511130H, NGV511553H, NGV512072H, NGV504821H, NGV504958H, NGV506170H, NGV506640H, NGV507710H, NGV509524H, NGV512488H, NGV503026H, NGV503502H, NGV509003H, NGV510547H, NGV507122H, NGV502212H, NGV501858H, NGV504033H, NGV506120H, NGV506219H, NGV506568H, NGV507297H, NGV507299H, NGV508206H, NGV510255H, NGV512851H, NGV513227H, NGV500382H, NGV503784H, NGV507699H, NGV504362H, NGV508463H, NGV509040H, NGV509294H, NGV510350H, NGV510834H, NGV502501H, NGV511464H, NGV506308H, NGV501040H, NGV512839H, NGV502448H, NGV506069H, NGV510576H, NGV508246H, NGV509801H, NGV504803H, NGV503694H, NGV506198H, NGV511802H, NGV511101H, NGV502437H, NGV503353H, NGV505846H, NGV511956H, NGV505881H, NGV502122H, NGV509042H, NGV512658H, NGV507324H, NGV504960H, NGV504962H, NGV506070H, NGV502754H, NGV509229H, NGV503168H, NGV504640H, NGV505969H, NGV505976H, NGV507836H, NGV508276H, NGV510957H, NGV510293H, NGV509033H, NGV505949H, NGV508118H, NGV506547H, NGV506299H, NGV511059H, NGV505583H, NGV506008H, NGV508107H, NGV509044H, NGV502991H, NGV503664H, NGV507637H, NGV506799H, NGV502277H, NGV505259H, NGV500980H, NGV506780H, NGV512639H, NGV506072H, NGV506200H, NGV512641H, NGV504497H, NGV502626H, NGV504239H, NGV503625H, NGV507651H, NGV501900H, NGV502178H, NGV501836H, NGV506691H, NGV510222H, NGV509619H, NGV508899H, NGV510143H, NGV504151H, NGV502335H, NGV508732H, NGV512543H, NGV502718H, NGV511108H, NGV502188H, NGV505284H, NGV512131H, NGV504449H, NGV502915H, NGV502916H, NGV509273H, NGV508090H, NGV501020H, NGV501099H, NGV504054H, NGV509015H, NGV506491H, NGV512683H, NGV510033H, NGV502975H, NGV510189H, NGV504304H, NGV504399H, NGV503063H, NGV506215H, NGV506216H, NGV508073H, NGV501083H, NGV504699H, NGV504734H, NGV504227H, NGV504232H, NGV502269H, NGV504627H, NGV512093H, NGV504351H, NGV504530H, NGV507056H, NGV510767H, NGV503691H, NGV508939H, NGV512969H; GTIN 00643169345218, Serial Numbers: NGV502310H, NGV503261H, NGV512927H, NGV501842H, NGV503108H, NGV502325H, NGV503314H, NGV503794H, NGV503795H, NGV503796H, NGV503266H, NGV506020H, NGV509254H, NGV503149H, NGV512049H, NGV512290H, NGV512511H, NGV512890H, NGV502315H, NGV510144H, NGV512906H, NGV503151H, NGV503747H, NGV508188H, NGV503733H, NGV505517H, NGV505867H, NGV508004H, NGV508011H, NGV508022H, NGV508205H, NGV508255H, NGV509209H, NGV509226H, NGV509688H, NGV509970H, NGV512884H, NGV503150H, NGV504463H, NGV500935H, NGV505087H, NGV505162H, NGV505163H, NGV509238H, NGV509575H, NGV509817H, NGV510140H, NGV511496H, NGV501906H, NGV506490H, NGV507930H, NGV512897H, NGV505543H, NGV506470H, NGV502120H, NGV508488H, NGV509156H, NGV511511H, NGV513179H, NGV503720H, NGV503740H, NGV505505H, NGV505521H, NGV505566H, NGV506572H, NGV506683H, NGV508512H, NGV503827H, NGV509790H, NGV512928H, NGV511819H, NGV505649H, NGV510653H, NGV512875H, NGV509030H, NGV512506H, NGV512803H, NGV512856H, NGV502836H, NGV502873H, NGV503356H, NGV512086H, NGV503133H, NGV512847H, NGV506543H, NGV500962H, NGV503876H, NGV505814H, NGV503799H, NGV505498H, NGV505540H, NGV509438H, NGV510019H, NGV510022H, NGV510181H, NGV510754H, NGV510912H, NGV512411H, NGV506638H, NGV507910H, NGV508017H, NGV509621H, NGV512887H, NGV513095H, NGV500360H, NGV501513H, NGV501608H, NGV502359H, NGV502369H, NGV503128H, NGV503723H, NGV503729H, NGV503754H, NGV505084H, NGV505547H, NGV505574H, NGV505581H, NGV505584H, NGV505661H, NGV505967H, NGV505994H, NGV506003H, NGV506280H, NGV506462H, NGV506646H, NGV507375H, NGV507432H, NGV507855H, NGV507931H, NGV507933H, NGV507948H, NGV508436H, NGV509164H, NGV509241H, NGV509304H, NGV509625H, NGV509992H, NGV510691H, NGV510762H, NGV511931H, NGV511933H, NGV511985H, NGV512507H, NGV512572H, NGV512773H, NGV512899H, NGV512934H, NGV512941H, NGV513175H, NGV503182H, NGV503191H, NGV503894H, NGV504040H, NGV505534H, NGV505537H, NGV505669H, NGV508050H, NGV510032H, NGV510711H, NGV512008H, NGV513207H, NGV511952H, NGV503852H, NGV507892H, NGV508456H, NGV510890H, NGV512754H, NGV502405H, NGV503866H, NGV503922H, NGV503923H, NGV506676H, NGV507457H, NGV508346H, NGV508467H, NGV510947H, NGV511534H, NGV511679H, NGV511986H, NGV512563H, NGV503825H, NGV503954H, NGV507983H, NGV512512H, NGV512541H, NGV501902H, NGV509633H, NGV510790H, NGV510919H, NGV504428H, NGV504472H, NGV504473H, NGV505587H, NGV513178H, NGV501822H, NGV502881H, NGV503269H, NGV503976H, NGV505556H, NGV505588H, NGV509125H, NGV511966H, NGV512238H, NGV503929H, NGV503931H, NGV504459H, NGV506644H, NGV500963H, NGV500969H, NGV504080H, NGV505184H, NGV505569H, NGV506237H, NGV506403H, NGV507888H, NGV507934H, NGV507965H, NGV509168H, NGV509205H, NGV509308H, NGV509593H, NGV509648H, NGV509661H, NGV511642H, NGV511682H, NGV512224H, NGV512870H, NGV513174H, NGV507951H, NGV509198H, NGV502742H, NGV502762H, NGV503160H, NGV505091H, NGV509227H, NGV501908H, NGV502876H, NGV505859H, NGV507954H, NGV511663H, NGV512018H, NGV512859H, NGV505159H, NGV509231H, NGV502848H, NGV502857H, NGV505665H, NGV505803H, NGV505811H, NGV508028H, NGV510996H, NGV512043H, NGV513116H, NGV502113H, NGV505417H, NGV510042H, NGV512020H, NGV505129H, NGV509983H, NGV507398H, NGV508027H, NGV508963H, NGV511936H, NGV512483H, NGV512489H, NGV513119H, NGV505178H, NGV507928H, NGV513133H, NGV500942H, NGV503147H, NGV505141H, NGV505145H, NGV506605H, NGV507919H, NGV508018H, NGV509596H, NGV510446H, NGV511865H, NGV512399H, NGV512830H, NGV513194H, NGV502803H, NGV510465H, NGV507909H, NGV501821H, NGV507163H, NGV503674H, NGV505561H, NGV510760H, NGV501895H, NGV502013H, NGV502401H, NGV503143H, NGV512881H, NGV505499H, NGV505996H, NGV505101H, NGV505942H, NGV505946H, NGV511734H, NGV503655H, NGV502025H, NGV502033H, NGV503994H, NGV505541H, NGV505557H, NGV506492H, NGV506593H, NGV503237H, NGV505678H, NGV503797H, NGV510090H, NGV511886H, NGV512589H, NGV503156H, NGV501899H, NGV506434H, NGV510884H, NGV505525H, NGV502766H, NGV505104H, NGV509588H, NGV512815H, NGV512817H, NGV512968H, NGV503239H, NGV503246H, NGV503998H, NGV507801H, NGV512034H, NGV512035H, NGV512213H, NGV512223H, NGV513204H, NGV513206H, NGV512810H, NGV503324H, NGV503328H, NGV507880H, NGV511687H, NGV512067H, NGV512068H, NGV512076H, NGV512902H, NGV506422H, NGV509102H, NGV511599H, NGV511811H, NGV503822H, NGV511680H, NGV507937H, NGV507972H, NGV505067H, NGV504935H, NGV502713H, NGV502832H, NGV508013H, NGV510004H, NGV512231H, NGV512910H, NGV513182H, NGV502831H, NGV507047H, NGV505559H, NGV502869H, NGV505120H, NGV501830H, NGV503736H, NGV503746H, NGV511608H, NGV506427H, NGV503517H, NGV503562H, NGV511818H, NGV505011H, NGV505092H, NGV505097H, NGV508476H, NGV508482H, NGV512053H, NGV505054H, NGV505057H, NGV510048H, NGV503778H, NGV511739H, NGV507953H, NGV502347H, NGV502757H, NGV502871H, NGV506575H, NGV507977H, NGV503490H, NGV505683H, NGV507442H, NGV507942H, NGV513196H, NGV503661H, NGV503666H, NGV503609H, NGV503690H, NGV505063H, NGV509676H, NGV509753H, NGV509972H, NGV500928H, NGV501897H, NGV502385H, NGV503255H, NGV505133H, NGV505536H, NGV505568H, NGV505571H, NGV505964H, NGV509165H, NGV509559H, NGV509572H, NGV510297H, NGV511727H, NGV512296H, NGV512711H, NGV509177H, NGV501559H, NGV505082H, NGV505623H, NGV506760H, NGV510680H, NGV507999H, NGV502403H, NGV512063H, NGV505551H, NGV503646H, NGV503992H, NGV504406H, NGV503763H, NGV503771H, NGV505018H, NGV505029H, NGV505031H, NGV505032H, NGV505034H, NGV505036H, NGV504411H, NGV504453H, NGV506504H, NGV506527H, NGV507924H, NGV508054H, NGV509183H, NGV506601H, NGV506639H, NGV509982H, NGV511517H, NGV501880H, NGV502035H, NGV502053H, NGV502834H, NGV504485H, NGV504610H, NGV505023H, NGV507024H, NGV510007H, NGV505009H, NGV502063H, NGV504444H, NGV503717H, NGV506414H, NGV508081H, NGV512733H, NGV503980H, NGV509178H, NGV512573H, NGV504458H, NGV510819H, NGV511055H, NGV511056H, NGV503214H, NGV503218H, NGV503975H, NGV505004H, NGV505586H, NGV506590H, NGV502244H, NGV502711H, NGV511827H, NGV505924H, NGV508303H, NGV508714H, NGV509981H, NGV510224H, NGV510676H, NGV511748H, NGV511984H, NGV502721H, NGV502736H, NGV502778H, NGV502786H, NGV502792H, NGV502805H, NGV502821H, NGV503696H, NGV503804H, NGV503812H, NGV503814H, NGV503890H, NGV505527H, NGV505558H, NGV507917H, NGV507925H, NGV509240H, NGV509427H, NGV511996H, NGV512479H, NGV512480H, NGV513186H, NGV513190H, NGV509208H, NGV502057H, NGV502795H, NGV504634H, NGV507980H, NGV512268H, NGV512272H, NGV502064H, NGV502072H, NGV503757H, NGV503993H, NGV504000H, NGV504001H, NGV505039H, NGV505183H, NGV507952H, NGV512205H, NGV507387H, NGV501879H, NGV505040H, NGV503820H, NGV501881H, NGV502376H, NGV507788H, NGV505481H, NGV505689H, NGV505929H, NGV505865H, NGV506245H, NGV512893H, NGV505570H, NGV506465H, NGV512892H, NGV501873H, NGV503272H, NGV510056H, NGV500975H, NGV504474H, NGV504480H, NGV505190H, NGV503817H, NGV505196H, NGV509993H, NGV510000H, NGV510873H, NGV512029H, NGV512030H, NGV512948H, NGV502825H, NGV506667H, NGV508076H, NGV513215H, NGV505360H, NGV506668H, NGV505685H, NGV507946H, NGV502060H, NGV508000H, NGV509193H, NGV509255H, NGV502352H, NGV507908H, NGV510008H, NGV507379H, NGV502374H, NGV503791H, NGV506603H, NGV512972H, NGV505562H, NGV510028H, NGV513198H, NGV507936H, NGV512726H, NGV510125H, NGV505925H, NGV512823H, NGV512814H, NGV502215H, NGV502337H, NGV504431H, NGV505526H, NGV505634H, NGV506419H, NGV510709H, NGV511518H, NGV511659H, NGV511660H, NGV512835H, NGV504456H, NGV504457H, NGV506458H, NGV503782H, NGV510015H, NGV512746H, NGV503106H, NGV505001H, NGV505560H, NGV510054H, NGV511035H, NGV512858H, NGV512227H, NGV506655H, NGV511909H, NGV505116H, NGV502708H, NGV503316H, NGV508365H, NGV509597H, NGV511839H, NGV508723H, NGV507969H, NGV505883H, NGV507883H, NGV502773H, NGV505688H, NGV506645H, NGV507378H, NGV508129H, NGV509653H, NGV512868H, NGV513213H, NGV502808H, NGV503275H, NGV503276H, NGV507435H, NGV511993H, NGV507913H, NGV503181H, NGV509223H, NGV510342H, NGV512232H, NGV512916H, NGV509918H, NGV510717H, NGV505595H, NGV510577H, NGV505579H, NGV505854H, NGV505928H, NGV508444H, NGV510367H, NGV511858H, NGV511907H, NGV512801H, NGV508110H, NGV512447H, NGV503142H, NGV505137H, NGV506477H, NGV509010H, NGV501846H, NGV500995H, NGV500921H, NGV502311H, NGV503258H, NGV503265H, NGV503789H, NGV505050H, NGV505062H, NGV505951H, NGV510147H, NGV510888H, NGV510025H, NGV502864H, NGV503291H, NGV505985H, NGV509191H, NGV509353H, NGV512376H, NGV504977H, NGV505530H, NGV506405H, NGV506408H, NGV511810H, NGV508459H, NGV508460H, NGV508461H, NGV512500H, NGV500511H, NGV502353H, NGV506479H, NGV507939H, NGV508435H, NGV500996H, NGV501868H, NGV503131H, NGV503154H, NGV505529H, NGV508505H, NGV509352H, NGV511459H, NGV511889H, NGV511894H, NGV505086H, NGV508419H, NGV500919H, NGV508493H, NGV512715H, NGV512727H, NGV509433H, NGV511937H, NGV511834H, NGV512595H, NGV505074H, NGV510047H, NGV500361H, NGV508061H, NGV505548H, NGV511653H, NGV512879H, NGV510115H, NGV512828H, NGV505048H, NGV506614H, NGV509179H, NGV509781H, NGV512796H, NGV502129H, NGV503100H, NGV506623H, NGV509058H, NGV508356H, NGV510030H, NGV501833H, NGV502183H, NGV502837H, NGV503699H, NGV507879H, NGV512531H, NGV503734H, NGV505051H, NGV505650H, NGV509180H, NGV505589H, NGV505591H, NGV508489H, NGV510542H, NGV511697H, NGV503383H, NGV503396H, NGV503420H, NGV503800H, NGV503969H, NGV503970H, NGV504420H, NGV505064H, NGV505085H, NGV505131H, NGV505506H, NGV505520H, NGV505563H, NGV505663H, NGV506633H, NGV507396H, NGV507955H, NGV507956H, NGV509556H, NGV509731H, NGV509969H, NGV510043H, NGV510433H, NGV510741H, NGV510744H, NGV510750H, NGV512778H, NGV512841H, NGV512842H, NGV513083H, NGV513128H, NGV513192H, NGV506423H, NGV508072H, NGV509974H, NGV503165H, NGV505509H, NGV507403H, NGV505518H, NGV505519H, NGV508065H, NGV503134H, NGV503793H, NGV505049H, NGV505836H, NGV512880H, NGV506609H, NGV501835H, NGV502012H, NGV511674H, NGV508720H, NGV503707H, NGV510091H, NGV505565H, NGV508025H, NGV510392H, NGV510423H, NGV512044H, NGV502235H, NGV509429H; GTIN 00643169345225, Serial Numbers: NGV505552H, NGV503213H, NGV505677H, NGV507449H, NGV507450H, NGV509444H, NGV504407H, NGV505179H, NGV505660H, NGV505891H, NGV506448H, NGV507065H, NGV507372H, NGV507918H, NGV507941H, NGV507964H, NGV508071H, NGV510789H, NGV510935H, NGV511866H, NGV512710H, NGV503755H, NGV503776H, NGV505686H, NGV501512H, NGV502393H, NGV503671H, NGV505100H, NGV505110H, NGV506999H, NGV507022H, NGV507920H, NGV508327H, NGV508475H, NGV510786H, NGV510944H, NGV511869H, NGV503818H, NGV505299H, NGV505448H, NGV505538H, NGV505555H, NGV505687H, NGV506022H, NGV506979H, NGV507440H, NGV509194H, NGV509214H, NGV509236H, NGV510156H, NGV510229H, NGV510939H, NGV511613H, NGV511705H, NGV511916H, NGV511921H, NGV512005H, NGV512062H, NGV512527H, NGV512748H, NGV512780H, NGV512878H, NGV512896H, NGV503865H, NGV504409H, NGV505674H, NGV505676H, NGV509639H, NGV511923H, NGV512728H, NGV512779H, NGV507932H, NGV512949H, NGV503991H, NGV508772H, NGV509577H, NGV512477H, NGV503730H, NGV504986H, NGV506016H, NGV510037H, NGV505155H, NGV512877H, NGV503322H, NGV508434H, NGV509419H, NGV510976H, NGV511587H, NGV511941H, NGV502196H, NGV502312H, NGV503801H, NGV503802H, NGV503870H, NGV505103H, NGV505670H, NGV505675H, NGV505695H, NGV506402H, NGV506406H, NGV506463H, NGV507904H, NGV507979H, NGV508030H, NGV508727H, NGV509153H, NGV509612H, NGV511032H, NGV511590H, NGV512747H, NGV512799H, NGV512805H, NGV512882H, NGV512777H, NGV505147H, NGV505297H, NGV507408H, NGV511507H, NGV511716H, NGV507940H, NGV509649H, NGV510086H, NGV508349H, NGV505169H, NGV512449H, NGV512798H, NGV503101H, NGV503680H, NGV510781H, NGV505624H, NGV507991H, NGV509605H, NGV510046H, NGV502384H, NGV502731H, NGV502858H, NGV502872H, NGV503144H, NGV503162H, NGV503197H, NGV503809H, NGV503815H, NGV505003H, NGV505024H, NGV505078H, NGV505130H, NGV505158H, NGV505170H, NGV505171H, NGV505233H, NGV505244H, NGV505593H, NGV505635H, NGV505640H, NGV505930H, NGV505970H, NGV507392H, NGV507889H, NGV507897H, NGV507970H, NGV507976H, NGV508049H, NGV508051H, NGV509022H, NGV509994H, NGV509996H, NGV510065H, NGV510088H, NGV511502H, NGV511970H, NGV512414H, NGV501894H, NGV509135H, NGV511840H, NGV503614H, NGV512096H, NGV503790H, NGV505549H, NGV506478H, NGV506661H, NGV512905H, NGV502379H, NGV508007H, NGV509363H, NGV507927H, NGV501601H, NGV502868H, NGV503926H, NGV503977H, NGV504002H, NGV505671H, NGV505673H, NGV508145H, NGV511039H, NGV512583H, NGV512594H, NGV512774H, NGV503687H, NGV505152H, NGV505646H, NGV507893H, NGV509579H, NGV512422H, NGV503235H, NGV503888H, NGV500354H, NGV503889H, NGV503924H, NGV503986H, NGV505901H, NGV508288H, NGV509219H, NGV505652H, NGV505874H, NGV506657H, NGV509143H, NGV512144H, NGV512697H, NGV512900H, NGV511604H, NGV512926H, NGV512935H, NGV506426H, NGV502375H, NGV503176H, NGV503193H, NGV503728H, NGV507437H, NGV505582H, NGV507868H, NGV508765H, NGV512013H, NGV512966H, NGV503703H, NGV505081H, NGV505102H, NGV505176H, NGV505185H, NGV507926H, NGV512172H, NGV503711H, NGV503713H, NGV505542H, NGV505697H, NGV505910H, NGV506432H, NGV506439H, NGV507446H, NGV508138H, NGV512529H, NGV509975H, NGV503126H, NGV505680H; GTIN 00643169345232, Serial Numbers: NGV502366H, NGV512822H, NGV512239H, NGV508344H, NGV508146H, NGV508225H, NGV510871H, NGV512240H, NGV502372H, NGV506981H, NGV507031H, NGV502637H, NGV507882H, NGV502321H, NGV506586H, NGV508338H, NGV510017H, NGV510883H, NGV510886H, NGV510892H, NGV510894H, NGV510895H, NGV510910H, NGV512088H, NGV510107H, NGV512001H, NGV502753H, NGV502827H, NGV502856H, NGV507911H, NGV508153H, NGV509230H, NGV509396H, NGV510830H, NGV510959H, NGV511898H, NGV506579H, NGV510932H, NGV501831H, NGV506625H, NGV507053H, NGV508329H, NGV510031H, NGV510137H, NGV510808H, NGV502822H, NGV502875H, NGV506532H, NGV507896H, NGV508376H, NGV508452H, NGV508454H, NGV511935H, NGV502370H, NGV502329H, NGV507248H, NGV508343H, NGV508354H, NGV508731H, NGV502362H, NGV509426H, NGV502397H, NGV502804H, NGV502846H, NGV507957H, NGV508114H, NGV508116H, NGV512807H, NGV512813H, NGV502339H, NGV502854H, NGV506975H, NGV509056H, NGV508699H, NGV500368H, NGV500412H, NGV502800H, NGV508136H, NGV509059H, NGV509109H, NGV509166H, NGV509248H, NGV511975H, NGV508106H, NGV509508H, NGV502878H, NGV503716H, NGV502236H, NGV501890H, NGV503783H, NGV500373H, NGV502318H, NGV503787H, NGV500385H, NGV503798H, NGV502812H; GTIN 00643169384170, Serial Numbers: NGV508547H, NGV508550H, NGV508551H, NGV508556H, NGV508558H, NGV508561H, NGV508958H, NGV508563H, NGV509999H, NGV510044H, NGV508957H, NGV505944H, NGV507204H, NGV506483H, NGV506624H, NGV502763H, NGV506656H, NGV500366H, NGV502400H, NGV506542H, NGV508031H, NGV508295H, NGV506632H, NGV506642H, NGV511901H, NGV508345H, NGV508351H, NGV507043H, NGV507199H, NGV502760H, NGV508340H, NGV502697H, NGV508059H, NGV512417H, NGV511486H, NGV511977H, NGV511978H, NGV511962H, NGV502195H, NGV502798H, NGV500380H, NGV508043H, NGV508069H, NGV506474H, NGV506620H, NGV502734H, NGV502746H, NGV502768H, NGV502777H, NGV502866H, NGV502890H, NGV505905H, NGV505923H, NGV505934H, NGV506468H, NGV506480H, NGV506481H, NGV506596H, NGV508036H, NGV508041H, NGV502578H, NGV508040H, NGV502398H, NGV508371H, NGV500387H, NGV502815H, NGV506466H, NGV506622H, NGV505657H, NGV508297H, NGV505941H, NGV511915H, NGV506584H, NGV506647H, NGV508337H, NGV502826H, NGV505516H, NGV505572H, NGV505653H, NGV505656H, NGV506460H, NGV506486H, NGV506592H, NGV508046H, NGV508047H, NGV508058H, NGV508363H, NGV510934H, NGV510958H, NGV511488H, NGV511998H, NGV512009H, NGV512073H, NGV512165H, NGV512177H, NGV512409H, NGV502358H, NGV508277H, NGV508284H, NGV511974H, NGV508020H, NGV508308H, NGV512404H, NGV502350H, NGV507907H, NGV502336H, NGV508298H, NGV511914H, NGV512024H, NGV506473H, NGV512421H, NGV511958H, NGV506649H, NGV508700H, NGV508701H, NGV508730H, NGV508016H, NGV511991H, NGV512081H, NGV502840H, NGV508364H, NGV508366H, NGV500371H, NGV502818H, NGV506595H, NGV508067H, NGV508068H, NGV508292H, NGV508360H, NGV510941H, NGV505644H, NGV506597H, NGV506598H, NGV502814H, NGV502859H, NGV505580H, NGV502391H, NGV505578H, NGV511904H, NGV502823H, NGV502824H, NGV505651H, NGV508026H, NGV511946H, NGV508300H, NGV508369H, NGV502323H, NGV502355H, NGV508033H, NGV502395H, NGV506546H, NGV502862H, NGV502917H, NGV511982H, NGV502813H, NGV502861H, NGV506626H, NGV511995H, NGV512173H, NGV507380H, NGV510960H, NGV506973H, NGV510911H, NGV505918H, NGV510893H, NGV505936H, NGV506024H, NGV510921H, NGV505914H, NGV508373H, NGV511953H, NGV507448H, NGV507004H, NGV505668H, NGV505672H, NGV510196H, NGV513244H, NGV513252H, NGV509956H, NGV513217H, NGV510844H, NGV506457H, NGV512579H, NGV505641H, NGV511761H, NGV506430H, NGV505658H, NGV506660H, NGV506664H, NGV506665H, NGV506666H, NGV508312H, NGV508315H, NGV508331H, NGV508334H, NGV508994H, NGV508995H, NGV509004H, NGV509028H, NGV509036H, NGV500970H, NGV500971H, NGV500979H, NGV500985H, NGV500998H, NGV501006H, NGV503196H, NGV503216H, NGV503220H, NGV503221H, NGV503222H, NGV503224H, NGV503226H, NGV503236H, NGV503244H, NGV503251H, NGV503257H, NGV503286H, NGV503334H, NGV503374H, NGV503375H, NGV503376H, NGV503403H, NGV503406H, NGV503408H, NGV503422H, NGV503429H, NGV503431H, NGV503435H, NGV503439H, NGV503440H, NGV503449H, NGV503450H, NGV503452H, NGV503454H, NGV503457H, NGV503465H, NGV503467H, NGV503471H, NGV503472H, NGV503474H, NGV503484H, NGV503493H, NGV503495H, NGV507461H, NGV510828H, NGV508132H, NGV506450H, NGV510192H, NGV510247H, NGV510973H, NGV500392H, NGV500930H, NGV500938H, NGV500952H, NGV502332H, NGV502333H, NGV510835H, NGV510839H, NGV500914H, NGV510825H, NGV505567H, NGV506637H, NGV505553H, NGV506631H, NGV506617H, NGV502781H, NGV502883H, NGV508294H, NGV511491H, NGV511928H, NGV511940H, NGV502882H, NGV511957H, NGV511983H, NGV503637H, NGV503640H, NGV503641H, NGV503642H, NGV505095H, NGV503891H, NGV505083H, NGV508374H, NGV505550H, NGV508989H, NGV503829H, NGV508959H, NGV509600H, NGV509698H, NGV503672H, NGV503974H, NGV508096H, NGV508121H, NGV508535H, NGV510897H, NGV505061H, NGV509355H, NGV505089H, NGV503669H, NGV503858H, NGV503912H, NGV505088H, NGV509725H, NGV503882H, NGV507902H, NGV508375H, NGV509726H, NGV503863H, NGV507377H, NGV509331H, NGV503826H, NGV508545H, NGV508896H, NGV503676H, NGV510840H, NGV503678H, NGV503682H, NGV505664H, NGV505981H, NGV505502H, NGV505504H, NGV505632H, NGV508100H, NGV508192H, NGV508764H, NGV511776H, NGV510885H, NGV512505H, NGV508281H, NGV510845H, NGV510879H, NGV503263H, NGV503302H, NGV503604H, NGV503611H, NGV505625H, NGV505904H, NGV505948H, NGV505983H, NGV505991H, NGV507385H, NGV507944H, NGV508123H, NGV508127H, NGV508268H, NGV508525H, NGV510850H, NGV510870H, NGV510875H, NGV510991H, NGV511792H, NGV512212H, NGV505603H.NGV503259H, NGV505638H, NGV505602H, NGV505610H, NGV510855H, NGV507988H, NGV503643H, NGV505066H, NGV505607H, NGV505609H, NGV511569H, NGV512544H, NGV508169H, NGV500403H, NGV505600H, NGV508101H, NGV506600H, NGV510968H, NGV505613H, NGV508158H, NGV508164H, NGV512523H, NGV507452H, NGV502328H, NGV508278H, NGV508279H, NGV510778H, NGV512513H, NGV512548H, NGV512556H, NGV505590H, NGV510822H, NGV507949H, NGV508533H, NGV505606H, NGV505629H, NGV510853H, NGV503635H, NGV503636H, NGV505075H, NGV501096H, NGV503199H, NGV503665H, NGV511020H, NGV500926H, NGV511632H, NGV505021H, NGV505044H, NGV505628H, NGV506447H, NGV507384H, NGV508217H, NGV508532H, NGV508230H, NGV508001H, NGV510878H, NGV511736H, NGV511760H, NGV501094H, NGV502286H, NGV502299H, NGV503644H, NGV503647H, NGV503650H, NGV503651H, NGV503652H, NGV503658H, NGV505353H, NGV505692H, NGV506268H, NGV507397H, NGV507402H, NGV507405H, NGV507406H, NGV507444H, NGV505058H, NGV511562H, NGV511899H, NGV501022H, NGV504636H, NGV503605H, NGV503701H, NGV503779H, NGV509986H, NGV509989H, NGV509998H, NGV510787H, NGV500959H, NGV500964H, NGV508241H, NGV502069H, NGV503624H, NGV503629H, NGV503692H, NGV503745H, NGV503751H, NGV504687H, NGV504999H, NGV505002H, NGV505007H, NGV505013H, NGV505014H, NGV505042H, NGV505043H, NGV506401H, NGV506416H, NGV510532H, NGV510612H, NGV510902H, NGV510906H, NGV510915H, NGV510916H, NGV512730H, NGV512788H, NGV503695H, NGV506577H, NGV509978H, NGV503677H, NGV501882H, NGV502124H, NGV506451H, NGV507261H, NGV507796H, NGV508010H, NGV511622H, NGV505027H, NGV505065H, NGV503606H, NGV505696H, NGV506428H, NGV506455H, NGV503785H, NGV503788H, NGV507891H, NGV508006H, NGV500390H, NGV500927H, NGV512804H, NGV512904H, NGV512930H, NGV505507H, NGV508168H, NGV503627H, NGV505019H, NGV507267H, NGV508085H, NGV501856H; GTIN 00643169384187, Serial Numbers: NGV502885H, NGV508497H, NGV509409H, NGV503466H, NGV506528H, NGV506615H, NGV506630H, NGV506641H, NGV511846H, NGV512070H, NGV509573H, NGV502838H, NGV509175H, NGV511831H, NGV502407H, NGV500936H, NGV502037H, NGV502884H, NGV508448H, NGV509096H, NGV509103H, NGV509114H, NGV509134H, NGV509184H, NGV509215H, NGV509220H, NGV509244H, NGV509403H, NGV509203H, NGV500947H, NGV502888H, NGV503243H, NGV503245H, NGV503254H, NGV509583H, NGV510369H, NGV511850H, NGV500923H, NGV501838H, NGV500973H, NGV503279H, NGV503366H, NGV503404H, NGV503462H, NGV500355H, NGV502324H, NGV503489H, NGV503256H; GTIN 00643169384194, Serial Numbers: NGV500966H, NGV503416H, NGV503423H, NGV508080H, NGV508176H, NGV508282H, NGV508342H, NGV509558H, NGV509599H, NGV509634H, NGV509788H, NGV510077H, NGV511483H, NGV511494H, NGV511500H, NGV511508H, NGV511554H, NGV511862H, NGV511867H, NGV511892H, NGV511906H, NGV508082H, NGV508094H, NGV508102H, NGV508109H, NGV508120H, NGV509604H, NGV509608H, NGV509609H, NGV509637H, NGV510152H, NGV510880H, NGV510953H, NGV510962H, NGV503411H, NGV503469H, NGV503473H, NGV507366H, NGV507984H, NGV508060H, NGV508147H, NGV508155H, NGV508427H, NGV509455H, NGV510114H, NGV511476H, NGV511934H, NGV511976H, NGV511980H, NGV511997H, NGV512036H, NGV512045H, NGV502031H, NGV509398H, NGV509401H, NGV509413H, NGV511844H, NGV512010H, NGV500981H, NGV501003H, NGV502052H, NGV503426H, NGV503442H, NGV507465H, NGV508088H, NGV508092H, NGV508492H, NGV509665H, NGV511487H, NGV511845H, NGV511972H, NGV503432H, NGV500957H, NGV503437H, NGV502049H, NGV503459H, NGV508086H, NGV502851H; GTIN 00643169384200, Serial Numbers: NGV507887H, NGV502020H, NGV502022H, NGV502373H, NGV503107H, NGV503135H, NGV503137H, NGV503138H, NGV503152H, NGV503161H, NGV503189H, NGV503190H, NGV503212H, NGV503238H, NGV505402H, NGV505420H, NGV505431H, NGV505435H, NGV505442H, NGV505445H, NGV505458H, NGV505459H, NGV505472H, NGV505488H, NGV505492H, NGV505681H, NGV506573H, NGV507454H, NGV507950H, NGV508339H, NGV509615H, NGV502024H, NGV502043H, NGV502047H, NGV502059H, NGV502068H, NGV503419H, NGV503448H, NGV503455H, NGV503477H, NGV503481H, NGV506231H, NGV506238H, NGV506411H, NGV506412H, NGV506431H, NGV506576H, NGV506578H, NGV506613H; GTIN 00643169384217, Serial Numbers: NGV503438H, NGV503623H, NGV503887H, NGV506998H, NGV501857H, NGV503908H, NGV507016H, NGV507019H, NGV503675H, NGV505980H, NGV505997H, NGV501073H, NGV500940H, NGV501068H, NGV501872H, NGV502737H, NGV505963H, NGV507063H, NGV503697H, NGV503753H, NGV504639H, NGV505037H, NGV505889H, NGV502632H, NGV502790H, NGV502810H, NGV502853H, NGV505096H, NGV505117H, NGV505642H, NGV507000H; GTIN 00643169384224, Serial Numbers: NGV510775H, NGV506257H, NGV506223H, NGV506282H, NGV506410H, NGV506449H, NGV506472H, NGV506487H, NGV506494H, NGV506602H, NGV506604H, NGV506607H, NGV506611H, NGV506658H, NGV503653H, NGV505903H, NGV508309H, NGV508326H, NGV509570H, NGV509618H, NGV509646H, NGV511917H, NGV511932H, NGV511971H; GTIN 00643169384231, Serial Numbers: NGV503424H, NGV500941H, NGV509991H, NGV506634H, NGV507961H, NGV507974H, NGV503925H, NGV503626H, NGV505655H, NGV512793H, NGV508432H, NGV507975H, NGV511658H, NGV512321H, NGV501907H, NGV502179H, NGV502186H, NGV502357H, NGV505637H, NGV506663H, NGV507994H, NGV506508H, NGV508287H, NGV510928H, NGV505022H, NGV506569H, NGV502334H, NGV506498H, NGV512458H, NGV503332H, NGV505041H, NGV505694H, NGV506413H, NGV506429H, NGV508002H, NGV510904H, NGV511793H, NGV512194H, NGV512434H, NGV510882H, NGV509210H, NGV510241H, NGV512117H, NGV503496H, NGV508293H, NGV506464H, NGV506574H, NGV511561H, NGV507443H, NGV503305H, NGV503323H, NGV502363H, NGV505647H, NGV511053H, NGV512156H, NGV503367H, NGV501878H, NGV503497H, NGV505627H, NGV512097H, NGV511944H, NGV505234H, NGV505682H; GTIN 00643169508156, Serial Numbers: NGV517372H, NGV517373H, NGV524369H, NGV524370H, NGV516484H, NGV516485H, NGV525786H, NGV527877H, NGV529697H, NGV513645H, NGV513646H, NGV513827H, NGV527375H, NGV515441H, NGV513647H, NGV513648H, NGV528834H, NGV513992H, NGV515196H, NGV526840H, NGV528259H, NGV529698H, NGV530102H, NGV518785H, NGV513993H, NGV515475H, NGV515155H, NGV515197H, NGV515324H, NGV515476H, NGV518499H, NGV524448H, NGV526838H, NGV528260H, NGV529966H, NGV550784H, NGV551825H, NGV524854H, NGV529055H, NGV513385H, NGV516123H, NGV529281H, NGV513309H, NGV523005H, NGV526087H, NGV530743H, NGV523749H, NGV526301H, NGV529443H, NGV551994H, NGV513944H, NGV516204H, NGV516397H, NGV516902H, NGV516912H, NGV517204H, NGV517427H, NGV518021H, NGV518348H, NGV518702H, NGV520382H, NGV520730H, NGV522512H, NGV524199H, NGV524200H, NGV524204H, NGV524205H, NGV524741H, NGV526101H, NGV526292H, NGV527379H, NGV528271H, NGV528598H, NGV528603H, NGV529183H, NGV529428H, NGV530137H, NGV530145H, NGV530443H, NGV530733H, NGV550501H, NGV550506H, NGV550966H, NGV550968H, NGV551132H, NGV551135H, NGV551368H, NGV551371H, NGV551378H, NGV551652H, NGV551794H, NGV552195H, NGV524407H, NGV513981H, NGV517153H, NGV517859H, NGV522910H, NGV523420H, NGV527294H, NGV528540H, NGV516430H, NGV526200H, NGV526025H, NGV527843H, NGV528745H, NGV529456H, NGV530204H, NGV527105H, NGV513923H, NGV516657H, NGV522622H, NGV530830H, NGV513005H, NGV514973H, NGV516976H, NGV521262H, NGV526588H, NGV518962H, NGV530603H, NGV529061H, NGV521107H, NGV527839H, NGV527934H, NGV528713H, NGV529423H, NGV530158H, NGV530227H, NGV550789H, NGV513422H, NGV550812H, NGV523241H, NGV526230H, NGV551778H, NGV522723H, NGV551842H, NGV513051H, NGV513775H, NGV514363H, NGV516953H, NGV517385H, NGV519071H, NGV520579H, NGV526096H, NGV526216H, NGV526429H, NGV526596H, NGV527981H, NGV528293H, NGV528543H, NGV530214H, NGV550510H, NGV513304H, NGV513392H, NGV514377H, NGV514511H, NGV515183H, NGV515840H, NGV516274H, NGV517137H, NGV517138H, NGV517195H, NGV519730H, NGV519915H, NGV520405H, NGV521276H, NGV522046H, NGV522354H, NGV522396H, NGV522780H, NGV522783H, NGV522832H, NGV523492H, NGV523886H, NGV524572H, NGV524679H, NGV525870H, NGV526125H, NGV526370H, NGV526531H, NGV526592H, NGV527003H, NGV527510H, NGV527513H, NGV527748H, NGV528864H, NGV550085H, NGV550090H, NGV550177H, NGV550662H, NGV550665H, NGV550870H, NGV550871H, NGV550888H, NGV522054H, NGV522985H, NGV526488H, NGV527350H, NGV527797H, NGV516988H, NGV514770H, NGV516629H, NGV518886H, NGV527505H, NGV551037H, NGV522806H, NGV526754H, NGV551494H, NGV513587H, NGV513778H, NGV516002H, NGV516682H, NGV518031H, NGV518727H, NGV518799H, NGV519255H, NGV520648H, NGV520951H, NGV521313H, NGV526214H, NGV526350H, NGV527344H, NGV527387H, NGV528694H, NGV529560H, NGV551656H, NGV524714H, NGV516470H, NGV524164H, NGV515944H, NGV514530H, NGV515552H, NGV515885H, NGV519565H, NGV520709H, NGV521951H, NGV526274H, NGV526343H, NGV526468H, NGV527481H, NGV527963H, NGV528695H, NGV528794H, NGV529523H, NGV530005H, NGV550809H, NGV551356H, NGV551996H, NGV552089H, NGV520630H, NGV521405H, NGV522669H, NGV528487H, NGV529974H, NGV519816H, NGV526007H, NGV526888H, NGV526944H, NGV527525H, NGV550121H, NGV550829H, NGV551541H, NGV514129H, NGV519002H, NGV525081H, NGV525275H, NGV525414H, NGV527654H, NGV509900H, NGV513916H, NGV514961H, NGV516393H, NGV516666H, NGV523860H, NGV527958H, NGV529441H, NGV514177H, NGV517074H, NGV517591H, NGV521014H, NGV524263H, NGV525145H, NGV525450H, NGV526145H, NGV527007H, NGV527008H, NGV527305H, NGV527880H, NGV529610H, NGV550422H, NGV550424H, NGV550425H, NGV550849H, NGV551551H, NGV513289H, NGV513692H, NGV513845H, NGV513881H, NGV513884H, NGV513948H, NGV517849H, NGV517851H, NGV520701H, NGV520742H, NGV524027H, NGV524029H, NGV524031H, NGV524033H, NGV524037H, NGV550555H, NGV550895H, NGV550897H, NGV525898H, NGV517034H, NGV519774H, NGV521730H, NGV524436H, NGV519315H, NGV528856H, NGV550748H, NGV551060H, NGV514060H, NGV514067H, NGV520530H, NGV522340H, NGV525030H, NGV525392H, NGV525836H, NGV526462H, NGV526562H, NGV528874H, NGV528982H, NGV524466H, NGV527395H, NGV528781H, NGV550622H, NGV515979H, NGV520050H, NGV520227H, NGV527369H, NGV529583H, NGV550625H, NGV513713H, NGV522059H, NGV529917H, NGV550434H, NGV550857H, NGV551412H, NGV551577H, NGV519686H, NGV520285H, NGV520788H, NGV521321H, NGV521322H, NGV521690H, NGV521778H, NGV524637H, NGV525765H, NGV526763H, NGV530157H, NGV530717H, NGV550634H, NGV522543H, NGV527012H, NGV528675H, NGV528676H, NGV528691H, NGV528710H, NGV528857H, NGV529126H, NGV530029H, NGV530872H, NGV551043H, NGV551741H, NGV551828H, NGV518985H, NGV519018H, NGV513029H, NGV521913H, NGV523104H, NGV527978H, NGV529407H, NGV530122H, NGV530644H, NGV515016H, NGV517727H, NGV517827H, NGV517940H, NGV521679H, NGV524570H, NGV525581H, NGV526237H, NGV526335H, NGV526396H, NGV527060H, NGV527297H, NGV528017H, NGV528930H, NGV530768H, NGV530824H, NGV550938H, NGV550977H, NGV551658H, NGV551914H, NGV518901H, NGV518949H, NGV520499H, NGV521620H, NGV529729H, NGV552027H, NGV520731H, NGV524730H, NGV550283H, NGV515468H, NGV520254H, NGV520258H, NGV521339H, NGV524579H, NGV526480H, NGV527426H, NGV527427H, NGV527428H, NGV527430H, NGV528973H, NGV529263H, NGV530762H, NGV550688H, NGV552011H, NGV515047H, NGV509711H, NGV516457H, NGV517093H, NGV523462H, NGV528098H, NGV528888H, NGV552052H, NGV527156H, NGV551053H, NGV524112H, NGV525355H, NGV526530H, NGV527227H, NGV527228H, NGV527230H, NGV530247H, NGV550389H, NGV550396H, NGV550668H, NGV518140H, NGV529385H, NGV529535H, NGV530731H, NGV526659H, NGV527373H, NGV513670H, NGV513671H, NGV513672H, NGV514955H, NGV520748H, NGV521266H, NGV521267H, NGV550128H, NGV550130H, NGV550131H, NGV550133H, NGV550134H, NGV513749H, NGV513752H, NGV515461H, NGV525797H, NGV528019H, NGV550880H, NGV551697H, NGV513806H, NGV514487H, NGV515739H, NGV516517H, NGV516528H, NGV517375H, NGV518484H, NGV518620H, NGV522210H, NGV522790H, NGV522791H, NGV523578H, NGV523838H, NGV524485H, NGV524487H, NGV524565H, NGV525073H, NGV525380H, NGV525656H, NGV525658H, NGV526759H, NGV527129H, NGV527722H, NGV527752H, NGV527754H, NGV528165H, NGV528168H, NGV528512H, NGV529009H, NGV529186H, NGV530622H, NGV550991H, NGV551773H, NGV551964H, NGV515368H, NGV518606H, NGV550706H, NGV552013H, NGV514369H, NGV517398H, NGV522444H, NGV525844H, NGV527218H, NGV514983H, NGV524756H, NGV525321H, NGV529837H, NGV513408H, NGV525567H, NGV521846H, NGV525863H, NGV529933H, NGV530079H, NGV516998H, NGV517292H, NGV518929H, NGV519053H, NGV522503H, NGV525878H, NGV527973H, NGV528217H, NGV528509H, NGV528740H, NGV530225H, NGV551185H, NGV551445H, NGV551574H, NGV551649H, NGV551655H, NGV551845H, NGV513655H, NGV513787H, NGV529342H, NGV521949H, NGV523025H, NGV525453H, NGV526889H, NGV526893H, NGV529990H, NGV551208H, NGV525402H, NGV514925H, NGV517500H, NGV517993H, NGV520576H, NGV522002H, NGV522759H, NGV526700H, NGV527282H, NGV527528H, NGV527884H, NGV528242H, NGV529372H, NGV529411H, NGV529901H, NGV550258H, NGV550615H, NGV550677H, NGV551399H, NGV551519H, NGV551903H, NGV552081H, NGV518801H, NGV525802H, NGV550917H, NGV523608H, NGV524009H, NGV525881H, NGV550879H, NGV513576H, NGV526266H, NGV514301H, NGV515252H, NGV522715H, NGV523650H, NGV525441H, NGV525443H, NGV525482H, NGV527612H, NGV527819H, NGV528257H, NGV529565H, NGV515712H, NGV516453H, NGV516993H, NGV518963H, NGV519212H, NGV519346H, NGV522206H, NGV522778H, NGV523337H, NGV524505H, NGV527673H, NGV527687H, NGV527688H, NGV527697H, NGV528920H, NGV529239H, NGV529251H, NGV529503H, NGV529569H, NGV530171H, NGV550300H, NGV550302H, NGV550721H, NGV551069H, NGV551091H, NGV551909H, NGV522198H, NGV522469H, NGV520264H, NGV524055H, NGV524376H, NGV525092H, NGV550716H, NGV551804H, NGV516986H, NGV527349H, NGV519845H, NGV520677H, NGV524311H, NGV524898H, NGV525976H, NGV527919H, NGV550100H, NGV552087H, NGV509667H, NGV513027H, NGV517342H, NGV517470H, NGV520435H, NGV520473H, NGV521851H, NGV523447H, NGV523952H, NGV524416H, NGV524644H, NGV527365H, NGV527546H, NGV528909H, NGV529519H, NGV529547H, NGV550271H, NGV550740H, NGV551205H, NGV551283H, NGV551387H, NGV551535H, NGV513470H, NGV519621H, NGV528007H, NGV515945H, NGV526651H, NGV551984H, NGV521067H, NGV522378H, NGV514297H, NGV514305H, NGV518369H, NGV526282H, NGV526691H, NGV525274H, NGV525831H, NGV528316H, NGV551629H, NGV523481H, NGV524097H, NGV526296H, NGV529267H, NGV530574H, NGV516324H, NGV550199H, NGV551063H, NGV551604H, NGV551635H, NGV551875H, NGV529740H, NGV516147H, NGV517456H, NGV518050H, NGV518056H, NGV519015H, NGV519021H, NGV519083H, NGV519110H, NGV519111H, NGV520272H, NGV520274H, NGV520289H, NGV521040H, NGV521441H, NGV522052H, NGV522889H, NGV523483H, NGV523530H, NGV523531H, NGV525873H, NGV525880H, NGV526154H, NGV526789H, NGV526790H, NGV526963H, NGV526966H, NGV526969H, NGV527674H, NGV527675H, NGV527676H, NGV527677H, NGV527681H, NGV527941H, NGV550352H, NGV550357H, NGV550358H, NGV551272H, NGV551274H, NGV551275H, NGV513508H, NGV517095H, NGV518493H, NGV519913H, NGV522913H, NGV529642H, NGV550681H, NGV524101H, NGV527933H, NGV550496H, NGV523310H, NGV514888H, NGV515521H, NGV530442H, NGV530885H, NGV525654H, NGV526527H, NGV509686H, NGV514197H, NGV517430H, NGV521144H, NGV521300H, NGV523502H, NGV524274H, NGV526997H, NGV527281H, NGV515190H, NGV516866H, NGV518874H, NGV518934H, NGV525305H, NGV525806H, NGV526680H, NGV526702H, NGV527440H, NGV528901H, NGV512996H, NGV513069H, NGV518945H, NGV518950H, NGV519418H, NGV520615H, NGV521980H, NGV522711H, NGV523974H, NGV524711H, NGV525752H, NGV527446H, NGV527598H, NGV528712H, NGV528853H, NGV529075H, NGV529726H, NGV550049H, NGV550285H, NGV550852H, NGV551429H, NGV551762H, NGV526035H, NGV520605H, NGV525445H, NGV529648H, NGV551247H, NGV527075H, NGV529068H, NGV525477H, NGV527538H, NGV523960H, NGV518447H, NGV519357H, NGV519725H, NGV522729H, NGV526026H, NGV530565H, NGV522649H, NGV526049H, NGV527135H, NGV529820H, NGV550695H, NGV551584H, NGV515986H, NGV520726H, NGV522761H, NGV523252H, NGV525855H, NGV526737H, NGV527515H, NGV527516H, NGV527517H, NGV528105H, NGV529132H, NGV551133H, NGV514746H, NGV515298H, NGV515692H, NGV527620H, NGV527866H, NGV529338H, NGV529906H, NGV530475H, NGV550901H, NGV551304H, NGV524542H, NGV528617H, NGV550728H, NGV527640H, NGV509749H, NGV525939H, NGV527969H, NGV528095H, NGV523594H, NGV515165H, NGV515362H, NGV516203H, NGV530592H, NGV551934H, NGV513494H, NGV516374H, NGV517883H, NGV521858H, NGV522939H, NGV524616H, NGV528861H, NGV529417H, NGV550401H, NGV550731H, NGV520954H, NGV521775H, NGV517636H, NGV518202H, NGV522609H, NGV522845H, NGV524911H, NGV525150H, NGV526456H, NGV528045H, NGV528237H, NGV551186H, NGV530434H, NGV530440H, NGV550511H, NGV525848H, NGV515395H, NGV527456H, NGV516559H, NGV518578H, NGV522213H, NGV522405H, NGV527850H, NGV515402H, NGV516259H, NGV551124H, NGV551198H, NGV516319H, NGV528885H, NGV529623H, NGV550923H, NGV509912H, NGV527766H, NGV517172H, NGV523374H, NGV527792H, NGV513037H, NGV513042H, NGV513414H, NGV516675H, NGV521853H, NGV528537H, NGV515517H, NGV550774H, NGV513660H, NGV520074H, NGV527196H, NGV527217H, NGV514149H, NGV515489H, NGV515943H, NGV519942H, NGV520556H, NGV520673H, NGV520872H, NGV522261H, NGV528703H, NGV514362H, NGV551369H, NGV515711H, NGV517652H, NGV518365H, NGV519082H, NGV520031H, NGV524293H, NGV524298H, NGV526135H, NGV528836H, NGV530201H, NGV550522H, NGV550527H, NGV550530H, NGV550534H, NGV550537H, NGV550538H, NGV509735H, NGV518153H, NGV521726H, NGV521927H, NGV523499H, NGV524138H, NGV526037H, NGV526038H, NGV526219H, NGV526220H, NGV526438H, NGV527761H, NGV527765H, NGV527789H, NGV528097H, NGV528302H, NGV528380H, NGV528546H, NGV528924H, NGV528927H, NGV529484H, NGV529908H, NGV530579H, NGV530586H, NGV550366H, NGV551137H, NGV551235H, NGV551527H, NGV551536H, NGV551549H, NGV552023H, NGV552025H, NGV521984H, NGV523379H, NGV516179H, NGV528453H, NGV529410H, NGV530107H, NGV530108H, NGV514513H, NGV523356H, NGV527683H, NGV528554H, NGV530878H, NGV513739H, NGV518175H, NGV518571H, NGV519131H, NGV527289H, NGV527925H, NGV529111H, NGV550782H, NGV550786H, NGV516312H, NGV528808H, NGV529393H, NGV525888H, NGV526114H, NGV522175H, NGV525448H, NGV529609H, NGV520246H, NGV521666H, NGV529158H, NGV529520H, NGV550705H, NGV551553H, NGV526783H, NGV528235H, NGV513338H, NGV525868H, NGV527807H, NGV550736H, NGV550995H, NGV551218H, NGV514353H, NGV515024H, NGV517254H, NGV519707H, NGV519709H, NGV522219H, NGV522764H, NGV523210H, NGV523227H, NGV523235H, NGV525551H, NGV527165H, NGV527166H, NGV527169H, NGV527177H, NGV527178H, NGV528479H, NGV529163H, NGV529670H, NGV529675H, NGV529687H, NGV529689H, NGV529692H, NGV529700H, NGV529701H, NGV529703H, NGV529719H, NGV530800H, NGV551983H, NGV520583H, NGV521270H, NGV528739H, NGV550630H, NGV528701H, NGV513663H, NGV518393H, NGV521386H, NGV523060H, NGV526148H, NGV528201H, NGV530596H, NGV530597H, NGV550426H, NGV551970H, NGV525538H, NGV516472H, NGV515358H, NGV520612H, NGV528980H, NGV509751H, NGV519285H, NGV523828H, NGV528687H, NGV529113H, NGV529119H, NGV550831H, NGV550832H, NGV551636H, NGV524841H, NGV524905H, NGV528301H, NGV515928H, NGV519154H, NGV519289H, NGV513523H, NGV515655H, NGV518379H, NGV519668H, NGV521840H, NGV522222H, NGV523049H, NGV523970H, NGV524235H, NGV524371H, NGV528845H, NGV528846H, NGV529256H, NGV529579H, NGV550194H, NGV550777H, NGV550778H, NGV516946H, NGV529848H, NGV550183H, NGV550835H, NGV551390H, NGV518991H, NGV520633H, NGV513626H, NGV515865H, NGV517593H, NGV518779H, NGV520092H, NGV551451H, NGV513535H, NGV521106H, NGV527566H, NGV527631H, NGV529314H, NGV530176H, NGV551869H, NGV516510H, NGV526883H, NGV528985H, NGV552020H, NGV522933H, NGV526039H, NGV518255H, NGV515819H, NGV527771H, NGV519717H, NGV522385H, NGV530616H, NGV514115H, NGV525712H, NGV529807H, NGV530812H, NGV550644H, NGV550645H, NGV551353H, NGV525249H, NGV550245H, NGV528025H, NGV513403H, NGV528811H, NGV527582H, NGV527959H, NGV529378H, NGV551257H, NGV526089H, NGV520458H, NGV518401H, NGV551284H, NGV518274H, NGV516987H, NGV517175H, NGV529405H, NGV529435H, NGV529492H, NGV527236H, NGV530542H, NGV514541H, NGV509758H, NGV516028H, NGV517009H, NGV522589H, NGV522828H, NGV522993H, NGV530105H, NGV525070H, NGV515524H, NGV516286H, NGV518327H, NGV518639H, NGV519632H, NGV519847H, NGV520599H, NGV521652H, NGV522033H, NGV522402H, NGV522847H, NGV524116H, NGV525820H, NGV525907H, NGV526491H, NGV526492H, NGV527900H, NGV528826H, NGV528892H, NGV528965H, NGV528968H, NGV529400H, NGV550410H, NGV551115H, NGV551116H, NGV551348H, NGV551812H, NGV551838H, NGV552044H, NGV515118H, NGV517573H, NGV527496H, NGV551532H, NGV551693H, NGV515989H, NGV518915H, NGV550672H, NGV551832H, NGV551966H, NGV550883H, NGV526123H, NGV514503H, NGV515443H, NGV518947H, NGV527091H, NGV527565H, NGV528109H, NGV529527H, NGV513318H, NGV524493H, NGV525086H, NGV513640H, NGV521762H, NGV521763H, NGV525478H, NGV527706H, NGV528417H, NGV525612H, NGV517114H, NGV519905H, NGV523901H, NGV524674H, NGV529589H, NGV529825H, NGV551158H, NGV551159H, NGV525908H, NGV550132H, NGV529235H, NGV509668H, NGV513455H, NGV515338H, NGV515355H, NGV515356H, NGV517229H, NGV518403H, NGV518951H, NGV519743H, NGV520066H, NGV521814H, NGV522760H, NGV525083H, NGV525084H, NGV526001H, NGV526256H, NGV528699H, NGV529645H, NGV530754H, NGV515615H, NGV517016H, NGV517786H, NGV521899H, NGV524618H, NGV528445H, NGV528908H, NGV516375H, NGV513614H, NGV529868H, NGV518982H, NGV550819H, NGV526471H, NGV530499H, NGV530503H, NGV550702H, NGV550846H, NGV527728H, NGV518928H, NGV520445H, NGV524081H, NGV526795H, NGV527662H, NGV530101H, NGV551799H, NGV513620H, NGV516639H, NGV525010H, NGV525407H, NGV551025H, NGV551398H, NGV513919H, NGV519521H, NGV521431H, NGV524390H, NGV530585H, NGV550772H, NGV513598H, NGV514449H, NGV517475H, NGV517476H, NGV519381H, NGV519382H, NGV522035H, NGV522229H, NGV525670H, NGV526829H, NGV528015H, NGV528953H, NGV529856H, NGV530182H, NGV551724H, NGV550279H, NGV513434H, NGV518631H, NGV527968H, NGV528004H, NGV528823H, NGV529371H, NGV519383H, NGV525951H, NGV526013H, NGV526217H, NGV514041H, NGV514440H, NGV515920H, NGV516367H, NGV518452H, NGV519012H, NGV522383H, NGV522384H, NGV524716H, NGV525167H, NGV526314H, NGV526695H, NGV526696H, NGV527655H, NGV529045H, NGV529103H, NGV550000H, NGV550002H, NGV550693H, NGV551734H, NGV551736H, NGV551739H, NGV551740H, NGV513796H, NGV515064H, NGV520314H, NGV509932H, NGV513492H, NGV515546H, NGV516743H, NGV518976H, NGV520181H, NGV520879H, NGV521786H, NGV524488H, NGV524955H, NGV525639H, NGV526697H, NGV526930H, NGV527815H, NGV528585H, NGV528700H, NGV528728H, NGV529109H, NGV529381H, NGV529521H, NGV529571H, NGV529795H, NGV530172H, NGV530419H, NGV550717H, NGV551453H, NGV551458H, NGV513536H, NGV526297H, NGV526694H, NGV527295H, NGV551925H, NGV523319H, NGV513397H, NGV517224H, NGV518632H, NGV518851H, NGV523349H, NGV525035H, NGV525889H, NGV528395H, NGV551163H, NGV522990H, NGV524658H, NGV525560H, NGV525790H, NGV525882H, NGV527667H, NGV528991H, NGV530113H, NGV550229H, NGV550841H, NGV551075H, NGV517344H, NGV529290H, NGV529454H, NGV529455H, NGV550601H, NGV550843H, NGV550848H, NGV551212H, NGV551452H, NGV528559H, NGV523876H, NGV515230H, NGV519880H, NGV520885H, NGV523047H, NGV525309H, NGV530262H, NGV551046H, NGV516015H, NGV513812H, NGV514475H, NGV514495H, NGV517925H, NGV519408H, NGV521000H, NGV523719H, NGV524244H, NGV524250H, NGV527101H, NGV528091H, NGV528092H, NGV529804H, NGV550502H, NGV513963H, NGV515734H, NGV516309H, NGV517067H, NGV528021H, NGV550179H, NGV550652H, NGV551696H, NGV518864H, NGV518938H, NGV524998H, NGV526060H, NGV527374H, NGV528347H, NGV529364H, NGV517359H, NGV515125H, NGV517249H, NGV522084H, NGV524417H, NGV525050H, NGV525128H, NGV525822H, NGV525903H, NGV526863H, NGV526864H, NGV527031H, NGV527312H, NGV527554H, NGV527874H, NGV529112H, NGV529122H, NGV529524H, NGV550197H, NGV550252H, NGV550517H, NGV551164H, NGV551381H, NGV551413H, NGV522197H, NGV551004H, NGV519402H, NGV513039H, NGV513597H, NGV516996H, NGV520566H, NGV524165H, NGV524302H, NGV524489H, NGV525808H, NGV526833H, NGV527781H, NGV529227H, NGV529401H, NGV551300H, NGV551363H, NGV518280H, NGV519791H, NGV520451H, NGV526816H, NGV527826H, NGV528424H, NGV529640H, NGV550776H, NGV519347H, NGV519722H, NGV551298H, NGV514007H, NGV514951H, NGV515162H, NGV515494H, NGV516729H, NGV516730H, NGV518507H, NGV519585H, NGV520372H, NGV520437H, NGV525994H, NGV526034H, NGV527817H, NGV527820H, NGV528550H, NGV528573H, NGV529635H, NGV529637H, NGV529763H, NGV530235H, NGV530236H, NGV550378H, NGV550610H, NGV550674H, NGV550714H, NGV550715H, NGV551072H, NGV551619H, NGV551624H, NGV551955H, NGV552277H, NGV552302H, NGV553043H, NGV513686H, NGV513760H, NGV515194H, NGV522863H, NGV529559H, NGV529950H, NGV513539H, NGV518491H, NGV528215H, NGV515212H, NGV519098H, NGV520024H, NGV524560H, NGV527944H, NGV529203H, NGV551664H, NGV551677H, NGV516019H, NGV523364H, NGV526358H, NGV526626H, NGV527737H, NGV527773H, NGV528771H, NGV529491H, NGV550747H, NGV530134H, NGV524936H, NGV517026H, NGV517027H, NGV518194H, NGV520665H, NGV522255H, NGV524292H, NGV525383H, NGV526538H, NGV518237H, NGV518637H, NGV526665H, NGV527487H, NGV529155H, NGV529505H, NGV529515H, NGV525410H, NGV515772H, NGV518588H, NGV509933H, NGV513009H, NGV513573H, NGV513678H, NGV515055H, NGV517045H, NGV517984H, NGV518376H, NGV518487H, NGV518489H, NGV518876H, NGV518895H, NGV519179H, NGV519180H, NGV519211H, NGV519596H, NGV520048H, NGV521667H, NGV522865H, NGV523009H, NGV523280H, NGV523819H, NGV524156H, NGV524157H, NGV524746H, NGV524795H, NGV524800H, NGV525353H, NGV525354H, NGV526493H, NGV526750H, NGV527705H, NGV527896H, NGV527990H, NGV528359H, NGV528897H, NGV528948H, NGV528949H, NGV529399H, NGV529942H, NGV529947H, NGV530010H, NGV530070H, NGV530076H, NGV550047H, NGV550236H, NGV550757H, NGV550760H, NGV550817H, NGV550864H, NGV550867H, NGV551151H, NGV551409H, NGV551923H, NGV519114H, NGV525874H, NGV525981H, NGV530472H, NGV530618H, NGV514276H, NGV514787H, NGV516428H, NGV519448H, NGV525156H, NGV525804H, NGV528186H, NGV529135H, NGV529576H, NGV530870H, NGV551466H, NGV551947H, NGV513352H, NGV529382H, NGV522400H, NGV515253H, NGV514277H, NGV518725H, NGV528374H, NGV528481H, NGV551280H, NGV521628H, NGV524922H, NGV524231H, NGV522719H, NGV513657H, NGV513825H, NGV515746H, NGV517771H, NGV518927H, NGV521335H, NGV521622H, NGV521811H, NGV522816H, NGV523123H, NGV525046H, NGV525065H, NGV526069H, NGV526346H, NGV528934H, NGV529873H, NGV530091H, NGV530110H, NGV530766H, NGV551101H, NGV513063H, NGV513339H, NGV518381H, NGV518742H, NGV520075H, NGV523919H, NGV526422H, NGV518782H, NGV530494H, NGV551907H, NGV525914H, NGV529421H, NGV551040H, NGV514059H, NGV521715H, NGV524517H, NGV529006H, NGV514340H, NGV522106H, NGV522801H, NGV523380H, NGV551209H, NGV514125H, NGV517213H, NGV517368H, NGV525833H, NGV527389H, NGV527396H, NGV527962H, NGV529490H, NGV530055H, NGV530233H, NGV530673H, NGV520752H, NGV522348H, NGV523085H, NGV526713H, NGV528896H, NGV513784H, NGV517824H, NGV520183H, NGV520211H, NGV521691H, NGV523064H, NGV523190H, NGV526723H, NGV527401H, NGV528120H, NGV528484H, NGV528780H, NGV529886H, NGV530629H, NGV530886H, NGV550045H, NGV550468H, NGV550583H, NGV550591H, NGV551817H, NGV551840H, NGV524324H, NGV527776H, NGV529962H, NGV550813H, NGV551286H, NGV523770H, NGV526173H, NGV528738H, NGV514106H, NGV515388H, NGV516012H, NGV516359H, NGV519060H, NGV519095H, NGV523622H, NGV525374H, NGV527307H, NGV527308H, NGV527479H, NGV529972H, NGV550184H, NGV550430H, NGV551743H, NGV513834H, NGV513921H, NGV514174H, NGV514566H, NGV515543H, NGV515900H, NGV517171H, NGV517532H, NGV517533H, NGV519090H, NGV521382H, NGV522257H, NGV522258H, NGV522805H, NGV523549H, NGV523555H, NGV523752H, NGV525544H, NGV525952H, NGV526203H, NGV527127H, NGV527653H, NGV529474H, NGV529476H, NGV530519H, NGV550143H, NGV550992H, NGV551220H, NGV551428H, NGV551467H, NGV551662H, NGV551767H, NGV551770H, NGV551870H, NGV552196H, NGV552986H, NGV517363H, NGV525039H, NGV550464H, NGV513036H, NGV513741H, NGV517800H, NGV518514H, NGV523924H, NGV530963H, NGV550855H, NGV517544H, NGV526032H, NGV527663H, NGV528539H, NGV551440H, NGV530148H, NGV514070H, NGV517352H, NGV519651H, NGV525592H, NGV528047H, NGV528052H, NGV528250H, NGV528251H, NGV529715H, NGV551447H, NGV551612H, NGV513785H, NGV515862H, NGV518205H, NGV524950H, NGV525103H, NGV526912H, NGV527104H, NGV527386H, NGV527691H, NGV528720H, NGV529964H, NGV530263H, NGV550099H, NGV550452H, NGV550683H, NGV550692H, NGV550753H, NGV551985H, NGV552084H, NGV552363H, NGV513748H, NGV529230H, NGV516957H, NGV523035H, NGV525000H, NGV529136H, NGV551252H, NGV551421H, NGV519567H, NGV516306H, NGV516460H, NGV517248H, NGV518294H, NGV520444H, NGV523038H, NGV525367H, NGV526063H, NGV526152H, NGV526904H, NGV527363H, NGV550146H, NGV529863H, NGV526073H, NGV523285H, NGV523475H, NGV524951H, NGV527443H, NGV551830H, NGV550013H, NGV550182H, NGV551883H, NGV515029H, NGV525520H, NGV528841H, NGV550198H, NGV513947H, NGV527221H, NGV530625H, NGV527415H, NGV519115H, NGV521311H, NGV527824H, NGV551968H, NGV517777H, NGV523171H, NGV523588H, NGV523589H, NGV527872H, NGV509719H, NGV509937H, NGV515682H, NGV520141H, NGV521912H, NGV522386H, NGV524280H, NGV527524H, NGV527879H, NGV529284H, NGV530155H, NGV550913H, NGV551462H, NGV551505H, NGV551509H, NGV509927H, NGV514535H, NGV515950H, NGV520271H, NGV520944H, NGV521675H, NGV521752H, NGV521764H, NGV528596H, NGV530244H, NGV551560H, NGV551714H, NGV515533H, NGV530742H, NGV515003H, NGV523986H, NGV525869H, NGV526712H, NGV551650H, NGV520330H, NGV514893H, NGV516900H, NGV517000H, NGV528314H, NGV528590H, NGV529720H, NGV530753H, NGV550293H, NGV513607H, NGV514540H, NGV514901H, NGV522727H, NGV524744H, NGV524883H, NGV525401H, NGV525884H, NGV526449H, NGV527399H, NGV529093H, NGV529346H, NGV550329H, NGV550412H, NGV551953H, NGV514260H, NGV514271H, NGV514579H, NGV514822H, NGV515069H, NGV515504H, NGV515993H, NGV516403H, NGV516405H, NGV517481H, NGV518409H, NGV518802H, NGV518807H, NGV518808H, NGV518809H, NGV519133H, NGV519738H, NGV520079H, NGV520105H, NGV520107H, NGV520178H, NGV520406H, NGV520687H, NGV521036H, NGV521646H, NGV521647H, NGV523107H, NGV523423H, NGV523424H, NGV524895H, NGV525161H, NGV525361H, NGV526107H, NGV526111H, NGV526576H, NGV527242H, NGV527629H, NGV528937H, NGV528938H, NGV528979H, NGV529939H, NGV529944H, NGV530086H, NGV530540H, NGV530594H, NGV530605H, NGV550205H, NGV550217H, NGV550249H, NGV550711H, NGV550712H, NGV550820H, NGV550878H, NGV551533H, NGV551534H, NGV551570H, NGV551688H, NGV552003H, NGV552007H, NGV552090H, NGV513499H, NGV517082H, NGV517516H, NGV518532H, NGV518566H, NGV519510H, NGV526053H, NGV526578H, NGV526887H, NGV526952H, NGV527356H, NGV527695H, NGV527943H, NGV528103H, NGV528104H, NGV529460H, NGV529619H, NGV551723H, NGV551950H, NGV523994H, NGV516520H, NGV516524H, NGV517671H, NGV517817H, NGV519851H, NGV523031H, NGV523937H, NGV524272H, NGV526479H, NGV526740H, NGV527648H, NGV528682H, NGV528962H, NGV550459H, NGV551146H, NGV551414H, NGV551691H, NGV550431H, NGV522850H, NGV522851H, NGV523435H, NGV526047H, NGV526476H, NGV529105H, NGV551711H, NGV526927H, NGV513463H, NGV529046H, NGV520411H, NGV520970H, NGV509832H, NGV520494H, NGV522068H, NGV551068H, NGV518324H, NGV521251H, NGV509722H, NGV509885H, NGV514026H, NGV514032H, NGV514180H, NGV514781H, NGV514783H, NGV514785H, NGV515694H, NGV515761H, NGV515875H, NGV515877H, NGV516751H, NGV516872H, NGV517068H, NGV517497H, NGV517504H, NGV518004H, NGV518293H, NGV518344H, NGV518467H, NGV518522H, NGV518525H, NGV518772H, NGV518859H, NGV518860H, NGV518905H, NGV518908H, NGV519030H, NGV519159H, NGV519420H, NGV519457H, NGV519654H, NGV519676H, NGV519678H, NGV520000H, NGV520290H, NGV520334H, NGV520585H, NGV520586H, NGV520631H, NGV520868H, NGV521234H, NGV521236H, NGV521272H, NGV521273H, NGV521320H, NGV521665H, NGV521685H, NGV521883H, NGV521884H, NGV521885H, NGV521886H, NGV521889H, NGV522005H, NGV522009H, NGV522330H, NGV522369H, NGV522836H, NGV523369H, NGV523726H, NGV523845H, NGV523849H, NGV523978H, NGV524186H, NGV524187H, NGV524688H, NGV524737H, NGV524915H, NGV524920H, NGV525118H, NGV525324H, NGV525682H, NGV525807H, NGV525941H, NGV526095H, NGV526099H, NGV526663H, NGV526787H, NGV526899H, NGV526900H, NGV526934H, NGV526938H, NGV526951H, NGV526953H, NGV526954H, NGV526957H, NGV527093H, NGV527264H, NGV527303H, NGV527519H, NGV527587H, NGV527588H, NGV527709H, NGV527710H, NGV527711H, NGV527712H, NGV527725H, NGV527811H, NGV527847H, NGV527991H, NGV527997H, NGV528037H, NGV528230H, NGV528232H, NGV528717H, NGV528719H, NGV528744H, NGV528748H, NGV528749H, NGV528792H, NGV529021H, NGV529035H, NGV529219H, NGV529274H, NGV529394H, NGV529615H, NGV529631H, NGV529639H, NGV529641H, NGV529651H, NGV529667H, NGV530039H, NGV530635H, NGV530636H, NGV550646H, NGV550658H, NGV550953H, NGV550955H, NGV550967H, NGV550971H, NGV550972H, NGV551001H, NGV551355H, NGV551357H, NGV551360H, NGV551361H, NGV551383H, NGV551394H, NGV551395H, NGV551397H, NGV551512H, NGV551513H, NGV551687H, NGV551745H, NGV551747H, NGV551748H, NGV551756H, NGV551851H, NGV551853H, NGV551855H, NGV552248H, NGV552344H, NGV552955H, NGV552960H, NGV553006H, NGV553048H, NGV520380H, NGV525666H, NGV527423H, NGV529211H, NGV530787H, NGV550837H, NGV552049H, NGV513697H, NGV514953H, NGV519530H, NGV520459H, NGV521309H, NGV527600H, NGV529138H, NGV550635H, NGV517313H, NGV517792H, NGV517793H, NGV526249H, NGV516899H, NGV525302H, NGV551704H, NGV551998H, NGV514188H, NGV515375H, NGV521218H, NGV521259H, NGV523119H, NGV526090H, NGV527050H, NGV527051H, NGV528262H, NGV530725H, NGV513608H, NGV514870H, NGV521035H, NGV522573H, NGV522576H, NGV524152H, NGV524154H, NGV527304H, NGV527316H, NGV527317H, NGV550414H, NGV550417H, NGV550419H, NGV517058H, NGV521656H, NGV514225H, NGV515050H, NGV520090H, NGV522720H, NGV523163H, NGV523358H, NGV524460H, NGV527416H, NGV527417H, NGV527419H, NGV528345H, NGV528599H, NGV528798H, NGV528799H, NGV550307H, NGV550872H, NGV550996H, NGV551332H, NGV551801H, NGV513512H, NGV514824H, NGV514849H, NGV515078H, NGV516091H, NGV516911H, NGV517796H, NGV518339H, NGV518343H, NGV519513H, NGV520127H, NGV521303H, NGV521643H, NGV521735H, NGV521876H, NGV522094H, NGV522721H, NGV522873H, NGV523027H, NGV523995H, NGV524283H, NGV524699H, NGV524872H, NGV525358H, NGV526245H, NGV526628H, NGV526849H, NGV527278H, NGV527471H, NGV527521H, NGV528057H, NGV528693H, NGV528708H, NGV529319H, NGV530186H, NGV530631H, NGV550007H, NGV550144H, NGV550215H, NGV550312H, NGV551227H, NGV551230H, NGV551294H, NGV551437H, NGV515661H, NGV519373H, NGV521012H, NGV522140H, NGV524267H, NGV524268H, NGV524269H, NGV524288H, NGV528464H, NGV550440H, NGV550442H, NGV550443H, NGV550445H, NGV527400H, NGV528866H, NGV530520H, NGV530575H, NGV551057H, NGV551544H, NGV514349H, NGV514351H, NGV514483H, NGV518999H, NGV520023H, NGV521669H, NGV522233H, NGV522295H, NGV522351H, NGV523100H, NGV524065H, NGV524073H, NGV525992H, NGV525993H, NGV526076H, NGV527947H, NGV530255H, NGV551679H, NGV551858H, NGV552039H, NGV525940H, NGV526501H, NGV514217H, NGV514372H, NGV515753H, NGV516543H, NGV524815H, NGV526022H, NGV526106H, NGV527461H, NGV529162H, NGV529841H, NGV530790H, NGV550788H, NGV550825H, NGV514429H, NGV514810H, NGV515182H, NGV516032H, NGV517071H, NGV517072H, NGV520656H, NGV520693H, NGV521325H, NGV521378H, NGV525359H, NGV526344H, NGV526443H, NGV527023H, NGV528020H, NGV528304H, NGV528311H, NGV528869H, NGV529050H, NGV530011H, NGV530198H, NGV530220H, NGV530477H, NGV530480H, NGV550701H, NGV551203H, NGV513563H, NGV518196H, NGV519663H, NGV523187H, NGV526165H, NGV526166H, NGV527126H, NGV529965H, NGV522561H, NGV523002H, NGV526498H, NGV528032H, NGV528188H, NGV529705H, NGV518197H, NGV522694H, NGV526116H, NGV526225H, NGV527326H, NGV527330H, NGV550487H, NGV550488H, NGV550489H, NGV550491H, NGV529671H, NGV551289H, NGV514484H, NGV528048H, NGV550629H, NGV550631H, NGV550650H, NGV513871H, NGV524174H, NGV526686H, NGV526924H, NGV527138H, NGV527380H, NGV528503H, NGV528505H, NGV529008H, NGV513800H, NGV528961H, NGV529774H, NGV515906H, NGV527606H, NGV526184H, NGV527805H, NGV528107H, NGV529487H, NGV530229H, NGV530436H, NGV552055H, NGV527200H, NGV551695H, NGV523412H, NGV527347H, NGV518387H, NGV519088H, NGV520225H, NGV521421H, NGV527895H, NGV528898H, NGV528899H, NGV529784H, NGV530774H, NGV514056H, NGV519627H, NGV522042H, NGV526085H, NGV530088H, NGV517542H, NGV550775H, NGV515038H, NGV517703H, NGV523375H, NGV524628H, NGV526746H, NGV528983H, NGV551566H, NGV551905H, NGV551927H, NGV551928H, NGV519729H, NGV523547H, NGV527215H, NGV527182H, NGV550833H, NGV517111H, NGV520078H, NGV529237H, NGV552017H, NGV552075H, NGV517390H, NGV517392H, NGV523488H, NGV526977H, NGV550915H, NGV526523H, NGV513989H, NGV522716H, NGV523998H, NGV524003H, NGV526812H, NGV527041H, NGV528571H, NGV550478H, NGV550479H, NGV551278H, NGV520778H, NGV516083H, NGV518407H, NGV527201H, NGV528116H, NGV514638H, NGV515622H, NGV521709H, NGV523273H, NGV524430H, NGV528023H, NGV528865H, NGV528871H, NGV528873H, NGV529951H, NGV530398H, NGV551173H, NGV551295H, NGV515374H, NGV526681H, NGV528454H, NGV528493H, NGV551755H, NGV509709H, NGV513000H, NGV514126H, NGV518406H, NGV550508H, NGV551196H, NGV514205H, NGV514208H, NGV517437H, NGV517439H, NGV521168H, NGV524230H, NGV524234H, NGV520888H, NGV522630H, NGV530894H, NGV550767H, NGV509760H, NGV513641H, NGV516227H, NGV517066H, NGV517547H, NGV519368H, NGV521274H, NGV523807H, NGV524380H, NGV524830H, NGV525577H, NGV526011H, NGV526044H, NGV526100H, NGV526298H, NGV527076H, NGV527354H, NGV527720H, NGV527914H, NGV528548H, NGV529173H, NGV529178H, NGV530173H, NGV530177H, NGV530259H, NGV530951H, NGV550444H, NGV550492H, NGV551455H, NGV551456H, NGV551564H, NGV551718H, NGV552034H, NGV513358H, NGV515307H, NGV522634H, NGV515899H, NGV520503H, NGV523683H, NGV523764H, NGV509744H, NGV517184H, NGV522817H, NGV522859H, NGV523467H, NGV527861H, NGV529040H, NGV550135H, NGV530218H, NGV551415H, NGV513852H, NGV515232H, NGV527580H, NGV529168H, NGV529220H, NGV529679H, NGV529893H, NGV530192H, NGV551120H, NGV551199H, NGV551215H, NGV551243H, NGV517429H, NGV514597H, NGV517211H, NGV520081H, NGV521792H, NGV523884H, NGV524923H, NGV526621H, NGV526981H, NGV527986H, NGV528263H, NGV528854H, NGV529356H, NGV529497H, NGV529498H, NGV530116H, NGV530495H, NGV550617H, NGV550619H, NGV550620H, NGV552040H, NGV513419H, NGV517780H, NGV550561H, NGV550949H, NGV551929H, NGV524443H, NGV550624H, NGV551022H, NGV519100H, NGV525090H, NGV520089H, NGV527241H, NGV527929H, NGV529305H, NGV529306H, NGV530199H, NGV551187H, NGV551291H, NGV551834H, NGV517689H, NGV526495H, NGV523573H, NGV525051H, NGV527407H, NGV517179H, NGV527162H, NGV528747H, NGV550127H, NGV513059H, NGV515919H, NGV516263H, NGV518268H, NGV518431H, NGV519324H, NGV522022H, NGV522834H, NGV523055H, NGV523057H, NGV523450H, NGV526247H, NGV527948H, NGV528172H, NGV528223H, NGV528491H, NGV530175H, NGV530844H, NGV550081H, NGV550084H, NGV551685H, NGV551846H, NGV551941H, NGV551965H, NGV525461H, NGV528587H, NGV517555H, NGV551561H, NGV518069H, NGV518884H, NGV528802H, NGV530265H, NGV550103H, NGV551425H, NGV526857H, NGV529177H, NGV513509H, NGV518326H, NGV530211H, NGV517134H, NGV520589H, NGV521938H, NGV522969H, NGV524880H, NGV526235H, NGV526444H, NGV526684H, NGV526797H, NGV527170H, NGV529877H, NGV530131H, NGV551599H, NGV526283H, NGV528683H, NGV529439H, NGV550584H, NGV521439H, NGV527364H, NGV527665H, NGV513541H, NGV515502H, NGV516478H, NGV519004H, NGV519986H, NGV520727H, NGV520728H, NGV524767H, NGV525819H, NGV525852H, NGV527291H, NGV528394H, NGV529387H, NGV529483H, NGV550863H, NGV521370H, NGV525004H, NGV525429H, NGV514594H, NGV520814H, NGV520067H, NGV514554H, NGV518966H, NGV519120H, NGV524653H, NGV525078H, NGV528206H, NGV514958H, NGV519850H, NGV519856H, NGV523426H, NGV524143H, NGV524145H, NGV527623H, NGV528870H, NGV528919H, NGV528941H, NGV528942H, NGV528943H, NGV528976H, NGV550173H, NGV550176H, NGV514359H, NGV516759H, NGV520192H, NGV525279H, NGV527628H, NGV514865H, NGV520722H, NGV526990H, NGV516983H, NGV526727H, NGV529948H, NGV550469H, NGV550633H, NGV521286H, NGV528824H, NGV529682H, NGV530527H, NGV519329H, NGV509683H, NGV513895H, NGV515413H, NGV516992H, NGV517715H, NGV518798H, NGV520961H, NGV520964H, NGV520965H, NGV520966H, NGV520969H, NGV520974H, NGV522429H, NGV522895H, NGV523010H, NGV523011H, NGV523674H, NGV524515H, NGV524860H, NGV525843H, NGV526514H, NGV527593H, NGV527646H, NGV528143H, NGV528192H, NGV528925H, NGV528926H, NGV529462H, NGV529470H, NGV550306H, NGV550471H, NGV550557H, NGV550844H, NGV551112H, NGV551420H, NGV551478H, NGV514635H, NGV519224H, NGV522364H, NGV522556H, NGV527651H, NGV522124H, NGV513404H, NGV513517H, NGV515996H, NGV520062H, NGV528228H, NGV552038H, NGV518232H, NGV515518H, NGV527322H, NGV527325H, NGV528122H, NGV528542H, NGV550476H, NGV513844H, NGV509707H, NGV515641H, NGV518855H, NGV521750H, NGV523518H, NGV523652H, NGV527960H, NGV528275H, NGV528704H, NGV550308H, NGV550845H, NGV551242H, NGV523877H, NGV517549H, NGV519325H, NGV527370H, NGV517571H, NGV522460H, NGV525809H, NGV528187H, NGV528194H, NGV528558H, NGV529624H, NGV530140H, NGV530143H, NGV550494H, NGV550768H, NGV551024H, NGV551028H, NGV551435H, NGV551588H, NGV551795H, NGV527769H, NGV513378H, NGV518638H, NGV520893H, NGV521711H, NGV522108H, NGV522248H, NGV522505H, NGV525532H, NGV527309H, NGV527361H, NGV527656H, NGV528140H, NGV528752H, NGV528922H, NGV550416H, NGV551319H, NGV551391H, NGV551554H, NGV518897H, NGV524428H, NGV526002H, NGV529222H, NGV514998H, NGV521326H, NGV521328H, NGV523731H, NGV527352H, NGV527854H, NGV528495H, NGV529304H, NGV530098H, NGV551726H, NGV527755H, NGV525446H, NGV550364H, NGV524993H, NGV530829H, NGV530833H, NGV550637H, NGV551646H, NGV551878H, NGV515975H, NGV524207H, NGV525958H, NGV527193H, NGV527434H, NGV528831H, NGV530072H, NGV527742H, NGV517479H, NGV529383H, NGV513980H, NGV514156H, NGV525734H, NGV527393H, NGV530435H, NGV550907H, NGV519214H, NGV520126H, NGV522310H, NGV524214H, NGV526115H, NGV528247H, NGV522275H, NGV527581H, NGV522717H, NGV526380H, NGV527844H, NGV529208H, NGV529209H, NGV515107H, NGV518681H, NGV521249H, NGV524987H, NGV528709H, NGV530210H, NGV530257H, NGV551384H, NGV514264H, NGV514269H, NGV516332H, NGV517139H, NGV517140H, NGV519559H, NGV519560H, NGV521284H, NGV522656H, NGV522666H, NGV523366H, NGV524941H, NGV525892H, NGV525895H, NGV526092H, NGV526093H, NGV527246H, NGV527250H, NGV550062H, NGV551343H, NGV551810H, NGV551814H, NGV551847H, NGV514334H, NGV521745H, NGV528721H, NGV513002H, NGV522664H, NGV525801H, NGV517521H, NGV521807H, NGV521852H, NGV523774H, NGV524934H, NGV529434H, NGV550639H, NGV550876H, NGV518340H, NGV520995H, NGV521683H, NGV526387H, NGV528572H, NGV550866H, NGV519272H, NGV515299H, NGV519062H, NGV520011H, NGV527834H, NGV529876H, NGV529963H, NGV551013H, NGV524387H, NGV514232H, NGV515172H, NGV520261H, NGV525990H, NGV526734H, NGV528574H, NGV529469H, NGV530690H, NGV551246H, NGV551248H, NGV551738H, NGV551774H, NGV522992H, NGV523189H, NGV525720H, NGV525960H, NGV528947H, NGV515264H, NGV520720H, NGV527466H, NGV528261H, NGV514823H, NGV516412H, NGV517662H, NGV528555H, NGV528561H, NGV550868H, NGV513545H, NGV513853H, NGV515762H, NGV516664H, NGV518653H, NGV518980H, NGV519101H, NGV520144H, NGV521706H, NGV522575H, NGV523580H, NGV523920H, NGV524665H, NGV525384H, NGV525596H, NGV525597H, NGV526144H, NGV526441H, NGV528569H, NGV530501H, NGV550148H, NGV550737H, NGV550749H, NGV550800H, NGV517678H, NGV519985H, NGV522434H, NGV522591H, NGV527080H, NGV528547H, NGV528671H, NGV529512H, NGV529513H, NGV530617H, NGV550704H, NGV551712H, NGV551849H, NGV551961H, NGV518222H, NGV526830H, NGV530267H, NGV519376H, NGV521803H, NGV526950H, NGV527905H, NGV528111H, NGV528326H, NGV530693H, NGV550922H, NGV551184H, NGV551676H, NGV521749H, NGV527870H, NGV527951H, NGV521376H, NGV524497H, NGV526719H, NGV527507H, NGV528298H, NGV530526H, NGV550230H, NGV552041H, NGV521753H, NGV509693H, NGV520823H, NGV527376H, NGV527867H, NGV528317H, NGV529822H, NGV514509H, NGV520663H, NGV524566H, NGV524567H, NGV525157H, NGV550679H, NGV514871H, NGV516447H, NGV516691H, NGV518693H, NGV520240H, NGV521005H, NGV522668H, NGV526406H, NGV526894H, NGV527542H, NGV527730H, NGV528279H, NGV528705H, NGV517952H, NGV520128H, NGV524945H, NGV527906H, NGV528478H, NGV514435H, NGV530243H, NGV524433H, NGV526529H, NGV523747H, NGV529612H, NGV529756H, NGV530099H, NGV551365H, NGV520538H, NGV527025H, NGV527130H, NGV529896H, NGV530413H, NGV551228H, NGV551992H, NGV514964H, NGV523243H, NGV527106H, NGV516438H, NGV516508H, NGV519509H, NGV520241H, NGV520243H, NGV520686H, NGV521364H, NGV521688H, NGV521707H, NGV524173H, NGV524971H, NGV525012H, NGV526800H, NGV526848H, NGV527563H, NGV527846H, NGV527856H, NGV527932H, NGV528458H, NGV528589H, NGV529013H, NGV529016H, NGV529590H, NGV530066H, NGV530123H, NGV551192H, NGV551293H, NGV516098H, NGV518070H, NGV525521H, NGV526127H, NGV526550H, NGV526885H, NGV527038H, NGV530119H, NGV517614H, NGV525698H, NGV528027H, NGV530598H, NGV551232H, NGV522069H, NGV530242H, NGV520387H, NGV551443H, NGV518839H, NGV519338H, NGV519872H, NGV520600H, NGV521124H, NGV522866H, NGV523081H, NGV523125H, NGV523519H, NGV523581H, NGV524286H, NGV524469H, NGV526068H, NGV526320H, NGV526520H, NGV527506H, NGV528621H, NGV528762H, NGV528763H, NGV528765H, NGV528936H, NGV529115H, NGV529606H, NGV530154H, NGV530168H, NGV530532H, NGV530535H, NGV550004H, NGV550005H, NGV550710H, NGV551639H, NGV517151H, NGV518138H, NGV528767H, NGV529264H, NGV529465H, NGV529611H, NGV550363H, NGV550752H, NGV529843H, NGV515144H, NGV523040H, NGV518210H, NGV525810H, NGV513966H, NGV518298H, NGV519370H, NGV529426H, NGV530373H, NGV550691H, NGV552009H, NGV529573H, NGV518272H, NGV519058H, NGV519538H, NGV520316H, NGV522811H, NGV518367H, NGV518794H, NGV519608H, NGV524914H, NGV526991H, NGV526998H, NGV529945H, NGV550012H, NGV550014H, NGV525503H, NGV513553H, NGV517293H, NGV523079H, NGV526871H, NGV527579H, NGV551665H, NGV529538H, NGV514491H, NGV520508H, NGV520667H, NGV530395H, NGV550928H, NGV526433H, NGV523564H, NGV529179H, NGV515202H, NGV517297H, NGV523984H, NGV526061H, NGV526862H, NGV529418H, NGV550499H, NGV521350H, NGV526619H, NGV519907H, NGV526124H, NGV550122H, NGV513560H, NGV517007H, NGV523415H, NGV523558H, NGV551175H, NGV551872H, NGV514084H, NGV528551H, NGV525871H, NGV522549H, NGV526250H, NGV519177H, NGV524318H, NGV529986H, NGV550321H, NGV550821H, NGV515041H, NGV515042H, NGV515932H, NGV517740H, NGV520410H, NGV520647H, NGV521409H, NGV522566H, NGV523902H, NGV524759H, NGV524761H, NGV526535H, NGV526537H, NGV528577H, NGV528580H, NGV528582H, NGV528584H, NGV529587H, NGV529592H, NGV529594H, NGV529598H, NGV529613H, NGV551260H, NGV551261H, NGV551276H, NGV551337H, NGV552050H, NGV514589H, NGV521757H, NGV551700H, NGV514152H, NGV517169H, NGV518106H, NGV526255H, NGV526270H, NGV527476H, NGV515249H, NGV518408H, NGV522921H, NGV527328H, NGV527488H, NGV529872H, NGV530063H, NGV530064H, NGV530206H, NGV550405H, NGV550627H, NGV550727H, NGV551225H, NGV551495H, NGV512975H, NGV527767H, NGV528205H, NGV513504H, NGV526197H, NGV528321H, NGV528324H, NGV550336H, NGV551372H, NGV551485H, NGV515032H, NGV516699H, NGV519527H, NGV521834H, NGV522678H, NGV522680H, NGV523812H, NGV526940H, NGV527319H, NGV528804H, NGV528807H, NGV528809H, NGV529741H, NGV529744H, NGV529752H, NGV530020H, NGV530111H, NGV530166H, NGV551219H, NGV552967H, NGV530114H, NGV513585H, NGV526609H, NGV550273H, NGV550433H, NGV551811H, NGV526454H, NGV550261H, NGV516995H, NGV552021H, NGV516985H, NGV522216H, NGV523969H, NGV513521H, NGV528900H, NGV551775H, NGV513917H, NGV521968H, NGV526182H, NGV528734H, NGV530090H, NGV530444H, NGV550066H, NGV550764H, NGV551097H, NGV551439H, NGV529167H, NGV528932H, NGV528622H, NGV530181H, NGV551498H, NGV551995H, NGV515922H, NGV515009H, NGV530112H, NGV551055H, NGV551058H, NGV551575H, NGV518698H, NGV523116H, NGV550294H, NGV520980H, NGV522531H, NGV525529H, NGV528954H, NGV528956H, NGV550298H, NGV550299H, NGV550881H, NGV551531H, NGV516760H, NGV525615H, NGV525635H, NGV527913H, NGV550900H, NGV551518H, NGV516423H, NGV520976H, NGV521017H, NGV521806H, NGV526489H, NGV513375H, NGV518558H, NGV522282H, NGV523947H, NGV524088H, NGV550735H, NGV523990H, NGV551188H, NGV513958H, NGV515433H, NGV528329H, NGV528333H, NGV528489H, NGV529766H, NGV522558H, NGV515021H, NGV513633H, NGV515008H, NGV515421H, NGV516507H, NGV516798H, NGV517024H, NGV518148H, NGV518778H, NGV518965H, NGV520337H, NGV521351H, NGV522931H, NGV523122H, NGV523177H, NGV524504H, NGV525101H, NGV526168H, NGV526187H, NGV527103H, NGV527377H, NGV527490H, NGV527535H, NGV527808H, NGV527858H, NGV528244H, NGV528566H, NGV530877H, NGV550317H, NGV551099H, NGV551555H, NGV519878H, NGV524076H, NGV525434H, NGV527158H, NGV528299H, NGV530207H, NGV530388H, NGV530619H, NGV522368H, NGV519704H, NGV515744H, NGV524654H, NGV550408H, NGV550656H, NGV551980H, NGV526641H, NGV529175H, NGV515958H, NGV522319H, NGV526627H, NGV528615H, NGV529599H, NGV550842H, NGV551265H, NGV515067H, NGV526764H, NGV528402H, NGV550262H, NGV550481H, NGV551127H, NGV551432H, NGV551988H, NGV520052H, NGV551287H, NGV551408H, NGV522370H, NGV529143H, NGV513680H, NGV525946H, NGV526813H, NGV514977H, NGV518815H, NGV520438H, NGV522355H, NGV525953H, NGV525988H, NGV528868H, NGV529288H, NGV515398H, NGV522083H, NGV523871H, NGV524374H, NGV524562H, NGV527954H, NGV528162H, NGV550552H, NGV550675H, NGV517964H, NGV530203H, NGV530861H, NGV519715H, NGV528911H, NGV526128H, NGV526631H, NGV526683H, NGV527465H, NGV518457H, NGV523863H, NGV523875H, NGV525293H, NGV526877H, NGV515511H, NGV517099H, NGV518600H, NGV521829H, NGV551680H, NGV514960H, NGV514994H, NGV526079H, NGV527199H, NGV513872H, NGV514996H, NGV551543H, NGV517221H, NGV527123H, NGV527922H, NGV528024H, NGV550158H, NGV551310H, NGV551973H, NGV517113H, NGV518922H, NGV523012H, NGV528761H, NGV528782H, NGV515659H, NGV521308H, NGV529444H, NGV529809H, NGV515967H, NGV520979H, NGV522044H, NGV513971H, NGV517485H, NGV518061H, NGV518585H, NGV527027H, NGV527030H, NGV513395H, NGV515636H, NGV516011H, NGV516496H, NGV517446H, NGV517469H, NGV517519H, NGV520161H, NGV520412H, NGV521362H, NGV521438H, NGV523382H, NGV523958H, NGV524881H, NGV525382H, NGV526221H, NGV526745H, NGV527391H, NGV528221H, NGV528225H, NGV529458H, NGV522872H, NGV513979H, NGV517810H, NGV523003H, NGV524786H, NGV527483H, NGV550125H, NGV551194H, NGV516758H, NGV526391H, NGV514311H, NGV514543H, NGV514544H, NGV514546H, NGV517404H, NGV519633H, NGV520913H, NGV514874H, NGV524626H, NGV525877H, NGV527013H, NGV527342H, NGV527500H, NGV527702H, NGV528141H, NGV528238H, NGV529486H, NGV529663H, NGV550255H, NGV552042H, NGV517335H, NGV527945H, NGV515242H, NGV515973H, NGV516934H, NGV519531H, NGV520352H, NGV525449H, NGV527833H, NGV527837H, NGV550874H, NGV552064H, NGV552067H, NGV517020H, NGV522714H, NGV526405H, NGV527418H, NGV530217H, NGV514409H, NGV515703H, NGV517360H, NGV527056H, NGV528743H, NGV530040H, NGV550581H, NGV550823H, NGV517281H, NGV520288H, NGV525325H, NGV550676H, NGV551140H, NGV552375H, NGV515898H, NGV527762H, NGV530634H, NGV517507H, NGV551563H, NGV526323H, NGV528368H, NGV528684H, NGV529419H, NGV515171H, NGV516518H, NGV514087H, NGV514088H, NGV518740H, NGV513731H, NGV513738H, NGV517330H, NGV520001H, NGV520904H, NGV520905H, NGV527258H, NGV527263H, NGV550104H, NGV550106H, NGV550107H, NGV550111H, NGV550112H, NGV550113H, NGV550116H, NGV550118H, NGV550120H, NGV515169H, NGV551681H, NGV528821H, NGV550314H, NGV514072H, NGV514877H, NGV515018H, NGV522858H, NGV527249H, NGV528005H, NGV529531H, NGV529959H, NGV550278H, NGV550632H, NGV550875H, NGV517747H, NGV519037H, NGV519358H, NGV520808H, NGV526418H, NGV527605H, NGV527813H, NGV528575H, NGV551211H, NGV551320H, NGV552048H, NGV509827H, NGV509905H, NGV514086H, NGV514095H, NGV514102H, NGV514802H, NGV515156H, NGV515175H, NGV515409H, NGV515887H, NGV516489H, NGV516502H, NGV516678H, NGV516735H, NGV517131H, NGV517132H, NGV517136H, NGV517386H, NGV517875H, NGV517877H, NGV517885H, NGV518601H, NGV519639H, NGV519928H, NGV519929H, NGV520064H, NGV520277H, NGV521089H, NGV521094H, NGV521809H, NGV521944H, NGV521945H, NGV521974H, NGV522038H, NGV522839H, NGV522840H, NGV523853H, NGV523854H, NGV524208H, NGV524213H, NGV524755H, NGV526179H, NGV526180H, NGV526332H, NGV526333H, NGV526598H, NGV526599H, NGV526807H, NGV526821H, NGV527575H, NGV527829H, NGV527831H, NGV527832H, NGV528072H, NGV528073H, NGV528132H, NGV528305H, NGV529551H, NGV529552H, NGV529556H, NGV530850H, NGV530851H, NGV530852H, NGV530904H, NGV550349H, NGV551094H, NGV551334H, NGV551538H, NGV551545H, NGV551547H, NGV551666H, NGV518550H, NGV523741H, NGV526964H, NGV551266H, NGV525466H, NGV525937H, NGV551792H, NGV552037H, NGV527493H, NGV530094H, NGV551586H, NGV518762H, NGV515467H, NGV521692H, NGV528838H, NGV513551H, NGV518313H, NGV527743H, NGV550856H, NGV516667H, NGV517376H, NGV519026H, NGV520280H, NGV525595H, NGV526291H, NGV528751H, NGV509892H, NGV513431H, NGV514834H, NGV518678H, NGV518840H, NGV521402H, NGV522732H, NGV523097H, NGV523162H, NGV525954H, NGV526150H, NGV527220H, NGV527887H, NGV529463H, NGV529829H, NGV530058H, NGV530060H, NGV550497H, NGV550997H, NGV551438H, NGV552062H, NGV551051H, NGV513824H, NGV517216H, NGV524393H, NGV526729H, NGV512988H, NGV514002H, NGV515736H, NGV522307H, NGV523570H, NGV524577H, NGV521404H, NGV524408H, NGV527967H, NGV527987H, NGV528958H, NGV530160H, NGV530509H, NGV518086H, NGV524947H, NGV527964H, NGV530165H, NGV521674H, NGV522335H, NGV526349H, NGV550320H, NGV525326H, NGV551769H, NGV513498H, NGV513805H, NGV519736H, NGV525830H, NGV530087H, NGV550588H, NGV550689H, NGV513056H, NGV527889H, NGV528156H, NGV512974H, NGV526312H, NGV528515H, NGV527134H, NGV528376H, NGV526473H, NGV527037H, NGV529857H, NGV527909H, NGV514257H, NGV523070H, NGV524329H, NGV524332H, NGV524333H, NGV527454H, NGV528218H, NGV529601H, NGV550415H, NGV513341H, NGV518277H, NGV526088H, NGV550390H, NGV514226H, NGV522697H, NGV521976H, NGV527437H, NGV528914H, NGV550219H, NGV551873H, NGV509878H, NGV520156H, NGV522553H, NGV522584H, NGV528124H, NGV528689H, NGV529216H, NGV529493H, NGV550092H, NGV552030H, NGV525423H, NGV513305H, NGV521173H, NGV530047H, NGV527618H, NGV551254H, NGV518666H, NGV519710H, NGV520202H, NGV520929H, NGV527097H, NGV527558H, NGV527689H, NGV528409H, NGV529852H, NGV529960H, NGV530676H, NGV550750H, NGV550860H, NGV551256H, NGV551591H, NGV527366H, NGV513574H, NGV518544H, NGV520324H, NGV551087H, NGV552070H, NGV514490H, NGV519641H, NGV523247H, NGV528995H, NGV517631H, NGV522876H, NGV528428H, NGV550890H, NGV551206H, NGV517207H, NGV517536H, NGV529530H, NGV550962H, NGV551253H, NGV520276H, NGV530221H, NGV551833H, NGV520476H, NGV529920H, NGV514940H, NGV518827H, NGV513766H, NGV524541H, NGV515547H, NGV517989H, NGV551065H, NGV551195H, NGV515704H, NGV516290H, NGV518940H, NGV518942H, NGV516905H, NGV525041H, NGV528795H, NGV527355H, NGV523625H, NGV518724H, NGV520053H, NGV527351H, NGV527876H, NGV529294H, NGV530080H, NGV515629H, NGV517163H, NGV526309H, NGV515722H, NGV525502H, NGV529072H, NGV530025H, NGV552035H, NGV517453H, NGV518267H, NGV523438H, NGV525447H, NGV526054H, NGV528825H, NGV550678H, NGV524313H, NGV515011H, NGV516266H, NGV516267H, NGV518998H, NGV519380H, NGV519935H, NGV522063H, NGV522814H, NGV524091H, NGV527245H, NGV528284H, NGV528872H, NGV529036H, NGV529039H, NGV529278H, NGV529826H, NGV550457H, NGV550804H, NGV550805H, NGV551250H, NGV551585H, NGV551637H, NGV551924H, NGV552031H, NGV552032H, NGV514221H, NGV517194H, NGV518318H, NGV518479H, NGV520553H, NGV521031H, NGV521033H, NGV521648H, NGV522039H, NGV523362H, NGV523961H, NGV526268H, NGV527011H, NGV527293H, NGV529412H, NGV551362H, NGV551522H, NGV551829H, NGV552990H, NGV515361H, NGV521240H, NGV522089H, NGV522263H, NGV523946H, NGV525015H, NGV525894H, NGV526021H, NGV526554H, NGV526555H, NGV527192H, NGV528500H, NGV529658H, NGV529664H, NGV529666H, NGV529707H, NGV530458H, NGV530522H, NGV530794H, NGV530797H, NGV551981H, NGV552000H, NGV515236H, NGV522361H, NGV524114H, NGV526051H, NGV526708H, NGV526836H, NGV527216H, NGV528679H, NGV529577H, NGV551107H, NGV519247H, NGV521902H, NGV525074H, NGV525168H, NGV525886H, NGV527243H, NGV527632H, NGV528059H, NGV529197H, NGV530139H, NGV551009H, NGV551155H, NGV522784H, NGV528013H, NGV550964H, NGV528457H, NGV516013H, NGV514830H, NGV521862H, NGV529224H, NGV521875H, NGV528685H, NGV516515H, NGV520294H, NGV521186H, NGV521780H, NGV551374H, NGV518618H, NGV518741H, NGV550448H, NGV515363H, NGV520911H, NGV524381H, NGV526884H, NGV527006H, NGV527459H, NGV528793H, NGV529160H, NGV551405H, NGV551587H, NGV518745H, NGV521021H, NGV515218H, NGV519118H, NGV519122H, NGV551012H, NGV529307H, NGV528757H, NGV513323H, NGV529228H, NGV518575H, NGV522577H, NGV524049H, NGV525306H, NGV528041H, NGV528294H, NGV528696H, NGV529496H, NGV530849H, NGV550154H, NGV550155H, NGV550156H, NGV551071H, NGV551074H, NGV551306H, NGV551318H, NGV551489H, NGV551791H, NGV551974H, NGV551403H, NGV525158H, NGV551406H, NGV515879H, NGV522809H, NGV550323H, NGV517051H, NGV520787H, NGV527438H, NGV550074H, NGV550077H, NGV519270H, NGV528306H, NGV529910H, NGV530831H, NGV551524H, NGV528174H, NGV526117H, NGV517364H, NGV519277H, NGV519631H, NGV525155H, NGV519569H, NGV520215H, NGV520275H, NGV529313H, NGV530926H, NGV551951H, NGV522621H, NGV518692H, NGV522099H, NGV521256H, NGV525488H, NGV527143H, NGV515706H, NGV517176H, NGV520960H, NGV523981H, NGV518360H, NGV514109H, NGV526146H, NGV526206H, NGV529242H, NGV530528H, NGV550263H, NGV525913H, NGV550815H, NGV515404H, NGV530657H, NGV522802H, NGV522693H, NGV528732H, NGV515422H, NGV519548H, NGV527526H, NGV551430H, NGV527116H, NGV520229H, NGV526435H, NGV529069H, NGV550003H, NGV520059H, NGV516301H, NGV517539H, NGV523173H, NGV529176H, NGV526996H, NGV527972H, NGV529588H, NGV530084H, NGV550101H, NGV550925H, NGV551204H, NGV551491H, NGV551492H, NGV551514H, NGV514976H, NGV516303H, NGV516351H, NGV516355H, NGV517339H, NGV519348H, NGV520167H, NGV526070H, NGV526896H, NGV529320H, NGV529537H, NGV530231H, NGV550884H, NGV529541H, NGV551690H, NGV526668H, NGV513820H, NGV517622H, NGV524483H, NGV550235H, NGV551590H, NGV551600H, NGV519054H, NGV514288H, NGV525924H, NGV530623H, NGV518939H, NGV521748H, NGV524673H, NGV509727H, NGV527444H, NGV528171H, NGV518423H, NGV528779H, NGV514417H, NGV530174H, NGV513789H, NGV514461H, NGV514464H, NGV514466H, NGV514469H, NGV515911H, NGV515912H, NGV515913H, NGV517918H, NGV517920H, NGV517921H, NGV519123H, NGV519945H, NGV519946H, NGV523771H, NGV523776H, NGV523855H, NGV526461H, NGV526513H, NGV526773H, NGV527324H, NGV528325H, NGV528534H, NGV528557H, NGV530121H, NGV530590H, NGV530868H, NGV550606H, NGV550607H, NGV551473H, NGV551476H, NGV514012H, NGV514014H, NGV514017H, NGV514019H, NGV514020H, NGV514022H, NGV514241H, NGV515528H, NGV516936H, NGV521024H, NGV521025H, NGV521028H, NGV521029H, NGV521030H, NGV524349H, NGV524355H, NGV524361H, NGV524363H, NGV526257H, NGV526258H, NGV526262H, NGV526263H, NGV527016H, NGV527062H, NGV527063H, NGV527064H, NGV527068H, NGV527069H, NGV527070H, NGV527071H, NGV527501H, NGV529308H, NGV529309H, NGV529316H, NGV529317H, NGV529475H, NGV529525H, NGV529646H, NGV530128H, NGV530129H, NGV530130H, NGV550040H, NGV550042H, NGV550983H, NGV550984H, NGV550986H, NGV550990H, NGV513468H, NGV516680H, NGV519609H, NGV526306H, NGV513489H, NGV524933H, NGV525396H, NGV527422H, NGV527970H, NGV528079H, NGV530141H, NGV551011H, NGV527733H, NGV527888H, NGV528513H, NGV529806H, NGV529912H, NGV525776H, NGV530038H, NGV521018H, NGV526181H, NGV529124H, NGV530630H, NGV550690H, NGV519741H, NGV524571H, NGV527845H, NGV550404H, NGV521684H, NGV526105H, NGV528698H, NGV551779H, NGV515391H, NGV518200H, NGV529187H, NGV551493H, NGV552004H, NGV518333H, NGV526064H, NGV526084H, NGV527685H, NGV525069H, NGV528850H, NGV530161H, NGV520278H, NGV522665H, NGV524948H, NGV514633H, NGV520457H, NGV521306H, NGV513417H, NGV550059H, NGV522325H, NGV529529H, NGV516054H, NGV513319H, NGV515496H, NGV516568H, NGV519079H, NGV524464H, NGV525902H, NGV526484H, NGV528214H, NGV529722H, NGV550824H, NGV551521H, NGV551949H, NGV520375H, NGV519642H, NGV525314H, NGV528033H, NGV528678H, NGV526010H, NGV514573H, NGV520850H, NGV526486H, NGV550801H, NGV514182H, NGV518825H, NGV521700H, NGV522220H, NGV522308H, NGV522842H, NGV525533H, NGV525975H, NGV527155H, NGV527638H, NGV527668H, NGV550095H, NGV550485H, NGV526926H, NGV529620H, NGV525890H, NGV552022H, NGV513410H, NGV513507H, NGV530811H, NGV514846H, NGV520510H, NGV513617H, NGV520101H, NGV530118H, NGV550369H, NGV550724H, NGV517015H, NGV518422H, NGV519066H, NGV518548H, NGV519042H, NGV523311H, NGV526141H, NGV509708H, NGV514134H, NGV514612H, NGV515665H, NGV517976H, NGV518149H, NGV518375H, NGV523743H, NGV524662H, NGV527024H, NGV527509H, NGV530096H, NGV530190H, NGV527658H, NGV551659H, NGV551661H, NGV551918H, NGV523921H, NGV527449H, NGV527855H, NGV526999H, NGV530045H, NGV513357H, NGV513600H, NGV516343H, NGV516516H, NGV517087H, NGV520143H, NGV523695H, NGV524072H, NGV525205H, NGV526777H, NGV528601H, NGV528692H, NGV550640H, NGV551098H, NGV521390H, NGV523191H, NGV527625H, NGV522819H, NGV518865H, NGV527405H, NGV528315H, NGV529935H, NGV514280H, NGV517370H, NGV519870H, NGV525261H, NGV515568H, NGV518769H, NGV521418H, NGV522853H, NGV526572H, NGV528952H, NGV515678H, NGV518547H, NGV527931H, NGV528686H, NGV526986H, NGV529969H, NGV513950H, NGV518481H, NGV528243H, NGV525126H, NGV521432H, NGV524989H, NGV525011H, NGV530226H, NGV550686H, NGV551506H, NGV551746H, NGV513628H, NGV514065H, NGV519017H, NGV520527H, NGV521097H, NGV522404H, NGV530085H, NGV515188H, NGV515278H, NGV515933H, NGV519386H, NGV520949H, NGV522673H, NGV522906H, NGV524300H, NGV526814H, NGV526818H, NGV527660H, NGV530147H, NGV551223H, NGV551479H, NGV524766H, NGV525972H, NGV518284H, NGV520477H, NGV522698H, NGV528227H, NGV519173H, NGV527484H, NGV550305H, NGV523843H, NGV519720H, NGV551921H, NGV529607H, NGV529812H, NGV527489H, NGV518828H, NGV521870H, NGV522311H, NGV524977H, NGV526563H, NGV528002H, NGV528820H, NGV550687H, NGV551297H, NGV552082H, NGV519113H, NGV519473H, NGV525038H, NGV528851H, NGV550297H, NGV551861H, NGV519694H, NGV528753H, NGV530665H, NGV518478H, NGV519542H, NGV526679H, NGV523186H, NGV551255H, NGV551920H, NGV523997H, NGV528471H, NGV550048H, NGV515925H, NGV517154H, NGV517165H, NGV529938H, NGV551448H, NGV515417H, NGV529751H, NGV550008H, NGV550264H, NGV550309H, NGV513452H, NGV520978H, NGV518114H, NGV527822H, NGV517428H, NGV524979H, NGV527431H, NGV528182H, NGV528224H, NGV528893H, NGV513596H, NGV514222H, NGV514476H, NGV515689H, NGV522843H, NGV523957H, NGV528082H, NGV528090H, NGV528830H, NGV551309H, NGV527499H, NGV514600H, NGV518850H, NGV520005H, NGV520301H, NGV520498H, NGV525979H, NGV526570H, NGV526908H, NGV528253H, NGV528285H, NGV528348H, NGV528902H, NGV528940H, NGV529232H, NGV529911H, NGV530209H, NGV530269H, NGV550975H, NGV551233H, NGV551916H, NGV513990H, NGV527626H, NGV550720H, NGV509874H, NGV521932H, NGV522826H, NGV524075H, NGV526769H, NGV528791H, NGV529147H, NGV519099H, NGV519239H, NGV519863H, NGV523188H, NGV524166H, NGV524940H, NGV527800H, NGV528248H, NGV528777H, NGV528844H, NGV528967H, NGV529940H, NGV550697H, NGV550961H, NGV551100H, NGV551706H, NGV513552H, NGV516211H, NGV517264H, NGV521424H, NGV523575H, NGV523831H, NGV518078H, NGV520932H, NGV524079H, NGV525930H, NGV525957H, NGV526164H, NGV526890H, NGV528946H, NGV529833H, NGV550732H, NGV551876H, NGV513327H, NGV525955H, NGV513954H, NGV525728H, NGV526993H, NGV550137H, NGV550882H, NGV550924H, NGV551576H, NGV516549H, NGV529980H, NGV519119H, NGV517371H, NGV526660H, NGV551288H, NGV521994H, NGV523026H, NGV527860H, NGV550593H, NGV514797H, NGV515143H, NGV509746H, NGV551148H, NGV513909H, NGV514769H, NGV515980H, NGV517528H, NGV517814H, NGV518260H, NGV526932H, NGV527164H, NGV551056H, NGV522506H, NGV515215H, NGV517820H, NGV529451H, NGV513791H, NGV526400H, NGV530170H, NGV517449H, NGV518805H, NGV529760H, NGV515908H, NGV524703H, NGV526046H, NGV527946H, NGV513839H, NGV515667H, NGV517441H, NGV520736H, NGV550807H, NGV551582H, NGV517116H, NGV517915H, NGV530646H, NGV528258H, NGV514981H, NGV550427H, NGV513864H, NGV517336H, NGV518611H, NGV518652H, NGV518654H, NGV520877H, NGV523304H, NGV526382H, NGV528071H, NGV529322H, NGV529574H, NGV529643H, NGV551708H, NGV551713H, NGV551715H, NGV551919H, NGV512995H, NGV519387H, NGV520045H, NGV523465H, NGV523687H, NGV527894H, NGV550680H, NGV522321H, NGV526796H, NGV551442H, NGV551373H, NGV551380H, NGV518655H, NGV551216H, NGV522286H, NGV520992H, NGV528096H, NGV550460H, NGV550461H, NGV551598H, NGV518969H, NGV513409H, NGV516092H, NGV517618H, NGV518504H, NGV520986H, NGV522519H, NGV523300H, NGV525996H, NGV526170H, NGV528858H, NGV529708H, NGV529757H, NGV529761H, NGV529887H, NGV550239H, NGV550862H, NGV551508H, NGV514471H, NGV529686H, NGV516525H, NGV519176H, NGV521621H, NGV522011H, NGV524445H, NGV526557H, NGV527557H, NGV527974H, NGV528724H, NGV528835H, NGV530124H, NGV530167H, NGV550943H, NGV551922H, NGV516753H, NGV522638H, NGV528674H, NGV528852H, NGV550612H, NGV522200H, NGV527010H, NGV527090H, NGV528904H, NGV529327H, NGV529336H, NGV529638H, NGV530184H, NGV530185H, NGV551039H, NGV551465H, NGV515382H, NGV514463H, NGV514935H, NGV527034H, NGV515751H, NGV517014H, NGV526463H, NGV522422H, NGV528038H, NGV513998H, NGV527384H, NGV552083H, NGV550123H, NGV550838H, NGV509812H, NGV523515H, NGV527194H, NGV528044H, NGV527463H, NGV515892H, NGV518342H, NGV518946H, NGV519254H, NGV519978H, NGV520142H, NGV522060H, NGV522613H, NGV522799H, NGV525206H, NGV527636H, NGV527873H, NGV529095H, NGV529097H, NGV550451H, NGV550696H, NGV514971H, NGV518776H, NGV524219H, NGV517567H, NGV529604H, NGV524315H, NGV515492H, NGV515741H, NGV516327H, NGV516328H, NGV516330H, NGV520208H, NGV520655H, NGV521105H, NGV521388H, NGV522027H, NGV523118H, NGV524085H, NGV525499H, NGV525664H, NGV525879H, NGV527451H, NGV527452H, NGV527692H, NGV528396H, NGV530706H, NGV530709H, NGV551054H, NGV523133H, NGV525334H, NGV551772H, NGV516674H, NGV518388H, NGV527421H, NGV528746H, NGV519038H, NGV520112H, NGV523004H, NGV523030H, NGV523160H, NGV523164H, NGV527044H, NGV527045H, NGV528945H, NGV529064H, NGV530178H, NGV550769H, NGV551200H, NGV551329H, NGV519616H, NGV526136H, NGV526347H, NGV526593H, NGV526806H, NGV528944H, NGV523760H, NGV526467H, NGV519612H, NGV526229H, NGV529283H, NGV530704H, NGV551719H, NGV552057H, NGV529581H, NGV529271H, NGV515154H, NGV515166H, NGV515917H, NGV516034H, NGV525297H, NGV529586H, NGV529602H, NGV529608H, NGV550231H, NGV551210H, NGV551231H, NGV514498H, NGV519911H, NGV521192H, NGV523045H, NGV527033H, NGV550145H, NGV550159H, NGV550180H, NGV551579H, NGV522416H, NGV513968H, NGV514157H, NGV513387H, NGV529011H, NGV550330H, NGV550439H, NGV517234H, NGV525812H, NGV527910H, NGV528516H, NGV521741H, NGV551402H, NGV526318H, NGV526319H, NGV529053H, NGV528786H, NGV525608H, NGV551464H, NGV522350H, NGV551703H, NGV521345H, NGV551258H, NGV519864H, NGV551241H, NGV520479H, NGV518064H, NGV527794H, NGV529480H, NGV517577H, NGV523013H, NGV525025H, NGV526134H, NGV527719H, NGV551502H, NGV550621H, NGV513384H, NGV527823H, NGV550806H, NGV523824H, NGV522309H, NGV519331H, NGV522636H, NGV514944H, NGV518932H, NGV525901H, NGV525966H, NGV525967H, NGV526931H, NGV527836H, NGV529172H, NGV529557H, NGV530258H, NGV550616H, NGV551134H, NGV513511H, NGV514000H, NGV513707H, NGV550914H, NGV524791H, NGV525933H, NGV527412H, NGV527721H, NGV515548H, NGV525997H, NGV528179H, NGV528680H, NGV515184H, NGV518752H, NGV521866H, NGV525316H, NGV525317H, NGV525318H, NGV525899H, NGV527458H, NGV528042H, NGV530189H, NGV550237H, NGV550402H, NGV519776H, NGV526083H, NGV529702H, NGV529970H, NGV514382H, NGV516097H, NGV519067H, NGV519251H, NGV525762H, NGV525791H, NGV525795H, NGV525798H, NGV526432H, NGV550453H, NGV550700H, NGV509854H, NGV518290H, NGV519074H, NGV520570H, NGV522049H, NGV522085H, NGV526019H, NGV528062H, NGV551092H, NGV526375H, NGV529597H, NGV520157H, NGV526841H, NGV528880H, NGV513582H, NGV551382H, NGV552033H, NGV529134H, NGV524539H, NGV530538H, NGV551805H, NGV514191H, NGV525154H, NGV528408H, NGV530095H, NGV550473H, NGV517159H, NGV517274H, NGV518530H, NGV518748H, NGV521406H, NGV521734H, NGV521959H, NGV522701H, NGV523427H, NGV524536H, NGV526018H, NGV526751H, NGV528758H, NGV530692H, NGV551213H, NGV551239H, NGV513373H, NGV516322H, NGV520546H, NGV521061H, NGV523089H, NGV523090H, NGV523394H, NGV523781H, NGV525008H, NGV525412H, NGV525565H, NGV525566H, NGV526196H, NGV527441H, NGV528822H, NGV529205H, NGV550406H, NGV551597H, NGV551710H, NGV515462H, NGV530756H, NGV550070H, NGV550438H, NGV550723H, NGV550941H, NGV551240H, NGV551468H, NGV517397H, NGV520807H, NGV526509H, NGV528295H, NGV523908H, NGV521755H, NGV527179H, NGV550946H, NGV516279H, NGV519571H, NGV520269H, NGV522838H, NGV524650H, NGV525803H, NGV526466H, NGV528415H, NGV528725H, NGV530082H, NGV530601H, NGV530648H, NGV550342H, NGV550345H, NGV513043H, NGV519866H, NGV526724H, NGV528410H, NGV526508H, NGV522804H, NGV523006H, NGV526658H, NGV528196H, NGV529478H, NGV530027H, NGV550654H, NGV519626H, NGV521009H, NGV526003H, NGV528065H, NGV515882H, NGV517513H, NGV519669H, NGV520196H, NGV520670H, NGV520671H, NGV522541H, NGV525617H, NGV526545H, NGV527335H, NGV527336H, NGV527339H, NGV527340H, NGV550565H, NGV550567H, NGV550568H, NGV513325H, NGV513554H, NGV517006H, NGV517053H, NGV518584H, NGV519327H, NGV519336H, NGV519589H, NGV519590H, NGV520977H, NGV523544H, NGV525152H, NGV525847H, NGV526252H, NGV528677H, NGV528729H, NGV529213H, NGV529352H, NGV529627H, NGV530106H, NGV530523H, NGV530550H, NGV550339H, NGV551657H, NGV551815H, NGV551897H, NGV551976H, NGV552014H, NGV527019H, NGV527251H, NGV528787H, NGV530240H, NGV530695H, NGV518299H, NGV518642H, NGV523314H, NGV523950H, NGV524007H, NGV524046H, NGV524048H, NGV524093H, NGV524516H, NGV524981H, NGV525663H, NGV525792H, NGV526595H, NGV530515H, NGV551520H, NGV522312H, NGV515147H, NGV518250H, NGV516003H, NGV520930H, NGV517587H, NGV530781H, NGV514762H, NGV528553H, NGV528069H, NGV551201H, NGV551936H, NGV516254H, NGV530943H, NGV514928H, NGV515043H, NGV518730H, NGV518800H, NGV519242H, NGV520771H, NGV522528H, NGV522774H, NGV523453H, NGV524108H, NGV524327H, NGV525085H, NGV525273H, NGV525409H, NGV525474H, NGV525643H, NGV526119H, NGV526120H, NGV527260H, NGV527261H, NGV527734H, NGV527738H, NGV527746H, NGV528552H, NGV528578H, NGV529479H, NGV529622H, NGV529916H, NGV530213H, NGV550098H, NGV550585H, NGV550597H, NGV527061H, NGV528101H, NGV530856H, NGV551548H, NGV518988H, NGV519670H, NGV518602H, NGV529292H, NGV529655H, NGV525991H, NGV526852H, NGV525284H, NGV525393H, NGV527508H, NGV520536H, NGV523175H, NGV524410H, NGV528903H, NGV529511H, NGV530628H, NGV530689H, NGV550087H, NGV517156H, NGV551089H, NGV518992H, NGV522062H, NGV523258H, NGV525189H, NGV525821H, NGV527828H, NGV529580H, NGV530504H, NGV551701H, NGV526195H, NGV527657H, NGV514587H, NGV522667H, NGV526607H, NGV524367H, NGV530241H, NGV523520H, NGV523654H, NGV517202H, NGV517942H, NGV518283H, NGV525099H, NGV527698H, NGV527878H, NGV528881H, NGV550196H, NGV550905H, NGV528923H, NGV509703H, NGV519577H, NGV521964H, NGV523318H, NGV525900H, NGV528690H, NGV550707H, NGV550930H, NGV517257H, NGV525346H, NGV528147H, NGV528385H, NGV550940H, NGV551578H, NGV551315H, NGV519716H, NGV519906H, NGV524845H, NGV528060H, NGV528467H, NGV529199H, NGV550790H, NGV529374H, NGV528272H, NGV528419H, NGV529644H, NGV550102H, NGV528195H, NGV528202H, NGV528203H, NGV530555H, NGV551249H, NGV551603H, NGV551967H, NGV552061H, NGV513416H, NGV515068H, NGV515200H, NGV522741H, NGV523042H, NGV523115H, NGV524835H, NGV526211H, NGV528034H, NGV527406H, NGV550429H, NGV518553H, NGV520393H, NGV527985H, NGV550411H, NGV550428H, NGV515033H, NGV517950H, NGV521879H, NGV523007H, NGV527891H, NGV550069H, NGV551507H, NGV519392H, NGV523606H, NGV516487H, NGV527392H, NGV522029H, NGV525989H, NGV528397H, NGV529449H, NGV550780H, NGV512987H, NGV513355H, NGV514574H, NGV517147H, NGV517148H, NGV517035H, NGV521680H, NGV526903H, NGV516512H, NGV521377H, NGV522262H, NGV551702H, NGV527809H, NGV530197H, NGV514299H, NGV514537H, NGV526377H, NGV520696H, NGV523772H, NGV519107H, NGV520378H, NGV528390H, NGV551010H, NGV509915H, NGV516410H, NGV517222H, NGV526080H, NGV513653H, NGV514452H, NGV519367H, NGV524482H, NGV526482H, NGV526542H, NGV527532H, NGV528959H, NGV521670H, NGV528797H, NGV529534H, NGV530202H, NGV529823H, NGV525406H, NGV527014H, NGV527394H, NGV527522H, NGV527534H, NGV529788H, NGV514082H, NGV516294H, NGV517378H, NGV519629H, NGV526507H, NGV526606H, NGV551032H, NGV519056H, NGV521923H, NGV525841H, NGV528366H, NGV530757H, NGV523043H, NGV528814H, NGV551497H, NGV525896H, NGV529504H, NGV524513H, NGV519733H, NGV518783H, NGV550853H, NGV551707H, NGV523141H, NGV523145H, NGV524422H, NGV525068H, NGV528533H, NGV514228H, NGV525813H, NGV527150H, NGV515509H, NGV516495H, NGV516527H, NGV520242H, NGV521825H, NGV521936H, NGV522642H, NGV525915H, NGV526356H, NGV526947H, NGV528806H, NGV529745H, NGV529750H, NGV552202H, NGV552212H, NGV552220H, NGV552224H, NGV552233H, NGV519375H, NGV523338H, NGV523850H, NGV527368H, NGV527409H, NGV551552H, NGV529236H, NGV522393H, NGV527571H, NGV530554H, NGV550666H, NGV551245H, NGV552046H, NGV523733H, NGV523738H, NGV528789H, NGV529628H, NGV529630H, NGV530701H, NGV530708H, NGV550516H, NGV551776H, NGV526193H, NGV529961H, NGV509737H, NGV519734H, NGV519735H, NGV526472H, NGV529543H, NGV551048H, NGV527048H, NGV527273H, NGV527645H, NGV528697H, NGV528867H, NGV529100H, NGV529325H, NGV550075H, NGV550974H, NGV551614H, NGV551573H, NGV526194H, NGV515939H, NGV550168H, NGV529083H, NGV527650H, NGV551017H, NGV514596H, NGV526865H, NGV530595H, NGV517227H, NGV529815H, NGV518224H, NGV527372H, NGV526402H, NGV514617H, NGV522617H, NGV522618H, NGV523093H, NGV517553H, NGV551879H, NGV551268H, NGV551312H, NGV523525H, NGV522869H, NGV527682H, NGV523532H, NGV550353H, NGV523996H, NGV513487H, NGV515738H, NGV516437H, NGV515204H, NGV515984H, NGV520236H, NGV514862H, NGV515048H, NGV514978H, NGV515953H, NGV514518H, NGV513797H, NGV518628H, NGV514949H, NGV514855H, NGV516544H, NGV525588H, NGV516250H, NGV523206H, NGV525375H, NGV514165H, NGV551221H, NGV515861H, NGV514036H, NGV516884H, NGV513466H, NGV519411H, NGV527231H, NGV528094H, NGV528832H, NGV529813H, NGV550503H, NGV518705H, NGV513067H, NGV517757H, NGV514559H, NGV514533H, NGV520745H, NGV522911H, NGV524039H, NGV529431H, NGV513908H, NGV513415H, NGV514330H, NGV526137H, NGV526326H, NGV519884H, NGV524004H, NGV529201H, NGV514840H, NGV550447H, NGV516880H, NGV519359H, NGV526577H, NGV513426H, NGV551113H, NGV515523H, NGV521242H, NGV526906H, NGV514121H, NGV524140H, NGV518300H, NGV520635H, NGV516252H, NGV514307H, NGV514310H, NGV513793H, NGV516255H, NGV517888H, NGV517892H, NGV519389H, NGV519849H, NGV519855H, NGV519861H, NGV523278H, NGV523334H, NGV523722H, NGV523756H, NGV524144H, NGV526161H, NGV520713H, NGV526354H, NGV527348H, NGV522855H, NGV527935H, NGV550181H, NGV509682H, NGV525865H, NGV513061H, NGV528211H, NGV527160H, NGV517742H, NGV520660H, NGV522619H, NGV513519H, NGV528518H, NGV524044H, NGV518919H, NGV518357H, NGV519937H, NGV521638H, NGV530149H, NGV518690H, NGV516336H, NGV515631H, NGV521367H, NGV550802H, NGV516269H, NGV520588H, NGV521737H, NGV523128H, NGV523866H, NGV550232H, NGV513819H, NGV526823H, NGV525928H, NGV516296H, NGV525949H, NGV529996H, NGV516407H, NGV516408H, NGV525849H, NGV518786H, NGV518788H, NGV526460H, NGV515387H, NGV514622H, NGV520049H, NGV523590H, NGV516739H, NGV527911H, NGV518349H, NGV518371H, NGV526481H, NGV550960H, NGV513709H, NGV550463H, NGV515113H, NGV521343H, NGV530510H, NGV514918H, NGV515150H, NGV525129H, NGV527551H, NGV530468H, NGV550269H, NGV551236H, NGV524312H, NGV514941H, NGV514945H, NGV523455H, NGV514803H, NGV524005H, NGV550771H, NGV514144H, NGV527388H, NGV529146H, NGV521414H; GTIN 00643169530126, Serial Numbers: NGV602706H, NGV602707H, NGV612436H, NGV602815H, NGV607268H, NGV604544H, NGV604552H, NGV604634H, NGV604640H, NGV604641H, NGV609482H, NGV609485H, NGV609491H, NGV609496H, NGV609507H, NGV609516H, NGV609529H, NGV609536H, NGV609944H, NGV609950H, NGV609953H, NGV609954H, NGV609957H, NGV609959H, NGV609971H, NGV609972H, NGV609974H, NGV609976H, NGV609977H, NGV609979H, NGV609985H, NGV609986H, NGV609991H, NGV609992H, NGV609998H, NGV610000H, NGV610001H, NGV610003H, NGV610009H, NGV610010H, NGV610014H, NGV610015H, NGV610019H, NGV610022H, NGV610026H, NGV610419H, NGV610420H, NGV610421H, NGV610444H, NGV610451H, NGV610460H, NGV610463H, NGV610469H, NGV610474H, NGV610475H, NGV610483H, NGV610497H, NGV610498H, NGV610501H, NGV610504H, NGV610509H, NGV610511H, NGV610512H, NGV610514H, NGV610517H, NGV610518H, NGV610519H, NGV610521H, NGV610527H, NGV610529H, NGV610530H, NGV610531H, NGV610532H, NGV610534H, NGV610536H, NGV610537H, NGV610538H, NGV610540H, NGV610541H, NGV610543H, NGV610544H, NGV610546H, NGV610548H, NGV610549H, NGV610553H, NGV610555H, NGV610560H, NGV610564H, NGV610567H, NGV610569H, NGV610573H, NGV610574H, NGV610575H, NGV610576H, NGV610577H, NGV610578H, NGV610579H, NGV610580H, NGV610581H, NGV610582H, NGV610587H, NGV610588H, NGV610591H, NGV610593H, NGV610595H, NGV610599H, NGV610603H, NGV610604H, NGV610608H, NGV610613H, NGV610690H, NGV610692H, NGV610693H, NGV610700H, NGV610701H, NGV610782H, NGV610783H, NGV610784H, NGV610785H, NGV611494H, NGV611496H, NGV611499H, NGV611520H, NGV611521H, NGV611541H, NGV611784H, NGV611794H, NGV611795H, NGV611796H, NGV611837H, NGV611852H, NGV611857H, NGV611858H, NGV611907H, NGV609060H, NGV609061H, NGV609062H, NGV609063H, NGV609064H, NGV609066H, NGV609059H, NGV602345H, NGV606211H, NGV607448H, NGV603121H, NGV604516H, NGV605284H, NGV605763H, NGV603131H, NGV604689H, NGV604511H, NGV552943H, NGV603127H, NGV602946H, NGV553604H, NGV601459H, NGV602425H, NGV603035H, NGV612857H, NGV612858H, NGV612866H, NGV608022H, NGV552607H, NGV554402H, NGV602222H, NGV602510H, NGV605233H, NGV605841H, NGV608927H, NGV611363H, NGV600927H, NGV602779H, NGV552761H, NGV554391H, NGV600007H, NGV600008H, NGV601326H, NGV601373H, NGV601775H, NGV601794H, NGV601798H, NGV602088H, NGV602152H, NGV602168H, NGV602421H, NGV602957H, NGV602990H, NGV603475H, NGV603690H, NGV603834H, NGV604497H, NGV604498H, NGV605675H, NGV605677H, NGV606077H, NGV606323H, NGV606432H, NGV608554H, NGV608578H, NGV608845H, NGV608920H, NGV609621H, NGV610215H, NGV610539H, NGV611042H, NGV611955H, NGV612821H, NGV612822H, NGV612823H, NGV612824H, NGV612825H, NGV612976H, NGV600515H, NGV602250H, NGV604012H, NGV604013H, NGV607354H, NGV609524H, NGV609989H, NGV604977H, NGV552600H, NGV552601H, NGV600517H, NGV600518H, NGV601505H, NGV607006H, NGV608093H, NGV608470H, NGV608472H, NGV608475H, NGV608476H, NGV612559H, NGV612561H, NGV612562H, NGV612563H, NGV612564H, NGV607144H, NGV610221H, NGV605738H, NGV554174H, NGV554427H, NGV554428H, NGV602091H, NGV605383H, NGV607397H, NGV608847H, NGV608849H, NGV608853H, NGV609417H, NGV609438H, NGV609441H, NGV610382H, NGV612102H, NGV600765H, NGV603497H, NGV601978H, NGV604041H, NGV613256H, NGV552927H, NGV553067H, NGV553237H, NGV553722H, NGV554414H, NGV600488H, NGV602426H, NGV604318H, NGV605711H, NGV605991H, NGV608576H, NGV608793H, NGV609181H, NGV609280H, NGV610360H, NGV611779H, NGV611781H, NGV613093H, NGV613187H, NGV600877H, NGV610293H, NGV612678H, NGV554418H, NGV600482H, NGV600519H, NGV554055H, NGV554179H, NGV554182H, NGV602026H, NGV602305H, NGV602804H, NGV603748H, NGV605231H, NGV606643H, NGV610949H, NGV553038H, NGV606321H, NGV608362H, NGV610624H, NGV613133H, NGV554356H, NGV600343H, NGV601933H, NGV602228H, NGV604161H, NGV604773H, NGV605334H, NGV606280H, NGV606679H, NGV606708H, NGV607551H, NGV608133H, NGV608811H, NGV552623H, NGV600158H, NGV600744H, NGV600975H, NGV601897H, NGV601898H, NGV602304H, NGV602321H, NGV602381H, NGV602882H, NGV603465H, NGV604032H, NGV605680H, NGV605681H, NGV606878H, NGV607172H, NGV607173H, NGV607309H, NGV607470H, NGV608441H, NGV609007H, NGV611269H, NGV612973H, NGV602970H, NGV600061H, NGV600299H, NGV606947H, NGV608834H, NGV608880H, NGV609265H, NGV612966H, NGV600674H, NGV607767H, NGV605556H, NGV606093H, NGV607241H, NGV607432H, NGV607800H, NGV609530H, NGV609534H, NGV610368H, NGV610806H, NGV612091H, NGV612164H, NGV612448H, NGV613020H, NGV553215H, NGV553675H, NGV553695H, NGV553712H, NGV554251H, NGV554276H, NGV600002H, NGV600134H, NGV600180H, NGV601732H, NGV602927H, NGV602950H, NGV604412H, NGV609472H, NGV609899H, NGV609901H, NGV610257H, NGV611490H, NGV612553H, NGV553186H, NGV600196H, NGV600520H, NGV602705H, NGV603290H, NGV608856H, NGV605280H, NGV607931H, NGV553152H, NGV553380H, NGV552637H, NGV552916H, NGV553323H, NGV600254H, NGV601737H, NGV601756H, NGV604527H, NGV608836H, NGV609267H, NGV609475H, NGV609868H, NGV611749H, NGV612456H, NGV600047H, NGV600986H, NGV608608H, NGV611599H, NGV612674H, NGV612740H, NGV602349H, NGV602632H, NGV603114H, NGV604384H, NGV604874H, NGV605015H, NGV605648H, NGV609820H, NGV552785H, NGV600616H, NGV600779H, NGV601053H, NGV601111H, NGV604150H, NGV605688H, NGV602891H, NGV605669H, NGV553681H, NGV600421H, NGV600638H, NGV602127H, NGV602981H, NGV603000H, NGV604559H, NGV605712H, NGV605725H, NGV606029H, NGV609241H, NGV553248H, NGV553355H, NGV553677H, NGV553699H, NGV554379H, NGV601696H, NGV602132H, NGV602366H, NGV603438H, NGV604988H, NGV604991H, NGV606756H, NGV609043H, NGV609744H, NGV610052H, NGV552932H, NGV554214H, NGV600252H, NGV601309H, NGV601403H, NGV601470H, NGV602170H, NGV602177H, NGV602781H, NGV602801H, NGV602943H, NGV604391H, NGV604392H, NGV604393H, NGV604410H, NGV604459H, NGV604679H, NGV604680H, NGV604877H, NGV604878H, NGV608854H, NGV608855H, NGV608862H, NGV552937H, NGV601883H, NGV602572H, NGV605716H, NGV607978H, NGV608883H, NGV600609H, NGV600611H, NGV601048H, NGV603761H, NGV605656H, NGV605916H, NGV606297H, NGV611780H, NGV611963H, NGV612609H, NGV603837H, NGV607406H, NGV602086H, NGV604162H, NGV604444H, NGV605158H, NGV605159H, NGV553648H, NGV553739H, NGV554291H, NGV600554H, NGV600555H, NGV601820H, NGV601821H, NGV604456H, NGV604513H, NGV606987H, NGV606988H, NGV610771H, NGV612957H, NGV554365H, NGV553670H, NGV602651H, NGV603926H, NGV552808H, NGV553116H, NGV606329H, NGV607366H, NGV610937H, NGV553673H, NGV605111H, NGV608653H, NGV609237H, NGV609594H, NGV610362H, NGV552717H, NGV553300H, NGV553425H, NGV553427H, NGV554142H, NGV600151H, NGV600787H, NGV600958H, NGV601491H, NGV601493H, NGV601494H, NGV601498H, NGV601500H, NGV601501H, NGV601575H, NGV601626H, NGV601628H, NGV605083H, NGV605086H, NGV605090H, NGV605093H, NGV605095H, NGV607912H, NGV601856H, NGV604730H, NGV605303H, NGV608495H, NGV610295H, NGV612109H, NGV612672H, NGV553117H, NGV553119H, NGV553273H, NGV553439H, NGV553442H, NGV600831H, NGV600917H, NGV602233H, NGV602249H, NGV604471H, NGV605174H, NGV605299H, NGV605300H, NGV605372H, NGV605379H, NGV605657H, NGV606721H, NGV607081H, NGV608024H, NGV608081H, NGV608618H, NGV608681H, NGV609299H, NGV610071H, NGV611417H, NGV612067H, NGV553669H, NGV600000H, NGV600373H, NGV600374H, NGV600623H, NGV603306H, NGV606327H, NGV608625H, NGV608636H, NGV608637H, NGV610272H, NGV610281H, NGV610297H, NGV612930H, NGV600037H, NGV602060H, NGV602255H, NGV602819H, NGV604305H, NGV606445H, NGV610280H, NGV610285H, NGV612105H, NGV612122H, NGV600326H, NGV600410H, NGV600411H, NGV610973H, NGV604843H, NGV553360H, NGV553388H, NGV554441H, NGV600204H, NGV600209H, NGV600453H, NGV600778H, NGV601804H, NGV602121H, NGV602206H, NGV604407H, NGV604633H, NGV605635H, NGV606025H, NGV606522H, NGV606646H, NGV606718H, NGV607102H, NGV607190H, NGV607780H, NGV608095H, NGV609430H, NGV609741H, NGV610347H, NGV613083H, NGV613443H, NGV552639H, NGV552739H, NGV552839H, NGV554273H, NGV601765H, NGV601869H, NGV604115H, NGV606118H, NGV607872H, NGV607967H, NGV610768H, NGV610769H, NGV610770H, NGV612572H, NGV612580H, NGV612591H, NGV612726H, NGV610023H, NGV552563H, NGV552779H, NGV552973H, NGV553072H, NGV553247H, NGV600471H, NGV600472H, NGV603528H, NGV604942H, NGV604943H, NGV605075H, NGV605424H, NGV606742H, NGV612555H, NGV612836H, NGV553378H, NGV553735H, NGV604905H, NGV604912H, NGV605524H, NGV612602H, NGV553204H, NGV553635H, NGV554364H, NGV600017H, NGV600041H, NGV600042H, NGV600449H, NGV600451H, NGV600452H, NGV604860H, NGV606125H, NGV606127H, NGV607148H, NGV607184H, NGV608890H, NGV608900H, NGV608901H, NGV608902H, NGV609885H, NGV602664H, NGV602887H, NGV603011H, NGV607108H, NGV608302H, NGV608311H, NGV608767H, NGV552784H, NGV553417H, NGV553453H, NGV600331H, NGV600959H, NGV601161H, NGV601872H, NGV602118H, NGV602364H, NGV603885H, NGV604127H, NGV604129H, NGV605270H, NGV605840H, NGV607329H, NGV607433H, NGV607954H, NGV605445H, NGV554060H, NGV607617H, NGV554079H, NGV600336H, NGV602070H, NGV604763H, NGV605993H, NGV611444H, NGV554203H, NGV554383H, NGV600296H, NGV600559H, NGV600706H, NGV601896H, NGV602423H, NGV602928H, NGV603148H, NGV604414H, NGV604996H, NGV605123H, NGV607408H, NGV607806H, NGV607828H, NGV610669H, NGV611434H, NGV613102H, NGV552714H, NGV553676H, NGV554125H, NGV600885H, NGV602095H, NGV605361H, NGV605668H, NGV606890H, NGV612859H, NGV553265H, NGV605976H, NGV612940H, NGV604575H, NGV606084H, NGV613431H, NGV552389H, NGV553032H, NGV600362H, NGV600386H, NGV600756H, NGV602105H, NGV602291H, NGV602410H, NGV603282H, NGV603283H, NGV603291H, NGV604360H, NGV604525H, NGV605674H, NGV606295H, NGV606654H, NGV606727H, NGV607333H, NGV607334H, NGV607337H, NGV607420H, NGV607813H, NGV607820H, NGV607890H, NGV608182H, NGV609499H, NGV610166H, NGV610401H, NGV610640H, NGV613068H, NGV554159H, NGV601860H, NGV603304H, NGV604492H, NGV605200H, NGV610034H, NGV610053H, NGV553615H, NGV553618H, NGV554248H, NGV554264H, NGV600678H, NGV601975H, NGV602770H, NGV602840H, NGV602977H, NGV606955H, NGV607015H, NGV607252H, NGV607615H, NGV608509H, NGV609194H, NGV609200H, NGV609469H, NGV609526H, NGV611040H, NGV611203H, NGV612635H, NGV612849H, NGV613375H, NGV554326H, NGV601918H, NGV612031H, NGV552663H, NGV602138H, NGV609844H, NGV605952H, NGV608527H, NGV608600H, NGV552657H, NGV608683H, NGV600595H, NGV609561H, NGV603904H, NGV605926H, NGV607846H, NGV608383H, NGV611685H, NGV553377H, NGV600532H.NGV600686H, NGV600911H, NGV601069H, NGV601802H, NGV601983H, NGV602821H, NGV604096H, NGV607443H, NGV610081H, NGV610811H, NGV612715H, NGV612723H, NGV552794H, NGV552802H, NGV552892H, NGV554172H, NGV600621H, NGV601950H, NGV602158H, NGV602416H, NGV605222H, NGV605224H, NGV605234H, NGV606112H, NGV606152H, NGV606723H, NGV606874H, NGV608512H, NGV610287H, NGV610288H, NGV610332H, NGV611029H, NGV611812H, NGV612656H, NGV612657H, NGV612688H, NGV608078H, NGV609174H, NGV610801H, NGV552952H, NGV553230H, NGV600287H, NGV603927H, NGV605031H, NGV605042H, NGV608366H, NGV611404H, NGV610815H, NGV552685H, NGV553124H, NGV601353H, NGV601361H, NGV601925H, NGV603245H, NGV605550H, NGV605552H, NGV610775H, NGV611804H, NGV612584H, NGV600435H, NGV609728H, NGV554346H, NGV604940H, NGV612459H, NGV605322H, NGV607191H, NGV607909H, NGV553368H, NGV554131H, NGV602568H, NGV608796H, NGV609429H, NGV611816H, NGV612522H, NGV612651H, NGV612658H, NGV603133H, NGV608008H, NGV600677H, NGV602516H, NGV600715H, NGV606000H, NGV608788H, NGV608807H, NGV610717H, NGV553652H, NGV601819H, NGV602343H, NGV611573H, NGV600268H, NGV602167H, NGV604321H, NGV604581H, NGV604973H, NGV609071H, NGV611803H, NGV612856H, NGV553079H, NGV600855H, NGV606315H, NGV609879H, NGV552575H, NGV552591H, NGV552596H, NGV552597H, NGV600694H, NGV601022H, NGV601024H, NGV601025H, NGV601034H, NGV601848H, NGV601849H, NGV607388H, NGV607447H, NGV608694H, NGV608695H, NGV608696H, NGV608697H, NGV608699H, NGV608700H, NGV608804H, NGV612262H, NGV612596H, NGV612597H, NGV612806H, NGV612903H, NGV612909H, NGV612910H, NGV613112H, NGV605510H, NGV611753H, NGV612481H, NGV554304H, NGV602360H, NGV603704H, NGV607458H, NGV607460H, NGV607464H, NGV609916H, NGV611972H, NGV608714H, NGV610202H, NGV553115H, NGV606342H, NGV607441H, NGV609250H, NGV604935H, NGV601258H, NGV602839H, NGV612023H, NGV553296H, NGV554154H, NGV600978H, NGV604109H, NGV604619H, NGV611210H, NGV553029H, NGV600281H, NGV610976H, NGV610381H, NGV602914H, NGV602983H, NGV603030H, NGV606111H, NGV606494H, NGV606536H, NGV607322H, NGV607459H, NGV600957H, NGV602653H, NGV603253H, NGV605593H, NGV606594H, NGV608001H, NGV608682H, NGV609838H, NGV609921H, NGV610309H, NGV611232H, NGV612041H, NGV552954H, NGV553408H, NGV553414H, NGV600096H, NGV600403H, NGV600636H, NGV601766H, NGV601855H, NGV602904H, NGV603340H, NGV603492H, NGV603999H, NGV604001H, NGV604367H, NGV607370H, NGV607423H, NGV607764H, NGV608088H, NGV609978H, NGV610061H, NGV610990H, NGV612489H, NGV612754H, NGV612981H, NGV613119H, NGV553562H, NGV606426H, NGV608118H, NGV610667H, NGV606352H, NGV605282H, NGV600332H, NGV600564H, NGV601428H, NGV603043H, NGV603770H, NGV604913H, NGV606703H, NGV607630H, NGV607661H, NGV609592H, NGV610750H, NGV610751H, NGV611747H, NGV611776H, NGV611800H, NGV553611H, NGV554118H, NGV554160H, NGV602317H, NGV605887H, NGV612292H, NGV613270H, NGV554328H, NGV603272H, NGV609474H, NGV609022H, NGV553686H, NGV553687H, NGV554162H, NGV600269H, NGV601476H, NGV601795H, NGV604085H, NGV606459H, NGV606462H, NGV607186H, NGV608140H, NGV610298H, NGV610962H, NGV611652H, NGV611740H, NGV612128H, NGV552862H, NGV553391H, NGV600819H, NGV601606H, NGV601769H, NGV602963H, NGV603714H, NGV603912H, NGV604073H, NGV606748H, NGV607074H, NGV609296H, NGV610130H, NGV610343H, NGV602107H, NGV605924H, NGV554202H, NGV600251H, NGV602399H, NGV606102H, NGV610384H, NGV611151H, NGV612571H, NGV553768H, NGV602189H, NGV602192H, NGV602624H, NGV604060H, NGV605005H, NGV608002H, NGV608325H, NGV609664H, NGV609828H, NGV609929H, NGV610170H, NGV610175H, NGV610178H, NGV610954H, NGV552809H, NGV608372H, NGV610947H, NGV609946H, NGV553177H, NGV552664H, NGV552796H, NGV552801H, NGV601348H, NGV607128H, NGV610708H, NGV612074H, NGV553245H, NGV600133H, NGV600632H, NGV552837H, NGV554188H, NGV600470H, NGV601311H, NGV601888H, NGV603096H, NGV603231H, NGV604766H, NGV605631H, NGV606302H, NGV607130H, NGV608251H, NGV608759H, NGV609949H, NGV611569H, NGV612080H, NGV612895H, NGV613300H, NGV607870H, NGV610661H, NGV611887H, NGV612684H, NGV553251H, NGV600603H, NGV605081H, NGV609320H, NGV610956H, NGV611574H, NGV612960H, NGV612984H, NGV613148H, NGV601778H, NGV607122H, NGV552825H, NGV553145H, NGV605609H, NGV552831H, NGV600578H, NGV601449H, NGV602277H, NGV602539H, NGV603781H, NGV604018H, NGV604753H, NGV605183H, NGV605184H, NGV606294H, NGV606532H, NGV607976H, NGV608970H, NGV611428H, NGV611898H, NGV612037H, NGV612230H, NGV552641H, NGV553149H, NGV602220H, NGV602221H, NGV602688H, NGV603301H, NGV604101H, NGV604508H, NGV605747H, NGV605930H, NGV605995H, NGV607430H, NGV607985H, NGV608066H, NGV610249H, NGV610250H, NGV611464H, NGV612581H, NGV553753H, NGV601578H, NGV601587H, NGV604682H, NGV605856H, NGV608555H, NGV609805H, NGV612370H, NGV602905H, NGV606333H, NGV608304H, NGV611442H, NGV612323H, NGV613050H, NGV613058H, NGV552678H, NGV552693H, NGV554398H, NGV600175H, NGV600651H, NGV601356H, NGV602665H, NGV602666H, NGV603702H, NGV603717H, NGV604790H, NGV606363H, NGV606458H, NGV606461H, NGV608102H, NGV608104H, NGV608105H, NGV608107H, NGV610212H, NGV610217H, NGV611437H, NGV611438H, NGV605519H, NGV607269H, NGV609051H, NGV609898H, NGV610135H, NGV611430H, NGV612950H, NGV554163H, NGV602845H, NGV603771H, NGV604153H, NGV605007H, NGV605996H, NGV606209H, NGV607248H, NGV600528H, NGV601773H, NGV602985H, NGV605626H, NGV609987H, NGV613193H, NGV600436H, NGV600662H, NGV602562H, NGV602925H, NGV603721H, NGV604928H, NGV605805H, NGV607115H, NGV608510H, NGV613201H, NGV553313H, NGV602446H, NGV612646H, NGV607776H, NGV605698H, NGV610171H, NGV612028H, NGV600708H, NGV552669H, NGV552709H, NGV554376H, NGV605824H, NGV609674H, NGV609679H, NGV612592H, NGV612989H, NGV552722H, NGV552769H, NGV606207H, NGV553193H, NGV554396H, NGV606762H, NGV607993H, NGV609965H, NGV552842H, NGV553176H, NGV554271H, NGV554272H, NGV600065H, NGV600859H, NGV600860H, NGV600861H, NGV601421H, NGV601422H, NGV602178H, NGV602184H, NGV602307H, NGV602308H, NGV602328H, NGV603268H, NGV603294H, NGV603295H, NGV603296H, NGV603297H, NGV603298H, NGV603764H, NGV603825H, NGV603827H, NGV603828H, NGV605002H, NGV605203H, NGV606226H, NGV606227H, NGV607736H, NGV608529H, NGV608950H, NGV608963H, NGV608964H, NGV610244H, NGV610291H, NGV610735H, NGV610948H, NGV610993H, NGV611060H, NGV611482H, NGV612236H, NGV612454H, NGV612906H, NGV613162H, NGV608782H, NGV611250H, NGV552580H, NGV552630H, NGV553160H, NGV553721H, NGV553747H, NGV600502H, NGV601427H, NGV602110H, NGV602369H, NGV604999H, NGV609584H, NGV611572H, NGV612538H, NGV613142H, NGV552688H, NGV553440H, NGV554095H, NGV554311H, NGV601455H, NGV602449H, NGV602789H, NGV603978H, NGV604457H, NGV604472H, NGV604772H, NGV604775H, NGV605943H, NGV606216H, NGV607988H, NGV608211H, NGV610094H, NGV610338H, NGV610983H, NGV612834H, NGV552721H, NGV554338H, NGV553047H, NGV554318H, NGV600513H, NGV602546H, NGV610188H, NGV611766H, NGV600215H, NGV607227H, NGV609725H, NGV611218H, NGV611321H, NGV612560H, NGV612569H, NGV612570H, NGV552587H, NGV553386H, NGV600190H, NGV600973H, NGV601685H, NGV605337H, NGV606120H, NGV607225H, NGV608526H, NGV608619H, NGV609706H, NGV610012H, NGV604536H, NGV553445H, NGV554347H, NGV600090H, NGV600249H, NGV600602H, NGV601073H, NGV601286H, NGV601393H, NGV601439H, NGV601452H, NGV601979H, NGV602332H, NGV602965H, NGV604366H, NGV605120H, NGV605268H, NGV605278H, NGV605339H, NGV607142H, NGV607145H, NGV607445H, NGV608129H, NGV609509H, NGV609722H, NGV610120H, NGV610845H, NGV610945H, NGV612270H, NGV612652H, NGV612660H, NGV612843H, NGV612844H, NGV613066H, NGV606116H, NGV609140H, NGV602171H, NGV606239H, NGV609772H, NGV610087H, NGV612426H, NGV608042H, NGV552953H, NGV600587H, NGV600610H, NGV608734H, NGV609982H, NGV610406H, NGV600230H, NGV608303H, NGV608729H, NGV602141H, NGV602773H, NGV603932H, NGV603995H, NGV605958H, NGV606260H, NGV609708H, NGV610007H, NGV610164H, NGV610201H, NGV612598H, NGV606620H, NGV607593H, NGV611601H, NGV612682H, NGV604315H, NGV611354H, NGV552766H, NGV600600H, NGV600993H, NGV601306H, NGV602269H, NGV604801H, NGV608830H, NGV553290H, NGV606402H, NGV607187H, NGV607914H, NGV612883H, NGV605784H, NGV554374H, NGV600072H, NGV602527H, NGV603997H, NGV604961H, NGV605602H, NGV605765H, NGV607235H, NGV607238H, NGV608039H, NGV608153H, NGV608225H, NGV609966H, NGV600086H, NGV602333H, NGV603070H, NGV603154H, NGV605365H, NGV606395H, NGV607907H, NGV608445H, NGV605640H, NGV554416H, NGV554319H, NGV604047H, NGV606875H, NGV552939H, NGV554066H, NGV601627H, NGV605653H, NGV605654H, NGV613047H, NGV605953H, NGV600751H, NGV600873H, NGV603975H, NGV605024H, NGV606440H, NGV607557H, NGV611286H, NGV602267H, NGV603200H, NGV604936H, NGV605703H, NGV611765H, NGV605262H, NGV611280H, NGV554399H, NGV600433H, NGV600987H, NGV607114H, NGV602213H, NGV604712H, NGV605286H, NGV607897H, NGV602675H, NGV610405H, NGV601845H, NGV603091H, NGV602356H, NGV607991H, NGV610626H, NGV611276H, NGV611755H, NGV612697H, NGV612703H, NGV612998H, NGV605967H, NGV609709H, NGV553084H, NGV553098H, NGV553282H, NGV553327H, NGV553576H, NGV553591H, NGV554297H, NGV600032H, NGV601366H, NGV601367H, NGV601529H, NGV602224H, NGV602259H, NGV603462H, NGV604917H, NGV604948H, NGV605289H, NGV607204H, NGV553704H, NGV606243H, NGV553631H, NGV553367H, NGV601866H, NGV610267H, NGV554138H, NGV600182H, NGV600630H, NGV601327H, NGV601876H, NGV603237H, NGV603348H, NGV603910H, NGV604524H, NGV606351H, NGV606361H, NGV606433H, NGV607968H, NGV552912H, NGV602436H, NGV554137H, NGV554153H, NGV600874H, NGV603594H, NGV605293H, NGV605791H, NGV606042H, NGV607884H, NGV608299H, NGV608481H, NGV610391H, NGV601030H, NGV600257H, NGV600416H, NGV600870H, NGV602855H, NGV604415H, NGV606074H, NGV608610H, NGV608870H, NGV609046H, NGV611240H, NGV612772H, NGV554380H, NGV554381H, NGV608139H, NGV603057H, NGV607213H, NGV553607H, NGV600161H, NGV602156H, NGV602157H, NGV602545H, NGV602554H, NGV607012H, NGV607017H, NGV607159H, NGV607891H, NGV609432H, NGV609433H, NGV610138H, NGV610186H, NGV554074H, NGV600062H, NGV600402H, NGV602481H, NGV604386H, NGV605432H, NGV603872H, NGV611806H, NGV602093H, NGV602724H, NGV603337H, NGV605441H, NGV605798H, NGV607885H, NGV609458H, NGV553095H, NGV553263H, NGV600725H, NGV600736H, NGV602034H, NGV602658H, NGV605104H, NGV607117H, NGV609745H, NGV609748H, NGV610987H, NGV611079H, NGV612189H, NGV607226H, NGV610119H, NGV612969H, NGV553578H, NGV554424H, NGV604379H, NGV612535H, NGV600141H, NGV554106H, NGV609797H, NGV611344H, NGV552665H, NGV554355H, NGV604112H, NGV610129H, NGV600817H, NGV604110H, NGV553403H, NGV553555H, NGV553587H, NGV600760H, NGV601768H, NGV602219H, NGV602428H, NGV602915H, NGV603156H, NGV603157H, NGV604592H, NGV607013H, NGV608829H, NGV608860H, NGV610264H, NGV612414H, NGV554090H, NGV602978H, NGV603395H, NGV604160H, NGV605755H, NGV605760H, NGV605877H, NGV607233H, NGV609569H, NGV609842H, NGV610090H, NGV610112H, NGV612400H, NGV612450H, NGV608151H, NGV609461H, NGV552928H, NGV553085H, NGV553434H, NGV605457H, NGV605614H, NGV605861H, NGV606986H, NGV607906H, NGV607994H, NGV552556H, NGV552557H, NGV601270H, NGV601276H, NGV601314H, NGV602775H, NGV602992H, NGV603032H, NGV607446H, NGV608126H, NGV610686H, NGV611028H, NGV553261H, NGV554123H, NGV602575H, NGV605942H, NGV608023H, NGV610935H, NGV612173H, NGV613408H, NGV607228H, NGV552716H, NGV605623H, NGV610035H, NGV610369H, NGV611063H, NGV602753H, NGV602755H, NGV552676H, NGV552677H, NGV553343H, NGV553344H, NGV553345H, NGV553346H, NGV600891H, NGV600892H, NGV601474H, NGV601475H, NGV602234H, NGV602245H, NGV602528H, NGV602529H, NGV603082H, NGV603471H, NGV603472H, NGV603474H, NGV603684H, NGV603706H, NGV603710H, NGV603935H, NGV603938H, NGV603943H, NGV604111H, NGV604518H, NGV604869H, NGV604885H, NGV604886H, NGV605787H, NGV606792H, NGV606794H, NGV606800H, NGV609470H, NGV609479H, NGV610218H, NGV610609H, NGV610792H, NGV610794H, NGV610795H, NGV610804H, NGV610994H, NGV610999H, NGV611008H, NGV611917H, NGV611922H, NGV612826H, NGV612827H, NGV612828H, NGV600276H, NGV552901H, NGV552908H, NGV553086H, NGV553242H, NGV553291H, NGV601484H, NGV601895H, NGV602032H, NGV602106H, NGV602314H, NGV602654H, NGV602835H, NGV603316H, NGV603801H, NGV603802H, NGV604025H, NGV604865H, NGV604887H, NGV604929H, NGV605137H, NGV605589H, NGV606514H, NGV606922H, NGV607064H, NGV607126H, NGV607185H.NGV607374H, NGV607454H, NGV607740H, NGV608052H, NGV608103H, NGV608130H, NGV608131H, NGV608277H, NGV608772H, NGV609240H, NGV609278H, NGV609424H, NGV610140H, NGV610199H, NGV611714H, NGV611716H, NGV612162H, NGV612937H, NGV602837H, NGV611365H, NGV607255H, NGV604924H, NGV553697H, NGV554081H, NGV600344H, NGV600346H, NGV600437H, NGV602453H, NGV605686H, NGV607543H, NGV609328H, NGV612093H, NGV612896H, NGV613049H, NGV552917H, NGV600220H, NGV603258H, NGV603261H, NGV603262H, NGV605786H, NGV602208H, NGV606002H, NGV613039H, NGV601818H, NGV602252H, NGV602254H, NGV602263H, NGV602734H, NGV606094H, NGV606311H, NGV610637H, NGV612787H, NGV612970H, NGV608805H, NGV553174H, NGV553734H, NGV600324H, NGV600365H, NGV601360H, NGV601362H, NGV602745H, NGV603192H, NGV604971H, NGV604972H, NGV608749H, NGV608752H, NGV608753H, NGV613006H, NGV613015H, NGV613019H, NGV601495H, NGV601629H, NGV605087H, NGV608313H, NGV608322H, NGV609031H, NGV553133H, NGV553135H, NGV553139H, NGV600335H, NGV600728H, NGV600942H, NGV601059H, NGV601359H, NGV601420H, NGV602490H, NGV602765H, NGV603324H, NGV603371H, NGV603620H, NGV603712H, NGV604848H, NGV607002H, NGV607003H, NGV607174H, NGV608114H, NGV608115H, NGV608116H, NGV608649H, NGV608773H, NGV609731H, NGV609732H, NGV609733H, NGV609734H, NGV610126H, NGV610127H, NGV610274H, NGV611247H, NGV603896H, NGV606258H, NGV606535H, NGV612472H, NGV554084H, NGV600683H, NGV601947H, NGV603415H, NGV603530H, NGV604927H, NGV605706H, NGV606937H, NGV607859H, NGV608954H, NGV611557H, NGV612612H, NGV613299H, NGV554185H, NGV554288H, NGV601522H, NGV603287H, NGV603659H, NGV603660H, NGV604844H, NGV605468H, NGV605757H, NGV605892H, NGV606490H, NGV607163H, NGV608082H, NGV608346H, NGV610075H, NGV610165H, NGV612467H, NGV612929H, NGV602646H, NGV604132H, NGV606088H, NGV608791H, NGV600675H, NGV602133H, NGV603981H, NGV553423H, NGV553708H, NGV552855H, NGV554337H, NGV602003H, NGV607107H, NGV552848H, NGV552940H, NGV553278H, NGV554194H, NGV601733H, NGV553384H, NGV553728H, NGV600398H, NGV600710H, NGV600909H, NGV601310H, NGV602313H, NGV602315H, NGV603468H, NGV603470H, NGV603811H, NGV604661H, NGV606137H, NGV606142H, NGV607523H, NGV607882H, NGV607887H, NGV609260H, NGV609284H, NGV609513H, NGV609518H, NGV612253H, NGV612267H, NGV553167H, NGV609767H, NGV610859H, NGV602581H, NGV600371H, NGV603970H, NGV553019H, NGV553663H, NGV554201H, NGV600213H, NGV603959H, NGV552976H, NGV553443H, NGV600071H, NGV600318H, NGV600354H, NGV600617H, NGV602841H, NGV603099H, NGV603278H, NGV605822H, NGV606205H, NGV609511H, NGV611309H, NGV611311H, NGV611343H, NGV607754H, NGV605782H, NGV608823H, NGV612056H, NGV612419H, NGV553627H, NGV600625H, NGV603998H, NGV605511H, NGV605580H, NGV606006H, NGV606007H, NGV607706H, NGV608850H, NGV609418H, NGV612251H, NGV600250H, NGV601585H, NGV603990H, NGV552821H, NGV552924H, NGV553259H, NGV553389H, NGV554443H, NGV554450H, NGV600084H, NGV600181H, NGV600291H, NGV600573H, NGV600764H, NGV600966H, NGV600998H, NGV601026H, NGV601027H, NGV602394H, NGV607945H, NGV608295H, NGV608348H, NGV610025H, NGV602475H, NGV552691H, NGV552774H, NGV600284H, NGV604086H, NGV604786H, NGV605615H, NGV605937H, NGV606662H, NGV607793H, NGV608087H, NGV608298H, NGV610346H, NGV552650H, NGV554448H, NGV600395H, NGV600635H, NGV601875H, NGV602374H, NGV602455H, NGV602489H, NGV603616H, NGV605319H, NGV605914H, NGV605929H, NGV606334H, NGV606335H, NGV606904H, NGV608579H, NGV608792H, NGV609032H, NGV609756H, NGV610312H, NGV613060H, NGV613197H, NGV611979H, NGV607942H, NGV607943H, NGV612111H, NGV612115H, NGV612136H, NGV602824H, NGV554417H, NGV608851H, NGV553106H, NGV553709H, NGV554429H, NGV602297H, NGV603110H, NGV606610H, NGV607247H, NGV553326H, NGV602246H, NGV600186H, NGV606290H, NGV610820H, NGV604033H, NGV605138H, NGV606702H, NGV607874H, NGV608959H, NGV609279H, NGV553317H, NGV600835H, NGV553717H, NGV610172H, NGV553083H, NGV600593H, NGV600664H, NGV600739H, NGV600740H, NGV601825H, NGV601965H, NGV601966H, NGV602352H, NGV602731H, NGV602823H, NGV602828H, NGV604043H, NGV604045H, NGV605799H, NGV606248H, NGV606285H, NGV606286H, NGV607496H, NGV607602H, NGV609248H, NGV610027H, NGV610260H, NGV610286H, NGV610402H, NGV611144H, NGV611788H, NGV611971H, NGV613062H, NGV607746H, NGV552921H, NGV553636H, NGV600379H, NGV600866H, NGV602680H, NGV605767H, NGV607011H, NGV608315H, NGV602414H, NGV600660H, NGV604833H, NGV610720H, NGV607607H, NGV607973H, NGV609287H, NGV552606H, NGV552734H, NGV553744H, NGV601823H, NGV601858H, NGV605643H, NGV606382H, NGV606385H, NGV606428H, NGV609302H, NGV609316H, NGV610222H, NGV610749H, NGV611076H, NGV611080H, NGV611396H, NGV553320H, NGV553573H, NGV601993H, NGV604147H, NGV606337H, NGV609665H, NGV612148H, NGV612548H, NGV554139H, NGV554140H, NGV552656H, NGV552674H, NGV552726H, NGV553142H, NGV600290H, NGV602210H, NGV602320H, NGV602324H, NGV603381H, NGV606683H, NGV610253H, NGV611576H, NGV611624H, NGV554198H, NGV600903H, NGV602341H, NGV605102H, NGV552567H, NGV552571H, NGV552618H, NGV553076H, NGV554312H, NGV554373H, NGV554393H, NGV554395H, NGV609597H, NGV553148H, NGV608076H, NGV552747H, NGV600006H, NGV600490H, NGV600596H, NGV602365H, NGV603286H, NGV603292H, NGV605746H, NGV609139H, NGV612603H, NGV613253H, NGV606068H, NGV553080H, NGV604750H, NGV607692H, NGV611381H, NGV612064H, NGV600207H, NGV600246H, NGV600247H, NGV600652H, NGV601307H, NGV601392H, NGV606645H, NGV608273H, NGV608638H, NGV608639H, NGV608640H, NGV608642H, NGV608644H, NGV608645H, NGV552696H, NGV553694H, NGV553748H, NGV600277H, NGV600553H, NGV600565H, NGV600598H, NGV600729H, NGV553170H, NGV602805H, NGV604728H, NGV612144H, NGV600035H, NGV600223H, NGV609494H, NGV552786H, NGV553379H, NGV600279H, NGV604108H, NGV612913H, NGV552680H, NGV554314H, NGV601540H, NGV601698H, NGV601764H, NGV602692H, NGV604807H, NGV605515H, NGV606602H, NGV607956H, NGV608833H, NGV609026H, NGV612873H, NGV608537H, NGV608541H, NGV610345H, NGV552911H, NGV600735H, NGV600851H, NGV603964H, NGV605966H, NGV605999H, NGV607901H, NGV609712H, NGV610004H, NGV610831H, NGV611156H, NGV611560H, NGV552588H, NGV552589H, NGV553285H, NGV553286H, NGV553372H, NGV553450H, NGV600718H, NGV611174H, NGV601885H, NGV602866H, NGV608111H, NGV608343H, NGV609715H, NGV611072H, NGV612077H, NGV552822H, NGV603756H, NGV604847H, NGV602200H, NGV604914H, NGV605768H, NGV606130H, NGV606726H, NGV607738H, NGV608029H, NGV608421H, NGV608497H, NGV612168H, NGV613005H, NGV606942H, NGV611754H, NGV612838H, NGV553446H, NGV554119H, NGV602647H, NGV603116H, NGV603703H, NGV609118H, NGV610317H, NGV610598H, NGV611357H, NGV611382H, NGV553147H, NGV609872H, NGV612540H, NGV601791H, NGV609505H, NGV603725H, NGV610203H, NGV554144H, NGV554147H, NGV600847H, NGV601782H, NGV602205H, NGV602676H, NGV603039H, NGV608407H, NGV608414H, NGV552555H, NGV553102H, NGV553394H, NGV553580H, NGV553726H, NGV600297H, NGV600312H, NGV600777H, NGV601810H, NGV602447H, NGV603222H, NGV603709H, NGV604555H, NGV605659H, NGV606070H, NGV606452H, NGV606720H, NGV608865H, NGV609300H, NGV609864H, NGV610342H, NGV610710H, NGV612207H, NGV612432H, NGV613184H, NGV554196H, NGV554275H, NGV600372H, NGV600419H, NGV600424H, NGV601839H, NGV601841H, NGV601851H, NGV601852H, NGV601853H, NGV603052H, NGV603574H, NGV607395H, NGV607398H, NGV607516H, NGV607879H, NGV607922H, NGV611322H, NGV611326H, NGV611335H, NGV602111H, NGV607132H, NGV610364H, NGV605318H, NGV552777H, NGV553240H, NGV604512H, NGV604577H, NGV605825H, NGV607794H, NGV610380H, NGV610662H, NGV605697H, NGV552620H, NGV552666H, NGV552772H, NGV553723H, NGV554218H, NGV600983H, NGV600995H, NGV601758H, NGV602515H, NGV604503H, NGV605835H, NGV606014H, NGV606420H, NGV606628H, NGV606629H, NGV608958H, NGV610713H, NGV610864H, NGV611342H, NGV611695H, NGV612557H, NGV553070H, NGV553136H, NGV553137H, NGV553150H, NGV554102H, NGV554434H, NGV600185H, NGV600439H, NGV602099H, NGV602448H, NGV604377H, NGV604677H, NGV604888H, NGV604894H, NGV605442H, NGV605920H, NGV606318H, NGV606425H, NGV607155H, NGV607983H, NGV608723H, NGV608725H, NGV608727H, NGV611329H, NGV611355H, NGV611718H, NGV613212H, NGV613222H, NGV613231H, NGV552608H, NGV553217H, NGV553334H, NGV553412H, NGV553589H, NGV554068H, NGV554101H, NGV554397H, NGV600388H, NGV600586H, NGV601301H, NGV601302H, NGV601450H, NGV601929H, NGV601963H, NGV602743H, NGV602939H, NGV603092H, NGV603135H, NGV603144H, NGV603149H, NGV603305H, NGV603561H, NGV603570H, NGV603676H, NGV605110H, NGV605628H, NGV605872H, NGV606312H, NGV606380H, NGV606434H, NGV606437H, NGV606623H, NGV607134H, NGV607139H, NGV607829H, NGV608071H, NGV608089H, NGV608955H, NGV609148H, NGV609462H, NGV609492H, NGV609823H, NGV610042H, NGV610116H, NGV610131H, NGV610818H, NGV611121H, NGV612132H, NGV612579H, NGV612829H, NGV613131H, NGV552920H, NGV600098H, NGV600684H, NGV602006H, NGV602174H, NGV602344H, NGV603123H, NGV603510H, NGV605951H, NGV606530H, NGV607344H, NGV609455H, NGV609994H, NGV553625H, NGV600053H, NGV600262H, NGV601807H, NGV602367H, NGV602406H, NGV602782H, NGV602862H, NGV603538H, NGV603557H, NGV604055H, NGV604805H, NGV605489H, NGV606485H, NGV607019H, NGV608329H, NGV608918H, NGV610141H, NGV610520H, NGV610652H, NGV612874H, NGV613455H, NGV553711H, NGV601893H, NGV602574H, NGV610966H, NGV552561H, NGV552582H, NGV553200H, NGV553202H, NGV600154H, NGV602075H, NGV602372H, NGV602607H, NGV602608H, NGV602609H, NGV602610H, NGV602642H, NGV603780H, NGV603803H, NGV603947H, NGV604049H, NGV604939H, NGV605809H, NGV608609H, NGV609450H, NGV609452H, NGV609460H, NGV610002H, NGV610006H, NGV611189H, NGV611261H, NGV611262H, NGV611263H, NGV611313H, NGV611314H, NGV611317H, NGV553114H, NGV602289H, NGV604477H, NGV607434H, NGV608680H, NGV612033H, NGV612100H, NGV612939H, NGV604660H, NGV605306H, NGV606624H, NGV606625H, NGV552609H, NGV601264H, NGV601354H, NGV610076H, NGV552858H, NGV552859H, NGV552860H, NGV552865H, NGV552866H, NGV552867H, NGV553641H, NGV553643H, NGV553644H, NGV553645H, NGV600020H, NGV600045H, NGV600046H, NGV600095H, NGV600146H, NGV600148H, NGV600167H, NGV600198H, NGV600201H, NGV600428H, NGV601054H, NGV601055H, NGV601056H, NGV601057H, NGV601090H, NGV601092H, NGV601093H, NGV602194H, NGV602195H, NGV602196H, NGV602279H, NGV602281H, NGV602282H, NGV602283H, NGV602284H, NGV602285H, NGV602286H, NGV602287H, NGV602309H, NGV602311H, NGV602623H, NGV602694H, NGV602791H, NGV602794H, NGV603624H, NGV603625H, NGV604454H, NGV604455H, NGV604464H, NGV604872H, NGV604873H, NGV604882H, NGV604883H, NGV606513H, NGV606515H, NGV606516H, NGV606746H, NGV607098H, NGV607099H, NGV607112H, NGV607307H, NGV607520H, NGV607526H, NGV607693H, NGV607694H, NGV607696H, NGV607697H, NGV607698H, NGV607699H, NGV608384H, NGV608389H, NGV608415H, NGV609024H, NGV609443H, NGV609447H, NGV610950H, NGV612106H, NGV612182H, NGV612685H, NGV612912H, NGV613074H, NGV613075H, NGV613078H, NGV613079H, NGV613107H, NGV613178H, NGV613190H, NGV613208H, NGV602679H, NGV604071H, NGV604300H, NGV605204H, NGV605290H, NGV606755H, NGV608272H, NGV609018H, NGV611594H, NGV612471H, NGV613105H, NGV554187H, NGV601429H, NGV602493H, NGV602682H, NGV604769H, NGV605326H, NGV607116H, NGV608112H, NGV610016H, NGV610089H, NGV610093H, NGV610682H, NGV611032H, NGV611051H, NGV611598H, NGV613160H, NGV552697H, NGV600480H, NGV600485H, NGV600503H, NGV601341H, NGV603842H, NGV603958H, NGV606416H, NGV606418H, NGV607315H, NGV608774H, NGV611483H, NGV611484H, NGV612641H, NGV612643H, NGV552904H, NGV553682H, NGV600327H, NGV601567H, NGV605919H, NGV607987H, NGV609036H, NGV609038H, NGV611409H, NGV611426H, NGV552724H, NGV600612H, NGV600871H, NGV602890H, NGV605026H, NGV605027H, NGV605030H, NGV607680H, NGV608158H, NGV611472H, NGV612647H, NGV612713H, NGV612730H, NGV612731H, NGV553211H, NGV553212H, NGV553214H, NGV554324H, NGV602080H, NGV602629H, NGV604762H, NGV605475H, NGV605478H, NGV605484H, NGV605486H, NGV608161H, NGV608162H, NGV608164H, NGV608671H, NGV608751H, NGV608757H, NGV608768H, NGV605285H, NGV610397H, NGV610773H, NGV613053H, NGV600143H, NGV602102H, NGV602684H, NGV602690H, NGV602756H, NGV602758H, NGV603279H, NGV603991H, NGV604686H, NGV604702H, NGV604703H, NGV607313H, NGV607316H, NGV607977H, NGV608835H, NGV608837H, NGV609467H, NGV609833H, NGV609834H, NGV609835H, NGV609849H, NGV609860H, NGV609862H, NGV609865H, NGV612983H, NGV600984H, NGV600991H.NGV605595H, NGV552662H, NGV552826H, NGV553218H, NGV553401H, NGV554145H, NGV554313H, NGV600081H, NGV600139H, NGV600214H, NGV600476H, NGV600620H, NGV600716H, NGV600965H, NGV601413H, NGV602154H, NGV602547H, NGV602742H, NGV602751H, NGV602883H, NGV603903H, NGV603916H, NGV604747H, NGV604770H, NGV604812H, NGV604863H, NGV605291H, NGV605666H, NGV605808H, NGV605859H, NGV605860H, NGV606632H, NGV606713H, NGV607151H, NGV607730H, NGV608269H, NGV608523H, NGV608583H, NGV609019H, NGV610242H, NGV610644H, NGV611163H, NGV611191H, NGV611786H, NGV611798H, NGV613106H, NGV600368H, NGV600676H, NGV601391H, NGV601394H, NGV601410H, NGV601412H, NGV605426H, NGV605435H, NGV606389H, NGV606391H, NGV606394H, NGV606396H, NGV606398H, NGV606408H, NGV606411H, NGV611590H, NGV613120H, NGV613122H, NGV613123H, NGV613126H, NGV613132H, NGV613137H, NGV613138H, NGV613147H, NGV601953H, NGV605927H, NGV606181H, NGV606441H, NGV606738H, NGV607797H, NGV552836H, NGV553586H, NGV554242H, NGV600352H, NGV603179H, NGV604099H, NGV604604H, NGV607805H, NGV610417H, NGV610542H, NGV610941H, NGV612649H, NGV605461H, NGV607980H, NGV610728H, NGV611270H, NGV613088H, NGV552710H, NGV552711H, NGV553216H, NGV553231H, NGV554184H, NGV600719H, NGV600730H, NGV601779H, NGV601786H, NGV602424H, NGV602558H, NGV605553H, NGV606328H, NGV606967H, NGV607921H, NGV609723H, NGV610245H, NGV611047H, NGV611811H, NGV612220H, NGV612331H, NGV612587H, NGV612590H, NGV552748H, NGV552752H, NGV553362H, NGV553426H, NGV553619H, NGV554093H, NGV554212H, NGV554305H, NGV600682H, NGV601516H, NGV602312H, NGV602395H, NGV602396H, NGV602412H, NGV603234H, NGV603235H, NGV603398H, NGV604963H, NGV604964H, NGV604965H, NGV604966H, NGV604968H, NGV604969H, NGV606469H, NGV606470H, NGV606618H, NGV607088H, NGV607427H, NGV607722H, NGV607728H, NGV608548H, NGV608561H, NGV608761H, NGV608763H, NGV608769H, NGV608770H, NGV611086H, NGV611137H, NGV611734H, NGV611735H, NGV611757H, NGV611758H, NGV612492H, NGV612582H, NGV612735H, NGV613246H, NGV554257H, NGV554266H, NGV554332H, NGV600030H, NGV602340H, NGV603993H, NGV606915H, NGV606923H, NGV609962H, NGV613035H, NGV613272H, NGV552938H, NGV607529H, NGV608846H, NGV609671H, NGV611742H, NGV612393H, NGV607785H, NGV553103H, NGV554306H, NGV605177H, NGV605979H, NGV606438H, NGV607325H, NGV607498H, NGV554255H, NGV600210H, NGV600444H, NGV600445H, NGV602013H, NGV602378H, NGV602379H, NGV602495H, NGV602810H, NGV603239H, NGV603915H, NGV604688H, NGV604944H, NGV606484H, NGV606694H, NGV607466H, NGV607769H, NGV607866H, NGV608274H, NGV608373H, NGV608500H, NGV608754H, NGV610237H, NGV611567H, NGV612165H, NGV612608H, NGV553626H, NGV601432H, NGV554071H, NGV602993H, NGV553416H, NGV553647H, NGV608035H, NGV553208H, NGV602334H, NGV603271H, NGV611387H, NGV612396H, NGV553288H, NGV554357H, NGV600333H, NGV600334H, NGV600960H, NGV601013H, NGV607007H, NGV608567H, NGV608571H, NGV608573H, NGV608584H, NGV608594H, NGV610813H, NGV611411H, NGV611418H, NGV611422H, NGV552728H, NGV553621H, NGV605639H, NGV609291H, NGV604421H, NGV613248H, NGV553314H, NGV553679H, NGV600724H, NGV601577H, NGV602030H, NGV602098H, NGV602331H, NGV603355H, NGV604983H, NGV604992H, NGV605100H, NGV605101H, NGV608231H, NGV608903H, NGV608907H, NGV608908H, NGV608914H, NGV612035H, NGV612185H, NGV612525H, NGV613204H, NGV613210H, NGV613264H, NGV613428H, NGV602908H, NGV601389H, NGV605731H, NGV608677H, NGV609644H, NGV610967H, NGV612138H, NGV612545H, NGV553579H, NGV600057H, NGV600091H, NGV600310H, NGV600575H, NGV601096H, NGV602162H, NGV602921H, NGV602942H, NGV603839H, NGV604806H, NGV605885H, NGV606935H, NGV607097H, NGV607733H, NGV608534H, NGV608848H, NGV609249H, NGV611148H, NGV611150H, NGV613096H, NGV613404H, NGV552829H, NGV553724H, NGV554389H, NGV602506H, NGV604837H, NGV604838H, NGV606895H, NGV552813H, NGV553028H, NGV604381H, NGV605397H, NGV608810H, NGV611989H, NGV553232H, NGV553448H, NGV554150H, NGV554197H, NGV601943H, NGV602432H, NGV602503H, NGV604148H, NGV604941H, NGV605357H, NGV605493H, NGV602585H, NGV554353H, NGV605452H, NGV607926H, NGV554283H, NGV553022H, NGV606466H, NGV606531H, NGV602172H, NGV603284H, NGV604159H, NGV605079H, NGV607270H, NGV609138H, NGV601854H, NGV605195H, NGV604923H, NGV610122H, NGV606016H, NGV607774H, NGV600948H, NGV601261H, NGV602976H, NGV606079H, NGV606521H, NGV607473H, NGV607483H, NGV607702H, NGV610447H, NGV611069H, NGV553091H, NGV553123H, NGV553432H, NGV601141H, NGV602595H, NGV602830H, NGV602831H, NGV604639H, NGV604738H, NGV604739H, NGV604741H, NGV607069H, NGV607090H, NGV607094H, NGV607425H, NGV607436H, NGV607461H, NGV608746H, NGV609922H, NGV611225H, NGV611227H, NGV613033H, NGV553120H, NGV553121H, NGV554368H, NGV601659H, NGV602542H, NGV606444H, NGV606927H, NGV607619H, NGV607757H, NGV553062H, NGV600380H, NGV602442H, NGV602499H, NGV602502H, NGV605828H, NGV600077H, NGV600078H, NGV600079H, NGV600156H, NGV604710H, NGV604714H, NGV604715H, NGV606864H, NGV608991H, NGV608993H, NGV608994H, NGV612995H, NGV612996H, NGV613003H, NGV600235H, NGV601365H, NGV602359H, NGV604154H, NGV606639H, NGV554444H, NGV609787H, NGV552838H, NGV601584H, NGV601593H, NGV606983H, NGV607920H, NGV608160H, NGV608932H, NGV609209H, NGV609517H, NGV609522H, NGV612911H, NGV554289H, NGV554339H, NGV601891H, NGV603577H, NGV604664H, NGV605557H, NGV606811H, NGV612088H, NGV612899H, NGV612947H, NGV612948H, NGV612955H, NGV612956H, NGV600216H, NGV600351H, NGV600529H, NGV600542H, NGV602136H, NGV602430H, NGV602519H, NGV604473H, NGV604476H, NGV608877H, NGV609758H, NGV611145H, NGV612941H, NGV608620H, NGV552342H, NGV552626H, NGV552659H, NGV552713H, NGV552776H, NGV552950H, NGV553629H, NGV554064H, NGV554075H, NGV602179H, NGV602316H, NGV602330H, NGV602696H, NGV602698H, NGV602875H, NGV603214H, NGV603216H, NGV603529H, NGV604051H, NGV605279H, NGV605730H, NGV606221H, NGV606498H, NGV606508H, NGV606553H, NGV606573H, NGV606583H, NGV606590H, NGV606684H, NGV607152H, NGV607253H, NGV607535H, NGV607731H, NGV607803H, NGV607875H, NGV608074H, NGV608084H, NGV608291H, NGV608292H, NGV608294H, NGV609476H, NGV609477H, NGV611348H, NGV612964H, NGV613382H, NGV600624H, NGV601889H, NGV602348H, NGV605664H, NGV610030H, NGV610757H, NGV611267H, NGV552554H, NGV600533H, NGV601503H, NGV603203H, NGV605449H, NGV605466H, NGV605795H, NGV605800H, NGV605802H, NGV612614H, NGV553250H, NGV610266H, NGV612483H, NGV600075H, NGV610744H, NGV613410H, NGV552565H, NGV552640H, NGV553433H, NGV554315H, NGV600005H, NGV600650H, NGV602517H, NGV602999H, NGV603019H, NGV603782H, NGV604450H, NGV605277H, NGV605897H, NGV606577H, NGV606884H, NGV607503H, NGV608099H, NGV608969H, NGV609182H, NGV609975H, NGV610231H, NGV610545H, NGV611346H, NGV611667H, NGV612575H, NGV612862H, NGV600961H, NGV606994H, NGV607782H, NGV610746H, NGV612679H, NGV612877H, NGV613040H, NGV553570H, NGV554092H, NGV600443H, NGV600464H, NGV600681H, NGV600862H, NGV600864H, NGV601945H, NGV602044H, NGV602050H, NGV602054H, NGV603388H, NGV603891H, NGV605756H, NGV605792H, NGV607858H, NGV608936H, NGV610756H, NGV610759H, NGV611095H, NGV611892H, NGV613305H, NGV613380H, NGV613402H, NGV605259H, NGV608730H, NGV605702H, NGV612889H, NGV612890H, NGV610324H, NGV600224H, NGV600226H, NGV600886H, NGV600887H, NGV605847H, NGV609729H, NGV600989H, NGV606245H, NGV606339H, NGV612908H, NGV601790H, NGV602405H, NGV604735H, NGV606353H, NGV607027H, NGV607137H, NGV607389H, NGV608590H, NGV612704H, NGV602605H, NGV603191H, NGV605599H, NGV607531H, NGV609952H, NGV606313H, NGV600618H, NGV612020H, NGV552584H, NGV554192H, NGV601847H, NGV602913H, NGV603908H, NGV607072H, NGV605871H, NGV600759H, NGV603274H, NGV605257H, NGV610192H, NGV554099H, NGV609420H, NGV552890H, NGV553175H, NGV553630H, NGV553668H, NGV554128H, NGV554401H, NGV600022H, NGV600667H, NGV601102H, NGV602268H, NGV602429H, NGV602677H, NGV602747H, NGV602748H, NGV602868H, NGV603686H, NGV603754H, NGV603786H, NGV604067H, NGV604752H, NGV604771H, NGV604784H, NGV605288H, NGV605692H, NGV605948H, NGV606450H, NGV606688H, NGV606784H, NGV607294H, NGV607729H, NGV608499H, NGV608868H, NGV609640H, NGV610045H, NGV610409H, NGV610643H, NGV610972H, NGV611167H, NGV611175H, NGV611797H, NGV612468H, NGV612574H, NGV612707H, NGV553413H, NGV603118H, NGV604724H, NGV606390H, NGV606397H, NGV606676H, NGV606677H, NGV608143H, NGV608146H, NGV554175H, NGV601842H, NGV603729H, NGV608033H, NGV552667H, NGV552778H, NGV601376H, NGV602079H, NGV602780H, NGV603421H, NGV605260H, NGV605717H, NGV612515H, NGV602454H, NGV605720H, NGV610146H, NGV613368H, NGV553059H, NGV600067H, NGV600413H, NGV600971H, NGV602123H, NGV602683H, NGV603079H, NGV603390H, NGV603675H, NGV604797H, NGV605733H, NGV605984H, NGV606661H, NGV608214H, NGV608223H, NGV608838H, NGV609902H, NGV612404H, NGV612513H, NGV612628H, NGV612922H, NGV602593H, NGV604485H, NGV552834H, NGV600431H, NGV601517H, NGV601519H, NGV601521H, NGV601569H, NGV601572H, NGV601622H, NGV601843H, NGV602659H, NGV602761H, NGV554286H, NGV600152H, NGV601850H, NGV602534H, NGV602535H, NGV602538H, NGV604840H, NGV607435H, NGV608794H, NGV609233H, NGV609239H, NGV610020H, NGV610414H, NGV612987H, NGV608692H, NGV608709H, NGV600829H, NGV610335H, NGV553280H, NGV600385H, NGV601712H, NGV605147H, NGV605956H, NGV609010H, NGV600741H, NGV610289H, NGV611427H, NGV552564H, NGV600132H, NGV600525H, NGV604398H, NGV604404H, NGV552811H, NGV600659H, NGV600837H, NGV553398H, NGV553404H, NGV553405H, NGV554359H, NGV554360H, NGV600238H, NGV600429H, NGV600945H, NGV600947H, NGV601325H, NGV602641H, NGV603619H, NGV603777H, NGV604831H, NGV605283H, NGV605743H, NGV605744H, NGV605969H, NGV606036H, NGV606039H, NGV607903H, NGV608536H, NGV608543H, NGV610252H, NGV610760H, NGV610761H, NGV613354H, NGV613390H, NGV611053H, NGV600484H, NGV605612H, NGV605613H, NGV609943H, NGV609955H, NGV611958H, NGV613429H, NGV603900H, NGV552923H, NGV553451H, NGV554062H, NGV554157H, NGV554158H, NGV554191H, NGV600382H, NGV601666H, NGV602982H, NGV603738H, NGV604389H, NGV607379H, NGV607944H, NGV610767H, NGV611476H, NGV612304H, NGV600653H, NGV601714H, NGV602645H, NGV603742H, NGV604077H, NGV604922H, NGV605970H, NGV609257H, NGV603501H, NGV604774H, NGV611722H, NGV606804H, NGV552781H, NGV553229H, NGV553272H, NGV553331H, NGV600462H, NGV600463H, NGV600954H, NGV601000H, NGV601003H, NGV601829H, NGV601831H, NGV601835H, NGV601836H, NGV602072H, NGV602417H, NGV602418H, NGV603101H, NGV603102H, NGV603633H, NGV604691H, NGV605352H, NGV605922H, NGV605923H, NGV606303H, NGV606304H, NGV606369H, NGV606604H, NGV607217H, NGV607904H, NGV608528H, NGV608652H, NGV611467H, NGV611633H, NGV611645H, NGV612855H, NGV554295H, NGV554301H, NGV554307H, NGV601527H, NGV601535H, NGV604117H, NGV604121H, NGV604505H, NGV608096H, NGV608815H, NGV609259H, NGV610191H, NGV611109H, NGV612214H, NGV612668H, NGV612932H, NGV612935H, NGV610975H, NGV600967H, NGV603350H, NGV603428H, NGV604821H, NGV606943H, NGV607264H, NGV609466H, NGV609495H, NGV612671H, NGV613041H, NGV603670H, NGV604283H, NGV607014H, NGV612962H, NGV553129H, NGV553130H, NGV602007H, NGV607135H, NGV607350H, NGV608061H, NGV610367H, NGV611821H, NGV612719H, NGV612898H, NGV552572H, NGV553693H, NGV600783H, NGV603003H, NGV603280H, NGV603288H, NGV603551H, NGV603626H, NGV603739H, NGV605533H, NGV607161H, NGV607251H, NGV607701H, NGV607893H, NGV611041H, NGV611341H, NGV611412H, NGV613282H, NGV613409H, NGV612850H, NGV553690H, NGV553691H, NGV554435H, NGV600219H, NGV601568H, NGV604737H, NGV605014H, NGV608365H, NGV612936H, NGV552560H, NGV552642H, NGV553596H, NGV553598H, NGV553599H, NGV600460H, NGV601731H, NGV601763H, NGV602350H, NGV602450H, NGV603269H, NGV604745H, NGV604891H, NGV607282H, NGV608125H, NGV609908H, NGV610236H, NGV611186H, NGV609800H, NGV554454H, NGV601469H, NGV612497H, NGV554086H, NGV600723H, NGV600921H, NGV600923H, NGV600925H, NGV601156H, NGV601905H, NGV602124H, NGV602467H, NGV602473H, NGV602991H, NGV603412H, NGV603422H, NGV603423H, NGV604106H, NGV605752H, NGV606071H, NGV606307H, NGV606371H, NGV608349H, NGV609136H, NGV609724H, NGV609951H, NGV610247H, NGV610248H, NGV610798H, NGV611207H, NGV611211H, NGV611582H, NGV611587H, NGV612101H, NGV612526H, NGV612527H, NGV612530H, NGV612531H, NGV613085H, NGV552576H, NGV601375H, NGV602920H, NGV604953H, NGV608301H, NGV612107H, NGV552615H, NGV552788H, NGV553203H, NGV554382H, NGV609391H.NGV609409H, NGV606191H, NGV607143H, NGV609008H, NGV600404H, NGV602486H, NGV605490H, NGV607182H, NGV554085H, NGV554249H, NGV600717H, NGV603671H, NGV607377H, NGV607972H, NGV600329H, NGV600300H, NGV601283H, NGV601890H, NGV602699H, NGV602746H, NGV602874H, NGV604499H, NGV607275H, NGV607279H, NGV612487H, NGV612934H, NGV613059H, NGV600992H, NGV604851H, NGV608148H, NGV608882H, NGV612695H, NGV604103H, NGV609792H, NGV611058H, NGV611571H, NGV554059H, NGV554269H, NGV602137H, NGV604447H, NGV606367H, NGV607383H, NGV600691H, NGV600692H, NGV602398H, NGV602979H, NGV604789H, NGV607371H, NGV608166H, NGV609012H, NGV611383H, NGV605022H, NGV605526H, NGV553603H, NGV552701H, NGV553410H, NGV600802H, NGV600824H, NGV602635H, NGV604722H, NGV604756H, NGV604845H, NGV605769H, NGV608713H, NGV608813H, NGV610978H, NGV611474H, NGV553241H, NGV553254H, NGV600166H, NGV600168H, NGV600169H, NGV602996H, NGV603014H, NGV605032H, NGV605033H, NGV605034H, NGV605035H, NGV605036H, NGV606156H, NGV607502H, NGV552888H, NGV602204H, NGV606364H, NGV612967H, NGV553039H, NGV553297H, NGV601955H, NGV610207H, NGV601805H, NGV604273H, NGV607150H, NGV610275H, NGV601781H, NGV608448H, NGV553650H, NGV553725H, NGV600088H, NGV601719H, NGV604915H, NGV605985H, NGV612240H, NGV553654H, NGV554452H, NGV600666H, NGV600731H, NGV600972H, NGV601672H, NGV602264H, NGV602864H, NGV604343H, NGV604902H, NGV604980H, NGV605663H, NGV601528H, NGV554063H, NGV603389H, NGV604665H, NGV605636H, NGV607789H, NGV612043H, NGV608079H, NGV608270H, NGV610764H, NGV600894H, NGV603929H, NGV606600H, NGV607113H, NGV608342H, NGV609736H, NGV611033H, NGV612019H, NGV606687H, NGV607165H, NGV608054H, NGV608521H, NGV612070H, NGV554088H, NGV554111H, NGV554316H, NGV600253H, NGV600272H, NGV600607H, NGV600613H, NGV601107H, NGV602559H, NGV604613H, NGV604615H, NGV604803H, NGV605287H, NGV606259H, NGV606719H, NGV607360H, NGV607394H, NGV608308H, NGV608320H, NGV608418H, NGV608545H, NGV608547H, NGV608715H, NGV608735H, NGV608998H, NGV609837H, NGV610160H, NGV610163H, NGV610642H, NGV610659H, NGV610862H, NGV612212H, NGV553752H, NGV600069H, NGV552895H, NGV552913H, NGV553060H, NGV553127H, NGV553128H, NGV553316H, NGV554277H, NGV601035H, NGV601036H, NGV601042H, NGV601045H, NGV601351H, NGV601352H, NGV602298H, NGV602790H, NGV604061H, NGV604064H, NGV604131H, NGV604143H, NGV604146H, NGV605263H, NGV605751H, NGV606534H, NGV606741H, NGV607378H, NGV607974H, NGV608505H, NGV608691H, NGV608704H, NGV608705H, NGV609742H, NGV609934H, NGV610963H, NGV610964H, NGV611338H, NGV611809H, NGV612558H, NGV612673H, NGV612729H, NGV601901H, NGV553184H, NGV602145H, NGV605851H, NGV606158H, NGV606231H, NGV608925H, NGV609027H, NGV610110H, NGV612110H, NGV600274H, NGV607352H, NGV607677H, NGV608330H, NGV613453H, NGV606083H, NGV608549H, NGV552792H, NGV553420H, NGV600387H, NGV600504H, NGV601884H, NGV604855H, NGV605779H, NGV606780H, NGV607475H, NGV608005H, NGV609519H, NGV609905H, NGV610562H, NGV610566H, NGV611617H, NGV612219H, NGV552604H, NGV554341H, NGV600157H, NGV600159H, NGV606733H, NGV607992H, NGV608306H, NGV609960H, NGV554344H, NGV603819H, NGV603838H, NGV605634H, NGV610352H, NGV612891H, NGV612083H, NGV552692H, NGV553293H, NGV554253H, NGV600259H, NGV601052H, NGV601277H, NGV601330H, NGV604889H, NGV605232H, NGV605252H, NGV605547H, NGV605642H, NGV605736H, NGV606667H, NGV606668H, NGV607351H, NGV609854H, NGV610911H, NGV613194H, NGV613265H, NGV600043H, NGV610036H, NGV600878H, NGV600879H, NGV600898H, NGV602431H, NGV608731H, NGV608733H, NGV612812H, NGV612814H, NGV612815H, NGV552789H, NGV553651H, NGV600049H, NGV601921H, NGV601924H, NGV602843H, NGV603205H, NGV606256H, NGV606660H, NGV607254H, NGV608324H, NGV608690H, NGV608779H, NGV609305H, NGV609525H, NGV609753H, NGV610854H, NGV611751H, NGV612260H, NGV612991H, NGV553612H, NGV554176H, NGV554327H, NGV600466H, NGV601468H, NGV601477H, NGV601590H, NGV601882H, NGV602078H, NGV602620H, NGV603920H, NGV604292H, NGV604674H, NGV604834H, NGV604984H, NGV605434H, NGV605513H, NGV605578H, NGV605806H, NGV606081H, NGV606267H, NGV606270H, NGV606414H, NGV606487H, NGV606790H, NGV606815H, NGV607078H, NGV607168H, NGV607169H, NGV607170H, NGV607568H, NGV607611H, NGV607744H, NGV608077H, NGV608662H, NGV608820H, NGV609653H, NGV609821H, NGV609822H, NGV609947H, NGV610098H, NGV610154H, NGV610177H, NGV610225H, NGV610351H, NGV610942H, NGV610943H, NGV612547H, NGV609034H, NGV600574H, NGV602407H, NGV604758H, NGV608614H, NGV611205H, NGV608952H, NGV603599H, NGV606144H, NGV606885H, NGV606993H, NGV610446H, NGV603613H, NGV606486H, NGV611059H, NGV606005H, NGV607587H, NGV609907H, NGV552931H, NGV553593H, NGV554219H, NGV554220H, NGV600417H, NGV600418H, NGV600426H, NGV600521H, NGV600569H, NGV600570H, NGV600572H, NGV600838H, NGV601787H, NGV602083H, NGV602704H, NGV603775H, NGV603776H, NGV603922H, NGV604005H, NGV604818H, NGV605462H, NGV606104H, NGV607431H, NGV610688H, NGV611217H, NGV605671H, NGV608597H, NGV610848H, NGV553324H, NGV554186H, NGV604046H, NGV606493H, NGV608467H, NGV609926H, NGV610394H, NGV611899H, NGV612401H, NGV606065H, NGV554422H, NGV611289H, NGV553438H, NGV605624H, NGV608520H, NGV611748H, NGV600350H, NGV605701H, NGV554333H, NGV554445H, NGV600825H, NGV603246H, NGV605061H, NGV605699H, NGV605950H, NGV607287H, NGV608070H, NGV610638H, NGV611439H, NGV612958H, NGV610863H, NGV601919H, NGV602630H, NGV609848H, NGV552583H, NGV601472H, NGV607995H, NGV612068H, NGV612701H, NGV612927H, NGV552610H, NGV554366H, NGV603276H, NGV602166H, NGV601507H, NGV604008H, NGV606673H, NGV610754H, NGV610755H, NGV553588H, NGV608790H, NGV608898H, NGV613314H, NGV554224H, NGV554250H, NGV613425H, NGV553109H, NGV603267H, NGV605719H, NGV605988H, NGV607409H, NGV610396H, NGV611304H, NGV553183H, NGV606519H, NGV611817H, NGV607417H, NGV612193H, NGV600164H, NGV609690H, NGV612194H, NGV554070H, NGV602119H, NGV605704H, NGV606338H, NGV606612H, NGV607799H, NGV608777H, NGV608822H, NGV610107H, NGV611083H, NGV612920H, NGV553283H, NGV605694H, NGV610803H, NGV552715H, NGV553556H, NGV612330H, NGV602878H, NGV602524H, NGV602590H, NGV602911H, NGV606636H, NGV601780H, NGV606293H, NGV606332H, NGV607971H, NGV611282H, NGV600241H, NGV600522H, NGV600658H, NGV601457H, NGV602016H, NGV604854H, NGV606451H, NGV609481H, NGV609489H, NGV609502H, NGV609983H, NGV610113H, NGV611859H, NGV612179H, NGV612382H, NGV612917H, NGV607212H, NGV602253H, NGV602257H, NGV602752H, NGV607024H, NGV607026H, NGV608044H, NGV608046H, NGV552568H, NGV553146H, NGV553253H, NGV553302H, NGV553351H, NGV553352H, NGV553392H, NGV554061H, NGV554369H, NGV554412H, NGV554413H, NGV600562H, NGV600747H, NGV600939H, NGV601863H, NGV603319H, NGV603634H, NGV603840H, NGV603841H, NGV604383H, NGV604409H, NGV604846H, NGV605844H, NGV605986H, NGV606373H, NGV606374H, NGV606375H, NGV606376H, NGV606378H, NGV608353H, NGV608354H, NGV608356H, NGV608357H, NGV608358H, NGV610255H, NGV610256H, NGV610416H, NGV610703H, NGV610705H, NGV610709H, NGV608200H, NGV554161H, NGV601032H, NGV602851H, NGV602852H, NGV608482H, NGV611092H, NGV611319H, NGV612376H, NGV612394H, NGV554130H, NGV600794H, NGV600977H, NGV601396H, NGV601894H, NGV603399H, NGV604595H, NGV605700H, NGV606015H, NGV606087H, NGV606471H, NGV607763H, NGV608098H, NGV610106H, NGV610363H, NGV610957H, NGV611224H, NGV611991H, NGV612501H, NGV553609H, NGV607990H, NGV612702H, NGV605474H, NGV612326H, NGV554404H, NGV554408H, NGV554415H, NGV600050H, NGV600051H, NGV601814H, NGV601816H, NGV601817H, NGV602441H, NGV602712H, NGV602959H, NGV604697H, NGV606136H, NGV606149H, NGV606150H, NGV606468H, NGV607387H, NGV607410H, NGV607411H, NGV607414H, NGV608332H, NGV612622H, NGV612544H, NGV604673H, NGV607878H, NGV602703H, NGV612206H, NGV613242H, NGV554447H, NGV606692H, NGV609938H, NGV611584H, NGV605314H, NGV605775H, NGV607997H, NGV610379H, NGV600685H, NGV608844H, NGV601686H, NGV606359H, NGV606442H, NGV552840H, NGV600689H, NGV601080H, NGV601920H, NGV604445H, NGV604469H, NGV606379H, NGV608156H, NGV608300H, NGV612886H, NGV553096H, NGV603824H, NGV605827H, NGV610234H, NGV610395H, NGV604011H, NGV605023H, NGV601750H, NGV605972H, NGV611221H, NGV552660H, NGV600695H, NGV608363H, NGV611238H, NGV554177H, NGV602521H, NGV603755H, NGV612985H, NGV613056H, NGV613057H, NGV553157H, NGV600321H, NGV600530H, NGV600531H, NGV601368H, NGV604149H, NGV605187H, NGV605531H, NGV609981H, NGV611676H, NGV552689H, NGV553364H, NGV601473H, NGV608675H, NGV602101H, NGV602631H, NGV603974H, NGV606739H, NGV607349H, NGV607786H, NGV601278H, NGV601759H, NGV602175H, NGV603899H, NGV604116H, NGV604607H, NGV608009H, NGV608040H, NGV610676H, NGV554361H, NGV600631H, NGV612986H, NGV554117H, NGV603124H, NGV607242H, NGV612464H, NGV553141H, NGV610946H, NGV611228H, NGV605096H, NGV613254H, NGV600083H, NGV600722H, NGV603029H, NGV603328H, NGV605794H, NGV607067H, NGV607798H, NGV609730H, NGV609816H, NGV610182H, NGV611002H, NGV553597H, NGV600150H, NGV600205H, NGV600543H, NGV602197H, NGV602258H, NGV602280H, NGV602474H, NGV602838H, NGV603831H, NGV604467H, NGV604858H, NGV604933H, NGV605665H, NGV605667H, NGV606003H, NGV606177H, NGV606188H, NGV607260H, NGV607892H, NGV607928H, NGV607930H, NGV608672H, NGV608707H, NGV609827H, NGV609839H, NGV610051H, NGV610115H, NGV610254H, NGV610375H, NGV610722H, NGV612861H, NGV613199H, NGV613266H, NGV600763H, NGV554148H, NGV608331H, NGV608798H, NGV554245H, NGV601465H, NGV601784H, NGV602948H, NGV603055H, NGV608832H, NGV612329H, NGV607422H, NGV553082H, NGV602873H, NGV602988H, NGV606499H, NGV606506H, NGV553585H, NGV553616H, NGV600074H, NGV553144H, NGV601932H, NGV612421H, NGV612683H, NGV600068H, NGV612439H, NGV552559H, NGV554329H, NGV601342H, NGV601346H, NGV604293H, NGV604345H, NGV554151H, NGV600547H, NGV600799H, NGV601525H, NGV601526H, NGV604985H, NGV604986H, NGV604987H, NGV612261H, NGV554406H, NGV604509H, NGV606406H, NGV609262H, NGV606076H, NGV553074H, NGV600338H, NGV600606H, NGV602067H, NGV602159H, NGV604385H, NGV604668H, NGV606383H, NGV607996H, NGV608090H, NGV612044H, NGV553395H, NGV602097H, NGV602909H, NGV605959H, NGV606650H, NGV607087H, NGV613067H, NGV613186H, NGV601879H, NGV606869H, NGV612060H, NGV609014H, NGV552849H, NGV554419H, NGV602526H, NGV603230H, NGV604526H, NGV605651H, NGV606272H, NGV606886H, NGV609313H, NGV609668H, NGV609684H, NGV609707H, NGV610189H, NGV610762H, NGV610812H, NGV610930H, NGV611390H, NGV611609H, NGV611721H, NGV612708H, NGV613095H, NGV602260H, NGV604713H, NGV607718H, NGV610584H, NGV601165H, NGV610048H, NGV553092H, NGV554348H, NGV606671H, NGV602401H, NGV609774H, NGV552592H, NGV600267H, NGV601094H, NGV602612H, NGV602813H, NGV612542H, NGV553371H, NGV554331H, NGV605296H, NGV552698H, NGV553594H, NGV554104H, NGV606596H, NGV552870H, NGV601447H, NGV601822H, NGV602001H, NGV602507H, NGV603171H, NGV605672H, NGV607597H, NGV608152H, NGV608687H, NGV608688H, NGV608827H, NGV609533H, NGV610841H, NGV612319H, NGV553292H, NGV554152H, NGV552655H, NGV600772H, NGV600773H, NGV600775H, NGV600955H, NGV602125H, NGV602128H, NGV609473H, NGV600328H, NGV552707H, NGV553125H, NGV600280H, NGV600384H, NGV600679H, NGV602435H, NGV602900H, NGV603307H, NGV603504H, NGV603511H, NGV605804H, NGV606004H, NGV606281H, NGV606953H, NGV607497H, NGV608278H, NGV608336H, NGV609893H, NGV610273H, NGV553632H, NGV600009H, NGV600505H, NGV603788H, NGV605375H, NGV605990H, NGV609773H, NGV610809H, NGV610846H, NGV552723H, NGV552893H, NGV553421H, NGV554231H, NGV600266H, NGV600459H, NGV601777H, NGV604344H, NGV605753H, NGV607835H, NGV607840H, NGV608137H, NGV609436H, NGV611727H, NGV601701H, NGV604134H, NGV604448H, NGV605275H, NGV610732H, NGV611308H, NGV608020H, NGV612238H, NGV553077H, NGV603961H, NGV606072H, NGV610664H, NGV606633H, NGV611192H, NGV600608H, NGV603601H, NGV606310H, NGV609829H, NGV611935H, NGV611940H, NGV613022H, NGV607070H, NGV553041H, NGV602469H, NGV602954H, NGV604916H, NGV605661H, NGV605714H, NGV606607H, NGV610377H, NGV604423H, NGV604436H, NGV606033H, NGV609521H, NGV610800H, NGV611054H, NGV605253H, NGV608350H, NGV610086H, NGV611100H, NGV606212H, NGV610714H, NGV610716H, NGV611720H, NGV605831H, NGV607369H, NGV553158H, NGV553159H, NGV553161H, NGV553163H, NGV553617H, NGV554392H, NGV602625H, NGV602626H, NGV602627H, NGV604053H, NGV604102H, NGV605932H, NGV606089H, NGV606091H, NGV606601H, NGV607092H.NGV607138H, NGV607338H, NGV607889H, NGV608738H, NGV608961H, NGV609784H, NGV609847H, NGV609912H, NGV609917H, NGV609918H, NGV610157H, NGV611602H, NGV612169H, NGV612171H, NGV612517H, NGV612872H, NGV603560H, NGV603186H, NGV605935H, NGV554166H, NGV554168H, NGV601317H, NGV609563H, NGV610866H, NGV604044H, NGV553730H, NGV553369H, NGV611239H, NGV554204H, NGV554206H, NGV554227H, NGV554352H, NGV600239H, NGV600244H, NGV602944H, NGV603669H, NGV603692H, NGV604782H, NGV607566H, NGV609914H, NGV610736H, NGV611591H, NGV611592H, NGV552732H, NGV552733H, NGV552799H, NGV552800H, NGV552900H, NGV553559H, NGV553560H, NGV553563H, NGV554430H, NGV554431H, NGV600010H, NGV600011H, NGV600012H, NGV600184H, NGV600536H, NGV600537H, NGV600538H, NGV600539H, NGV600748H, NGV600853H, NGV601009H, NGV601828H, NGV601833H, NGV601997H, NGV602069H, NGV602113H, NGV602114H, NGV602116H, NGV602459H, NGV602468H, NGV602484H, NGV603521H, NGV603523H, NGV603531H, NGV603533H, NGV603535H, NGV604543H, NGV604646H, NGV604648H, NGV604654H, NGV605003H, NGV605261H, NGV605617H, NGV605621H, NGV605910H, NGV605912H, NGV605913H, NGV605915H, NGV606067H, NGV606686H, NGV606990H, NGV606992H, NGV607086H, NGV607101H, NGV607308H, NGV607402H, NGV608060H, NGV608134H, NGV608135H, NGV608136H, NGV608360H, NGV608364H, NGV608806H, NGV608967H, NGV609425H, NGV609426H, NGV609427H, NGV609434H, NGV609877H, NGV609878H, NGV609881H, NGV610096H, NGV610103H, NGV610224H, NGV610327H, NGV610621H, NGV610979H, NGV610981H, NGV611926H, NGV612296H, NGV612605H, NGV612607H, NGV612610H, NGV612738H, NGV613459H, NGV613468H, NGV553710H, NGV602829H, NGV604141H, NGV607812H, NGV608041H, NGV608718H, NGV610300H, NGV600315H, NGV601395H, NGV601400H, NGV603688H, NGV603689H, NGV600890H, NGV606881H, NGV606883H, NGV607368H, NGV607359H, NGV552679H, NGV553056H, NGV554200H, NGV602389H, NGV604038H, NGV611026H, NGV611942H, NGV552672H, NGV552902H, NGV553312H, NGV553419H, NGV553437H, NGV554116H, NGV554221H, NGV600171H, NGV600208H, NGV600524H, NGV600534H, NGV601350H, NGV601437H, NGV601952H, NGV602353H, NGV602649H, NGV602650H, NGV602655H, NGV602656H, NGV602871H, NGV604094H, NGV604656H, NGV605320H, NGV605538H, NGV606108H, NGV606989H, NGV607421H, NGV607894H, NGV607947H, NGV607960H, NGV608157H, NGV608398H, NGV608670H, NGV608921H, NGV609669H, NGV609688H, NGV609930H, NGV611078H, NGV611323H, NGV611976H, NGV612177H, NGV612184H, NGV612965H, NGV613051H, NGV613458H, NGV607129H, NGV552569H, NGV553298H, NGV600629H, NGV603606H, NGV606381H, NGV608487H, NGV553156H, NGV600883H, NGV600884H, NGV601445H, NGV602543H, NGV602783H, NGV604723H, NGV600888H, NGV608147H, NGV609055H, NGV609058H, NGV604510H, NGV606316H, NGV612864H, NGV552760H, NGV603064H, NGV607330H, NGV610350H, NGV612485H, NGV603317H, NGV603715H, NGV611353H, NGV552827H, NGV602500H, NGV604281H, NGV604295H, NGV604957H, NGV611403H, NGV611454H, NGV601770H, NGV606439H, NGV604506H, NGV552804H, NGV600492H, NGV604152H, NGV604746H, NGV604861H, NGV607939H, NGV609791H, NGV610136H, NGV610393H, NGV611960H, NGV607385H, NGV601961H, NGV602749H, NGV604084H, NGV604307H, NGV609717H, NGV552598H, NGV553347H, NGV553348H, NGV600311H, NGV600804H, NGV601153H, NGV601460H, NGV601461H, NGV601462H, NGV604369H, NGV604370H, NGV604371H, NGV604372H, NGV604374H, NGV604376H, NGV604871H, NGV606701H, NGV608737H, NGV608740H, NGV608741H, NGV608742H, NGV552894H, NGV602445H, NGV606349H, NGV606430H, NGV612631H, NGV552757H, NGV552787H, NGV554317H, NGV600356H, NGV604092H, NGV554082H, NGV554420H, NGV600048H, NGV601722H, NGV601723H, NGV604137H, NGV604946H, NGV605004H, NGV605842H, NGV609663H, NGV609925H, NGV610399H, NGV611863H, NGV612061H, NGV613094H, NGV613098H, NGV554423H, NGV552634H, NGV553382H, NGV607249H, NGV611629H, NGV613191H, NGV601987H, NGV601996H, NGV602693H, NGV603386H, NGV603637H, NGV604800H, NGV605980H, NGV605997H, NGV607393H, NGV607908H, NGV609428H, NGV612846H, NGV612878H, NGV612879H, NGV553407H, NGV609995H, NGV554057H, NGV553026H, NGV553075H, NGV600438H, NGV606213H, NGV610721H, NGV612025H, NGV553252H, NGV603678H, NGV604809H, NGV606740H, NGV606933H, NGV608072H, NGV608924H, NGV610354H, NGV600516H, NGV605323H, NGV609268H, NGV610796H, NGV602916H, NGV603005H, NGV607181H, NGV608048H, NGV609755H, NGV610779H, NGV612398H, NGV613091H, NGV613294H, NGV553220H, NGV554246H, NGV604100H, NGV610971H, NGV612938H, NGV600544H, NGV603048H, NGV605781H, NGV602674H, NGV607089H, NGV609874H, NGV610726H, NGV610977H, NGV611478H, NGV612615H, NGV600509H, NGV601293H, NGV552896H, NGV600577H, NGV608580H, NGV605577H, NGV604308H, NGV605685H, NGV601960H, NGV604035H, NGV604876H, NGV605939H, NGV610663H, NGV603277H, NGV600192H, NGV602640H, NGV554173H, NGV606813H, NGV607220H, NGV607245H, NGV610927H, NGV612664H, NGV602226H, NGV606386H, NGV553436H, NGV553622H, NGV553623H, NGV553624H, NGV600024H, NGV600025H, NGV600026H, NGV600028H, NGV600176H, NGV600454H, NGV600455H, NGV600473H, NGV600703H, NGV601060H, NGV601063H, NGV601064H, NGV601067H, NGV601081H, NGV601499H, NGV601623H, NGV602274H, NGV602648H, NGV603349H, NGV603680H, NGV604056H, NGV605911H, NGV606167H, NGV607478H, NGV607479H, NGV607480H, NGV608821H, NGV609322H, NGV609323H, NGV609457H, NGV610356H, NGV611091H, NGV611559H, NGV612506H, NGV612556H, NGV612619H, NGV612621H, NGV552835H, NGV552934H, NGV553210H, NGV554354H, NGV554363H, NGV554387H, NGV602043H, NGV603002H, NGV604602H, NGV604620H, NGV604811H, NGV604823H, NGV606277H, NGV606905H, NGV607065H, NGV608318H, NGV609157H, NGV609159H, NGV609549H, NGV609973H, NGV610557H, NGV611312H, NGV611446H, NGV612327H, NGV612751H, NGV612782H, NGV612835H, NGV612854H, NGV612993H, NGV612999H, NGV552648H, NGV552944H, NGV553093H, NGV553742H, NGV554367H, NGV600066H, NGV600136H, NGV600461H, NGV600755H, NGV600757H, NGV602276H, NGV602673H, NGV603697H, NGV604744H, NGV604899H, NGV605008H, NGV605011H, NGV605012H, NGV605013H, NGV605693H, NGV607556H, NGV607773H, NGV607899H, NGV608123H, NGV608314H, NGV608891H, NGV608892H, NGV608893H, NGV608895H, NGV609836H, NGV611414H, NGV611463H, NGV611692H, NGV611736H, NGV612062H, NGV612134H, NGV612445H, NGV612954H, NGV612968H, NGV610931H, NGV607205H, NGV608032H, NGV611181H, NGV611017H, NGV605689H, NGV605771H, NGV607562H, NGV609042H, NGV609052H, NGV609771H, NGV602541H, NGV606080H, NGV610183H, NGV610279H, NGV606724H, NGV552681H, NGV552741H, NGV553071H, NGV600322H, NGV602318H, NGV602339H, NGV607353H, NGV607886H, NGV608128H, NGV608881H, NGV600391H, NGV601800H, NGV602239H, NGV605301H, NGV607272H, NGV607721H, NGV609866H, NGV609999H, NGV605423H, NGV601697H, NGV601710H, NGV606358H, NGV609251H, NGV600375H, NGV600669H, NGV609939H, NGV602130H, NGV552929H, NGV613115H, NGV612852H, NGV607025H, NGV602433H, NGV606222H, NGV606736H, NGV607881H, NGV606009H, NGV607741H, NGV612191H, NGV552933H, NGV553671H, NGV553740H, NGV600070H, NGV600273H, NGV603783H, NGV606350H, NGV606704H, NGV608203H, NGV611014H, NGV612297H, NGV604720H, NGV606613H, NGV554243H, NGV612438H, NGV602413H, NGV602509H, NGV607919H, NGV602294H, NGV605894H, NGV607457H, NGV554083H, NGV601774H, NGV602571H, NGV607020H, NGV607463H, NGV609270H, NGV552872H, NGV554437H, NGV600187H, NGV600749H, NGV600936H, NGV601940H, NGV602064H, NGV603682H, NGV603728H, NGV603733H, NGV603751H, NGV605401H, NGV605774H, NGV607136H, NGV607565H, NGV608108H, NGV608109H, NGV609153H, NGV609942H, NGV610788H, NGV610790H, NGV610791H, NGV611746H, NGV612016H, NGV612076H, NGV612217H, NGV612362H, NGV612576H, NGV607691H, NGV602505H, NGV604380H, NGV606330H, NGV608017H, NGV610390H, NGV552694H, NGV607675H, NGV611391H, NGV603062H, NGV606117H, NGV552909H, NGV553665H, NGV600217H, NGV600842H, NGV601902H, NGV602011H, NGV602261H, NGV602777H, NGV602778H, NGV603193H, NGV603593H, NGV604113H, NGV604896H, NGV607527H, NGV607528H, NGV608064H, NGV609726H, NGV611258H, NGV611867H, NGV613424H, NGV607382H, NGV607880H, NGV610816H, NGV610959H, NGV613401H, NGV553718H, NGV600990H, NGV602251H, NGV602347H, NGV602492H, NGV602555H, NGV603117H, NGV603720H, NGV607109H, NGV607777H, NGV612287H, NGV612494H, NGV609625H, NGV609647H, NGV610765H, NGV607962H, NGV604133H, NGV552883H, NGV554207H, NGV610283H, NGV606013H, NGV602910H, NGV605650H, NGV553411H, NGV601922H, NGV602400H, NGV605179H, NGV605271H, NGV607131H, NGV608075H, NGV608359H, NGV609721H, NGV601942H, NGV607124H, NGV601837H, NGV609285H, NGV609096H, NGV553287H, NGV601900H, NGV601934H, NGV602615H, NGV603247H, NGV604340H, NGV604606H, NGV605691H, NGV609044H, NGV609206H, NGV609298H, NGV610880H, NGV612710H, NGV603639H, NGV610850H, NGV610961H, NGV611715H, NGV612870H, NGV609421H, NGV612200H, NGV606001H, NGV552897H, NGV553271H, NGV554190H, NGV601986H, NGV602387H, NGV602619H, NGV605931H, NGV607214H, NGV607381H, NGV607895H, NGV608000H, NGV611968H, NGV600758H, NGV603404H, NGV605447H, NGV605928H, NGV612654H, NGV602876H, NGV604050H, NGV604903H, NGV607743H, NGV600044H, NGV602628H, NGV552812H, NGV552922H, NGV552926H, NGV553385H, NGV600031H, NGV601915H, NGV602300H, NGV602301H, NGV604463H, NGV604794H, NGV606296H, NGV606410H, NGV606412H, NGV607119H, NGV607120H, NGV607127H, NGV608339H, NGV610251H, NGV612529H, NGV612532H, NGV612977H, NGV612978H, NGV613415H, NGV613418H, NGV605978H, NGV605574H, NGV606200H, NGV609023H, NGV612221H, NGV552832H, NGV554386H, NGV610155H, NGV604522H, NGV552935H, NGV553357H, NGV612199H, NGV610797H, NGV607830H, NGV610929H, NGV602216H, NGV605118H, NGV612905H, NGV554076H, NGV601116H, NGV601119H, NGV601295H, NGV601466H, NGV602736H, NGV602893H, NGV602894H, NGV602895H, NGV602896H, NGV602897H, NGV602899H, NGV602924H, NGV603699H, NGV603719H, NGV603933H, NGV603937H, NGV603940H, NGV604687H, NGV604693H, NGV604895H, NGV604898H, NGV607004H, NGV607008H, NGV607010H, NGV607068H, NGV607548H, NGV607563H, NGV610044H, NGV610630H, NGV610636H, NGV610641H, NGV612245H, NGV552577H, NGV552578H, NGV552579H, NGV552581H, NGV553656H, NGV553658H, NGV553660H, NGV553661H, NGV554132H, NGV554279H, NGV601085H, NGV601089H, NGV601121H, NGV601122H, NGV601125H, NGV601127H, NGV601482H, NGV601483H, NGV601485H, NGV601486H, NGV601487H, NGV601489H, NGV601490H, NGV604022H, NGV604023H, NGV604024H, NGV604027H, NGV604028H, NGV604030H, NGV604612H, NGV604632H, NGV605037H, NGV605038H, NGV605039H, NGV605040H, NGV605041H, NGV607210H, NGV607211H, NGV607216H, NGV607223H, NGV607847H, NGV607851H, NGV607853H, NGV608423H, NGV608435H, NGV608937H, NGV608938H, NGV608939H, NGV608942H, NGV608943H, NGV608944H, NGV608946H, NGV608948H, NGV608949H, NGV612040H, NGV612045H, NGV612048H, NGV613002H, NGV613007H, NGV613009H, NGV613010H, NGV613011H, NGV613012H, NGV613027H, NGV613030H, NGV600363H, NGV554115H, NGV553715H, NGV612307H, NGV600848H, NGV601941H, NGV604623H, NGV606737H, NGV606930H, NGV609456H, NGV609620H, NGV613084H, NGV607221H, NGV552846H, NGV553088H, NGV553666H, NGV601004H, NGV601870H, NGV609880H, NGV612155H, NGV612630H, NGV603948H, NGV604743H, NGV610290H, NGV600040H, NGV600355H, NGV600663H, NGV604542H, NGV600323H, NGV611085H, NGV607178H, NGV601857H, NGV607418H, NGV600178H, NGV604748H, NGV604749H, NGV605807H, NGV611031H, NGV612477H, NGV602525H, NGV554080H, NGV603004H, NGV609964H, NGV610108H, NGV610109H, NGV611774H, NGV612232H, NGV612411H, NGV606261H, NGV605331H, NGV605793H, NGV610353H, NGV610683H, NGV554407H, NGV602303H, NGV602438H, NGV605619H, NGV606779H, NGV554377H, NGV553431H, NGV601992H, NGV602214H, NGV604052H, NGV607951H, NGV602544H, NGV602759H, NGV603623H, NGV604793H, NGV606322H, NGV606449H, NGV606651H, NGV606653H, NGV607428H, NGV608086H, NGV608113H, NGV608748H, NGV609017H, NGV609232H, NGV610991H, NGV612423H, NGV603873H, NGV602826H, NGV603166H, NGV603167H, NGV603894H, NGV605186H, NGV605724H, NGV605728H, NGV611098H, NGV612029H, NGV612583H, NGV609884H, NGV610241H, NGV613101H, NGV605977H, NGV613218H, NGV600414H, NGV601583H, NGV604138H, NGV606540H, NGV608550H, NGV610187H, NGV607157H, NGV609623H, NGV601530H, NGV601531H, NGV601533H, NGV601630H, NGV603914H, NGV608340H, NGV608517H, NGV608518H, NGV609126H, NGV603220H, NGV604417H, NGV604424H, NGV604440H, NGV604993H, NGV605047H, NGV605051H, NGV605056H, NGV605067H, NGV552675H, NGV601380H, NGV601383H, NGV601384H, NGV601386H, NGV601388H, NGV603314H, NGV606022H, NGV606503H, NGV606543H, NGV606545H, NGV606547H, NGV606548H.NGV606550H, NGV606555H, NGV606561H, NGV606562H, NGV606563H, NGV606568H, NGV606581H, NGV606585H, NGV606591H, NGV606592H, NGV606593H, NGV607348H, NGV552817H, NGV600364H, NGV600665H, NGV601372H, NGV601407H, NGV601409H, NGV601425H, NGV601426H, NGV601430H, NGV601431H, NGV601433H, NGV601435H, NGV601436H, NGV601440H, NGV601443H, NGV601444H, NGV601448H, NGV601451H, NGV601981H, NGV602232H, NGV603105H, NGV604007H, NGV604104H, NGV604585H, NGV604675H, NGV604716H, NGV605611H, NGV606090H, NGV606691H, NGV607484H, NGV607487H, NGV607490H, NGV607492H, NGV607493H, NGV607572H, NGV607577H, NGV607578H, NGV607581H, NGV607584H, NGV607588H, NGV607589H, NGV607596H, NGV607633H, NGV607635H, NGV607643H, NGV607650H, NGV607652H, NGV607657H, NGV607664H, NGV607666H, NGV607668H, NGV607669H, NGV607670H, NGV607672H, NGV609388H, NGV609390H, NGV609394H, NGV609395H, NGV609401H, NGV609402H, NGV609404H, NGV609410H, NGV609415H, NGV609552H, NGV609553H, NGV609554H, NGV609555H, NGV609557H, NGV609560H, NGV609562H, NGV609564H, NGV609576H, NGV609577H, NGV609581H, NGV609582H, NGV609583H, NGV609586H, NGV609587H, NGV609589H, NGV609590H, NGV609591H, NGV609593H, NGV609596H, NGV609601H, NGV609602H, NGV610307H, NGV611756H, NGV611777H, NGV601631H, NGV601633H, NGV601634H, NGV601636H, NGV601637H, NGV601638H, NGV601642H, NGV601643H, NGV601645H, NGV601646H, NGV601648H, NGV601650H, NGV601652H, NGV601653H, NGV601670H, NGV601684H, NGV601694H, NGV601705H, NGV609137H, NGV609242H, NGV609255H, NGV609261H, NGV609266H, NGV609273H, NGV609274H, NGV609275H, NGV609292H, NGV609315H, NGV609603H, NGV609604H, NGV609605H, NGV609606H, NGV609607H, NGV609611H, NGV609612H, NGV609613H, NGV609616H, NGV609618H, NGV609619H, NGV609624H, NGV609627H, NGV609629H, NGV609630H, NGV609633H, NGV609635H, NGV609636H, NGV601481H, NGV600406H, NGV601581H, NGV601586H, NGV601588H, NGV602009H, NGV602846H, NGV607052H, NGV607054H, NGV607055H, NGV607059H, NGV607060H, NGV609208H, NGV609228H, NGV609295H, NGV609386H, NGV609537H, NGV609542H, NGV609544H, NGV609546H, NGV609547H, NGV609548H, NGV609550H, NGV552731H, NGV605869H, NGV612979H, NGV606939H, NGV610724H, NGV611356H, NGV600094H, NGV602273H, NGV611271H, NGV606564H, NGV607911H, NGV601536H, NGV601537H, NGV601538H, NGV601539H, NGV601542H, NGV601546H, NGV601547H, NGV601549H, NGV601550H, NGV601551H, NGV601552H, NGV601553H, NGV601554H, NGV601556H, NGV601560H, NGV601561H, NGV601595H, NGV601596H, NGV601599H, NGV601601H, NGV601602H, NGV601603H, NGV601607H, NGV601608H, NGV601609H, NGV601616H, NGV601617H, NGV601620H, NGV601621H, NGV601767H, NGV601811H, NGV601873H, NGV604031H, NGV604810H, NGV606171H, NGV606208H, NGV606317H, NGV606454H, NGV606819H, NGV606826H, NGV606827H, NGV606832H, NGV606838H, NGV606843H, NGV606848H, NGV606852H, NGV606853H, NGV606855H, NGV606857H, NGV606859H, NGV606862H, NGV606964H, NGV606966H, NGV606968H, NGV606972H, NGV606973H, NGV606974H, NGV606975H, NGV606977H, NGV606978H, NGV606980H, NGV606982H, NGV606984H, NGV606985H, NGV607031H, NGV607032H, NGV607033H, NGV607034H, NGV607036H, NGV607038H, NGV607040H, NGV607041H, NGV607043H, NGV607044H, NGV607045H, NGV607046H, NGV607048H, NGV607062H, NGV607063H, NGV609073H, NGV609075H, NGV609077H, NGV609080H, NGV609081H, NGV609082H, NGV609083H, NGV609085H, NGV609086H, NGV609087H, NGV609097H, NGV609099H, NGV609100H, NGV609101H, NGV609102H, NGV609106H, NGV609112H, NGV609116H, NGV609117H, NGV609124H, NGV609128H, NGV609130H, NGV609132H, NGV609134H, NGV609165H, NGV609493H, NGV553370H, NGV554155H, NGV554378H, NGV600872H, NGV602633H, NGV603677H, NGV605882H, NGV606635H, NGV602185H, NGV602686H, NGV606880H, NGV609909H, NGV601378H, NGV603023H, NGV603044H, NGV610802H, NGV602930H, NGV611284H, NGV552764H, NGV553306H, NGV600450H, NGV600619H, NGV602010H, NGV602319H, NGV602573H, NGV602847H, NGV603071H, NGV603072H, NGV603073H, NGV603913H, NGV605655H, NGV605803H, NGV605843H, NGV605881H, NGV606132H, NGV606151H, NGV607685H, NGV608480H, NGV608728H, NGV609414H, NGV609803H, NGV609980H, NGV610083H, NGV610220H, NGV610938H, NGV612258H, NGV612904H, NGV613001H, NGV613337H, NGV603285H, NGV612820H, NGV612831H, NGV600337H, NGV611750H, NGV552877H, NGV612892H, NGV613329H, NGV552910H, NGV607404H, NGV606997H, NGV611927H, NGV553221H, NGV600056H, NGV600064H, NGV601742H, NGV607384H, NGV608027H, NGV609850H, NGV609853H, NGV611065H, NGV605610H, NGV609680H, NGV552876H, NGV553069H, NGV553262H, NGV553680H, NGV554241H, NGV600342H, NGV600734H, NGV600828H, NGV602818H, NGV602989H, NGV603578H, NGV604076H, NGV604105H, NGV605813H, NGV606512H, NGV608905H, NGV610193H, NGV610766H, NGV610883H, NGV611891H, NGV600055H, NGV600232H, NGV601962H, NGV603320H, NGV603833H, NGV604125H, NGV605193H, NGV607449H, NGV607451H, NGV608598H, NGV609244H, NGV608264H, NGV609508H, NGV602638H, NGV607262H, NGV553600H, NGV602183H, NGV553557H, NGV600306H, NGV601730H, NGV602411H, NGV603736H, NGV605982H, NGV609523H, NGV610558H, NGV553601H, NGV553649H, NGV554171H, NGV554261H, NGV600172H, NGV600798H, NGV600806H, NGV602094H, NGV602427H, NGV602439H, NGV605211H, NGV605216H, NGV605709H, NGV606306H, NGV606606H, NGV610142H, NGV610173H, NGV611406H, NGV612885H, NGV604026H, NGV600140H, NGV608797H, NGV611469H, NGV608279H, NGV612347H, NGV600357H, NGV602859H, NGV607471H, NGV608825H, NGV552725H, NGV554169H, NGV602104H, NGV602867H, NGV606387H, NGV606674H, NGV606675H, NGV608142H, NGV608719H, NGV609812H, NGV612112H, NGV552611H, NGV552994H, NGV554217H, NGV554274H, NGV600721H, NGV601935H, NGV601938H, NGV602207H, NGV602478H, NGV602479H, NGV604480H, NGV604482H, NGV605017H, NGV605018H, NGV607176H, NGV607198H, NGV607201H, NGV608334H, NGV608560H, NGV608586H, NGV611279H, NGV611606H, NGV611733H, NGV612887H, NGV554294H, NGV600550H, NGV601744H, NGV604830H, NGV604832H, NGV604849H, NGV604859H, NGV607442H, NGV601023H, NGV610180H, NGV613043H, NGV604438H, NGV605973H, NGV612032H, NGV552585H, NGV607194H, NGV612875H, NGV554094H, NGV600604H, NGV600163H, NGV610777H, NGV604098H, NGV612158H, NGV603778H, NGV607192H, NGV607305H, NGV608659H, NGV603086H, NGV607076H, NGV553169H, NGV554098H, NGV554350H, NGV600994H, NGV601441H, NGV602889H, NGV603482H, NGV603737H, NGV605545H, NGV607162H, NGV607171H, NGV610111H, NGV611413H, NGV611458H, NGV612676H, NGV552873H, NGV611212H, NGV553736H, NGV553737H, NGV553738H, NGV554109H, NGV600932H, NGV601014H, NGV601015H, NGV601016H, NGV601017H, NGV601020H, NGV601021H, NGV601104H, NGV601106H, NGV606355H, NGV606356H, NGV607933H, NGV607934H, NGV607936H, NGV608888H, NGV608896H, NGV608897H, NGV612520H, NGV612521H, NGV603213H, NGV606240H, NGV608030H, NGV610910H, NGV609330H, NGV610731H, NGV553646H, NGV600407H, NGV600614H, NGV602689H, NGV610167H, NGV612869H, NGV553219H, NGV610774H, NGV607506H, NGV604666H, NGV605000H, NGV605001H, NGV606424H, NGV612853H, NGV552612H, NGV609256H, NGV604567H, NGV604814H, NGV610258H, NGV611315H, NGV613092H, NGV553207H, NGV554069H, NGV602361H, NGV602402H, NGV603448H, NGV605936H, NGV608486H, NGV609444H, NGV554149H, NGV601464H, NGV607704H, NGV610105H, NGV612901H, NGV552622H, NGV605902H, NGV610597H, NGV554213H, NGV600018H, NGV603553H, NGV603635H, NGV607843H, NGV611583H, NGV611588H, NGV613117H, NGV554135H, NGV604037H, NGV608015H, NGV611620H, NGV600236H, NGV600029H, NGV600080H, NGV600135H, NGV602809H, NGV602807H, NGV606932H, NGV610079H, NGV610677H, NGV612283H, NGV612284H, NGV612289H, NGV612290H, NGV600283H, NGV603090H, NGV611066H, NGV552945H, NGV553698H, NGV601908H, NGV602126H, NGV602199H, NGV602667H, NGV603752H, NGV604594H, NGV604787H, NGV605850H, NGV606096H, NGV606279H, NGV606696H, NGV606698H, NGV607745H, NGV608003H, NGV608057H, NGV608338H, NGV613169H, NGV604785H, NGV552898H, NGV603447H, NGV611810H, NGV552730H, NGV552737H, NGV554065H, NGV600628H, NGV605625H, NGV605695H, NGV609207H, NGV610763H, NGV553165H, NGV553356H, NGV553678H, NGV553741H, NGV554239H, NGV600298H, NGV601007H, NGV602888H, NGV603735H, NGV604144H, NGV604734H, NGV604979H, NGV605542H, NGV605776H, NGV605778H, NGV606176H, NGV606308H, NGV607540H, NGV608091H, NGV609490H, NGV609666H, NGV610169H, NGV554403H, NGV554280H, NGV554290H, NGV600401H, NGV603887H, NGV604829H, NGV610067H, NGV611253H, NGV612166H, NGV600826H, NGV610305H, NGV553031H, NGV553073H, NGV553166H, NGV553180H, NGV553749H, NGV600557H, NGV553387H, NGV602355H, NGV602397H, NGV552695H, NGV553309H, NGV600832H, NGV604296H, NGV605971H, NGV603240H, NGV608271H, NGV608665H, NGV609924H, NGV611011H, NGV612234H, NGV613284H, NGV608207H, NGV606495H, NGV600394H, NGV610310H, NGV552851H, NGV602103H, NGV602391H, NGV602393H, NGV602941H, NGV603197H, NGV604742H, NGV604908H, NGV606263H, NGV606324H, NGV607546H, NGV607762H, NGV610322H, NGV612578H, NGV612732H, NGV552844H, NGV608812H, NGV600240H, NGV600304H, NGV600308H, NGV553667H, NGV553328H, NGV554343H, NGV601711H, NGV602760H, NGV603321H, NGV603832H, NGV604390H, NGV604736H, NGV610711H, NGV609727H, NGV609737H, NGV602392H, NGV608016H, NGV602552H, NGV605518H, NGV607121H, NGV602420H, NGV601727H, NGV553610H, NGV606924H, NGV611760H, NGV553569H, NGV606276H, NGV607965H, NGV552793H, NGV553571H, NGV602271H, NGV603243H, NGV603387H, NGV605940H, NGV606273H, NGV608097H, NGV608149H, NGV608337H, NGV608647H, NGV608775H, NGV608873H, NGV610158H, NGV610674H, NGV610689H, NGV610857H, NGV611233H, NGV612491H, NGV612897H, NGV606808H, NGV610159H, NGV603242H, NGV553064H, NGV554209H, NGV600097H, NGV601949H, NGV601999H, NGV602180H, NGV602508H, NGV605180H, NGV605644H, NGV606447H, NGV606616H, NGV606669H, NGV606871H, NGV607110H, NGV607467H, NGV607491H, NGV608275H, NGV608686H, NGV608744H, NGV609389H, NGV609405H, NGV610214H, NGV611686H, NGV612057H, NGV612227H, NGV613046H, NGV613372H, NGV552719H, NGV552884H, NGV554436H, NGV604347H, NGV604352H, NGV610308H, NGV609498H, NGV602744H, NGV607153H, NGV607250H, NGV610407H, NGV553094H, NGV553275H, NGV600657H, NGV602444H, NGV603134H, NGV603628H, NGV603629H, NGV603963H, NGV604698H, NGV604862H, NGV604892H, NGV604904H, NGV606743H, NGV606807H, NGV607167H, NGV608094H, NGV608755H, NGV610657H, NGV610660H, NGV554195H, NGV601799H, NGV602202H, NGV603818H, NGV603876H, NGV604039H, NGV605360H, NGV607787H, NGV608732H, NGV609006H, NGV609794H, NGV610055H, NGV610366H, NGV612086H, NGV613037H, NGV553270H, NGV600563H, NGV600580H, NGV601471H, NGV601480H, NGV602820H, NGV603257H, NGV605904H, NGV607532H, NGV608296H, NGV608316H, NGV609817H, NGV605297H, NGV600349H, NGV603817H, NGV604981H, NGV606319H, NGV610296H, NGV552670H, NGV553089H, NGV553100H, NGV553189H, NGV553191H, NGV553192H, NGV601812H, NGV602325H, NGV602409H, NGV603944H, NGV603951H, NGV604163H, NGV604269H, NGV604842H, NGV605918H, NGV606752H, NGV607237H, NGV608013H, NGV608710H, NGV608711H, NGV608716H, NGV611251H, NGV611252H, NGV611260H, NGV612982H, NGV612990H, NGV612994H, NGV602774H, NGV607386H, NGV607765H, NGV612871H, NGV553700H, NGV602022H, NGV605472H, NGV606028H, NGV606340H, NGV606341H, NGV607808H, NGV607811H, NGV610344H, NGV603727H, NGV606751H, NGV606996H, NGV607276H, NGV552914H, NGV554351H, NGV600697H, NGV604537H, NGV601688H, NGV602404H, NGV606526H, NGV607265H, NGV609692H, NGV609768H, NGV606728H, NGV608011H, NGV602419H, NGV612942H, NGV611285H, NGV552573H, NGV552740H, NGV600833H, NGV602215H, NGV602776H, NGV603325H, NGV605324H, NGV605679H, NGV607482H, NGV607515H, NGV607724H, NGV608056H, NGV609935H, NGV610656H, NGV554330H, NGV600197H, NGV600408H, NGV600434H, NGV602787H, NGV606758H, NGV607861H, NGV608080H, NGV610139H, NGV612440H, NGV612594H, NGV613175H, NGV553683H, NGV602370H, NGV606409H, NGV606417H, NGV607229H, NGV607230H, NGV608828H, NGV608843H, NGV613255H, NGV552299H, NGV553305H, NGV600233H, NGV601806H, NGV602636H, NGV602643H, NGV602856H, NGV602858H, NGV603142H, NGV604036H, NGV604515H, NGV604783H, NGV604911H, NGV604954H, NGV604994H, NGV605113H, NGV605115H, NGV605316H, NGV605317H, NGV606101H, NGV606404H, NGV607239H, NGV607842H, NGV607946H, NGV610238H, NGV610403H, NGV610535H, NGV610729H, NGV610882H, NGV610885H, NGV610890H, NGV610944H, NGV611102H, NGV611104H, NGV611347H, NGV600093H, NGV607133H, NGV600203H, NGV600023H, NGV602662H, NGV608876H, NGV612108H, NGV552863H, NGV601864H, NGV602295H, NGV602984H, NGV603794H, NGV603987H, NGV604529H, NGV606729H, NGV606730H, NGV606925H, NGV607419H.NGV609185H, NGV609205H, NGV610095H, NGV610330H, NGV610410H, NGV611220H, NGV612050H, NGV612051H, NGV612052H, NGV612309H, NGV553107H, NGV608154H, NGV613239H, NGV554293H, NGV602487H, NGV604493H, NGV609790H, NGV612334H, NGV600500H, NGV608739H, NGV553340H, NGV554236H, NGV608100H, NGV612306H, NGV612453H, NGV553140H, NGV553374H, NGV604719H, NGV601715H, NGV553099H, NGV601355H, NGV601868H, NGV603402H, NGV605917H, NGV607390H, NGV610123H, NGV610780H, NGV612648H, NGV600696H, NGV602443H, NGV602462H, NGV607953H, NGV610733H, NGV612700H, NGV612705H, NGV612915H, NGV612916H, NGV612176H, NGV612403H, NGV609057H, NGV600441H, NGV601524H, NGV604079H, NGV604118H, NGV609911H, NGV600495H, NGV611158H, NGV611610H, NGV553257H, NGV600347H, NGV600540H, NGV602553H, NGV602702H, NGV612565H, NGV552628H, NGV607676H, NGV608776H, NGV608043H, NGV552971H, NGV607256H, NGV601912H, NGV605933H, NGV612813H, NGV607125H, NGV606251H, NGV553065H, NGV600389H, NGV602377H, NGV609435H, NGV600243H, NGV600430H, NGV602354H, NGV609608H, NGV602844H, NGV605215H, NGV603051H, NGV604975H, NGV609969H, NGV611773H, NGV612951H, NGV602637H, NGV611697H, NGV603085H, NGV553295H, NGV553315H, NGV600469H, NGV602936H, NGV603015H, NGV607801H, NGV607802H, NGV608293H, NGV609478H, NGV610240H, NGV610737H, NGV611020H, NGV611206H, NGV611324H, NGV606229H, NGV609304H, NGV609919H, NGV606541H, NGV552586H, NGV607621H, NGV607720H, NGV600584H, NGV602588H, NGV608635H, NGV553195H, NGV553452H, NGV602697H, NGV603983H, NGV600174H, NGV600293H, NGV600753H, NGV602270H, NGV602586H, NGV603807H, NGV604599H, NGV604776H, NGV605019H, NGV606024H, NGV606269H, NGV608141H, NGV610091H, NGV612655H, NGV613082H, NGV605077H, NGV553602H, NGV603791H, NGV606948H, NGV600003H, NGV600366H, NGV603251H, NGV603263H, NGV603265H, NGV607504H, NGV608504H, NGV609041H, NGV553239H, NGV604938H, NGV606783H, NGV608006H, NGV604069H, NGV601463H, NGV605960H, NGV553108H, NGV606232H, NGV608577H, NGV611778H, NGV554112H, NGV554230H, NGV554235H, NGV554240H, NGV600345H, NGV600399H, NGV600639H, NGV600642H, NGV600643H, NGV600767H, NGV600769H, NGV601404H, NGV601405H, NGV601406H, NGV601411H, NGV601414H, NGV601419H, NGV601434H, NGV602850H, NGV602934H, NGV602935H, NGV602937H, NGV603031H, NGV603038H, NGV603053H, NGV603162H, NGV603587H, NGV603588H, NGV603590H, NGV603861H, NGV603865H, NGV603917H, NGV604063H, NGV604074H, NGV604075H, NGV604078H, NGV604088H, NGV604382H, NGV604658H, NGV604765H, NGV604925H, NGV604931H, NGV604934H, NGV604998H, NGV605105H, NGV605107H, NGV605226H, NGV605237H, NGV605241H, NGV605310H, NGV605311H, NGV605312H, NGV605313H, NGV605347H, NGV605348H, NGV605350H, NGV605351H, NGV605356H, NGV605961H, NGV606047H, NGV606048H, NGV606049H, NGV606059H, NGV606060H, NGV606858H, NGV607560H, NGV607957H, NGV607958H, NGV608287H, NGV608289H, NGV608564H, NGV608566H, NGV608575H, NGV608587H, NGV608999H, NGV609000H, NGV609001H, NGV609002H, NGV609652H, NGV609656H, NGV609657H, NGV609659H, NGV609660H, NGV609662H, NGV609735H, NGV609759H, NGV609760H, NGV609762H, NGV609763H, NGV609764H, NGV609765H, NGV610302H, NGV610314H, NGV610315H, NGV610316H, NGV610323H, NGV610894H, NGV610902H, NGV610924H, NGV610925H, NGV610952H, NGV610955H, NGV611441H, NGV611449H, NGV611450H, NGV611451H, NGV611452H, NGV611453H, NGV611455H, NGV611456H, NGV611457H, NGV611459H, NGV611470H, NGV612153H, NGV612365H, NGV612366H, NGV612369H, NGV612588H, NGV612632H, NGV612691H, NGV612692H, NGV612693H, NGV553400H, NGV600432H, NGV605992H, NGV606998H, NGV611632H, NGV552720H, NGV552880H, NGV553235H, NGV600231H, NGV601242H, NGV601907H, NGV602726H, NGV602735H, NGV605819H, NGV607096H, NGV607188H, NGV607712H, NGV608393H, NGV609253H, NGV609288H, NGV609471H, NGV610125H, NGV610149H, NGV610748H, NGV612157H, NGV612568H, NGV612633H, NGV613183H, NGV600567H, NGV601734H, NGV601991H, NGV602494H, NGV603846H, NGV604356H, NGV606399H, NGV607023H, NGV607123H, NGV607796H, NGV607856H, NGV608615H, NGV608617H, NGV608819H, NGV608953H, NGV609442H, NGV609988H, NGV610628H, NGV610951H, NGV612698H, NGV553002H, NGV553402H, NGV554384H, NGV600073H, NGV600191H, NGV601605H, NGV601751H, NGV601761H, NGV601772H, NGV601783H, NGV601903H, NGV602160H, NGV603050H, NGV603793H, NGV603800H, NGV609324H, NGV609327H, NGV609810H, NGV609897H, NGV610650H, NGV612226H, NGV613170H, NGV554425H, NGV605006H, NGV600661H, NGV601453H, NGV612907H, NGV609314H, NGV600581H, NGV602092H, NGV603066H, NGV600314H, NGV612801H, NGV604825H, NGV552646H, NGV600709H, NGV600849H, NGV600951H, NGV602669H, NGV602671H, NGV603211H, NGV604156H, NGV605329H, NGV605332H, NGV605333H, NGV605427H, NGV606664H, NGV607118H, NGV612642H, NGV605766H, NGV552647H, NGV601397H, NGV612484H, NGV608840H, NGV600303H, NGV602357H, NGV602362H, NGV603507H, NGV607380H, NGV607438H, NGV607902H, NGV609030H, NGV609958H, NGV612172H, NGV554210H, NGV600390H, NGV600791H, NGV604009H, NGV604010H, NGV604792H, NGV606995H, NGV612848H, NGV613353H, NGV554260H, NGV613054H, NGV554262H, NGV600511H, NGV600592H, NGV601874H, NGV602147H, NGV602150H, NGV602971H, NGV603299H, NGV603303H, NGV604816H, NGV604820H, NGV605754H, NGV606453H, NGV606457H, NGV606898H, NGV607839H, NGV610825H, NGV611473H, NGV611657H, NGV611690H, NGV613061H, NGV552758H, NGV554281H, NGV600137H, NGV600944H, NGV602848H, NGV603008H, NGV603312H, NGV603695H, NGV603698H, NGV603757H, NGV603853H, NGV603923H, NGV603928H, NGV604136H, NGV604158H, NGV604692H, NGV604699H, NGV604918H, NGV604920H, NGV605267H, NGV605594H, NGV605745H, NGV606031H, NGV606944H, NGV607288H, NGV607290H, NGV607376H, NGV607501H, NGV607707H, NGV607715H, NGV607949H, NGV608282H, NGV608283H, NGV608285H, NGV608327H, NGV608676H, NGV608787H, NGV609423H, NGV609431H, NGV609437H, NGV609439H, NGV609777H, NGV609781H, NGV609782H, NGV609875H, NGV610263H, NGV610730H, NGV610752H, NGV610805H, NGV611202H, NGV612551H, NGV612606H, NGV612616H, NGV612863H, NGV612988H, NGV613267H, NGV607614H, NGV612119H, NGV553249H, NGV554215H, NGV601985H, NGV610304H, NGV553322H, NGV554252H, NGV604827H, NGV612670H, NGV553406H, NGV553713H, NGV602434H, NGV602886H, NGV605559H, NGV607613H, NGV606362H, NGV554114H, NGV608778H, NGV609796H, NGV609993H, NGV601661H, NGV603026H, NGV607234H, NGV603017H, NGV553126H, NGV605517H, NGV601566H, NGV602657H, NGV604662H, NGV606401H, NGV612663H, NGV600549H, NGV600594H, NGV600626H, NGV600648H, NGV601006H, NGV602140H, NGV602243H, NGV604070H, NGV604365H, NGV604781H, NGV604901H, NGV605687H, NGV606419H, NGV606422H, NGV606448H, NGV606496H, NGV607940H, NGV608310H, NGV608391H, NGV608624H, NGV608655H, NGV608663H, NGV609047H, NGV609048H, NGV609050H, NGV610294H, NGV610632H, NGV611215H, NGV611275H, NGV611763H, NGV612135H, NGV612543H, NGV612677H, NGV612952H, NGV608781H, NGV611551H, NGV607753H, NGV609786H, NGV552771H, NGV607750H, NGV610758H, NGV552999H, NGV602472H, NGV604120H, NGV604780H, NGV605522H, NGV606193H, NGV607900H, NGV608816H, NGV609948H, NGV554193H, NGV600058H, NGV600142H, NGV600465H, NGV602384H, NGV602385H, NGV602386H, NGV602678H, NGV603554H, NGV605955H, NGV607314H, NGV608674H, NGV609463H, NGV609811H, NGV613103H, NGV600341H, NGV607104H, NGV607923H, NGV608957H, NGV612595H, NGV552882H, NGV606128H, NGV554256H, NGV604082H, NGV604336H, NGV604708H, NGV608623H, NGV608966H, NGV610974H, NGV601957H, NGV609714H, NGV610228H, NGV600805H, NGV604080H, NGV611575H, NGV553714H, NGV601497H, NGV602288H, NGV605021H, NGV605085H, NGV608864H, NGV608871H, NGV608872H, NGV608874H, NGV608875H, NGV608884H, NGV613159H, NGV605084H, NGV612620H, NGV606731H, NGV552704H, NGV607416H, NGV607932H, NGV607472H, NGV607937H, NGV609643H, NGV610049H, NGV610050H, NGV612121H, NGV612131H, NGV612625H, NGV613108H, NGV613413H, NGV552819H, NGV606372H, NGV607778H, NGV553732H, NGV554165H, NGV600680H, NGV602496H, NGV602498H, NGV603259H, NGV603260H, NGV610740H, NGV612709H, NGV602146H, NGV602953H, NGV606345H, NGV610969H, NGV609573H, NGV606075H, NGV612933H, NGV602491H, NGV553335H, NGV602278H, NGV602326H, NGV603627H, NGV608551H, NGV613269H, NGV611358H, NGV602591H, NGV553112H, NGV553227H, NGV553365H, NGV600785H, NGV602708H, NGV603596H, NGV605192H, NGV605544H, NGV607804H, NGV610210H, NGV607195H, NGV606095H, NGV601344H, NGV552653H, NGV554223H, NGV601776H, NGV604517H, NGV610013H, NGV603138H, NGV552617H, NGV602142H, NGV553243H, NGV553705H, NGV604850H, NGV606991H, NGV608622H, NGV609815H, NGV612524H, NGV600457H, NGV601926H, NGV602926H, NGV603765H, NGV613114H, NGV608381H, NGV609686H, NGV610054H, NGV610986H, NGV600442H, NGV553258H, NGV600145H, NGV600815H, NGV605989H, NGV606647H, NGV611101H, NGV613288H, NGV602236H, NGV602201H, NGV604788H, NGV604857H, NGV606699H, NGV612081H, NGV613245H, NGV613283H, NGV554320H, NGV600601H, NGV600732H, NGV603228H, NGV603705H, NGV605346H, NGV606658H, NGV607295H, NGV608511H, NGV612015H, NGV603488H, NGV554164H, NGV600633H, NGV608120H, NGV610008H, NGV610011H, NGV610208H, NGV612971H, NGV612627H, NGV600506H, NGV552998H, NGV602563H, NGV610648H, NGV613089H, NGV610824H, NGV602135H, NGV605866H, NGV607481H, NGV611128H, NGV612518H, NGV554178H, NGV601363H, NGV601374H, NGV602211H, NGV605276H, NGV606659H, NGV608117H, NGV609311H, NGV610318H, NGV600325H, NGV600712H, NGV602223H, NGV606098H, NGV610743H, NGV612387H, NGV607916H, NGV601438H, NGV601446H, NGV602144H, NGV603965H, NGV607405H, NGV608083H, NGV609387H, NGV610849H, NGV611745H, NGV600737H, NGV609310H, NGV612293H, NGV600222H, NGV601509H, NGV601512H, NGV601514H, NGV602165H, NGV603595H, NGV604006H, NGV604486H, NGV604488H, NGV607077H, NGV607514H, NGV607755H, NGV607963H, NGV608063H, NGV609254H, NGV600856H, NGV601748H, NGV602860H, NGV607336H, NGV608842H, NGV606777H, NGV601788H, NGV602966H, NGV603986H, NGV609069H, NGV613176H, NGV607984H, NGV552686H, NGV603789H, NGV606254H, NGV608344H, NGV609222H, NGV605660H, NGV607913H, NGV602074H, NGV600319H, NGV600360H, NGV603120H, NGV613044H, NGV554410H, NGV554411H, NGV602711H, NGV602713H, NGV602714H, NGV602715H, NGV602718H, NGV602955H, NGV602956H, NGV603988H, NGV604956H, NGV605645H, NGV606133H, NGV606134H, NGV607075H, NGV607079H, NGV607080H, NGV607082H, NGV607083H, NGV610174H, NGV610246H, NGV612624H, NGV554120H, NGV600654H, NGV602806H, NGV603238H, NGV604568H, NGV605103H, NGV606314H, NGV607924H, NGV612229H, NGV611479H, NGV603630H, NGV604122H, NGV606909H, NGV602265H, NGV602879H, NGV606665H, NGV609685H, NGV553260H, NGV554222H, NGV612974H, NGV602148H, NGV611242H, NGV602120H, NGV604663H, NGV553021H, NGV600597H, NGV606388H, NGV600687H, NGV600688H, NGV602903H, NGV603658H, NGV604388H, NGV608049H, NGV608789H, NGV608802H, NGV609888H, NGV610926H, NGV611723H, NGV612085H, NGV612841H, NGV612845H, NGV603065H, NGV611933H, NGV602710H, NGV602716H, NGV602717H, NGV604058H, NGV604545H, NGV604551H, NGV604573H, NGV604625H, NGV606429H, NGV608037H, NGV609412H, NGV609419H, NGV610161H, NGV611739H, NGV612073H, NGV612257H, NGV600854H, NGV607982H, NGV613097H, NGV552636H, NGV554124H, NGV554237H, NGV602306H, NGV604468H, NGV605480H, NGV605506H, NGV605582H, NGV605883H, NGV606199H, NGV606476H, NGV606477H, NGV606482H, NGV606595H, NGV607341H, NGV608101H, NGV608684H, NGV608800H, NGV610313H, NGV610370H, NGV610960H, NGV612519H, NGV612549H, NGV604348H, NGV611769H, NGV611001H, NGV604151H, NGV610928H, NGV601881H, NGV605112H, NGV605647H, NGV605673H, NGV606107H, NGV608276H, NGV609378H, NGV610734H, NGV610936H, NGV552374H, NGV604731H, NGV604978H, NGV605925H, NGV608195H, NGV608552H, NGV613080H, NGV602403H, NGV603723H, NGV604959H, NGV605975H, NGV608783H, NGV608784H, NGV608785H, NGV608786H, NGV609915H, NGV610037H, NGV610039H, NGV610365H, NGV612075H, NGV612460H, NGV602550H, NGV609809H, NGV611614H, NGV612925H, NGV613207H, NGV609967H, NGV610552H, NGV606436H, NGV606717H, NGV607918H, NGV610753H, NGV612690H, NGV605273H, NGV602827H, NGV607197H, NGV612480H, NGV604519H, NGV604135H, NGV606249H, NGV606400H, NGV611295H, NGV608574H, NGV612371H, NGV612508H, NGV612837H, NGV607975H, NGV601755H, NGV553238H, NGV553415H, NGV608720H, NGV608722H, NGV608159H, NGV600673H, NGV600422H, NGV553155H, NGV605416H, NGV607538H, NGV609527H, NGV601038H, NGV601844H, NGV601859H, NGV604465H, NGV609793H, NGV607066H, NGV604520H, NGV601563H, NGV554089H, NGV602912H, NGV609676H, NGV601557H, NGV554325H, NGV553178H, NGV600060H, NGV608031H.NGV552574H, NGV554133H, NGV609937H, NGV600752H, NGV603076H, NGV601279H, NGV608051H, NGV552566H, NGV602870H, NGV609483H, NGV600869H, NGV608361H, NGV604533H, NGV602235H, NGV606803H, NGV602518H, NGV602520H, NGV608532H, NGV608809H, NGV608021H, NGV609290H, NGV553134H, NGV605161H, NGV606346H, NGV553277H, NGV604453H, NGV609770H, NGV552886H, NGV552925H, NGV554087H, NGV553777H, NGV601266H, NGV602005H, NGV604732H, NGV605682H, NGV605683H, NGV602351H, NGV604930H, NGV612599H, NGV609317H, NGV600309H, NGV603189H, NGV603334H, NGV608492H, NGV600076H, NGV603759H, NGV609710H, NGV611248H, NGV607111H, NGV609067H, NGV600427H, NGV602310H, NGV607292H, NGV610038H, NGV552906H, NGV601504H, NGV553209H, NGV604091H, NGV612699H, NGV602256H, NGV604726H, NGV604937H, NGV604962H, NGV608858H, NGV608569H, NGV608572H, NGV553063H, NGV600919H, NGV608065H, NGV602832H, NGV602833H, NGV607105H, NGV608747H, NGV601801H, NGV602342H, NGV606489H, NGV604478H, NGV554388H, NGV609682H, NGV608309H, NGV553653H, NGV610211H, NGV611995H, NGV610413H, NGV604395H, NGV603716H, NGV605632H, NGV607207H, NGV602108H, NGV554110H, NGV610235H, NGV604590H, NGV609739H, NGV610385H, NGV610386H, NGV604145H, NGV600875H, NGV607854H, NGV603229H, NGV611045H, NGV604727H, NGV610687H, NGV602161H, NGV554091H, NGV606377H, NGV553040H, NGV604678H, NGV604778H, NGV612216H, NGV609894H, NGV601098H, NGV605974H, NGV607898H, NGV553110H, NGV608951H, NGV600582H, NGV607964H, NGV602476H, NGV607179H, NGV607180H, NGV607183H, NGV608923H, NGV554056H, NGV552930H, NGV601785H, NGV605165H, NGV605852H, NGV552797H, NGV552798H, NGV553564H, NGV602482H, NGV608058H, NGV604139H, NGV600880H, NGV606326H, NGV601071H, NGV552749H, NGV554143H, NGV601508H, NGV601910H, NGV611566H, NGV600668H, NGV602754H, NGV603774H, NGV601827H, NGV552712H, NGV607869H, NGV612600H, NGV612601H, NGV604799H, NGV601126H, NGV602457H, NGV600713H, NGV606761H, NGV552729H, NGV604425H, NGV606507H, NGV609398H, NGV609403H, NGV609408H, NGV609556H, NGV609578H, NGV601689H, NGV609281H, NGV609628H, NGV607058H, NGV609179H, NGV609551H, NGV601615H, NGV607042H, NGV607061H, NGV609076H, NGV609107H, NGV609114H, NGV605213H, NGV611265H, NGV605858H, NGV600348H, NGV611044H, NGV600212H, NGV612767H, NGV600189H, NGV607028H, NGV609416H, NGV552602H, NGV553206H, NGV604798H, NGV608717H, NGV612975H, NGV602589H, NGV602720H, NGV607320H, NGV603013H, NGV600285H, NGV606170H, NGV600700H, NGV603336H, NGV602198H, NGV611222H, NGV608150H, NGV609512H, NGV605109H, NGV606046H, NGV605721H, NGV554432H, NGV552907H, NGV609970H, NGV552742H, NGV554299H, NGV610340H, NGV600637H, NGV613166H, NGV608886H, NGV553113H, NGV603400H, NGV606236H, NGV611231H, NGV553775H, NGV607196H, NGV554370H, NGV607671H, NGV600493H, NGV600985H, NGV612659H, NGV604491H, NGV608668H, NGV610970H, NGV600211H; GTIN 00643169630512, Serial Numbers: NGV703680H, NGV703828H, NGV703832H, NGV704329H, NGV704396H, NGV704455H, NGV704456H, NGV704944H, NGV705025H, NGV705672H, NGV706275H, NGV706276H, NGV706277H, NGV707072H, NGV707073H, NGV707074H, NGV707075H, NGV707534H, NGV707535H, NGV707622H, NGV707623H, NGV707625H, NGV707870H, NGV707871H, NGV707872H, NGV707873H, NGV708492H, NGV708493H, NGV708494H, NGV708495H, NGV708496H, NGV708497H, NGV708668H, NGV708669H, NGV708670H, NGV708671H, NGV708672H, NGV708673H, NGV708675H, NGV708676H, NGV708678H, NGV708679H, NGV708680H, NGV708682H, NGV708683H, NGV708684H, NGV708685H, NGV708688H, NGV708689H, NGV708690H, NGV708692H, NGV708693H, NGV708694H, NGV708695H, NGV708696H, NGV708697H, NGV708698H, NGV708699H, NGV708700H, NGV708701H, NGV708702H, NGV708703H, NGV708704H, NGV708705H, NGV708706H, NGV708707H, NGV708708H, NGV708709H, NGV708710H, NGV708711H, NGV708713H, NGV708714H, NGV708715H, NGV708716H, NGV708717H, NGV708718H, NGV708719H, NGV708720H, NGV708721H, NGV708722H, NGV708723H, NGV708724H, NGV708725H, NGV708726H, NGV708727H, NGV708728H, NGV708729H, NGV708730H, NGV708731H, NGV708733H, NGV708734H, NGV708735H, NGV708736H, NGV708737H, NGV708738H, NGV708739H, NGV708740H, NGV708741H, NGV708742H, NGV708745H, NGV708747H, NGV708748H, NGV708749H, NGV708750H, NGV708751H, NGV708752H, NGV708753H, NGV708754H, NGV708755H, NGV708756H, NGV708757H, NGV708758H, NGV708759H, NGV708760H, NGV708761H, NGV708762H, NGV708763H, NGV708764H, NGV708875H, NGV708877H, NGV708878H, NGV708879H, NGV708880H, NGV708881H, NGV708894H, NGV708895H, NGV709134H, NGV709157H, NGV709159H, NGV709160H, NGV709251H, NGV709266H, NGV709298H, NGV709564H, NGV709569H, NGV709613H, NGV709958H, NGV710146H, NGV710147H, NGV710148H, NGV710149H, NGV710151H, NGV710152H, NGV710153H, NGV710155H, NGV710157H, NGV710158H, NGV710159H, NGV710160H, NGV710162H, NGV710163H, NGV710164H, NGV710165H, NGV710166H, NGV710167H, NGV710168H, NGV710169H, NGV710170H, NGV710172H, NGV710173H, NGV710174H, NGV710175H, NGV710176H, NGV710177H, NGV710179H, NGV710180H, NGV710181H, NGV710182H, NGV710183H, NGV713621H, NGV713622H, NGV713623H, NGV713628H, NGV713629H, NGV713630H, NGV713631H, NGV713895H, NGV717680H, NGV717681H, NGV717771H, NGV717777H, NGV717779H, NGV717787H, NGV717790H, NGV718724H, NGV718725H, NGV718730H, NGV718739H, NGV718744H, NGV718750H, NGV718754H, NGV718758H, NGV718759H, NGV718765H, NGV718767H, NGV718768H, NGV718769H, NGV718770H, NGV718771H, NGV718772H, NGV718773H, NGV718775H, NGV718776H, NGV718777H, NGV718784H, NGV718785H, NGV718788H, NGV718799H, NGV718801H, NGV718802H, NGV718804H, NGV718805H, NGV718806H, NGV718807H, NGV718809H, NGV718810H, NGV718811H, NGV719749H, NGV719750H, NGV719751H, NGV719752H, NGV720274H, NGV720275H, NGV720278H, NGV720279H, NGV720281H, NGV720282H, NGV720283H, NGV720286H, NGV720289H, NGV721427H, NGV722101H, NGV722105H, NGV722109H, NGV722110H, NGV722111H, NGV722112H, NGV722113H, NGV722114H, NGV722115H, NGV722116H, NGV722117H, NGV722118H, NGV722119H, NGV722120H, NGV722122H, NGV722123H, NGV722124H, NGV722125H, NGV722126H, NGV722171H, NGV722178H, NGV722182H, NGV722185H, NGV722190H, NGV722191H, NGV722194H, NGV722198H, NGV722200H, NGV722202H, NGV722216H, NGV722222H, NGV722226H, NGV722230H, NGV722232H, NGV722273H, NGV722283H, NGV722290H, NGV722296H, NGV722312H, NGV722314H, NGV722330H, NGV722333H, NGV722343H, NGV722344H, NGV722348H, NGV722350H, NGV722353H, NGV722615H, NGV723474H, NGV723475H, NGV723476H, NGV723780H, NGV723781H, NGV723784H, NGV723785H, NGV723788H, NGV723789H, NGV723790H, NGV723791H, NGV725125H, NGV725127H, NGV725131H, NGV725133H, NGV725135H, NGV725136H, NGV725138H, NGV725139H, NGV725140H, NGV725141H, NGV725142H, NGV725144H, NGV725145H, NGV725146H, NGV725147H, NGV725148H, NGV725449H, NGV725451H, NGV725452H, NGV725453H, NGV725455H, NGV725456H, NGV726257H, NGV726315H, NGV726437H, NGV726808H, NGV726809H, NGV726811H, NGV726816H, NGV726817H, NGV726818H, NGV726825H, NGV726828H, NGV726830H, NGV727080H, NGV727084H, NGV727085H, NGV727086H, NGV727087H, NGV727088H, NGV727089H, NGV727090H, NGV727091H, NGV727094H, NGV727095H, NGV727096H, NGV727097H, NGV727098H, NGV727101H, NGV727103H, NGV727105H, NGV727521H, NGV727522H, NGV727523H, NGV727713H, NGV727714H, NGV727715H, NGV727716H, NGV727717H, NGV727718H, NGV727719H, NGV727720H, NGV727721H, NGV727722H, NGV727723H, NGV727724H, NGV727725H, NGV727726H, NGV727727H, NGV726807H, NGV726810H, NGV726814H, NGV726819H, NGV726815H, NGV725923H, NGV726805H, NGV726806H, NGV726813H, NGV725910H, NGV708201H, NGV708202H, NGV720933H, NGV722380H, NGV725902H, NGV725879H, NGV651515H, NGV707078H, NGV709631H, NGV710357H, NGV722381H, NGV725886H, NGV725887H, NGV725888H, NGV710917H, NGV725364H, NGV725881H, NGV707450H, NGV707689H, NGV708203H, NGV710055H, NGV725883H, NGV725904H, NGV000170H, NGV650188H, NGV650578H, NGV700713H, NGV703374H, NGV704247H, NGV705358H, NGV707158H, NGV708782H, NGV711157H, NGV711358H, NGV717262H, NGV717352H, NGV718356H, NGV718763H, NGV721242H, NGV721548H, NGV715936H, NGV716690H, NGV719846H, NGV725313H, NGV001219H, NGV650313H, NGV704086H, NGV721898H, NGV727604H, NGV001289H, NGV704826H, NGV713336H, NGV720849H, NGV722736H, NGV722999H, NGV723046H, NGV001848H, NGV707171H, NGV000282H, NGV000814H, NGV001077H, NGV001129H, NGV002057H, NGV003104H, NGV004229H, NGV005145H, NGV650124H, NGV650128H, NGV651314H, NGV651477H, NGV651562H, NGV651575H, NGV652233H, NGV700314H, NGV700688H, NGV700726H, NGV701951H, NGV701954H, NGV702203H, NGV702775H, NGV703346H, NGV703653H, NGV704645H, NGV705287H, NGV705288H, NGV705580H, NGV705915H, NGV706272H, NGV707019H, NGV707079H, NGV707172H, NGV707942H, NGV708584H, NGV709412H, NGV709438H, NGV709811H, NGV709977H, NGV709986H, NGV711017H, NGV712808H, NGV714385H, NGV715616H, NGV716955H, NGV716958H, NGV716959H, NGV716966H, NGV716967H, NGV717831H, NGV718294H, NGV719269H, NGV719917H, NGV720427H, NGV720938H, NGV723190H, NGV708194H, NGV713626H, NGV717651H, NGV723718H, NGV725942H, NGV001894H, NGV003838H, NGV003901H, NGV004850H, NGV702139H, NGV702143H, NGV702990H, NGV705092H, NGV706165H, NGV706252H, NGV711545H, NGV713833H, NGV714252H, NGV714253H, NGV714260H, NGV714623H, NGV717062H, NGV717867H, NGV718740H, NGV721250H, NGV722559H, NGV726094H, NGV726971H, NGV001034H, NGV002342H, NGV652078H, NGV701824H, NGV705028H, NGV720638H, NGV723484H, NGV004948H, NGV719043H, NGV001867H, NGV003360H, NGV004432H, NGV702115H, NGV703246H, NGV705077H, NGV708822H, NGV709843H, NGV709866H, NGV711631H, NGV711632H, NGV711635H, NGV712508H, NGV714310H, NGV714314H, NGV714315H, NGV715829H, NGV715843H, NGV719176H, NGV719734H, NGV719957H, NGV719958H, NGV719986H, NGV720383H, NGV720690H, NGV722874H, NGV723499H, NGV725259H, NGV725474H, NGV725504H, NGV001116H, NGV721930H, NGV725618H, NGV716875H, NGV717953H, NGV718778H, NGV725283H, NGV725381H, NGV726611H, NGV000654H, NGV709248H, NGV002360H, NGV002960H, NGV003710H, NGV003903H, NGV004066H, NGV004325H, NGV004360H, NGV005036H, NGV005269H, NGV700799H, NGV701958H, NGV702388H, NGV702984H, NGV703055H, NGV703103H, NGV703952H, NGV704637H, NGV704638H, NGV705924H, NGV707946H, NGV708589H, NGV709790H, NGV711227H, NGV711848H, NGV711873H, NGV712746H, NGV714358H, NGV714604H, NGV716698H, NGV716926H, NGV717017H, NGV717018H, NGV717211H, NGV717520H, NGV722711H, NGV723249H, NGV723266H, NGV723267H, NGV723635H, NGV724661H, NGV725864H, NGV726621H, NGV726704H, NGV727106H, NGV727780H, NGV002688H, NGV714421H, NGV720382H, NGV651407H, NGV652345H, NGV702401H, NGV705921H, NGV705970H, NGV707221H, NGV708048H, NGV712559H, NGV713888H, NGV714469H, NGV714472H, NGV714486H, NGV720907H, NGV725535H, NGV003749H, NGV704400H, NGV707954H, NGV718615H, NGV724308H, NGV724497H, NGV723678H, NGV725282H, NGV725563H, NGV004975H, NGV004976H, NGV705436H, NGV713992H, NGV713993H, NGV719018H, NGV719748H, NGV723199H, NGV726744H, NGV001126H, NGV001238H, NGV001354H, NGV001849H, NGV002200H, NGV003364H, NGV003365H, NGV005246H, NGV650652H, NGV651021H, NGV651116H, NGV652537H, NGV702572H, NGV702733H, NGV703280H, NGV705342H, NGV705371H, NGV705749H, NGV705750H, NGV706442H, NGV707478H, NGV708976H, NGV709718H, NGV710003H, NGV710053H, NGV711051H, NGV711165H, NGV711617H, NGV712004H, NGV712005H, NGV712213H, NGV712694H, NGV712738H, NGV713994H, NGV716222H, NGV716923H, NGV717489H, NGV720442H, NGV720661H, NGV720728H, NGV720729H, NGV721908H, NGV721909H, NGV722096H, NGV723177H, NGV723209H, NGV723241H, NGV725224H, NGV725350H, NGV725432H, NGV726535H, NGV726800H, NGV726801H, NGV726802H, NGV727059H, NGV000079H, NGV001014H, NGV001064H, NGV001313H, NGV001596H, NGV001764H, NGV001953H, NGV001955H, NGV002336H, NGV002959H, NGV003587H, NGV004314H, NGV004541H, NGV004986H, NGV650414H, NGV650613H, NGV650828H, NGV650903H, NGV652015H, NGV652547H, NGV701080H, NGV701557H, NGV702126H, NGV702243H, NGV702475H, NGV702691H, NGV703137H, NGV704304H, NGV705327H, NGV705434H, NGV705440H, NGV707092H, NGV707093H, NGV707219H, NGV707318H, NGV708591H, NGV708826H, NGV708864H, NGV708957H, NGV709855H, NGV710068H, NGV710281H, NGV710386H, NGV711541H, NGV711569H, NGV711679H, NGV712410H, NGV712648H, NGV712667H, NGV712756H, NGV713302H, NGV713839H, NGV713840H, NGV714201H, NGV714342H, NGV714619H, NGV714620H, NGV715469H, NGV716560H, NGV716841H, NGV717395H, NGV717755H, NGV717848H, NGV718166H, NGV718312H, NGV718892H, NGV719249H, NGV719998H, NGV720343H, NGV720398H, NGV720469H, NGV720470H, NGV720542H, NGV720549H, NGV722914H, NGV722941H, NGV723330H, NGV723694H, NGV723695H, NGV724456H, NGV724476H, NGV725426H, NGV725427H, NGV725580H, NGV725581H, NGV726105H, NGV726142H, NGV726227H, NGV726418H, NGV726889H, NGV727569H, NGV727570H, NGV705127H, NGV707273H, NGV708000H, NGV711684H, NGV711877H, NGV717754H, NGV721686H, NGV700635H, NGV700636H, NGV702076H, NGV702474H, NGV002532H, NGV706349H, NGV715245H, NGV715524H, NGV716831H, NGV717583H, NGV717821H, NGV717887H, NGV718366H, NGV720384H, NGV720884H, NGV722990H, NGV726366H, NGV650699H, NGV725530H, NGV725600H, NGV004812H, NGV000438H, NGV001713H, NGV004337H, NGV701464H, NGV705347H, NGV705348H, NGV708950H, NGV709140H, NGV709153H, NGV710076H, NGV712757H, NGV713890H, NGV717653H, NGV718817H, NGV719894H, NGV721313H, NGV721893H, NGV000234H, NGV000533H, NGV000554H, NGV000655H, NGV001060H, NGV002354H, NGV003159H, NGV004604H, NGV703200H, NGV703203H, NGV709092H, NGV709764H, NGV709912H, NGV711952H, NGV720193H, NGV724494H, NGV726245H, NGV000365H, NGV000694H, NGV001173H, NGV001174H, NGV001865H, NGV003668H, NGV004928H, NGV005251H, NGV650358H, NGV650383H, NGV650385H, NGV650477H, NGV650852H, NGV650992H, NGV651211H, NGV651237H, NGV651440H, NGV651563H, NGV703642H, NGV703692H, NGV703839H, NGV703919H, NGV705161H, NGV705355H, NGV705596H, NGV706507H, NGV706632H, NGV707388H, NGV707498H, NGV709108H, NGV709968H, NGV710361H, NGV711192H, NGV711193H, NGV711312H, NGV712564H, NGV712566H, NGV712763H, NGV712770H, NGV714248H, NGV715536H, NGV715945H, NGV715947H, NGV717482H, NGV718011H, NGV718012H, NGV718014H, NGV718234H, NGV718266H, NGV719297H, NGV719299H, NGV719399H, NGV721164H, NGV721865H, NGV721867H, NGV723008H, NGV723333H, NGV723372H, NGV724258H, NGV724261H, NGV724862H, NGV725370H, NGV725591H, NGV726411H, NGV726903H, NGV727107H, NGV727108H, NGV004941H, NGV651863H, NGV702679H, NGV702690H, NGV704989H, NGV705015H, NGV707122H, NGV709244H, NGV709293H, NGV710013H, NGV710187H, NGV710189H, NGV712430H, NGV715477H, NGV715479H, NGV716996H, NGV719736H, NGV719738H, NGV720756H, NGV723274H, NGV725333H, NGV727029H, NGV727037H, NGV727603H, NGV650663H, NGV651036H, NGV706405H, NGV710178H, NGV718241H, NGV719510H, NGV004747H, NGV005069H, NGV005255H, NGV707228H, NGV714311H, NGV715549H, NGV003070H, NGV707468H, NGV711096H, NGV719123H, NGV725613H, NGV000996H, NGV001892H, NGV002512H, NGV002578H, NGV002683H, NGV002684H, NGV004997H, NGV005014H, NGV650098H, NGV650203H, NGV651155H, NGV651408H, NGV652642H, NGV701701H, NGV701788H, NGV701983H, NGV702485H, NGV703323H, NGV703565H, NGV703579H, NGV703587H, NGV703652H, NGV704892H, NGV705301H, NGV707035H, NGV707554H, NGV707854H, NGV709661H, NGV710578H, NGV714001H, NGV714002H, NGV716146H, NGV717016H, NGV717485H, NGV717823H, NGV718436H, NGV718448H, NGV719295H, NGV721323H, NGV721812H, NGV722814H, NGV723597H, NGV723664H, NGV725294H, NGV725691H, NGV726386H, NGV726467H, NGV726953H, NGV727690H, NGV003994H, NGV651202H, NGV652104H, NGV700427H, NGV701725H, NGV703375H, NGV705318H, NGV706905H, NGV707555H, NGV707571H, NGV710211H, NGV710213H, NGV710405H, NGV713506H, NGV715940H, NGV716142H, NGV716218H, NGV719989H, NGV720582H, NGV722025H, NGV722848H, NGV723016H, NGV724646H, NGV725447H, NGV727018H, NGV003404H, NGV650416H, NGV650708H, NGV651077H, NGV700122H, NGV701094H, NGV701907H, NGV705619H, NGV707142H, NGV707347H, NGV708319H, NGV711897H, NGV712488H, NGV714152H, NGV714155H, NGV715717H, NGV716894H, NGV717677H, NGV717969H, NGV721758H, NGV003444H, NGV703901H, NGV712775H, NGV716123H, NGV718345H, NGV720721H, NGV722027H, NGV725329H, NGV725330H, NGV725433H, NGV700859H, NGV710144H, NGV708043H, NGV710125H, NGV717931H, NGV724924H, NGV726896H, NGV000268H, NGV002394H, NGV005237H, NGV005274H, NGV650715H, NGV650720H, NGV700470H, NGV704460H, NGV705225H, NGV710390H, NGV711132H, NGV711133H, NGV716251H, NGV716717H, NGV718378H, NGV718383H, NGV718447H, NGV719246H, NGV719275H, NGV719703H, NGV721014H, NGV723925H, NGV725495H, NGV727580H, NGV720471H, NGV001847H, NGV002031H, NGV652481H, NGV700273H, NGV703223H, NGV704353H, NGV705722H, NGV706090H, NGV710271H, NGV711023H, NGV712552H, NGV712863H, NGV700824H, NGV703073H, NGV703697H, NGV706088H, NGV706982H, NGV708977H, NGV711396H, NGV717599H, NGV718970H, NGV719436H, NGV719785H, NGV721662H, NGV721682H, NGV723585H, NGV726506H, NGV002698H, NGV002699H, NGV003457H, NGV004126H, NGV004130H, NGV004133H, NGV004742H, NGV650471H, NGV650472H, NGV651055H, NGV651069H, NGV651376H, NGV700101H, NGV700105H, NGV700226H, NGV700613H, NGV701131H, NGV702588H, NGV703312H, NGV703313H, NGV703321H, NGV704719H, NGV705255H, NGV705624H, NGV705885H, NGV705950H, NGV707452H, NGV710759H, NGV710760H, NGV710974H, NGV711251H, NGV711618H, NGV711795H, NGV715000H, NGV715008H, NGV717324H, NGV001075H, NGV702632H, NGV705886H, NGV705929H, NGV706423H, NGV709795H, NGV711985H, NGV714424H, NGV718814H, NGV719341H, NGV720662H, NGV726870H, NGV720591H, NGV003393H, NGV004719H, NGV705758H, NGV706688H, NGV712439H, NGV712646H, NGV712981H, NGV714023H, NGV714024H, NGV716582H, NGV716659H, NGV722440H, NGV722993H, NGV723875H, NGV708857H, NGV003305H, NGV712054H, NGV716843H, NGV001799H, NGV651100H, NGV651107H, NGV706224H, NGV709093H, NGV710126H, NGV710127H, NGV710323H, NGV711180H, NGV711844H, NGV714567H, NGV718954H, NGV719177H, NGV720692H, NGV722741H, NGV723024H, NGV723344H, NGV725522H, NGV727602H, NGV723677H, NGV726673H, NGV001127H, NGV002231H, NGV003568H, NGV003569H, NGV004056H, NGV004965H, NGV004966H, NGV651204H, NGV651351H, NGV651355H, NGV652090H, NGV652234H, NGV652469H, NGV701516H, NGV702686H, NGV703077H, NGV703248H, NGV703249H, NGV703250H, NGV703922H, NGV703923H, NGV705215H, NGV705918H, NGV705919H, NGV707351H, NGV708192H, NGV709231H, NGV711748H, NGV711805H, NGV712223H, NGV712574H, NGV715862H, NGV721347H, NGV721870H, NGV723003H, NGV723020H, NGV724499H, NGV724747H, NGV725430H, NGV725639H, NGV725640H, NGV727323H, NGV727395H, NGV727710H, NGV727711H, NGV727809H, NGV715558H, NGV719010H, NGV720328H, NGV724073H, NGV000212H, NGV001132H, NGV003874H, NGV004562H, NGV703047H, NGV705196H, NGV709666H, NGV710344H, NGV713535H, NGV717639H, NGV723794H, NGV726264H, NGV726494H, NGV726304H, NGV000207H, NGV001318H, NGV650842H, NGV700322H, NGV720653H, NGV650204H, NGV650277H, NGV651850H, NGV700106H, NGV706628H, NGV707036H, NGV709517H, NGV711311H, NGV711469H, NGV715299H, NGV716267H, NGV716587H, NGV716697H, NGV717254H, NGV722976H, NGV650135H, NGV721667H, NGV722979H, NGV727613H, NGV650698H, NGV705175H, NGV705345H, NGV707298H, NGV707935H, NGV713246H, NGV713646H, NGV713712H, NGV722841H, NGV702723H, NGV704616H, NGV709965H, NGV715502H, NGV651935H, NGV651938H, NGV651939H, NGV651942H, NGV652092H, NGV701299H, NGV702112H, NGV702397H, NGV703859H, NGV704930H, NGV704947H, NGV704950H, NGV705639H, NGV707238H, NGV707240H, NGV707243H, NGV707869H, NGV708326H, NGV712208H, NGV716010H, NGV716093H, NGV716095H, NGV718196H, NGV718918H, NGV718948H, NGV719358H, NGV720452H, NGV720453H, NGV720676H, NGV720680H, NGV725578H, NGV725916H, NGV725965H, NGV727048H, NGV727557H, NGV727706H, NGV727708H, NGV709207H, NGV717043H, NGV001135H, NGV001980H, NGV002084H, NGV002110H, NGV002490H, NGV003152H, NGV003781H, NGV004227H, NGV004466H, NGV650034H, NGV651066H, NGV651254H, NGV651384H, NGV701361H, NGV701879H, NGV702183H, NGV702442H, NGV702848H, NGV703146H, NGV703355H, NGV708184H, NGV708187H, NGV708188H, NGV712499H, NGV712530H, NGV712742H, NGV715308H, NGV715856H, NGV716583H, NGV716585H, NGV716589H, NGV717507H, NGV721389H, NGV721392H, NGV721581H, NGV721841H, NGV722680H, NGV722905H, NGV723873H, NGV724669H, NGV725296H, NGV725679H, NGV725680H, NGV725786H, NGV726216H, NGV726986H, NGV727671H, NGV001986H, NGV002147H, NGV003187H, NGV003188H, NGV003189H, NGV003191H, NGV004273H, NGV004399H, NGV004854H, NGV650825H, NGV700179H, NGV700728H, NGV704382H, NGV705093H, NGV705226H, NGV706229H, NGV706335H, NGV707292H, NGV708818H, NGV709906H, NGV709938H, NGV711851H, NGV712980H, NGV713617H, NGV715925H, NGV715971H, NGV716709H, NGV718206H, NGV719059H, NGV719402H, NGV719498H, NGV719598H, NGV719645H, NGV719718H, NGV720688H, NGV721177H, NGV721691H, NGV722099H, NGV722100H, NGV722428H, NGV723141H, NGV724729H, NGV724730H, NGV724731H, NGV725998H, NGV725999H, NGV726768H, NGV726845H, NGV702144H, NGV000633H, NGV004455H, NGV707063H, NGV707373H, NGV709653H, NGV712640H, NGV720986H, NGV721946H, NGV725217H, NGV651617H, NGV711720H, NGV726139H, NGV721588H, NGV000337H, NGV000672H, NGV001196H, NGV001739H, NGV004025H, NGV004867H, NGV005027H, NGV005028H, NGV650583H, NGV650586H, NGV650606H, NGV650874H, NGV652347H, NGV702925H, NGV703398H, NGV703401H, NGV705034H, NGV705520H, NGV705540H, NGV705673H, NGV705674H, NGV705675H, NGV708180H, NGV708952H, NGV709714H, NGV710240H, NGV711102H, NGV711286H, NGV716914H, NGV719061H, NGV720248H, NGV720252H, NGV720253H, NGV721175H, NGV724469H, NGV724473H, NGV724474H, NGV726112H, NGV726113H, NGV726115H, NGV726116H, NGV726117H, NGV726981H, NGV726983H, NGV000825H, NGV002977H, NGV003435H, NGV003718H, NGV701219H, NGV701364H, NGV705277H, NGV707152H, NGV708631H, NGV711219H, NGV713618H, NGV713619H, NGV713989H, NGV713991H, NGV715959H, NGV715967H, NGV717636H, NGV717829H, NGV718907H, NGV719491H, NGV719500H, NGV720400H, NGV721144H, NGV721251H, NGV722929H, NGV722931H, NGV722932H, NGV723038H, NGV723039H, NGV723586H, NGV725308H, NGV725560H, NGV726095H, NGV726096H, NGV709488H, NGV000887H, NGV000998H, NGV001663H, NGV002863H, NGV002864H, NGV004504H, NGV651038H, NGV652460H, NGV652473H, NGV700209H, NGV701053H, NGV701054H, NGV704104H, NGV706635H, NGV706636H, NGV715964H, NGV715976H, NGV715978H, NGV716997H, NGV717663H, NGV718887H, NGV720049H, NGV720838H, NGV720875H, NGV721362H, NGV721364H, NGV721365H, NGV721792H, NGV725187H, NGV716703H, NGV706633H, NGV714073H, NGV717074H, NGV718787H, NGV721579H, NGV722831H, NGV725571H, NGV708593H, NGV712097H, NGV652271H, NGV652273H, NGV652274H, NGV652315H, NGV703171H, NGV711204H, NGV712587H, NGV714586H, NGV717496H, NGV717497H, NGV718836H, NGV720942H, NGV722534H, NGV722934H, NGV723921H, NGV724753H, NGV725115H, NGV703149H, NGV705889H, NGV705939H, NGV705940H, NGV705942H, NGV711307H, NGV718523H, NGV721594H, NGV722981H, NGV724992H, NGV725758H, NGV727010H, NGV000036H, NGV000197H, NGV000425H, NGV000427H, NGV000447H, NGV001521H, NGV001791H, NGV002229H, NGV004836H, NGV004837H, NGV651848H, NGV702108H, NGV702295H, NGV703694H, NGV704174H, NGV705373H, NGV708603H, NGV708606H, NGV708617H, NGV708995H, NGV709762H, NGV709889H, NGV710984H, NGV714022H, NGV716570H, NGV716571H, NGV717189H, NGV717623H, NGV719346H, NGV720621H, NGV722439H, NGV722510H, NGV722677H, NGV723246H, NGV723289H, NGV723290H, NGV700788H, NGV726154H, NGV701468H, NGV722678H, NGV000649H, NGV716101H, NGV727813H, NGV003403H, NGV650925H, NGV712351H, NGV716796H, NGV719944H, NGV720092H, NGV720321H, NGV720878H, NGV722067H, NGV726587H, NGV000695H, NGV003280H, NGV003282H, NGV003939H, NGV004457H, NGV005187H, NGV700940H, NGV701454H, NGV701535H, NGV701584H, NGV701891H, NGV702498H, NGV704014H, NGV708861H, NGV711257H, NGV711821H, NGV712197H, NGV715534H, NGV715538H, NGV718937H, NGV719301H, NGV719546H, NGV719795H, NGV719797H, NGV721332H, NGV721951H, NGV723517H, NGV724740H, NGV724741H, NGV724742H, NGV726415H, NGV726888H, NGV000650H, NGV002240H, NGV700406H, NGV701972H, NGV702540H, NGV706830H, NGV709056H, NGV709532H, NGV709984H, NGV710111H, NGV711223H, NGV711842H, NGV712035H, NGV712194H, NGV716194H, NGV720261H, NGV720928H, NGV722085H, NGV001112H, NGV001676H, NGV710314H, NGV711077H, NGV711816H, NGV714456H, NGV714458H, NGV717603H, NGV714309H, NGV703162H, NGV704249H, NGV709060H, NGV715620H, NGV715672H, NGV715676H, NGV724068H, NGV725253H, NGV725477H, NGV726940H, NGV000535H, NGV000855H, NGV000951H, NGV650094H, NGV700411H, NGV708909H, NGV711909H, NGV712951H, NGV714189H, NGV714191H, NGV717391H, NGV719021H, NGV722385H, NGV000854H, NGV001012H, NGV001749H, NGV001750H, NGV002167H, NGV002896H, NGV003283H, NGV003438H, NGV003945H, NGV651146H, NGV700169H, NGV700789H, NGV701693H, NGV703193H, NGV703953H, NGV704488H, NGV704710H, NGV704839H, NGV705047H, NGV705076H, NGV705220H, NGV705230H, NGV705239H, NGV705364H, NGV705807H, NGV705974H, NGV707155H, NGV707345H, NGV707355H, NGV707467H, NGV707481H, NGV708661H, NGV708959H, NGV709421H, NGV710099H, NGV711003H, NGV711786H, NGV712061H, NGV712310H, NGV712726H, NGV713023H, NGV713690H, NGV714576H, NGV714977H, NGV715588H, NGV715597H, NGV715604H, NGV716908H, NGV716912H, NGV716993H, NGV717203H, NGV718608H, NGV718609H, NGV718610H, NGV720345H, NGV720401H, NGV720594H, NGV720636H, NGV720891H, NGV720989H, NGV722014H, NGV723612H, NGV723842H, NGV723883H, NGV725612H, NGV726514H, NGV726516H, NGV726957H, NGV727153H, NGV652199H, NGV705397H, NGV712724H, NGV004201H, NGV713957H, NGV713960H, NGV716989H, NGV724451H, NGV725273H, NGV700352H, NGV700451H, NGV702150H, NGV719001H, NGV726990H, NGV000381H, NGV000720H, NGV001765H, NGV001977H, NGV002540H, NGV002709H, NGV003659H, NGV003728H, NGV004141H, NGV004426H, NGV004761H, NGV004814H, NGV650483H, NGV650666H, NGV650709H, NGV700237H, NGV700637H, NGV701799H, NGV702054H, NGV702425H, NGV702576H, NGV702841H, NGV706284H, NGV706417H, NGV706469H, NGV706473H, NGV706648H, NGV706808H, NGV706937H, NGV706957H, NGV709072H, NGV709628H, NGV710060H, NGV712454H, NGV712797H, NGV714066H, NGV714070H, NGV715315H, NGV715405H, NGV715422H, NGV716998H, NGV717247H, NGV717668H, NGV717669H, NGV717875H, NGV719625H, NGV719682H, NGV720861H, NGV721819H, NGV723057H, NGV724048H, NGV724192H, NGV724269H, NGV725875H, NGV726655H, NGV726932H, NGV727006H, NGV727703H, NGV003881H, NGV650095H, NGV708608H, NGV719552H, NGV715119H, NGV704874H, NGV709282H, NGV716595H, NGV004064H, NGV002184H, NGV701737H, NGV707297H, NGV709289H, NGV711181H, NGV715652H, NGV722779H, NGV723527H, NGV726646H, NGV716557H, NGV716886H, NGV720914H, NGV000063H, NGV651370H, NGV700296H, NGV702697H, NGV705233H, NGV719712H, NGV721207H, NGV721344H, NGV723824H, NGV710225H, NGV716457H, NGV719730H, NGV000973H, NGV001047H, NGV003056H, NGV004357H, NGV652533H, NGV652534H, NGV700374H, NGV701536H, NGV703320H, NGV704978H, NGV706024H, NGV707069H, NGV709018H, NGV709422H, NGV709934H, NGV710128H, NGV710440H, NGV711903H, NGV711904H, NGV711905H, NGV711906H, NGV711908H, NGV713701H, NGV713702H, NGV715840H, NGV715861H, NGV715868H, NGV719363H, NGV719366H, NGV723292H, NGV723598H, NGV724213H, NGV725772H, NGV726143H, NGV726144H, NGV726145H, NGV726626H, NGV000413H, NGV000481H, NGV000699H, NGV000800H, NGV001922H, NGV001923H, NGV002904H, NGV002905H, NGV003654H, NGV004411H, NGV004412H, NGV651882H, NGV700383H, NGV700552H, NGV701896H, NGV701897H, NGV702738H, NGV703154H, NGV703155H, NGV703278H, NGV705247H, NGV705935H, NGV706619H, NGV706621H, NGV707003H, NGV707004H, NGV708779H, NGV709237H, NGV709238H, NGV709322H, NGV709937H, NGV710228H, NGV710347H, NGV711340H, NGV711341H, NGV711342H, NGV712362H, NGV712368H, NGV712494H, NGV712496H, NGV712743H, NGV712810H, NGV712824H, NGV714383H, NGV714605H, NGV714606H, NGV716550H, NGV716551H, NGV716623H, NGV716631H, NGV716896H, NGV717165H, NGV717168H, NGV717218H, NGV717398H, NGV717621H, NGV717868H, NGV717869H, NGV718210H, NGV718796H, NGV718839H, NGV719218H, NGV719222H, NGV719930H, NGV719932H, NGV719933H, NGV720152H, NGV720159H, NGV720698H, NGV720892H, NGV721298H, NGV721465H, NGV721493H, NGV722887H, NGV722968H, NGV723011H, NGV723315H, NGV723697H, NGV724305H, NGV724644H, NGV724719H, NGV725035H, NGV725102H, NGV725957H, NGV725959H, NGV725960H, NGV726226H, NGV726278H, NGV727383H, NGV727385H, NGV727561H, NGV650462H, NGV723193H, NGV705483H, NGV705349H, NGV002324H, NGV003873H, NGV004763H, NGV703569H, NGV714337H, NGV715749H, NGV715750H, NGV715753H, NGV716301H, NGV716845H, NGV717581H, NGV717956H, NGV720084H, NGV722910H, NGV723253H, NGV726265H, NGV726266H, NGV726268H, NGV726916H, NGV703811H, NGV706613H, NGV712407H, NGV717858H, NGV000390H, NGV651163H, NGV001082H, NGV001728H, NGV004671H, NGV650081H, NGV650648H, NGV700259H, NGV700325H, NGV703977H, NGV704717H, NGV705654H, NGV706986H, NGV706987H, NGV711504H, NGV712063H, NGV712745H, NGV714301H, NGV715738H, NGV716186H, NGV718017H, NGV720116H, NGV720648H, NGV721198H, NGV722493H, NGV726294H, NGV726659H, NGV726695H, NGV708837H, NGV003368H, NGV004128H, NGV651000H, NGV701169H, NGV705600H, NGV715619H, NGV720477H, NGV725662H, NGV702236H, NGV715854H, NGV719735H, NGV725126H, NGV723000H, NGV723863H, NGV001181H, NGV001275H, NGV003931H, NGV700253H, NGV700291H, NGV701705H, NGV704697H, NGV707123H, NGV716736H, NGV717373H, NGV717550H, NGV717559H, NGV719655H, NGV721412H, NGV722879H, NGV724040H, NGV726064H, NGV004989H, NGV725592H, NGV707980H, NGV650242H, NGV650323H, NGV652230H, NGV709734H, NGV711385H, NGV717298H, NGV717403H, NGV004011H, NGV702607H, NGV705404H, NGV709716H, NGV719993H, NGV719995H, NGV651693H, NGV652091H, NGV652130H, NGV704239H, NGV704292H, NGV706684H, NGV707169H, NGV710237H, NGV710238H, NGV713052H, NGV714194H, NGV716519H, NGV717746H, NGV719311H, NGV720062H, NGV720145H, NGV720425H, NGV721880H, NGV725653H, NGV725715H, NGV725955H, NGV000873H, NGV650004H, NGV707338H, NGV719721H, NGV719722H, NGV720905H, NGV650405H, NGV706316H, NGV716994H, NGV000670H, NGV000693H, NGV001837H, NGV002306H, NGV651765H, NGV700262H, NGV700620H, NGV708919H, NGV709802H, NGV715546H, NGV717263H, NGV002226H, NGV720474H, NGV725484H, NGV001285H, NGV002177H, NGV002772H, NGV002877H, NGV004372H, NGV650145H, NGV702123H, NGV706357H, NGV709873H, NGV713258H, NGV715993H, NGV716000H, NGV717103H, NGV717510H, NGV718291H, NGV721201H, NGV725733H, NGV726618H, NGV727678H, NGV723955H, NGV001758H, NGV001913H, NGV002424H, NGV004214H, NGV004815H, NGV004816H, NGV004883H, NGV004884H, NGV004885H, NGV004886H, NGV650774H, NGV650775H, NGV650776H, NGV700399H, NGV700400H, NGV700402H, NGV700403H, NGV700959H, NGV700961H, NGV700962H, NGV700964H, NGV700966H, NGV700968H, NGV700969H, NGV702635H, NGV702638H, NGV702654H, NGV702802H, NGV702977H, NGV703667H, NGV703668H, NGV704183H, NGV704185H, NGV704405H, NGV704410H, NGV705426H, NGV705930H, NGV705931H, NGV705933H, NGV705944H, NGV705945H, NGV705947H, NGV706314H, NGV706315H, NGV706524H, NGV706525H, NGV707030H, NGV707031H, NGV707032H, NGV707472H, NGV707474H, NGV709999H, NGV710438H, NGV710439H, NGV711119H, NGV711429H, NGV711430H, NGV711727H, NGV711827H, NGV711993H, NGV712373H, NGV712381H, NGV712426H, NGV712534H, NGV712599H, NGV712601H, NGV713294H, NGV713314H, NGV714436H, NGV714437H, NGV714438H, NGV714598H, NGV715724H, NGV715732H, NGV716240H, NGV716269H, NGV716273H, NGV716283H, NGV716707H, NGV716720H, NGV716729H, NGV716888H, NGV716889H, NGV716939H, NGV717006H, NGV717117H, NGV717118H, NGV717415H, NGV717591H, NGV717839H, NGV717965H, NGV717966H, NGV717971H, NGV717972H, NGV718569H, NGV718570H, NGV718674H, NGV718962H, NGV719185H, NGV719445H, NGV719554H, NGV719555H, NGV719733H, NGV720381H, NGV720519H, NGV720647H, NGV720853H, NGV720997H, NGV721085H, NGV721247H, NGV721559H, NGV722445H, NGV722565H, NGV723303H, NGV723453H, NGV723467H, NGV723468H, NGV723806H, NGV724228H, NGV724684H, NGV726205H, NGV726320H, NGV726357H, NGV726375H, NGV726545H, NGV726559H, NGV726774H, NGV707483H, NGV711103H, NGV711688H, NGV713961H, NGV719725H, NGV723077H, NGV725016H, NGV725307H, NGV725310H, NGV725311H, NGV726519H, NGV726894H, NGV704296H, NGV712676H, NGV713339H, NGV719041H, NGV726867H, NGV003962H, NGV004992H, NGV650411H, NGV700154H, NGV700247H, NGV700541H, NGV701650H, NGV701651H, NGV702419H, NGV703004H, NGV706342H, NGV706666H, NGV707224H, NGV711820H, NGV714176H, NGV714951H, NGV716833H, NGV718626H, NGV719594H, NGV720654H, NGV721249H, NGV721409H, NGV723566H, NGV723702H, NGV725261H, NGV726180H, NGV726733H, NGV727226H, NGV727674H, NGV704700H, NGV706860H, NGV711106H, NGV717959H, NGV001864H, NGV651346H, NGV700986H, NGV701692H, NGV705960H, NGV712760H, NGV713606H, NGV715570H, NGV718042H, NGV718300H, NGV720703H, NGV709228H, NGV712727H, NGV717122H, NGV002696H, NGV003694H, NGV708962H, NGV716360H, NGV718707H, NGV719429H, NGV720633H, NGV727629H, NGV002243H, NGV651315H, NGV651639H, NGV705799H, NGV706743H, NGV712231H, NGV713248H, NGV714291H, NGV715657H, NGV717566H, NGV717683H, NGV718005H, NGV719672H, NGV721726H, NGV000688H, NGV000228H, NGV001430H, NGV715912H, NGV725015H, NGV001887H, NGV701893H, NGV725160H, NGV725490H, NGV726476H, NGV000325H, NGV000357H, NGV000411H, NGV000541H, NGV000734H, NGV001286H, NGV001587H, NGV001677H, NGV002705H, NGV002785H, NGV003309H, NGV003995H, NGV650608H, NGV651909H, NGV704108H, NGV704642H, NGV705263H, NGV706985H, NGV707368H, NGV708616H, NGV708862H, NGV709876H, NGV709894H, NGV711034H, NGV711211H, NGV712203H, NGV712877H, NGV714172H, NGV714175H, NGV716624H, NGV717584H, NGV718624H, NGV718625H, NGV718677H, NGV718790H, NGV719325H, NGV721031H, NGV722394H, NGV723183H, NGV723939H, NGV724800H, NGV725128H, NGV703427H, NGV713345H, NGV714000H, NGV000340H, NGV002363H, NGV002615H, NGV002740H, NGV002995H, NGV002996H, NGV003057H, NGV004017H, NGV004196H, NGV004519H, NGV650076H, NGV650302H, NGV650844H, NGV651331H, NGV700398H, NGV701078H, NGV701826H, NGV702088H, NGV702602H, NGV703254H, NGV704620H, NGV705227H, NGV705352H, NGV705572H, NGV705689H, NGV705897H, NGV706337H, NGV706429H, NGV706440H, NGV706447H, NGV706993H, NGV707060H, NGV707085H, NGV707503H, NGV709059H, NGV709625H, NGV709679H, NGV709692H, NGV709775H, NGV709786H, NGV709793H, NGV709913H, NGV709917H, NGV709982H, NGV710011H, NGV710265H, NGV710282H, NGV710302H, NGV710530H, NGV710775H, NGV711280H, NGV711717H, NGV711853H, NGV711954H, NGV712649H, NGV713300H, NGV713820H, NGV713821H, NGV713823H, NGV713824H, NGV714397H, NGV714402H, NGV714409H, NGV715911H, NGV717083H, NGV717295H, NGV717320H, NGV717903H, NGV717905H, NGV717906H, NGV718276H, NGV718278H, NGV718665H, NGV719791H, NGV720060H, NGV720559H, NGV720622H, NGV720976H, NGV720977H, NGV720978H, NGV721780H, NGV721851H, NGV722041H, NGV722042H, NGV723536H, NGV723903H, NGV724332H, NGV724896H, NGV724899H, NGV725491H, NGV726553H, NGV726758H, NGV726874H, NGV727547H, NGV727595H, NGV705019H, NGV717637H, NGV707064H, NGV718533H, NGV002903H, NGV004512H, NGV005022H, NGV706045H, NGV002326H, NGV700131H, NGV704601H, NGV719866H, NGV720315H, NGV722944H, NGV000449H, NGV001248H, NGV719769H, NGV721761H, NGV001933H, NGV700243H, NGV723298H, NGV713238H, NGV719688H, NGV000487H, NGV000545H, NGV004291H, NGV004508H, NGV650077H, NGV704421H, NGV704787H, NGV705115H, NGV715733H, NGV716747H, NGV718004H, NGV726841H, NGV001766H, NGV001852H, NGV003154H, NGV650172H, NGV712217H, NGV716264H, NGV721341H, NGV722833H, NGV724283H, NGV725506H, NGV725547H, NGV725548H, NGV703682H, NGV709055H, NGV001803H, NGV001888H, NGV002708H, NGV003657H, NGV003671H, NGV004395H, NGV004397H, NGV650456H, NGV650646H, NGV652114H, NGV700992H, NGV702846H, NGV703143H, NGV703144H, NGV703647H, NGV705031H, NGV705462H, NGV705987H, NGV706952H, NGV707146H, NGV708231H, NGV712663H, NGV714076H, NGV715273H, NGV715457H, NGV715506H, NGV717255H, NGV717463H, NGV718141H, NGV718143H, NGV718144H, NGV718600H, NGV719830H, NGV719936H, NGV719942H, NGV721457H, NGV721459H, NGV721960H, NGV725322H, NGV725419H, NGV726078H, NGV000209H, NGV003844H, NGV004112H, NGV004135H, NGV004147H, NGV005209H, NGV650088H, NGV652425H, NGV652462H, NGV700524H, NGV700751H, NGV702667H, NGV704097H, NGV704173H, NGV706836H, NGV706873H, NGV706992H, NGV707087H, NGV707308H, NGV708951H, NGV710435H, NGV714074H, NGV714075H, NGV716838H, NGV722938H, NGV723706H, NGV002001H, NGV707990H, NGV000557H, NGV652171H, NGV703314H, NGV706367H, NGV714331H, NGV718199H, NGV718587H, NGV722649H, NGV725537H, NGV726853H, NGV000295H, NGV000510H, NGV004090H, NGV004640H, NGV005282H, NGV650178H, NGV650830H, NGV651090H, NGV652265H, NGV652372H, NGV700441H, NGV702160H, NGV702317H, NGV703080H, NGV703113H, NGV703274H, NGV703285H, NGV703287H, NGV703288H, NGV703945H, NGV703946H, NGV704094H, NGV704635H, NGV705209H, NGV705862H, NGV706159H, NGV706160H, NGV706723H, NGV707197H, NGV707392H, NGV707585H, NGV708772H, NGV708887H, NGV708889H, NGV709104H, NGV709680H, NGV709683H, NGV710087H, NGV710092H, NGV710093H, NGV710094H, NGV710383H, NGV711018H, NGV711449H, NGV711683H, NGV712334H, NGV712519H, NGV712520H, NGV712954H, NGV713276H, NGV713279H, NGV713675H, NGV713676H, NGV713975H, NGV713976H, NGV713977H, NGV713979H, NGV714292H, NGV714295H, NGV714296H, NGV714297H, NGV714366H, NGV714439H, NGV716255H, NGV716257H, NGV716662H, NGV716718H, NGV716919H, NGV717094H, NGV717151H, NGV717177H, NGV717667H, NGV718183H, NGV719073H, NGV719207H, NGV719254H, NGV719256H, NGV719621H, NGV719756H, NGV720006H, NGV721111H, NGV721160H, NGV721161H, NGV721163H, NGV721617H, NGV721619H, NGV722083H, NGV722084H, NGV722411H, NGV722412H, NGV722946H, NGV722948H, NGV723036H, NGV723175H, NGV723255H, NGV723511H, NGV723540H, NGV724005H, NGV724019H, NGV724097H, NGV724099H, NGV724678H, NGV724699H, NGV725415H, NGV725417H, NGV725645H, NGV725937H, NGV726187H, NGV726872H, NGV726873H, NGV727554H, NGV727620H, NGV727881H, NGV727885H, NGV727890H, NGV702595H, NGV705803H, NGV708165H, NGV718902H, NGV709473H, NGV711484H, NGV715988H, NGV724743H, NGV726586H, NGV651862H, NGV704070H, NGV706093H, NGV000049H, NGV000706H, NGV002146H, NGV003121H, NGV003860H, NGV004212H, NGV004494H, NGV004764H, NGV650571H, NGV650954H, NGV703388H, NGV705464H, NGV705470H, NGV706476H, NGV707277H, NGV707524H, NGV708503H, NGV709406H, NGV710065H, NGV711488H, NGV712402H, NGV712509H, NGV715272H, NGV715294H, NGV715304H, NGV717853H, NGV720931H, NGV720932H, NGV723884H, NGV725924H, NGV001083H, NGV702703H, NGV715392H, NGV650367H, NGV702604H, NGV709516H, NGV709638H, NGV711822H, NGV723275H, NGV724829H, NGV727701H, NGV703383H, NGV713974H, NGV719441H, NGV724291H, NGV726855H, NGV004995H, NGV701566H, NGV716701H, NGV000024H, NGV002385H, NGV004521H, NGV005234H, NGV651113H, NGV652248H, NGV703018H, NGV703238H, NGV703284H, NGV703937H, NGV704917H, NGV705508H, NGV705850H, NGV707432H, NGV707455H, NGV707636H, NGV708170H, NGV710113H, NGV710114H, NGV712037H, NGV712608H, NGV712626H, NGV713025H, NGV715438H, NGV715954H, NGV715957H, NGV716541H, NGV717537H, NGV717539H, NGV719758H, NGV722452H, NGV725459H, NGV726778H, NGV726869H, NGV727694H, NGV003697H, NGV703842H, NGV705731H, NGV725155H, NGV726570H, NGV726596H, NGV001106H, NGV004802H, NGV716730H, NGV000897H, NGV003300H, NGV003691H, NGV004086H, NGV004195H, NGV650814H, NGV651497H, NGV651931H, NGV701875H, NGV702146H, NGV702777H, NGV702784H, NGV702946H, NGV705306H, NGV705828H, NGV707033H, NGV707882H, NGV707883H, NGV707886H, NGV708660H, NGV709112H, NGV712198H, NGV712199H, NGV712376H, NGV712621H, NGV713620H, NGV716794H, NGV719892H, NGV720626H, NGV720879H, NGV721350H, NGV722392H, NGV723879H, NGV724112H, NGV726385H, NGV727024H, NGV000987H, NGV003158H, NGV003846H, NGV650823H, NGV700977H, NGV702731H, NGV705368H, NGV717071H, NGV719040H, NGV720796H, NGV723028H, NGV724651H, NGV726197H, NGV727661H, NGV004791H, NGV704777H, NGV709230H, NGV715548H, NGV718310H, NGV719760H, NGV004016H, NGV727646H, NGV712057H, NGV722396H, NGV725205H, NGV725211H, NGV726965H, NGV002495H, NGV004663H, NGV703453H, NGV711074H, NGV727654H, NGV721356H, NGV000369H, NGV003726H, NGV003773H, NGV650928H, NGV716885H, NGV004950H, NGV713698H, NGV713716H, NGV720596H, NGV704467H, NGV716202H, NGV721486H, NGV000724H, NGV004707H, NGV004834H, NGV005186H, NGV650452H, NGV650754H, NGV651201H, NGV651224H, NGV651544H, NGV652068H, NGV701964H, NGV702439H, NGV702687H, NGV703207H, NGV703245H, NGV703949H, NGV705169H, NGV705489H, NGV707181H, NGV708943H, NGV709084H, NGV709324H, NGV709621H, NGV709717H, NGV710998H, NGV712189H, NGV713331H, NGV713332H, NGV714209H, NGV714246H, NGV717990H, NGV717991H, NGV717993H, NGV718479H, NGV718632H, NGV721479H, NGV721950H, NGV721983H, NGV723029H, NGV723168H, NGV723575H, NGV725092H, NGV726208H, NGV726209H, NGV726891H, NGV727075H, NGV001706H, NGV703112H, NGV001684H, NGV700850H, NGV701445H, NGV702121H, NGV702844H, NGV706095H, NGV707119H, NGV707275H, NGV707342H, NGV707442H, NGV707567H, NGV707860H, NGV711700H, NGV712693H, NGV714388H, NGV716219H, NGV716220H, NGV716227H, NGV717004H, NGV719792H, NGV723014H, NGV723946H, NGV724542H, NGV726477H, NGV726649H, NGV726650H, NGV000139H, NGV000146H, NGV000147H, NGV000150H, NGV000151H, NGV000152H, NGV000155H, NGV002545H, NGV002549H, NGV002556H, NGV002558H, NGV002560H, NGV002561H, NGV002562H, NGV002563H, NGV005006H, NGV005008H, NGV650413H, NGV650700H, NGV652336H, NGV701220H, NGV701222H, NGV701223H, NGV701224H, NGV701226H, NGV701227H, NGV701228H, NGV704632H, NGV706100H, NGV706109H, NGV706846H, NGV706847H, NGV707025H, NGV707026H, NGV708575H, NGV708576H, NGV708578H, NGV709995H, NGV709996H, NGV710122H, NGV713987H, NGV716232H, NGV716917H, NGV719747H, NGV719954H, NGV719955H, NGV720814H, NGV720815H, NGV720821H, NGV721020H, NGV721021H, NGV721317H, NGV721318H, NGV721320H, NGV721887H, NGV721889H, NGV650568H, NGV651531H, NGV700417H, NGV701376H, NGV706228H, NGV707436H, NGV725482H, NGV000674H, NGV003409H, NGV650025H, NGV652140H, NGV700419H, NGV700420H, NGV700421H, NGV702391H, NGV708590H, NGV709320H, NGV715655H, NGV715661H, NGV717522H, NGV718073H, NGV723770H, NGV723772H, NGV726422H, NGV727627H, NGV703591H, NGV001422H, NGV003682H, NGV004109H, NGV701401H, NGV702313H, NGV704565H, NGV708157H, NGV708498H, NGV712669H, NGV712758H, NGV714625H, NGV714627H, NGV720865H, NGV722574H, NGV723034H, NGV723356H, NGV726502H, NGV726503H, NGV001443H, NGV001564H, NGV004391H, NGV650312H, NGV650851H, NGV651208H, NGV651837H, NGV702497H, NGV706345H, NGV707344H, NGV712969H, NGV715939H, NGV719959H, NGV720900H, NGV721331H, NGV723025H, NGV725121H, NGV726905H, NGV003000H, NGV705413H, NGV709265H, NGV709645H, NGV001117H, NGV001880H, NGV004243H, NGV650049H, NGV702757H, NGV707440H, NGV708251H, NGV709261H, NGV709737H, NGV709738H, NGV720456H, NGV721790H, NGV723704H, NGV723847H, NGV726398H, NGV727621H, NGV651080H, NGV706698H, NGV650119H, NGV000911H, NGV004092H, NGV005151H, NGV650987H, NGV700941H, NGV701310H, NGV701314H, NGV701821H, NGV704784H, NGV705665H, NGV707471H, NGV709519H, NGV710912H, NGV712854H, NGV715535H, NGV717267H, NGV717506H, NGV718509H, NGV719615H, NGV722978H, NGV725987H, NGV726034H, NGV726256H, NGV726258H, NGV726472H, NGV726547H, NGV726584H, NGV726585H, NGV726831H, NGV726997H, NGV727601H, NGV727624H, NGV727626H, NGV708833H, NGV712260H, NGV712261H, NGV712567H, NGV712569H, NGV714144H, NGV714207H, NGV714613H, NGV716608H, NGV717843H, NGV717880H, NGV717881H, NGV717883H, NGV718704H, NGV718781H, NGV718819H, NGV719057H, NGV722078H, NGV722079H, NGV725760H, NGV002450H, NGV002451H, NGV700204H, NGV707159H, NGV714306H, NGV717411H, NGV717504H, NGV720358H, NGV004378H, NGV717674H, NGV718745H, NGV720740H, NGV720742H, NGV720941H, NGV721019H, NGV721022H, NGV721586H, NGV721773H, NGV721793H, NGV721832H, NGV721906H, NGV721907H, NGV721997H, NGV721998H, NGV721999H, NGV722058H, NGV722059H, NGV722575H, NGV722576H, NGV722577H, NGV722578H, NGV722579H, NGV722581H, NGV722582H, NGV722583H, NGV725556H, NGV725561H, NGV726184H, NGV726185H, NGV726484H, NGV726485H, NGV726633H, NGV726634H, NGV726787H, NGV000305H, NGV703624H, NGV703625H, NGV705994H, NGV710014H, NGV714491H, NGV715265H, NGV719925H, NGV724509H, NGV724734H, NGV650325H, NGV652393H, NGV715921H, NGV723916H, NGV715507H, NGV001080H, NGV001085H, NGV001727H, NGV003050H, NGV003483H, NGV003591H, NGV003996H, NGV004009H, NGV650031H, NGV651534H, NGV703192H, NGV709904H, NGV712658H, NGV715556H, NGV719054H, NGV723163H, NGV724135H, NGV726273H, NGV725744H, NGV721829H, NGV726556H, NGV650046H, NGV706244H, NGV000945H, NGV004982H, NGV705252H, NGV705253H, NGV705800H, NGV705927H, NGV705936H, NGV705938H, NGV706219H, NGV706306H, NGV706434H, NGV706435H, NGV707267H, NGV707268H, NGV707443H, NGV709103H, NGV709105H, NGV709106H, NGV709109H, NGV709110H, NGV709250H, NGV709252H, NGV709254H, NGV710253H, NGV710908H, NGV710969H, NGV710970H, NGV710971H, NGV711238H, NGV711586H, NGV711879H, NGV711880H, NGV711881H, NGV711882H, NGV712701H, NGV712703H, NGV712816H, NGV712818H, NGV713613H, NGV714621H, NGV714622H, NGV715893H, NGV715931H, NGV716562H, NGV717140H, NGV717141H, NGV717312H, NGV717907H, NGV717908H, NGV717911H, NGV717912H, NGV718094H, NGV718691H, NGV718693H, NGV718694H, NGV718845H, NGV718846H, NGV718926H, NGV719937H, NGV719970H, NGV719971H, NGV720403H, NGV720405H, NGV721707H, NGV721709H, NGV721710H, NGV721920H, NGV721921H, NGV722509H, NGV722512H, NGV722757H, NGV722758H, NGV722759H, NGV723588H, NGV723618H, NGV723619H, NGV723708H, NGV723720H, NGV724864H, NGV725353H, NGV725820H, NGV725821H, NGV725822H, NGV725834H, NGV726149H, NGV726150H, NGV726376H, NGV726490H, NGV726703H, NGV727000H, NGV727123H, NGV727370H, NGV727803H, NGV727804H, NGV000888H, NGV002954H, NGV004974H, NGV701114H, NGV702922H, NGV707088H, NGV708609H, NGV708766H, NGV708787H, NGV710310H, NGV724240H, NGV000371H, NGV001002H, NGV001524H, NGV001569H, NGV001570H, NGV001681H, NGV001899H, NGV001932H, NGV002500H, NGV002810H, NGV002849H, NGV003341H, NGV003564H, NGV004136H, NGV004171H, NGV004749H, NGV650072H, NGV650075H, NGV650440H, NGV650585H, NGV650587H, NGV650730H, NGV650829H, NGV651132H, NGV651197H, NGV651359H, NGV651890H, NGV652354H, NGV700224H, NGV700281H, NGV700414H, NGV700495H, NGV700812H, NGV700839H, NGV701096H, NGV701277H, NGV701559H, NGV701918H, NGV702166H, NGV702393H, NGV702664H, NGV703093H, NGV703094H, NGV703378H, NGV703608H, NGV703902H, NGV704447H, NGV704606H, NGV704893H, NGV704895H, NGV704896H, NGV704897H, NGV704898H, NGV704899H, NGV704900H, NGV704901H, NGV704902H, NGV704904H, NGV704905H, NGV704906H, NGV704908H, NGV704909H, NGV706551H, NGV707149H, NGV707150H, NGV707621H, NGV708532H, NGV708533H, NGV708534H, NGV708535H, NGV708537H, NGV708538H, NGV708539H, NGV708540H, NGV708541H, NGV708543H, NGV708544H, NGV708545H, NGV708547H, NGV708548H, NGV708549H, NGV708551H, NGV708552H, NGV708553H, NGV708554H, NGV708555H, NGV712776H, NGV714303H, NGV714304H, NGV715312H, NGV715492H, NGV715700H, NGV715838H, NGV715841H, NGV716115H, NGV716179H, NGV717193H, NGV717957H, NGV718016H, NGV718118H, NGV718388H, NGV718390H, NGV718391H, NGV718565H, NGV718568H, NGV718687H, NGV718816H, NGV719051H, NGV720631H, NGV721145H, NGV721462H, NGV721463H, NGV722450H, NGV722746H, NGV722844H, NGV723001H, NGV723210H, NGV723244H, NGV723605H, NGV723888H, NGV724143H, NGV724147H, NGV724457H, NGV724458H, NGV724459H, NGV725104H, NGV725168H, NGV725741H, NGV726259H, NGV726260H, NGV726262H, NGV726263H, NGV727544H, NGV700525H, NGV709020H, NGV711722H, NGV713343H, NGV715280H, NGV719770H, NGV001114H, NGV004436H, NGV650877H, NGV704201H, NGV706650H, NGV000553H, NGV651701H, NGV700483H, NGV702650H, NGV702943H, NGV703943H, NGV704811H, NGV706394H, NGV707363H, NGV709076H, NGV709953H, NGV710936H, NGV711273H, NGV714324H, NGV714326H, NGV717236H, NGV717280H, NGV717837H, NGV718444H, NGV718972H, NGV718973H, NGV719046H, NGV719359H, NGV719562H, NGV719610H, NGV720020H, NGV720095H, NGV720337H, NGV721119H, NGV721471H, NGV723041H, NGV723359H, NGV723693H, NGV723826H, NGV724058H, NGV724066H, NGV724092H, NGV724235H, NGV724927H, NGV725161H, NGV726656H, NGV726679H, NGV000061H, NGV000913H, NGV004407H, NGV004408H, NGV004509H, NGV005001H, NGV650292H, NGV651144H, NGV651231H, NGV700372H, NGV700816H, NGV701435H, NGV701455H, NGV701965H, NGV703095H, NGV703570H, NGV703681H, NGV703944H, NGV704789H, NGV704973H, NGV705174H, NGV705188H, NGV705189H, NGV705905H, NGV707112H, NGV707476H, NGV709279H, NGV709403H, NGV709846H, NGV710039H, NGV710040H, NGV710218H, NGV710343H, NGV710387H, NGV711491H, NGV712759H, NGV712958H, NGV713032H, NGV713980H, NGV713982H, NGV714594H, NGV714595H, NGV715622H, NGV716229H, NGV717523H, NGV717872H, NGV717873H, NGV718139H, NGV718324H, NGV718824H, NGV718825H, NGV718833H, NGV720605H, NGV721480H, NGV721481H, NGV725293H, NGV725327H, NGV725420H, NGV726206H, NGV726712H, NGV727806H, NGV720050H, NGV650332H, NGV701649H, NGV701946H, NGV702915H, NGV705210H, NGV705211H, NGV711357H, NGV714257H, NGV719948H, NGV724712H, NGV727743H, NGV003577H, NGV717810H, NGV718872H, NGV721222H, NGV000874H, NGV000883H, NGV004415H, NGV651956H, NGV652216H, NGV701892H, NGV701904H, NGV705623H, NGV706383H, NGV708886H, NGV711696H, NGV718302H, NGV721468H, NGV726612H, NGV002484H, NGV000570H, NGV003540H, NGV003792H, NGV004060H, NGV650060H, NGV652279H, NGV700320H, NGV700618H, NGV700847H, NGV701481H, NGV701538H, NGV705920H, NGV705961H, NGV706376H, NGV706549H, NGV706741H, NGV706806H, NGV707293H, NGV709688H, NGV711025H, NGV712062H, NGV718130H, NGV718202H, NGV718977H, NGV719391H, NGV719395H, NGV722634H, NGV727876H, NGV001917H, NGV002064H, NGV005019H, NGV650382H, NGV650916H, NGV651319H, NGV700065H, NGV700353H, NGV704399H, NGV706717H, NGV706718H, NGV708855H, NGV710342H, NGV711087H, NGV711403H, NGV711955H, NGV712286H, NGV714571H, NGV722785H, NGV722788H, NGV723762H, NGV724796H, NGV726020H, NGV726577H, NGV726645H, NGV726915H, NGV708381H, NGV724770H, NGV650577H, NGV000103H, NGV000375H, NGV000494H, NGV000495H, NGV001685H, NGV001686H, NGV001762H, NGV002789H, NGV002792H, NGV002793H, NGV003670H, NGV004783H, NGV005174H, NGV650259H, NGV650261H, NGV650262H, NGV650614H, NGV650615H, NGV650618H, NGV650948H, NGV651472H, NGV651476H, NGV651545H, NGV651950H, NGV652054H, NGV700148H, NGV700809H, NGV701023H, NGV702142H, NGV702171H, NGV703157H, NGV703408H, NGV704425H, NGV705153H, NGV705154H, NGV705333H, NGV705406H, NGV705472H, NGV706764H, NGV706925H, NGV707047H, NGV708971H, NGV709007H, NGV709008H, NGV709095H, NGV709222H, NGV709258H, NGV710036H, NGV710961H, NGV712080H, NGV712221H, NGV712377H, NGV712378H, NGV712602H, NGV712603H, NGV712700H, NGV712959H, NGV712962H, NGV712964H, NGV715618H, NGV715754H, NGV715762H, NGV715780H, NGV715877H, NGV717096H, NGV717838H, NGV717915H, NGV717916H, NGV718049H, NGV718475H, NGV718481H, NGV719144H, NGV719145H, NGV719829H, NGV719939H, NGV719940H, NGV720406H, NGV720444H, NGV720730H, NGV720940H, NGV721826H, NGV722687H, NGV722817H, NGV723252H, NGV723270H, NGV723603H, NGV724868H, NGV727060H, NGV727731H, NGV004829H, NGV704383H, NGV002415H, NGV652109H, NGV705408H, NGV705409H, NGV705585H, NGV712016H, NGV712212H, NGV727113H, NGV002162H, NGV003867H, NGV707430H, NGV713053H, NGV714053H, NGV717162H, NGV717184H, NGV721583H, NGV723492H, NGV703880H, NGV707952H, NGV711298H, NGV709526H, NGV005032H, NGV652136H, NGV711982H, NGV721393H, NGV705085H, NGV703178H, NGV726408H, NGV000004H, NGV001784H, NGV004374H, NGV701656H, NGV702932H, NGV703551H, NGV703556H, NGV703599H, NGV705954H, NGV708940H, NGV708941H, NGV708956H, NGV709426H, NGV714398H, NGV714403H, NGV719704H, NGV719764H, NGV726302H, NGV726592H, NGV727358H, NGV004304H, NGV650565H, NGV701714H, NGV701828H, NGV703236H, NGV705884H, NGV706206H, NGV708179H, NGV709227H, NGV712609H, NGV714488H, NGV718232H, NGV720485H, NGV721131H, NGV721760H, NGV724489H, NGV725964H, NGV000468H, NGV002894H, NGV004054H, NGV650609H, NGV701727H, NGV705299H, NGV706710H, NGV721345H, NGV701886H, NGV716546H, NGV718861H, NGV719850H, NGV723807H, NGV001732H, NGV001497H, NGV004253H, NGV709296H, NGV718349H, NGV725759H, NGV727858H, NGV717289H, NGV720741H, NGV721610H, NGV721820H, NGV726011H, NGV726747H, NGV726749H, NGV709573H, NGV716943H, NGV001572H, NGV003794H, NGV001752H, NGV002085H, NGV650138H, NGV650759H, NGV651983H, NGV652032H, NGV701870H, NGV702322H, NGV704117H, NGV704743H, NGV711587H, NGV717279H, NGV721923H, NGV725985H, NGV726136H, NGV726654H, NGV650403H, NGV714415H, NGV704370H, NGV709352H, NGV709983H, NGV710085H, NGV712336H, NGV712831H, NGV714261H, NGV715513H, NGV717536H, NGV717866H, NGV718321H, NGV718542H, NGV718679H, NGV719005H, NGV719125H, NGV719255H, NGV721311H, NGV722384H, NGV722889H, NGV723823H, NGV725340H, NGV726865H, NGV000206H, NGV000540H, NGV709053H, NGV722769H, NGV725616H, NGV001054H, NGV702552H, NGV717290H, NGV717291H, NGV001197H, NGV004335H, NGV709502H, NGV712340H, NGV004776H, NGV005215H, NGV650085H, NGV651499H, NGV700716H, NGV702008H, NGV702300H, NGV702762H, NGV703915H, NGV704694H, NGV706475H, NGV707480H, NGV708081H, NGV713588H, NGV715282H, NGV715541H, NGV718008H, NGV718009H, NGV718966H, NGV719696H, NGV719698H, NGV720532H, NGV721400H, NGV722982H, NGV724266H, NGV725266H, NGV722090H, NGV722435H, NGV724809H, NGV718736H, NGV720482H, NGV001965H, NGV002537H, NGV650989H, NGV651232H, NGV700373H, NGV709551H, NGV722691H, NGV652132H, NGV706274H, NGV706281H, NGV709276H, NGV709882H, NGV709971H, NGV715721H, NGV718275H, NGV719223H, NGV721858H, NGV725737H, NGV725944H, NGV726443H, NGV004334H, NGV700368H, NGV002157H, NGV003927H, NGV721280H, NGV725993H, NGV706983H, NGV712624H, NGV713988H, NGV719497H, NGV722012H, NGV723951H, NGV725298H, NGV000077H, NGV004751H, NGV650457H, NGV702151H, NGV702428H, NGV704453H, NGV706872H, NGV715266H, NGV718207H, NGV718208H, NGV719724H, NGV002477H, NGV700722H, NGV711322H, NGV715247H, NGV717121H, NGV717164H, NGV721023H, NGV722827H, NGV722828H, NGV725116H, NGV725117H, NGV708110H, NGV651249H, NGV718976H, NGV719000H, NGV720395H, NGV722648H, NGV725501H, NGV726851H, NGV707595H, NGV002009H, NGV651164H, NGV651480H, NGV706343H, NGV706420H, NGV707846H, NGV002151H, NGV003120H, NGV003122H, NGV652042H, NGV704362H, NGV704497H, NGV705279H, NGV711463H, NGV712610H, NGV713963H, NGV713964H, NGV713965H, NGV713966H, NGV713967H, NGV718397H, NGV719589H, NGV721553H, NGV722006H, NGV723710H, NGV726985H, NGV727567H, NGV704395H, NGV001781H, NGV004445H, NGV651354H, NGV701728H, NGV702695H, NGV703552H, NGV704486H, NGV704979H, NGV709225H, NGV711742H, NGV714401H, NGV718611H, NGV719607H, NGV724999H, NGV726092H, NGV701620H, NGV704354H, NGV705876H, NGV707532H, NGV709314H, NGV701925H, NGV003193H, NGV722553H, NGV003981H, NGV650168H, NGV700712H, NGV714615H, NGV721418H, NGV723494H, NGV000267H, NGV701883H, NGV702610H, NGV703881H, NGV705309H, NGV711793H, NGV713034H, NGV713304H, NGV714077H, NGV715484H, NGV716863H, NGV702426H, NGV712341H, NGV712343H, NGV720981H, NGV652379H, NGV725623H, NGV727109H, NGV003756H, NGV703015H, NGV004712H, NGV700709H, NGV702568H, NGV704113H, NGV704114H, NGV706624H, NGV706626H, NGV706843H, NGV713480H, NGV723480H, NGV725428H, NGV000157H, NGV001048H, NGV001387H, NGV001840H, NGV005220H, NGV005224H, NGV701963H, NGV702987H, NGV704711H, NGV705090H, NGV711268H, NGV712740H, NGV712741H, NGV718791H, NGV002416H, NGV005178H, NGV650130H, NGV650636H, NGV705729H, NGV705846H, NGV707286H, NGV709709H, NGV712858H, NGV718862H, NGV723257H, NGV725625H, NGV003912H, NGV001957H, NGV002509H, NGV002910H, NGV002911H, NGV002950H, NGV003119H, NGV005199H, NGV005200H, NGV715866H, NGV715876H, NGV715881H, NGV717220H, NGV720685H, NGV725377H, NGV000078H, NGV710340H, NGV715938H, NGV715973H, NGV716625H, NGV717394H, NGV718470H, NGV718961H, NGV719503H, NGV720025H, NGV720026H, NGV720486H, NGV720957H, NGV721032H, NGV721334H, NGV721410H, NGV721764H, NGV722545H, NGV722698H, NGV723240H, NGV723481H, NGV723769H, NGV723796H, NGV723880H, NGV725395H, NGV725568H, NGV727615H, NGV727683H, NGV727796H, NGV002165H, NGV705359H, NGV705609H, NGV707988H, NGV717820H, NGV718339H, NGV724937H, NGV003904H, NGV651676H, NGV706399H, NGV720962H, NGV000475H, NGV001141H, NGV002080H, NGV003361H, NGV004700H, NGV005243H, NGV005244H, NGV651218H, NGV700315H, NGV700349H, NGV704999H, NGV705201H, NGV705701H, NGV706753H, NGV709845H, NGV711083H, NGV712186H, NGV712712H, NGV716835H, NGV716865H, NGV717285H, NGV717471H, NGV717919H, NGV719189H, NGV719281H, NGV720524H, NGV720824H, NGV721058H, NGV721931H, NGV721970H, NGV722639H, NGV723233H, NGV723851H, NGV725384H, NGV726478H, NGV726949H, NGV000096H, NGV000373H, NGV000871H, NGV004929H, NGV650424H, NGV651367H, NGV651372H, NGV700216H, NGV700252H, NGV700348H, NGV700350H, NGV704067H, NGV705362H, NGV706360H, NGV706728H, NGV707256H, NGV708654H, NGV708814H, NGV709800H, NGV710031H, NGV710070H, NGV710091H, NGV711677H, NGV713036H, NGV714334H, NGV715417H, NGV717470H, NGV718190H, NGV719190H, NGV720394H, NGV720522H, NGV722873H, NGV723691H, NGV723920H, NGV723935H, NGV724064H, NGV726481H, NGV727069H, NGV003127H, NGV711749H, NGV714179H, NGV714181H, NGV714183H, NGV715531H, NGV717206H, NGV717207H, NGV717505H, NGV717615H, NGV718756H, NGV719019H, NGV719066H, NGV719069H, NGV721156H, NGV722706H, NGV723198H, NGV723340H, NGV723557H, NGV724475H, NGV724510H, NGV726270H, NGV726271H, NGV726531H, NGV726552H, NGV726588H, NGV726644H, NGV726975H, NGV700280H, NGV700302H, NGV720755H, NGV725869H, NGV000714H, NGV000981H, NGV003914H, NGV004979H, NGV702282H, NGV703356H, NGV704006H, NGV705017H, NGV706690H, NGV717114H, NGV717116H, NGV719978H, NGV720587H, NGV723287H, NGV704092H, NGV707635H, NGV710287H, NGV710395H, NGV713525H, NGV713596H, NGV714200H, NGV714204H, NGV715514H, NGV715831H, NGV717676H, NGV720236H, NGV720246H, NGV720746H, NGV722077H, NGV726419H, NGV726420H, NGV726630H, NGV727584H, NGV002992H, NGV004431H, NGV704310H, NGV709895H, NGV714193H, NGV717068H, NGV720123H, NGV720593H, NGV002153H, NGV704744H, NGV715966H, NGV719186H, NGV725780H, NGV704934H, NGV704935H, NGV704962H, NGV704963H, NGV709514H, NGV715982H, NGV722588H, NGV001409H, NGV001414H, NGV002034H, NGV002073H, NGV002565H, NGV700469H, NGV700632H, NGV700732H, NGV701461H, NGV703170H, NGV703935H, NGV704459H, NGV706232H, NGV706354H, NGV707604H, NGV708791H, NGV712049H, NGV712050H, NGV712051H, NGV715554H, NGV718880H, NGV724103H, NGV726693H, NGV002058H, NGV726497H, NGV001214H, NGV001779H, NGV002389H, NGV002812H, NGV004114H, NGV652577H, NGV701975H, NGV704691H, NGV705356H, NGV705790H, NGV707333H, NGV707564H, NGV709327H, NGV710210H, NGV711732H, NGV713337H, NGV713534H, NGV713541H, NGV713639H, NGV713641H, NGV713910H, NGV716299H, NGV716303H, NGV716346H, NGV718375H, NGV718607H, NGV718688H, NGV720031H, NGV720701H, NGV722921H, NGV726521H, NGV726600H, NGV727635H, NGV000322H, NGV000333H, NGV003665H, NGV003695H, NGV003696H, NGV003989H, NGV004077H, NGV005012H, NGV650144H, NGV650336H, NGV650338H, NGV650787H, NGV700453H, NGV701856H, NGV701858H, NGV701859H, NGV701860H, NGV701861H, NGV701862H, NGV702880H, NGV703610H, NGV704205H, NGV704699H, NGV707324H, NGV708222H, NGV708223H, NGV708224H, NGV708225H, NGV709120H, NGV711492H, NGV712007H, NGV712070H, NGV712568H, NGV712589H, NGV713664H, NGV713665H, NGV714082H, NGV714083H, NGV714084H, NGV715470H, NGV715498H, NGV715499H, NGV715505H, NGV717159H, NGV718146H, NGV718597H, NGV718598H, NGV718599H, NGV719508H, NGV719967H, NGV720005H, NGV721310H, NGV722864H, NGV722865H, NGV722880H, NGV723655H, NGV723846H, NGV723848H, NGV725928H, NGV726124H, NGV726125H, NGV709268H, NGV001099H, NGV003177H, NGV003340H, NGV005235H, NGV650625H, NGV702512H, NGV706271H, NGV709033H, NGV709909H, NGV717253H, NGV717422H, NGV719405H, NGV720943H, NGV721107H, NGV722869H, NGV723865H, NGV724487H, NGV725289H, NGV726672H, NGV727219H, NGV727220H, NGV727641H, NGV001673H, NGV001949H, NGV650841H, NGV711555H, NGV717125H, NGV004312H, NGV700563H, NGV711045H, NGV711222H, NGV718843H, NGV722026H, NGV004062H, NGV652051H, NGV700207H, NGV700299H, NGV701171H, NGV701172H, NGV701458H, NGV703956H, NGV705244H, NGV705454H, NGV705822H, NGV707148H, NGV707220H, NGV709155H, NGV709808H, NGV711305H, NGV712065H, NGV712405H, NGV714018H, NGV714019H, NGV716716H, NGV717076H, NGV718298H, NGV719211H, NGV719904H, NGV720378H, NGV724053H, NGV727606H, NGV704887H, NGV705643H, NGV706214H, NGV709417H, NGV000906H, NGV002422H, NGV002432H, NGV002433H, NGV002434H, NGV003288H, NGV003290H, NGV003291H, NGV003293H, NGV003295H, NGV004010H, NGV004012H, NGV004013H, NGV004014H, NGV004826H, NGV004827H, NGV651714H, NGV651719H, NGV702184H, NGV702188H, NGV702190H, NGV702192H, NGV703164H, NGV703165H, NGV703166H, NGV703167H, NGV704940H, NGV704942H, NGV704943H, NGV705965H, NGV705966H, NGV705967H, NGV707005H, NGV707974H, NGV707975H, NGV707976H, NGV707977H, NGV707978H, NGV709746H, NGV709747H, NGV709748H, NGV709750H, NGV709765H, NGV711123H, NGV711124H, NGV711125H, NGV711126H, NGV711949H, NGV712013H, NGV712014H, NGV715910H, NGV716522H, NGV716524H, NGV716525H, NGV716695H, NGV717081H, NGV717299H, NGV717844H, NGV717854H, NGV717934H, NGV717935H, NGV718992H, NGV718993H, NGV718996H, NGV719540H, NGV720800H, NGV720801H, NGV720802H, NGV721329H, NGV721333H, NGV721352H, NGV721762H, NGV722413H, NGV722414H, NGV722415H, NGV722416H, NGV722418H, NGV722796H, NGV723744H, NGV723745H, NGV723746H, NGV723747H, NGV723748H, NGV725403H, NGV725406H, NGV725407H, NGV725408H, NGV725409H, NGV725410H, NGV725411H, NGV725412H, NGV725413H, NGV725414H, NGV727793H, NGV727799H, NGV001284H, NGV002203H, NGV715827H, NGV726473H, NGV726482H, NGV721469H, NGV722653H, NGV723896H, NGV723915H, NGV725442H, NGV000889H, NGV702648H, NGV703270H, NGV712970H, NGV721099H, NGV000247H, NGV000448H, NGV001100H, NGV001271H, NGV001592H, NGV001593H, NGV001594H, NGV001915H, NGV001916H, NGV002878H, NGV003260H, NGV004049H, NGV004204H, NGV004205H, NGV004209H, NGV004347H, NGV004442H, NGV005065H, NGV650392H, NGV650867H, NGV651412H, NGV651580H, NGV651607H, NGV651700H, NGV652331H, NGV700092H, NGV700265H, NGV700277H, NGV700473H, NGV700496H, NGV700497H, NGV700547H, NGV700714H, NGV701148H, NGV701253H, NGV701442H, NGV701823H, NGV702312H, NGV702386H, NGV702593H, NGV702699H, NGV703140H, NGV703183H, NGV703425H, NGV703807H, NGV703837H, NGV703879H, NGV704101H, NGV704102H, NGV704541H, NGV704747H, NGV704760H, NGV705074H, NGV705264H, NGV706089H, NGV706207H, NGV706285H, NGV706739H, NGV706979H, NGV707486H, NGV707598H, NGV707972H, NGV707973H, NGV709507H, NGV709541H, NGV709881H, NGV709902H, NGV710360H, NGV710902H, NGV711111H, NGV711115H, NGV711142H, NGV711428H, NGV711849H, NGV711871H, NGV712393H, NGV712399H, NGV712463H, NGV712465H, NGV712780H, NGV712822H, NGV712832H, NGV713024H, NGV713260H, NGV713342H, NGV715401H, NGV715402H, NGV715403H, NGV715406H, NGV715407H, NGV715408H, NGV715409H, NGV716262H, NGV716530H, NGV716586H, NGV717137H, NGV717155H, NGV717417H, NGV717418H, NGV717741H, NGV718476H, NGV718477H, NGV719146H, NGV719147H, NGV719530H, NGV721171H, NGV721172H, NGV721173H, NGV721632H, NGV721815H, NGV721816H, NGV722793H, NGV722806H, NGV723138H, NGV723140H, NGV723497H, NGV724132H, NGV724138H, NGV724140H, NGV724277H, NGV724460H, NGV724737H, NGV724749H, NGV725271H, NGV725299H, NGV725401H, NGV725764H, NGV725787H, NGV725802H, NGV725803H, NGV726269H, NGV726425H, NGV726520H, NGV726562H, NGV726675H, NGV726892H, NGV727072H, NGV727304H, NGV727736H, NGV000782H, NGV000783H, NGV704718H, NGV704736H, NGV704929H, NGV710368H, NGV000816H, NGV002991H, NGV003425H, NGV003426H, NGV004530H, NGV650634H, NGV650635H, NGV650637H, NGV651007H, NGV651009H, NGV651012H, NGV651013H, NGV651614H, NGV651628H, NGV651883H, NGV700565H, NGV702641H, NGV702666H, NGV703620H, NGV703815H, NGV704621H, NGV704888H, NGV705122H, NGV705262H, NGV706202H, NGV708990H, NGV709311H, NGV709778H, NGV711349H, NGV712562H, NGV712696H, NGV712697H, NGV713372H, NGV713502H, NGV714459H, NGV715413H, NGV717409H, NGV719713H, NGV720335H, NGV720896H, NGV721681H, NGV721800H, NGV721801H, NGV721824H, NGV722541H, NGV722609H, NGV722738H, NGV723042H, NGV726563H, NGV727296H, NGV727660H, NGV001822H, NGV705431H, NGV705432H, NGV718451H, NGV004497H, NGV704322H, NGV709471H, NGV710432H, NGV711162H, NGV711166H, NGV717638H, NGV718618H, NGV720739H, NGV722056H, NGV723687H, NGV725378H, NGV722061H, NGV003169H, NGV712333H, NGV719028H, NGV000531H, NGV002319H, NGV003640H, NGV003732H, NGV003835H, NGV003837H, NGV650523H, NGV700200H, NGV702390H, NGV703000H, NGV704001H, NGV704286H, NGV704336H, NGV704411H, NGV713305H, NGV717341H, NGV718703H, NGV721012H, NGV721146H, NGV721419H, NGV721420H, NGV721421H, NGV721422H, NGV721423H, NGV721424H, NGV722800H, NGV722801H, NGV725444H, NGV726021H, NGV726909H, NGV727663H, NGV003702H, NGV715712H, NGV715913H, NGV715919H, NGV718454H, NGV000572H, NGV001870H, NGV003166H, NGV004023H, NGV651205H, NGV652363H, NGV700492H, NGV700498H, NGV701452H, NGV703673H, NGV703846H, NGV703976H, NGV705328H, NGV705387H, NGV705817H, NGV707632H, NGV709301H, NGV712774H, NGV715671H, NGV721110H, NGV721314H, NGV723776H, NGV725768H, NGV726159H, NGV727122H, NGV000041H, NGV001571H, NGV002383H, NGV002646H, NGV003059H, NGV004040H, NGV706094H, NGV715277H, NGV718812H, NGV720901H, NGV721948H, NGV002811H, NGV004306H, NGV004695H, NGV701632H, NGV701993H, NGV702504H, NGV704558H, NGV704927H, NGV704928H, NGV708249H, NGV708250H, NGV711066H, NGV711721H, NGV714487H, NGV714489H, NGV715284H, NGV715545H, NGV717898H, NGV718245H, NGV719768H, NGV720644H, NGV721098H, NGV721330H, NGV722537H, NGV722809H, NGV722824H, NGV725721H, NGV702510H, NGV709447H, NGV725434H, NGV000555H, NGV000963H, NGV000972H, NGV001861H, NGV004256H, NGV651242H, NGV651392H, NGV651400H, NGV700313H, NGV700370H, NGV701244H, NGV701528H, NGV701529H, NGV702286H, NGV702290H, NGV702351H, NGV702539H, NGV702578H, NGV703414H, NGV703415H, NGV704081H, NGV704705H, NGV704707H, NGV705135H, NGV705191H, NGV705192H, NGV707201H, NGV707202H, NGV707203H, NGV707205H, NGV707206H, NGV707207H, NGV709652H, NGV709655H, NGV709656H, NGV709659H, NGV709660H, NGV710236H, NGV710243H, NGV711508H, NGV711576H, NGV711579H, NGV711864H, NGV712369H, NGV712514H, NGV712524H, NGV712860H, NGV715596H, NGV718396H, NGV721214H, NGV721215H, NGV721218H, NGV726844H, NGV000436H, NGV001089H, NGV004093H, NGV703059H, NGV704359H, NGV704506H, NGV704630H, NGV709857H, NGV709858H, NGV712077H, NGV717269H, NGV718897H, NGV000776H, NGV002457H, NGV004404H, NGV650695H, NGV651447H, NGV651841H, NGV651913H, NGV701712H, NGV702225H, NGV705278H, NGV705595H, NGV707109H, NGV711187H, NGV712641H, NGV712717H, NGV717928H, NGV717949H, NGV720601H, NGV721574H, NGV722456H, NGV725001H, NGV727110H, NGV000174H, NGV000503H, NGV001023H, NGV003382H, NGV003960H, NGV004097H, NGV004116H, NGV650310H, NGV650454H, NGV650807H, NGV651142H, NGV651385H, NGV700046H, NGV700633H, NGV701438H, NGV701622H, NGV701908H, NGV702058H, NGV702724H, NGV703247H, NGV703319H, NGV703354H, NGV704198H, NGV704258H, NGV704348H, NGV705187H, NGV705427H, NGV705603H, NGV705827H, NGV707428H, NGV707560H, NGV708811H, NGV708832H, NGV709288H, NGV709784H, NGV709785H, NGV710078H, NGV710926H, NGV710953H, NGV711686H, NGV711692H, NGV712022H, NGV712423H, NGV712489H, NGV713510H, NGV713715H, NGV715782H, NGV715783H, NGV716458H, NGV716556H, NGV716641H, NGV716804H, NGV717389H, NGV717493H, NGV718681H, NGV719951H, NGV719981H, NGV720511H, NGV720980H, NGV721231H, NGV721255H, NGV723361H, NGV723365H, NGV723573H, NGV723622H, NGV723804H, NGV726790H, NGV726973H, NGV727277H, NGV727278H, NGV727572H, NGV727786H, NGV005182H, NGV701298H, NGV706842H, NGV710426H, NGV712069H, NGV724690H, NGV001024H, NGV714656H, NGV717190H, NGV712629H, NGV716200H, NGV001573H, NGV001574H, NGV651675H, NGV652581H, NGV652582H, NGV652584H, NGV703016H, NGV705634H, NGV706216H, NGV707568H, NGV707849H, NGV708876H, NGV714387H, NGV714389H, NGV714394H, NGV716221H, NGV717865H, NGV726362H, NGV709667H, NGV650058H, NGV701427H, NGV651626H, NGV000017H, NGV000220H, NGV000471H, NGV000483H, NGV000502H, NGV000966H, NGV000983H, NGV000989H, NGV001801H, NGV002846H, NGV002848H, NGV004078H, NGV004079H, NGV004418H, NGV004419H, NGV004420H, NGV650920H, NGV650922H, NGV651816H, NGV651824H, NGV700346H, NGV701485H, NGV701488H, NGV701801H, NGV702155H, NGV702763H, NGV702764H, NGV702765H, NGV703196H, NGV703618H, NGV704260H, NGV704261H, NGV704262H, NGV704263H, NGV704264H, NGV704265H, NGV704446H, NGV705066H, NGV705213H, NGV705322H, NGV705468H, NGV705469H, NGV705507H, NGV705587H, NGV705645H, NGV705646H, NGV705647H, NGV706459H, NGV706460H, NGV706775H, NGV707000H, NGV707001H, NGV707600H, NGV707601H, NGV707602H, NGV708677H, NGV708993H, NGV708994H, NGV709531H, NGV709592H, NGV709593H, NGV709594H, NGV710050H, NGV710972H, NGV711084H, NGV711178H, NGV711239H, NGV711520H, NGV711521H, NGV711577H, NGV712650H, NGV712666H, NGV712718H, NGV712856H, NGV714005H, NGV714006H, NGV714010H, NGV714824H, NGV715390H, NGV715851H, NGV716211H, NGV717075H, NGV717249H, NGV718867H, NGV719139H, NGV719821H, NGV719837H, NGV719877H, NGV719878H, NGV720433H, NGV720434H, NGV720435H, NGV720846H, NGV720882H, NGV721183H, NGV721184H, NGV721286H, NGV721287H, NGV722064H, NGV722568H, NGV722589H, NGV722590H, NGV722829H, NGV722830H, NGV723293H, NGV723294H, NGV724706H, NGV724717H, NGV724858H, NGV725727H, NGV725728H, NGV725729H, NGV726384H, NGV726707H, NGV727684H, NGV727685H, NGV727734H, NGV720871H, NGV721041H, NGV722766H, NGV723394H, NGV725583H, NGV713590H, NGV714410H, NGV714411H, NGV715892H, NGV727709H, NGV709684H, NGV718689H, NGV723613H, NGV651094H, NGV704358H, NGV714327H, NGV717271H, NGV716985H, NGV004659H, NGV650328H, NGV650431H, NGV650607H, NGV651223H, NGV651619H, NGV703302H, NGV704211H, NGV704236H, NGV704554H, NGV705339H, NGV705340H, NGV705363H, NGV705775H, NGV706217H, NGV707865H, NGV710445H, NGV711095H, NGV711852H, NGV712485H, NGV712493H, NGV714145H, NGV716603H, NGV716941H, NGV717845H, NGV718702H, NGV720497H, NGV722651H, NGV722908H, NGV722909H, NGV723134H, NGV725365H, NGV727596H, NGV726274H, NGV714308H, NGV725207H, NGV727050H, NGV712506H, NGV005228H, NGV002986H, NGV005285H, NGV702748H, NGV703329H, NGV651923H, NGV700227H, NGV704722H, NGV704723H, NGV706268H, NGV709004H, NGV709039H, NGV711041H, NGV718261H, NGV718351H, NGV723269H, NGV000344H, NGV004803H, NGV004804H, NGV652039H, NGV700989H, NGV700990H, NGV702179H, NGV704017H, NGV708530H, NGV711432H, NGV714276H, NGV715643H, NGV715760H, NGV715765H, NGV716565H, NGV718116H, NGV718121H, NGV718124H, NGV719323H, NGV719717H, NGV001891H, NGV711319H, NGV713672H, NGV723595H, NGV002693H, NGV003378H, NGV003464H, NGV003471H, NGV003595H, NGV650660H, NGV700063H, NGV701046H, NGV702067H, NGV702483H, NGV707014H, NGV708655H, NGV708659H, NGV710291H, NGV711395H, NGV711698H, NGV712154H, NGV712748H, NGV713307H, NGV713507H, NGV718971H, NGV719681H, NGV720579H, NGV720830H, NGV721324H, NGV722388H, NGV722877H, NGV725638H, NGV727712H, NGV000642H, NGV000909H, NGV002430H, NGV002953H, NGV004332H, NGV004873H, NGV650429H, NGV650926H, NGV651858H, NGV700691H, NGV702396H, NGV702923H, NGV704088H, NGV706021H, NGV707209H, NGV708900H, NGV710079H, NGV712038H, NGV712455H, NGV712570H, NGV716362H, NGV716441H, NGV716442H, NGV718705H, NGV719428H, NGV720751H, NGV722563H, NGV722852H, NGV723231H, NGV723553H, NGV723604H, NGV723889H, NGV724453H, NGV725968H, NGV726232H, NGV726979H, NGV727550H, NGV727552H, NGV727693H, NGV000818H, NGV001130H, NGV001242H, NGV001857H, NGV001866H, NGV002566H, NGV002938H, NGV003096H, NGV003865H, NGV003971H, NGV004036H, NGV004180H, NGV004232H, NGV004765H, NGV004781H, NGV005043H, NGV650354H, NGV651527H, NGV651656H, NGV700294H, NGV700484H, NGV702992H, NGV703818H, NGV705261H, NGV705343H, NGV705531H, NGV705533H, NGV706198H, NGV706205H, NGV707340H, NGV708158H, NGV708780H, NGV709634H, NGV709673H, NGV710015H, NGV711441H, NGV711594H, NGV711603H, NGV711986H, NGV712679H, NGV713527H, NGV714381H, NGV714823H, NGV716879H, NGV717397H, NGV720893H, NGV720991H, NGV721692H, NGV721925H, NGV724998H, NGV725735H, NGV725813H, NGV726605H, NGV726963H, NGV726984H, NGV727058H, NGV727394H, NGV727590H, NGV701731H, NGV722977H, NGV702643H, NGV714338H, NGV650176H, NGV726350H, NGV711475H, NGV716203H, NGV004874H, NGV650812H, NGV652072H, NGV003967H, NGV001063H, NGV650602H, NGV650881H, NGV724347H, NGV704277H, NGV003686H, NGV001140H, NGV706804H, NGV709664H, NGV709735H, NGV711819H, NGV721817H, NGV722052H, NGV725481H, NGV001733H, NGV704168H, NGV711188H, NGV003705H, NGV700329H, NGV704385H, NGV705518H, NGV705904H, NGV707331H, NGV707596H, NGV708989H, NGV709979H, NGV709980H, NGV711009H, NGV711392H, NGV712026H, NGV714426H, NGV717210H, NGV718627H, NGV719187H, NGV719515H, NGV720913H, NGV722980H, NGV723901H, NGV727688H, NGV717202H, NGV725608H, NGV718283H, NGV719975H, NGV723574H, NGV000679H, NGV001824H, NGV003565H, NGV003930H, NGV650870H, NGV651493H, NGV701956H, NGV702394H, NGV702431H, NGV702796H, NGV702823H, NGV702989H, NGV703277H, NGV703368H, NGV703586H, NGV704368H, NGV704430H, NGV704468H, NGV705579H, NGV705795H, NGV705847H, NGV706176H, NGV708622H, NGV708983H, NGV709877H, NGV710294H, NGV710977H, NGV710979H, NGV711464H, NGV712541H, NGV712835H, NGV718048H, NGV723032H, NGV724301H, NGV651424H, NGV703535H, NGV708621H, NGV712404H, NGV712652H, NGV720750H, NGV000644H, NGV000662H, NGV001267H, NGV003842H, NGV004679H, NGV004817H, NGV700192H, NGV701568H, NGV703757H, NGV705330H, NGV705331H, NGV708871H, NGV716356H, NGV716683H, NGV717107H, NGV717589H, NGV726991H, NGV727039H, NGV002997H, NGV003430H, NGV005067H, NGV701726H, NGV702103H, NGV703894H, NGV704280H, NGV704303H, NGV705259H, NGV705323H, NGV706377H, NGV706462H, NGV706707H, NGV706984H, NGV707287H, NGV707309H, NGV707539H, NGV708987H, NGV709058H, NGV709295H, NGV709511H, NGV709512H, NGV709693H, NGV709851H, NGV709870H, NGV710029H, NGV710139H, NGV710142H, NGV710358H, NGV710372H, NGV710763H, NGV710892H, NGV711005H, NGV711282H, NGV711321H, NGV711493H, NGV711596H, NGV711883H, NGV712345H, NGV712346H, NGV712417H, NGV712594H, NGV712722H, NGV712805H, NGV712821H, NGV712872H, NGV713817H, NGV713826H, NGV713828H, NGV714404H, NGV714405H, NGV714412H, NGV714413H, NGV714414H, NGV714418H, NGV714419H, NGV715295H, NGV715904H, NGV715906H, NGV717297H, NGV717904H, NGV718280H, NGV719790H, NGV719793H, NGV719973H, NGV720292H, NGV720558H, NGV720827H, NGV721158H, NGV721166H, NGV721167H, NGV721300H, NGV721302H, NGV721922H, NGV722050H, NGV722241H, NGV722269H, NGV722770H, NGV722890H, NGV722935H, NGV723638H, NGV723654H, NGV723898H, NGV723913H, NGV724000H, NGV724897H, NGV725564H, NGV725621H, NGV726537H, NGV726538H, NGV726636H, NGV726687H, NGV726897H, NGV726898H, NGV726959H, NGV726960H, NGV727391H, NGV727548H, NGV727662H, NGV727737H, NGV705030H, NGV001516H, NGV004441H, NGV651921H, NGV701562H, NGV701576H, NGV704124H, NGV704209H, NGV002109H, NGV004759H, NGV004839H, NGV701203H, NGV701872H, NGV704072H, NGV704230H, NGV705653H, NGV705780H, NGV706114H, NGV707403H, NGV707405H, NGV707422H, NGV707664H, NGV707669H, NGV707678H, NGV707680H, NGV710915H, NGV710916H, NGV711134H, NGV712547H, NGV717167H, NGV717171H, NGV717173H, NGV717178H, NGV718215H, NGV719025H, NGV719027H, NGV721047H, NGV721049H, NGV722438H, NGV722858H, NGV722859H, NGV723665H, NGV723960H, NGV724791H, NGV725325H, NGV726157H, NGV727375H, NGV706344H, NGV001122H, NGV002946H, NGV003717H, NGV003719H, NGV003727H, NGV003760H, NGV004616H, NGV650596H, NGV650735H, NGV651378H, NGV700717H, NGV701830H, NGV702344H, NGV702921H, NGV703050H, NGV703054H, NGV703177H, NGV703181H, NGV704884H, NGV705167H, NGV705692H, NGV705814H, NGV706317H, NGV706639H, NGV706640H, NGV707214H, NGV707216H, NGV709038H, NGV709067H, NGV709069H, NGV709898H, NGV711387H, NGV711515H, NGV711531H, NGV711729H, NGV711936H, NGV717687H, NGV717690H, NGV718449H, NGV718460H, NGV718753H, NGV722019H, NGV722021H, NGV722022H, NGV723764H, NGV723835H, NGV723870H, NGV725796H, NGV726283H, NGV726699H, NGV651140H, NGV712613H, NGV717516H, NGV718506H, NGV721120H, NGV723447H, NGV723867H, NGV004414H, NGV004635H, NGV650737H, NGV700327H, NGV704441H, NGV705283H, NGV705972H, NGV706954H, NGV707337H, NGV708604H, NGV711105H, NGV712535H, NGV717521H, NGV718393H, NGV720904H, NGV002195H, NGV702651H, NGV703025H, NGV706616H, NGV707459H, NGV716856H, NGV717100H, NGV719677H, NGV721426H, NGV722742H, NGV725974H, NGV726199H, NGV726429H, NGV726466H, NGV714569H, NGV715768H, NGV721187H, NGV721370H, NGV721371H, NGV004319H, NGV004993H, NGV004994H, NGV651535H, NGV700422H, NGV700423H, NGV700424H, NGV700425H, NGV714325H, NGV715504H, NGV718459H, NGV719640H, NGV721461H, NGV723840H, NGV723841H, NGV000252H, NGV001349H, NGV703145H, NGV708874H, NGV711220H, NGV711472H, NGV717110H, NGV717111H, NGV723773H, NGV725767H, NGV002189H, NGV002955H, NGV003049H, NGV650479H, NGV650690H, NGV702164H, NGV703001H, NGV707620H, NGV712045H, NGV712326H, NGV712721H, NGV716207H, NGV716870H, NGV718539H, NGV719762H, NGV719799H, NGV725606H, NGV726899H, NGV001065H, NGV001270H, NGV002124H, NGV700974H, NGV701587H, NGV702675H, NGV703827H, NGV705320H, NGV708865H, NGV711225H, NGV001578H, NGV001958H, NGV002515H, NGV651124H, NGV651126H, NGV700100H, NGV701797H, NGV702214H, NGV703987H, NGV704331H, NGV706520H, NGV707043H, NGV711807H, NGV718041H, NGV719798H, NGV720038H, NGV720044H, NGV723952H, NGV724925H, NGV726383H, NGV726681H, NGV726682H, NGV726689H, NGV000907H, NGV001186H, NGV709089H, NGV720320H, NGV723338H, NGV003010H, NGV004806H, NGV703411H, NGV711878H, NGV716891H, NGV707979H, NGV715475H, NGV703606H, NGV725445H, NGV003310H, NGV709143H, NGV711654H, NGV716823H, NGV717120H, NGV724209H, NGV726137H, NGV727224H, NGV000094H, NGV000105H, NGV000469H, NGV001827H, NGV002564H, NGV002980H, NGV003377H, NGV650469H, NGV650470H, NGV651411H, NGV701580H, NGV702392H, NGV703687H, NGV703690H, NGV704289H, NGV704586H, NGV704587H, NGV706170H, NGV706171H, NGV709273H, NGV710098H, NGV710367H, NGV711173H, NGV711175H, NGV711589H, NGV717135H, NGV717170H, NGV717252H, NGV721627H, NGV721706H, NGV722028H, NGV722093H, NGV723004H, NGV724929H, NGV725105H, NGV725324H, NGV725326H, NGV717805H, NGV724220H, NGV724221H, NGV003855H, NGV650750H, NGV651949H, NGV701526H, NGV702213H, NGV704413H, NGV706226H, NGV706373H, NGV706424H, NGV709460H, NGV709495H, NGV702669H, NGV704142H, NGV708650H, NGV711200H, NGV714444H, NGV717325H, NGV717326H, NGV717946H, NGV724933H, NGV725091H, NGV726827H, NGV726829H, NGV000182H, NGV001070H, NGV003067H, NGV003752H, NGV003754H, NGV003953H, NGV004248H, NGV004249H, NGV651353H, NGV702219H, NGV702556H, NGV702616H, NGV702617H, NGV702685H, NGV702854H, NGV703814H, NGV703899H, NGV703928H, NGV704210H, NGV704326H, NGV704333H, NGV704393H, NGV704794H, NGV706029H, NGV706032H, NGV706037H, NGV706039H, NGV706197H, NGV706209H, NGV706455H, NGV706740H, NGV707861H, NGV708003H, NGV708161H, NGV708664H, NGV708665H, NGV709123H, NGV709133H, NGV709150H, NGV709223H, NGV709990H, NGV710073H, NGV710303H, NGV710304H, NGV710306H, NGV710333H, NGV711039H, NGV711047H, NGV711062H, NGV712056H, NGV712187H, NGV712188H, NGV712331H, NGV712498H, NGV712526H, NGV712575H, NGV712576H, NGV712577H, NGV712952H, NGV714186H, NGV714187H, NGV715434H, NGV715439H, NGV715659H, NGV715663H, NGV716615H, NGV716932H, NGV718027H, NGV718028H, NGV718029H, NGV718030H, NGV718639H, NGV718647H, NGV718701H, NGV719140H, NGV719333H, NGV719335H, NGV719438H, NGV720726H, NGV721272H, NGV721340H, NGV721987H, NGV722725H, NGV722726H, NGV722862H, NGV723730H, NGV724777H, NGV724786H, NGV724787H, NGV724902H, NGV726345H, NGV703677H, NGV707058H, NGV718025H, NGV721401H, NGV723510H, NGV726566H, NGV727585H, NGV001055H, NGV001120H, NGV005023H, NGV700813H, NGV700815H, NGV701497H, NGV702423H, NGV702496H, NGV702858H, NGV708079H, NGV708519H, NGV709607H, NGV709608H, NGV709668H, NGV709774H, NGV710939H, NGV713344H, NGV720083H, NGV723045H, NGV726911H, NGV714380H, NGV719138H, NGV721200H, NGV723234H, NGV709685H, NGV718858H, NGV001525H, NGV002975H, NGV651063H, NGV702340H, NGV702726H, NGV713287H, NGV716947H, NGV725288H, NGV003884H, NGV650159H, NGV651627H, NGV703151H, NGV709951H, NGV716002H, NGV725315H, NGV726368H, NGV719779H, NGV721589H, NGV724196H, NGV725095H, NGV001831H, NGV001919H, NGV004797H, NGV004798H, NGV650371H, NGV714568H, NGV722875H, NGV726893H, NGV000181H, NGV000652H, NGV001105H, NGV001138H, NGV001232H, NGV650804H, NGV700343H, NGV701236H, NGV702405H, NGV702438H, NGV703111H, NGV703273H, NGV704375H, NGV705863H, NGV708939H, NGV711183H, NGV711544H, NGV711580H, NGV713290H, NGV713293H, NGV715305H, NGV715849H, NGV715852H, NGV716706H, NGV718514H, NGV718851H, NGV718852H, NGV721390H, NGV721580H, NGV721838H, NGV721840H, NGV725436H, NGV726946H, NGV651712H, NGV726624H, NGV002590H, NGV652220H, NGV703848H, NGV703878H, NGV707315H, NGV710436H, NGV711078H, NGV716504H, NGV717258H, NGV717383H, NGV718384H, NGV719511H, NGV719702H, NGV720731H, NGV722425H, NGV723653H, NGV725628H, NGV725788H, NGV725789H, NGV725790H, NGV725791H, NGV726772H, NGV727199H, NGV727582H, NGV000472H, NGV003433H, NGV650731H, NGV710755H, NGV712968H, NGV000539H, NGV000544H, NGV712480H, NGV717196H, NGV722842H, NGV724302H, NGV000458H, NGV000739H, NGV001742H, NGV003113H, NGV651801H, NGV700730H, NGV700731H, NGV701595H, NGV701597H, NGV703623H, NGV705165H, NGV705416H, NGV706054H, NGV706265H, NGV706286H, NGV706536H, NGV708353H, NGV708362H, NGV711733H, NGV711843H, NGV712538H, NGV713243H, NGV715551H, NGV715553H, NGV715557H, NGV715566H, NGV715567H, NGV716544H, NGV717264H, NGV718684H, NGV718882H, NGV718952H, NGV720349H, NGV720598H, NGV720718H, NGV721035H, NGV721386H, NGV721938H, NGV721977H, NGV722918H, NGV723400H, NGV723470H, NGV723877H, NGV723906H, NGV723950H, NGV724086H, NGV724622H, NGV725301H, NGV725418H, NGV726642H, NGV001990H, NGV650181H, NGV720625H, NGV702050H, NGV714184H, NGV004773H, NGV650772H, NGV652128H, NGV652383H, NGV700396H, NGV700397H, NGV701841H, NGV709624H, NGV710928H, NGV713841H, NGV714203H, NGV714577H, NGV717610H, NGV717715H, NGV718866H, NGV723201H, NGV724983H, NGV725948H, NGV727855H, NGV727856H, NGV000260H, NGV001985H, NGV002232H, NGV002421H, NGV002479H, NGV003567H, NGV003968H, NGV004499H, NGV004844H, NGV004846H, NGV651350H, NGV651404H, NGV651830H, NGV651877H, NGV700161H, NGV700162H, NGV700168H, NGV700555H, NGV700571H, NGV701459H, NGV702569H, NGV703027H, NGV703110H, NGV703160H, NGV703941H, NGV704317H, NGV704605H, NGV704880H, NGV704981H, NGV705243H, NGV705906H, NGV706091H, NGV707385H, NGV707437H, NGV707458H, NGV707999H, NGV708324H, NGV708903H, NGV708904H, NGV709021H, NGV709241H, NGV709247H, NGV709260H, NGV709262H, NGV710937H, NGV710993H, NGV711624H, NGV712024H, NGV712036H, NGV712048H, NGV712349H, NGV712366H, NGV713508H, NGV714148H, NGV714149H, NGV714150H, NGV714151H, NGV714153H, NGV714154H, NGV714158H, NGV714159H, NGV714160H, NGV716600H, NGV716601H, NGV716714H, NGV717092H, NGV717093H, NGV717498H, NGV718323H, NGV718350H, NGV718999H, NGV719137H, NGV720263H, NGV720505H, NGV720577H, NGV721205H, NGV721206H, NGV721258H, NGV721654H, NGV722008H, NGV722063H, NGV723376H, NGV723758H, NGV724584H, NGV724585H, NGV724588H, NGV724617H, NGV725362H, NGV725379H, NGV726421H, NGV726438H, NGV726713H, NGV726714H, NGV727874H, NGV003594H, NGV704923H, NGV715828H, NGV715833H, NGV717041H, NGV721843H, NGV000309H, NGV000339H, NGV000565H, NGV001175H, NGV001179H, NGV001265H, NGV001498H, NGV001499H, NGV001505H, NGV001582H, NGV001589H, NGV002883H, NGV002941H, NGV002981H, NGV003588H, NGV003589H, NGV003590H, NGV003905H, NGV004178H, NGV004179H, NGV005157H, NGV005158H, NGV005201H, NGV005202H, NGV005252H, NGV650351H, NGV650679H, NGV650839H, NGV651772H, NGV652397H, NGV652398H, NGV700217H, NGV700234H, NGV700288H, NGV700289H, NGV700479H, NGV700480H, NGV700481H, NGV700482H, NGV701011H, NGV701278H, NGV702477H, NGV702478H, NGV702479H, NGV702480H, NGV702591H, NGV702592H, NGV703643H, NGV703644H, NGV703645H, NGV704428H, NGV704429H, NGV704855H, NGV704856H, NGV704857H, NGV705324H, NGV705810H, NGV705948H, NGV708594H, NGV711264H, NGV711467H, NGV712192H, NGV712359H, NGV712571H, NGV713298H, NGV714407H, NGV714417H, NGV715908H, NGV718666H, NGV719423H, NGV719789H, NGV720298H, NGV720304H, NGV721052H, NGV721165H, NGV724662H, NGV724707H, NGV726359H, NGV726539H, NGV726761H, NGV727681H, NGV000372H, NGV000675H, NGV001215H, NGV001217H, NGV002313H, NGV002325H, NGV002328H, NGV002375H, NGV003286H, NGV003451H, NGV004401H, NGV004762H, NGV005122H, NGV650835H, NGV651406H, NGV651502H, NGV651504H, NGV700222H, NGV701520H, NGV701911H, NGV702506H, NGV702727H, NGV704196H, NGV704197H, NGV704418H, NGV704737H, NGV705133H, NGV705458H, NGV706221H, NGV707433H, NGV708744H, NGV708859H, NGV709472H, NGV709612H, NGV709853H, NGV710135H, NGV711224H, NGV712755H, NGV713669H, NGV714271H, NGV714272H, NGV715393H, NGV718974H, NGV719657H, NGV720342H, NGV722031H, NGV722032H, NGV722888H, NGV723162H, NGV723696H, NGV725170H, NGV000412H, NGV000828H, NGV002099H, NGV002536H, NGV002711H, NGV003505H, NGV003647H, NGV004703H, NGV004801H, NGV005180H, NGV651095H, NGV651279H, NGV700316H, NGV701653H, NGV702278H, NGV702672H, NGV702904H, NGV704433H, NGV705145H, NGV705606H, NGV706287H, NGV706374H, NGV706892H, NGV707305H, NGV707461H, NGV707684H, NGV708582H, NGV709696H, NGV709850H, NGV710002H, NGV710196H, NGV710267H, NGV710449H, NGV711080H, NGV711265H, NGV711556H, NGV711570H, NGV712164H, NGV712307H, NGV712391H, NGV712830H, NGV716344H, NGV716969H, NGV716970H, NGV717452H, NGV719081H, NGV719082H, NGV721048H, NGV721244H, NGV721396H, NGV721397H, NGV721539H, NGV722035H, NGV722036H, NGV722870H, NGV722917H, NGV723012H, NGV723457H, NGV725593H, NGV725756H, NGV725761H, NGV725763H, NGV726881H, NGV726962H, NGV727005H, NGV727634H, NGV727705H, NGV002398H, NGV001310H, NGV001312H, NGV002581H, NGV002587H, NGV002588H, NGV002794H, NGV002795H, NGV004099H, NGV004100H, NGV651162H, NGV651892H, NGV651901H, NGV700527H, NGV700562H, NGV701815H, NGV702195H, NGV703197H, NGV703199H, NGV704588H, NGV706082H, NGV706112H, NGV706113H, NGV706407H, NGV706408H, NGV706409H, NGV706410H, NGV706436H, NGV706437H, NGV706813H, NGV707492H, NGV708633H, NGV708636H, NGV708637H, NGV708638H, NGV708639H, NGV709619H, NGV709620H, NGV711326H, NGV711329H, NGV711333H, NGV711887H, NGV712470H, NGV712472H, NGV712753H, NGV713244H, NGV714629H, NGV714630H, NGV714631H, NGV714633H, NGV715800H, NGV717046H, NGV717047H, NGV717693H, NGV717694H, NGV717697H, NGV717698H, NGV717699H, NGV717701H, NGV717747H, NGV717751H, NGV718564H, NGV718998H, NGV719396H, NGV719398H, NGV720172H, NGV721112H, NGV721223H, NGV721225H, NGV721877H, NGV721878H, NGV721879H, NGV722951H, NGV722952H, NGV722953H, NGV722954H, NGV722955H, NGV723531H, NGV724905H, NGV724908H, NGV724919H, NGV724920H, NGV725494H, NGV725976H, NGV727371H, NGV002190H, NGV004711H, NGV705808H, NGV000201H, NGV004882H, NGV005003H, NGV701234H, NGV702137H, NGV709036H, NGV710770H, NGV711601H, NGV711606H, NGV712525H, NGV712539H, NGV722768H, NGV705420H, NGV708312H, NGV710407H, NGV710409H, NGV712800H, NGV712866H, NGV713582H, NGV715637H, NGV717688H, NGV720725H, NGV722018H, NGV722506H, NGV723166H, NGV723544H, NGV723680H, NGV723682H, NGV726282H, NGV727121H, NGV003071H, NGV652249H, NGV701237H, NGV707065H, NGV712195H, NGV713491H, NGV716824H, NGV717239H, NGV717307H, NGV717355H, NGV001335H, NGV719127H, NGV708613H, NGV717166H, NGV002782H, NGV651518H, NGV652365H, NGV704314H, NGV717611H, NGV721276H, NGV721308H, NGV725492H, NGV725493H, NGV002274H, NGV003653H, NGV003951H, NGV004766H, NGV700700H, NGV701637H, NGV702447H, NGV703898H, NGV711609H, NGV717924H, NGV707945H, NGV709085H, NGV709090H, NGV709756H, NGV717702H, NGV714565H, NGV000415H, NGV000416H, NGV000422H, NGV000479H, NGV000496H, NGV000497H, NGV000560H, NGV000561H, NGV000567H, NGV002152H, NGV003303H, NGV003973H, NGV003974H, NGV004032H, NGV004365H, NGV004366H, NGV004367H, NGV650136H, NGV650161H, NGV650162H, NGV650377H, NGV650379H, NGV650580H, NGV650589H, NGV650590H, NGV651284H, NGV651287H, NGV651291H, NGV651844H, NGV651847H, NGV652115H, NGV652119H, NGV652120H, NGV652124H, NGV700182H, NGV700184H, NGV701387H, NGV701431H, NGV701776H, NGV701777H, NGV702220H, NGV702221H, NGV702310H, NGV703242H, NGV703251H, NGV703252H, NGV703253H, NGV703960H, NGV704061H, NGV704062H, NGV704407H, NGV704408H, NGV705088H, NGV705089H, NGV705249H, NGV718483H, NGV719282H, NGV719802H, NGV719803H, NGV719804H, NGV719805H, NGV719806H, NGV719807H, NGV719808H, NGV719809H, NGV719810H, NGV726390H, NGV726391H, NGV726392H, NGV726393H, NGV726394H, NGV726395H, NGV726396H, NGV726397H, NGV000451H, NGV000454H, NGV001853H, NGV002050H, NGV003573H, NGV650376H, NGV650427H, NGV650691H, NGV650706H, NGV651520H, NGV700360H, NGV701392H, NGV704318H, NGV705658H, NGV705662H, NGV710012H, NGV711559H, NGV711560H, NGV711638H, NGV711639H, NGV711640H, NGV712098H, NGV712584H, NGV713575H, NGV713705H, NGV713996H, NGV714195H, NGV714450H, NGV715568H, NGV715579H, NGV715583H, NGV716091H, NGV716854H, NGV717008H, NGV719015H, NGV719129H, NGV720480H, NGV720556H, NGV723200H, NGV724217H, NGV724851H, NGV726198H, NGV726658H, NGV726740H, NGV727004H, NGV651451H, NGV704938H, NGV700364H, NGV719279H, NGV000191H, NGV000317H, NGV001037H, NGV001731H, NGV002511H, NGV003922H, NGV004238H, NGV650187H, NGV650542H, NGV650639H, NGV650803H, NGV651256H, NGV651390H, NGV651452H, NGV651846H, NGV652416H, NGV700937H, NGV701304H, NGV701894H, NGV702382H, NGV702509H, NGV703122H, NGV705185H, NGV705635H, NGV705877H, NGV707012H, NGV707513H, NGV708981H, NGV711528H, NGV711726H, NGV715547H, NGV719619H, NGV720984H, NGV721678H, NGV726977H, NGV726998H, NGV000505H, NGV650021H, NGV650805H, NGV702851H, NGV705621H, NGV705622H, NGV705852H, NGV706161H, NGV709552H, NGV709635H, NGV710960H, NGV712862H, NGV723600H, NGV725096H, NGV725097H, NGV000168H, NGV000327H, NGV000334H, NGV001249H, NGV001927H, NGV002095H, NGV003940H, NGV003941H, NGV004102H, NGV004351H, NGV004529H, NGV004789H, NGV004831H, NGV004960H, NGV650744H, NGV651443H, NGV652630H, NGV702167H, NGV704066H, NGV705474H, NGV705871H, NGV705872H, NGV707282H, NGV707794H, NGV707956H, NGV708191H, NGV708825H, NGV709035H, NGV709066H, NGV709259H, NGV709771H, NGV709849H, NGV710137H, NGV710951H, NGV711217H, NGV712706H, NGV714572H, NGV714596H, NGV717099H, NGV717136H, NGV717161H, NGV718289H, NGV718488H, NGV718853H, NGV719666H, NGV719820H, NGV719990H, NGV720222H, NGV720226H, NGV720448H, NGV720449H, NGV721367H, NGV721368H, NGV721369H, NGV723097H, NGV723312H, NGV723469H, NGV725026H, NGV725027H, NGV725031H, NGV725877H, NGV726281H, NGV726297H, NGV726298H, NGV726620H, NGV726622H, NGV727192H, NGV000137H, NGV003195H, NGV003196H, NGV650996H, NGV651856H, NGV652293H, NGV700251H, NGV701521H, NGV701863H, NGV701864H, NGV701865H, NGV705718H, NGV707262H, NGV707453H, NGV707500H, NGV715767H, NGV720035H, NGV720886H, NGV721008H, NGV721068H, NGV721373H, NGV721961H, NGV722049H, NGV723539H, NGV723721H, NGV723813H, NGV724910H, NGV726163H, NGV001274H, NGV003888H, NGV003897H, NGV004862H, NGV651629H, NGV651631H, NGV652008H, NGV652011H, NGV652014H, NGV652029H, NGV652030H, NGV652036H, NGV702941H, NGV702942H, NGV703335H, NGV703338H, NGV704023H, NGV704026H, NGV704030H, NGV704034H, NGV704035H, NGV705651H, NGV707855H, NGV707856H, NGV707857H, NGV707858H, NGV707859H, NGV711337H, NGV711338H, NGV711797H, NGV711798H, NGV711800H, NGV711803H, NGV711804H, NGV716183H, NGV716188H, NGV716828H, NGV721913H, NGV722371H, NGV722375H, NGV722379H, NGV723749H, NGV723750H, NGV723752H, NGV723753H, NGV723754H, NGV723755H, NGV724158H, NGV724159H, NGV724181H, NGV724182H, NGV727536H, NGV700351H, NGV711678H, NGV711889H, NGV716634H, NGV718197H, NGV720003H, NGV723265H, NGV725165H, NGV725383H, NGV725486H, NGV725487H, NGV000028H, NGV000035H, NGV000988H, NGV000990H, NGV000997H, NGV001184H, NGV001517H, NGV002223H, NGV002682H, NGV002685H, NGV003956H, NGV004134H, NGV004730H, NGV004967H, NGV004996H, NGV005016H, NGV005035H, NGV650480H, NGV650897H, NGV650904H, NGV651054H, NGV651178H, NGV651362H, NGV651687H, NGV652634H, NGV652659H, NGV700107H, NGV700170H, NGV701717H, NGV701906H, NGV702002H, NGV702060H, NGV702090H, NGV702091H, NGV702421H, NGV702618H, NGV703566H, NGV703567H, NGV703918H, NGV704553H, NGV704797H, NGV704800H, NGV704891H, NGV705081H, NGV705125H, NGV705300H, NGV706870H, NGV706966H, NGV706969H, NGV707017H, NGV707018H, NGV707316H, NGV707983H, NGV708525H, NGV708565H, NGV708567H, NGV708805H, NGV708806H, NGV708827H, NGV709158H, NGV710105H, NGV710129H, NGV710130H, NGV710132H, NGV710209H, NGV711316H, NGV711318H, NGV711835H, NGV711876H, NGV711942H, NGV712714H, NGV712825H, NGV712828H, NGV713986H, NGV713999H, NGV714462H, NGV715517H, NGV715520H, NGV715810H, NGV715942H, NGV717549H, NGV717822H, NGV717861H, NGV717862H, NGV717894H, NGV717988H, NGV717989H, NGV718430H, NGV718644H, NGV718654H, NGV718657H, NGV719743H, NGV719788H, NGV719920H, NGV719961H, NGV719987H, NGV719988H, NGV720207H, NGV720347H, NGV720713H, NGV721128H, NGV721129H, NGV721195H, NGV721196H, NGV721630H, NGV721752H, NGV721789H, NGV721810H, NGV721811H, NGV721833H, NGV723027H, NGV723271H, NGV723272H, NGV723273H, NGV723723H, NGV724051H, NGV724714H, NGV724745H, NGV724746H, NGV725394H, NGV725730H, NGV727190H, NGV727612H, NGV727702H, NGV727853H, NGV000834H, NGV003042H, NGV650170H, NGV700279H, NGV701683H, NGV702420H, NGV705840H, NGV707981H, NGV709776H, NGV712542H, NGV713607H, NGV714362H, NGV716921H, NGV719624H, NGV720397H, NGV722473H, NGV722474H, NGV722475H, NGV722489H, NGV722586H, NGV000453H, NGV000455H, NGV000837H, NGV000838H, NGV000858H, NGV001188H, NGV001528H, NGV001577H, NGV001738H, NGV002051H, NGV002571H, NGV003107H, NGV003722H, NGV003758H, NGV004095H, NGV004146H, NGV004236H, NGV004237H, NGV004302H, NGV004303H, NGV004739H, NGV004771H, NGV004772H, NGV005150H, NGV005155H, NGV650765H, NGV650766H, NGV650778H, NGV650932H, NGV651318H, NGV651322H, NGV651521H, NGV651745H, NGV651805H, NGV652339H, NGV652618H, NGV700363H, NGV700458H, NGV700466H, NGV700502H, NGV701262H, NGV701281H, NGV701723H, NGV703046H, NGV703914H, NGV704253H, NGV704288H, NGV704315H, NGV704316H, NGV704351H, NGV704373H, NGV704490H, NGV704492H, NGV704493H, NGV704867H, NGV704868H, NGV705116H, NGV705696H, NGV705761H, NGV706767H, NGV706773H, NGV706899H, NGV706904H, NGV707102H, NGV707126H, NGV707319H, NGV707321H, NGV708557H, NGV708831H, NGV709205H, NGV709274H, NGV709291H, NGV709326H, NGV709498H, NGV709499H, NGV710102H, NGV710217H, NGV710396H, NGV710909H, NGV711641H, NGV711776H, NGV712553H, NGV712585H, NGV712752H, NGV714574H, NGV715614H, NGV715649H, NGV716851H, NGV716855H, NGV717577H, NGV717954H, NGV718129H, NGV718307H, NGV718584H, NGV718965H, NGV719017H, NGV719516H, NGV719766H, NGV719870H, NGV719956H, NGV720082H, NGV720478H, NGV720628H, NGV720894H, NGV721698H, NGV721808H, NGV721809H, NGV721900H, NGV722434H, NGV723157H, NGV723410H, NGV723413H, NGV723450H, NGV723917H, NGV724293H, NGV724503H, NGV725496H, NGV725497H, NGV725498H, NGV726542H, NGV726823H, NGV727543H, NGV727622H, NGV650723H, NGV651726H, NGV651727H, NGV651730H, NGV651731H, NGV651732H, NGV651734H, NGV651740H, NGV651749H, NGV652201H, NGV652206H, NGV652209H, NGV652211H, NGV652214H, NGV706850H, NGV706851H, NGV706852H, NGV706853H, NGV706855H, NGV706856H, NGV706857H, NGV706858H, NGV707082H, NGV707141H, NGV710820H, NGV713340H, NGV714288H, NGV715482H, NGV715645H, NGV716567H, NGV717571H, NGV717573H, NGV718831H, NGV720797H, NGV720798H, NGV721051H, NGV721057H, NGV722602H, NGV722608H, NGV723136H, NGV723184H, NGV723627H, NGV725290H, NGV726318H, NGV726319H, NGV726321H, NGV726322H, NGV726323H, NGV726325H, NGV726326H, NGV726327H, NGV726329H, NGV726331H, NGV726332H, NGV726333H, NGV726337H, NGV726340H, NGV726341H, NGV726343H, NGV000884H, NGV000947H, NGV000993H, NGV001038H, NGV002091H, NGV003767H, NGV004878H, NGV650791H, NGV651067H, NGV651494H, NGV701955H, NGV704011H, NGV704134H, NGV707139H, NGV707374H, NGV709550H, NGV709632H, NGV709869H, NGV710441H, NGV711685H, NGV712191H, NGV714188H, NGV716617H, NGV717392H, NGV717863H, NGV719022H, NGV719385H, NGV720985H, NGV723263H, NGV723339H, NGV727740H, NGV000158H, NGV001115H, NGV650213H, NGV651878H, NGV652252H, NGV703052H, NGV704096H, NGV704911H, NGV705430H, NGV706701H, NGV706754H, NGV706755H, NGV708807H, NGV709791H, NGV711718H, NGV711785H, NGV714449H, NGV715909H, NGV717419H, NGV719204H, NGV719838H, NGV720350H, NGV720719H, NGV721647H, NGV722572H, NGV723909H, NGV723911H, NGV725284H, NGV725295H, NGV726089H, NGV726090H, NGV727213H, NGV001021H, NGV001113H, NGV005192H, NGV650052H, NGV701661H, NGV704352H, NGV706110H, NGV706237H, NGV706875H, NGV711148H, NGV711249H, NGV714340H, NGV718316H, NGV719947H, NGV720635H, NGV720656H, NGV721092H, NGV722080H, NGV722754H, NGV723281H, NGV724454H, NGV726945H, NGV727563H, NGV727565H, NGV000019H, NGV000080H, NGV001723H, NGV001726H, NGV002440H, NGV003053H, NGV003484H, NGV003485H, NGV004002H, NGV004005H, NGV651305H, NGV652400H, NGV700260H, NGV701428H, NGV701430H, NGV702335H, NGV704716H, NGV706548H, NGV707588H, NGV708944H, NGV708947H, NGV708958H, NGV709559H, NGV709637H, NGV711462H, NGV711471H, NGV712029H, NGV712355H, NGV713699H, NGV713700H, NGV715037H, NGV715533H, NGV716666H, NGV717366H, NGV718018H, NGV718119H, NGV718493H, NGV719056H, NGV719538H, NGV719787H, NGV720053H, NGV720388H, NGV721001H, NGV721769H, NGV722069H, NGV722442H, NGV722966H, NGV723857H, NGV724065H, NGV725228H, NGV725317H, NGV725483H, NGV725521H, NGV725542H, NGV726070H, NGV726194H, NGV726272H, NGV726518H, NGV727807H, NGV001926H, NGV002410H, NGV650955H, NGV651724H, NGV703070H, NGV703071H, NGV703072H, NGV703074H, NGV703075H, NGV703076H, NGV703215H, NGV703216H, NGV703217H, NGV703218H, NGV703219H, NGV704372H, NGV709918H, NGV711205H, NGV711206H, NGV711207H, NGV711209H, NGV711343H, NGV711345H, NGV711450H, NGV711452H, NGV711453H, NGV711454H, NGV711455H, NGV711456H, NGV711458H, NGV711459H, NGV711461H, NGV711534H, NGV711535H, NGV711536H, NGV711537H, NGV711538H, NGV711539H, NGV711540H, NGV712255H, NGV714319H, NGV717215H, NGV717217H, NGV717995H, NGV718517H, NGV718592H, NGV719420H, NGV720500H, NGV720503H, NGV720733H, NGV721210H, NGV721415H, NGV721416H, NGV721417H, NGV721853H, NGV721856H, NGV722402H, NGV722403H, NGV722751H, NGV722752H, NGV722926H, NGV722927H, NGV722928H, NGV723142H, NGV723958H, NGV723959H, NGV724688H, NGV724695H, NGV724698H, NGV725946H, NGV726043H, NGV726060H, NGV726444H, NGV726775H, NGV727658H, NGV002823H, NGV004119H, NGV701634H, NGV702383H, NGV702384H, NGV703240H, NGV703594H, NGV705148H, NGV706366H, NGV708649H, NGV708653H, NGV711355H, NGV003073H, NGV003074H, NGV003075H, NGV003077H, NGV003078H, NGV003079H, NGV003081H, NGV003084H, NGV001909H, NGV651420H, NGV701375H, NGV704414H, NGV706233H, NGV706715H, NGV710269H, NGV710750H, NGV711503H, NGV712723H, NGV716228H, NGV717343H, NGV719873H, NGV721380H, NGV723894H, NGV725479H, NGV725480H, NGV706708H, NGV706359H, NGV711828H, NGV723532H, NGV002388H, NGV652529H, NGV001998H, NGV003109H, NGV003110H, NGV003111H, NGV003112H, NGV005063H, NGV651872H, NGV651874H, NGV651876H, NGV701008H, NGV701009H, NGV701015H, NGV701017H, NGV701155H, NGV701215H, NGV702106H, NGV702230H, NGV702492H, NGV702936H, NGV703049H, NGV704477H, NGV704478H, NGV704479H, NGV704480H, NGV704481H, NGV704482H, NGV704483H, NGV704484H, NGV708568H, NGV708569H, NGV708570H, NGV708571H, NGV711563H, NGV711564H, NGV711565H, NGV711566H, NGV713708H, NGV717058H, NGV719563H, NGV720535H, NGV720603H, NGV721585H, NGV722451H, NGV722554H, NGV722555H, NGV722558H, NGV723256H, NGV723259H, NGV723261H, NGV723852H, NGV724615H, NGV725397H, NGV726993H, NGV708586H, NGV705925H, NGV707496H, NGV704313H, NGV705821H, NGV710773H, NGV711574H, NGV711588H, NGV713266H, NGV717640H, NGV723900H, NGV725156H, NGV726241H, NGV726779H, NGV727653H, NGV002160H, NGV004149H, NGV005270H, NGV702387H, NGV703370H, NGV703372H, NGV704325H, NGV705824H, NGV707961H, NGV003148H, NGV000210H, NGV002527H, NGV650664H, NGV651339H, NGV702713H, NGV704320H, NGV709818H, NGV718471H, NGV726496H, NGV003677H, NGV703674H, NGV707226H, NGV709325H, NGV711912H, NGV717396H, NGV719708H, NGV719710H, NGV720439H, NGV722081H, NGV723778H, NGV650182H, NGV702507H, NGV705407H, NGV722975H, NGV000965H, NGV000974H, NGV000979H, NGV001323H, NGV001509H, NGV005047H, NGV005050H, NGV005051H, NGV005052H, NGV005053H, NGV005054H, NGV005056H, NGV005057H, NGV005059H, NGV650308H, NGV650819H, NGV650820H, NGV650821H, NGV650824H, NGV707843H, NGV707845H, NGV709636H, NGV709835H, NGV709836H, NGV709837H, NGV711915H, NGV711916H, NGV711917H, NGV711918H, NGV711919H, NGV711920H, NGV711957H, NGV717576H, NGV717634H, NGV719949H, NGV722405H, NGV722406H, NGV722407H, NGV722410H, NGV707488H, NGV724637H, NGV002362H, NGV004658H, NGV652079H, NGV652086H, NGV706689H, NGV707531H, NGV709698H, NGV710080H, NGV711156H, NGV711276H, NGV716604H, NGV716607H, NGV718365H, NGV720008H, NGV721986H, NGV722790H, NGV726409H, NGV727814H, NGV002246H, NGV003766H, NGV650349H, NGV700737H, NGV700750H, NGV702429H, NGV703584H, NGV704361H, NGV705956H, NGV720499H, NGV726453H, NGV001131H, NGV002102H, NGV002156H, NGV002158H, NGV003161H, NGV003162H, NGV004297H, NGV650019H, NGV652245H, NGV700172H, NGV702124H, NGV703190H, NGV704068H, NGV704596H, NGV704968H, NGV705200H, NGV705219H, NGV705405H, NGV705642H, NGV705860H, NGV706752H, NGV707559H, NGV708232H, NGV708306H, NGV708308H, NGV710227H, NGV710365H, NGV710771H, NGV711279H, NGV711315H, NGV711590H, NGV713044H, NGV713560H, NGV716094H, NGV716836H, NGV716948H, NGV716950H, NGV717468H, NGV717920H, NGV719654H, NGV722638H, NGV723937H, NGV725385H, NGV726902H, NGV727553H, NGV727560H, NGV003099H, NGV650278H, NGV711468H, NGV004105H, NGV650042H, NGV651031H, NGV703091H, NGV704698H, NGV707574H, NGV708867H, NGV711232H, NGV713997H, NGV722916H, NGV727677H, NGV001185H, NGV002983H, NGV003445H, NGV005026H, NGV650398H, NGV650444H, NGV650983H, NGV651006H, NGV700312H, NGV702001H, NGV702749H, NGV703999H, NGV705155H, NGV705669H, NGV706428H, NGV710112H, NGV711099H, NGV712431H, NGV712859H, NGV713532H, NGV713533H, NGV713616H, NGV714593H, NGV717152H, NGV717154H, NGV723454H, NGV725366H, NGV003880H, NGV004750H, NGV701707H, NGV702284H, NGV709710H, NGV717464H, NGV718037H, NGV720870H, NGV723547H, NGV723548H, NGV727221H, NGV727222H, NGV003775H, NGV651819H, NGV652645H, NGV702427H, NGV703058H, NGV703179H, NGV703381H, NGV703627H, NGV704631H, NGV705123H, NGV706048H, NGV707301H, NGV707311H, NGV708093H, NGV708925H, NGV709233H, NGV709234H, NGV710034H, NGV711269H, NGV711649H, NGV711973H, NGV712015H, NGV712074H, NGV713296H, NGV714185H, NGV714697H, NGV715694H, NGV717270H, NGV718510H, NGV718511H, NGV719601H, NGV720960H, NGV721597H, NGV721598H, NGV722885H, NGV722886H, NGV723197H, NGV724972H, NGV724976H, NGV724978H, NGV725421H, NGV725422H, NGV725423H, NGV726076H, NGV726295H, NGV002594H, NGV002940H, NGV002969H, NGV650894H, NGV651812H, NGV651813H, NGV701910H, NGV702511H, NGV704470H, NGV704703H, NGV704932H, NGV705839H, NGV706444H, NGV706445H, NGV706699H, NGV706700H, NGV711010H, NGV711011H, NGV711489H, NGV712630H, NGV712631H, NGV715674H, NGV717370H, NGV717630H, NGV718287H, NGV708155H, NGV717761H, NGV718837H, NGV720863H, NGV722883H, NGV725567H, NGV706781H, NGV711666H, NGV711831H, NGV712503H, NGV720536H, NGV725975H, NGV713516H, NGV708924H, NGV720319H, NGV703665H, NGV706774H, NGV000801H, NGV651290H, NGV701531H, NGV704432H, NGV708804H, NGV711195H, NGV713245H, NGV713671H, NGV717338H, NGV001925H, NGV650327H, NGV651239H, NGV651263H, NGV652357H, NGV703297H, NGV704324H, NGV711914H, NGV712591H, NGV712664H, NGV716627H, NGV717138H, NGV727214H, NGV727217H, NGV652181H, NGV713709H, NGV717524H, NGV718939H, NGV724917H, NGV724918H, NGV725367H, NGV725368H, NGV000203H, NGV001332H, NGV701084H, NGV701583H, NGV707115H, NGV707320H, NGV709561H, NGV711262H, NGV712283H, NGV712412H, NGV715491H, NGV719070H, NGV723459H, NGV711941H, NGV712771H, NGV004580H, NGV000081H, NGV000341H, NGV000723H, NGV000984H, NGV001518H, NGV002442H, NGV002702H, NGV004746H, NGV005021H, NGV651798H, NGV702194H, NGV702288H, NGV703418H, NGV706203H, NGV708893H, NGV709583H, NGV710319H, NGV712067H, NGV713998H, NGV716699H, NGV717332H, NGV719225H, NGV719226H, NGV720646H, NGV725114H, NGV001768H, NGV001769H, NGV001770H, NGV002154H, NGV002155H, NGV002690H, NGV004050H, NGV004051H, NGV652000H, NGV652001H, NGV652005H, NGV700195H, NGV700196H, NGV701152H, NGV701153H, NGV702916H, NGV702917H, NGV703968H, NGV703969H, NGV708166H, NGV708168H, NGV708169H, NGV709961H, NGV709962H, NGV714440H, NGV717308H, NGV718268H, NGV720805H, NGV722037H, NGV722040H, NGV722834H, NGV722837H, NGV722839H, NGV723508H, NGV723941H, NGV723943H, NGV724121H, NGV725695H, NGV725697H, NGV725698H, NGV725701H, NGV000199H, NGV000537H, NGV002737H, NGV005097H, NGV652034H, NGV702066H, NGV703120H, NGV703129H, NGV704285H, NGV706794H, NGV709051H, NGV709641H, NGV711568H, NGV717005H, NGV719857H, NGV721315H, NGV722386H, NGV722664H, NGV722734H, NGV722843H, NGV722967H, NGV725106H, NGV725858H, NGV726528H, NGV703664H, NGV704931H, NGV717791H, NGV717792H, NGV723353H, NGV001841H, NGV002485H, NGV002487H, NGV004670H, NGV702729H, NGV705865H, NGV708821H, NGV710895H, NGV721609H, NGV721611H, NGV721612H, NGV723887H, NGV726106H, NGV002865H, NGV002871H, NGV705615H, NGV705616H, NGV705617H, NGV708896H, NGV708897H, NGV721613H, NGV721614H, NGV721615H, NGV721616H, NGV722000H, NGV722001H, NGV722002H, NGV727386H, NGV727387H, NGV727388H, NGV000846H, NGV004333H, NGV650840H, NGV702745H, NGV721375H, NGV721387H, NGV721466H, NGV725978H, NGV726191H, NGV726862H, NGV001911H, NGV717377H, NGV718943H, NGV718944H, NGV719414H, NGV000880H, NGV001068H, NGV003676H, NGV003678H, NGV700235H, NGV700630H, NGV701877H, NGV702599H, NGV703808H, NGV705304H, NGV705490H, NGV705538H, NGV708882H, NGV716639H, NGV719627H, NGV719883H, NGV719890H, NGV722832H, NGV724479H, NGV725868H, NGV726857H, NGV001330H, NGV001543H, NGV001942H, NGV001951H, NGV003955H, NGV004127H, NGV004927H, NGV004985H, NGV700951H, NGV706309H, NGV707390H, NGV707399H, NGV707475H, NGV707523H, NGV707614H, NGV708929H, NGV709987H, NGV710103H, NGV716817H, NGV716819H, NGV717085H, NGV717143H, NGV717467H, NGV718387H, NGV721431H, NGV721813H, NGV721814H, NGV723864H, NGV725448H, NGV727805H, NGV000156H, NGV000737H, NGV001679H, NGV001680H, NGV002493H, NGV700955H, NGV701127H, NGV704881H, NGV704882H, NGV705451H, NGV705962H, NGV708523H, NGV708524H, NGV715889H, NGV719488H, NGV719489H, NGV719490H, NGV720921H, NGV723538H, NGV724750H, NGV726752H, NGV726929H, NGV727205H, NGV727605H, NGV651918H, NGV708820H, NGV004674H, NGV000364H, NGV001808H, NGV003852H, NGV652112H, NGV706439H, NGV706725H, NGV710930H, NGV714278H, NGV718123H, NGV718936H, NGV725718H, NGV000200H, NGV000467H, NGV001198H, NGV001952H, NGV711116H, NGV726483H, NGV000237H, NGV000274H, NGV001273H, NGV001280H, NGV001869H, NGV001876H, NGV001877H, NGV001881H, NGV001882H, NGV003343H, NGV003349H, NGV003350H, NGV003418H, NGV003419H, NGV003423H, NGV003559H, NGV003976H, NGV003992H, NGV003993H, NGV004174H, NGV004244H, NGV004283H, NGV004938H, NGV004944H, NGV005086H, NGV005089H, NGV005096H, NGV650023H, NGV650040H, NGV650045H, NGV651040H, NGV651402H, NGV652332H, NGV652335H, NGV702231H, NGV702289H, NGV702311H, NGV702334H, NGV702404H, NGV702580H, NGV702693H, NGV702694H, NGV702701H, NGV702702H, NGV702868H, NGV703258H, NGV703260H, NGV703809H, NGV703820H, NGV703821H, NGV703882H, NGV704123H, NGV704419H, NGV704502H, NGV705011H, NGV705012H, NGV705013H, NGV705143H, NGV706242H, NGV706261H, NGV706480H, NGV706960H, NGV707259H, NGV708301H, NGV708784H, NGV709032H, NGV709077H, NGV709082H, NGV709308H, NGV709425H, NGV709579H, NGV709581H, NGV709814H, NGV710362H, NGV711147H, NGV711263H, NGV711393H, NGV711394H, NGV712060H, NGV712325H, NGV712692H, NGV712801H, NGV713427H, NGV713467H, NGV713837H, NGV715301H, NGV715472H, NGV716512H, NGV716515H, NGV716640H, NGV717146H, NGV717156H, NGV717980H, NGV718060H, NGV718942H, NGV719471H, NGV719568H, NGV719582H, NGV719700H, NGV720087H, NGV720088H, NGV720353H, NGV720897H, NGV721043H, NGV721044H, NGV721091H, NGV721604H, NGV723282H, NGV724348H, NGV724349H, NGV724966H, NGV724967H, NGV725193H, NGV726309H, NGV726310H, NGV726316H, NGV726317H, NGV726432H, NGV726433H, NGV726434H, NGV726435H, NGV726512H, NGV726513H, NGV003688H, NGV703134H, NGV000332H, NGV651857H, NGV651908H, NGV651912H, NGV702898H, NGV703813H, NGV710280H, NGV715669H, NGV718478H, NGV725110H, NGV725476H, NGV726939H, NGV726978H, NGV727642H, NGV717359H, NGV724461H, NGV651683H, NGV700116H, NGV700245H, NGV704077H, NGV709124H, NGV714395H, NGV715483H, NGV723247H, NGV725743H, NGV725808H, NGV650408H, NGV716596H, NGV002137H, NGV003489H, NGV003748H, NGV650811H, NGV701838H, NGV702488H, NGV705245H, NGV707323H, NGV714078H, NGV715697H, NGV716883H, NGV719982H, NGV720650H, NGV721763H, NGV725576H, NGV726079H, NGV000922H, NGV004819H, NGV709094H, NGV712616H, NGV000176H, NGV000177H, NGV001258H, NGV001854H, NGV003970H, NGV004439H, NGV004794H, NGV652069H, NGV701440H, NGV701959H, NGV702412H, NGV702869H, NGV703107H, NGV704216H, NGV704294H, NGV704369H, NGV704381H, NGV704728H, NGV704759H, NGV705027H, NGV705097H, NGV705709H, NGV705953H, NGV707418H, NGV708086H, NGV708491H, NGV709899H, NGV711108H, NGV711197H, NGV712409H, NGV712500H, NGV712588H, NGV714274H, NGV714279H, NGV714287H, NGV716945H, NGV716972H, NGV716975H, NGV718103H, NGV718104H, NGV718109H, NGV721487H, NGV721491H, NGV721753H, NGV721837H, NGV721864H, NGV722771H, NGV722772H, NGV723189H, NGV723956H, NGV724565H, NGV725254H, NGV726351H, NGV726961H, NGV727525H, NGV727526H, NGV727689H, NGV727800H, NGV717864H, NGV003898H, NGV004877H, NGV650572H, NGV650573H, NGV651423H, NGV704048H, NGV704343H, NGV706750H, NGV706805H, NGV707170H, NGV707317H, NGV708964H, NGV709798H, NGV709901H, NGV710285H, NGV711769H, NGV711770H, NGV711773H, NGV712068H, NGV715035H, NGV715571H, NGV715586H, NGV715587H, NGV715589H, NGV715590H, NGV715598H, NGV716687H, NGV716688H, NGV716842H, NGV716844H, NGV716848H, NGV717244H, NGV717386H, NGV717940H, NGV718044H, NGV718045H, NGV718301H, NGV719493H, NGV719495H, NGV720052H, NGV720111H, NGV720115H, NGV721937H, NGV725186H, NGV000992H, NGV001973H, NGV002456H, NGV002458H, NGV002502H, NGV003857H, NGV004522H, NGV004523H, NGV650097H, NGV650697H, NGV650962H, NGV651434H, NGV651590H, NGV651781H, NGV702226H, NGV702227H, NGV704629H, NGV705594H, NGV705875H, NGV705900H, NGV707095H, NGV707107H, NGV707110H, NGV707111H, NGV709285H, NGV709686H, NGV710894H, NGV711185H, NGV712622H, NGV712699H, NGV713523H, NGV714347H, NGV714348H, NGV714356H, NGV715805H, NGV717349H, NGV717927H, NGV719035H, NGV719658H, NGV719661H, NGV720158H, NGV720602H, NGV720831H, NGV720836H, NGV721383H, NGV721384H, NGV721573H, NGV721575H, NGV723194H, NGV723645H, NGV723717H, NGV724939H, NGV724940H, NGV725656H, NGV726868H, NGV727111H, NGV727739H, NGV004113H, NGV700683H, NGV725838H, NGV004185H, NGV702395H, NGV705399H, NGV707417H, NGV707515H, NGV707644H, NGV712428H, NGV720492H, NGV720493H, NGV721582H, NGV722082H, NGV724985H, NGV727891H, NGV709797H, NGV703817H, NGV705539H, NGV714801H, NGV716206H, NGV716961H, NGV717832H, NGV720432H, NGV720585H, NGV721926H, NGV721927H, NGV725810H, NGV725814H, NGV726750H, NGV727051H, NGV727331H, NGV727332H, NGV727704H, NGV000440H, NGV651566H, NGV714329H, NGV719287H, NGV722640H, NGV726901H, NGV002924H, NGV002925H, NGV002926H, NGV002927H, NGV002929H, NGV002931H, NGV652550H, NGV652553H, NGV652554H, NGV652556H, NGV652557H, NGV652559H, NGV652561H, NGV652562H, NGV700461H, NGV708206H, NGV708207H, NGV708208H, NGV708209H, NGV708210H, NGV708211H, NGV708213H, NGV709557H, NGV002706H, NGV002707H, NGV002876H, NGV701919H, NGV707877H, NGV709305H, NGV710004H, NGV710429H, NGV711182H, NGV711233H, NGV711532H, NGV716223H, NGV716235H, NGV717487H, NGV717488H, NGV718474H, NGV719196H, NGV719745H, NGV721796H, NGV723843H, NGV724594H, NGV724598H, NGV726753H, NGV726754H, NGV727148H, NGV720930H, NGV723180H, NGV001710H, NGV715271H, NGV722996H, NGV727300H, NGV719443H, NGV004634H, NGV707021H, NGV723602H, NGV725655H, NGV004131H, NGV710254H, NGV712855H, NGV717421H, NGV718735H, NGV719188H, NGV723657H, NGV726762H, NGV726690H, NGV651466H, NGV700197H, NGV702605H, NGV704498H, NGV710059H, NGV711040H, NGV717039H, NGV719243H, NGV700135H, NGV723479H, NGV726882H, NGV712398H, NGV715932H, NGV727150H, NGV000808H, NGV721693H, NGV724996H, NGV000204H, NGV000849H, NGV001341H, NGV002405H, NGV002570H, NGV003561H, NGV003740H, NGV004020H, NGV004048H, NGV004189H, NGV004843H, NGV650110H, NGV650415H, NGV650770H, NGV650934H, NGV651078H, NGV651246H, NGV651325H, NGV651337H, NGV652340H, NGV701305H, NGV701391H, NGV701579H, NGV701724H, NGV702223H, NGV702224H, NGV702493H, NGV702561H, NGV703045H, NGV703619H, NGV704194H, NGV704287H, NGV704319H, NGV704339H, NGV704387H, NGV704422H, NGV704714H, NGV705652H, NGV707022H, NGV707114H, NGV707143H, NGV707907H, NGV707924H, NGV708016H, NGV708198H, NGV708247H, NGV708248H, NGV709206H, NGV709292H, NGV709563H, NGV710101H, NGV710140H, NGV710345H, NGV710421H, NGV710422H, NGV711140H, NGV711390H, NGV711597H, NGV711598H, NGV711792H, NGV712411H, NGV712544H, NGV712802H, NGV713380H, NGV713707H, NGV714162H, NGV714163H, NGV714202H, NGV715569H, NGV715578H, NGV715630H, NGV715631H, NGV715639H, NGV715648H, NGV716089H, NGV717010H, NGV717086H, NGV717272H, NGV717955H, NGV718052H, NGV718053H, NGV718305H, NGV718629H, NGV718673H, NGV718964H, NGV719071H, NGV719130H, NGV719163H, NGV719517H, NGV719869H, NGV719914H, NGV720508H, NGV720895H, NGV720983H, NGV721564H, NGV721699H, NGV721700H, NGV722433H, NGV722786H, NGV722789H, NGV722988H, NGV723155H, NGV723156H, NGV723444H, NGV723498H, NGV724150H, NGV724504H, NGV724911H, NGV724912H, NGV724913H, NGV724914H, NGV724915H, NGV726541H, NGV726627H, NGV726635H, NGV726791H, NGV726794H, NGV726799H, NGV727001H, NGV000456H, NGV650303H, NGV650458H, NGV652656H, NGV702487H, NGV707278H, NGV708528H, NGV000418H, NGV652037H, NGV706327H, NGV721282H, NGV722712H, NGV722957H, NGV002499H, NGV003950H, NGV701888H, NGV702003H, NGV702006H, NGV713691H, NGV700436H, NGV706910H, NGV708542H, NGV725431H, NGV725435H, NGV000894H, NGV703638H, NGV709691H, NGV711567H, NGV003134H, NGV005167H, NGV700791H, NGV702122H, NGV702237H, NGV706313H, NGV709098H, NGV713309H, NGV713830H, NGV714049H, NGV001983H, NGV004687H, NGV651962H, NGV709114H, NGV712634H, NGV723931H, NGV726546H, NGV722715H, NGV001674H, NGV704559H, NGV710756H, NGV715269H, NGV721026H, NGV725686H, NGV000034H, NGV000055H, NGV001489H, NGV001907H, NGV002999H, NGV003001H, NGV003002H, NGV650105H, NGV651230H, NGV700250H, NGV700945H, NGV701418H, NGV701421H, NGV703562H, NGV705604H, NGV706864H, NGV707295H, NGV708774H, NGV709933H, NGV710263H, NGV712707H, NGV714267H, NGV714270H, NGV716547H, NGV716811H, NGV717836H, NGV720086H, NGV721608H, NGV721982H, NGV723116H, NGV725525H, NGV725785H, NGV726413H, NGV727057H, NGV727597H, NGV727623H, NGV714321H, NGV715970H, NGV720716H, NGV001565H, NGV002952H, NGV004202H, NGV004279H, NGV650569H, NGV651169H, NGV651179H, NGV703352H, NGV703992H, NGV707190H, NGV707365H, NGV708244H, NGV708245H, NGV708499H, NGV709530H, NGV709832H, NGV709833H, NGV710769H, NGV711398H, NGV712527H, NGV712615H, NGV712786H, NGV712787H, NGV712788H, NGV712789H, NGV716821H, NGV716822H, NGV717234H, NGV718272H, NGV718273H, NGV718524H, NGV718588H, NGV718589H, NGV720417H, NGV720446H, NGV720475H, NGV720572H, NGV720573H, NGV720574H, NGV721711H, NGV723286H, NGV723719H, NGV724110H, NGV724752H, NGV726108H, NGV726110H, NGV726847H, NGV726848H, NGV726850H, NGV723578H, NGV726994H, NGV701764H, NGV714256H, NGV714262H, NGV725287H, NGV001087H, NGV003820H, NGV650361H, NGV650453H, NGV652587H, NGV705223H, NGV706385H, NGV706746H, NGV706788H, NGV707529H, NGV712560H, NGV713311H, NGV715693H, NGV716453H, NGV716935H, NGV722393H, NGV000976H, NGV701510H, NGV706195H, NGV001255H, NGV004440H, NGV702446H, NGV718793H, NGV002836H, NGV002900H, NGV002901H, NGV002912H, NGV003266H, NGV003267H, NGV003854H, NGV004034H, NGV004035H, NGV005112H, NGV005114H, NGV650624H, NGV651172H, NGV651181H, NGV651183H, NGV651198H, NGV651199H, NGV651200H, NGV700625H, NGV700626H, NGV701086H, NGV701087H, NGV703629H, NGV703631H, NGV704837H, NGV704838H, NGV704889H, NGV704890H, NGV706768H, NGV709309H, NGV710964H, NGV710965H, NGV711109H, NGV711112H, NGV711113H, NGV711161H, NGV711164H, NGV716537H, NGV718376H, NGV718635H, NGV719166H, NGV719174H, NGV719662H, NGV719663H, NGV720076H, NGV720077H, NGV720136H, NGV720137H, NGV721633H, NGV721875H, NGV721876H, NGV721991H, NGV721992H, NGV723761H, NGV724641H, NGV724642H, NGV724663H, NGV724664H, NGV725792H, NGV727201H, NGV727202H, NGV727203H, NGV720529H, NGV002710H, NGV004161H, NGV004738H, NGV650792H, NGV651118H, NGV702608H, NGV702754H, NGV703348H, NGV703585H, NGV705958H, NGV705963H, NGV718955H, NGV003709H, NGV004796H, NGV650626H, NGV650627H, NGV702038H, NGV702938H, NGV703607H, NGV718285H, NGV720440H, NGV000898H, NGV701978H, NGV703672H, NGV706246H, NGV726784H, NGV712507H, NGV719076H, NGV001825H, NGV001904H, NGV002361H, NGV004838H, NGV650856H, NGV651374H, NGV651711H, NGV652170H, NGV652173H, NGV652184H, NGV701031H, NGV702336H, NGV703057H, NGV703343H, NGV703402H, NGV703942H, NGV709044H, NGV709908H, NGV718855H, NGV719678H, NGV722701H, NGV722798H, NGV723179H, NGV725973H, NGV723477H, NGV004087H, NGV004824H, NGV650785H, NGV651255H, NGV706798H, NGV710413H, NGV711571H, NGV712058H, NGV714584H, NGV715847H, NGV718480H, NGV719715H, NGV719816H, NGV720411H, NGV725281H, NGV703205H, NGV721676H, NGV002916H, NGV000635H, NGV001194H, NGV002134H, NGV002148H, NGV002149H, NGV002150H, NGV003394H, NGV003429H, NGV004410H, NGV004823H, NGV005213H, NGV650189H, NGV651492H, NGV651496H, NGV652070H, NGV700283H, NGV701732H, NGV701733H, NGV702562H, NGV703106H, NGV703464H, NGV703985H, NGV704125H, NGV704452H, NGV704628H, NGV704869H, NGV704952H, NGV704953H, NGV704970H, NGV704971H, NGV706677H, NGV706712H, NGV706713H, NGV709212H, NGV709213H, NGV709215H, NGV709243H, NGV709520H, NGV709539H, NGV709574H, NGV709575H, NGV709841H, NGV709963H, NGV709964H, NGV710904H, NGV711834H, NGV712384H, NGV712385H, NGV713030H, NGV713578H, NGV713579H, NGV713580H, NGV714089H, NGV714091H, NGV714092H, NGV714093H, NGV714452H, NGV714453H, NGV714454H, NGV715646H, NGV715696H, NGV718550H, NGV718590H, NGV718591H, NGV718928H, NGV718929H, NGV719472H, NGV719476H, NGV719523H, NGV721185H, NGV721186H, NGV721260H, NGV722669H, NGV722728H, NGV723144H, NGV723146H, NGV723461H, NGV723628H, NGV723629H, NGV723631H, NGV723632H, NGV723633H, NGV725108H, NGV725122H, NGV725123H, NGV726080H, NGV726081H, NGV726082H, NGV727389H, NGV727390H, NGV727639H, NGV000087H, NGV002239H, NGV002322H, NGV652243H, NGV700155H, NGV700405H, NGV701149H, NGV705913H, NGV706810H, NGV706811H, NGV707445H, NGV707446H, NGV711404H, NGV712113H, NGV712115H, NGV715857H, NGV715858H, NGV715859H, NGV715860H, NGV719060H, NGV719313H, NGV720061H, NGV721113H, NGV721114H, NGV721116H, NGV726598H, NGV001076H, NGV001339H, NGV002678H, NGV005261H, NGV650107H, NGV703037H, NGV705117H, NGV709554H, NGV712791H, NGV714164H, NGV720956H, NGV001819H, NGV004405H, NGV700459H, NGV702414H, NGV703831H, NGV705152H, NGV705798H, NGV716104H, NGV718299H, NGV721378H, NGV721786H, NGV726763H, NGV727210H, NGV003566H, NGV004848H, NGV650121H, NGV651265H, NGV700467H, NGV701663H, NGV701928H, NGV702182H, NGV703574H, NGV704039H, NGV704583H, NGV705670H, NGV707235H, NGV709766H, NGV712332H, NGV713029H, NGV713031H, NGV713983H, NGV713985H, NGV715709H, NGV717929H, NGV717930H, NGV719547H, NGV719548H, NGV719648H, NGV722690H, NGV725336H, NGV727578H, NGV005263H, NGV700395H, NGV706988H, NGV707335H, NGV708257H, NGV708890H, NGV711836H, NGV713625H, NGV713918H, NGV716952H, NGV717235H, NGV718361H, NGV718669H, NGV718670H, NGV718671H, NGV718890H, NGV720671H, NGV720672H, NGV720674H, NGV720675H, NGV723569H, NGV723626H, NGV723819H, NGV726126H, NGV726127H, NGV726128H, NGV726129H, NGV726131H, NGV726132H, NGV726133H, NGV726134H, NGV726135H, NGV723801H, NGV003114H, NGV003116H, NGV700223H, NGV701699H, NGV703033H, NGV703265H, NGV703339H, NGV703972H, NGV705473H, NGV707938H, NGV715713H, NGV718606H, NGV725776H, NGV000225H, NGV001745H, NGV003490H, NGV650871H, NGV652241H, NGV703087H, NGV704593H, NGV705819H, NGV705973H, NGV707191H, NGV707984H, NGV707985H, NGV708641H, NGV709929H, NGV710929H, NGV711935H, NGV712483H, NGV712668H, NGV712715H, NGV712790H, NGV713047H, NGV714280H, NGV714608H, NGV715708H, NGV716566H, NGV718945H, NGV718946H, NGV718947H, NGV719315H, NGV719317H, NGV719462H, NGV719918H, NGV721016H, NGV721018H, NGV721475H, NGV721476H, NGV721779H, NGV721782H, NGV722709H, NGV723609H, NGV724273H, NGV725719H, NGV727864H, NGV726380H, NGV702275H, NGV705765H, NGV708813H, NGV711611H, NGV716991H, NGV721226H, NGV721759H, NGV726533H, NGV000663H, NGV001724H, NGV002530H, NGV003351H, NGV003356H, NGV003672H, NGV003673H, NGV003699H, NGV004160H, NGV004173H, NGV004190H, NGV650104H, NGV650193H, NGV650650H, NGV650651H, NGV700793H, NGV701779H, NGV701783H, NGV701785H, NGV701988H, NGV702136H, NGV703194H, NGV703264H, NGV703966H, NGV703967H, NGV705120H, NGV705270H, NGV705460H, NGV705527H, NGV706458H, NGV707307H, NGV707356H, NGV707591H, NGV708960H, NGV710335H, NGV711310H, NGV712353H, NGV714298H, NGV714299H, NGV714302H, NGV714305H, NGV716535H, NGV717941H, NGV717942H, NGV718120H, NGV719197H, NGV719200H, NGV719449H, NGV719477H, NGV719786H, NGV720859H, NGV721897H, NGV722561H, NGV723505H, NGV726200H, NGV726237H, NGV726657H, NGV726843H, NGV000045H, NGV002474H, NGV700817H, NGV705170H, NGV705805H, NGV707350H, NGV708907H, NGV712379H, NGV712720H, NGV715462H, NGV720089H, NGV722011H, NGV726416H, NGV727022H, NGV727586H, NGV004704H, NGV701102H, NGV715559H, NGV715561H, NGV715562H, NGV715795H, NGV716230H, NGV717233H, NGV717250H, NGV717596H, NGV718456H, NGV718538H, NGV719012H, NGV719154H, NGV719361H, NGV719371H, NGV719782H, NGV720098H, NGV720208H, NGV720531H, NGV720538H, NGV720539H, NGV720639H, NGV721359H, NGV721601H, NGV721603H, NGV722074H, NGV723552H, NGV723561H, NGV723683H, NGV723821H, NGV723822H, NGV725339H, NGV725460H, NGV725510H, NGV725512H, NGV720898H, NGV723291H, NGV001042H, NGV004981H, NGV005254H, NGV710324H, NGV715728H, NGV723651H, NGV726766H, NGV001856H, NGV002949H, NGV701150H, NGV702999H, NGV717801H, NGV717874H, NGV002208H, NGV004169H, NGV004220H, NGV004453H, NGV651236H, NGV651852H, NGV700108H, NGV700132H, NGV700477H, NGV705424H, NGV706554H, NGV707512H, NGV708666H, NGV711585H, NGV712563H, NGV714243H, NGV715537H, NGV718248H, NGV718264H, NGV719796H, NGV720127H, NGV725734H, NGV727583H, NGV000414H, NGV002043H, NGV004697H, NGV004932H, NGV004933H, NGV651666H, NGV702413H, NGV703153H, NGV703563H, NGV704093H, NGV705158H, NGV709313H, NGV709503H, NGV711159H, NGV711466H, NGV712209H, NGV712210H, NGV714021H, NGV714026H, NGV716572H, NGV716575H, NGV718860H, NGV718931H, NGV718934H, NGV719305H, NGV720619H, NGV721628H, NGV725553H, NGV725555H, NGV726555H, NGV727212H, NGV651651H, NGV723152H, NGV711088H, NGV716681H, NGV716846H, NGV003243H, NGV650355H, NGV650360H, NGV650445H, NGV706381H, NGV708246H, NGV717309H, NGV717311H, NGV719044H, NGV722401H, NGV723149H, NGV723310H, NGV725363H, NGV001895H, NGV703939H, NGV705582H, NGV714259H, NGV715962H, NGV724522H, NGV725183H, NGV725334H, NGV005024H, NGV701366H, NGV723961H, NGV726236H, NGV726866H, NGV002943H, NGV004215H, NGV651488H, NGV701569H, NGV701575H, NGV703662H, NGV705139H, NGV705861H, NGV705874H, NGV707094H, NGV709304H, NGV710404H, NGV712713H, NGV716924H, NGV717600H, NGV721089H, NGV727738H, NGV711234H, NGV712288H, NGV716511H, NGV719033H, NGV720567H, NGV720974H, NGV000224H, NGV000756H, NGV001744H, NGV002951H, NGV003493H, NGV003902H, NGV004876H, NGV652375H, NGV652542H, NGV652579H, NGV701719H, NGV702210H, NGV702991H, NGV703085H, NGV703211H, NGV703429H, NGV708640H, NGV710308H, NGV712647H, NGV712807H, NGV713048H, NGV713714H, NGV715759H, NGV716212H, NGV718522H, NGV720039H, NGV720583H, NGV720584H, NGV723617H, NGV725088H, NGV725797H, NGV726667H, NGV727152H, NGV000552H, NGV001193H, NGV001523H, NGV650234H, NGV650305H, NGV650406H, NGV650644H, NGV651195H, NGV705126H, NGV705332H, NGV705350H, NGV711213H, NGV711292H, NGV711496H, NGV711497H, NGV711498H, NGV711499H, NGV711500H, NGV711812H, NGV713526H, NGV714597H, NGV715414H, NGV716606H, NGV717973H, NGV717974H, NGV718364H, NGV718672H, NGV718889H, NGV720745H, NGV721868H, NGV721989H, NGV721990H, NGV722667H, NGV722668H, NGV723212H, NGV723803H, NGV724887H, NGV724888H, NGV726151H, NGV726152H, NGV726153H, NGV727672H, NGV004943H, NGV701075H, NGV701100H, NGV702867H, NGV702875H, NGV704501H, NGV704802H, NGV706964H, NGV708302H, NGV708783H, NGV709031H, NGV709224H, NGV710325H, NGV712059H, NGV712329H, NGV712806H, NGV715260H, NGV716852H, NGV717145H, NGV719868H, NGV719881H, NGV720910H, NGV720995H, NGV721952H, NGV722431H, NGV722432H, NGV723811H, NGV702676H, NGV001415H, NGV003639H, NGV650592H, NGV650751H, NGV651652H, NGV702424H, NGV706111H, NGV706390H, NGV706393H, NGV709088H, NGV710349H, NGV710973H, NGV711507H, NGV715836H, NGV716287H, NGV716288H, NGV716294H, NGV719326H, NGV719422H, NGV721228H, NGV722767H, NGV703168H, NGV703174H, NGV703959H, NGV704106H, NGV704980H, NGV706303H, NGV712270H, NGV722924H, NGV705525H, NGV724776H, NGV001974H, NGV652463H, NGV702759H, NGV719160H, NGV706729H, NGV709027H, NGV727302H, NGV003131H, NGV003949H, NGV701476H, NGV703184H, NGV703412H, NGV708917H, NGV709578H, NGV712020H, NGV713387H, NGV715632H, NGV716990H, NGV719008H, NGV719142H, NGV719994H, NGV720472H, NGV720548H, NGV721688H, NGV721942H, NGV725472H, NGV004675H, NGV700536H, NGV000744H, NGV000745H, NGV000748H, NGV002423H, NGV004292H, NGV651221H, NGV700240H, NGV700339H, NGV701711H, NGV702499H, NGV703612H, NGV704290H, NGV704291H, NGV706416H, NGV707180H, NGV709486H, NGV712275H, NGV712545H, NGV716801H, NGV717201H, NGV717384H, NGV717586H, NGV722656H, NGV723300H, NGV725815H, NGV726492H, NGV727686H, NGV727810H, NGV000428H, NGV650010H, NGV650026H, NGV652105H, NGV704966H, NGV707370H, NGV707371H, NGV710000H, NGV714285H, NGV717641H, NGV719667H, NGV727730H, NGV000768H, NGV003287H, NGV003342H, NGV651577H, NGV703306H, NGV706162H, NGV708930H, NGV711067H, NGV711702H, NGV712055H, NGV715609H, NGV719656H, NGV721747H, NGV005275H, NGV650428H, NGV704397H, NGV704398H, NGV711619H, NGV711872H, NGV713689H, NGV718628H, NGV721783H, NGV721932H, NGV727027H, NGV005172H, NGV723500H, NGV724633H, NGV727588H, NGV650430H, NGV652177H, NGV705302H, NGV705437H, NGV706227H, NGV706999H, NGV710326H, NGV711863H, NGV715944H, NGV715946H, NGV715950H, NGV717484H, NGV718007H, NGV719911H, NGV720963H, NGV720964H, NGV721088H, NGV721193H, NGV721194H, NGV721825H, NGV724963H, NGV725392H, NGV726174H, NGV727524H, NGV704308H, NGV704309H, NGV709500H, NGV001820H, NGV002600H, NGV002674H, NGV003124H, NGV003397H, NGV004140H, NGV004368H, NGV004685H, NGV005183H, NGV650957H, NGV650969H, NGV651529H, NGV702280H, NGV704958H, NGV704960H, NGV705190H, NGV706115H, NGV706333H, NGV706481H, NGV707145H, NGV707603H, NGV707605H, NGV709126H, NGV709689H, NGV711523H, NGV711527H, NGV712492H, NGV712674H, NGV714312H, NGV714313H, NGV714316H, NGV715555H, NGV715990H, NGV717502H, NGV717936H, NGV717939H, NGV718847H, NGV718883H, NGV718884H, NGV719042H, NGV719133H, NGV724665H, NGV724666H, NGV003140H, NGV003143H, NGV701987H, NGV711891H, NGV715274H, NGV716861H, NGV000052H, NGV000054H, NGV000516H, NGV000518H, NGV000523H, NGV000826H, NGV002885H, NGV002890H, NGV004533H, NGV004534H, NGV004535H, NGV651252H, NGV651905H, NGV651928H, NGV651929H, NGV651930H, NGV652589H, NGV700109H, NGV700112H, NGV700114H, NGV702740H, NGV702766H, NGV703005H, NGV703006H, NGV704933H, NGV705830H, NGV706466H, NGV706467H, NGV706468H, NGV707449H, NGV707643H, NGV708088H, NGV708480H, NGV709264H, NGV709586H, NGV709708H, NGV711252H, NGV711253H, NGV711299H, NGV711300H, NGV712185H, NGV712632H, NGV715169H, NGV715525H, NGV716529H, NGV716532H, NGV716534H, NGV716629H, NGV716630H, NGV716632H, NGV717030H, NGV717031H, NGV718764H, NGV719079H, NGV719080H, NGV719817H, NGV719818H, NGV720265H, NGV720595H, NGV720637H, NGV720936H, NGV720939H, NGV721305H, NGV722569H, NGV722570H, NGV722571H, NGV722587H, NGV723610H, NGV725306H, NGV725641H, NGV726085H, NGV726086H, NGV726088H, NGV727860H, NGV727866H, NGV003656H, NGV003347H, NGV003389H, NGV650941H, NGV651695H, NGV702083H, NGV702205H, NGV702341H, NGV702342H, NGV704049H, NGV705168H, NGV706861H, NGV707950H, NGV708511H, NGV709896H, NGV710410H, NGV712864H, NGV717658H, NGV717661H, NGV717827H, NGV718463H, NGV718749H, NGV722020H, NGV722505H, NGV722507H, NGV723833H, NGV725795H, NGV725870H, NGV726529H, NGV726701H, NGV001678H, NGV650600H, NGV651092H, NGV651427H, NGV651927H, NGV700241H, NGV700452H, NGV703182H, NGV705156H, NGV705690H, NGV706236H, NGV706319H, NGV706320H, NGV708508H, NGV712196H, NGV712799H, NGV718185H, NGV718925H, NGV719231H, NGV722017H, NGV722508H, NGV723681H, NGV723928H, NGV724279H, NGV726500H, NGV726501H, NGV727117H, NGV003617H, NGV004156H, NGV652525H, NGV701525H, NGV708585H, NGV717090H, NGV725328H, NGV001123H, NGV700203H, NGV719612H, NGV652082H, NGV702930H, NGV705686H, NGV725300H, NGV004197H, NGV703255H, NGV705415H, NGV706387H, NGV708351H, NGV717119H, NGV001507H, NGV003372H, NGV700991H, NGV703088H, NGV703908H, NGV705818H, NGV721778H, NGV716691H, NGV719650H, NGV000975H, NGV650880H, NGV650937H, NGV651017H, NGV706215H, NGV706334H, NGV708234H, NGV708235H, NGV708237H, NGV708238H, NGV708240H, NGV708241H, NGV708242H, NGV708243H, NGV708948H, NGV717300H, NGV717301H, NGV719525H, NGV719526H, NGV719527H, NGV720267H, NGV720269H, NGV720270H, NGV721576H, NGV721954H, NGV721955H, NGV723350H, NGV723351H, NGV723352H, NGV723593H, NGV723594H, NGV724107H, NGV724108H, NGV726000H, NGV726001H, NGV726736H, NGV726738H, NGV726739H, NGV727668H, NGV727669H, NGV727670H, NGV003583H, NGV705398H, NGV710219H, NGV711171H, NGV717281H, NGV713341H, NGV716920H, NGV719512H, NGV720437H, NGV720618H, NGV001740H, NGV001813H, NGV003877H, NGV003879H, NGV003911H, NGV652519H, NGV705753H, NGV705826H, NGV710431H, NGV717547H, NGV720122H, NGV721394H, NGV721596H, NGV723684H, NGV725664H, NGV001563H, NGV003424H, NGV004208H, NGV650733H, NGV652116H, NGV652117H, NGV700465H, NGV704409H, NGV707378H, NGV707730H, NGV707731H, NGV716354H, NGV717045H, NGV718187H, NGV720209H, NGV725722H, NGV726921H, NGV726958H, NGV715635H, NGV710033H, NGV001579H, NGV002965H, NGV002966H, NGV002968H, NGV004320H, NGV004735H, NGV004756H, NGV005130H, NGV005230H, NGV650330H, NGV650704H, NGV651304H, NGV652373H, NGV700435H, NGV702564H, NGV704472H, NGV705422H, NGV706829H, NGV706975H, NGV706976H, NGV707658H, NGV707864H, NGV708195H, NGV708196H, NGV708197H, NGV711893H, NGV711921H, NGV719928H, NGV721220H, NGV717499H, NGV002634H, NGV707101H, NGV712793H, NGV000389H, NGV001118H, NGV001121H, NGV005011H, NGV005015H, NGV650443H, NGV651190H, NGV652590H, NGV652623H, NGV700819H, NGV700820H, NGV701368H, NGV701377H, NGV701378H, NGV706981H, NGV709028H, NGV709029H, NGV711943H, NGV711944H, NGV714463H, NGV719437H, NGV720488H, NGV720695H, NGV721263H, NGV721264H, NGV721265H, NGV721984H, NGV726683H, NGV726684H, NGV726685H, NGV715485H, NGV717225H, NGV650907H, NGV701119H, NGV725990H, NGV003750H, NGV003751H, NGV004052H, NGV650114H, NGV650908H, NGV651334H, NGV651336H, NGV700099H, NGV700727H, NGV701108H, NGV701721H, NGV702626H, NGV703028H, NGV704589H, NGV707489H, NGV708770H, NGV708979H, NGV712804H, NGV713055H, NGV713352H, NGV715885H, NGV716581H, NGV717282H, NGV718856H, NGV720407H, NGV720410H, NGV721306H, NGV721929H, NGV724970H, NGV724971H, NGV726373H, NGV726832H, NGV003666H, NGV708966H, NGV726405H, NGV000547H, NGV001850H, NGV001851H, NGV003486H, NGV004224H, NGV004495H, NGV004498H, NGV005189H, NGV005247H, NGV650079H, NGV651001H, NGV651022H, NGV651026H, NGV651240H, NGV700185H, NGV700975H, NGV701245H, NGV701636H, NGV701645H, NGV701646H, NGV701647H, NGV702030H, NGV702031H, NGV702069H, NGV702563H, NGV702614H, NGV702615H, NGV703304H, NGV703430H, NGV703446H, NGV703532H, NGV704202H, NGV704321H, NGV704337H, NGV704967H, NGV706358H, NGV706384H, NGV706388H, NGV706769H, NGV706912H, NGV707547H, NGV707569H, NGV707576H, NGV707694H, NGV708490H, NGV708588H, NGV708619H, NGV708795H, NGV708797H, NGV708798H, NGV708868H, NGV708869H, NGV709042H, NGV709521H, NGV709538H, NGV709827H, NGV709970H, NGV710226H, NGV710377H, NGV710957H, NGV710958H, NGV710959H, NGV711231H, NGV711272H, NGV711592H, NGV711593H, NGV712034H, NGV712184H, NGV712234H, NGV712364H, NGV712513H, NGV712846H, NGV713322H, NGV713326H, NGV713414H, NGV714196H, NGV715999H, NGV716521H, NGV716549H, NGV717315H, NGV717579H, NGV717812H, NGV717856H, NGV718337H, NGV718338H, NGV718512H, NGV718915H, NGV719337H, NGV719340H, NGV719451H, NGV719550H, NGV719693H, NGV719694H, NGV719741H, NGV720124H, NGV720460H, NGV720461H, NGV720696H, NGV722695H, NGV722702H, NGV722704H, NGV722792H, NGV722807H, NGV722915H, NGV722973H, NGV722974H, NGV723031H, NGV723614H, NGV723639H, NGV723640H, NGV723914H, NGV723947H, NGV723949H, NGV724658H, NGV726378H, NGV726379H, NGV726403H, NGV726404H, NGV726619H, NGV726964H, NGV726968H, NGV726996H, NGV001751H, NGV703947H, NGV718630H, NGV004181H, NGV701217H, NGV704845H, NGV704846H, NGV708407H, NGV715725H, NGV720108H, NGV721680H, NGV722443H, NGV722846H, NGV723284H, NGV725798H, NGV727011H, NGV001736H, NGV002321H, NGV702138H, NGV702601H, NGV702927H, NGV705224H, NGV717950H, NGV721941H, NGV725485H, NGV725994H, NGV000499H, NGV000779H, NGV651046H, NGV708482H, NGV715961H, NGV717534H, NGV721312H, NGV721822H, NGV725157H, NGV650324H, NGV701789H, NGV704018H, NGV709927H, NGV710339H, NGV711022H, NGV700130H, NGV704256H, NGV707120H, NGV000525H, NGV000526H, NGV001061H, NGV001125H, NGV001237H, NGV002438H, NGV002575H, NGV004327H, NGV004328H, NGV004515H, NGV004808H, NGV005250H, NGV650685H, NGV650693H, NGV651609H, NGV651642H, NGV700450H, NGV700800H, NGV701199H, NGV701200H, NGV701201H, NGV701248H, NGV701249H, NGV701250H, NGV701251H, NGV701252H, NGV703131H, NGV703541H, NGV703547H, NGV703679H, NGV704306H, NGV704496H, NGV705130H, NGV705204H, NGV705205H, NGV705702H, NGV705873H, NGV706101H, NGV706164H, NGV706421H, NGV706422H, NGV707027H, NGV707034H, NGV707272H, NGV707866H, NGV707991H, NGV707993H, NGV707994H, NGV707997H, NGV707998H, NGV708872H, NGV710124H, NGV711226H, NGV711439H, NGV712193H, NGV712206H, NGV713666H, NGV713692H, NGV714476H, NGV715615H, NGV716290H, NGV716538H, NGV717473H, NGV717475H, NGV717508H, NGV718131H, NGV718133H, NGV718200H, NGV718223H, NGV718293H, NGV718593H, NGV718595H, NGV718708H, NGV719860H, NGV719905H, NGV720117H, NGV720441H, NGV720575H, NGV720589H, NGV720681H, NGV720785H, NGV720823H, NGV720835H, NGV720872H, NGV721004H, NGV721232H, NGV721402H, NGV721425H, NGV721532H, NGV721716H, NGV722389H, NGV722645H, NGV722647H, NGV722939H, NGV723258H, NGV723260H, NGV723307H, NGV723349H, NGV724183H, NGV724187H, NGV724498H, NGV724984H, NGV725166H, NGV725270H, NGV725323H, NGV726091H, NGV726291H, NGV726423H, NGV726595H, NGV726676H, NGV726710H, NGV726711H, NGV726955H, NGV726956H, NGV727209H, NGV727628H, NGV727732H, NGV727733H, NGV727811H, NGV727812H, NGV727877H, NGV001908H, NGV701652H, NGV702934H, NGV704489H, NGV707572H, NGV709300H, NGV710783H, NGV711479H, NGV712339H, NGV712678H, NGV713255H, NGV713270H, NGV717890H, NGV717891H, NGV719247H, NGV719609H, NGV720018H, NGV720717H, NGV725874H, NGV701822H, NGV702452H, NGV702768H, NGV703826H, NGV703903H, NGV706351H, NGV707540H, NGV707542H, NGV709323H, NGV709769H, NGV710138H, NGV711339H, NGV712453H, NGV715960H, NGV717387H, NGV718381H, NGV718382H, NGV720876H, NGV721036H, NGV721039H, NGV723073H, NGV723501H, NGV724521H, NGV724736H, NGV724891H, NGV724909H, NGV727328H, NGV715607H, NGV716355H, NGV726146H, NGV700711H, NGV701878H, NGV709063H, NGV709232H, NGV712636H, NGV719532H, NGV719872H, NGV722672H, NGV723563H, NGV727030H, NGV001797H, NGV004434H, NGV650990H, NGV703169H, NGV705959H, NGV711179H, NGV003474H, NGV650579H, NGV650741H, NGV650742H, NGV652341H, NGV700561H, NGV701903H, NGV704089H, NGV704090H, NGV704283H, NGV707168H, NGV707211H, NGV709142H, NGV711693H, NGV703020H, NGV707051H, NGV709043H, NGV709312H, NGV709315H, NGV715743H, NGV717399H, NGV722446H, NGV650981H, NGV652113H, NGV705282H, NGV705338H, NGV712578H, NGV712579H, NGV715664H, NGV001931H, NGV709627H, NGV709749H, NGV709989H, NGV721140H, NGV727534H, NGV000236H, NGV000346H, NGV000717H, NGV000910H, NGV002730H, NGV002732H, NGV002734H, NGV651015H, NGV700317H, NGV700319H, NGV700356H, NGV703810H, NGV704245H, NGV705008H, NGV712357H, NGV712358H, NGV712839H, NGV716834H, NGV717465H, NGV717857H, NGV719966H, NGV720120H, NGV721002H, NGV725342H, NGV725391H, NGV652267H, NGV710016H, NGV711026H, NGV723556H, NGV000831H, NGV716613H, NGV718779H, NGV720010H, NGV720496H, NGV723148H, NGV723932H, NGV725375H, NGV727742H, NGV650628H, NGV701782H, NGV001098H, NGV650027H, NGV702661H, NGV704128H, NGV715281H, NGV717266H, NGV710019H, NGV700418H, NGV702141H, NGV709318H, NGV721818H, NGV723473H, NGV721281H, NGV721283H, NGV703333H, NGV703834H, NGV704511H, NGV705305H, NGV706069H, NGV706686H, NGV712597H, NGV712598H, NGV723299H, NGV723854H, NGV724649H, NGV725692H, NGV004416H, NGV651079H, NGV705602H, NGV719927H, NGV725014H, NGV000187H, NGV000636H, NGV000056H, NGV650165H, NGV651635H, NGV700230H, NGV700232H, NGV702709H, NGV703133H, NGV705292H, NGV707606H, NGV711042H, NGV712416H, NGV714157H, NGV715310H, NGV716719H, NGV717346H, NGV717977H, NGV723545H, NGV723546H, NGV713275H, NGV721768H, NGV002202H, NGV002970H, NGV004123H, NGV652359H, NGV713368H, NGV713679H, NGV717197H, NGV717833H, NGV717834H, NGV722857H, NGV726204H, NGV726764H, NGV713046H, NGV712308H, NGV001956H, NGV700836H, NGV705457H, NGV706300H, NGV709270H, NGV709479H, NGV709590H, NGV710397H, NGV711813H, NGV712548H, NGV714197H, NGV715741H, NGV715742H, NGV715745H, NGV715746H, NGV716003H, NGV717466H, NGV717578H, NGV721860H, NGV721861H, NGV722368H, NGV723182H, NGV724645H, NGV724659H, NGV724784H, NGV726430H, NGV726793H, NGV726879H, NGV726880H, NGV726930H, NGV727741H, NGV001773H, NGV001964H, NGV002454H, NGV003465H, NGV705612H, NGV706885H, NGV725374H, NGV714289H, NGV714581H, NGV004406H, NGV700707H, NGV701912H, NGV709994H, NGV001591H, NGV003390H, NGV005284H, NGV706168H, NGV706647H, NGV709576H, NGV717671H, NGV720476H, NGV720521H, NGV723037H, NGV723953H, NGV000802H, NGV002366H, NGV002695H, NGV003924H, NGV004755H, NGV005236H, NGV650967H, NGV705578H, NGV711073H, NGV712450H, NGV713693H, NGV713847H, NGV715411H, NGV716271H, NGV716275H, NGV716704H, NGV717088H, NGV719446H, NGV720959H, NGV721083H, NGV725185H, NGV727687H, NGV705312H, NGV705581H, NGV715915H, NGV717158H, NGV717900H, NGV709920H, NGV652190H, NGV703014H, NGV004280H, NGV700455H, NGV702597H, NGV709483H, NGV709485H, NGV714057H, NGV001191H, NGV001200H, NGV001202H, NGV001755H, NGV002037H, NGV002039H, NGV002041H, NGV002042H, NGV002459H, NGV002998H, NGV003963H, NGV003964H, NGV003965H, NGV003966H, NGV650721H, NGV650725H, NGV650727H, NGV650728H, NGV652235H, NGV652237H, NGV652240H, NGV700438H, NGV700543H, NGV700705H, NGV700743H, NGV700744H, NGV700745H, NGV701898H, NGV701899H, NGV701900H, NGV703098H, NGV703099H, NGV703100H, NGV703119H, NGV704704H, NGV705274H, NGV705275H, NGV705276H, NGV705629H, NGV705630H, NGV705631H, NGV705632H, NGV707160H, NGV707161H, NGV707162H, NGV707193H, NGV707194H, NGV709482H, NGV709566H, NGV709568H, NGV709571H, NGV709941H, NGV709942H, NGV709944H, NGV710150H, NGV711098H, NGV711100H, NGV711101H, NGV711894H, NGV711895H, NGV711896H, NGV713572H, NGV716527H, NGV716528H, NGV716694H, NGV717052H, NGV717053H, NGV717054H, NGV718696H, NGV718697H, NGV718698H, NGV719036H, NGV719038H, NGV719039H, NGV720807H, NGV720809H, NGV720810H, NGV721289H, NGV721291H, NGV721965H, NGV721967H, NGV721968H, NGV722529H, NGV722531H, NGV722532H, NGV723164H, NGV723165H, NGV723326H, NGV723327H, NGV725667H, NGV725700H, NGV725740H, NGV725771H, NGV725828H, NGV727193H, NGV727194H, NGV727198H, NGV001051H, NGV004795H, NGV705951H, NGV722740H, NGV714294H, NGV723648H, NGV705273H, NGV001016H, NGV001019H, NGV001022H, NGV003175H, NGV003176H, NGV003248H, NGV003866H, NGV651632H, NGV651633H, NGV651646H, NGV702198H, NGV703651H, NGV705844H, NGV705845H, NGV706382H, NGV708972H, NGV709703H, NGV710239H, NGV712387H, NGV716204H, NGV717001H, NGV717644H, NGV717705H, NGV718616H, NGV719522H, NGV720045H, NGV720046H, NGV720415H, NGV721360H, NGV721940H, NGV723335H, NGV723642H, NGV724188H, NGV724623H, NGV724635H, NGV725232H, NGV725478H, NGV725936H, NGV000069H, NGV000285H, NGV000360H, NGV000457H, NGV001950H, NGV002130H, NGV003308H, NGV004145H, NGV004163H, NGV004235H, NGV004778H, NGV650643H, NGV650848H, NGV651356H, NGV651361H, NGV702055H, NGV702178H, NGV703156H, NGV704111H, NGV704926H, NGV705892H, NGV706863H, NGV707213H, NGV709199H, NGV709200H, NGV709201H, NGV709202H, NGV711229H, NGV712447H, NGV712475H, NGV713057H, NGV713599H, NGV715246H, NGV715250H, NGV715256H, NGV715293H, NGV716906H, NGV717390H, NGV717662H, NGV718986H, NGV719485H, NGV719753H, NGV720604H, NGV721178H, NGV721179H, NGV721180H, NGV721181H, NGV722632H, NGV724070H, NGV724447H, NGV724488H, NGV724490H, NGV724606H, NGV725935H, NGV725961H, NGV725962H, NGV725963H, NGV726210H, NGV727618H, NGV727630H, NGV716738H, NGV717884H, NGV726040H, NGV002087H, NGV001291H, NGV703032H, NGV710943H, NGV712614H, NGV701643H, NGV706687H, NGV706995H, NGV709642H, NGV714003H, NGV723853H, NGV706411H, NGV000450H, NGV652081H, NGV000376H, NGV000378H, NGV000719H, NGV002712H, NGV002936H, NGV650460H, NGV650461H, NGV650465H, NGV651439H, NGV651442H, NGV651444H, NGV709753H, NGV712813H, NGV712814H, NGV712815H, NGV718988H, NGV718991H, NGV721988H, NGV722775H, NGV724627H, NGV714011H, NGV716213H, NGV725601H, NGV650718H, NGV717378H, NGV705237H, NGV711427H, NGV712323H, NGV712337H, NGV713695H, NGV716795H, NGV002689H, NGV705598H, NGV705854H, NGV705964H, NGV707274H, NGV711203H, NGV715704H, NGV718490H, NGV718731H, NGV719063H, NGV720600H, NGV721028H, NGV721029H, NGV721934H, NGV722010H, NGV722499H, NGV719664H, NGV721964H, NGV705509H, NGV705688H, NGV701059H, NGV706949H, NGV003296H, NGV004906H, NGV005217H, NGV651150H, NGV651896H, NGV701734H, NGV703991H, NGV704215H, NGV704495H, NGV707131H, NGV709210H, NGV709239H, NGV711397H, NGV720972H, NGV721962H, NGV725629H, NGV725635H, NGV716871H, NGV716956H, NGV716963H, NGV716964H, NGV719916H, NGV724797H, NGV002987H, NGV003952H, NGV004261H, NGV702185H, NGV703148H, NGV707076H, NGV710434H, NGV000244H, NGV708618H, NGV710081H, NGV711081H, NGV715972H, NGV717385H, NGV724681H, NGV725983H, NGV651641H, NGV725341H, NGV004810H, NGV651756H, NGV707328H, NGV715943H, NGV720829H, NGV726387H, NGV700615H, NGV705172H, NGV706617H, NGV717474H, NGV718916H, NGV722940H, NGV726367H, NGV701168H, NGV000051H, NGV000088H, NGV000131H, NGV003851H, NGV650905H, NGV650913H, NGV651853H, NGV652353H, NGV703191H, NGV704199H, NGV704298H, NGV705271H, NGV706222H, NGV706241H, NGV709779H, NGV710355H, NGV714251H, NGV715497H, NGV717925H, NGV717926H, NGV718733H, NGV719979H, NGV721775H, NGV723195H, NGV723495H, NGV723592H, NGV724540H, NGV724932H, NGV725832H, NGV727598H, NGV000229H, NGV652087H, NGV005125H, NGV720624H, NGV002858H, NGV002860H, NGV651479H, NGV652259H, NGV652261H, NGV701514H, NGV701515H, NGV704853H, NGV704854H, NGV716516H, NGV716517H, NGV719665H, NGV720799H, NGV721168H, NGV724726H, NGV002364H, NGV703929H, NGV706199H, NGV706201H, NGV714601H, NGV714603H, NGV716594H, NGV716598H, NGV705902H, NGV718727H, NGV000165H, NGV000821H, NGV001269H, NGV004514H, NGV651222H, NGV702120H, NGV707958H, NGV711869H, NGV711870H, NGV713509H, NGV715933H, NGV715935H, NGV716832H, NGV717222H, NGV720916H, NGV721620H, NGV722092H, NGV003920H, NGV003957H, NGV705079H, NGV717429H, NGV717446H, NGV717501H, NGV717851H, NGV717852H, NGV726230H, NGV002623H, NGV719643H, NGV719706H, NGV004856H, NGV725675H, NGV726181H, NGV727145H, NGV650991H, NGV705353H, NGV726358H, NGV000511H, NGV000789H, NGV000792H, NGV000886H, NGV004166H, NGV650826H, NGV650834H, NGV700210H, NGV701985H, NGV703300H, NGV704095H, NGV705212H, NGV705813H, NGV707294H, NGV710193H, NGV710311H, NGV710313H, NGV711697H, NGV712749H, NGV718177H, NGV718832H, NGV720007H, NGV720606H, NGV725316H, NGV725586H, NGV727031H, NGV707329H, NGV707349H, NGV714033H, NGV714062H, NGV719328H, NGV722670H, NGV724030H, NGV724109H, NGV725179H, NGV727649H, NGV001873H, NGV003882H, NGV700354H, NGV700487H, NGV717276H, NGV717519H, NGV718536H, NGV650816H, NGV001584H, NGV001939H, NGV004117H, NGV004255H, NGV004510H, NGV004511H, NGV650601H, NGV651699H, NGV700695H, NGV700696H, NGV701091H, NGV701092H, NGV701793H, NGV701974H, NGV702071H, NGV702072H, NGV702073H, NGV702473H, NGV702914H, NGV704371H, NGV704394H, NGV705357H, NGV705611H, NGV706282H, NGV706415H, NGV706879H, NGV706883H, NGV709860H, NGV711070H, NGV711071H, NGV711846H, NGV715687H, NGV717588H, NGV717618H, NGV718309H, NGV719513H, NGV719727H, NGV721007H, NGV721248H, NGV722764H, NGV722884H, NGV722997H, NGV723554H, NGV723860H, NGV723861H, NGV725687H, NGV725689H, NGV726564H, NGV727319H, NGV727322H, NGV727789H, NGV651907H, NGV652075H, NGV003048H, NGV003381H, NGV709646H, NGV710346H, NGV719855H, NGV726229H, NGV000025H, NGV000280H, NGV000426H, NGV001761H, NGV001832H, NGV001943H, NGV002067H, NGV002118H, NGV003285H, NGV003795H, NGV004934H, NGV651674H, NGV652452H, NGV701363H, NGV701662H, NGV702156H, NGV703065H, NGV703067H, NGV703632H, NGV703965H, NGV704165H, NGV704293H, NGV705401H, NGV706108H, NGV706297H, NGV706302H, NGV706308H, NGV706449H, NGV707285H, NGV707288H, NGV707510H, NGV708085H, NGV709543H, NGV709893H, NGV709960H, NGV711518H, NGV711519H, NGV711610H, NGV712010H, NGV712321H, NGV712395H, NGV712396H, NGV713573H, NGV713684H, NGV714269H, NGV714273H, NGV716692H, NGV716803H, NGV716913H, NGV717148H, NGV717198H, NGV717199H, NGV717679H, NGV717806H, NGV717807H, NGV717808H, NGV717835H, NGV717981H, NGV718543H, NGV719006H, NGV719064H, NGV719143H, NGV719518H, NGV719519H, NGV719652H, NGV719997H, NGV720085H, NGV720727H, NGV721689H, NGV721787H, NGV723219H, NGV723779H, NGV725321H, NGV725611H, NGV725878H, NGV726212H, NGV726671H, NGV726887H, NGV727155H, NGV727290H, NGV727291H, NGV727295H, NGV727614H, NGV651128H, NGV705532H, NGV705895H, NGV709803H, NGV710442H, NGV710935H, NGV713524H, NGV716438H, NGV716444H, NGV716446H, NGV720391H, NGV725969H, NGV712617H, NGV717456H, NGV717642H, NGV717672H, NGV719220H, NGV000067H, NGV710283H, NGV001071H, NGV001072H, NGV001290H, NGV001446H, NGV001910H, NGV002755H, NGV002817H, NGV003255H, NGV003774H, NGV004753H, NGV651187H, NGV651215H, NGV651462H, NGV652530H, NGV701578H, NGV701616H, NGV702170H, NGV702800H, NGV703040H, NGV703365H, NGV703917H, NGV703984H, NGV704487H, NGV705856H, NGV706644H, NGV708991H, NGV710044H, NGV711048H, NGV711049H, NGV711584H, NGV712660H, NGV713051H, NGV713306H, NGV713310H, NGV713603H, NGV717814H, NGV717816H, NGV719400H, NGV720689H, NGV720899H, NGV721122H, NGV721153H, NGV721866H, NGV723306H, NGV724251H, NGV725371H, NGV725940H, NGV726593H, NGV702435H, NGV704363H, NGV707522H, NGV003892H, NGV701073H, NGV705671H, NGV708485H, NGV717109H, NGV717517H, NGV717518H, NGV723342H, NGV725346H, NGV726864H, NGV727792H, NGV711057H, NGV001794H, NGV700681H, NGV707944H, NGV710186H, NGV713265H, NGV727014H, NGV725816H, NGV719633H, NGV720385H, NGV001839H, NGV702271H, NGV000676H, NGV000705H, NGV000731H, NGV002068H, NGV701041H, NGV701280H, NGV703416H, NGV703581H, NGV703650H, NGV705573H, NGV708092H, NGV708775H, NGV708777H, NGV709429H, NGV715924H, NGV719215H, NGV720134H, NGV720135H, NGV003092H, NGV003095H, NGV003170H, NGV650591H, NGV651387H, NGV651393H, NGV701443H, NGV701542H, NGV701543H, NGV701555H, NGV702722H, NGV708089H, NGV711068H, NGV711533H, NGV711645H, NGV711826H, NGV711850H, NGV711937H, NGV712309H, NGV712639H, NGV712736H, NGV713600H, NGV716193H, NGV716199H, NGV716208H, NGV716234H, NGV716642H, NGV716925H, NGV719584H, NGV719586H, NGV720042H, NGV720630H, NGV722448H, NGV722660H, NGV722911H, NGV723525H, NGV725304H, NGV726360H, NGV727573H, NGV727033H, NGV703830H, NGV713301H, NGV721015H, NGV722400H, NGV000850H, NGV004259H, NGV701976H, NGV704749H, NGV704916H, NGV708140H, NGV709481H, NGV711284H, NGV711604H, NGV716965H, NGV718498H, NGV718863H, NGV718864H, NGV722675H, NGV722906H, NGV726365H, NGV001823H, NGV701504H, NGV702970H, NGV703911H, NGV704085H, NGV704087H, NGV707348H, NGV707366H, NGV709146H, NGV711246H, NGV711335H, NGV711825H, NGV713638H, NGV713896H, NGV713898H, NGV717626H, NGV719003H, NGV720016H, NGV720017H, NGV721830H, NGV722066H, NGV723930H, NGV001826H, NGV004221H, NGV651149H, NGV702751H, NGV704248H, NGV705683H, NGV725731H, NGV651192H, NGV651532H, NGV704415H, NGV715473H, NGV726591H, NGV721123H, NGV726388H, NGV701106H, NGV707593H, NGV712380H, NGV712809H, NGV000978H, NGV001319H, NGV001875H, NGV002021H, NGV003818H, NGV651064H, NGV651510H, NGV651546H, NGV651547H, NGV652031H, NGV700840H, NGV701885H, NGV703316H, NGV709269H, NGV718676H, NGV719288H, NGV719876H, NGV723226H, NGV726773H, NGV000227H, NGV004254H, NGV650167H, NGV651774H, NGV702681H, NGV704237H, NGV707617H, NGV715825H, NGV717126H, NGV718531H, NGV721376H, NGV723345H, NGV723159H, NGV651044H, NGV001788H, NGV713650H, NGV717888H, NGV723768H, NGV702080H, NGV702169H, NGV703364H, NGV703580H, NGV704112H, NGV704876H, NGV706267H, NGV712432H, NGV715991H, NGV717102H, NGV703833H, NGV722682H, NGV001800H, NGV003063H, NGV003064H, NGV651154H, NGV702330H, NGV702794H, NGV706412H, NGV706747H, NGV709209H, NGV720376H, NGV721891H, NGV722686H, NGV004858H, NGV705599H, NGV707153H, NGV715863H, NGV716361H, NGV718320H, NGV718492H, NGV719505H, NGV719507H, NGV721239H, NGV724733H, NGV725210H, NGV000781H, NGV001703H, NGV003991H, NGV004154H, NGV004720H, NGV650153H, NGV650588H, NGV651065H, NGV700198H, NGV700549H, NGV701743H, NGV701760H, NGV702165H, NGV704364H, NGV704799H, NGV705001H, NGV705014H, NGV705193H, NGV706370H, NGV706379H, NGV706868H, NGV710084H, NGV712202H, NGV713027H, NGV713585H, NGV713696H, NGV718174H, NGV718473H, NGV718500H, NGV720590H, NGV722681H, NGV722913H, NGV725274H, NGV725355H, NGV726613H, NGV704127H, NGV717035H, NGV651371H, NGV652528H, NGV701214H, NGV702953H, NGV704110H, NGV001256H, NGV003586H, NGV704426H, NGV708883H, NGV708967H, NGV711215H, NGV712659H, NGV712731H, NGV718313H, NGV721770H, NGV722753H, NGV723153H, NGV725192H, NGV725538H, NGV726895H, NGV000385H, NGV002236H, NGV709424H, NGV715969H, NGV718508H, NGV725239H, NGV726353H, NGV000328H, NGV000843H, NGV003249H, NGV004891H, NGV004998H, NGV652101H, NGV701138H, NGV703435H, NGV706431H, NGV706849H, NGV707049H, NGV711437H, NGV711438H, NGV712204H, NGV712687H, NGV712819H, NGV712820H, NGV714587H, NGV714591H, NGV715519H, NGV715611H, NGV717849H, NGV717850H, NGV723608H, NGV726574H, NGV727366H, NGV005194H, NGV000465H, NGV004164H, NGV004288H, NGV005196H, NGV651004H, NGV651964H, NGV705106H, NGV715259H, NGV720534H, NGV726708H, NGV000591H, NGV001772H, NGV704416H, NGV705512H, NGV706084H, NGV707583H, NGV708099H, NGV709558H, NGV711218H, NGV720374H, NGV720375H, NGV721385H, NGV725103H, NGV002191H, NGV652483H, NGV701397H, NGV704213H, NGV704214H, NGV001338H, NGV002078H, NGV001073H, NGV001885H, NGV002644H, NGV003875H, NGV004037H, NGV005066H, NGV701354H, NGV701439H, NGV708299H, NGV713501H, NGV717183H, NGV718001H, NGV719319H, NGV719711H, NGV003707H, NGV651366H, NGV651419H, NGV651485H, NGV704726H, NGV706878H, NGV710208H, NGV710309H, NGV711477H, NGV712301H, NGV717859H, NGV722794H, NGV723509H, NGV725314H, NGV705793H, NGV000488H, NGV001568H, NGV650713H, NGV706294H, NGV714177H, NGV724943H, NGV001566H, NGV001567H, NGV002055H, NGV004799H, NGV650180H, NGV702586H, NGV703017H, NGV704115H, NGV704207H, NGV704619H, NGV708199H, NGV709057H, NGV711699H, NGV712190H, NGV714390H, NGV716225H, NGV717003H, NGV720081H, NGV722542H, NGV722544H, NGV724941H, NGV726856H, NGV727283H, NGV727652H, NGV001221H, NGV001260H, NGV001262H, NGV000905H, NGV003058H, NGV003194H, NGV004528H, NGV005225H, NGV650337H, NGV651791H, NGV651795H, NGV700220H, NGV701160H, NGV701730H, NGV705864H, NGV705883H, NGV708488H, NGV708834H, NGV709154H, NGV711149H, NGV711561H, NGV714578H, NGV717087H, NGV717612H, NGV717613H, NGV717616H, NGV717712H, NGV718911H, NGV725867H, NGV726414H, NGV000240H, NGV002045H, NGV004186H, NGV704507H, NGV712254H, NGV002448H, NGV002694H, NGV706472H, NGV706474H, NGV710136H, NGV711605H, NGV716108H, NGV727695H, NGV000221H, NGV003145H, NGV650331H, NGV650359H, NGV702201H, NGV702503H, NGV709647H, NGV710045H, NGV721277H, NGV723637H, NGV002315H, NGV004936H, NGV700806H, NGV703369H, NGV705372H, NGV709823H, NGV709900H, NGV710307H, NGV711212H, NGV712851H, NGV000039H, NGV000830H, NGV001858H, NGV002138H, NGV002140H, NGV002346H, NGV002379H, NGV002691H, NGV003149H, NGV003178H, NGV003638H, NGV004027H, NGV004047H, NGV004524H, NGV650117H, NGV650266H, NGV650344H, NGV650684H, NGV650971H, NGV650982H, NGV650986H, NGV651206H, NGV651313H, NGV651342H, NGV651694H, NGV652076H, NGV652408H, NGV700255H, NGV700499H, NGV700957H, NGV700978H, NGV702229H, NGV702855H, NGV702861H, NGV703175H, NGV703341H, NGV703957H, NGV704107H, NGV704212H, NGV704957H, NGV705217H, NGV705272H, NGV705450H, NGV706321H, NGV706324H, NGV706556H, NGV707024H, NGV707313H, NGV707314H, NGV707381H, NGV707426H, NGV707427H, NGV707867H, NGV707960H, NGV708558H, NGV708646H, NGV708817H, NGV708954H, NGV709010H, NGV709767H, NGV709886H, NGV710054H, NGV711013H, NGV711037H, NGV711210H, NGV711824H, NGV712086H, NGV712087H, NGV712092H, NGV713609H, NGV713610H, NGV713611H, NGV713612H, NGV713891H, NGV715033H, NGV716712H, NGV717585H, NGV717889H, NGV717983H, NGV717984H, NGV718894H, NGV718912H, NGV719544H, NGV719689H, NGV720100H, NGV720101H, NGV720655H, NGV721079H, NGV721595H, NGV722449H, NGV723002H, NGV723228H, NGV723301H, NGV724017H, NGV724018H, NGV724100H, NGV724934H, NGV724936H, NGV726732H, NGV720487H, NGV724569H, NGV725222H, NGV725757H, NGV701446H, NGV714015H, NGV718540H, NGV000718H, NGV704116H, NGV711061H, NGV714182H, NGV717646H, NGV651049H, NGV710906H, NGV714086H, NGV715510H, NGV652196H, NGV713594H, NGV715735H, NGV002994H, NGV702324H, NGV708163H, NGV712762H, NGV002518H, NGV651606H, NGV700180H, NGV706722H, NGV708610H, NGV708712H, NGV711892H, NGV711956H, NGV717020H, NGV717287H, NGV723615H, NGV725215H, NGV725220H, NGV725318H, NGV725320H, NGV725398H, NGV725938H, NGV726055H, NGV727169H, NGV000920H, NGV001830H, NGV004446H, NGV650125H, NGV700333H, NGV700334H, NGV704434H, NGV704910H, NGV705666H, NGV706018H, NGV706098H, NGV706477H, NGV707389H, NGV707454H, NGV709589H, NGV709610H, NGV711015H, NGV711028H, NGV711029H, NGV712662H, NGV712665H, NGV712836H, NGV712837H, NGV714635H, NGV714636H, NGV714995H, NGV717794H, NGV717795H, NGV719409H, NGV723579H, NGV723581H, NGV726175H, NGV726177H, NGV727894H, NGV000235H, NGV000562H, NGV000578H, NGV000866H, NGV001025H, NGV001030H, NGV001088H, NGV001096H, NGV001243H, NGV001244H, NGV001245H, NGV001734H, NGV001809H, NGV002854H, NGV003043H, NGV003044H, NGV003478H, NGV003479H, NGV003480H, NGV004206H, NGV004207H, NGV004969H, NGV004970H, NGV650065H, NGV650067H, NGV650134H, NGV650399H, NGV650400H, NGV651151H, NGV652093H, NGV652094H, NGV652095H, NGV700446H, NGV701039H, NGV701145H, NGV701285H, NGV701287H, NGV701288H, NGV701291H, NGV701613H, NGV701624H, NGV701999H, NGV702093H, NGV702094H, NGV702095H, NGV702096H, NGV703114H, NGV703116H, NGV703124H, NGV703397H, NGV704063H, NGV704473H, NGV704474H, NGV704476H, NGV704701H, NGV704746H, NGV705248H, NGV705676H, NGV705677H, NGV705914H, NGV706169H, NGV706432H, NGV706433H, NGV707055H, NGV707056H, NGV707103H, NGV707108H, NGV707189H, NGV708136H, NGV708970H, NGV708988H, NGV709317H, NGV709663H, NGV709702H, NGV709704H, NGV709844H, NGV709884H, NGV709976H, NGV710230H, NGV710231H, NGV710920H, NGV710923H, NGV710966H, NGV711127H, NGV711191H, NGV711621H, NGV711861H, NGV712008H, NGV712105H, NGV712433H, NGV712533H, NGV712683H, NGV712686H, NGV712766H, NGV712767H, NGV713295H, NGV713518H, NGV713520H, NGV713604H, NGV713605H, NGV713711H, NGV713886H, NGV713889H, NGV714600H, NGV715740H, NGV717007H, NGV717514H, NGV717593H, NGV719880H, NGV720379H, NGV721905H, NGV722004H, NGV722005H, NGV722630H, NGV723799H, NGV725084H, NGV725086H, NGV725614H, NGV725615H, NGV726305H, NGV726616H, NGV726838H, NGV727177H, NGV727189H, NGV001218H, NGV003870H, NGV003871H, NGV651225H, NGV651226H, NGV707959H, NGV710235H, NGV711578H, NGV715518H, NGV716277H, NGV716279H, NGV716280H, NGV718401H, NGV721756H, NGV721834H, NGV725775H, NGV726217H, NGV726218H, NGV726219H, NGV726220H, NGV650363H, NGV651871H, NGV701388H, NGV705529H, NGV707023H, NGV707619H, NGV712561H, NGV718491H, NGV719913H, NGV719984H, NGV725773H, NGV725774H, NGV706258H, NGV713313H, NGV716938H, NGV719252H, NGV720851H, NGV701265H, NGV703380H, NGV705867H, NGV001227H, NGV705637H, NGV003151H, NGV722464H, NGV725843H, NGV000006H, NGV709633H, NGV003681H, NGV004218H, NGV650882H, NGV702448H, NGV703208H, NGV704457H, NGV705928H, NGV707508H, NGV709614H, NGV709615H, NGV710066H, NGV710067H, NGV712392H, NGV716971H, NGV716973H, NGV717323H, NGV718519H, NGV718924H, NGV719754H, NGV720862H, NGV721050H, NGV722697H, NGV723938H, NGV724599H, NGV725372H, NGV726781H, NGV726782H, NGV000161H, NGV001503H, NGV001753H, NGV001777H, NGV001896H, NGV003125H, NGV003315H, NGV003318H, NGV003547H, NGV004338H, NGV004869H, NGV004956H, NGV004957H, NGV005273H, NGV650694H, NGV650864H, NGV651045H, NGV651301H, NGV651516H, NGV651889H, NGV652414H, NGV652491H, NGV701161H, NGV701395H, NGV701462H, NGV701463H, NGV702113H, NGV702747H, NGV703561H, NGV703691H, NGV703994H, NGV704267H, NGV704625H, NGV704627H, NGV706178H, NGV706179H, NGV706180H, NGV706371H, NGV706770H, NGV706772H, NGV706939H, NGV706940H, NGV706942H, NGV706943H, NGV706945H, NGV708935H, NGV708936H, NGV708937H, NGV708938H, NGV711170H, NGV711172H, NGV712183H, NGV712481H, NGV715013H, NGV715017H, NGV715394H, NGV715396H, NGV715397H, NGV715399H, NGV715956H, NGV716256H, NGV716542H, NGV716568H, NGV716826H, NGV716877H, NGV716878H, NGV716881H, NGV717055H, NGV717057H, NGV717490H, NGV717491H, NGV717492H, NGV717494H, NGV719292H, NGV720149H, NGV720923H, NGV720925H, NGV721587H, NGV721902H, NGV721903H, NGV722891H, NGV723217H, NGV723739H, NGV723741H, NGV723742H, NGV723743H, NGV724738H, NGV725171H, NGV725181H, NGV725182H, NGV725184H, NGV725190H, NGV726508H, NGV726509H, NGV727162H, NGV727166H, NGV727167H, NGV727168H, NGV002971H, NGV005044H, NGV712025H, NGV714352H, NGV725811H, NGV003415H, NGV650782H, NGV652364H, NGV701116H, NGV701483H, NGV701655H, NGV709954H, NGV716249H, NGV003933H, NGV004108H, NGV700445H, NGV701020H, NGV701097H, NGV702406H, NGV704871H, NGV707582H, NGV707628H, NGV707629H, NGV707904H, NGV709097H, NGV709475H, NGV710051H, NGV711006H, NGV711650H, NGV711651H, NGV717050H, NGV717051H, NGV719152H, NGV719217H, NGV719819H, NGV719879H, NGV720451H, NGV720777H, NGV720944H, NGV721634H, NGV727207H, NGV727698H, NGV000712H, NGV000716H, NGV004842H, NGV700323H, NGV702753H, NGV703038H, NGV705238H, NGV705284H, NGV707327H, NGV709437H, NGV710106H, NGV710119H, NGV719864H, NGV719906H, NGV726561H, NGV727665H, NGV727666H, NGV002378H, NGV650083H, NGV650425H, NGV651076H, NGV708884H, NGV719646H, NGV719854H, NGV721914H, NGV723944H, NGV004427H, NGV000918H, NGV651650H, NGV708986H, NGV710899H, NGV711832H, NGV720380H, NGV721125H, NGV701239H, NGV716558H, NGV723225H, NGV726507H, NGV718292H, NGV002107H, NGV715647H, NGV716907H, NGV720864H, NGV002475H, NGV718680H, NGV718359H, NGV726179H, NGV709648H, NGV711931H, NGV718941H, NGV723321H, NGV725087H, NGV000089H, NGV001324H, NGV650946H, NGV707129H, NGV708953H, NGV711196H, NGV715500H, NGV717400H, NGV718369H, NGV721637H, NGV725469H, NGV701761H, NGV713990H, NGV000661H, NGV002397H, NGV650035H, NGV700432H, NGV702105H, NGV704069H, NGV705825H, NGV707010H, NGV707011H, NGV707646H, NGV708628H, NGV716569H, NGV721728H, NGV723662H, NGV716995H, NGV719553H, NGV719853H, NGV721850H, NGV722033H, NGV722860H, NGV722861H, NGV722863H, NGV001689H, NGV001690H, NGV001691H, NGV001693H, NGV001695H, NGV003319H, NGV003332H, NGV003333H, NGV004369H, NGV004639H, NGV703089H, NGV703347H, NGV704921H, NGV704982H, NGV707071H, NGV707435H, NGV711091H, NGV717163H, NGV719078H, NGV724616H, NGV726665H, NGV650567H, NGV000359H, NGV002094H, NGV002215H, NGV650177H, NGV650595H, NGV650630H, NGV652358H, NGV700213H, NGV700214H, NGV701206H, NGV702109H, NGV702161H, NGV702997H, NGV703237H, NGV704936H, NGV705294H, NGV712032H, NGV712750H, NGV712843H, NGV714293H, NGV716231H, NGV716901H, NGV716902H, NGV717147H, NGV719074H, NGV719135H, NGV723325H, NGV723549H, NGV723858H, NGV723866H, NGV725861H, NGV721299H, NGV650244H, NGV717401H, NGV000335H, NGV002844H, NGV003089H, NGV003091H, NGV651104H, NGV652103H, NGV652429H, NGV652436H, NGV700827H, NGV700828H, NGV700829H, NGV703889H, NGV703890H, NGV704436H, NGV704437H, NGV704438H, NGV704439H, NGV704440H, NGV705888H, NGV706008H, NGV708336H, NGV709204H, NGV709469H, NGV709529H, NGV710204H, NGV711487H, NGV713593H, NGV714014H, NGV714016H, NGV714017H, NGV716197H, NGV719212H, NGV721899H, NGV723319H, NGV725209H, NGV726083H, NGV726839H, NGV726840H, NGV727206H, NGV705693H, NGV002411H, NGV004828H, NGV004870H, NGV650673H, NGV700639H, NGV703981H, NGV708183H, NGV720104H, NGV722810H, NGV722856H, NGV723663H, NGV724653H, NGV724930H, NGV702688H, NGV713713H, NGV717992H, NGV720377H, NGV001213H, NGV651489H, NGV703671H, NGV705087H, NGV705923H, NGV709611H, NGV709925H, NGV711512H, NGV712064H, NGV712502H, NGV713670H, NGV715391H, NGV723328H, NGV726202H, NGV001968H, NGV004168H, NGV702696H, NGV703376H, NGV704692H, NGV712383H, NGV719536H, NGV724958H, NGV702444H, NGV002961H, NGV706641H, NGV000506H, NGV000507H, NGV002781H, NGV003578H, NGV004774H, NGV004775H, NGV651302H, NGV652125H, NGV700558H, NGV702491H, NGV704284H, NGV704713H, NGV704886H, NGV706406H, NGV711245H, NGV717371H, NGV718026H, NGV719815H, NGV720732H, NGV720994H, NGV721221H, NGV727645H, NGV650155H, NGV711689H, NGV709542H, NGV709972H, NGV651097H, NGV651648H, NGV709777H, NGV705206H, NGV709249H, NGV709730H, NGV707516H, NGV720455H, NGV720754H, NGV723893H, NGV000321H, NGV000875H, NGV650092H, NGV650423H, NGV651280H, NGV651572H, NGV651624H, NGV700355H, NGV700846H, NGV704020H, NGV704328H, NGV704894H, NGV705052H, NGV705258H, NGV713257H, NGV717543H, NGV717918H, NGV001252H, NGV001007H, NGV001581H, NGV651506H, NGV652139H, NGV652144H, NGV652146H, NGV652257H, NGV701447H, NGV701451H, NGV701465H, NGV702712H, NGV702721H, NGV704367H, NGV704500H, NGV707462H, NGV707463H, NGV707659H, NGV707863H, NGV711114H, NGV711607H, NGV713959H, NGV718134H, NGV704330H, NGV706288H, NGV711287H, NGV723656H, NGV706299H, NGV709891H, NGV718513H, NGV714464H, NGV723623H, NGV004403H, NGV651855H, NGV717922H, NGV719844H, NGV723373H, NGV707013H, NGV713026H, NGV713028H, NGV717182H, NGV003893H, NGV702216H, NGV703048H, NGV705650H, NGV706204H, NGV707360H, NGV707361H, NGV709434H, NGV718950H, NGV719907H, NGV721406H, NGV722497H, NGV725806H, NGV001222H, NGV004022H, NGV651987H, NGV700304H, NGV702065H, NGV707121H, NGV715416H, NGV717893H, NGV722024H, NGV725694H, NGV005253H, NGV702064H, NGV710214H, NGV716554H, NGV717892H, NGV720093H, NGV721086H, NGV723490H, NGV725331H, NGV726250H, NGV726504H, NGV005238H, NGV652053H, NGV703578H, NGV709047H, NGV709810H, NGV711094H, NGV711283H, NGV711543H, NGV712747H, NGV713308H, NGV715592H, NGV716136H, NGV717089H, NGV717652H, NGV720203H, NGV720970H, NGV003137H, NGV003138H, NGV005002H, NGV005005H, NGV650712H, NGV652219H, NGV652223H, NGV702994H, NGV702995H, NGV703224H, NGV707578H, NGV710962H, NGV710963H, NGV712732H, NGV712744H, NGV719416H, NGV719679H, NGV000288H, NGV001190H, NGV001812H, NGV002932H, NGV002937H, NGV002958H, NGV002985H, NGV003155H, NGV003156H, NGV003832H, NGV004153H, NGV004339H, NGV004394H, NGV004503H, NGV650818H, NGV651052H, NGV651397H, NGV651760H, NGV652474H, NGV700074H, NGV700357H, NGV700358H, NGV700749H, NGV700823H, NGV701470H, NGV703042H, NGV703241H, NGV703324H, NGV703539H, NGV703648H, NGV704254H, NGV704255H, NGV705388H, NGV705590H, NGV706730H, NGV706771H, NGV706871H, NGV706977H, NGV707185H, NGV707955H, NGV709001H, NGV709073H, NGV709651H, NGV709992H, NGV710433H, NGV711154H, NGV712484H, NGV715287H, NGV715288H, NGV715290H, NGV715807H, NGV715808H, NGV715812H, NGV715813H, NGV715884H, NGV717500H, NGV717664H, NGV717665H, NGV718038H, NGV718614H, NGV718683H, NGV719161H, NGV720413H, NGV721271H, NGV723320H, NGV724668H, NGV724670H, NGV724713H, NGV724789H, NGV725699H, NGV725765H, NGV725925H, NGV726625H, NGV726637H, NGV726876H, NGV000822H, NGV003142H, NGV003182H, NGV003183H, NGV003184H, NGV003185H, NGV003716H, NGV650353H, NGV651123H, NGV651194H, NGV651524H, NGV652057H, NGV652061H, NGV652065H, NGV701847H, NGV701848H, NGV704355H, NGV704356H, NGV705593H, NGV706329H, NGV706696H, NGV707332H, NGV707509H, NGV707565H, NGV707566H, NGV709501H, NGV709674H, NGV709761H, NGV709789H, NGV709852H, NGV710255H, NGV710334H, NGV711046H, NGV711583H, NGV711599H, NGV711600H, NGV711602H, NGV712083H, NGV712690H, NGV712812H, NGV714095H, NGV714096H, NGV714097H, NGV714422H, NGV716890H, NGV718878H, NGV720033H, NGV720885H, NGV721138H, NGV721774H, NGV721996H, NGV722356H, NGV723316H, NGV723533H, NGV723562H, NGV723671H, NGV725446H, NGV726007H, NGV710948H, NGV651592H, NGV709283H, NGV710305H, NGV000952H, NGV650929H, NGV702159H, NGV703298H, NGV703816H, NGV707062H, NGV707151H, NGV708630H, NGV711008H, NGV712708H, NGV716616H, NGV717923H, NGV721945H, NGV722466H, NGV722666H, NGV723167H, NGV726988H, NGV727559H, NGV000196H, NGV002672H, NGV004718H, NGV650740H, NGV651621H, NGV703657H, NGV704103H, NGV704431H, NGV704965H, NGV705625H, NGV705698H, NGV705699H, NGV706714H, NGV708605H, NGV708607H, NGV708615H, NGV708974H, NGV713351H, NGV715636H, NGV716580H, NGV716655H, NGV717160H, NGV717357H, NGV717619H, NGV717629H, NGV719239H, NGV719308H, NGV722992H, NGV723874H, NGV725981H, NGV000588H, NGV000653H, NGV003975H, NGV652080H, NGV652338H, NGV702633H, NGV705894H, NGV706122H, NGV708932H, NGV709277H, NGV709848H, NGV711139H, NGV712501H, NGV712531H, NGV712823H, NGV713829H, NGV715839H, NGV716588H, NGV721391H, NGV721748H, NGV723018H, NGV727535H, NGV000223H, NGV721208H, NGV002505H, NGV004547H, NGV718142H, NGV719509H, NGV720340H, NGV721252H, NGV004736H, NGV703564H, NGV709480H, NGV711020H, NGV717322H, NGV726426H, NGV000829H, NGV707289H, NGV001013H, NGV003841H, NGV004678H, NGV706515H, NGV723318H, NGV726493H, NGV726610H, NGV713956H, NGV720332H, NGV720389H, NGV726729H, NGV722991H, NGV725163H, NGV726364H, NGV702149H, NGV707284H, NGV711024H, NGV715248H, NGV715509H, NGV718985H, NGV651752H, NGV706375H, NGV711516H, NGV000093H, NGV651549H, NGV701048H, NGV717187H, NGV719871H, NGV726900H, NGV720929H, NGV725006H, NGV727114H, NGV650133H, NGV652004H, NGV704715H, NGV705354H, NGV709513H, NGV710038H, NGV711327H, NGV713633H, NGV713900H, NGV720323H, NGV721405H, NGV700146H, NGV706652H, NGV710268H, NGV716895H, NGV718219H, NGV718227H, NGV722692H, NGV722693H, NGV722694H, NGV722854H, NGV700622H, NGV706785H, NGV714012H, NGV714834H, NGV004502H, NGV700825H, NGV701089H, NGV718461H, NGV720958H, NGV721197H, NGV650320H, NGV718840H, NGV719924H, NGV721894H, NGV702584H, NGV702606H, NGV717544H, NGV718394H, NGV651904H, NGV708583H, NGV710988H, NGV716929H, NGV718235H, NGV719732H, NGV724057H, NGV725624H, NGV705281H, NGV705618H, NGV708946H, NGV723711H, NGV705218H, NGV709307H, NGV712486H, NGV721307H, NGV704417H, NGV000489H, NGV700212H, NGV703121H, NGV709840H, NGV711405H, NGV718019H, NGV000185H, NGV000186H, NGV000459H, NGV000529H, NGV000556H, NGV000681H, NGV000683H, NGV000689H, NGV000690H, NGV000691H, NGV002355H, NGV003741H, NGV650011H, NGV650038H, NGV650655H, NGV650858H, NGV650935H, NGV651084H, NGV701267H, NGV703135H, NGV704225H, NGV704335H, NGV704378H, NGV705613H, NGV705664H, NGV705882H, NGV708154H, NGV708829H, NGV709026H, NGV709137H, NGV709152H, NGV709585H, NGV709725H, NGV711060H, NGV711202H, NGV711701H, NGV713649H, NGV716241H, NGV716523H, NGV716620H, NGV716621H, NGV718031H, NGV719331H, NGV719334H, NGV719343H, NGV719435H, NGV721273H, NGV721274H, NGV722673H, NGV722674H, NGV723661H, NGV725180H, NGV726247H, NGV726770H, NGV000820H, NGV717188H, NGV718334H, NGV718335H, NGV718641H, NGV724928H, NGV726925H, NGV719919H, NGV715804H, NGV715809H, NGV715981H, NGV719257H, NGV719606H, NGV721993H, NGV721994H, NGV721995H, NGV722593H, NGV722699H, NGV722700H, NGV724631H, NGV724632H, NGV725018H, NGV725019H, NGV727696H, NGV727697H, NGV723171H, NGV702100H, NGV702101H, NGV001924H, NGV004937H, NGV706470H, NGV710089H, NGV727802H, NGV707982H, NGV722912H, NGV005256H, NGV720723H, NGV723712H, NGV721227H, NGV722747H, NGV723551H, NGV000259H, NGV000446H, NGV651691H, NGV000656H, NGV001562H, NGV650570H, NGV720592H, NGV720883H, NGV721685H, NGV726976H, NGV650080H, NGV725585H, NGV718960H, NGV700229H, NGV704725H, NGV709005H, NGV709132H, NGV709297H, NGV711097H, NGV715626H, NGV715867H, NGV003278H, NGV650961H, NGV651789H, NGV651797H, NGV703235H, NGV706670H, NGV706671H, NGV707165H, NGV710121H, NGV712852H, NGV713250H, NGV714190H, NGV714192H, NGV715601H, NGV715606H, NGV716679H, NGV722799H, NGV703829H, NGV700189H, NGV004935H, NGV704644H, NGV711261H, NGV712783H, NGV714592H, NGV703441H, NGV705303H, NGV707592H, NGV709487H, NGV716654H, NGV716867H, NGV716868H, NGV718006H, NGV720355H, NGV725995H, NGV726603H, NGV703036H, NGV706646H, NGV722795H, NGV723254H, NGV723844H, NGV726450H, NGV002052H, NGV002414H, NGV701425H, NGV702111H, NGV703920H, NGV707879H, NGV715496H, NGV716708H, NGV716722H, NGV716728H, NGV718576H, NGV722657H, NGV726544H, NGV001940H, NGV001941H, NGV651196H, NGV700193H, NGV703262H, NGV706880H, NGV707042H, NGV707876H, NGV720168H, NGV703139H, NGV709116H, NGV714141H, NGV719559H, NGV721494H, NGV726602H, NGV726906H, NGV718664H, NGV002017H, NGV004059H, NGV650919H, NGV651449H, NGV706527H, NGV710032H, NGV718596H, NGV722635H, NGV726290H, NGV000659H, NGV000805H, NGV003703H, NGV004104H, NGV702287H, NGV702755H, NGV703161H, NGV704073H, NGV705195H, NGV705768H, NGV708626H, NGV708794H, NGV710437H, NGV711297H, NGV711317H, NGV715613H, NGV718002H, NGV718535H, NGV721296H, NGV721297H, NGV722995H, NGV724568H, NGV724775H, NGV726084H, NGV001672H, NGV002339H, NGV004821H, NGV701697H, NGV716292H, NGV719085H, NGV721411H, NGV722395H, NGV725441H, NGV725488H, NGV718380H, NGV714199H, NGV714208H, NGV001041H, NGV706389H, NGV712551H, NGV001325H, NGV000368H, NGV001886H, NGV720091H, NGV713973H, NGV714607H, NGV722436H, NGV724033H, NGV724945H, NGV726643H, NGV002446H, NGV650033H, NGV650055H, NGV651561H, NGV652549H, NGV700454H, NGV701836H, NGV702070H, NGV703628H, NGV703835H, NGV704010H, NGV704608H, NGV707464H, NGV709410H, NGV709535H, NGV709822H, NGV710118H, NGV712394H, NGV712606H, NGV712829H, NGV713263H, NGV713531H, NGV715580H, NGV717365H, NGV717962H, NGV718055H, NGV718257H, NGV721243H, NGV722454H, NGV723341H, NGV726249H, NGV000226H, NGV002524H, NGV002525H, NGV002942H, NGV004959H, NGV650139H, NGV650846H, NGV651399H, NGV651833H, NGV652476H, NGV701572H, NGV701660H, NGV701718H, NGV702168H, NGV704603H, NGV712546H, NGV716148H, NGV722739H, NGV002131H, NGV713643H, NGV718347H, NGV721240H, NGV724675H, NGV000042H, NGV716977H, NGV718518H, NGV721754H, NGV003816H, NGV718722H, NGV718723H, NGV718968H, NGV719387H, NGV000085H, NGV000501H, NGV001134H, NGV001230H, NGV001512H, NGV002244H, NGV002520H, NGV002681H, NGV002750H, NGV003436H, NGV003476H, NGV003575H, NGV004118H, NGV004329H, NGV651209H, NGV651388H, NGV652006H, NGV652355H, NGV700324H, NGV700416H, NGV700725H, NGV700947H, NGV701158H, NGV701297H, NGV701681H, NGV702212H, NGV702495H, NGV702598H, NGV702850H, NGV703022H, NGV703361H, NGV703384H, NGV704016H, NGV704060H, NGV704297H, NGV704612H, NGV704984H, NGV704985H, NGV704986H, NGV705291H, NGV705620H, NGV706311H, NGV706312H, NGV706347H, NGV706368H, NGV706369H, NGV707266H, NGV707302H, NGV707367H, NGV707424H, NGV707425H, NGV709219H, NGV709220H, NGV709221H, NGV709772H, NGV710197H, NGV710932H, NGV711016H, NGV711141H, NGV711364H, NGV711823H, NGV712084H, NGV712085H, NGV712764H, NGV713615H, NGV713653H, NGV714168H, NGV714169H, NGV715512H, NGV718909H, NGV719691H, NGV722867H, NGV723397H, NGV723398H, NGV723464H, NGV000263H, NGV001711H, NGV650335H, NGV705313H, NGV001976H, NGV651251H, NGV701654H, NGV709050H, NGV712319H, NGV716268H, NGV727655H, NGV004289H, NGV700638H, NGV702436H, NGV702508H, NGV703222H, NGV705231H, NGV705234H, NGV706401H, NGV708643H, NGV709669H, NGV709670H, NGV711089H, NGV713913H, NGV715419H, NGV715468H, NGV717181H, NGV719813H, NGV000362H, NGV706572H, NGV706936H, NGV710289H, NGV710991H, NGV721784H, NGV721037H, NGV724938H, NGV702708H, NGV702928H, NGV001062H, NGV650879H, NGV651473H, NGV651828H, NGV651829H, NGV652127H, NGV702125H, NGV702415H, NGV702919H, NGV705380H, NGV705383H, NGV706392H, NGV706822H, NGV707334H, NGV710190H, NGV710191H, NGV711817H, NGV712682H, NGV712796H, NGV714255H, NGV714588H, NGV714590H, NGV717811H, NGV718114H, NGV718115H, NGV718188H, NGV718484H, NGV718486H, NGV719201H, NGV719668H, NGV719670H, NGV719822H, NGV723280H, NGV723724H, NGV723725H, NGV724211H, NGV725119H, NGV726486H, NGV726487H, NGV711347H, NGV725373H, NGV003804H, NGV708942H, NGV711324H, NGV711847H, NGV719483H, NGV724602H, NGV725739H, NGV705494H, NGV004757H, NGV651509H, NGV703421H, NGV706909H, NGV700444H, NGV703823H, NGV705524H, NGV705601H, NGV713694H, NGV715689H, NGV720700H, NGV002237H, NGV003983H, NGV701216H, NGV701820H, NGV707615H, NGV000134H, NGV700159H, NGV700515H, NGV701539H, NGV702689H, NGV720467H, NGV726138H, NGV650166H, NGV711275H, NGV722585H, NGV722750H, NGV723172H, NGV726662H, NGV712849H, NGV713530H, NGV716652H, NGV721303H, NGV721304H, NGV723535H, NGV724676H, NGV725863H, NGV707495H, NGV713299H, NGV723186H, NGV002756H, NGV702172H, NGV705176H, NGV705319H, NGV706097H, NGV723506H, NGV004851H, NGV701884H, NGV705922H, NGV706352H, NGV710184H, NGV712627H, NGV712772H, NGV715716H, NGV726346H, NGV004251H, NGV004252H, NGV650748H, NGV650783H, NGV705268H, NGV709337H, NGV716350H, NGV716448H, NGV717257H, NGV719759H, NGV000003H, NGV702603H, NGV703886H, NGV703936H, NGV710320H, NGV712681H, NGV712842H, NGV718170H, NGV726886H, NGV000324H, NGV001301H, NGV002399H, NGV002400H, NGV002402H, NGV002508H, NGV004682H, NGV004787H, NGV004788H, NGV651160H, NGV651175H, NGV702560H, NGV703102H, NGV703105H, NGV703128H, NGV703130H, NGV704614H, NGV704974H, NGV704975H, NGV707044H, NGV707046H, NGV707394H, NGV708838H, NGV708839H, NGV708840H, NGV710020H, NGV710022H, NGV710023H, NGV710296H, NGV711351H, NGV711352H, NGV711353H, NGV711442H, NGV711529H, NGV711551H, NGV711552H, NGV711665H, NGV711929H, NGV712388H, NGV712490H, NGV715806H, NGV715816H, NGV715818H, NGV718194H, NGV718203H, NGV718573H, NGV718574H, NGV721103H, NGV721104H, NGV721105H, NGV721106H, NGV724755H, NGV724756H, NGV724757H, NGV724758H, NGV724759H, NGV001049H, NGV001303H, NGV001305H, NGV001377H, NGV001380H, NGV001554H, NGV001934H, NGV001935H, NGV001936H, NGV001937H, NGV002460H, NGV002461H, NGV002462H, NGV002463H, NGV002464H, NGV002465H, NGV002466H, NGV002467H, NGV003232H, NGV003233H, NGV003234H, NGV003236H, NGV003239H, NGV003242H, NGV651414H, NGV651415H, NGV652322H, NGV652404H, NGV652406H, NGV652508H, NGV652512H, NGV652513H, NGV652514H, NGV652516H, NGV652518H, NGV701177H, NGV701179H, NGV701180H, NGV701181H, NGV701183H, NGV701184H, NGV701185H, NGV701186H, NGV701187H, NGV701188H, NGV701189H, NGV701191H, NGV701193H, NGV701194H, NGV701195H, NGV701196H, NGV701197H, NGV703865H, NGV703867H, NGV703869H, NGV703876H, NGV704766H, NGV704767H, NGV704769H, NGV704770H, NGV704771H, NGV704773H, NGV704774H, NGV704776H, NGV705514H, NGV706120H, NGV706627H, NGV706731H, NGV707187H, NGV708214H, NGV708215H, NGV708216H, NGV708217H, NGV708219H, NGV708220H, NGV708221H, NGV709420H, NGV710327H, NGV710336H, NGV710337H, NGV710941H, NGV711494H, NGV711704H, NGV711705H, NGV711707H, NGV711708H, NGV711709H, NGV711711H, NGV711712H, NGV711714H, NGV711715H, NGV711716H, NGV711978H, NGV711979H, NGV711980H, NGV711981H, NGV712870H, NGV713054H, NGV713315H, NGV714143H, NGV714146H, NGV715830H, NGV715832H, NGV716953H, NGV717034H, NGV717037H, NGV717038H, NGV717040H, NGV717133H, NGV717134H, NGV717342H, NGV718426H, NGV718427H, NGV718979H, NGV718980H, NGV718982H, NGV718983H, NGV719352H, NGV719354H, NGV719355H, NGV719357H, NGV720560H, NGV720561H, NGV720562H, NGV720564H, NGV720565H, NGV720566H, NGV720955H, NGV725972H, NGV726041H, NGV001682H, NGV706989H, NGV715544H, NGV725214H, NGV721319H, NGV725755H, NGV000173H, NGV000666H, NGV001225H, NGV706253H, NGV706263H, NGV711031H, NGV712777H, NGV717361H, NGV718169H, NGV722989H, NGV723131H, NGV723507H, NGV001730H, NGV002033H, NGV003541H, NGV650255H, NGV701640H, NGV702314H, NGV702680H, NGV703267H, NGV704462H, NGV705316H, NGV706208H, NGV715803H, NGV716743H, NGV717347H, NGV717968H, NGV726286H, NGV726565H, NGV726576H, NGV650924H, NGV709801H, NGV711244H, NGV003684H, NGV651410H, NGV700332H, NGV703907H, NGV707607H, NGV707608H, NGV709002H, NGV711703H, NGV702998H, NGV700433H, NGV726771H, NGV003923H, NGV005277H, NGV650316H, NGV701241H, NGV702770H, NGV707061H, NGV651299H, NGV707080H, NGV711612H, NGV715624H, NGV651673H, NGV700091H, NGV713318H, NGV714386H, NGV724589H, NGV726967H, NGV001522H, NGV004723H, NGV651112H, NGV703840H, NGV711575H, NGV717356H, NGV723872H, NGV000423H, NGV004725H, NGV651620H, NGV702104H, NGV704279H, NGV704281H, NGV705389H, NGV705990H, NGV708984H, NGV716578H, NGV726306H, NGV000669H, NGV000901H, NGV003108H, NGV003687H, NGV004769H, NGV651122H, NGV651327H, NGV700158H, NGV701151H, NGV701208H, NGV702087H, NGV704195H, NGV704852H, NGV711789H, NGV711791H, NGV713363H, NGV718125H, NGV718306H, NGV702486H, NGV716226H, NGV717483H, NGV721888H, NGV723072H, NGV726498H, NGV716611H, NGV727852H, NGV000977H, NGV001843H, NGV001846H, NGV650657H, NGV706356H, NGV004042H, NGV005153H, NGV651796H, NGV704969H, NGV705445H, NGV709956H, NGV710192H, NGV721677H, NGV727052H, NGV717597H, NGV716847H, NGV651343H, NGV000477H, NGV000566H, NGV000715H, NGV001212H, NGV003005H, NGV004137H, NGV704191H, NGV704992H, NGV706291H, NGV707006H, NGV709065H, NGV709739H, NGV709740H, NGV710205H, NGV711928H, NGV725003H, NGV001833H, NGV003345H, NGV650518H, NGV700629H, NGV705728H, NGV707090H, NGV709321H, NGV709536H, NGV710017H, NGV000127H, NGV000680H, NGV002108H, NGV003563H, NGV004452H, NGV650192H, NGV652540H, NGV652543H, NGV701207H, NGV702530H, NGV705909H, NGV708599H, NGV709623H, NGV712347H, NGV714634H, NGV717802H, NGV718204H, NGV718617H, NGV721904H, NGV722998H, NGV723726H, NGV723827H, NGV726399H, NGV721777H, NGV725996H, NGV004155H, NGV005279H, NGV000485H, NGV000726H, NGV002680H, NGV004176H, NGV004301H, NGV651025H, NGV706350H, NGV708434H, NGV708624H, NGV709138H, NGV710401H, NGV718355H, NGV719631H, NGV720776H, NGV720780H, NGV722504H, NGV722724H, NGV726833H, NGV000241H, NGV001128H, NGV717405H, NGV717535H, NGV718886H, NGV719270H, NGV719426H, NGV719543H, NGV720490H, NGV721199H, NGV725463H, NGV002024H, NGV722057H, NGV000593H, NGV002543H, NGV701205H, NGV701682H, NGV704391H, NGV704392H, NGV706612H, NGV707186H, NGV710145H, NGV715690H, NGV715710H, NGV723227H, NGV723243H, NGV723302H, NGV002395H, NGV000326H, NGV001298H, NGV003370H, NGV701524H, NGV701648H, NGV700834H, NGV708481H, NGV708484H, NGV708486H, NGV716304H, NGV718062H, NGV718908H, NGV002700H, NGV701762H, NGV701766H, NGV701808H, NGV706930H, NGV706932H, NGV706973H, NGV706974H, NGV005207H, NGV651657H, NGV702061H, NGV704078H, NGV005037H, NGV005038H, NGV005046H, NGV650391H, NGV652187H, NGV652326H, NGV708327H, NGV708328H, NGV710369H, NGV710370H, NGV710371H, NGV652202H, NGV652203H, NGV652208H, NGV652244H, NGV652253H, NGV652308H, NGV652311H, NGV652312H, NGV652313H, NGV652329H, NGV703405H, NGV703406H, NGV703407H, NGV705907H, NGV705908H, NGV705911H, NGV705912H, NGV707231H, NGV707232H, NGV707234H, NGV707236H, NGV708318H, NGV708320H, NGV708322H, NGV708330H, NGV716573H, NGV716806H, NGV717302H, NGV717303H, NGV718216H, NGV718220H, NGV718794H, NGV718795H, NGV719564H, NGV719566H, NGV720946H, NGV720947H, NGV720949H, NGV720950H, NGV720952H, NGV723377H, NGV725037H, NGV725038H, NGV725807H, NGV001871H, NGV001883H, NGV003417H, NGV003420H, NGV004142H, NGV004242H, NGV004245H, NGV004341H, NGV004945H, NGV650164H, NGV701074H, NGV701101H, NGV701329H, NGV701592H, NGV701977H, NGV702402H, NGV702403H, NGV702872H, NGV703023H, NGV703588H, NGV704120H, NGV704764H, NGV708300H, NGV708303H, NGV708304H, NGV708305H, NGV708311H, NGV708313H, NGV708314H, NGV708315H, NGV708317H, NGV708667H, NGV708870H, NGV709580H, NGV710374H, NGV711530H, NGV719191H, NGV719572H, NGV720012H, NGV723283H, NGV000492H, NGV000498H, NGV002935H, NGV005029H, NGV650560H, NGV650701H, NGV701037H, NGV701110H, NGV701316H, NGV701319H, NGV701320H, NGV701322H, NGV701324H, NGV701325H, NGV701326H, NGV701328H, NGV701331H, NGV701332H, NGV701333H, NGV701334H, NGV701335H, NGV701338H, NGV701339H, NGV701340H, NGV701341H, NGV701342H, NGV701343H, NGV701344H, NGV701349H, NGV701350H, NGV701351H, NGV701352H, NGV701353H, NGV701355H, NGV701357H, NGV701360H, NGV703180H, NGV704756H, NGV706243H, NGV708094H, NGV708096H, NGV708097H, NGV708100H, NGV708102H, NGV708103H, NGV708104H, NGV708105H, NGV708106H, NGV708107H, NGV708108H, NGV708109H, NGV708111H, NGV708112H, NGV708113H, NGV708114H, NGV708119H, NGV708120H, NGV708153H, NGV708171H, NGV708172H, NGV708173H, NGV708174H, NGV708175H, NGV708176H, NGV709781H, NGV710995H, NGV723369H, NGV724820H, NGV725348H, NGV725349H, NGV725351H, NGV725356H, NGV725357H, NGV725359H, NGV725361H, NGV725676H, NGV726934H, NGV727797H, NGV727798H, NGV652192H, NGV652319H, NGV652321H, NGV652323H, NGV652325H, NGV707244H, NGV707248H, NGV708159H, NGV708162H, NGV710110H, NGV001515H, NGV652148H, NGV652149H, NGV652150H, NGV652154H, NGV652156H, NGV652158H, NGV652159H, NGV652162H, NGV652163H, NGV652165H, NGV652166H, NGV652169H, NGV701666H, NGV701668H, NGV701669H, NGV701674H, NGV701676H, NGV701677H, NGV704360H, NGV706823H, NGV706824H, NGV706826H, NGV707249H, NGV707251H, NGV707252H, NGV707253H, NGV707254H, NGV707255H, NGV707400H, NGV707401H, NGV707402H, NGV707404H, NGV707406H, NGV707408H, NGV707409H, NGV707410H, NGV707413H, NGV707416H, NGV707419H, NGV707421H, NGV707423H, NGV707660H, NGV707662H, NGV707666H, NGV707668H, NGV707671H, NGV707675H, NGV707676H, NGV707679H, NGV707682H, NGV707683H, NGV707685H, NGV711129H, NGV711241H, NGV711400H, NGV711662H, NGV711900H, NGV714277H, NGV714281H, NGV714284H, NGV714609H, NGV718205H, NGV719026H, NGV719302H, NGV719304H, NGV719306H, NGV719307H, NGV719575H, NGV719576H, NGV720065H, NGV720066H, NGV720067H, NGV720069H, NGV720070H, NGV720071H, NGV720072H, NGV720073H, NGV721482H, NGV721483H, NGV721484H, NGV721488H, NGV721489H, NGV721490H, NGV721836H, NGV723666H, NGV723668H, NGV723669H, NGV723670H, NGV724313H, NGV724314H, NGV724315H, NGV724317H, NGV724318H, NGV724477H, NGV725230H, NGV725233H, NGV725242H, NGV725245H, NGV725250H, NGV725252H, NGV725526H, NGV726572H, NGV726573H, NGV726796H, NGV726797H, NGV727355H, NGV727356H, NGV000222H, NGV002814H, NGV650688H, NGV651071H, NGV700264H, NGV701926H, NGV702152H, NGV709639H, NGV713984H, NGV717627H, NGV704485H, NGV716216H, NGV716884H, NGV000160H, NGV001091H, NGV701603H, NGV701605H, NGV701606H, NGV701607H, NGV703456H, NGV703473H, NGV703477H, NGV703706H, NGV703726H, NGV704217H, NGV704219H, NGV704220H, NGV704223H, NGV704228H, NGV704729H, NGV709078H, NGV709079H, NGV709080H, NGV709081H, NGV715633H, NGV720576H, NGV650722H, NGV651041H, NGV700932H, NGV720887H, NGV721127H, NGV000195H, NGV001052H, NGV002125H, NGV003604H, NGV003608H, NGV650631H, NGV651109H, NGV702127H, NGV702532H, NGV703455H, NGV704344H, NGV708786H, NGV708799H, NGV709492H, NGV709805H, NGV709806H, NGV711146H, NGV712361H, NGV712476H, NGV712512H, NGV714268H, NGV717580H, NGV717620H, NGV719248H, NGV726246H, NGV726222H, NGV710952H, NGV651047H, NGV651048H, NGV714254H, NGV000273H, NGV003448H, NGV003861H, NGV700624H, NGV705484H, NGV705485H, NGV705487H, NGV705488H, NGV705491H, NGV705492H, NGV705493H, NGV705495H, NGV705496H, NGV705497H, NGV705498H, NGV705499H, NGV705500H, NGV705501H, NGV705502H, NGV705503H, NGV705511H, NGV705515H, NGV705528H, NGV705534H, NGV705536H, NGV705542H, NGV705543H, NGV705544H, NGV705546H, NGV705547H, NGV705548H, NGV705550H, NGV705551H, NGV705552H, NGV705553H, NGV705555H, NGV705556H, NGV705557H, NGV705558H, NGV705565H, NGV705566H, NGV705569H, NGV707520H, NGV707888H, NGV707892H, NGV707893H, NGV707895H, NGV707897H, NGV707898H, NGV708080H, NGV708082H, NGV708083H, NGV708084H, NGV709367H, NGV709991H, NGV710893H, NGV714064H, NGV714065H, NGV714067H, NGV714071H, NGV714072H, NGV716109H, NGV716119H, NGV716122H, NGV716181H, NGV716637H, NGV716638H, NGV716643H, NGV716837H, NGV717072H, NGV717604H, NGV717605H, NGV717737H, NGV717933H, NGV717970H, NGV718198H, NGV719673H, NGV719674H, NGV719675H, NGV719676H, NGV719884H, NGV719885H, NGV719886H, NGV719889H, NGV719891H, NGV721697H, NGV721701H, NGV721702H, NGV721915H, NGV721916H, NGV721918H, NGV721919H, NGV722513H, NGV722515H, NGV722517H, NGV722519H, NGV722520H, NGV722521H, NGV722522H, NGV722523H, NGV722525H, NGV722526H, NGV723051H, NGV723052H, NGV723053H, NGV723054H, NGV723055H, NGV723537H, NGV723902H, NGV725047H, NGV725049H, NGV725174H, NGV725176H, NGV725177H, NGV725178H, NGV726118H, NGV726119H, NGV726120H, NGV726121H, NGV726122H, NGV726495H, NGV726858H, NGV726860H, NGV727376H, NGV727379H, NGV727381H, NGV000062H, NGV000777H, NGV001598H, NGV003406H, NGV003408H, NGV004860H, NGV651380H, NGV651425H, NGV702575H, NGV702791H, NGV703069H, NGV704875H, NGV709052H, NGV710088H, NGV715412H, NGV716592H, NGV716602H, NGV716909H, NGV719251H, NGV721229H, NGV726389H, NGV713589H, NGV720525H, NGV721006H, NGV724764H, NGV726632H, NGV000433H, NGV719227H, NGV725199H, NGV727676H, NGV719912H, NGV004063H, NGV005259H, NGV005260H, NGV704550H, NGV708310H, NGV710185H, NGV714085H, NGV715628H, NGV715882H, NGV004348H, NGV005278H, NGV652141H, NGV714566H, NGV717084H, NGV724095H, NGV700545H, NGV711050H, NGV715466H, NGV717261H, NGV717363H, NGV717364H, NGV702434H, NGV703373H, NGV704302H, NGV000057H, NGV652498H, NGV001818H, NGV725684H, NGV000179H, NGV000218H, NGV002020H, NGV002420H, NGV004389H, NGV651274H, NGV651281H, NGV700856H, NGV700857H, NGV701477H, NGV701635H, NGV701867H, NGV701969H, NGV703975H, NGV704300H, NGV704993H, NGV704994H, NGV705452H, NGV706262H, NGV706266H, NGV707081H, NGV707479H, NGV707504H, NGV707587H, NGV709864H, NGV710035H, NGV711285H, NGV712778H, NGV716450H, NGV716451H, NGV716462H, NGV716745H, NGV716849H, NGV719852H, NGV720743H, NGV721279H, NGV721804H, NGV721805H, NGV721845H, NGV721847H, NGV721848H, NGV722029H, NGV722030H, NGV722945H, NGV723834H, NGV723878H, NGV725626H, NGV726875H, NGV715801H, NGV707964H, NGV715540H, NGV000107H, NGV000108H, NGV000113H, NGV000114H, NGV000115H, NGV000117H, NGV000118H, NGV000119H, NGV000120H, NGV000122H, NGV650853H, NGV708253H, NGV708254H, NGV708255H, NGV708256H, NGV708258H, NGV708295H, NGV708297H, NGV709665H, NGV712040H, NGV714027H, NGV718935H, NGV723871H, NGV724062H, NGV724063H, NGV724225H, NGV725979H, NGV725980H, NGV726213H, NGV726308H, NGV000710H, NGV001767H, NGV701658H, NGV703437H, NGV705257H, NGV708965H, NGV713033H, NGV726173H, NGV727140H, NGV727861H, NGV727862H, NGV722710H, NGV725742H, NGV705505H, NGV708645H, NGV717947H, NGV652063H, NGV716700H, NGV723698H, NGV726301H, NGV727632H, NGV650640H, NGV716746H, NGV717978H, NGV004065H, NGV712635H, NGV715471H, NGV726162H, NGV717899H, NGV652381H, NGV000060H, NGV000391H, NGV004350H, NGV702583H, NGV711295H, NGV713632H, NGV707375H, NGV000205H, NGV000410H, NGV003550H, NGV003551H, NGV003683H, NGV005064H, NGV650836H, NGV651459H, NGV702433H, NGV702505H, NGV702866H, NGV703582H, NGV704420H, NGV704785H, NGV705952H, NGV708483H, NGV709271H, NGV709275H, NGV709854H, NGV711184H, NGV711743H, NGV714206H, NGV714341H, NGV715755H, NGV715758H, NGV718066H, NGV718747H, NGV718748H, NGV718783H, NGV718786H, NGV720543H, NGV720747H, NGV721605H, NGV721606H, NGV721607H, NGV722855H, NGV725914H, NGV725997H, NGV726183H, NGV004958H, NGV000573H, NGV705589H, NGV705657H, NGV706106H, NGV707269H, NGV707312H, NGV717251H, NGV717793H, NGV718209H, NGV718640H, NGV721294H, NGV721295H, NGV721572H, NGV725809H, NGV726540H, NGV726837H, NGV708776H, NGV000575H, NGV001792H, NGV650089H, NGV651803H, NGV703244H, NGV705151H, NGV715564H, NGV724015H, NGV000254H, NGV000286H, NGV003887H, NGV701233H, NGV703271H, NGV703318H, NGV713816H, NGV001722H, NGV001988H, NGV004961H, NGV700233H, NGV702579H, NGV706372H, NGV716293H, NGV718024H, NGV721969H, NGV723686H, NGV724892H, NGV000167H, NGV000704H, NGV001754H, NGV001821H, NGV002816H, NGV700297H, NGV700553H, NGV705181H, NGV705583H, NGV706307H, NGV714343H, NGV714344H, NGV715748H, NGV716705H, NGV717374H, NGV718792H, NGV718963H, NGV719943H, NGV725094H, NGV725099H, NGV726201H, NGV726231H, NGV000249H, NGV000251H, NGV001005H, NGV001333H, NGV003859H, NGV004602H, NGV004779H, NGV651268H, NGV701051H, NGV702222H, NGV705375H, NGV705708H, NGV706238H, NGV708602H, NGV708842H, NGV709009H, NGV709011H, NGV716153H, NGV717424H, NGV717426H, NGV719084H, NGV719900H, NGV721191H, NGV723621H, NGV723792H, NGV723793H, NGV710134H, NGV719701H, NGV001207H, NGV705644H, NGV700215H, NGV704724H, NGV709782H, NGV002194H, NGV004627H, NGV707280H, NGV707380H, NGV717633H, NGV722062H, NGV725661H, NGV706245H, NGV706139H, NGV717294H, NGV722399H, NGV724090H, NGV725093H, NGV001320H, NGV701098H, NGV703138H, NGV703342H, NGV707506H, NGV712607H, NGV715493H, NGV716899H, NGV726780H, NGV726824H, NGV001040H, NGV650422H, NGV652217H, NGV717985H, NGV719011H, NGV001879H, NGV002922H, NGV003198H, NGV003199H, NGV003202H, NGV003205H, NGV003207H, NGV003209H, NGV003211H, NGV003213H, NGV003214H, NGV003218H, NGV003219H, NGV003222H, NGV003223H, NGV003225H, NGV003226H, NGV003227H, NGV003251H, NGV003259H, NGV003270H, NGV003271H, NGV003274H, NGV652611H, NGV652614H, NGV652620H, NGV652621H, NGV652624H, NGV652627H, NGV652631H, NGV652633H, NGV652640H, NGV652641H, NGV652643H, NGV702430H, NGV704646H, NGV704647H, NGV704648H, NGV704649H, NGV704652H, NGV704653H, NGV704655H, NGV704656H, NGV704657H, NGV704658H, NGV704662H, NGV704664H, NGV704667H, NGV704668H, NGV704669H, NGV704671H, NGV704672H, NGV704673H, NGV704678H, NGV704679H, NGV704680H, NGV704682H, NGV704685H, NGV704686H, NGV705037H, NGV707223H, NGV710233H, NGV710251H, NGV711169H, NGV711407H, NGV711408H, NGV711411H, NGV711418H, NGV711419H, NGV711420H, NGV711423H, NGV714320H, NGV714582H, NGV715316H, NGV715395H, NGV715428H, NGV715488H, NGV718633H, NGV718841H, NGV718987H, NGV719580H, NGV720075H, NGV720351H, NGV720502H, NGV720615H, NGV720734H, NGV721473H, NGV722847H, NGV723132H, NGV726030H, NGV726044H, NGV726046H, NGV726047H, NGV726048H, NGV726050H, NGV726051H, NGV726053H, NGV726056H, NGV726057H, NGV726058H, NGV726061H, NGV726062H, NGV726066H, NGV726067H, NGV726068H, NGV720333H, NGV001747H, NGV705178H, NGV715530H, NGV716214H, NGV717759H, NGV704731H, NGV711250H, NGV715952H, NGV721101H, NGV727016H, NGV000684H, NGV001094H, NGV002066H, NGV700569H, NGV711473H, NGV004299H, NGV710488H, NGV723881H, NGV710981H, NGV712619H, NGV004085H, NGV723170H, NGV000921H, NGV003597H, NGV706839H, NGV650115H, NGV707270H, NGV710934H, NGV711150H, NGV711435H, NGV714094H, NGV718050H, NGV721067H, NGV004809H, NGV702856H, NGV705136H, NGV705391H, NGV706709H, NGV726382H, NGV726505H, NGV650629H, NGV707393H, NGV711546H, NGV705086H, NGV725012H, NGV702062H, NGV703266H, NGV703698H, NGV708662H, NGV709229H, NGV710924H, NGV707599H, NGV710461H, NGV721241H, NGV001989H, NGV002993H, NGV651360H, NGV651926H, NGV702308H, NGV703051H, NGV705816H, NGV706862H, NGV707215H, NGV707951H, NGV709897H, NGV710200H, NGV710202H, NGV710411H, NGV711513H, NGV712557H, NGV713683H, NGV717828H, NGV718184H, NGV721828H, NGV723543H, NGV723836H, NGV723992H, NGV725872H, NGV726499H, NGV726530H, NGV727118H, NGV004018H, NGV717879H, NGV720889H, NGV724654H, NGV724981H, NGV001930H, NGV003087H, NGV704955H, NGV650329H, NGV713513H, NGV716151H, NGV722023H, NGV003157H, NGV650419H, NGV650420H, NGV650421H, NGV651262H, NGV651920H, NGV652250H, NGV700308H, NGV700336H, NGV700468H, NGV700852H, NGV701532H, NGV701533H, NGV706427H, NGV707009H, NGV710777H, NGV710779H, NGV710780H, NGV711107H, NGV711628H, NGV711629H, NGV715955H, NGV716685H, NGV717328H, NGV717329H, NGV717330H, NGV717331H, NGV717368H, NGV720421H, NGV720423H, NGV720918H, NGV720919H, NGV720920H, NGV722631H, NGV725746H, NGV725747H, NGV725749H, NGV725750H, NGV725751H, NGV725752H, NGV725753H, NGV725754H, NGV004359H, NGV650276H, NGV651512H, NGV702422H, NGV709892H, NGV711557H, NGV711884H, NGV712109H, NGV723414H, NGV724311H, NGV703938H, NGV705697H, NGV705878H, NGV708612H, NGV716976H, NGV718000H, NGV719347H, NGV001236H, NGV702075H, NGV703229H, NGV710057H, NGV710090H, NGV711359H, NGV712637H, NGV713515H, NGV719192H, NGV719728H, NGV723795H, NGV000486H, NGV003723H, NGV005120H, NGV651289H, NGV707040H, NGV711818H, NGV713592H, NGV714266H, NGV714573H, NGV715653H, NGV716004H, NGV717375H, NGV717376H, NGV719899H, NGV720557H, NGV721806H, NGV724870H, NGV725989H, NGV000692H, NGV002105H, NGV722665H, NGV723521H, NGV714004H, NGV707096H, NGV000367H, NGV652142H, NGV723805H, NGV724630H, NGV003061H, NGV004177H, NGV651212H, NGV652545H, NGV700837H, NGV701117H, NGV703024H, NGV703597H, NGV704074H, NGV704988H, NGV707290H, NGV708844H, NGV711033H, NGV713845H, NGV714167H, NGV714171H, NGV714174H, NGV718544H, NGV718910H, NGV719950H, NGV721192H, NGV721976H, NGV724640H, NGV727593H, NGV004305H, NGV650794H, NGV704058H, NGV705521H, NGV705523H, NGV650183H, NGV701994H, NGV705095H, NGV705823H, NGV706269H, NGV708323H, NGV713500H, NGV715594H, NGV721257H, NGV721757H, NGV704444H, NGV709023H, NGV714328H, NGV720096H, NGV704218H, NGV719430H, NGV704865H, NGV711725H, NGV715752H, NGV726548H, NGV651665H, NGV651672H, NGV708945H, NGV002161H, NGV702622H, NGV717572H, NGV717598H, NGV720934H, NGV724959H, NGV726765H, NGV004192H, NGV703825H, NGV001872H, NGV003487H, NGV706096H, NGV725717H, NGV002492H, NGV650911H, NGV707460H, NGV726402H, NGV717304H, NGV717305H, NGV721751H, NGV721869H, NGV721882H, NGV722097H, NGV722098H, NGV650801H, NGV651430H, NGV652226H, NGV701057H, NGV701659H, NGV701981H, NGV704738H, NGV705029H, NGV706010H, NGV707497H, NGV710967H, NGV711483H, NGV712657H, NGV713815H, NGV713970H, NGV713971H, NGV713972H, NGV716270H, NGV716272H, NGV716284H, NGV718400H, NGV719156H, NGV719271H, NGV719442H, NGV719867H, NGV720454H, NGV720971H, NGV721346H, NGV724141H, NGV724700H, NGV725400H, NGV725499H, NGV726449H, NGV726950H, NGV727218H, NGV727867H, NGV704380H, NGV706456H, NGV002072H, NGV003916H, NGV004376H, NGV650368H, NGV651019H, NGV651086H, NGV700318H, NGV700845H, NGV701413H, NGV701417H, NGV701657H, NGV702218H, NGV703202H, NGV703654H, NGV704244H, NGV704404H, NGV705080H, NGV706042H, NGV706735H, NGV706736H, NGV706789H, NGV707493H, NGV708597H, NGV709054H, NGV709687H, NGV710063H, NGV710064H, NGV711356H, NGV714333H, NGV715263H, NGV717019H, NGV717513H, NGV718876H, NGV719250H, NGV719591H, NGV720182H, NGV720194H, NGV720634H, NGV720873H, NGV720874H, NGV720967H, NGV721134H, NGV721353H, NGV721533H, NGV721684H, NGV721823H, NGV722007H, NGV722930H, NGV723825H, NGV724126H, NGV724492H, NGV724493H, NGV725393H, NGV726244H, NGV727159H, NGV727551H, NGV003007H, NGV717515H, NGV002357H, NGV004930H, NGV650964H, NGV700201H, NGV700721H, NGV700733H, NGV704954H, NGV707077H, NGV717407H, NGV726944H, NGV721958H, NGV721959H, NGV721544H, NGV727783H, NGV000637H, NGV002542H, NGV003456H, NGV701889H, NGV701890H, NGV706400H, NGV707029H, NGV707677H, NGV711076H, NGV713038H, NGV714618H, NGV715420H, NGV715853H, NGV719320H, NGV719321H, NGV719814H, NGV720491H, NGV723013H, NGV723815H, NGV723862H, NGV724889H, NGV724964H, NGV725265H, NGV002062H, NGV701071H, NGV701560H, NGV702206H, NGV702207H, NGV704057H, NGV705815H, NGV709068H, NGV709537H, NGV711384H, NGV711730H, NGV713319H, NGV713320H, NGV717412H, NGV717413H, NGV719229H, NGV723763H, NGV723837H, NGV002471H, NGV709091H, NGV724679H, NGV721971H, NGV004684H, NGV005071H, NGV005257H, NGV700849H, NGV708193H, NGV711522H, NGV715552H, NGV715998H, NGV719238H, NGV719345H, NGV720528H, NGV720794H, NGV721593H, NGV727815H, NGV702440H, NGV001329H, NGV002962H, NGV650857H, NGV700330H, NGV703909H, NGV706685H, NGV708812H, NGV704817H, NGV716857H, NGV718501H, NGV721261H, NGV723173H, NGV707359H, NGV713657H, NGV002409H, NGV005191H, NGV702611H, NGV705955H, NGV717999H, NGV719923H, NGV726015H, NGV727728H, NGV723248H, NGV723541H, NGV001397H, NGV004191H, NGV701243H, NGV708836H, NGV709695H, NGV708651H, NGV005211H, NGV702051H, NGV705926H, NGV706807H, NGV709046H, NGV716900H, NGV717245H, NGV717246H, NGV718503H, NGV718504H, NGV720860H, NGV722359H, NGV723015H, NGV723150H, NGV715309H, NGV715688H, NGV716789H, NGV726569H, NGV711553H, NGV701437H, NGV711001H, NGV721679H, NGV720708H, NGV727863H, NGV001139H, NGV705792H, NGV716561H, NGV707511H, NGV717670H, NGV719962H, NGV724038H, NGV003988H, NGV705036H, NGV705285H, NGV722696H, NGV724281H, NGV651401H, NGV002818H, NGV651102H, NGV705065H, NGV706471H, NGV717455H, NGV718228H, NGV713303H, NGV000542H, NGV003439H, NGV651391H, NGV651450H, NGV651897H, NGV652521H, NGV652523H, NGV702114H, NGV702860H, NGV703173H, NGV703742H, NGV704204H, NGV704790H, NGV705166H, NGV705510H, NGV705627H, NGV706146H, NGV707937H, NGV708587H, NGV708908H, NGV709418H, NGV711137H, NGV713346H, NGV715283H, NGV715285H, NGV715291H, NGV715307H, NGV717601H, NGV718842H, NGV718844H, NGV721040H, NGV721675H, NGV722803H, NGV004849H, NGV711004H, NGV712795H, NGV716591H, NGV716739H, NGV001241H, NGV001229H, NGV003925H, NGV004041H, NGV004228H, NGV651919H, NGV704842H, NGV704843H, NGV706251H, NGV707068H, NGV709682H, NGV710069H, NGV712677H, NGV715766H, NGV717025H, NGV717026H, NGV717028H, NGV717029H, NGV719148H, NGV719835H, NGV721935H, NGV714264H, NGV001420H, NGV706196H, NGV001513H, NGV000338H, NGV000864H, NGV001205H, NGV002128H, NGV003466H, NGV003934H, NGV004143H, NGV004144H, NGV004954H, NGV650273H, NGV650747H, NGV704556H, NGV705455H, NGV713844H, NGV717617H, NGV718751H, NGV704494H, NGV709509H, NGV712792H, NGV714198H, NGV003637H, NGV651671H, NGV702812H, NGV703007H, NGV709113H, NGV709115H, NGV717602H, NGV720911H, NGV721885H, NGV710901H, NGV005188H, NGV709926H, NGV714317H, NGV717840H, NGV720336H, NGV000342H, NGV000343H, NGV001067H, NGV002049H, NGV002713H, NGV002729H, NGV003715H, NGV003889H, NGV003891H, NGV004246H, NGV005266H, NGV005268H, NGV651323H, NGV651916H, NGV651933H, NGV651934H, NGV700551H, NGV700720H, NGV701259H, NGV702217H, NGV702327H, NGV702573H, NGV703892H, NGV703893H, NGV703979H, NGV705575H, NGV705576H, NGV718102H, NGV720586H, NGV723963H, NGV003150H, NGV004699H, NGV703576H, NGV710774H, NGV707890H, NGV716626H, NGV701880H, NGV701882H, NGV703824H, NGV708663H, NGV003416H, NGV000885H, NGV001585H, NGV002441H, NGV002472H, NGV002593H, NGV002973H, NGV003139H, NGV004706H, NGV004732H, NGV650071H, NGV651011H, NGV651213H, NGV651352H, NGV651622H, NGV700559H, NGV701512H, NGV701747H, NGV701845H, NGV702081H, NGV703622H, NGV704795H, NGV705160H, NGV705591H, NGV706301H, NGV706380H, NGV707218H, NGV707525H, NGV708098H, NGV708835H, NGV709107H, NGV709493H, NGV709494H, NGV709856H, NGV709878H, NGV709880H, NGV709939H, NGV710082H, NGV710639H, NGV710785H, NGV712201H, NGV712367H, NGV712420H, NGV712680H, NGV712874H, NGV713042H, NGV713043H, NGV713284H, NGV714430H, NGV714432H, NGV714433H, NGV715794H, NGV715799H, NGV718904H, NGV718923H, NGV719374H, NGV719561H, NGV719638H, NGV719781H, NGV720344H, NGV721141H, NGV721399H, NGV723658H, NGV724792H, NGV724793H, NGV724825H, NGV710376H, NGV726751H, NGV720643H, NGV725269H, NGV726207H, NGV721142H, NGV725343H, NGV725509H, NGV719034H, NGV001532H, NGV001845H, NGV004686H, NGV709040H, NGV713727H, NGV725376H, NGV720339H, NGV709025H, NGV711745H, NGV713328H, NGV727281H, NGV702293H, NGV707091H, NGV712651H, NGV715489H, NGV715784H, NGV718296H, NGV724484H, NGV726424H, NGV651688H, NGV000132H, NGV001782H, NGV001804H, NGV001859H, NGV003643H, NGV700267H, NGV704311H, NGV705411H, NGV706519H, NGV709522H, NGV712205H, NGV716285H, NGV717131H, NGV717132H, NGV718442H, NGV719002H, NGV720055H, NGV724628H, NGV725438H, NGV725619H, NGV726601H, NGV726651H, NGV000083H, NGV000084H, NGV001783H, NGV002349H, NGV003006H, NGV003644H, NGV650689H, NGV651219H, NGV651417H, NGV700566H, NGV702787H, NGV704004H, NGV704376H, NGV704542H, NGV704543H, NGV704545H, NGV704546H, NGV705513H, NGV705605H, NGV708233H, NGV709975H, NGV710062H, NGV710338H, NGV711501H, NGV711502H, NGV715258H, NGV715521H, NGV715779H, NGV715922H, NGV715929H, NGV718015H, NGV718057H, NGV718058H, NGV719902H, NGV720506H, NGV722491H, NGV722492H, NGV724935H, NGV725098H, NGV725234H, NGV725279H, NGV725345H, NGV725475H, NGV004855H, NGV708793H, NGV000640H, NGV001817H, NGV002007H, NGV004058H, NGV004089H, NGV650662H, NGV651455H, NGV652350H, NGV700211H, NGV700301H, NGV701909H, NGV702235H, NGV702332H, NGV705969H, NGV707441H, NGV710331H, NGV710375H, NGV712006H, NGV712011H, NGV714583H, NGV718997H, NGV721584H, NGV721694H, NGV722901H, NGV723262H, NGV726160H, NGV720386H, NGV003803H, NGV709477H, NGV002054H, NGV001235H, NGV003947H, NGV004165H, NGV004300H, NGV004526H, NGV650622H, NGV651153H, NGV700409H, NGV700429H, NGV703332H, NGV705870H, NGV707156H, NGV707157H, NGV707164H, NGV707382H, NGV707874H, NGV708004H, NGV711230H, NGV713514H, NGV713640H, NGV716352H, NGV716353H, NGV716435H, NGV716839H, NGV717574H, NGV717921H, NGV721639H, NGV723599H, NGV724052H, NGV725801H, NGV726077H, NGV726924H, NGV710116H, NGV713333H, NGV004057H, NGV703204H, NGV004641H, NGV651082H, NGV651542H, NGV652601H, NGV652603H, NGV652607H, NGV700735H, NGV701628H, NGV701629H, NGV701630H, NGV702010H, NGV702019H, NGV702021H, NGV716128H, NGV718834H, NGV720169H, NGV720171H, NGV723797H, NGV724285H, NGV725505H, NGV703951H, NGV705414H, NGV705802H, NGV712042H, NGV712356H, NGV712382H, NGV713247H, NGV713656H, NGV715481H, NGV716198H, NGV716205H, NGV717645H, NGV719369H, NGV719384H, NGV722743H, NGV722744H, NGV723393H, NGV725007H, NGV726071H, NGV726072H, NGV726073H, NGV726075H, NGV712312H, NGV003763H, NGV004500H, NGV005034H, NGV650322H, NGV651309H, NGV702716H, NGV702717H, NGV702939H, NGV704885H, NGV705094H, NGV705597H, NGV705995H, NGV706833H, NGV708866H, NGV709556H, NGV709768H, NGV711983H, NGV712000H, NGV712002H, NGV712769H, NGV712861H, NGV713286H, NGV715965H, NGV716897H, NGV717481H, NGV717745H, NGV718013H, NGV721212H, NGV724507H, NGV701708H, NGV701709H, NGV702409H, NGV702683H, NGV702853H, NGV703660H, NGV704995H, NGV704997H, NGV705221H, NGV705222H, NGV705517H, NGV717194H, NGV717256H, NGV717358H, NGV719458H, NGV719487H, NGV719801H, NGV719909H, NGV719910H, NGV721731H, NGV721746H, NGV719847H, NGV726123H, NGV701466H, NGV710332H, NGV712429H, NGV712479H, NGV713647H, NGV716342H, NGV716345H, NGV718158H, NGV718372H, NGV719945H, NGV720484H, NGV721237H, NGV723192H, NGV724656H, NGV724931H, NGV726788H, NGV727041H, NGV651365H, NGV701165H, NGV702416H, NGV702437H, NGV702476H, NGV702623H, NGV702637H, NGV703268H, NGV705137H, NGV705293H, NGV705298H, NGV705678H, NGV705679H, NGV707097H, NGV709064H, NGV709671H, NGV710141H, NGV710378H, NGV720513H, NGV727282H, NGV003454H, NGV650120H, NGV651120H, NGV703158H, NGV714265H, NGV715651H, NGV715654H, NGV719171H, NGV719260H, NGV719261H, NGV720763H, NGV722089H, NGV701772H, NGV002449H, NGV702786H, NGV707322H, NGV709809H, NGV711625H, NGV721033H, NGV003776H, NGV702107H, NGV702500H, NGV706211H, NGV706213H, NGV706231H, NGV706247H, NGV706249H, NGV707067H, NGV707851H, NGV707852H, NGV707853H, NGV708226H, NGV708227H, NGV708228H, NGV708229H, NGV708230H, NGV708389H, NGV709545H, NGV709546H, NGV709547H, NGV709548H, NGV709549H, NGV710275H, NGV710278H, NGV710764H, NGV710765H, NGV710767H, NGV710949H, NGV710950H, NGV711669H, NGV711670H, NGV711672H, NGV711674H, NGV711922H, NGV711938H, NGV711939H, NGV716559H, NGV716978H, NGV716979H, NGV716981H, NGV716982H, NGV716983H, NGV716984H, NGV717024H, NGV717027H, NGV717590H, NGV717963H, NGV717964H, NGV718303H, NGV718709H, NGV719175H, NGV719457H, NGV719460H, NGV719634H, NGV719862H, NGV719952H, NGV719983H, NGV720109H, NGV720530H, NGV720580H, NGV720581H, NGV720909H, NGV721348H, NGV721407H, NGV721895H, NGV721896H, NGV722495H, NGV722496H, NGV722853H, NGV723488H, NGV723515H, NGV723516H, NGV723534H, NGV724624H, NGV724625H, NGV724626H, NGV725042H, NGV725685H, NGV725804H, NGV725930H, NGV725932H, NGV726164H, NGV726372H, NGV726465H, NGV726742H, NGV727656H, NGV727657H, NGV000253H, NGV003253H, NGV652351H, NGV702482H, NGV703345H, NGV708595H, NGV711052H, NGV000568H, NGV002978H, NGV003674H, NGV705592H, NGV718585H, NGV712012H, NGV717333H, NGV719032H, NGV716936H, NGV002517H, NGV002621H, NGV003850H, NGV005154H, NGV005242H, NGV700464H, NGV701050H, NGV701917H, NGV702620H, NGV702859H, NGV704346H, NGV708985H, NGV709478H, NGV710955H, NGV711436H, NGV712505H, NGV712798H, NGV713536H, NGV713644H, NGV714455H, NGV715421H, NGV717360H, NGV717382H, NGV717819H, NGV719686H, NGV719687H, NGV720106H, NGV720848H, NGV723524H, NGV724009H, NGV724817H, NGV004111H, NGV652532H, NGV700205H, NGV701444H, NGV701449H, NGV705711H, NGV705713H, NGV705714H, NGV706254H, NGV712803H, NGV714007H, NGV714465H, NGV716217H, NGV716555H, NGV717265H, NGV717706H, NGV717711H, NGV718502H, NGV719504H, NGV719859H, NGV719938H, NGV719941H, NGV720346H, NGV721182H, NGV721204H, NGV721460H, NGV721979H, NGV722760H, NGV722762H, NGV722763H, NGV723636H, NGV723679H, NGV724989H, NGV725158H, NGV725159H, NGV725977H, NGV726097H, NGV726098H, NGV726099H, NGV708901H, NGV710212H, NGV710980H, NGV712363H, NGV716682H, NGV717649H, NGV719376H, NGV719695H, NGV722502H, NGV722503H, NGV726410H, NGV000065H, NGV003467H, NGV004286H, NGV651158H, NGV703419H, NGV704617H, NGV711999H, NGV718319H, NGV721696H, NGV723224H, NGV711002H, NGV004694H, NGV720988H, NGV723143H, NGV725582H, NGV726511H, NGV000239H, NGV000470H, NGV000733H, NGV000986H, NGV001133H, NGV001136H, NGV001699H, NGV002413H, NGV002481H, NGV002483H, NGV002739H, NGV003163H, NGV003164H, NGV003165H, NGV003442H, NGV650186H, NGV650495H, NGV650612H, NGV651189H, NGV652083H, NGV652084H, NGV652085H, NGV652384H, NGV700390H, NGV700493H, NGV700950H, NGV701914H, NGV702116H, NGV702306H, NGV703029H, NGV703159H, NGV703261H, NGV703769H, NGV704591H, NGV704597H, NGV704872H, NGV704873H, NGV704878H, NGV704879H, NGV705446H, NGV707100H, NGV707199H, NGV707398H, NGV708559H, NGV708560H, NGV708561H, NGV708563H, NGV708771H, NGV711201H, NGV711243H, NGV711391H, NGV711474H, NGV711958H, NGV711959H, NGV711960H, NGV711962H, NGV711963H, NGV711964H, NGV711965H, NGV712704H, NGV713375H, NGV713903H, NGV715423H, NGV715848H, NGV717066H, NGV717112H, NGV718072H, NGV718097H, NGV718317H, NGV718577H, NGV718728H, NGV718818H, NGV718857H, NGV719164H, NGV719264H, NGV719332H, NGV719723H, NGV719963H, NGV719965H, NGV719976H, NGV720599H, NGV720748H, NGV720749H, NGV720843H, NGV720844H, NGV721072H, NGV721143H, NGV721202H, NGV721771H, NGV721933H, NGV722658H, NGV723017H, NGV723188H, NGV723222H, NGV723601H, NGV725819H, NGV726093H, NGV726195H, NGV726312H, NGV726590H, NGV726639H, NGV727225H, NGV727729H, NGV001741H, NGV726215H, NGV001702H, NGV718662H, NGV725738H, NGV002443H, NGV004548H, NGV705183H, NGV705890H, NGV705957H, NGV725090H, NGV716289H, NGV714579H, NGV717614H, NGV702678H, NGV703980H, NGV704599H, NGV000091H, NGV650158H, NGV650656H, NGV651755H, NGV651758H, NGV651759H, NGV703353H, NGV705667H, NGV708581H, NGV710133H, NGV710328H, NGV712033H, NGV715722H, NGV719520H, NGV719611H, NGV720230H, NGV721042H, NGV725040H, NGV726661H, NGV651495H, NGV705091H, NGV710030H, NGV712688H, NGV715627H, NGV726377H, NGV000824H, NGV002166H, NGV650649H, NGV652360H, NGV652596H, NGV706812H, NGV707210H, NGV707533H, NGV711086H, NGV712782H, NGV717334H, NGV717353H, NGV717354H, NGV720632H, NGV720682H, NGV720684H, NGV721117H, NGV702147H, NGV715261H, NGV716702H, NGV722897H, NGV703575H, NGV713912H, NGV713916H, NGV719726H, NGV720030H, NGV726371H, NGV720512H, NGV727049H, NGV701990H, NGV705208H, NGV000890H, NGV000902H, NGV701795H, NGV703598H, NGV717448H, NGV726276H, NGV726999H, NGV002088H, NGV720473H, NGV002121H, NGV702596H, NGV721844H, NGV722748H, NGV723611H, NGV710074H, NGV707133H, NGV710104H, NGV725984H, NGV000480H, NGV001104H, NGV001259H, NGV001778H, NGV002133H, NGV003725H, NGV704877H, NGV705976H, NGV709741H, NGV710330H, NGV712401H, NGV718659H, NGV720927H, NGV721975H, NGV724946H, NGV724979H, NGV000129H, NGV001231H, NGV001233H, NGV002418H, NGV002419H, NGV003229H, NGV003230H, NGV003231H, NGV651811H, NGV706295H, NGV711969H, NGV711970H, NGV713475H, NGV713608H, NGV715747H, NGV717393H, NGV720973H, NGV700795H, NGV715734H, NGV721557H, NGV704192H, NGV723178H, NGV000845H, NGV002063H, NGV003642H, NGV004354H, NGV004964H, NGV651043H, NGV700143H, NGV701264H, NGV703359H, NGV703366H, NGV704401H, NGV705626H, NGV705829H, NGV706831H, NGV706832H, NGV707066H, NGV707396H, NGV710379H, NGV719545H, NGV720510H, NGV723572H, NGV725941H, NGV706425H, NGV717129H, NGV000128H, NGV000133H, NGV650676H, NGV702628H, NGV706906H, NGV721351H, NGV722441H, NGV725332H, NGV727735H, NGV709294H, NGV717080H, NGV726370H, NGV726783H, NGV708899H, NGV711616H, NGV723296H, NGV727115H, NGV650680H, NGV713297H, NGV727631H, NGV000797H, NGV001257H, NGV001969H, NGV702920H, NGV711514H, NGV716349H, NGV717773H, NGV717855H, NGV723583H, NGV725167H, NGV700478H, NGV707887H, NGV650965H, NGV711433H, NGV706726H, NGV714173H, NGV721076H, NGV711221H, NGV719167H, NGV723777H, NGV711927H, NGV712047H, NGV712294H, NGV724586H, NGV706738H, NGV002139H, NGV004167H, NGV711177H, NGV715543H, NGV717897H, NGV718010H, NGV719697H, NGV719767H, NGV719780H, NGV720855H, NGV720856H, NGV726233H, NGV651264H, NGV651869H, NGV705280H, NGV708856H, NGV713893H, NGV717982H, NGV721121H, NGV721755H, NGV722871H, NGV723940H, NGV726731H, NGV727267H, NGV651248H, NGV703013H, NGV707491H, NGV708997H, NGV715845H, NGV724988H, NGV003496H, NGV004857H, NGV702162H, NGV704463H, NGV711687H, NGV651686H, NGV704914H, NGV704915H, NGV707594H, NGV716968H, NGV718255H, NGV724212H, NGV724452H, NGV720354H, NGV720903H, NGV721321H, NGV725390H, NGV001615H, NGV003515H, NGV004832H, NGV004833H, NGV651074H, NGV651087H, NGV700531H, NGV703201H, NGV704403H, NGV704745H, NGV705480H, NGV708847H, NGV717476H, NGV717817H, NGV718036H, NGV718663H, NGV719502H, NGV721059H, NGV002081H, NGV702743H, NGV712644H, NGV651431H, NGV712039H, NGV712338H, NGV713885H, NGV002126H, NGV002127H, NGV704702H, NGV712335H, NGV721328H, NGV727790H, NGV712725H, NGV002428H, NGV005185H, NGV004175H, NGV004226H, NGV651445H, NGV702581H, NGV707379H, NGV708614H, NGV708632H, NGV712811H, NGV717142H, NGV717885H, NGV717886H, NGV718678H, NGV719276H, NGV719567H, NGV720996H, NGV721883H, NGV721884H, NGV721953H, NGV724619H, NGV726225H, NGV726431H, NGV702333H, NGV710072H, NGV706996H, NGV708828H, NGV713595H, NGV716618H, NGV723660H, NGV727667H, NGV704423H, NGV711431H, NGV005280H, NGV708982H, NGV719126H, NGV719150H, NGV723313H, NGV724621H, NGV002948H, NGV706293H, NGV707469H, NGV710947H, NGV713330H, NGV652027H, NGV702792H, NGV720888H, NGV650938H, NGV700953H, NGV706270H, NGV001938H, NGV651394H, NGV711865H, NGV716540H, NGV719653H, NGV719740H, NGV721890H, NGV723856H, NGV726471H, NGV726567H, NGV727638H, NGV000558H, NGV001621H, NGV001721H, NGV650054H, NGV650795H, NGV651880H, NGV652399H, NGV652541H, NGV700138H, NGV702734H, NGV704983H, NGV705843H, NGV705996H, NGV706760H, NGV707291H, NGV707514H, NGV709955H, NGV716827H, NGV713505H, NGV651059H, NGV723022H, NGV001984H, NGV700485H, NGV706279H, NGV706280H, NGV726439H, NGV726440H, NGV727568H, NGV722562H, NGV712293H, NGV722851H, NGV003401H, NGV004380H, NGV650390H, NGV650393H, NGV652043H, NGV700491H, NGV701022H, NGV701422H, NGV701971H, NGV704278H, NGV707339H, NGV707362H, NGV708922H, NGV710052H, NGV710389H, NGV711573H, NGV711744H, NGV714034H, NGV715855H, NGV719318H, NGV720450H, NGV722054H, NGV725089H, NGV725188H, NGV727647H, NGV727648H, NGV000409H, NGV650381H, NGV718182H, NGV651428H, NGV702636H, NGV704449H, NGV704964H, NGV706997H, NGV709102H, NGV709118H, NGV709842H, NGV710058H, NGV711079H, NGV711547H, NGV712705H, NGV713045H, NGV715501H, NGV718774H, NGV719845H, NGV720483H, NGV724094H, NGV724620H, NGV724807H, NGV725191H, NGV727301H, NGV000703H, NGV651579H, NGV704105H, NGV710215H, NGV718358H, NGV721795H, NGV727008H, NGV723268H, NGV000064H, NGV000460H, NGV707222H, NGV712424H, NGV000213H, NGV000214H, NGV000216H, NGV000217H, NGV000219H, NGV000430H, NGV000431H, NGV000432H, NGV000439H, NGV000441H, NGV000442H, NGV000576H, NGV000579H, NGV000581H, NGV000582H, NGV000583H, NGV000584H, NGV000585H, NGV000586H, NGV001107H, NGV001203H, NGV001204H, NGV001351H, NGV001352H, NGV001353H, NGV001356H, NGV001557H, NGV001558H, NGV001559H, NGV001687H, NGV001688H, NGV001807H, NGV001992H, NGV001993H, NGV001994H, NGV002135H, NGV002141H, NGV002142H, NGV002143H, NGV002145H, NGV002906H, NGV002907H, NGV002909H, NGV002918H, NGV002919H, NGV002920H, NGV002921H, NGV003171H, NGV003172H, NGV003173H, NGV003174H, NGV003769H, NGV003771H, NGV003772H, NGV004067H, NGV004068H, NGV004069H, NGV004072H, NGV004074H, NGV004075H, NGV004076H, NGV004385H, NGV004386H, NGV004387H, NGV004388H, NGV004790H, NGV005162H, NGV005164H, NGV005165H, NGV005166H, NGV650435H, NGV650437H, NGV650439H, NGV650441H, NGV650786H, NGV650833H, NGV650885H, NGV650889H, NGV650890H, NGV651029H, NGV651165H, NGV651166H, NGV651167H, NGV651168H, NGV651180H, NGV651270H, NGV651283H, NGV651295H, NGV651296H, NGV651300H, NGV651454H, NGV651460H, NGV651465H, NGV651469H, NGV651471H, NGV651543H, NGV651554H, NGV651557H, NGV651559H, NGV651560H, NGV652276H, NGV652281H, NGV652284H, NGV652285H, NGV652286H, NGV652287H, NGV652289H, NGV652290H, NGV652291H, NGV652292H, NGV700718H, NGV700719H, NGV701062H, NGV701063H, NGV701064H, NGV701065H, NGV701066H, NGV701067H, NGV701930H, NGV701933H, NGV701934H, NGV701936H, NGV701941H, NGV701943H, NGV701944H, NGV702128H, NGV702129H, NGV702130H, NGV702132H, NGV702133H, NGV702134H, NGV702135H, NGV702263H, NGV702518H, NGV702531H, NGV702551H, NGV702862H, NGV703002H, NGV703061H, NGV703063H, NGV703078H, NGV703231H, NGV703232H, NGV703233H, NGV703340H, NGV703590H, NGV703600H, NGV703601H, NGV703603H, NGV703604H, NGV703605H, NGV703649H, NGV703683H, NGV703684H, NGV703685H, NGV703924H, NGV703925H, NGV703927H, NGV704268H, NGV704269H, NGV704270H, NGV704271H, NGV704272H, NGV704273H, NGV704274H, NGV704275H, NGV704276H, NGV704577H, NGV704579H, NGV704580H, NGV704581H, NGV704582H, NGV704860H, NGV704861H, NGV704862H, NGV704863H, NGV705335H, NGV705336H, NGV705337H, NGV705393H, NGV705475H, NGV705476H, NGV705477H, NGV705478H, NGV705479H, NGV705703H, NGV705704H, NGV705705H, NGV705706H, NGV705707H, NGV705773H, NGV705831H, NGV705832H, NGV705833H, NGV705834H, NGV705975H, NGV706001H, NGV706102H, NGV706103H, NGV706105H, NGV706107H, NGV706116H, NGV706172H, NGV706182H, NGV706184H, NGV706186H, NGV706188H, NGV706190H, NGV706310H, NGV706325H, NGV706338H, NGV706339H, NGV706340H, NGV706623H, NGV706702H, NGV706703H, NGV706704H, NGV706705H, NGV706706H, NGV706733H, NGV706818H, NGV706820H, NGV706913H, NGV706914H, NGV706919H, NGV706920H, NGV706922H, NGV706923H, NGV706924H, NGV707225H, NGV707407H, NGV707549H, NGV707551H, NGV707552H, NGV707553H, NGV708550H, NGV708566H, NGV708815H, NGV708848H, NGV708849H, NGV708850H, NGV708851H, NGV708852H, NGV708853H, NGV708854H, NGV709013H, NGV709014H, NGV709015H, NGV709017H, NGV709019H, NGV709169H, NGV709284H, NGV709395H, NGV709598H, NGV709599H, NGV709600H, NGV709602H, NGV709605H, NGV709675H, NGV709676H, NGV709677H, NGV709678H, NGV709751H, NGV709752H, NGV709754H, NGV709755H, NGV709946H, NGV709947H, NGV709948H, NGV709949H, NGV709950H, NGV710046H, NGV710047H, NGV710048H, NGV710049H, NGV710244H, NGV710245H, NGV710246H, NGV710247H, NGV710248H, NGV710249H, NGV710256H, NGV710257H, NGV710258H, NGV710259H, NGV710260H, NGV710261H, NGV711053H, NGV711054H, NGV711055H, NGV711056H, NGV711120H, NGV711121H, NGV711334H, NGV711735H, NGV711736H, NGV711738H, NGV711739H, NGV711740H, NGV711741H, NGV711974H, NGV711975H, NGV711976H, NGV711977H, NGV712313H, NGV712315H, NGV712529H, NGV712554H, NGV712581H, NGV712582H, NGV712583H, NGV712620H, NGV712653H, NGV712655H, NGV712656H, NGV712833H, NGV712838H, NGV712841H, NGV712850H, NGV712853H, NGV713039H, NGV713897H, NGV713905H, NGV713906H, NGV713908H, NGV715424H, NGV715644H, NGV715682H, NGV715706H, NGV715723H, NGV715737H, NGV715865H, NGV715869H, NGV715891H, NGV715895H, NGV715896H, NGV715903H, NGV715905H, NGV715914H, NGV715917H, NGV715918H, NGV715920H, NGV715926H, NGV716564H, NGV716805H, NGV716814H, NGV716898H, NGV717067H, NGV717070H, NGV717200H, NGV717241H, NGV717260H, NGV717344H, NGV717654H, NGV717655H, NGV717824H, NGV717825H, NGV717826H, NGV718020H, NGV718022H, NGV718023H, NGV718077H, NGV718137H, NGV718138H, NGV718155H, NGV718175H, NGV718179H, NGV718180H, NGV718237H, NGV718242H, NGV718247H, NGV718256H, NGV718341H, NGV718343H, NGV718373H, NGV718374H, NGV718389H, NGV718398H, NGV718528H, NGV718530H, NGV718602H, NGV718603H, NGV718604H, NGV718605H, NGV718729H, NGV718741H, NGV718752H, NGV718760H, NGV718782H, NGV718932H, NGV718933H, NGV718949H, NGV718957H, NGV718959H, NGV719131H, NGV719684H, NGV719685H, NGV719893H, NGV720090H, NGV720254H, NGV720255H, NGV720256H, NGV720257H, NGV720258H, NGV720356H, NGV720429H, NGV720458H, NGV720481H, NGV720640H, NGV720641H, NGV720642H, NGV720657H, NGV720658H, NGV720659H, NGV720660H, NGV720665H, NGV720666H, NGV720667H, NGV720668H, NGV720669H, NGV720670H, NGV721266H, NGV721267H, NGV721268H, NGV721269H, NGV721621H, NGV721622H, NGV721623H, NGV721624H, NGV721625H, NGV721626H, NGV721794H, NGV721831H, NGV721871H, NGV721872H, NGV721873H, NGV721874H, NGV722045H, NGV722046H, NGV722047H, NGV722430H, NGV722659H, NGV722958H, NGV722959H, NGV722960H, NGV722961H, NGV722962H, NGV722963H, NGV722964H, NGV722965H, NGV723202H, NGV723203H, NGV723204H, NGV723205H, NGV723206H, NGV723207H, NGV723417H, NGV723418H, NGV723421H, NGV723422H, NGV723423H, NGV723424H, NGV723425H, NGV723434H, NGV723438H, NGV724872H, NGV724873H, NGV724874H, NGV724875H, NGV724877H, NGV724879H, NGV724880H, NGV724881H, NGV724882H, NGV724883H, NGV724884H, NGV724885H, NGV724886H, NGV725021H, NGV725025H, NGV725028H, NGV725032H, NGV725034H, NGV725036H, NGV725043H, NGV725044H, NGV725048H, NGV726454H, NGV726456H, NGV726457H, NGV726459H, NGV726718H, NGV726719H, NGV726720H, NGV726721H, NGV726724H, NGV726725H, NGV726726H, NGV726727H, NGV727124H, NGV727126H, NGV727131H, NGV727133H, NGV727135H, NGV727138H, NGV727528H, NGV727529H, NGV727530H, NGV727531H, NGV727532H, NGV727533H, NGV003461H, NGV652056H, NGV707384H, NGV711059H, NGV725164H, NGV001404H, NGV002974H, NGV004689H, NGV713682H, NGV722663H, NGV723558H, NGV004080H, NGV715916H, NGV000500H, NGV001431H, NGV713288H, NGV714055H, NGV717259H, NGV717932H, NGV722896H, NGV723802H, NGV726267H, NGV001757H, NGV001774H, NGV002523H, NGV003123H, NGV003618H, NGV003713H, NGV004326H, NGV004433H, NGV004701H, NGV004710H, NGV650365H, NGV651233H, NGV651316H, NGV700306H, NGV701109H, NGV701382H, NGV702196H, NGV702554H, NGV702555H, NGV703152H, NGV703626H, NGV704688H, NGV704689H, NGV705694H, NGV705695H, NGV705868H, NGV706223H, NGV706418H, NGV707200H, NGV707299H, NGV707519H, NGV708005H, NGV709101H, NGV709148H, NGV709306H, NGV709402H, NGV710944H, NGV711064H, NGV711695H, NGV712873H, NGV713349H, NGV715268H, NGV715814H, NGV715815H, NGV715821H, NGV715822H, NGV716244H, NGV717180H, NGV717901H, NGV718353H, NGV718651H, NGV718734H, NGV720078H, NGV720079H, NGV720080H, NGV720866H, NGV720867H, NGV722087H, NGV722088H, NGV722398H, NGV722560H, NGV723523H, NGV724203H, NGV724268H, NGV725312H, NGV726666H, NGV727112H, NGV003126H, NGV650352H, NGV650641H, NGV651174H, NGV701518H, NGV702157H, NGV702158H, NGV702316H, NGV704015H, NGV705869H, NGV705899H, NGV713253H, NGV716513H, NGV721233H, NGV722444H, NGV723703H, NGV723922H, NGV723924H, NGV706734H, NGV000896H, NGV716850H, NGV001276H, NGV715523H, NGV002824H, NGV002880H, NGV003348H, NGV003862H, NGV003863H, NGV004744H, NGV004879H, NGV005210H, NGV650220H, NGV700798H, NGV701047H, NGV711913H, NGV716957H, NGV704082H, NGV712623H, NGV723181H, NGV650205H, NGV702677H, NGV706235H, NGV707517H, NGV712867H, NGV724280H, NGV003374H, NGV004940H, NGV712403H, NGV703812H, NGV723250H, NGV705346H, NGV726877H, NGV716459H, NGV716460H, NGV716791H, NGV717362H, NGV718186H, NGV709662H, NGV712511H, NGV716005H, NGV651099H, NGV704951H, NGV710321H, NGV718567H, NGV721054H, NGV724894H, NGV723238H, NGV707448H, NGV721262H, NGV723133H, NGV726638H, NGV718541H, NGV718780H, NGV721053H, NGV725812H, NGV000262H, NGV001046H, NGV001400H, NGV702663H, NGV705103H, NGV725440H, NGV708652H, NGV708843H, NGV710301H, NGV710911H, NGV711829H, NGV713688H, NGV714434H, NGV716721H, NGV716734H, NGV717410H, NGV718914H, NGV722898H, NGV724828H, NGV650900H, NGV650734H, NGV700382H, NGV000842H, NGV000985H, NGV650196H, NGV650200H, NGV650598H, NGV651282H, NGV651778H, NGV651861H, NGV703276H, NGV706642H, NGV709045H, NGV709723H, NGV709780H, NGV003883H, NGV004800H, NGV703257H, NGV703931H, NGV704883H, NGV705417H, NGV705449H, NGV705519H, NGV707271H, NGV709303H, NGV709506H, NGV709794H, NGV711426H, NGV711731H, NGV712330H, NGV712728H, NGV717229H, NGV719383H, NGV720167H, NGV720514H, NGV720516H, NGV721947H, NGV723899H, NGV724491H, NGV725598H, NGV725859H, NGV726728H, NGV726989H, NGV001520H, NGV651866H, NGV652088H, NGV701638H, NGV701710H, NGV704641H, NGV707264H, NGV707265H, NGV710996H, NGV000549H, NGV001795H, NGV004505H, NGV004520H, NGV651881H, NGV651884H, NGV704305H, NGV704561H, NGV714013H, NGV715786H, NGV000135H, NGV001510H, NGV001796H, NGV002359H, NGV005239H, NGV005241H, NGV652102H, NGV700199H, NGV700244H, NGV700347H, NGV702730H, NGV703491H, NGV703791H, NGV704071H, NGV704991H, NGV705574H, NGV706292H, NGV706719H, NGV706721H, NGV707140H, NGV707358H, NGV707651H, NGV707962H, NGV709208H, NGV709433H, NGV709690H, NGV709744H, NGV712133H, NGV712375H, NGV712504H, NGV714580H, NGV715454H, NGV715464H, NGV715465H, NGV715528H, NGV715641H, NGV718110H, NGV718828H, NGV718829H, NGV719314H, NGV719533H, NGV719535H, NGV719671H, NGV719757H, NGV720164H, NGV721046H, NGV721093H, NGV721094H, NGV721717H, NGV721718H, NGV722043H, NGV722919H, NGV722923H, NGV723323H, NGV725004H, NGV725201H, NGV725291H, NGV725716H, NGV726475H, NGV726628H, NGV726936H, NGV727334H, NGV727335H, NGV727784H, NGV000149H, NGV000265H, NGV000266H, NGV000269H, NGV000270H, NGV000336H, NGV000569H, NGV000793H, NGV000892H, NGV000912H, NGV001206H, NGV001287H, NGV001326H, NGV001350H, NGV001421H, NGV001433H, NGV001434H, NGV001576H, NGV001599H, NGV001683H, NGV001855H, NGV002185H, NGV002187H, NGV002404H, NGV002408H, NGV002497H, NGV002510H, NGV002771H, NGV002820H, NGV002913H, NGV002914H, NGV002934H, NGV002984H, NGV003545H, NGV003546H, NGV003630H, NGV003660H, NGV003661H, NGV003708H, NGV003762H, NGV003811H, NGV003906H, NGV003935H, NGV004044H, NGV004170H, NGV004188H, NGV004239H, NGV004421H, NGV004422H, NGV004501H, NGV004792H, NGV004811H, NGV004820H, NGV004835H, NGV004901H, NGV004983H, NGV004984H, NGV004990H, NGV005111H, NGV005142H, NGV005149H, NGV005159H, NGV005179H, NGV005205H, NGV005214H, NGV005272H, NGV650309H, NGV650315H, NGV650384H, NGV650407H, NGV650433H, NGV650434H, NGV650849H, NGV651186H, NGV651363H, NGV651375H, NGV651432H, NGV651481H, NGV651654H, NGV651706H, NGV651885H, NGV652033H, NGV652138H, NGV652477H, NGV700272H, NGV700337H, NGV700359H, NGV700494H, NGV700544H, NGV700796H, NGV700803H, NGV700855H, NGV700858H, NGV700973H, NGV701043H, NGV701076H, NGV701077H, NGV701315H, NGV701586H, NGV701598H, NGV701715H, NGV701729H, NGV701869H, NGV701922H, NGV701923H, NGV701940H, NGV702092H, NGV702118H, NGV702315H, NGV702408H, NGV702449H, NGV702489H, NGV702544H, NGV702600H, NGV702640H, NGV702682H, NGV702692H, NGV702704H, NGV702735H, NGV702736H, NGV702752H, NGV702993H, NGV703185H, NGV703213H, NGV703225H, NGV703227H, NGV703272H, NGV703439H, NGV703442H, NGV703444H, NGV703537H, NGV703554H, NGV703571H, NGV703583H, NGV703593H, NGV703916H, NGV704002H, NGV704180H, NGV704234H, NGV704301H, NGV704458H, NGV704504H, NGV704585H, NGV704622H, NGV704639H, NGV704687H, NGV704712H, NGV704739H, NGV704781H, NGV704841H, NGV704870H, NGV705098H, NGV705104H, NGV705110H, NGV705111H, NGV705112H, NGV705114H, NGV705132H, NGV705180H, NGV705240H, NGV705241H, NGV705269H, NGV705321H, NGV705402H, NGV705403H, NGV705410H, NGV705419H, NGV705423H, NGV705516H, NGV705530H, NGV705660H, NGV705776H, NGV705777H, NGV705778H, NGV705783H, NGV705880H, NGV705881H, NGV706057H, NGV706087H, NGV706173H, NGV706174H, NGV706230H, NGV706259H, NGV706283H, NGV706414H, NGV706454H, NGV706461H, NGV706615H, NGV706629H, NGV706634H, NGV706663H, NGV706697H, NGV706716H, NGV706724H, NGV706834H, NGV707084H, NGV707105H, NGV707175H, NGV707182H, NGV707310H, NGV707341H, NGV707352H, NGV707353H, NGV707354H, NGV707369H, NGV707465H, NGV707466H, NGV707953H, NGV707963H, NGV707967H, NGV707968H, NGV708091H, NGV708648H, NGV708800H, NGV708801H, NGV708802H, NGV708803H, NGV708816H, NGV708911H, NGV708913H, NGV708914H, NGV708915H, NGV708975H, NGV708999H, NGV709006H, NGV709236H, NGV709489H, NGV709508H, NGV709515H, NGV709584H, NGV709587H, NGV709595H, NGV709597H, NGV709643H, NGV709650H, NGV709700H, NGV709701H, NGV709733H, NGV709757H, NGV709758H, NGV709770H, NGV709773H, NGV709865H, NGV709911H, NGV709935H, NGV709952H, NGV709957H, NGV710007H, NGV710008H, NGV710009H, NGV710024H, NGV710042H, NGV710115H, NGV710195H, NGV710199H, NGV710264H, NGV710273H, NGV710315H, NGV710364H, NGV710898H, NGV710910H, NGV710914H, NGV710919H, NGV710925H, NGV710997H, NGV711000H, NGV711085H, NGV711235H, NGV711237H, NGV711260H, NGV711289H, NGV711306H, NGV711309H, NGV711314H, NGV711346H, NGV711376H, NGV711572H, NGV711627H, NGV711655H, NGV711774H, NGV711806H, NGV711859H, NGV711868H, NGV711875H, NGV711898H, NGV711899H, NGV711951H, NGV712046H, NGV712155H, NGV712174H, NGV712182H, NGV712327H, NGV712437H, NGV712466H, NGV712487H, NGV712497H, NGV712510H, NGV712516H, NGV712528H, NGV726835H, NGV000785H, NGV002945H, NGV003388H, NGV004622H, NGV651093H, NGV701704H, NGV705421H, NGV706234H, NGV706638H, NGV707124H, NGV002367H, NGV002572H, NGV004158H, NGV004506H, NGV651381H, NGV705999H, NGV707590H, NGV711007H, NGV711653H, NGV719534H, NGV721683H, NGV721886H, NGV723005H, NGV719136H, NGV003246H, NGV004752H, NGV005184H, NGV702119H, NGV702609H, NGV702795H, NGV703557H, NGV704350H, NGV704557H, NGV707300H, NGV709966H, NGV716840H, NGV717195H, NGV719898H, NGV000786H, NGV002061H, NGV003721H, NGV702797H, NGV703921H, NGV709931H, NGV712642H, NGV700236H, NGV700976H, NGV704365H, NGV709829H, NGV709830H, NGV710077H, NGV710086H, NGV712441H, NGV713334H, NGV702199H, NGV719851H, NGV727292H, NGV000665H, NGV701257H, NGV711190H, NGV713577H, NGV718440H, NGV711446H, NGV714575H, NGV720054H, NGV707395H, NGV711814H, NGV713655H, NGV715251H, NGV715526H, NGV717587H, NGV719083H, NGV704282H, NGV712222H, NGV000478H, NGV000673H, NGV001039H, NGV001314H, NGV001315H, NGV002803H, NGV002804H, NGV002805H, NGV002807H, NGV002808H, NGV002809H, NGV004082H, NGV004120H, NGV004122H, NGV004125H, NGV004150H, NGV004151H, NGV005020H, NGV651623H, NGV651782H, NGV700996H, NGV700997H, NGV700998H, NGV701000H, NGV701001H, NGV712018H, NGV712019H, NGV713512H, NGV722598H, NGV722601H, NGV004841H, NGV702298H, NGV705121H, NGV707357H, NGV719556H, NGV719557H, NGV721159H, NGV723355H, NGV723850H, NGV725443H, NGV711987H, NGV711988H, NGV711989H, NGV711990H, NGV725194H, NGV725195H, NGV000101H, NGV000104H, NGV001084H, NGV001102H, NGV002032H, NGV002676H, NGV003245H, NGV003641H, NGV003982H, NGV004413H, NGV004875H, NGV650087H, NGV650141H, NGV651057H, NGV651523H, NGV651832H, NGV652368H, NGV701633H, NGV701843H, NGV702331H, NGV703636H, NGV703841H, NGV703930H, NGV703962H, NGV704012H, NGV705866H, NGV706386H, NGV706478H, NGV706980H, NGV706990H, NGV707184H, NGV708918H, NGV709847H, NGV709868H, NGV710188H, NGV711032H, NGV711152H, NGV711984H, NGV712517H, NGV714420H, NGV714425H, NGV715317H, NGV715773H, NGV716987H, NGV718658H, NGV718661H, NGV722942H, NGV724947H, NGV711130H, NGV712344H, NGV720533H, NGV722729H, NGV722774H, NGV723245H, NGV726417H, NGV726757H, NGV727644H, NGV003679H, NGV712857H, NGV713962H, NGV704918H, NGV710408H, NGV714307H, NGV715730H, NGV717689H, NGV720357H, NGV720724H, NGV721827H, NGV723869H, NGV651653H, NGV002096H, NGV002097H, NGV002120H, NGV003807H, NGV003828H, NGV003830H, NGV004552H, NGV004558H, NGV004559H, NGV004560H, NGV652098H, NGV700610H, NGV700611H, NGV700934H, NGV702097H, NGV702098H, NGV702318H, NGV702321H, NGV702399H, NGV705588H, NGV706894H, NGV707325H, NGV708765H, NGV708767H, NGV709450H, NGV709565H, NGV710005H, NGV712073H, NGV712977H, NGV716009H, NGV716102H, NGV718917H, NGV718920H, NGV719324H, NGV719699H, NGV721253H, NGV725502H, NGV000166H, NGV000646H, NGV001970H, NGV004527H, NGV701887H, NGV703973H, NGV705809H, NGV000050H, NGV650623H, NGV651297H, NGV651625H, NGV702211H, NGV703068H, NGV703963H, NGV704922H, NGV705636H, NGV706848H, NGV710341H, NGV711065H, NGV711888H, NGV717350H, NGV719946H, NGV723023H, NGV723050H, NGV723816H, NGV724467H, NGV725429H, NGV726279H, NGV727289H, NGV726597H, NGV002436H, NGV004728H, NGV004729H, NGV702656H, NGV703854H, NGV704832H, NGV705384H, NGV707016H, NGV709528H, NGV709622H, NGV709681H, NGV710561H, NGV712066H, NGV712826H, NGV717486H, NGV718438H, NGV718645H, NGV718655H, NGV719298H, NGV719742H, NGV720645H, NGV720965H, NGV721322H, NGV721326H, NGV721629H, NGV722878H, NGV723176H, NGV726468H, NGV703410H, NGV706210H, NGV708529H, NGV717314H, NGV724296H, NGV725162H, NGV726355H, NGV723357H, NGV726608H, NGV001302H, NGV704924H, NGV708980H, NGV651595H, NGV703655H, NGV708156H, NGV708928H, NGV712397H, NGV706510H, NGV004715H, NGV707579H, NGV707634H, NGV709787H, NGV721209H, NGV002047H, NGV003363H, NGV001044H, NGV708863H, NGV709981H, NGV715264H, NGV001506H, NGV001531H, NGV003468H, NGV652594H, NGV700126H, NGV702197H, NGV702996H, NGV709867H, NGV709907H, NGV713517H, NGV717296H, NGV720553H, NGV723395H, NGV723513H, NGV726743H, NGV001018H, NGV002533H, NGV650674H, NGV700556H, NGV702585H, NGV000230H, NGV712593H, NGV719705H, NGV722655H, NGV724124H, NGV726469H, NGV727067H, NGV727636H, NGV727637H, NGV707556H, NGV707557H, NGV712537H, NGV726641H, NGV000861H, NGV003088H, NGV003572H, NGV003829H, NGV651640H, NGV651690H, NGV700612H, NGV701300H, NGV701303H, NGV702319H, NGV702320H, NGV702398H, NGV702513H, NGV702619H, NGV703858H, NGV705640H, NGV705641H, NGV706889H, NGV707563H, NGV708656H, NGV709049H, NGV709567H, NGV710006H, NGV711854H, NGV712540H, NGV716100H, NGV718726H, NGV719739H, NGV719765H, NGV721590H, NGV722498H, NGV722866H, NGV723926H, NGV725967H, NGV726280H, NGV727555H, NGV702309H, NGV715718H, NGV723220H, NGV003690H, NGV003853H, NGV650749H, NGV705016H, NGV706991H, NGV724961H, NGV724962H, NGV702400H, NGV702924H, NGV001884H, NGV702059H, NGV717098H, NGV000047H, NGV002238H, NGV003302H, NGV004926H, NGV650412H, NGV651170H, NGV651541H, NGV700248H, NGV700392H, NGV702728H, NGV702760H, NGV702774H, NGV703096H, NGV703101H, NGV703592H, NGV704503H, NGV705721H, NGV706361H, NGV706362H, NGV707132H, NGV707212H, NGV707304H, NGV709086H, NGV710399H, NGV710400H, NGV711198H, NGV712633H, NGV712638H, NGV715410H, NGV715726H, NGV716510H, NGV716610H, NGV716612H, NGV716645H, NGV716951H, NGV716986H, NGV717022H, NGV717044H, NGV717286H, NGV717525H, NGV718032H, NGV718033H, NGV718737H, NGV718738H, NGV720160H, NGV720735H, NGV721284H, NGV722654H, NGV723026H, NGV723620H, NGV726826H, NGV727650H, NGV002380H, NGV700156H, NGV700723H, NGV704753H, NGV719539H, NGV001836H, NGV706803H, NGV706438H, NGV708531H, NGV711934H, NGV718122H, NGV721017H, NGV001095H, NGV001211H, NGV004963H, NGV650347H, NGV704307H, NGV717871H, NGV719244H, NGV721149H, NGV652055H, NGV700982H, NGV711591H, NGV718848H, NGV726543H, NGV726947H, NGV703970H, NGV004172H, NGV651244H, NGV705649H, NGV710446H, NGV720552H, NGV724900H, NGV725654H, NGV726101H, NGV726623H, NGV726678H, NGV726692H, NGV726948H, NGV718192H, NGV719592H, NGV723021H, NGV710100H, NGV711158H, NGV702325H, NGV709527H, NGV717379H, NGV717509H, NGV718149H, NGV718260H, NGV721772H, NGV723782H, NGV725683H, NGV726969H, NGV726970H, NGV720466H, NGV708489H, NGV715937H, NGV721309H, NGV723800H, NGV705879H, NGV706463H, NGV706483H, NGV707939H, NGV708785H, NGV709255H, NGV720011H, NGV720494H, NGV720495H, NGV721013H, NGV721150H, NGV723491H, NGV725457H, NGV725473H, NGV726158H, NGV707613H, NGV707690H, NGV707693H, NGV707695H, NGV715253H, NGV715928H, NGV715930H, NGV715974H, NGV717648H, NGV719013H, NGV722072H, NGV722546H, NGV722549H, NGV722550H, NGV727619H, NGV702758H, NGV703420H, NGV708846H, NGV710317H, NGV711313H, NGV712406H, NGV719157H, NGV720954H, NGV721799H, NGV722086H, NGV727009H, NGV001295H, NGV704295H, NGV650006H, NGV702935H, NGV709945H, NGV717380H, NGV717469H, NGV717472H, NGV724115H, NGV724804H, NGV726995H, NGV005000H, NGV005208H, NGV718746H, NGV702814H, NGV710083H, NGV715785H, NGV719303H, NGV720094H, NGV720196H, NGV720527H, NGV721398H, NGV723343H, NGV723659H, NGV725590H, NGV702174H, NGV650687H, NGV701839H, NGV704374H, NGV705400H, NGV705663H, NGV706121H, NGV708963H, NGV711796H, NGV711856H, NGV711857H, NGV713037H, NGV713574H, NGV717809H, NGV717961H, NGV718622H, NGV720013H, NGV721189H, NGV721190H, NGV721943H, NGV721980H, NGV722689H, NGV722730H, NGV726284H, NGV726285H, NGV000162H, NGV002822H, NGV710918H, NGV715480H, NGV718586H, NGV719861H, NGV726575H, NGV726609H, NGV702911H, NGV707048H, NGV707383H, NGV707941H, NGV708644H, NGV709111H, NGV710108H, NGV710262H, NGV710356H, NGV710416H, NGV712272H, NGV712661H, NGV712709H, NGV712710H, NGV712719H, NGV712768H, NGV712966H, NGV713049H, NGV713522H, NGV713528H, NGV713660H, NGV714210H, NGV714241H, NGV714244H, NGV714245H, NGV714247H, NGV714249H, NGV714250H, NGV714617H, NGV715563H, NGV715695H, NGV715719H, NGV715963H, NGV716866H, NGV716869H, NGV716872H, NGV716880H, NGV716887H, NGV716893H, NGV716903H, NGV716904H, NGV716905H, NGV716918H, NGV716934H, NGV717013H, NGV717015H, NGV717059H, NGV717061H, NGV717063H, NGV717064H, NGV717104H, NGV717105H, NGV717316H, NGV717317H, NGV717318H, NGV717335H, NGV717336H, NGV717337H, NGV717339H, NGV717340H, NGV717420H, NGV717477H, NGV717478H, NGV717479H, NGV717480H, NGV717512H, NGV717526H, NGV717528H, NGV717529H, NGV717530H, NGV717531H, NGV717532H, NGV717813H, NGV718250H, NGV718325H, NGV718327H, NGV718328H, NGV718329H, NGV718330H, NGV718367H, NGV718370H, NGV718371H, NGV718552H, NGV718710H, NGV718713H, NGV718714H, NGV718716H, NGV718717H, NGV718718H, NGV718719H, NGV718720H, NGV718721H, NGV718757H, NGV718761H, NGV718762H, NGV718938H, NGV718967H, NGV718969H, NGV719182H, NGV719183H, NGV719199H, NGV719213H, NGV719284H, NGV719285H, NGV719286H, NGV719368H, NGV719386H, NGV720128H, NGV720131H, NGV720133H, NGV720359H, NGV720361H, NGV720362H, NGV720363H, NGV720364H, NGV720365H, NGV720366H, NGV720367H, NGV720369H, NGV720370H, NGV720371H, NGV720372H, NGV720607H, NGV720608H, NGV720609H, NGV720610H, NGV720611H, NGV720612H, NGV720614H, NGV720616H, NGV720617H, NGV720623H, NGV720854H, NGV720969H, NGV721151H, NGV721152H, NGV721259H, NGV721395H, NGV721414H, NGV721430H, NGV721565H, NGV721567H, NGV721568H, NGV721569H, NGV721570H, NGV721571H, NGV721766H, NGV721767H, NGV721901H, NGV722459H, NGV722461H, NGV722462H, NGV722463H, NGV722892H, NGV722893H, NGV722894H, NGV722895H, NGV722987H, NGV723043H, NGV723058H, NGV723059H, NGV723213H, NGV723215H, NGV723277H, NGV723278H, NGV723304H, NGV723305H, NGV723332H, NGV723519H, NGV723522H, NGV723528H, NGV723675H, NGV723688H, NGV723766H, NGV723818H, NGV724242H, NGV724253H, NGV724256H, NGV724257H, NGV724271H, NGV724480H, NGV724482H, NGV724833H, NGV724843H, NGV724852H, NGV724861H, NGV724865H, NGV724903H, NGV725276H, NGV725280H, NGV725458H, NGV725545H, NGV725732H, NGV725939H, NGV726148H, NGV726166H, NGV726167H, NGV726169H, NGV726170H, NGV726172H, NGV726255H, NGV726293H, NGV726522H, NGV726523H, NGV726525H, NGV726526H, NGV726527H, NGV726582H, NGV726589H, NGV726923H, NGV727538H, NGV727539H, NGV727540H, NGV727541H, NGV727542H, NGV727854H, NGV002453H, NGV652304H, NGV704366H, NGV005264H, NGV703596H, NGV000387H, NGV003277H, NGV004043H, NGV004847H, NGV004872H, NGV005148H, NGV700238H, NGV701030H, NGV704721H, NGV705164H, NGV705366H, NGV706809H, NGV709525H, NGV710913H, NGV711866H, NGV711867H, NGV714180H, NGV718755H, NGV719020H, NGV719067H, NGV719380H, NGV719613H, NGV719856H, NGV720056H, NGV721154H, NGV721157H, NGV722707H, NGV723309H, NGV726025H, NGV726741H, NGV726974H, NGV708629H, NGV000702H, NGV001979H, NGV703616H, NGV704636H, NGV711199H, NGV713884H, NGV716265H, NGV718888H, NGV722060H, NGV001961H, NGV003011H, NGV003025H, NGV003026H, NGV650394H, NGV702741H, NGV703081H, NGV703595H, NGV706645H, NGV712735H, NGV717495H, NGV718612H, NGV722594H, NGV722784H, NGV727791H, NGV705932H, NGV708527H, NGV708556H, NGV714585H, NGV715270H, NGV715673H, NGV715872H, NGV719622H, NGV720412H, NGV723229H, NGV723891H, NGV726992H, NGV707002H, NGV711174H, NGV711214H, NGV714614H, NGV004987H, NGV700488H, NGV702140H, NGV702301H, NGV703933H, NGV705307H, NGV707037H, NGV707163H, NGV707257H, NGV707260H, NGV707261H, NGV707397H, NGV708611H, NGV710276H, NGV711910H, NGV712973H, NGV715714H, NGV716295H, NGV717113H, NGV718489H, NGV719062H, NGV720014H, NGV720015H, NGV722773H, NGV723196H, NGV725152H, NGV725818H, NGV726196H, NGV726400H, NGV714628H, NGV001057H, NGV650096H, NGV651235H, NGV702767H, NGV706452H, NGV709824H, NGV712291H, NGV719293H, NGV725387H, NGV726412H, NGV701700H, NGV708961H, NGV001552H, NGV002065H, NGV002491H, NGV004103H, NGV005276H, NGV704122H, NGV704238H, NGV706948H, NGV712785H, NGV713056H, NGV715297H, NGV716858H, NGV726192H, NGV719486H, NGV722075H, NGV709887H, NGV709890H, NGV001972H, NGV002056H, NGV700689H, NGV701120H, NGV701691H, NGV703291H, NGV703294H, NGV703611H, NGV703641H, NGV704100H, NGV704549H, NGV705461H, NGV705762H, NGV706364H, NGV706840H, NGV707144H, NGV708573H, NGV708574H, NGV709272H, NGV709496H, NGV710043H, NGV710903H, NGV711021H, NGV711302H, NGV711509H, NGV711890H, NGV711932H, NGV711946H, NGV712440H, NGV714286H, NGV714346H, NGV716954H, NGV718295H, NGV718562H, NGV718563H, NGV720436H, NGV721337H, NGV721862H, NGV721863H, NGV722904H, NGV723560H, NGV723689H, NGV726102H, NGV727592H, NGV000106H, NGV001947H, NGV002247H, NGV002254H, NGV003475H, NGV003477H, NGV003648H, NGV003915H, NGV004139H, NGV005181H, NGV700685H, NGV700740H, NGV702187H, NGV702191H, NGV702193H, NGV705163H, NGV705806H, NGV707052H, NGV707053H, NGV707054H, NGV707439H, NGV709003H, NGV709122H, NGV710194H, NGV710322H, NGV715720H, NGV717345H, NGV718076H, NGV718377H, NGV718450H, NGV719128H, NGV720457H, NGV721254H, NGV721636H, NGV722566H, NGV722567H, NGV722984H, NGV722985H, NGV723279H, NGV723542H, NGV723756H, NGV725337H, NGV725461H, NGV725470H, NGV725471H, NGV726789H, NGV709226H, NGV715888H, NGV718333H, NGV718352H, NGV722382H, NGV725303H, NGV720961H, NGV652536H, NGV724643H, NGV725720H, NGV726606H, NGV705425H, NGV707038H, NGV003455H, NGV003610H, NGV004615H, NGV703234H, NGV709553H, NGV716357H, NGV723820H, NGV725437H, NGV002356H, NGV004132H, NGV707364H, NGV716999H, NGV720324H, NGV701791H, NGV004768H, NGV712868H, NGV716300H, NGV701486H, NGV712556H, NGV716873H, NGV717069H, NGV718526H, NGV718527H, NGV718529H, NGV718690H, NGV718692H, NGV718732H, NGV718885H, NGV719004H, NGV719162H, NGV719180H, NGV719262H, NGV719336H, NGV719628H, NGV719629H, NGV719839H, NGV719840H, NGV720627H, NGV720845H, NGV720880H, NGV720881H, NGV720990H, NGV721592H, NGV721765H, NGV721972H, NGV721973H, NGV722391H, NGV722447H, NGV723449H, NGV726235H, NGV727026H, NGV727599H, NGV727600H, NGV718168H, NGV722749H, NGV715415H, NGV715650H, NGV000868H, NGV004233H, NGV004352H, NGV700568H, NGV704118H, NGV705186H, NGV705412H, NGV708200H, NGV709804H, NGV711930H, NGV712672H, NGV716455H, NGV716463H, NGV717666H, NGV719542H, NGV721030H, NGV721335H, NGV721802H, NGV722034H, NGV723322H, NGV726299H, NGV727139H, NGV719974H, NGV004887H, NGV703954H, NGV708809H, NGV708810H, NGV712604H, NGV000462H, NGV000463H, NGV000841H, NGV000930H, NGV001863H, NGV002799H, NGV002947H, NGV003399H, NGV004952H, NGV004972H, NGV004978H, NGV650171H, NGV650901H, NGV651486H, NGV701213H, NGV701519H, NGV701778H, NGV702269H, NGV707949H, NGV708902H, NGV708996H, NGV709022H, NGV709523H, NGV709903H, NGV710071H, NGV715756H, NGV719372H, NGV720021H, NGV720520H, NGV720982H, NGV720737H, NGV717996H, NGV721270H, NGV721492H, NGV722755H, NGV724950H, NGV724954H, NGV726451H, NGV727329H, NGV651273H, NGV702102H, NGV709024H, NGV711972H, NGV712027H, NGV002129H, NGV002132H, NGV003768H, NGV003937H, NGV004045H, NGV005060H, NGV652256H, NGV700407H, NGV700426H, NGV700739H, NGV702077H, NGV702079H, NGV702268H, NGV702912H, NGV704847H, NGV704848H, NGV704850H, NGV705977H, NGV707482H, NGV710402H, NGV711288H, NGV712848H, NGV715527H, NGV715532H, NGV715897H, NGV716988H, NGV717951H, NGV717952H, NGV718135H, NGV718532H, NGV719506H, NGV720331H, NGV722731H, NGV723295H, NGV723559H, NGV724672H, NGV726401H, NGV726972H, NGV711151H, NGV713251H, NGV714147H, NGV717870H, NGV718667H, NGV725219H, NGV001272H, NGV726111H, NGV721924H, NGV726445H, NGV720201H, NGV651953H, NGV652401H, NGV711176H, NGV713584H, NGV714599H, NGV721124H, NGV726324H, NGV701345H, NGV706177H, NGV712671H, NGV721358H, NGV725264H, NGV001331H, NGV650862H, NGV706240H, NGV712611H, NGV722637H, NGV726479H, NGV719014H, NGV703041H, NGV711728H, NGV719763H, NGV720468H, NGV651692H, NGV709872H, NGV650112H, NGV709742H, NGV716859H, NGV702180H, NGV004597H, NGV705159H, NGV717033H, NGV711945H, NGV712869H, NGV717321H, NGV727040H, NGV721139H, NGV723048H, NGV003894H, NGV704390H, NGV707502H, NGV707840H, NGV708634H, NGV714323H, NGV715751H, NGV717095H, NGV718453H, NGV718515H, NGV720338H, NGV722765H, NGV723705H, NGV715623H, NGV708819H, NGV711044H, NGV718322H, NGV717192H, NGV718815H, NGV720858H, NGV725380H, NGV725425H, NGV726238H, NGV708923H, NGV702085H, NGV710201H, NGV710203H, NGV712558H, NGV000379H, NGV000848H, NGV000852H, NGV000856H, NGV003701H, NGV651436H, NGV707057H, NGV711266H, NGV711510H, NGV712871H, NGV650705H, NGV704796H, NGV709474H, NGV710224H, NGV711348H, NGV711691H, NGV715789H, NGV715791H, NGV715797H, NGV716731H, NGV717594H, NGV717595H, NGV717847H, NGV718262H, NGV718981H, NGV720890H, NGV725933H, NGV706838H, NGV712021H, NGV712023H, NGV712491H, NGV716788H, NGV717351H, NGV721821H, NGV720597H, NGV721188H, NGV722986H, NGV726834H, NGV001898H, NGV708592H, NGV709281H, NGV003827H, NGV713321H, NGV713627H, NGV722936H, NGV725817H, NGV725986H, NGV000461H, NGV004285H, NGV702668H, NGV705433H, NGV713583H, NGV721045H, NGV000163H, NGV001263H, NGV651081H, NGV651341H, NGV709859H, NGV712071H, NGV718267H, NGV725189H, NGV726674H, NGV001268H, NGV003743H, NGV712955H, NGV721285H, NGV723137H, NGV001220H, NGV701960H, NGV717734H, NGV004392H, NGV004825H, NGV650357H, NGV651851H, NGV651966H, NGV652108H, NGV652651H, NGV652655H, NGV702673H, NGV715796H, NGV717678H, NGV725389H, NGV725396H, NGV704126H, NGV707343H, NGV708321H, NGV716942H, NGV650106H, NGV700249H, NGV002671H, NGV002882H, NGV003102H, NGV004668H, NGV005146H, NGV652131H, NGV700747H, NGV701834H, NGV703060H, NGV705149H, NGV705289H, NGV705648H, NGV705898H, NGV707059H, NGV707948H, NGV708287H, NGV711947H, NGV715515H, NGV716864H, NGV716911H, NGV716962H, NGV718499H, NGV720541H, NGV720924H, NGV720993H, NGV721496H, NGV721501H, NGV722902H, NGV723446H, NGV725793H, NGV000383H, NGV004760H, NGV650906H, NGV703912H, NGV704384H, NGV708598H, NGV710097H, NGV710380H, NGV710381H, NGV710382H, NGV711517H, NGV712371H, NGV002529H, NGV703982H, NGV704155H, NGV705202H, NGV706225H, NGV712030H, NGV714300H, NGV717367H, NGV720179H, NGV001987H, NGV002899H, NGV004330H, NGV004780H, NGV650057H, NGV651891H, NGV700634H, NGV703009H, NGV703126H, NGV703637H, NGV704424H, NGV704611H, NGV704690H, NGV706749H, NGV706998H, NGV707387H, NGV707875H, NGV707971H, NGV709792H, NGV711258H, NGV716545H, NGV703661H, NGV705656H, NGV706837H, NGV004681H, NGV703012H, NGV712053H, NGV716693H, NGV723230H, NGV715593H, NGV715934H, NGV719897H, NGV720847H, NGV722735H, NGV723701H, NGV726211H, NGV726550H, NGV708769H, NGV717643H, NGV715907H, NGV723757H, NGV002703H, NGV003985H, NGV713634H, NGV713637H, NGV713686H, NGV726300H, NGV704203H, NGV708205H, NGV000126H, NGV000382H, NGV001918H, NGV002337H, NGV650073H, NGV651177H, NGV651721H, NGV651723H, NGV701688H, NGV701698H, NGV704450H, NGV705194H, NGV705978H, NGV710318H, NGV710354H, NGV712729H, NGV717106H, NGV717319H, NGV717896H, NGV718286H, NGV718631H, NGV720975H, NGV723707H, NGV723765H, NGV723767H, NGV724294H, NGV724739H, NGV726242H, NGV726243H, NGV726248H, NGV701866H, NGV705113H, NGV707457H, NGV725319H, NGV652129H, NGV702613H, NGV720125H, NGV726878H, NGV705786H, NGV708500H, NGV708501H, NGV708502H, NGV708504H, NGV708506H, NGV708507H, NGV708509H, NGV708510H, NGV708514H, NGV708515H, NGV708516H, NGV708517H, NGV708521H, NGV711830H, NGV711833H, NGV711838H, NGV711839H, NGV711840H, NGV711841H, NGV718404H, NGV718410H, NGV718411H, NGV718416H, NGV718418H, NGV718422H, NGV718650H, NGV719660H, NGV722716H, NGV722717H, NGV722718H, NGV722719H, NGV722721H, NGV722722H, NGV723731H, NGV723732H, NGV723734H, NGV723736H, NGV723737H, NGV725113H, NGV725227H, NGV725462H, NGV725464H, NGV725465H, NGV725466H, NGV725467H, NGV725468H, NGV725705H, NGV725706H, NGV725707H, NGV725708H, NGV725709H, NGV725710H, NGV725711H, NGV725712H, NGV725713H, NGV725714H, NGV725895H, NGV725896H, NGV725898H, NGV725900H, NGV701522H, NGV706330H, NGV707134H, NGV707135H, NGV714142H, NGV720498H, NGV725002H, NGV707447H, NGV001905H, NGV001906H, NGV002059H, NGV002798H, NGV003337H, NGV003338H, NGV003339H, NGV004323H, NGV651474H, NGV652059H, NGV701966H, NGV701967H, NGV701968H, NGV703670H, NGV705466H, NGV706194H, NGV707050H, NGV707609H, NGV707610H, NGV707611H, NGV708658H, NGV709974H, NGV710292H, NGV711554H, NGV711581H, NGV711582H, NGV712754H, NGV713697H, NGV714099H, NGV714139H, NGV715056H, NGV716710H, NGV717381H, NGV718051H, NGV718879H, NGV718921H, NGV719072H, NGV719259H, NGV719590H, NGV719692H, NGV720032H, NGV720037H, NGV721467H, NGV722882H, NGV724686H, NGV726369H, NGV726581H, NGV001011H, NGV003938H, NGV702960H, NGV703424H, NGV710222H, NGV715699H, NGV720704H, NGV711794H, NGV001526H, NGV001527H, NGV003003H, NGV003004H, NGV651308H, NGV709969H, NGV001209H, NGV002939H, NGV650838H, NGV701716H, NGV713470H, NGV713968H, NGV715591H, NGV718395H, NGV724556H, NGV702177H, NGV703084H, NGV704792H, NGV719463H, NGV721781H, NGV724657H, NGV650661H, NGV719151H, NGV000904H, NGV001187H, NGV002569H, NGV003402H, NGV003462H, NGV005258H, NGV650633H, NGV651229H, NGV651377H, NGV700748H, NGV704402H, NGV704990H, NGV705360H, NGV705610H, NGV707039H, NGV707117H, NGV707147H, NGV709831H, NGV711996H, NGV711997H, NGV724611H, NGV726087H, NGV727808H, NGV701230H, NGV705720H, NGV711482H, NGV652484H, NGV707177H, NGV711350H, NGV001321H, NGV001842H, NGV002036H, NGV003757H, NGV003944H, NGV004007H, NGV004008H, NGV004980H, NGV650642H, NGV651307H, NGV651491H, NGV651867H, NGV651868H, NGV700175H, NGV700379H, NGV701505H, NGV701871H, NGV702612H, NGV702621H, NGV703008H, NGV703010H, NGV704464H, NGV704465H, NGV705719H, NGV705797H, NGV705849H, NGV705858H, NGV706796H, NGV706797H, NGV707015H, NGV707538H, NGV707545H, NGV709875H, NGV709879H, NGV711308H, NGV713041H, NGV713356H, NGV714428H, NGV714429H, NGV714431H, NGV715487H, NGV715798H, NGV718034H, NGV718821H, NGV718922H, NGV719141H, NGV719310H, NGV721797H, NGV721798H, NGV722076H, NGV723808H, NGV724779H, NGV724781H, NGV725169H, NGV726442H, NGV727288H, NGV726617H, NGV001590H, NGV703978H, NGV706793H, NGV650064H, NGV707501H, NGV709267H, NGV717976H, NGV722095H, NGV726557H, NGV651959H, NGV726100H, NGV716740H, NGV720710H, NGV725388H, NGV002796H, NGV701068H, NGV706255H, NGV706656H, NGV707106H, NGV708906H, NGV709070H, NGV711511H, NGV714332H, NGV716258H, NGV716946H, NGV717408H, NGV719353H, NGV720715H, NGV720908H, NGV720968H, NGV721132H, NGV722756H, NGV723945H, NGV725866H, NGV727679H, NGV727795H, NGV722055H, NGV722797H, NGV720438H, NGV709370H, NGV703092H, NGV709263H, NGV717909H, NGV717910H, NGV718269H, NGV718270H, NGV719149H, NGV720570H, NGV720571H, NGV726951H, NGV704720H, NGV712643H, NGV715834H, NGV715844H, NGV717268H, NGV718035H, NGV719209H, NGV719210H, NGV725647H, NGV726551H, NGV727007H, NGV005156H, NGV700462H, NGV704119H, NGV706727H, NGV711144H, NGV713846H, NGV716931H, NGV723317H, NGV724921H, NGV726277H, NGV002330H, NGV703064H, NGV703163H, NGV703349H, NGV710393H, NGV720029H, NGV726275H, NGV706732H, NGV715494H, NGV715777H, NGV706403H, NGV707136H, NGV712573H, NGV714467H, NGV714468H, NGV715958H, NGV717943H, NGV718344H, NGV718868H, NGV719009H, NGV721010H, NGV725010H, NGV725020H, NGV725062H, NGV712536H, NGV710430H, NGV714570H, NGV004430H, NGV707041H, NGV711296H, NGV711860H, NGV712590H, NGV713662H, NGV713687H, NGV714080H, NGV715495H, NGV718601H, NGV720004H, NGV722881H, NGV723530H, NGV723849H, NGV727036H, NGV718331H, NGV726785H, NGV726786H, NGV704945H, NGV717273H, NGV719969H, NGV723236H, NGV723237H, NGV723408H, NGV711885H, NGV713958H, NGV723366H, NGV719179H, NGV707183H, NGV707530H, NGV708743H, NGV708792H, NGV709978H, NGV710025H, NGV710366H, NGV711808H, NGV711811H, NGV711966H, NGV713636H, NGV716296H, NGV716748H, NGV716786H, NGV718571H, NGV719348H, NGV720199H, NGV720518H, NGV720792H, NGV720850H, NGV720998H, NGV721081H, NGV721910H, NGV726462H, NGV717562H, NGV722922H, NGV726629H, NGV707178H, NGV710350H, NGV712360H, NGV717248H, NGV717275H, NGV717278H, NGV718537H, NGV719484H, NGV720399H, NGV722714H, NGV723455H, NGV723685H, NGV726614H, NGV727875H, NGV717878H, NGV720793H, NGV726705H, NGV703609H, NGV705917H, NGV720652H, NGV721534H, NGV726980H, NGV722745H, NGV703838H, NGV705021H, NGV706404H, NGV707989H, NGV711189H, NGV713587H, NGV722818H, NGV704075H, NGV704349H, NGV725971H, NGV726677H, NGV723885H, NGV727330H, NGV714322H, NGV715757H, NGV716862H, NGV718457H, NGV718940H, NGV719373H, NGV719469H, NGV719639H, NGV719991H, NGV719992H, NGV720877H, NGV723567H, NGV723568H, NGV723954H, NGV725358H, NGV716590H, NGV715467H, NGV717124H, NGV718951H, NGV719492H, NGV720752H, NGV724516H, NGV716343H, NGV720103H, NGV720181H, NGV725285H, NGV725286H, NGV725399H, NGV726599H, NGV726640H, NGV715707H, NGV716715H, NGV718572H, NGV719134H, NGV720462H, NGV720463H, NGV720489H, NGV720922H, NGV721245H, NGV721246H, NGV721949H, NGV722535H, NGV722536H, NGV723214H, NGV723641H, NGV723964H, NGV725544H, NGV726239H, NGV726670H, NGV727294H, NGV718903H, NGV720761H, NGV721560H, NGV721807H, NGV723239H, NGV724761H, NGV724850H, NGV718265H, NGV724272H, NGV715584H, NGV715878H, NGV717078H, NGV717123H, NGV717127H, NGV717128H, NGV717545H, NGV718441H, NGV719050H, NGV719155H, NGV719865H, NGV722501H, NGV722661H, NGV723831H, NGV724895H, NGV726836H, NGV720312H, NGV723722H, NGV724705H, NGV725595H, NGV726554H, NGV726604H, NGV719560H, NGV721278H, NGV723555H, NGV720271H, NGV727042H, NGV714161H, NGV714602H, NGV716593H, NGV716711H, NGV716915H, NGV717979H, NGV718743H, NGV721638H, NGV722009H, NGV723381H, NGV713348H, NGV727574H, NGV726669H, NGV725508H, NGV727673H, NGV722539H, NGV726464H, NGV725745H, NGV726407H, NGV724986H, NGV724993H, NGV726461H, NGV003815H, NGV004315H, NGV724926H, NGV723512H, NGV722737H, NGV702842H, NGV726534H, NGV726647H, NGV003778H, NGV001053H, NGV706953H, NGV712031H, NGV708927H, NGV000832H, NGV652394H, NGV715581H, NGV709796H, NGV002596H, NGV650132H, NGV000872H, NGV652450H, NGV001109H, NGV710758H, NGV707242H, NGV726549H, NGV708190H, NGV003055H, NGV708657H, NGV708858H, NGV708860H, NGV726002H, NGV726003H, NGV724468H, NGV726114H, NGV721363H, NGV716110H, NGV722872H, NGV711787H, NGV000732H, NGV702339H, NGV705820H, NGV711924H, NGV724968H, NGV725770H, NGV706618H, NGV718578H, NGV721963H, NGV712477H, NGV714624H, NGV002117H, NGV002567H, NGV004213H, NGV651413H, NGV712794H, NGV700988H, NGV722564H, NGV716744H, NGV003153H, NGV650173H, NGV712219H, NGV719707H, NGV720163H, NGV726314H, NGV701895H, NGV713673H, NGV713978H, NGV716949H, NGV717174H, NGV726188H, NGV707429H, NGV720307H, NGV721005H, NGV701523H, NGV004211H, NGV000917H, NGV702175H, NGV719720H, NGV721100H, NGV706099H, NGV709997H, NGV713995H, NGV000484H, NGV722584H, NGV725559H, NGV705251H, NGV726069H, NGV712983H, NGV718140H, NGV718217H, NGV723334H, NGV003367H, NGV704956H, NGV712592H, NGV705681H, NGV721703H, NGV721704H, NGV721705H, NGV726224H, NGV001293H, NGV700708H, NGV726536H, NGV722778H, NGV003913H, NGV720686H, NGV721956H, NGV721957H, NGV722419H, NGV722420H, NGV722421H, NGV707020H, NGV703861H, NGV000314H, NGV723829H, NGV650736H, NGV718789H, NGV726193H, NGV723386H, NGV004607H, NGV703228H, NGV707007H, NGV716191H, NGV705682H, NGV727651H, NGV003136H, NGV725702H, NGV719887H, NGV708160H, NGV719350H, NGV713289H, NGV001776H, NGV003735H, NGV710485H, NGV719232H, NGV704912H, NGV702670H, NGV708536H, NGV721275H, NGV724788H, NGV652337H, NGV724638H, NGV708926H, NGV705812H, NGV723624H, NGV702624H, NGV714339H, NGV711911H, NGV704121H, NGV709588H, NGV717703H, NGV701386H, NGV705250H, NGV715490H, NGV000504H, NGV651210H, NGV719811H, NGV004423H, NGV721631H, NGV721791H, NGV722487H, NGV700428H, NGV720722H, NGV725109H, NGV712734H, NGV719415H, NGV702204H, NGV650306H, NGV000919H, NGV717139H, NGV719242H, NGV725696H, NGV701385H, NGV717799H, NGV717800H, NGV721361H, NGV723354H, NGV706304H, NGV003777H, NGV717622H, NGV704019H, NGV004507H, NGV712328H, NGV002390H, NGV702852H, NGV707414H, NGV721485H, NGV726311H, NGV002928H, NGV002930H, NGV003980H, NGV004138H, NGV707562H, NGV650463H, NGV714165H, NGV004081H, NGV704466H, NGV724871H, NGV726846H, NGV718385H, NGV718636H, NGV719169H, NGV702659H, NGV650621H, NGV708181H, NGV718290H, NGV722685H, NGV709644H, NGV725107H, NGV718288H, NGV707456H, NGV717404H, NGV719934H, NGV726517H, NGV705804H, NGV701258H, NGV701970H, NGV700724H, NGV719456H, NGV004110H, NGV651709H, NGV717938H, NGV718849H, NGV708087H, NGV725305H, NGV650659H, NGV001043H, NGV718482H, NGV715870H, NGV722650H, NGV650404H, NGV651191H, NGV700822H, NGV721599H, NGV651003H, NGV711168H, NGV713685H, NGV723838H, NGV004953H, NGV005177H, NGV718311H, NGV722777H, NGV723264H, NGV651247H, NGV652110H, NGV650726H, NGV700741H, NGV701901H, NGV703053H, NGV707195H, NGV727196H, NGV706457H, NGV714336H, NGV721342H, NGV720664H, NGV001560H, NGV702739H, NGV652263H, NGV721169H, NGV708973H, NGV717232H, NGV005203H, NGV721055H, NGV004358H, NGV726510H, NGV651285H, NGV004182H, NGV000725H, NGV000728H, NGV717186H, NGV702543H, NGV703082H, NGV721377H, NGV721464H, NGV000559H, NGV708920H, NGV718638H, NGV703621H, NGV725588H, NGV650574H, NGV713642H, NGV722548H, NGV001247H, NGV709415H, NGV718392H, NGV003919H, NGV702407H, NGV727700H, NGV652275H, NGV700442H, NGV702186H, NGV652444H, NGV725550H, NGV727002H, NGV652147H, NGV710359H, NGV702323H, NGV707577H, NGV719418H, NGV719419H, NGV719680H, NGV708505H, NGV715625H, NGV718682H, NGV721219H, NGV716574H, NGV000835H, NGV002111H, NGV718238H, NGV722972H, NGV712389H, NGV715820H, NGV718575H, NGV718798H, NGV718820H, NGV721102H, NGV001050H, NGV711706H, NGV003655H, NGV650100H, NGV712781H, NGV704851H, NGV651117H, NGV714335H, NGV701680H, NGV711274H, NGV706928H, NGV722811H, NGV652185H, NGV652330H, NGV708325H, NGV720951H, NGV705429H, NGV652167H, NGV701670H, NGV701673H, NGV707411H, NGV714282H, NGV720063H, NGV724322H, NGV705486H, NGV705560H, NGV705561H, NGV707894H, NGV721917H, NGV725175H, NGV726861H, NGV727380H, NGV724990H, NGV001806H, NGV000110H, NGV707847H, NGV704677H, NGV712104H, NGV726054H, NGV721339H, NGV727617H, NGV724508H, NGV721881H, NGV721133H, NGV719240H, NGV650999H, NGV000697H, NGV003714H, NGV004447H, NGV712044H, NGV719528H, NGV718462H, NGV705584H, NGV001760H, NGV002488H, NGV723223H, NGV652586H, NGV725045H, NGV705197H, NGV726607H, NGV720912H, NGV707881H, NGV650886H, NGV712580H, NGV712844H, NGV717292H, NGV722429H, NGV707970H, NGV722044H, NGV709316H, NGV001124H, NGV004863H, NGV710412H, NGV003917H, NGV700999H, NGV722605H, NGV722627H, NGV722628H, NGV711991H, NGV725100H, NGV725111H, NGV725134H, NGV718580H, NGV002686H, NGV004382H, NGV703568H, NGV721354H, NGV727043H, NGV727045H, NGV724727H, NGV724728H, NGV715595H, NGV651843H, NGV650099H, NGV726448H, NGV707618H, NGV722903H, NGV004240H, NGV721543H, NGV719224H, NGV720707H, NGV706692H, NGV714807H, NGV719659H, NGV726141H, NGV709924H, NGV724610H, NGV718642H, NGV726147H, NGV002780H, NGV004545H, NGV718823H, NGV726615H, NGV720326H; GTIN 00643169701007, Serial Numbers: NGV519252H, NGV520381H, NGV520816H, NGV522425H, NGV528369H, NGV517848H, NGV526015H, NGV513907H, NGV514266H, NGV516659H, NGV520147H, NGV520367H, NGV521104H, NGV521227H, NGV523368H, NGV524855H, NGV525215H, NGV527183H, NGV518156H, NGV519210H, NGV515637H, NGV515961H, NGV516738H, NGV517603H, NGV519806H, NGV520033H, NGV520777H, NGV520829H, NGV524751H, NGV525684H, NGV520835H, NGV523569H, NGV525729H, NGV515280H, NGV516136H, NGV516783H, NGV523202H, NGV523224H, NGV523431H, NGV524828H, NGV524972H, NGV526674H, NGV527583H, NGV519227H, NGV519429H, NGV520037H, NGV524771H, NGV525422H, NGV515844H, NGV515852H, NGV517919H, NGV518557H, NGV518823H, NGV519267H, NGV521155H, NGV523232H, NGV523317H, NGV523398H, NGV523922H, NGV525163H, NGV525562H, NGV527610H, NGV515290H, NGV517612H, NGV519958H, NGV515091H, NGV515848H, NGV516833H, NGV517943H, NGV521183H, NGV523934H, NGV515569H, NGV515570H, NGV515572H, NGV517554H, NGV517922H, NGV518142H, NGV518459H, NGV518914H, NGV518916H, NGV520364H, NGV520456H, NGV520513H, NGV520704H, NGV522191H, NGV522451H, NGV523795H, NGV524781H, NGV525735H, NGV526423H, NGV527774H, NGV527778H, NGV515613H, NGV525131H, NGV518829H, NGV520239H, NGV520279H, NGV515435H, NGV516094H, NGV516106H, NGV522991H, NGV525537H, NGV515805H, NGV515808H, NGV515812H, NGV517763H, NGV517900H, NGV517961H, NGV518793H, NGV520130H, NGV520515H, NGV521122H, NGV523417H, NGV523931H, NGV524216H, NGV524862H, NGV524983H, NGV525514H, NGV525519H, NGV526672H, NGV527035H, NGV527146H, NGV527300H, NGV527555H, NGV515849H, NGV517143H, NGV523640H, NGV514313H, NGV516576H, NGV516579H, NGV517895H, NGV520323H, NGV520900H, NGV521098H, NGV523556H, NGV523559H, NGV524193H, NGV524426H, NGV525106H, NGV525680H, NGV526058H, NGV527791H, NGV514946H, NGV519578H, NGV519944H, NGV523359H, NGV516095H, NGV523827H, NGV515854H, NGV516574H, NGV525340H, NGV526656H, NGV527591H, NGV528361H, NGV528406H, NGV515594H, NGV516067H, NGV520580H, NGV523546H, NGV524175H, NGV525212H, NGV525583H, NGV527573H, NGV515134H, NGV515589H, NGV516311H, NGV523389H, NGV516723H, NGV520034H, NGV523808H, NGV524985H, NGV515430H, NGV515431H, NGV518168H, NGV519809H, NGV519826H, NGV522121H, NGV522123H, NGV522957H, NGV514453H, NGV515255H, NGV516133H, NGV517955H, NGV518911H, NGV520057H, NGV521202H, NGV521231H, NGV523507H, NGV515793H, NGV522961H, NGV524899H, NGV515567H, NGV520140H, NGV521073H, NGV523286H, NGV524201H, NGV525751H, NGV525754H, NGV525781H, NGV514424H, NGV514451H, NGV514659H, NGV514876H, NGV515151H, NGV516036H, NGV516050H, NGV516058H, NGV516130H, NGV516134H, NGV516477H, NGV516845H, NGV517584H, NGV517629H, NGV517642H, NGV517672H, NGV517899H, NGV517939H, NGV518023H, NGV518560H, NGV518629H, NGV518708H, NGV519292H, NGV519983H, NGV520287H, NGV520377H, NGV520453H, NGV520497H, NGV520509H, NGV520613H, NGV520636H, NGV520753H, NGV520886H, NGV520938H, NGV520940H, NGV521083H, NGV521086H, NGV521727H, NGV522034H, NGV522096H, NGV522403H, NGV522908H, NGV522919H, NGV523155H, NGV523289H, NGV523340H, NGV523355H, NGV523388H, NGV523543H, NGV523618H, NGV523659H, NGV523664H, NGV523826H, NGV524227H, NGV524238H, NGV524253H, NGV524763H, NGV524846H, NGV525336H, NGV525510H, NGV525578H, NGV525681H, NGV525683H, NGV525685H, NGV525686H, NGV525757H, NGV526233H, NGV526273H, NGV526307H, NGV526386H, NGV526408H, NGV526678H, NGV527259H, NGV527607H, NGV527796H, NGV527907H, NGV528413H, NGV513977H, NGV516026H, NGV516033H, NGV516811H, NGV517813H, NGV520468H, NGV520626H, NGV522445H, NGV522986H, NGV524816H, NGV525339H, NGV526725H, NGV513799H, NGV516694H, NGV516850H, NGV517923H, NGV519552H, NGV520523H, NGV520534H, NGV522167H, NGV522414H, NGV523205H, NGV523601H, NGV524325H, NGV514337H, NGV514927H, NGV515288H, NGV516721H, NGV516862H, NGV521095H, NGV522030H, NGV522138H, NGV522174H, NGV522932H, NGV523626H, NGV523627H, NGV524627H, NGV524723H, NGV525730H, NGV526365H, NGV526395H, NGV527603H, NGV516096H, NGV516297H, NGV517628H, NGV517907H, NGV516001H, NGV518673H, NGV522448H, NGV522954H, NGV516791H, NGV523442H, NGV515377H, NGV523557H, NGV514439H, NGV519246H, NGV519248H, NGV519815H, NGV522349H, NGV522988H, NGV523324H, NGV525076H, NGV515373H, NGV516756H, NGV525827H, NGV516148H, NGV516565H, NGV523378H, NGV523480H, NGV524733H, NGV525014H, NGV525141H, NGV525143H, NGV525151H, NGV525660H, NGV516620H, NGV518151H, NGV526413H, NGV514368H, NGV514441H, NGV515313H, NGV515822H, NGV519321H, NGV519322H, NGV519472H, NGV519790H, NGV519798H, NGV519940H, NGV520162H, NGV520514H, NGV520516H, NGV520593H, NGV523172H, NGV523732H, NGV524775H, NGV524959H, NGV524982H, NGV524984H, NGV525522H, NGV526409H, NGV526620H, NGV528309H, NGV528399H, NGV516817H, NGV521230H, NGV522162H, NGV522164H, NGV516683H, NGV521164H, NGV522179H, NGV522406H, NGV520218H, NGV515991H, NGV516102H, NGV517805H, NGV517901H, NGV516789H, NGV516847H, NGV517619H, NGV518089H, NGV519825H, NGV520383H, NGV522951H, NGV522952H, NGV523255H, NGV523264H, NGV523411H, NGV523679H, NGV523833H, NGV524047H, NGV524171H, NGV525589H, NGV525593H, NGV525671H, NGV525702H, NGV525783H, NGV526564H, NGV527306H, NGV524718H, NGV527271H, NGV515317H, NGV515813H, NGV518490H, NGV520489H, NGV521093H, NGV524264H, NGV524304H, NGV524927H, NGV525146H, NGV525234H, NGV514873H, NGV516318H, NGV517978H, NGV518568H, NGV520032H, NGV520306H, NGV520309H, NGV520311H, NGV522452H, NGV523333H, NGV523436H, NGV524697H, NGV524925H, NGV525524H, NGV525620H, NGV525630H, NGV525746H, NGV526303H, NGV524990H, NGV527237H, NGV527256H, NGV516692H, NGV517945H, NGV516073H, NGV516276H, NGV516749H, NGV517751H, NGV523830H, NGV525546H, NGV525559H, NGV518109H, NGV517999H, NGV522322H, NGV523533H, NGV515759H, NGV515799H, NGV519470H, NGV520019H, NGV522379H, NGV523308H, NGV526067H, NGV527775H, NGV516809H, NGV520003H, NGV520171H, NGV520391H, NGV522937H, NGV525737H, NGV520292H, NGV522624H, NGV522984H, NGV523315H, NGV515351H, NGV520622H, NGV514410H, NGV519504H, NGV527214H, NGV516821H, NGV518170H, NGV523351H, NGV527122H, NGV527763H, NGV527871H, NGV526261H, NGV517337H, NGV514428H, NGV516075H, NGV517711H, NGV525652H, NGV523616H, NGV515453H, NGV515778H, NGV515976H, NGV516822H, NGV516830H, NGV521133H, NGV522960H, NGV523474H, NGV523783H, NGV524247H, NGV525570H, NGV525722H, NGV526288H, NGV526929H, NGV514433H, NGV514740H, NGV516701H, NGV514725H, NGV514741H, NGV520918H, NGV524838H, NGV526671H, NGV527597H, NGV515790H, NGV516857H, NGV522447H, NGV520890H, NGV517713H, NGV515380H, NGV518870H, NGV514906H, NGV515643H, NGV522093H, NGV523274H, NGV523610H, NGV524867H, NGV517898H, NGV513914H, NGV514937H, NGV516644H, NGV516815H, NGV519286H, NGV521148H, NGV521153H, NGV523257H, NGV523339H, NGV523353H, NGV525501H, NGV525550H, NGV527859H, NGV526275H, NGV514863H, NGV516575H, NGV518621H, NGV518627H, NGV520281H, NGV520327H, NGV520797H, NGV520810H, NGV522674H, NGV524242H, NGV524243H, NGV525017H, NGV525049H, NGV525098H, NGV525628H, NGV527788H, NGV515086H, NGV516856H, NGV523964H, NGV518981H, NGV520286H, NGV521161H, NGV521201H, NGV522116H, NGV524790H, NGV518555H, NGV522944H, NGV523805H, NGV515082H, NGV517967H, NGV520210H, NGV526412H, NGV526521H, NGV515319H, NGV520015H, NGV519939H, NGV520257H, NGV520307H, NGV526430H, NGV526431H, NGV526440H, NGV516247H, NGV519522H, NGV519541H, NGV520341H, NGV523612H, NGV525032H, NGV527572H, NGV523628H, NGV527180H, NGV518104H, NGV526458H, NGV515293H, NGV516763H, NGV516764H, NGV520305H, NGV515392H, NGV521331H, NGV523223H, NGV515119H, NGV519175H, NGV523655H, NGV526937H, NGV526939H, NGV515325H, NGV523179H, NGV524869H, NGV525539H, NGV515274H, NGV521212H, NGV522978H, NGV524607H, NGV515273H, NGV520569H, NGV521332H, NGV515654H, NGV526573H, NGV514724H, NGV522153H, NGV522157H, NGV522159H, NGV525452H, NGV525490H, NGV525493H, NGV514282H, NGV514445H, NGV514450H, NGV515445H, NGV517906H, NGV519518H, NGV521211H, NGV525535H, NGV525647H, NGV523644H, NGV524798H, NGV514738H, NGV516755H, NGV519262H, NGV523207H, NGV524689H, NGV525159H, NGV514936H, NGV525350H, NGV525370H, NGV516564H, NGV516858H, NGV520422H, NGV524753H, NGV525748H, NGV516803H, NGV519520H, NGV522190H, NGV528216H, NGV514342H, NGV514345H, NGV514379H, NGV514387H, NGV514395H, NGV514399H, NGV514407H, NGV514422H, NGV516782H, NGV522935H, NGV522938H, NGV522958H, NGV522966H, NGV522979H, NGV522994H, NGV523037H, NGV523066H, NGV519987H, NGV520825H, NGV521238H, NGV521341H, NGV525542H, NGV516142H, NGV517856H, NGV519744H, NGV514950H, NGV514682H, NGV516792H, NGV516794H, NGV517634H, NGV517980H, NGV517981H, NGV517982H, NGV519421H, NGV520564H, NGV524802H, NGV524822H, NGV525846H, NGV526415H, NGV527115H, NGV527447H, NGV520488H, NGV516841H, NGV522377H, NGV523816H, NGV527569H, NGV519279H, NGV521237H, NGV527902H, NGV516938H, NGV515364H, NGV518179H, NGV520452H, NGV514942H, NGV515281H, NGV516826H, NGV516875H, NGV519195H, NGV521140H, NGV523658H, NGV524878H, NGV527001H, NGV518838H, NGV520335H, NGV520533H, NGV525689H, NGV526421H, NGV516086H, NGV518145H, NGV520416H, NGV523611H, NGV516736H, NGV517979H, NGV522401H, NGV525164H, NGV513986H, NGV514016H, NGV516724H, NGV516725H, NGV517579H, NGV517949H, NGV517953H, NGV520501H, NGV520676H, NGV520716H, NGV521208H, NGV523321H, NGV524832H, NGV526560H, NGV527536H, NGV513911H, NGV513930H, NGV514706H, NGV514712H, NGV514885H, NGV515256H, NGV515262H, NGV515268H, NGV516624H, NGV516626H, NGV517585H, NGV517597H, NGV520901H, NGV521157H, NGV521158H, NGV522016H, NGV522076H, NGV522148H, NGV522149H, NGV523253H, NGV523254H, NGV523267H, NGV523698H, NGV524181H, NGV524196H, NGV524903H, NGV524938H, NGV525192H, NGV525201H, NGV525296H, NGV525311H, NGV525508H, NGV526510H, NGV526546H, NGV527147H, NGV527168H, NGV527268H, NGV524255H, NGV527617H, NGV520858H, NGV515827H, NGV525132H, NGV514722H, NGV514723H, NGV523192H, NGV525278H, NGV525497H, NGV515810H, NGV516909H, NGV516921H, NGV520776H, NGV521121H, NGV523354H, NGV524249H, NGV524738H, NGV524794H, NGV525526H, NGV526687H, NGV527032H, NGV527739H, NGV527810H, NGV528029H, NGV514754H, NGV514444H, NGV514736H, NGV515306H, NGV516139H, NGV516855H, NGV519811H, NGV520320H, NGV523669H, NGV525633H, NGV518709H, NGV526304H, NGV514364H, NGV514947H, NGV514968H, NGV515057H, NGV515066H, NGV515818H, NGV516236H, NGV516825H, NGV519980H, NGV520325H, NGV520601H, NGV522972H, NGV518843H, NGV521092H, NGV515828H, NGV520541H, NGV523386H, NGV523208H, NGV515787H, NGV516638H, NGV516802H, NGV520875H, NGV525033H, NGV516814H, NGV519821H, NGV520430H, NGV520749H, NGV523298H, NGV515820H, NGV515824H, NGV523185H, NGV515437H, NGV519990H, NGV521182H, NGV523962H, NGV524245H, NGV515565H, NGV515804H, NGV515831H, NGV515838H, NGV520616H, NGV520617H, NGV520828H, NGV520837H, NGV523645H, NGV523653H, NGV524808H, NGV525619H, NGV517644H, NGV522302H, NGV524721H, NGV527267H, NGV515378H, NGV520414H, NGV524609H, NGV521051H, NGV526027H, NGV527284H, NGV519824H, NGV514967H, NGV520543H, NGV520548H, NGV520621H, NGV524890H, NGV526416H, NGV520502H, NGV527577H, NGV516101H, NGV522914H, NGV520118H, NGV522172H, NGV514319H, NGV520425H, NGV523477H, NGV524258H, NGV517649H, NGV517972H, NGV519117H, NGV519606H, NGV520333H, NGV523178H, NGV523686H, NGV518516H, NGV518656H, NGV519149H, NGV520310H, NGV520577H, NGV523700H, NGV524473H, NGV526968H, NGV514864H, NGV522660H, NGV525433H, NGV523372H, NGV524694H, NGV524836H, NGV514700H, NGV515963H, NGV516816H, NGV519503H, NGV520472H, NGV522120H, NGV523643H, NGV524180H, NGV525672H, NGV527082H, NGV514737H, NGV520582H, NGV522604H, NGV523832H, NGV524260H, NGV515396H, NGV521108H, NGV521163H, NGV522996H, NGV523265H, NGV524499H, NGV525071H, NGV526450H, NGV514714H, NGV514370H, NGV527585H, NGV514391H, NGV514432H, NGV525372H, NGV521220H, NGV523487H, NGV528245H, NGV514914H, NGV525117H, NGV525642H, NGV515839H, NGV518686H, NGV518696H, NGV518699H, NGV521145H, NGV524762H, NGV528349H, NGV523370H, NGV514883H, NGV514912H, NGV514924H, NGV518513H, NGV523892H, NGV517941H, NGV522949H, NGV519763H, NGV523604H, NGV525138H, NGV526988H, NGV527287H, NGV516786H, NGV517890H, NGV519994H, NGV522472H, NGV524251H, NGV524853H, NGV526385H, NGV514897H, NGV519433H, NGV520461H, NGV520912H, NGV522122H, NGV522963H, NGV522987H, NGV518617H, NGV520684H, NGV525564H, NGV526427H, NGV515620H, NGV517927H, NGV524865H, NGV514378H, NGV522104H, NGV523967H, NGV515292H, NGV514909H, NGV516768H, NGV516774H, NGV516775H, NGV517008H, NGV523028H, NGV523053H, NGV523102H, NGV523120H, NGV525062H, NGV525067H, NGV525637H, NGV526342H, NGV526842H, NGV526897H, NGV526913H, NGV523335H, NGV520245H, NGV515791H, NGV523703H, NGV516084H, NGV526426H, NGV526905H, NGV517640H, NGV515955H, NGV515957H, NGV518441H, NGV518918H, NGV519447H, NGV521176H, NGV521221H, NGV522409H, NGV525054H, NGV525112H, NGV516099H, NGV525718H, NGV520424H, NGV517574H, NGV514666H, NGV520068H, NGV520611H, NGV519751H, NGV516860H, NGV519950H, NGV525667H, NGV516120H, NGV516122H, NGV516125H, NGV516127H, NGV516129H, NGV515406H, NGV515424H, NGV516116H, NGV520507H, NGV520511H, NGV526791H, NGV514919H, NGV516070H, NGV516071H, NGV521181H, NGV521232H, NGV522141H, NGV522909H, NGV520897H, NGV520899H, NGV523225H, NGV523309H, NGV517686H, NGV517878H, NGV514427H, NGV524803H, NGV520483H, NGV518178H, NGV520792H, NGV525553H, NGV525557H, NGV525721H, NGV517674H, NGV519454H, NGV519482H, NGV516113H, NGV513543H, NGV520363H, NGV525717H, NGV526589H, NGV519967H, NGV523639H, NGV526398H, NGV522102H, NGV524646H, NGV525405H, NGV516140H, NGV519432H, NGV524202H, NGV522920H, NGV516886H, NGV516337H, NGV516339H, NGV517565H, NGV517891H, NGV519760H, NGV520552H, NGV523779H, NGV525701H, NGV514910H, NGV515074H, NGV525323H, NGV525531H, NGV514683H, NGV514904H, NGV516808H, NGV519758H, NGV520404H, NGV523245H, NGV523671H, NGV515331H, NGV515333H, NGV516615H, NGV517826H, NGV518390H, NGV518536H, NGV518541H, NGV518567H, NGV519505H, NGV520449H, NGV520572H, NGV522018H, NGV522185H, NGV522304H, NGV522496H, NGV523347H, NGV523634H, NGV523802H, NGV523894H, NGV525607H, NGV525613H, NGV525692H, NGV525699H, NGV525750H, NGV525700H, NGV522463H, NGV518088H, NGV521141H, NGV522107H, NGV522902H, NGV516143H, NGV514752H, NGV515649H, NGV522980H, NGV524437H, NGV524477H, NGV515966H, NGV516085H, NGV523222H, NGV516065H, NGV517562H, NGV520021H, NGV525124H, NGV514884H, NGV518091H, NGV515450H, NGV523534H, NGV524474H, NGV524801H, NGV519962H, NGV515841H, NGV526138H, NGV520568H, NGV514744H, NGV513755H; GTIN 00643169701014, Serial Numbers: NGV515305H, NGV515336H, NGV516409H, NGV519174H, NGV519435H, NGV522441H, NGV523260H, NGV523336H, NGV524988H, NGV525650H, NGV526336H, NGV526361H, NGV526961H, NGV514013H, NGV514667H, NGV514668H, NGV515352H, NGV515353H, NGV517752H, NGV519203H, NGV523302H, NGV527315H, NGV514716H, NGV516411H, NGV517698H, NGV517726H, NGV517887H, NGV518098H, NGV518116H, NGV519185H, NGV523306H, NGV525759H, NGV526698H, NGV527314H, NGV514303H, NGV514726H, NGV515266H, NGV515270H, NGV516704H, NGV517782H, NGV518141H, NGV518517H, NGV519139H, NGV519603H, NGV520013H, NGV520267H, NGV523281H, NGV523316H, NGV523371H, NGV524710H, NGV524956H, NGV526259H, NGV526689H, NGV526972H, NGV517754H, NGV518561H, NGV519153H, NGV519182H, NGV519460H, NGV522388H, NGV527568H, NGV528219H, NGV528267H, NGV528530H, NGV514676H, NGV515428H, NGV523648H, NGV514733H, NGV518540H, NGV518569H, NGV519144H, NGV523857H, NGV523865H, NGV524308H, NGV524818H, NGV526447H, NGV526613H, NGV515412H, NGV525310H, NGV526305H, NGV514414H, NGV522169H, NGV523203H, NGV523680H, NGV526452H, NGV517694H, NGV517944H, NGV518912H, NGV519309H, NGV524748H, NGV524839H, NGV526404H, NGV526731H, NGV514376H, NGV514911H, NGV515065H, NGV515075H, NGV517897H, NGV517911H, NGV519158H, NGV519168H, NGV519220H, NGV519293H, NGV521103H, NGV523221H, NGV523240H, NGV523360H, NGV523361H, NGV523444H, NGV523459H, NGV523571H, NGV524501H, NGV526424H, NGV526496H, NGV527552H, NGV527567H, NGV526910H, NGV514408H, NGV517781H, NGV523390H, NGV524939H, NGV517778H, NGV519299H, NGV524986H, NGV515342H, NGV515785H, NGV522391H, NGV523839H, NGV515432H, NGV528398H, NGV515126H, NGV515335H, NGV517787H, NGV519441H, NGV519620H, NGV523271H, NGV515269H, NGV517692H, NGV519200H, NGV523587H, NGV523702H, NGV524840H, NGV524852H, NGV524970H, NGV525568H, NGV528290H, NGV515136H, NGV519476H, NGV523062H, NGV525506H, NGV525778H, NGV526156H, NGV526445H, NGV514324H, NGV514333H, NGV514673H, NGV514678H, NGV514679H, NGV514684H, NGV515291H, NGV515302H, NGV515339H, NGV515788H, NGV515809H, NGV516671H, NGV517586H, NGV517766H, NGV518436H, NGV518454H, NGV519193H, NGV519296H, NGV519941H, NGV519989H, NGV520008H, NGV520018H, NGV521179H, NGV521196H, NGV521207H, NGV521235H, NGV522101H, NGV522163H, NGV523293H, NGV523295H, NGV523350H, NGV523357H, NGV523609H, NGV523623H, NGV523635H, NGV524314H, NGV524844H, NGV525042H, NGV525079H, NGV525674H, NGV525719H, NGV526357H, NGV526662H, NGV527590H, NGV527616H, NGV528362H, NGV528392H, NGV517738H, NGV527131H, NGV515121H, NGV515286H, NGV515381H, NGV517913H, NGV521076H, NGV522394H, NGV522936H, NGV523586H, NGV523708H, NGV524809H, NGV524995H, NGV525511H, NGV526541H, NGV526638H, NGV515784H, NGV519753H, NGV522183H, NGV522743H, NGV520734H, NGV515347H, NGV519307H, NGV521169H, NGV523449H, NGV524727H, NGV517675H, NGV520248H, NGV521738H, NGV523291H, NGV524719H, NGV525657H, NGV525775H, NGV527594H, NGV514298H, NGV516444H, NGV518603H, NGV524885H, NGV525186H, NGV513991H, NGV515326H, NGV521152H, NGV521159H, NGV523665H, NGV524886H, NGV525320H, NGV525676H, NGV525677H, NGV525763H, NGV528401H, NGV522901H, NGV515440H, NGV518842H, NGV519136H, NGV517734H, NGV517735H, NGV519303H, NGV519304H, NGV523856H, NGV515090H, NGV515093H, NGV515520H, NGV517930H, NGV523811H, NGV524704H, NGV525080H, NGV525087H, NGV522998H, NGV526363H, NGV513906H, NGV514672H, NGV514715H, NGV515310H, NGV515755H, NGV517661H, NGV522399H, NGV522439H, NGV523391H, NGV526958H, NGV519831H, NGV525516H, NGV525586H, NGV526167H, NGV514692H, NGV514693H, NGV515384H, NGV515400H, NGV515405H, NGV516748H, NGV518577H, NGV519634H, NGV519764H, NGV519786H, NGV521080H, NGV521081H, NGV524423H, NGV524871H, NGV523399H, NGV520006H, NGV514694H, NGV514697H, NGV518117H; GTIN 00643169701021, Serial Numbers: NGV516406H, NGV516925H, NGV518391H, NGV526381H, NGV526410H, NGV514758H, NGV514760H, NGV516268H, NGV516713H, NGV516744H, NGV516836H, NGV517926H, NGV517959H, NGV518591H, NGV523290H, NGV523348H, NGV524784H, NGV524811H, NGV524824H, NGV516650H, NGV516781H, NGV517299H, NGV517570H, NGV519570H, NGV520460H, NGV522676H, NGV524669H, NGV524764H, NGV524856H, NGV525471H, NGV527279H, NGV514805H, NGV515779H, NGV515780H, NGV517886H, NGV517908H, NGV520223H, NGV520535H, NGV520540H, NGV520606H, NGV521058H, NGV521087H, NGV521112H, NGV522128H, NGV522907H, NGV522912H, NGV524696H, NGV524742H, NGV524868H, NGV525244H, NGV525527H, NGV527439H, NGV513960H, NGV513964H, NGV515720H, NGV515766H, NGV521135H, NGV524640H, NGV524826H, NGV525431H, NGV525525H, NGV514806H, NGV514820H, NGV515017H, NGV515133H, NGV515832H, NGV516823H, NGV517572H, NGV517578H, NGV517592H, NGV518438H, NGV520471H, NGV520525H, NGV520821H, NGV522976H, NGV523526H, NGV524698H, NGV524735H, NGV525006H, NGV525399H, NGV525419H, NGV525462H, NGV514766H, NGV514771H, NGV514794H, NGV518597H, NGV522925H, NGV522956H, NGV522970H, NGV522981H, NGV524780H, NGV524879H, NGV524996H, NGV520526H, NGV513959H, NGV514585H, NGV514734H, NGV515801H, NGV515845H, NGV516769H, NGV516949H, NGV518614H, NGV520204H, NGV522139H, NGV522974H, NGV523238H, NGV523363H, NGV523377H, NGV523395H, NGV524876H, NGV525435H, NGV525574H, NGV518476H, NGV519222H, NGV523729H, NGV514732H, NGV514990H, NGV515010H, NGV515346H, NGV515608H, NGV523161H, NGV524823H, NGV525209H, NGV516772H, NGV517643H, NGV522131H, NGV525475H, NGV514729H, NGV515847H, NGV516018H, NGV516041H, NGV516700H, NGV520763H, NGV520798H, NGV514558H, NGV514609H, NGV515781H, NGV516778H, NGV520646H, NGV520762H, NGV521756H, NGV522066H, NGV522953H, NGV523624H, NGV524702H, NGV524717H, NGV524819H, NGV525002H, NGV525259H, NGV525573H, NGV526512H, NGV514743H, NGV516813H, NGV516828H, NGV516838H, NGV517666H, NGV519126H, NGV519145H, NGV520822H, NGV520934H, NGV527269H, NGV514801H, NGV516773H, NGV520936H, NGV521139H, NGV521170H, NGV524701H, NGV516342H, NGV524647H, NGV514831H, NGV514839H, NGV523663H, NGV523676H, NGV524599H, NGV524604H, NGV524833H, NGV525555H, NGV513956H, NGV514727H, NGV514777H, NGV514811H, NGV514813H, NGV514818H, NGV515823H, NGV516703H, NGV517609H, NGV517816H, NGV518394H, NGV518630H, NGV519810H, NGV523540H, NGV523773H, NGV523798H, NGV523814H, NGV525442H, NGV514595H, NGV514608H, NGV514614H, NGV515834H, NGV515835H, NGV515998H, NGV516652H, NGV516660H, NGV516767H, NGV516797H, NGV517663H, NGV517664H, NGV520642H, NGV521979H, NGV515148H, NGV515301H, NGV520919H, NGV524897H, NGV525268H, NGV525571H, NGV527285H, NGV514027H, NGV513982H, NGV514043H, NGV518483H, NGV522127H, NGV520482H, NGV520521H, NGV521054H, NGV524991H, NGV524992H, NGV514623H, NGV514711H, NGV522603H, NGV523023H, NGV524821H, NGV524896H, NGV514695H, NGV514703H, NGV514843H, NGV515842H, NGV524529H, NGV524552H, NGV528340H, NGV519808H, NGV521056H, NGV521127H, NGV521130H, NGV521195H, NGV525174H, NGV525178H, NGV525198H, NGV515837H, NGV514718H, NGV514720H, NGV514853H, NGV514926H, NGV516837H, NGV519756H, NGV519771H, NGV519783H, NGV519796H, NGV519799H, NGV516839H, NGV516851H, NGV516861H, NGV516894H, NGV523600H, NGV523605H; GTIN 00643169701038, Serial Numbers: NGV514768H, NGV519465H, NGV522695H, NGV527531H, NGV514763H, NGV520831H, NGV526310H, NGV526917H, NGV526983H, NGV527229H, NGV526551H, NGV526272H, NGV520802H, NGV519462H, NGV523067H, NGV525977H, NGV514789H, NGV525496H, NGV526566H, NGV514796H, NGV519515H, NGV520799H, NGV526532H, NGV514756H, NGV514780H, NGV514788H, NGV519316H, NGV520273H, NGV525492H, NGV526285H, NGV526337H, NGV526567H, NGV526984H, NGV527226H, NGV527453H, NGV514759H, NGV527039H, NGV514660H, NGV514764H, NGV520283H, NGV525333H, NGV526511H, NGV519817H, NGV520811H, NGV522115H, NGV525761H, NGV514664H, NGV516859H, NGV519427H, NGV523170H, NGV520880H, NGV516771H, NGV520637H, NGV520926H, NGV520933H, NGV522950H, NGV516853H, NGV516854H, NGV520490H, NGV520805H, NGV520907H, NGV520922H, NGV514690H, NGV519792H, NGV524398H, NGV522926H, NGV524375H, NGV515320H, NGV514671H, NGV514980H, NGV514611H, NGV515563H, NGV520939H, NGV516832H, NGV515341H, NGV515999H, NGV520342H, NGV520426H, NGV523147H, NGV523778H, NGV524372H, NGV516636H, NGV516940H, NGV524639H, NGV514375H, NGV514686H, NGV519273H, NGV514704H, NGV514705H, NGV514974H, NGV515276H, NGV515985H, NGV518121H, NGV519430H, NGV519502H, NGV520440H, NGV520796H, NGV520896H, NGV522130H, NGV523183H, NGV523212H, NGV523791H, NGV525736H, NGV525740H, NGV514675H, NGV520699H, NGV514583H, NGV520765H, NGV524364H, NGV514665H, NGV516777H, NGV516787H, NGV523325H, NGV520467H, NGV514670H, NGV516645H, NGV521252H, NGV514599H, NGV519423H, NGV515332H, NGV524402H, NGV514730H, NGV514735H, NGV514751H, NGV514753H, NGV515315H, NGV515334H, NGV515647H, NGV516242H, NGV516271H, NGV516648H, NGV516834H, NGV518096H, NGV518122H, NGV519803H, NGV519818H, NGV519823H, NGV520036H, NGV520317H, NGV521217H, NGV521222H, NGV522129H, NGV522616H, NGV522942H, NGV523269H, NGV523297H, NGV524337H, NGV524420H, NGV524617H, NGV516262H, NGV518158H, NGV520941H, NGV514661H, NGV515004H, NGV516649H, NGV516923H, NGV518113H, NGV523158H, NGV524631H, NGV514610H, NGV514707H, NGV516231H, NGV516270H, NGV516658H, NGV516799H, NGV520455H, NGV523174H, NGV524633H, NGV514593H, NGV524346H, NGV515559H, NGV516829H, NGV514564H, NGV515648H, NGV516954H, NGV520347H, NGV521334H, NGV521342H, NGV522959H, NGV524326H, NGV524400H, NGV514689H, NGV518123H, NGV520942H, NGV520945H, NGV523782H, NGV525749H, NGV514021H, NGV519833H, NGV520678H, NGV520946H, NGV521282H, NGV524316H, NGV524389H, NGV524666H, NGV525288H, NGV525344H, NGV522971H, NGV515350H, NGV520338H, NGV521225H, NGV524331H, NGV522126H, NGV515323H, NGV519281H, NGV519313H, NGV522195H, NGV524396H, NGV520761H, NGV518132H, NGV514685H, NGV519271H, NGV519838H, NGV524320H, NGV524641H, NGV516224H, NGV516848H, NGV518115H, NGV519274H, NGV519789H, NGV519836H, NGV520784H, NGV520795H, NGV520827H, NGV524695H, NGV514042H, NGV514987H, NGV524279H, NGV521224H, NGV524651H, NGV520847H, NGV524634H, NGV515581H, NGV521233H, NGV522989H, NGV525747H, NGV516795H, NGV520431H, NGV514717H, NGV515322H, NGV516251H, NGV516762H, NGV519280H, NGV519778H, NGV519814H, NGV520256H, NGV520469H, NGV520812H, NGV522132H, NGV522623H, NGV524323H, NGV524354H, NGV524391H, NGV516031H, NGV514384H, NGV514710H, NGV516820H, NGV516874H, NGV516770H, NGV514604H, NGV516840H, NGV519835H, NGV522928H, NGV519431H, NGV524675H, NGV519284H, NGV519828H, NGV524447H, NGV517772H, NGV519491H, NGV519463H, NGV522184H, NGV524894H, NGV525561H, NGV525602H, NGV525604H, NGV525610H, NGV525614H, NGV525695H, NGV516581H, NGV519311H, NGV519827H, NGV524657H, NGV514308H, NGV519287H, NGV519301H, NGV519426H, NGV525634H, NGV525704H, NGV525623H, NGV514295H, NGV519294H, NGV519500H, NGV522188H, NGV519490H, NGV524259H, NGV519446H, NGV525638H, NGV518097H, NGV519497H, NGV519425H, NGV524446H, NGV525616H, NGV525618H, NGV525708H, NGV523699H, NGV525668H, NGV519434H, NGV519487H, NGV519691H, NGV519477H, NGV519488H, NGV519506H, NGV519496H, NGV519494H, NGV519508H, NGV525270H, NGV514350H, NGV524491H, NGV519442H, NGV519424H, NGV515716H, NGV516634H, NGV519469H, NGV519512H, NGV519819H, NGV524494H, NGV524655H, NGV525693H, NGV525665H, NGV519499H, NGV519495H, NGV519310H, NGV514328H, NGV515726H, NGV516627H, NGV516637H, NGV517765H, NGV517768H, NGV518100H, NGV519437H, NGV519601H, NGV522182H, NGV523704H, NGV523713H, NGV524248H, NGV524276H, NGV524421H, NGV524424H, NGV524887H, NGV524892H, NGV525271H, NGV516612H, NGV514371H, NGV522397H, NGV522410H, NGV513920H, NGV513928H, NGV515071H, NGV520362H, NGV525028H, NGV525364H, NGV514625H, NGV515591H, NGV520865H, NGV522651H, NGV516663H, NGV519517H, NGV521156H, NGV519574H, NGV520874H, NGV519584H, NGV520836H, NGV520846H, NGV522417H, NGV525587H, NGV525591H, NGV518146H, NGV518152H, NGV519516H, NGV519519H, NGV519551H, NGV519858H, NGV520262H, NGV520843H, NGV520844H, NGV522644H, NGV525352H, NGV519561H, NGV516566H, NGV519597H, NGV514419H, NGV514425H, NGV514436H, NGV514645H, NGV515628H, NGV516863H, NGV521150H, NGV521151H, NGV525601H, NGV514430H, NGV519576H, NGV520259H, NGV522641H, NGV525465H, NGV525388H, NGV516137H, NGV516138H, NGV516806H, NGV516827H, NGV521146H, NGV521147H, NGV522655H, NGV522675H, NGV515669H, NGV516867H, NGV516150H, NGV516152H, NGV519591H, NGV520862H, NGV521171H, NGV516578H, NGV520260H, NGV514640H, NGV522450H, NGV514624H, NGV516831H, NGV514390H, NGV515560H, NGV525338H, NGV514412H, NGV520559H, NGV520590H, NGV520619H, NGV522916H, NGV526371H, NGV521128H, NGV527830H, NGV516917H, NGV528403H, NGV518166H, NGV514283H, NGV518181H, NGV520793H, NGV516582H, NGV516623H, NGV516864H, NGV519801H, NGV519812H, NGV520517H, NGV523825H, NGV525329H, NGV528093H, NGV518160H, NGV522659H, NGV514050H, NGV514052H, NGV514137H, NGV514292H, NGV514293H, NGV515758H, NGV515962H, NGV516072H, NGV516077H, NGV518159H, NGV518163H, NGV518173H, NGV519805H, NGV520768H, NGV520785H, NGV520859H, NGV525031H, NGV525281H, NGV525766H, NGV525770H, NGV525779H, NGV526503H, NGV528332H, NGV515770H, NGV515817H, NGV518155H, NGV519802H, NGV520786H, NGV522657H, NGV525771H, NGV523799H, NGV523818H, NGV515777H, NGV518154H, NGV523823H, NGV514454H, NGV514455H, NGV514474H, NGV514499H, NGV514560H, NGV514561H, NGV514563H, NGV514584H, NGV514586H, NGV514607H, NGV514616H, NGV514655H, NGV514656H, NGV514658H, NGV515773H, NGV515775H, NGV515786H, NGV515789H, NGV515792H, NGV515811H, NGV515825H, NGV515829H, NGV515853H, NGV516007H, NGV516093H, NGV516109H, NGV516114H, NGV516117H, NGV518461H, NGV518477H, NGV518497H, NGV518688H, NGV518695H, NGV518717H, NGV518832H, NGV518848H, NGV518871H, NGV518875H, NGV518893H, NGV518909H, NGV519740H, NGV519746H, NGV519747H, NGV519750H, NGV519754H, NGV519755H, NGV519761H, NGV519765H, NGV519766H, NGV519768H, NGV519769H, NGV519770H, NGV519772H, NGV519773H, NGV519775H, NGV519779H, NGV519780H, NGV519781H, NGV519784H, NGV519787H, NGV519788H, NGV513905H, NGV515137H, NGV517985H, NGV523959H, NGV523963H, NGV523966H, NGV525127H, NGV525135H, NGV525165H, NGV513932H, NGV518112H, NGV519183H, NGV519793H, NGV520421H, NGV524069H, NGV524622H, NGV514836H, NGV515089H, NGV515092H, NGV515095H, NGV515096H, NGV515097H, NGV519197H, NGV519208H, NGV519216H, NGV521125H, NGV524178H, NGV524182H, NGV524211H, NGV524215H, NGV525055H, NGV525162H, NGV524635H, NGV514357H, NGV515007H, NGV516876H, NGV523110H, NGV514854H, NGV514842H, NGV514374H, NGV514385H, NGV514833H, NGV516910H, NGV514358H, NGV516897H, NGV515139H, NGV515149H, NGV516197H, NGV516199H, NGV516215H, NGV516218H, NGV516780H, NGV517647H, NGV517659H, NGV517668H, NGV520550H, NGV520551H, NGV525709H, NGV525673H, NGV513938H, NGV514005H, NGV516948H, NGV513940H, NGV513935H, NGV520779H, NGV514038H, NGV514039H, NGV514669H, NGV515277H, NGV515283H, NGV516229H, NGV522633H, NGV522637H, NGV524289H, NGV515297H, NGV516225H, NGV523127H, NGV525267H, NGV518824H, NGV519127H, NGV519129H, NGV519132H, NGV519135H, NGV517681H, NGV517682H, NGV517685H, NGV517704H, NGV517710H, NGV517724H, NGV517729H, NGV517912H, NGV524220H, NGV514406H, NGV517322H, NGV525221H, NGV525193H, NGV525199H, NGV525204H, NGV525211H, NGV521043H, NGV525064H, NGV525082H, NGV525110H, NGV525115H, NGV525116H, NGV525120H, NGV525248H, NGV525252H, NGV514701H, NGV514651H, NGV528334H, NGV514674H, NGV519832H, NGV522158H, NGV528087H, NGV516605H, NGV514699H, NGV519456H, NGV519830H, NGV520291H, NGV520755H, NGV521172H, NGV521177H, NGV523796H, NGV523797H, NGV525330H, NGV516604H, NGV516610H, NGV518511H, NGV518524H, NGV519317H, NGV514702H, NGV514721H, NGV514728H, NGV514747H, NGV514750H, NGV514845H, NGV514847H, NGV517749H, NGV519658H, NGV520188H, NGV520300H, NGV520707H, NGV520710H, NGV520714H, NGV520760H, NGV520769H, NGV520770H, NGV521191H, NGV522151H, NGV522160H, NGV525675H, NGV525678H, NGV525711H, NGV525727H, NGV527624H, NGV528076H, NGV514687H, NGV516600H, NGV525716H, NGV527592H, NGV527757H, NGV516585H, NGV527753H, NGV514713H, NGV514742H, NGV516598H, NGV518415H, NGV518443H, NGV519453H, NGV521194H, NGV521210H, NGV522075H, NGV522135H, NGV521199H, NGV519449H, NGV516602H, NGV517730H, NGV519829H, NGV522133H, NGV522146H, NGV520712H, NGV516596H, NGV514719H, NGV518502H, NGV525645H, NGV522150H, NGV520754H, NGV522178H, NGV518161H, NGV522155H, NGV522161H, NGV524997H, NGV528331H, NGV516209H, NGV514838H, NGV519171H, NGV524831H, NGV523905H, NGV520396H, NGV514663H, NGV515583H, NGV516115H, NGV518087H, NGV520194H, NGV520361H, NGV524527H, NGV515135H, NGV518095H, NGV520030H, NGV519191H, NGV522317H, NGV516063H, NGV520154H, NGV514745H, NGV516074H, NGV517951H, NGV518439H, NGV519161H, NGV519189H, NGV519205H, NGV521787H, NGV522186H, NGV523929H, NGV524810H, NGV524843H, NGV525190H, NGV525245H, NGV514356H, NGV514448H, NGV514568H, NGV514571H, NGV514657H, NGV516076H, NGV516128H, NGV516357H, NGV519166H, NGV519170H, NGV519178H, NGV519181H, NGV521082H, NGV521085H, NGV521118H, NGV522000H, NGV524850H, NGV524858H, NGV524866H, NGV525175H, NGV525242H, NGV525260H, NGV526379H, NGV526401H, NGV526420H, NGV526451H, NGV514680H, NGV517326H, NGV522067H, NGV523951H, NGV519244H, NGV514784H, NGV519800H, NGV525123H, NGV516168H, NGV516160H, NGV514814H, NGV519795H, NGV522357H, NGV524870H, NGV523942H, NGV514688H, NGV515059H, NGV515606H, NGV521123H, NGV514355H, NGV524573H, NGV514446H, NGV514592H, NGV517604H, NGV518080H, NGV522326H, NGV523775H, NGV523777H, NGV525182H, NGV525842H, NGV520191H, NGV523878H, NGV516155H, NGV516202H, NGV523918H, NGV523954H, NGV514696H, NGV520170H, NGV516189H, NGV514790H, NGV523941H, NGV524909H, NGV514807H, NGV514841H, NGV516087H, NGV516108H, NGV524848H, NGV519199H, NGV514792H, NGV519797H, NGV522461H, NGV528055H, NGV519217H, NGV516045H, NGV514613H, NGV517639H, NGV522915H, NGV520220H, NGV514650H, NGV520195H, NGV523943H, NGV528086H, NGV513866H, NGV518691H, NGV519152H, NGV516121H, NGV515087H, NGV516164H, NGV519165H, NGV514817H, NGV514739H, NGV514895H, NGV519186H, NGV524787H, NGV524793H, NGV527544H, NGV522100H, NGV528407H, NGV517556H, NGV517561H, NGV520547H, NGV525237H, NGV514555H, NGV516060H, NGV516156H, NGV516166H, NGV517598H, NGV517620H, NGV517624H, NGV528341H, NGV515371H, NGV519221H, NGV519483H, NGV523613H, NGV513865H, NGV515103H, NGV515105H, NGV516132H, NGV519192H, NGV520004H, NGV520359H, NGV520369H, NGV520385H, NGV520614H, NGV522905H, NGV523740H, NGV523614H, NGV514434H, NGV514649H, NGV519150H, NGV519957H, NGV523767H, NGV515588H, NGV515604H, NGV515152H, NGV515259H, NGV515260H, NGV515261H, NGV515265H, NGV515267H, NGV515271H, NGV524557H, NGV524569H, NGV516186H, NGV516192H, NGV520190H, NGV522109H, NGV523755H, NGV523762H, NGV524683H, NGV524684H, NGV515579H, NGV514437H, NGV523472H, NGV514907H, NGV519151H, NGV515080H, NGV515997H, NGV522923H, NGV522995H, NGV523063H, NGV516172H, NGV514886H, NGV515455H, NGV515575H, NGV517914H, NGV517935H, NGV520197H, NGV522114H, NGV515272H, NGV523862H, NGV515061H, NGV515063H, NGV515070H, NGV516170H, NGV516418H, NGV516790H, NGV516812H, NGV517551H, NGV517569H, NGV517588H, NGV519259H, NGV519260H, NGV519265H, NGV519269H, NGV520390H, NGV520413H, NGV523638H, NGV523667H, NGV524584H, NGV524605H, NGV524606H, NGV524611H, NGV524619H, NGV526540H, NGV526552H, NGV527533H, NGV527537H, NGV527609H, NGV527614H, NGV527619H, NGV524503H, NGV524507H, NGV522112H, NGV515968H, NGV515990H, NGV520355H, NGV521136H, NGV523737H, NGV524709H, NGV524773H, NGV521138H, NGV520233H, NGV518419H, NGV519468H, NGV514416H, NGV514442H, NGV519206H, NGV523400H, NGV523461H, NGV523691H, NGV526455H, NGV527578H, NGV527584H, NGV527694H, NGV515120H, NGV515122H, NGV515123H, NGV517638H, NGV520557H, NGV525829H, NGV519201H, NGV520628H, NGV520703H, NGV515100H, NGV516124H, NGV516180H, NGV514413H, NGV520597H, NGV516376H; GTIN 00643169701045, Serial Numbers: NGV514731H, NGV515385H, NGV516352H, NGV523341H, NGV527798H, NGV514866H, NGV514889H, NGV515263H, NGV516310H, NGV519190H, NGV526362H, NGV526390H, NGV526411H, NGV518107H, NGV515300H, NGV515992H, NGV518094H, NGV520826H, NGV526279H, NGV526446H, NGV519202H, NGV524707H, NGV524875H, NGV525554H, NGV526372H, NGV526923H, NGV528355H, NGV528363H, NGV518165H, NGV518180H, NGV518862H, NGV515303H, NGV515304H, NGV515964H, NGV516426H, NGV516655H, NGV517680H, NGV518105H, NGV518133H, NGV520434H, NGV521845H, NGV523541H, NGV523619H, NGV523817H, NGV525640H, NGV525646H, NGV526299H, NGV526902H, NGV526967H, NGV516591H, NGV516618H, NGV514748H, NGV516459H, NGV518090H, NGV518102H, NGV525624H, NGV526428H, NGV514421H, NGV519229H, NGV525832H, NGV516800H, NGV516804H, NGV516805H, NGV516846H, NGV516877H, NGV516889H, NGV516916H, NGV516929H, NGV525495H, NGV525594H, NGV526600H, NGV514851H, NGV515343H, NGV515348H, NGV518126H, NGV518127H, NGV518130H, NGV518137H, NGV518139H, NGV522918H, NGV523126H, NGV525805H, NGV525872H, NGV526645H, NGV526832H, NGV525579H, NGV515318H, NGV515446H, NGV515774H, NGV515795H, NGV515843H, NGV516035H, NGV516043H, NGV516068H, NGV516368H, NGV517564H, NGV524805H, NGV525489H, NGV525764H, NGV526065H, NGV526831H, NGV526955H, NGV528411H, NGV516370H, NGV525481H, NGV515763H, NGV518124H, NGV520758H, NGV525121H; GTIN 00643169701052, Serial Numbers: NGV514662H, NGV514861H, NGV514882H, NGV514923H, NGV515275H, NGV515294H, NGV515316H, NGV515327H, NGV515337H, NGV515345H, NGV515354H, NGV516765H, NGV516784H, NGV516842H, NGV516956H, NGV517958H, NGV517973H, NGV518099H, NGV518537H, NGV518549H, NGV518592H, NGV518599H, NGV523135H, NGV523346H, NGV524000H, NGV524059H, NGV527161H, NGV527190H, NGV527195H, NGV527208H, NGV527257H, NGV527276H, NGV527292H, NGV527318H, NGV514401H, NGV515284H, NGV515312H, NGV515752H, NGV515760H, NGV516943H, NGV516952H, NGV516955H, NGV517779H, NGV517916H, NGV518563H, NGV520315H, NGV520462H, NGV520824H, NGV521185H, NGV521209H, NGV523914H, NGV523955H, NGV524240H, NGV524368H, NGV524395H, NGV524827H, NGV524847H, NGV526867H, NGV526868H, NGV515282H, NGV515609H, NGV515619H, NGV515749H, NGV515794H, NGV516926H, NGV518017H, NGV518052H, NGV518157H, NGV518171H, NGV520573H, NGV520766H, NGV520834H, NGV520838H, NGV520840H, NGV522194H, NGV522605H, NGV522663H, NGV522964H, NGV523103H, NGV523105H, NGV526276H, NGV526646H, NGV526704H, NGV526749H, NGV526753H, NGV526794H, NGV526948H, NGV526959H, NGV526994H, NGV527110H, NGV527132H, NGV527174H, NGV527323H, NGV528246H, NGV528310H, NGV514881H, NGV514932H, NGV514943H, NGV515295H, NGV515389H, NGV515415H, NGV515418H, NGV515427H, NGV521114H, NGV521119H, NGV521214H, NGV521329H, NGV522125H, NGV522152H, NGV525500H, NGV525558H, NGV525644H, NGV525648H, NGV525669H, NGV526269H, NGV524760H, NGV526654H, NGV516059H, NGV516146H, NGV516175H, NGV516633H, NGV516807H, NGV517722H, NGV517733H, NGV518067H, NGV518177H, NGV520833H, NGV521766H, NGV522168H, NGV522380H, NGV522390H, NGV524693H, NGV524750H, NGV525498H, NGV526843H, NGV527154H, NGV515329H, NGV516590H, NGV516603H, NGV517963H, NGV518005H, NGV527543H, NGV527547H, NGV516606H, NGV516631H, NGV517714H, NGV524873H, NGV514381H, NGV516089H, NGV518184H, NGV523130H; GTIN 00643169701069, Serial Numbers: NGV518685H, NGV518707H, NGV517119H, NGV517351H, NGV517423H, NGV517531H, NGV517550H, NGV517696H, NGV517701H, NGV517716H, NGV517717H, NGV517721H, NGV517725H, NGV519283H, NGV519291H, NGV519298H, NGV523806H, NGV523829H, NGV523842H, NGV523910H, NGV524661H, NGV524825H, NGV525139H, NGV525534H, NGV525986H, NGV526328H, NGV526611H, NGV526644H, NGV526650H, NGV526655H, NGV526673H, NGV526677H, NGV527119H, NGV527125H, NGV527142H, NGV527149H, NGV527197H, NGV527223H, NGV527283H, NGV527311H, NGV527321H, NGV527780H, NGV515765H, NGV515767H, NGV515769H, NGV516055H, NGV516057H, NGV516066H, NGV516069H, NGV516081H, NGV516103H, NGV516118H, NGV516151H, NGV516154H, NGV516570H, NGV516572H, NGV516611H, NGV516621H, NGV516622H, NGV516625H, NGV516630H, NGV516643H, NGV516651H, NGV516883H, NGV516922H, NGV516950H, NGV517301H, NGV517303H, NGV517310H, NGV517607H, NGV517650H, NGV517687H, NGV517690H, NGV517693H, NGV517695H, NGV517728H, NGV517736H, NGV517750H, NGV517755H, NGV517761H, NGV517773H, NGV517774H, NGV517783H, NGV517784H, NGV517785H, NGV518413H, NGV518463H, NGV518711H, NGV518830H, NGV518847H, NGV518879H, NGV518891H, NGV518902H, NGV518906H, NGV518913H, NGV519263H, NGV519546H, NGV519749H, NGV519757H, NGV519807H, NGV522631H, NGV519305H, NGV519536H, NGV522653H, NGV523800H, NGV523912H, NGV524708H, NGV524863H, NGV525378H, NGV525400H, NGV525456H, NGV519822H; GTIN 00643169701076, Serial Numbers: NGV516421H, NGV521111H, NGV522136H, NGV523602H, NGV525572H, NGV526295H, NGV526353H, NGV527411H, NGV522943H, NGV516673H, NGV515296H, NGV516562H, NGV516583H, NGV518183H, NGV519238H, NGV523567H, NGV523651H, NGV524502H, NGV524758H, NGV526516H, NGV516220H, NGV516243H, NGV522137H, NGV522316H, NGV522347H, NGV514915H, NGV517928H, NGV517936H, NGV518111H, NGV521064H, NGV527735H, NGV525269H, NGV525724H, NGV525732H, NGV515806H, NGV515807H, NGV516810H, NGV521167H, NGV523367H, NGV526919H, NGV515632H, NGV515640H, NGV515652H, NGV520409H, NGV521226H, NGV522373H, NGV523660H, NGV525487H, NGV525556H, NGV525741H, NGV526384H, NGV526478H, NGV514891H, NGV514922H, NGV518167H, NGV519533H, NGV520750H, NGV523237H, NGV523642H, NGV524629H, NGV525052H, NGV525197H, NGV526442H, NGV516285H, NGV521069H, NGV521077H, NGV521109H, NGV524725H, NGV524796H, NGV524889H, NGV526414H; GTIN 00643169701083, Serial Numbers: NGV515014H, NGV516586H, NGV520172H, NGV525390H, NGV525463H, NGV514986H, NGV515833H, NGV520360H, NGV526640H, NGV526649H, NGV526652H, NGV526726H, NGV526741H, NGV526747H, NGV526757H, NGV527109H, NGV527114H, NGV527207H, NGV527222H, NGV527224H, NGV514477H, NGV514557H, NGV515000H, NGV515049H, NGV515051H, NGV520189H, NGV520205H, NGV520354H, NGV520356H, NGV525982H, NGV526023H, NGV514478H, NGV514485H, NGV514522H, NGV514643H, NGV514677H, NGV514779H, NGV514991H, NGV514993H, NGV515015H, NGV515020H, NGV515028H, NGV515308H, NGV520365H, NGV520384H, NGV520394H, NGV520403H, NGV520417H, NGV520418H, NGV520420H, NGV520427H, NGV520428H, NGV520429H, NGV520433H, NGV520870H, NGV520891H, NGV520931H, NGV526024H, NGV526036H, NGV526050H, NGV526604H, NGV519752H, NGV520400H, NGV520441H, NGV520925H, NGV523284H, NGV523916H, NGV524768H, NGV524842H, NGV518700H, NGV517910H, NGV527302H, NGV516601H, NGV520920H, NGV520923H; GTIN 00643169701090, Serial Numbers: NGV525224H, NGV515111H, NGV526213H, NGV517938H, NGV526935H, NGV515129H, NGV515099H, NGV517947H, NGV525005H, NGV514681H, NGV526943H, NGV524807H, NGV517656H, NGV520110H, NGV525576H, NGV522113H, NGV524857H, NGV524882H, NGV525114H, NGV514691H, NGV521219H, NGV525231H, NGV525236H, NGV525753H, NGV526502H, NGV526074H, NGV521047H, NGV521327H, NGV518442H, NGV520304H, NGV528438H, NGV528448H, NGV552485H, NGV524479H, NGV521198H, NGV527172H, NGV528266H, NGV528482H, NGV521075H, NGV522095H, NGV522111H, NGV522196H, NGV517279H, NGV521059H, NGV526945H, NGV552462H, NGV519820H, NGV528239H, NGV520674H, NGV524691H, NGV517296H, NGV528318H, NGV525096H, NGV526765H, NGV528501H, NGV517566H, NGV519785H, NGV521771H, NGV518445H, NGV523229H, NGV523307H, NGV524347H, NGV528463H, NGV518414H, NGV527561H, NGV528517H, NGV518398H, NGV517284H, NGV518826H, NGV519837H, NGV523332H, NGV524432H, NGV523326H, NGV519602H, NGV521048H, NGV518675H, NGV517287H, NGV528434H, NGV521046H, NGV522967H, NGV527002H; GTIN 00643169732346, Serial Numbers: NGV551684H, NGV552450H, NGV529870H, NGV550544H, NGV520889H, NGV520906H, NGV528054H, NGV529501H, NGV550904H, NGV551613H, NGV552550H, NGV551018H, NGV553004H, NGV529125H, NGV529207H, NGV530409H, NGV550770H, NGV552155H, NGV529929H, NGV520809H, NGV530284H, NGV530779H, NGV530837H, NGV550351H, NGV552122H, NGV552125H, NGV552148H, NGV552154H, NGV528955H, NGV550056H, NGV550572H, NGV551483H, NGV551648H, NGV552386H, NGV553033H, NGV529347H, NGV529987H, NGV530511H, NGV530945H, NGV550947H, NGV551675H, NGV552355H, NGV552371H, NGV528592H, NGV551526H, NGV529773H, NGV529943H, NGV516212H, NGV528969H, NGV530044H, NGV530364H, NGV530660H, NGV552365H, NGV529066H, NGV529098H, NGV529157H, NGV530279H, NGV550523H, NGV551651H, NGV530312H, NGV530465H, NGV551327H, NGV552149H, NGV552276H, NGV552346H, NGV530944H, NGV552266H, NGV529085H, NGV551557H, NGV529260H, NGV529341H, NGV530936H, NGV530938H, NGV551131H, NGV551160H, NGV551161H, NGV551168H, NGV553046H, NGV552510H, NGV553018H, NGV530001H, NGV530337H, NGV550954H, NGV552080H, NGV528178H, NGV528992H, NGV550898H, NGV551110H, NGV526839H, NGV529091H, NGV530295H, NGV530464H, NGV530670H, NGV551084H, NGV551119H, NGV552105H, NGV515783H, NGV516226H, NGV516253H, NGV524638H, NGV528127H, NGV528137H, NGV528144H, NGV528154H, NGV528510H, NGV528604H, NGV528770H, NGV528783H, NGV528989H, NGV528993H, NGV529056H, NGV529108H, NGV529359H, NGV529384H, NGV529452H, NGV529860H, NGV529898H, NGV529925H, NGV529952H, NGV530023H, NGV530056H, NGV530061H, NGV530067H, NGV530294H, NGV530321H, NGV530351H, NGV530396H, NGV530454H, NGV530460H, NGV530700H, NGV530713H, NGV530715H, NGV530716H, NGV530728H, NGV530834H, NGV530882H, NGV530917H, NGV550031H, NGV550096H, NGV550105H, NGV550301H, NGV551007H, NGV551121H, NGV551128H, NGV551643H, NGV551653H, NGV552126H, NGV552238H, NGV552246H, NGV552254H, NGV552381H, NGV552383H, NGV552465H, NGV552470H, NGV552491H, NGV552498H, NGV552517H, NGV553053H, NGV553054H, NGV530322H, NGV530374H, NGV552439H, NGV552528H, NGV529182H, NGV530401H, NGV530663H, NGV530854H, NGV552169H, NGV552388H, NGV530576H, NGV530703H, NGV550979H, NGV551027H, NGV550600H, NGV528135H, NGV529461H, NGV530397H, NGV551785H, NGV530314H, NGV530319H, NGV550950H, NGV550957H, NGV551323H, NGV530405H, NGV529076H, NGV529079H, NGV529909H, NGV529930H, NGV530004H, NGV530365H, NGV530783H, NGV551341H, NGV551638H, NGV552421H, NGV530531H, NGV550903H, NGV528465H, NGV528970H, NGV529262H, NGV529684H, NGV529688H, NGV529989H, NGV530318H, NGV530326H, NGV530491H, NGV550110H, NGV550524H, NGV551285H, NGV551750H, NGV552100H, NGV552140H, NGV552208H, NGV552392H, NGV552477H, NGV552329H, NGV551668H, NGV520856H, NGV530399H, NGV530940H, NGV552466H, NGV530428H, NGV530467H, NGV552171H, NGV516162H, NGV516221H, NGV517962H, NGV520767H, NGV530677H, NGV550548H, NGV552305H, NGV552394H, NGV553044H, NGV530271H, NGV530841H, NGV550152H, NGV552339H, NGV552364H, NGV552372H, NGV527102H, NGV530589H, NGV530915H, NGV550060H, NGV550744H, NGV551375H, NGV516158H, NGV529358H, NGV550310H, NGV550348H, NGV551102H, NGV551396H, NGV552231H, NGV552237H, NGV528173H, NGV552119H, NGV529973H, NGV550379H, NGV550509H, NGV530352H, NGV528163H, NGV529080H, NGV529315H, NGV529681H, NGV529905H, NGV530921H, NGV530935H, NGV552351H, NGV530426H, NGV550944H, NGV552284H, NGV530057H, NGV551768H, NGV553023H, NGV529955H, NGV529318H, NGV523768H, NGV530730H, NGV552330H, NGV552377H, NGV553030H, NGV530847H, NGV530925H, NGV552261H, NGV552442H, NGV528545H, NGV528383H, NGV529058H, NGV529063H, NGV529834H, NGV529975H, NGV530967H, NGV551790H, NGV551874H, NGV552098H, NGV552106H, NGV552112H, NGV552120H, NGV552130H, NGV528560H, NGV528813H, NGV552102H, NGV552288H, NGV528350H, NGV529174H, NGV529532H, NGV529865H, NGV530310H, NGV551066H, NGV552380H, NGV530546H, NGV552526H, NGV553037H, NGV529782H, NGV550400H, NGV523999H, NGV530342H, NGV530658H, NGV551126H, NGV551993H, NGV520853H, NGV528490H, NGV528504H, NGV528974H, NGV529218H, NGV530308H, NGV550204H, NGV550207H, NGV551516H, NGV552133H, NGV552137H, NGV552138H, NGV552139H, NGV552347H, NGV517965H, NGV529445H, NGV530490H, NGV550088H, NGV552267H, NGV552287H, NGV528443H, NGV529746H, NGV550376H, NGV552063H, NGV527187H, NGV529275H, NGV529376H, NGV530368H, NGV530568H, NGV529976H, NGV528627H, NGV529985H, NGV550586H, NGV551865H, NGV552338H, NGV552391H, NGV530366H, NGV522170H, NGV552215H, NGV529558H, NGV530675H, NGV550254H, NGV551511H, NGV552235H, NGV552350H, NGV552551H, NGV530573H, NGV530558H, NGV550347H, NGV552326H, NGV552343H, NGV529718H, NGV530769H, NGV552164H, NGV552172H, NGV520881H, NGV530377H, NGV552191H, NGV552316H, NGV552419H, NGV552480H, NGV530346H, NGV530832H, NGV528461H, NGV529754H, NGV530479H, NGV552458H, NGV529226H, NGV550018H, NGV550386H, NGV552099H, NGV552493H, NGV530485H, NGV550289H, NGV551033H, NGV552966H, NGV528485H, NGV551628H, NGV552174H, NGV528971H, NGV528999H, NGV552115H, NGV529062H, NGV529430H, NGV529923H, NGV551104H, NGV523694H, NGV529408H, NGV551070H, NGV529001H, NGV529004H, NGV529017H, NGV529020H, NGV529024H, NGV529042H, NGV529073H, NGV529084H, NGV529257H, NGV528387H, NGV551311H, NGV530745H, NGV516234H, NGV528466H, NGV528475H, NGV529032H, NGV529034H, NGV529240H, NGV529241H, NGV530036H, NGV551317H, NGV551621H, NGV552199H, NGV552529H, NGV551999H, NGV528114H, NGV528117H, NGV552545H, NGV516191H, NGV530937H, NGV551077H, NGV552129H, NGV552160H, NGV529165H, NGV552094H, NGV552097H, NGV528609H, NGV529060H, NGV551546H, NGV552209H, NGV552217H, NGV552469H, NGV530021H, NGV530755H, NGV551376H, NGV552214H, NGV552453H, NGV552533H, NGV553020H, NGV528435H, NGV528436H, NGV530016H, NGV530043H, NGV530273H, NGV530332H, NGV530334H, NGV530367H, NGV530422H, NGV530427H, NGV530429H, NGV530587H, NGV530613H, NGV530627H, NGV530795H, NGV530933H, NGV550142H, NGV550147H, NGV550187H, NGV550560H, NGV550562H, NGV552184H, NGV552187H, NGV527151H, NGV551917H, NGV550549H, NGV530002H, NGV550276H, NGV529695H, NGV550952H, NGV528291H, NGV529054H, NGV530018H, NGV530361H, NGV530469H, NGV552432H, NGV517419H, NGV552527H, NGV552992H, NGV528816H, NGV528984H, NGV550528H, NGV516232H, NGV529957H, NGV530372H, NGV552979H, NGV530054H, NGV552317H, NGV552536H, NGV552312H, NGV552206H, NGV516193H, NGV551179H, NGV552997H, NGV528529H, NGV528531H, NGV529507H, NGV529854H, NGV529859H, NGV530430H, NGV530431H, NGV530450H, NGV530451H, NGV550251H, NGV550257H, NGV530552H, NGV530569H, NGV530911H, NGV552156H, NGV552157H, NGV530905H, NGV529618H, NGV528522H, NGV551114H, NGV530739H, NGV551038H, NGV552991H, NGV530391H, NGV530820H, NGV527173H, NGV528565H, NGV529189H, NGV552454H, NGV529094H, NGV530508H, NGV551047H, NGV552472H, NGV553003H, NGV520884H, NGV529145H, NGV530846H, NGV550359H, NGV525622H, NGV528189H, NGV551580H, NGV550891H, NGV529738H, NGV529743H, NGV529849H, NGV552259H, NGV552414H, NGV552426H, NGV528382H, NGV530268H, NGV529694H, NGV530360H, NGV550570H, NGV552240H, NGV552464H, NGV552522H, NGV550989H, NGV551324H, NGV528451H, NGV529926H, NGV530313H, NGV552325H, NGV529787H, NGV520943H, NGV551737H, NGV529932H, NGV530390H, NGV530019H, NGV529742H, NGV529827H, NGV530948H, NGV550911H, NGV552521H, NGV528198H, NGV530274H, NGV552260H, NGV552982H, NGV552985H, NGV530343H, NGV530763H, NGV551470H, NGV552311H, NGV550590H, NGV550325H, NGV551339H, NGV529438H, NGV529636H, NGV530230H, NGV530287H, NGV530316H, NGV530317H, NGV530432H, NGV530433H, NGV530455H, NGV552397H, NGV552410H, NGV552417H, NGV551003H, NGV528455H, NGV552245H, NGV550532H, NGV551103H, NGV551177H, NGV527118H, NGV526744H, NGV530283H, NGV529361H, NGV551336H, NGV552969H, NGV526614H, NGV528312H, NGV528788H, NGV550959H, NGV552230H, NGV528167H, NGV529889H, NGV529433H, NGV530934H, NGV551111H, NGV552294H, NGV526666H, NGV526711H, NGV530331H, NGV530654H, NGV551034H, NGV520791H, NGV529983H, NGV530932H, NGV551149H, NGV551347H, NGV552512H, NGV530897H, NGV550065H, NGV552358H, NGV552272H, NGV550097H, NGV552001H, NGV552296H, NGV530406H, NGV553005H, NGV523917H, NGV527128H, NGV553017H, NGV551475H, NGV526642H, NGV526748H, NGV552382H, NGV528028H, NGV529680H, NGV527136H, NGV528229H, NGV528474H, NGV529190H, NGV529247H, NGV529265H, NGV529711H, NGV529717H, NGV529778H, NGV529897H, NGV529931H, NGV530305H, NGV530577H, NGV530949H, NGV550340H, NGV550987H, NGV551608H, NGV551609H, NGV552179H, NGV552352H, NGV552353H, NGV552530H, NGV553007H, NGV530771H, NGV552984H, NGV529762H, NGV550755H, NGV552262H, NGV530339H, NGV530340H, NGV529764H, NGV551171H, NGV551316H, NGV530329H, NGV530335H, NGV529904H, NGV550078H, NGV552225H, NGV552319H, NGV552474H, NGV527124H, NGV530392H; GTIN 00643169732353, Serial Numbers: NGV529000H, NGV529025H, NGV530782H, NGV530918H, NGV550242H, NGV550636H, NGV550761H, NGV551095H, NGV551333H, NGV528149H, NGV529088H, NGV530408H, NGV552186H, NGV552192H, NGV552430H, NGV529180H, NGV526738H, NGV528429H, NGV529156H, NGV529992H, NGV529995H, NGV550214H, NGV551856H, NGV552091H, NGV530421H, NGV530656H, NGV551866H, NGV552492H, NGV529029H, NGV529861H, NGV529949H, NGV530482H, NGV550240H, NGV551183H, NGV528790H, NGV528486H, NGV530354H, NGV551335H, NGV529002H, NGV552131H, NGV552369H, NGV526755H, NGV530276H, NGV530899H, NGV550222H, NGV550579H, NGV550589H, NGV552134H, NGV529194H, NGV529204H, NGV529997H, NGV530037H, NGV530299H, NGV530437H, NGV530909H, NGV530960H, NGV551642H, NGV551692H, NGV551982H, NGV552158H, NGV552255H, NGV552291H, NGV530891H, NGV529882H, NGV529977H, NGV530412H, NGV551259H, NGV528847H, NGV529116H, NGV552069H, NGV552271H, NGV550505H, NGV551153H, NGV529792H, NGV530000H, NGV530764H, NGV530893H, NGV529027H, NGV529386H, NGV529928H, NGV530355H, NGV530386H, NGV530452H, NGV530549H, NGV530638H, NGV530696H, NGV530699H, NGV530721H, NGV530821H, NGV530922H, NGV530969H, NGV550260H, NGV550290H, NGV550569H, NGV551867H, NGV552146H, NGV552162H, NGV552204H, NGV552412H, NGV552424H, NGV552451H, NGV528480H, NGV550657H, NGV551147H, NGV529010H, NGV529895H, NGV552327H, NGV552448H, NGV530195H, NGV530315H, NGV530336H, NGV530772H, NGV530898H, NGV551167H, NGV530964H, NGV552457H, NGV552476H, NGV550296H, NGV550247H, NGV529369H, NGV528468H, NGV528986H, NGV529442H, NGV529847H, NGV551105H, NGV551504H, NGV552143H, NGV529982H, NGV530051H, NGV530415H, NGV550602H, NGV550988H, NGV551899H, NGV552249H, NGV525983H, NGV526608H, NGV528981H, NGV530069H, NGV530298H, NGV530438H, NGV552495H, NGV552501H, NGV529037H, NGV550587H, NGV528459H, NGV528994H, NGV528998H, NGV530645H, NGV530712H, NGV530735H, NGV530736H, NGV530759H, NGV550603H, NGV551542H, NGV551616H, NGV552513H, NGV552515H, NGV551852H, NGV529978H, NGV530062H, NGV530913H, NGV551169H, NGV529853H, NGV551556H, NGV551596H, NGV552121H, NGV530278H, NGV529922H, NGV530621H, NGV529450H, NGV530593H, NGV551987H, NGV529059H, NGV551338H, NGV552354H, NGV528508H, NGV528612H, NGV529026H, NGV530136H, NGV530512H, NGV530734H, NGV530760H, NGV550545H, NGV526553H, NGV530301H; GTIN 00643169732360, Serial Numbers: NGV552481H, NGV530661H, NGV530741H, NGV552173H, NGV529019H, NGV529683H, NGV529781H, NGV530285H, NGV530929H, NGV550932H, NGV551109H, NGV530607H, NGV516213H, NGV527159H, NGV529673H, NGV530325H, NGV530536H, NGV550167H, NGV550174H, NGV550200H, NGV552229H, NGV550165H, NGV550894H, NGV552182H, NGV525985H, NGV528370H, NGV529850H, NGV550150H, NGV526437H, NGV528018H, NGV528084H, NGV528388H, NGV528414H, NGV528421H, NGV529023H, NGV530683H, NGV530702H, NGV530956H, NGV550887H, NGV550893H, NGV550956H, NGV550965H, NGV528338H, NGV528339H, NGV529844H, NGV550010H, NGV526436H, NGV528102H, NGV528365H, NGV528625H, NGV529482H, NGV529818H, NGV530400H, NGV550397H, NGV550403H, NGV526369H, NGV527622H, NGV527701H, NGV550937H, NGV552116H, NGV552150H, NGV529891H, NGV550384H, NGV529772H, NGV529678H, NGV530666H, NGV528356H, NGV529884H, NGV552103H, NGV527976H, NGV528404H, NGV550108H, NGV526732H, NGV527916H, NGV530954H, NGV550387H, NGV552159H, NGV552161H, NGV526602H, NGV528378H, NGV528492H, NGV529099H, NGV529363H, NGV529855H, NGV529991H, NGV550244H, NGV528360H, NGV528988H, NGV529022H, NGV529086H, NGV529759H, NGV529777H, NGV530330H, NGV530357H, NGV530370H, NGV530445H, NGV530653H, NGV550268H, NGV550292H, NGV550303H, NGV550919H, NGV551841H, NGV552167H, NGV552178H, NGV527835H, NGV528357H, NGV528476H, NGV529028H, NGV529584H, NGV529704H, NGV530896H, NGV550195H, NGV550395H, NGV552257H, NGV516210H, NGV526623H, NGV528400H, NGV529797H, NGV528337H, NGV528343H, NGV529015H, NGV529709H, NGV550226H, NGV528391H, NGV528423H, NGV528431H, NGV529732H, NGV528469H, NGV528473H, NGV551819H; GTIN 00643169732377, Serial Numbers NGV530867H, NGV552108H, NGV530697H, NGV530845H, NGV530858H, NGV552109H, NGV552111H, NGV530747H, NGV530859H, NGV530862H, NGV530860H, NGV530719H, NGV550332H, NGV550267H, NGV530448H, NGV552504H, NGV530746H, NGV526692H, NGV526772H, NGV528180H, NGV552323H, NGV552532H, NGV528220H, NGV530453H, NGV530456H, NGV550188H, NGV550149H, NGV550220H, NGV550233H, NGV528420H, NGV528335H, NGV526017H, NGV528386H, NGV550153H, NGV530363H, NGV530705H, NGV552336H, NGV530827H, NGV550138H, NGV550089H, NGV550969H, NGV528138H, NGV528197H, NGV528209H, NGV528425H, NGV528607H, NGV530347H, NGV530381H, NGV530382H, NGV551765H, NGV528204H, NGV530776H, NGV528297H, NGV530822H, NGV550304H, NGV551771H, NGV530457H, NGV550958H, NGV528208H, NGV528276H, NGV528614H, NGV550086H, NGV551005H, NGV551682H, NGV530836H, NGV550265H, NGV528373H, NGV530770H, NGV530375H, NGV530738H, NGV550227H, NGV551735H, NGV526720H, NGV526703H, NGV528367H, NGV528372H, NGV528377H, NGV528613H, NGV530344H, NGV530449H, NGV530798H, NGV550170H, NGV550322H, NGV550365H, NGV552324H, NGV552361H, NGV550318H, NGV552321H, NGV526033H, NGV551000H, NGV528190H, NGV528193H, NGV526767H, NGV552054H, NGV552124H, NGV552118H, NGV526756H, NGV527210H, NGV527313H, NGV550248H, NGV550266H, NGV528185H, NGV552341H, NGV530694H, NGV550201H, NGV551730H, NGV530789H, NGV550191H, NGV528123H, NGV528200H, NGV528313H, NGV551786H, NGV552328H, NGV550126H, NGV551686H, NGV530385H, NGV550243H, NGV552113H, NGV551820H, NGV551821H, NGV551848H, NGV552177H, NGV552180H, NGV552193H, NGV530604H, NGV530608H, NGV551868H, NGV551798H, NGV530647H, NGV552144H, NGV552128H, NGV530461H, NGV530462H, NGV530463H, NGV551797H, NGV551881H, NGV552135H, NGV552136H, NGV552151H, NGV552163H, NGV552176H, NGV552132H, NGV530633H, NGV529600H, NGV530507H, NGV552226H, NGV530362H, NGV528536H, NGV528541H, NGV530356H, NGV550380H, NGV551181H, NGV551182H, NGV528523H, NGV550392H, NGV550393H, NGV550399H, NGV530359H, NGV528519H, NGV528521H, NGV528588H, NGV552170H, NGV552175H, NGV530353H, NGV530389H, NGV530410H, NGV528586H, NGV550385H, NGV550372H, NGV550388H, NGV528524H, NGV530296H, NGV530297H, NGV530300H, NGV530302H, NGV530306H, NGV550722H, NGV550739H, NGV550742H, NGV551377H, NGV552404H, NGV552420H, NGV552428H, NGV552440H, NGV552441H, NGV552456H, NGV530498H, NGV528430H, NGV528433H, NGV528437H, NGV528440H, NGV528444H, NGV530311H, NGV529312H, NGV529326H, NGV530333H, NGV530411H, NGV552367H, NGV552378H, NGV528511H, NGV529353H, NGV529366H, NGV530320H, NGV530439H, NGV530446H, NGV529351H, NGV529324H, NGV529367H, NGV530420H, NGV529277H, NGV529279H, NGV529161H, NGV530280H, NGV529123H, NGV529153H, NGV529193H, NGV550536H, NGV529114H, NGV550563H, NGV525862H, NGV529248H, NGV529250H, NGV529254H, NGV529880H, NGV529953H, NGV529967H, NGV529981H, NGV530042H, NGV530376H, NGV550028H, NGV550030H, NGV530402H, NGV528828H, NGV530669H, NGV529269H, NGV530651H, NGV530887H, NGV530889H, NGV528498H, NGV528756H, NGV528472H, NGV530866H, NGV550206H, NGV552362H, NGV552366H, NGV529446H, NGV529459H, NGV529791H, NGV529793H, NGV529798H, NGV529799H, NGV529808H, NGV529810H, NGV529814H, NGV529816H, NGV529830H, NGV529661H, NGV552165H, NGV528736H, NGV552286H, NGV552166H, NGV551602H, NGV552304H, NGV529767H, NGV529730H, NGV551606H, NGV551617H, NGV552258H, NGV552264H, NGV551618H, NGV552275H, NGV552285H, NGV551567H, NGV552306H, NGV528593H, NGV529677H, NGV552168H, NGV552298H, NGV529710H, NGV551632H, NGV551601H, NGV552301H, NGV528722H, NGV552289H, NGV529727H, NGV552318H, NGV551640H, NGV552490H, NGV528233H, NGV530493H, NGV550019H, NGV551166H, NGV528131H, NGV530281H, NGV530541H, NGV552478H, NGV552541H, NGV551956H, NGV528160H, NGV528462H, NGV530876H, NGV551901H, NGV527816H, NGV528142H, NGV528145H, NGV528148H, NGV528151H, NGV528175H, NGV529078H, NGV529133H, NGV529144H, NGV529148H, NGV529150H, NGV529485H, NGV529553H, NGV529900H, NGV529903H, NGV529927H, NGV530879H, NGV530890H, NGV530901H, NGV530902H, NGV530907H, NGV550072H, NGV551130H, NGV551290H, NGV551301H, NGV551314H, NGV551322H, NGV551342H, NGV551482H, NGV551490H, NGV551517H, NGV552503H, NGV552523H, NGV552525H, NGV552531H, NGV528110H, NGV550945H, NGV516198H, NGV526776H, NGV530545H, NGV550892H, NGV530572H, NGV527140H, NGV550571H, NGV526778H, NGV550529H, NGV550920H, NGV550939H, NGV528133H, NGV530380H, NGV550574H, NGV552499H, NGV550553H, NGV550743H, NGV551029H, NGV552552H, NGV552987H, NGV529245H, NGV552546H, NGV529107H, NGV529117H, NGV530910H, NGV551472H, NGV551644H, NGV550926H, NGV550933H, NGV552395H, NGV550566H, NGV552427H, NGV528118H, NGV550948H, NGV551082H, NGV529031H, NGV529038H, NGV529934H, NGV530006H, NGV530017H, NGV550596H, NGV552963H, NGV527539H, NGV550032H, NGV527814H, NGV552974H, NGV530916H, NGV528307H, NGV529102H, NGV529106H, NGV529629H, NGV530962H, NGV551986H, NGV530583H, NGV528152H, NGV550746H, NGV551162H, NGV551416H, NGV530394H, NGV550061H, NGV528449H, NGV530277H, NGV530289H, NGV530513H, NGV552993H, NGV529051H, NGV529946H, NGV530489H, NGV530873H, NGV551449H, NGV528113H, NGV552959H, NGV530132H, NGV552197H, NGV530525H, NGV530543H, NGV530547H, NGV530548H, NGV530923H, NGV530928H, NGV551086H, NGV551090H, NGV551118H, NGV551123H, NGV551136H, NGV551165H, NGV529200H, NGV529914H, NGV529915H, NGV550896H, NGV528278H, NGV529291H, NGV529375H, NGV529409H, NGV529427H, NGV530425H, NGV530447H, NGV530466H, NGV550978H, NGV550993H, NGV551085H, NGV551129H, NGV551150H, NGV551152H, NGV551156H, NGV551351H, NGV552213H, NGV552216H, NGV552218H, NGV552221H, NGV529432H, NGV550533H, NGV550518H, NGV552236H, NGV552239H, NGV529152H, NGV529789H, NGV529790H, NGV550001H, NGV550024H, NGV550025H, NGV550546H, NGV550034H, NGV528153H, NGV528447H, NGV528176H, NGV530553H, NGV530591H, NGV550759H, NGV552234H, NGV529268H, NGV550038H, NGV550058H, NGV550751H, NGV550910H, NGV550963H, NGV551663H, NGV529956H, NGV530046H, NGV530068H, NGV530958H, NGV530961H, NGV550064H, NGV550067H, NGV550559H, NGV550970H, NGV550973H, NGV551385H, NGV552433H, NGV552348H; GTIN 00643169732384, Serial Numbers: NGV528527H, NGV552210H, NGV530035H, NGV529656H, NGV529721H, NGV550055H, NGV550091H, NGV528439H, NGV529181H, NGV550063H, NGV551633H, NGV551660H, NGV551782H, NGV552024H, NGV552547H, NGV550238H, NGV528591H, NGV528726H, NGV529018H, NGV550208H, NGV550209H, NGV550212H, NGV552077H, NGV552408H, NGV552446H, NGV552449H, NGV550270H, NGV550280H, NGV552452H, NGV552473H, NGV526374H, NGV529626H, NGV550295H, NGV527247H, NGV528441H, NGV528446H, NGV530871H, NGV550027H, NGV550051H, NGV550281H, NGV552207H, NGV552494H, NGV552958H, NGV529737H, NGV529758H, NGV550020H, NGV550057H, NGV550213H, NGV552251H, NGV552279H, NGV552282H, NGV552309H, NGV552310H, NGV552370H, NGV552384H, NGV552401H, NGV552429H, NGV552463H, NGV530384H, NGV550021H, NGV550291H, NGV551671H, NGV528159H, NGV528352H, NGV528416H, NGV528532H, NGV529192H, NGV529195H, NGV529253H, NGV529575H, NGV529596H, NGV529614H, NGV530050H, NGV551296H, NGV552093H, NGV552127H, NGV552219H, NGV552425H, NGV552548H, NGV552227H, NGV552514H; GTIN 00643169732391, Serial Numbers: NGV530059H, NGV530270H, NGV550026H, NGV550082H, NGV550250H, NGV551067H, NGV551117H, NGV551122H, NGV551593H, NGV551605H, NGV551654H, NGV551732H, NGV516222H, NGV526366H, NGV526771H, NGV527171H, NGV528442H, NGV529561H, NGV529713H, NGV530282H, NGV530292H, NGV530304H, NGV550367H, NGV550382H, NGV529554H, NGV529647H, NGV529649H, NGV529716H, NGV529765H, NGV529768H, NGV529771H, NGV529794H, NGV529811H, NGV529817H, NGV530275H, NGV530349H, NGV550277H, NGV550521H, NGV550673H, NGV551461H, NGV552483H, NGV552497H, NGV552349H, NGV552376H, NGV552461H, NGV530959H, NGV552222H, NGV516235H, NGV529555H, NGV530341H, NGV530369H, NGV530952H, NGV550043H, NGV550331H, NGV550361H, NGV550525H, NGV550542H, NGV550976H, NGV551021H, NGV552505H, NGV530358H, NGV551050H, NGV551145H, NGV530309H, NGV551049H; GTIN 00643169732407, Serial Numbers: NGV526378H, NGV526605H, NGV526785H, NGV527137H, NGV528514H, NGV529272H, NGV529299H, NGV529425H, NGV529440H, NGV529464H, NGV529676H, NGV529723H, NGV529775H, NGV529776H, NGV529779H, NGV529780H, NGV529783H, NGV529785H, NGV529805H, NGV529828H, NGV529866H, NGV550016H, NGV550017H, NGV550023H, NGV550033H, NGV550039H, NGV550041H, NGV550054H, NGV550073H, NGV550094H, NGV550114H, NGV550169H, NGV550223H, NGV550550H, NGV550564H, NGV550578H, NGV551583H, NGV516238H, NGV550341H, NGV550368H, NGV550371H, NGV550373H, NGV550383H, NGV550889H, NGV551030H, NGV516208H, NGV516216H; GTIN 00643169732414, Serial Numbers: NGV550324H, NGV550514H, NGV552406H, NGV552445H, NGV552486H, NGV528526H, NGV529296H, NGV530729H, NGV550117H, NGV551076H, NGV551178H, NGV552360H, NGV530698H, NGV529065H, NGV551871H, NGV551607H, NGV551634H, NGV553015H, NGV550531H, NGV552398H, NGV526394H, NGV553024H, NGV520402H, NGV552487H, NGV552519H, NGV552516H; GTIN 00643169732421, Serial Numbers: NGV551020H, NGV516237H; GTIN 00643169732438, Serial Numbers: NGV530254H, NGV529217H, NGV530551H, NGV552402H, NGV530441H, NGV530416H, NGV530417H, NGV529899H, NGV529918H, NGV551041H, NGV551174H, NGV551565H, NGV551569H, NGV529402H, NGV550119H, NGV530272H, NGV530691H, NGV552396H, NGV530030H, NGV552332H, NGV552153H, NGV552283H, NGV552337H, NGV530888H, NGV530414H, NGV530324H, NGV552110H, NGV530327H, NGV530013H, NGV552101H, NGV530682H, NGV552198H, NGV552232H, NGV530801H, NGV552390H, NGV529344H, NGV530350H, NGV550573H, NGV529468H, NGV530881H, NGV529030H, NGV530813H, NGV530914H, NGV530387H, NGV530955H, NGV550556H, NGV530707H, NGV530718H, NGV530686H, NGV530927H, NGV552357H, NGV528997H, NGV528966H, NGV530649H; GTIN 00643169984226, Serial Numbers: NGV613584H, NGV613330H, NGV613490H, NGV613524H, NGV613499H, NGV613476H, NGV613582H, NGV613628H, NGV613614H, NGV613394H, NGV613395H, NGV613494H, NGV613527H, NGV613651H, NGV613319H, NGV613320H, NGV613321H, NGV613322H, NGV613561H, NGV613506H, NGV613236H, NGV613237H, NGV613244H, NGV613495H, NGV612882H, NGV613827H, NGV613828H, NGV613436H, NGV613312H, NGV613583H, NGV613633H, NGV612546H, NGV612884H, NGV613371H, NGV613466H, NGV613470H, NGV610082H, NGV613501H, NGV613860H, NGV613613H, NGV613369H, NGV613533H, NGV613543H, NGV613571H, NGV613432H, NGV612914H, NGV613654H, NGV613407H, NGV613493H, NGV613483H, NGV613484H, NGV613405H, NGV613838H, NGV613451H, NGV613612H, NGV613775H, NGV612458H, NGV612536H, NGV613292H, NGV613542H, NGV613768H, NGV613195H, NGV613465H, NGV613565H, NGV613579H, NGV613809H, NGV613787H, NGV613481H, NGV613316H, NGV613801H, NGV613794H, NGV613863H, NGV613625H, NGV613293H, NGV613298H, NGV613325H, NGV613326H, NGV613498H, NGV613442H, NGV613641H, NGV613279H, NGV613356H, NGV613850H, NGV613540H, NGV613491H, NGV610057H, NGV610064H, NGV613352H, NGV613386H, NGV613766H, NGV613812H, NGV613440H, NGV613586H, NGV611923H, NGV613179H, NGV613403H, NGV613799H, NGV613538H, NGV613858H, NGV613275H, NGV613859H, NGV613450H, NGV613513H, NGV613636H, NGV613808H, NGV613469H, NGV613694H, NGV613700H, NGV614116H, NGV613286H, NGV613327H, NGV613396H, NGV613400H, NGV613464H, NGV613770H, NGV613825H, NGV613249H, NGV613660H, NGV613662H, NGV613515H, NGV613449H, NGV613597H, NGV612880H, NGV613384H, NGV613385H, NGV613798H, NGV613604H, NGV613076H, NGV613492H, NGV613562H, NGV613359H, NGV613381H, NGV613578H, NGV612442H, NGV613864H, NGV613587H, NGV613502H, NGV613536H, NGV613537H, NGV613556H, NGV613573H, NGV613596H, NGV613346H, NGV613832H, NGV613835H, NGV613848H, NGV613590H, NGV613445H, NGV613358H, NGV613576H, NGV613862H, NGV613357H, NGV613560H, NGV613441H, NGV613514H, NGV613516H, NGV613539H, NGV613643H, NGV613648H, NGV613553H, NGV613856H, NGV613259H, NGV613263H, NGV613774H, NGV613854H, NGV613792H, NGV613426H, NGV613276H, NGV613278H, NGV613776H, NGV613778H, NGV613779H, NGV613505H, NGV613324H, NGV613295H, NGV613624H, NGV613670H, NGV613668H, NGV613472H, NGV613783H, NGV612739H, NGV612755H, NGV612757H, NGV612770H, NGV612803H, NGV613214H, NGV613419H, NGV613520H, NGV612533H, NGV612541H, NGV613837H, NGV613839H, NGV613435H, NGV613557H, NGV613558H, NGV613608H, NGV613844H, NGV613216H, NGV613217H, NGV613220H, NGV613223H, NGV613224H, NGV613226H, NGV613228H, NGV613229H, NGV613233H, NGV613995H, NGV613598H, NGV613215H, NGV610118H, NGV610137H, NGV610147H, NGV612585H, NGV613508H, NGV613588H, NGV613845H, NGV613568H, NGV613471H, NGV613343H, NGV613370H, NGV613430H, NGV613350H, NGV613373H, NGV613623H, NGV613444H, NGV613831H, NGV613602H, NGV613478H, NGV613379H, NGV613599H, NGV613781H, NGV613577H, NGV613496H, NGV613437H, NGV612714H, NGV613551H, NGV613434H, NGV613342H, NGV613619H, NGV613601H, NGV613840H, NGV613656H, NGV613281H, NGV613531H, NGV613842H, NGV613378H, NGV613594H, NGV613824H, NGV613285H, NGV613301H, NGV613416H, NGV613417H, NGV613365H, NGV613335H, NGV610074H, NGV613629H, NGV613663H, NGV613659H, NGV613607H, NGV613317H, NGV613313H, NGV612665H, NGV613252H, NGV613296H, NGV610144H, NGV613667H, NGV613302H, NGV613807H, NGV613603H, NGV613646H, NGV613649H, NGV613673H, NGV613674H, NGV613681H, NGV613682H, NGV613690H, NGV613691H, NGV613701H, NGV613707H, NGV613709H, NGV613723H, NGV613732H, NGV613741H, NGV613744H, NGV613747H, NGV613760H, NGV613761H, NGV613867H, NGV613868H, NGV613872H, NGV613873H, NGV613894H, NGV613896H, NGV613897H, NGV613909H, NGV613911H, NGV613912H, NGV613913H, NGV613920H, NGV613924H, NGV613925H, NGV613927H, NGV613928H, NGV613932H, NGV613935H, NGV613937H, NGV613940H, NGV613943H, NGV613953H, NGV613954H, NGV613957H, NGV613964H, NGV613979H, NGV613987H, NGV613990H, NGV614001H, NGV614006H, NGV614010H, NGV614012H, NGV614019H, NGV614020H, NGV614026H, NGV614032H, NGV614035H, NGV614055H, NGV614061H, NGV614063H, NGV614065H, NGV614071H, NGV614079H, NGV614085H, NGV614087H, NGV614090H, NGV614108H, NGV614141H, NGV613644H, NGV613482H, NGV613615H, NGV613479H, NGV613791H, NGV613519H, NGV613853H, NGV613348H, NGV613541H, NGV613525H, NGV613554H, NGV610068H, NGV613802H, NGV613447H, NGV612788H, NGV613606H, NGV613795H, NGV613414H, NGV613830H, NGV613497H, NGV613833H, NGV613780H, NGV613782H, NGV613487H, NGV613328H, NGV613504H, NGV613310H, NGV613366H, NGV613550H, NGV613605H, NGV613626H, NGV613509H, NGV613338H, NGV613951H, NGV614089H, NGV613570H, NGV613851H, NGV613788H, NGV613475H, NGV613836H, NGV613785H, NGV613640H, NGV612503H, NGV613422H, NGV612718H, NGV613273H, NGV613247H, NGV613821H, NGV613460H, NGV613463H, NGV613815H, NGV613816H, NGV613818H, NGV613820H, NGV613592H, NGV613574H, NGV613323H, NGV613529H, NGV613411H, NGV613232H, NGV613485H, NGV613287H, NGV613591H, NGV613593H, NGV613855H, NGV613622H, NGV613639H, NGV613391H, NGV613767H, NGV613610H, NGV613427H, NGV613486H, NGV613367H, NGV612728H, NGV613311H, NGV613595H, NGV613517H, NGV613364H, NGV613544H, NGV613383H, NGV614038H, NGV614044H, NGV613620H, NGV613810H, NGV613784H, NGV613507H, NGV613803H, NGV612783H, NGV613221H, NGV613988H, NGV613989H, NGV613611H, NGV613834H, NGV613817H; GTIN 00643169999930, Serial Numbers: NGV601913H, NGV603962H, NGV605053H, NGV607267H, NGV607300H, NGV607332H, NGV611165H, NGV611169H, NGV601726H, NGV605820H, NGV607663H, NGV610479H, NGV610919H, NGV611257H, NGV611691H, NGV611986H, NGV612114H, NGV603453H, NGV603458H, NGV603956H, NGV606865H, NGV608347H, NGV611410H, NGV612736H, NGV612756H, NGV604289H, NGV604285H, NGV607610H, NGV607770H, NGV610781H, NGV612721H, NGV612749H, NGV603972H, NGV604284H, NGV605601H, NGV606187H, NGV608496H, NGV610847H, NGV611264H, NGV611399H, NGV607748H, NGV608879H, NGV609201H, NGV609802H, NGV612746H, NGV612810H, NGV601558H, NGV606195H, NGV606806H, NGV607091H, NGV610047H, NGV611188H, NGV611328H, NGV611650H, NGV612278H, NGV613241H, NGV603427H, NGV605597H, NGV612763H, NGV601571H, NGV601574H, NGV601579H, NGV604538H, NGV606058H, NGV606289H, NGV608192H, NGV608240H, NGV608513H, NGV609020H, NGV609025H, NGV611106H, NGV611627H, NGV612006H, NGV612269H, NGV613023H, NGV607509H, NGV604326H, NGV601980H, NGV603354H, NGV603813H, NGV608218H, NGV608935H, NGV609171H, NGV610355H, NGV612000H, NGV612725H, NGV604358H, NGV609198H, NGV610645H, NGV611214H, NGV611647H, NGV601753H, NGV603132H, NGV604325H, NGV604328H, NGV607440H, NGV607795H, NGV608190H, NGV608247H, NGV608915H, NGV609190H, NGV610844H, NGV611510H, NGV611636H, NGV603119H, NGV605406H, NGV607549H, NGV608525H, NGV609738H, NGV610046H, NGV610358H, NGV611005H, NGV611131H, NGV611597H, NGV611914H, NGV605052H, NGV603368H, NGV606920H, NGV613171H, NGV603366H, NGV603960H, NGV605150H, NGV608262H, NGV608484H, NGV610884H, NGV610898H, NGV601619H, NGV601973H, NGV605572H, NGV608217H, NGV608508H, NGV611116H, NGV611199H, NGV605189H, NGV611067H, NGV613545H, NGV604272H, NGV605713H, NGV606061H, NGV608386H, NGV609223H, NGV609814H, NGV609936H, NGV611638H, NGV611651H, NGV605772H, NGV606945H, NGV609217H, NGV601144H, NGV601576H, NGV603495H, NGV603498H, NGV606196H, NGV607486H, NGV607710H, NGV608205H, NGV608220H, NGV611141H, NGV611622H, NGV613167H, NGV613180H, NGV613189H, NGV613274H, NGV601192H, NGV606582H, NGV609183H, NGV610373H, NGV611155H, NGV600897H, NGV601182H, NGV601196H, NGV601198H, NGV601687H, NGV601716H, NGV602602H, NGV602951H, NGV603153H, NGV603160H, NGV603380H, NGV603463H, NGV603464H, NGV603662H, NGV603713H, NGV603744H, NGV603852H, NGV603882H, NGV604268H, NGV605063H, NGV605064H, NGV605238H, NGV605281H, NGV605488H, NGV605505H, NGV605740H, NGV605785H, NGV605814H, NGV606140H, NGV606611H, NGV606775H, NGV606835H, NGV606867H, NGV606956H, NGV607620H, NGV607622H, NGV607737H, NGV608260H, NGV608396H, NGV608544H, NGV609039H, NGV609642H, NGV609889H, NGV609945H, NGV610867H, NGV610896H, NGV611003H, NGV611036H, NGV611111H, NGV611230H, NGV611241H, NGV611246H, NGV611268H, NGV611405H, NGV611678H, NGV611687H, NGV611830H, NGV611994H, NGV612273H, NGV612389H, NGV612733H, NGV613289H, NGV613665H, NGV613671H, NGV601247H, NGV601662H, NGV601708H, NGV602020H, NGV603069H, NGV603863H, NGV603980H, NGV605070H, NGV606252H, NGV606309H, NGV607760H, NGV609528H, NGV613055H, NGV613341H, NGV607781H, NGV601167H, NGV601338H, NGV603068H, NGV605886H, NGV606153H, NGV607601H, NGV607605H, NGV608196H, NGV608490H, NGV608494H, NGV610268H, NGV610502H, NGV611563H, NGV611593H, NGV602037H, NGV603413H, NGV603417H, NGV603881H, NGV604280H, NGV605122H, NGV605124H, NGV612413H, NGV603199H, NGV603892H, NGV605089H, NGV611006H, NGV603515H, NGV610387H, NGV610822H, NGV611177H, NGV606298H, NGV606244H, NGV613181H, NGV613182H, NGV612712H, NGV603356H, NGV603500H, NGV603661H, NGV605722H, NGV608498H, NGV609169H, NGV610852H, NGV610916H, NGV611634H, NGV605569H, NGV603223H, NGV603855H, NGV603857H, NGV605576H, NGV606559H, NGV609204H, NGV609566H, NGV610787H, NGV611333H, NGV611589H, NGV607357H, NGV601200H, NGV602995H, NGV603137H, NGV603357H, NGV604291H, NGV604317H, NGV604322H, NGV605266H, NGV605566H, NGV605832H, NGV606023H, NGV606705H, NGV606785H, NGV607751H, NGV607815H, NGV607825H, NGV608171H, NGV608253H, NGV608388H, NGV609984H, NGV610404H, NGV611159H, NGV611245H, NGV611936H, NGV611981H, NGV612771H, NGV613149H, NGV605274H, NGV607637H, NGV611660H, NGV611978H, NGV603552H, NGV603555H, NGV603848H, NGV606026H, NGV607346H, NGV607570H, NGV610629H, NGV603372H, NGV603950H, NGV607598H, NGV611024H, NGV612773H, NGV601591H, NGV601612H, NGV601640H, NGV601964H, NGV602055H, NGV603396H, NGV603541H, NGV603835H, NGV605088H, NGV605440H, NGV605464H, NGV606693H, NGV606907H, NGV607281H, NGV607585H, NGV608197H, NGV608202H, NGV609147H, NGV610043H, NGV610435H, NGV610466H, NGV611595H, NGV611693H, NGV611893H, NGV601721H, NGV601994H, NGV606206H, NGV607489H, NGV607649H, NGV613638H, NGV601086H, NGV603147H, NGV607328H, NGV611670H, NGV605811H, NGV601713H, NGV601738H, NGV601877H, NGV601988H, NGV602757H, NGV603329H, NGV603532H, NGV604761H, NGV605607H, NGV606866H, NGV606882H, NGV606893H, NGV607756H, NGV609234H, NGV611655H, NGV613087H, NGV604311H, NGV606796H, NGV609154H, NGV611197H, NGV613143H, NGV601624H, NGV602932H, NGV607838H, NGV610905H, NGV601075H, NGV601984H, NGV603408H, NGV603888H, NGV605097H, NGV608252H, NGV610478H, NGV613157H, NGV604287H, NGV607511H, NGV607656H, NGV610040H, NGV608941H, NGV609219H, NGV610810H, NGV605463H, NGV602533H, NGV608175H, NGV611487H, NGV611983H, NGV612324H, NGV612363H, NGV603347H, NGV603352H, NGV603851H, NGV604017H, NGV604286H, NGV605564H, NGV605853H, NGV606189H, NGV606220H, NGV606824H, NGV607356H, NGV607667H, NGV609637H, NGV609808H, NGV610858H, NGV610932H, NGV611140H, NGV611488H, NGV612807H, NGV612722H, NGV607312H, NGV609160H, NGV609700H, NGV610799H, NGV611120H, NGV601193H, NGV601695H, NGV605092H, NGV605487H, NGV607723H, NGV605223H, NGV612375H, NGV601201H, NGV603359H, NGV609841H, NGV601166H, NGV608266H, NGV609056H, NGV609177H, NGV611056H, NGV601203H, NGV603792H, NGV606198H, NGV607623H, NGV609108H, NGV605455H, NGV607625H, NGV609164H, NGV606682H, NGV611050H, NGV612186H, NGV607624H, NGV603202H, NGV611154H, NGV605418H, NGV611110H, NGV606754H, NGV607725H, NGV608553H, NGV601172H, NGV601202H, NGV601237H, NGV601690H, NGV603513H, NGV603604H, NGV603607H, NGV607324H, NGV607711H, NGV607727H, NGV611143H, NGV613042H, NGV607285H, NGV603067H, NGV605219H, NGV605225H, NGV605230H, NGV606035H, NGV607845H, NGV605543H, NGV601082H, NGV604329H, NGV605059H, NGV610378H, NGV601580H, NGV611070H, NGV612867H, NGV603022H, NGV607768H, NGV609176H, NGV609187H, NGV609188H, NGV609189H, NGV609197H, NGV611999H, NGV605884H, NGV601665H, NGV605528H, NGV606899H, NGV611928H, NGV601218H, NGV605307H, NGV607361H, NGV611962H, NGV612181H, NGV605094H, NGV606759H, NGV608263H, NGV608613H, NGV612317H, NGV611297H, NGV605304H, NGV605308H, NGV612427H, NGV604363H, NGV606055H, NGV611565H, NGV608242H, NGV609235H, NGV603481H, NGV607673H, NGV607681H, NGV601238H, NGV605142H, NGV601709H, NGV604405H, NGV606805H, NGV606919H, NGV609167H, NGV611497H, NGV613303H, NGV606566H, NGV611122H, NGV603430H, NGV605166H, NGV606030H, NGV610570H, NGV611162H, NGV611538H, NGV611948H, NGV612311H, NGV612795H, NGV601180H, NGV606044H, NGV611243H, NGV613616H, NGV605575H, NGV606822H, NGV606846H, NGV610482H, NGV610494H, NGV607688H, NGV608236H, NGV608408H, NGV608542H, NGV610829H, NGV610833H, NGV612190H, NGV611919H, NGV602038H, NGV609065H, NGV609641H, NGV603977H, NGV604768H, NGV609806H, NGV609168H, NGV605245H, NGV607567H, NGV609145H, NGV609769H, NGV611034H, NGV611954H, NGV601170H, NGV603209H, NGV603215H, NGV605404H, NGV607266H, NGV607541H, NGV606173H, NGV611291H, NGV611292H, NGV611952H, NGV610441H, NGV610878H, NGV611648H, NGV613140H, NGV601959H, NGV603397H, NGV603459H, NGV603750H, NGV603767H, NGV604351H, NGV605154H, NGV605555H, NGV605710H, NGV607363H, NGV608430H, NGV608436H, NGV611043H, NGV601618H, NGV608370H, NGV606786H, NGV610793H, NGV603177H, NGV603178H, NGV603187H, NGV612748H, NGV605197H, NGV605411H, NGV606778H, NGV605228H, NGV605762H, NGV606069H, NGV601592H, NGV601614H, NGV601667H, NGV601671H, NGV601673H, NGV601674H, NGV606264H, NGV606265H, NGV607518H, NGV607519H, NGV607525H, NGV608437H, NGV608456H, NGV608459H, NGV605244H, NGV607283H, NGV610480H, NGV603487H, NGV603491H, NGV612059H, NGV603170H, NGV603335H, NGV603429H, NGV604671H, NGV608189H, NGV609614H, NGV609615H, NGV611219H, NGV611564H, NGV604341H, NGV611290H, NGV607306H, NGV608581H, NGV611682H, NGV603172H, NGV612510H, NGV604815H, NGV605073H, NGV605476H, NGV611547H, NGV611553H, NGV603611H, NGV603664H, NGV605585H, NGV608996H, NGV609003H, NGV611176H, NGV612762H, NGV603976H, NGV601692H, NGV605938H, NGV607271H, NGV608889H, NGV609236H, NGV611493H, NGV601128H, NGV601174H, NGV601179H, NGV601245H, NGV601562H, NGV601704H, NGV601706H, NGV605302H, NGV605495H, NGV605501H, NGV605532H, NGV605535H, NGV606900H, NGV609920H, NGV609928H, NGV611132H, NGV611138H, NGV612790H, NGV606038H, NGV606931H, NGV609211H, NGV610889H, NGV600827H, NGV611134H, NGV611135H, NGV610723H, NGV612800H, NGV601916H, NGV608643H, NGV608658H, NGV610456H, NGV611819H, NGV601282H, NGV601284H, NGV601936H, NGV603103H, NGV603473H, NGV606626H, NGV608899H, NGV611149H, NGV601598H, NGV603640H, NGV601990H, NGV606787H, NGV603083H, NGV603183H, NGV603484H, NGV606255H, NGV606797H, NGV607998H, NGV611996H, NGV609788H, NGV603617H, NGV606598H, NGV609150H, NGV601320H, NGV605864H, NGV606902H, NGV607732H, NGV607539H, NGV612353H, NGV606056H, NGV607642H, NGV609687H, NGV611378H, NGV600834H, NGV601095H, NGV601745H, NGV603795H, NGV607362H, NGV607522H, NGV607705H, NGV611160H, NGV603550H, NGV606234H, NGV606700H, NGV606773H, NGV607273H, NGV607569H, NGV601651H, NGV606849H, NGV611646H, NGV611129H, NGV611402H, NGV612222H, NGV611147H, NGV603810H, NGV608651H, NGV609141H, NGV611841H, NGV601274H, NGV606823H, NGV601664H, NGV608929H, NGV601340H, NGV605171H, NGV605408H, NGV607289H, NGV607687H, NGV607989H, NGV609825H, NGV612333H, NGV610647H, NGV610408H, NGV605888H, NGV608483H, NGV606129H, NGV606897H, NGV607327H, NGV611397H, NGV612322H, NGV612349H, NGV602570H, NGV606237H, NGV607686H, NGV611010H, NGV612001H, NGV603341H, NGV601186H, NGV610865H, NGV612154H, NGV612175H, NGV608004H, NGV608911H, NGV608913H, NGV609681H, NGV612443H, NGV601625H, NGV603433H, NGV603449H, NGV603450H, NGV605868H, NGV611119H, NGV611977H, NGV613209H, NGV601600H, NGV606949H, NGV607274H, NGV607317H, NGV609203H, NGV612255H, NGV612724H, NGV613163H, NGV601644H, NGV605983H, NGV607761H, NGV613128H, NGV601115H, NGV601611H, NGV608489H, NGV603049H, NGV601259H, NGV607335H, NGV609851H, NGV611126H, NGV611421H, NGV601682H, NGV606266H, NGV609186H, NGV611298H, NGV612124H, NGV603643H, NGV603651H, NGV603652H, NGV606652H, NGV609180H, NGV601246H, NGV604349H, NGV604653H, NGV604696H, NGV601291H, NGV601304H, NGV608565H, NGV603467H, NGV611974H, NGV606860H, NGV603087H, NGV604316H, NGV607559H, NGV609693H, NGV603638H, NGV605218H, NGV603182H, NGV611213H, NGV613158H, NGV612727H, NGV603338H, NGV605485H, NGV607788H, NGV607809H, NGV609968H, NGV611993H, NGV605181H, NGV613136H, NGV603522H, NGV603525H, NGV603540H, NGV603583H, NGV603586H, NGV605893H, NGV605895H, NGV605899H, NGV605901H, NGV605909H, NGV608976H, NGV608984H, NGV608985H, NGV608990H, NGV611190H, NGV611194H, NGV612789H, NGV612802H, NGV612809H, NGV603580H, NGV606190H, NGV605098H, NGV606906H, NGV605816H, NGV606799H, NGV611229H, NGV611425H, NGV604333H, NGV607766H, NGV607364H, NGV609672H, NGV606253H, NGV612769H, NGV613627H, NGV611039H, NGV604312H, NGV608975H, NGV601683H, NGV602569H, NGV601678H, NGV610440H, NGV603576H, NGV609713H, NGV604364H, NGV604645H, NGV611064H, NGV605210H, NGV607709H, NGV611649H, NGV612003H, NGV612008H, NGV612797H, NGV611195H, NGV612118H, NGV602028H, NGV607439H, NGV607574H, NGV608616H, NGV611012H, NGV611068H, NGV604353H, NGV603768H, NGV603112H, NGV605954H, NGV608259H, NGV611061H, NGV611680H, NGV601177H, NGV601195H, NGV603376H, NGV611556H, NGV612473H, NGV612507H, NGV607508H, NGV610455H, NGV610490H, NGV610778H, NGV610821H, NGV611184H, NGV611316H, NGV606019H, NGV606204H, NGV603478H, NGV611447H, NGV606018H, NGV610741H, NGV603059H, NGV603164H, NGV605562H, NGV605812H, NGV605407H, NGV612775H, NGV603973H, NGV605065H, NGV607609H, NGV611848H, NGV605987H, NGV606008H, NGV606135H, NGV608261H, NGV603163H, NGV604016H, NGV605239H, NGV606247H, NGV608256H, NGV603519H, NGV603897H, NGV606032H, NGV601658H, NGV603108H, NGV603126H, NGV603330H, NGV603432H, NGV603653H, NGV604004H, NGV606301H, NGV606488H, NGV606903H, NGV606962H, NGV607665H, NGV608474H, NGV608477H, NGV610828H, NGV611114H, NGV611637H, NGV611669H, NGV613125H, NGV603089H, NGV603061H, NGV604331H, NGV606165H, NGV612010H, NGV606192H, NGV602599H, NGV605395H, NGV608394H, NGV613635H, NGV608219H, NGV603169H, NGV606690H, NGV607358H, NGV603493H, NGV604309H, NGV607595H, NGV609890H, NGV611107H, NGV611492H, NGV613361H, NGV613362H, NGV608461H, NGV601604H, NGV611201H, NGV601189H, NGV601582H, NGV603095H, NGV609173H, NGV611534H, NGV611672H; GTIN 00643169999947, Serial Numbers: NGV602186H, NGV603942H, NGV603945H, NGV608428H, NGV609224H, NGV612014H, NGV613150H, NGV606012H, NGV606052H, NGV608328H, NGV610696H, NGV601677H, NGV603460H, NGV605439H, NGV607591H, NGV603342H, NGV603351H, NGV606057H, NGV607644H, NGV608014H, NGV608186H, NGV609152H, NGV611959H, NGV606570H, NGV612417H, NGV601109H, NGV601235H, NGV601564H, NGV601691H, NGV602242H, NGV602244H, NGV604539H, NGV605235H, NGV605417H, NGV605900H, NGV606043H, NGV606589H, NGV606712H, NGV606831H, NGV607342H, NGV608667H, NGV609156H, NGV609931H, NGV610843H, NGV610892H, NGV610899H, NGV611568H, NGV612208H, NGV612779H, NGV612796H, NGV603159H, NGV606064H, NGV601159H, NGV601676H, NGV601699H, NGV602169H, NGV603364H, NGV606041H, NGV608403H, NGV608522H, NGV609172H, NGV611615H, NGV612013H, NGV612498H, NGV601148H, NGV603326H, NGV607646H, NGV611639H, NGV612753H, NGV605845H, NGV607474H, NGV601757H, NGV605071H, NGV605878H, NGV601570H, NGV603363H, NGV603796H, NGV606670H, NGV608248H, NGV609694H, NGV601173H, NGV601316H, NGV601641H, NGV602014H, NGV603125H, NGV603843H, NGV603845H, NGV605579H, NGV605876H, NGV606749H, NGV606772H, NGV607573H, NGV607651H, NGV608028H, NGV608198H, NGV601088H, NGV602051H, NGV609199H, NGV609698H, NGV605898H, NGV606695H, NGV609558H, NGV610992H, NGV612390H, NGV612408H, NGV613090H, NGV606082H, NGV607576H, NGV606268H, NGV601597H, NGV612174H, NGV603382H, NGV608244H, NGV611216H, NGV602229H, NGV605863H, NGV606159H, NGV610853H, NGV606793H, NGV607674H, NGV610277H, NGV610442H, NGV612505H, NGV601281H, NGV601290H, NGV601654H, NGV601655H, NGV601656H, NGV601702H, NGV601720H, NGV601739H, NGV603122H, NGV603207H, NGV603331H, NGV603480H, NGV603483H, NGV603485H, NGV605068H, NGV605436H, NGV605473H, NGV605491H, NGV606109H, NGV606678H, NGV606689H, NGV607296H, NGV607505H, NGV607534H, NGV607626H, NGV607679H, NGV607700H, NGV608145H, NGV608201H, NGV608234H, NGV608257H, NGV608333H, NGV609699H, NGV609716H, NGV609747H, NGV609840H, NGV609910H, NGV610601H, NGV611223H, NGV611244H, NGV611287H, NGV611644H, NGV611988H, NGV612024H, NGV612402H, NGV612431H, NGV612435H, NGV612502H, NGV613154H, NGV603358H, NGV607297H, NGV612478H, NGV606818H, NGV608559H, NGV610888H, NGV612465H, NGV601743H, NGV603332H, NGV608053H, NGV610915H, NGV601752H, NGV602057H, NGV603884H, NGV607586H, NGV607969H, NGV610860H, NGV611408H, NGV611951H, NGV613111H, NGV612742H, NGV603367H, NGV601968H, NGV610893H, NGV611654H, NGV601157H, NGV601589H, NGV606040H, NGV611886H, NGV601268H, NGV601792H, NGV602027H, NGV603416H, NGV606282H, NGV611666H, NGV613173H, NGV601208H, NGV601610H, NGV602015H, NGV603383H, NGV603879H, NGV606050H, NGV606053H, NGV606725H, NGV608464H, NGV608556H, NGV609598H, NGV610876H, NGV611433H, NGV611998H, NGV602033H, NGV603502H, NGV603509H, NGV603883H, NGV606037H, NGV608488H, NGV609163H, NGV611577H, NGV610465H, NGV610425H, NGV606960H, NGV613331H, NGV606891H, NGV610336H, NGV610605H, NGV611293H, NGV606027H, NGV606868H, NGV606941H, NGV608531H, NGV611294H, NGV611546H, NGV612198H, NGV610635H, NGV612211H, NGV603362H, NGV606863H, NGV613200H, NGV610477H, NGV611642H, NGV613344H, NGV605565H, NGV607521H, NGV606912H, NGV605516H, NGV601967H, NGV605062H, NGV607783H, NGV608438H, NGV601977H, NGV603442H, NGV603452H, NGV605833H, NGV605838H, NGV606809H, NGV609900H, NGV610438H, NGV609225H, NGV602218H, NGV606224H, NGV603353H, NGV603499H, NGV602212H, NGV604304H, NGV604313H, NGV605099H, NGV606017H, NGV606034H, NGV611997H, NGV611461H; GTIN 00643169999954, Serial Numbers: NGV601077H, NGV601131H, NGV602023H, NGV603784H, NGV608385H, NGV608392H, NGV609215H, NGV610594H, NGV611526H, NGV612367H, NGV612388H, NGV612758H, NGV600934H, NGV601190H, NGV601675H, NGV604856H, NGV604881H, NGV605394H, NGV605568H, NGV605573H, NGV606179H, NGV606215H, NGV608449H, NGV609622H, NGV612030H, NGV606305H, NGV602047H, NGV600931H, NGV600933H, NGV604270H, NGV605143H, NGV605151H, NGV605879H, NGV606166H, NGV606172H, NGV608400H, NGV609658H, NGV609677H, NGV609830H, NGV609867H, NGV612418H, NGV613152H, NGV601974H, NGV601008H, NGV601613H, NGV605563H, NGV606114H, NGV606124H, NGV606169H, NGV604945H, NGV604951H, NGV605044H, NGV605340H, NGV605603H, NGV605849H, NGV606299H, NGV608419H, NGV612243H, NGV612357H, NGV612441H, NGV600836H, NGV600876H, NGV601176H, NGV602048H, NGV602551H, NGV602556H, NGV602587H, NGV602738H, NGV602803H, NGV603806H, NGV603812H, NGV603847H, NGV605908H, NGV608378H, NGV608382H, NGV608406H, NGV608452H, NGV608454H, NGV608910H, NGV608912H, NGV608917H, NGV608928H, NGV608978H, NGV608980H, NGV610464H, NGV610895H, NGV611023H, NGV611071H, NGV611495H, NGV611512H, NGV611668H, NGV611829H, NGV611924H, NGV612022H, NGV612385H, NGV612412H, NGV612422H, NGV612470H, NGV612474H.NGV612504H, NGV612509H, NGV612512H, NGV612716H, NGV613144H, NGV602603H, NGV603868H, NGV603941H, NGV606110H, NGV608457H, NGV610881H, NGV611234H, NGV611535H, NGV612383H, NGV612386H, NGV601143H, NGV601171H, NGV601995H, NGV602812H, NGV605126H, NGV605541H, NGV608425H, NGV608434H, NGV608451H, NGV608971H, NGV608988H, NGV609202H, NGV609873H, NGV610056H, NGV610914H, NGV612294H, NGV603871H, NGV605880H, NGV606113H, NGV608410H, NGV609595H, NGV610453H, NGV611477H, NGV612263H, NGV612429H, NGV602021H, NGV602531H, NGV604684H, NGV605815H, NGV608376H, NGV608420H, NGV608982H, NGV612151H, NGV612223H, NGV612303H, NGV612662H, NGV602725H, NGV609883H, NGV609891H, NGV611288H, NGV612416H, NGV600908H, NGV605873H, NGV605874H, NGV601647H, NGV603708H, NGV603741H, NGV603745H, NGV603746H, NGV605539H, NGV611578H, NGV611662H, NGV611663H, NGV601178H, NGV601188H, NGV601251H, NGV602025H, NGV602036H, NGV608375H, NGV610515H, NGV610842H, NGV611517H, NGV612358H, NGV600926H, NGV604278H, NGV606162H, NGV608374H, NGV610427H, NGV611226H, NGV611640H, NGV612791H, NGV608977H, NGV610449H, NGV610450H, NGV600920H, NGV606148H, NGV606219H, NGV610923H, NGV602600H, NGV605587H, NGV605591H, NGV605606H, NGV605790H, NGV605797H, NGV611013H, NGV611360H, NGV611374H, NGV605467H, NGV612611H, NGV600918H, NGV602012H, NGV603724H, NGV604429H, NGV605128H, NGV605737H, NGV608930H, NGV608940H, NGV611035H, NGV612277H, NGV612285H, NGV605857H, NGV606300H, NGV608433H, NGV609144H, NGV612332H, NGV612523H, NGV611513H, NGV605830H, NGV606238H, NGV606284H, NGV609632H, NGV611683H, NGV601669H, NGV608404H, NGV605836H, NGV605837H, NGV600937H, NGV601749H, NGV608444H, NGV609143H, NGV609146H, NGV609191H, NGV609230H, NGV612017H, NGV612047H, NGV612215H, NGV612316H, NGV612420H, NGV612793H, NGV602017H, NGV608085H, NGV608369H, NGV612036H, NGV612127H, NGV605826H, NGV608167H, NGV608179H, NGV608181H, NGV608188H, NGV608232H, NGV608249H, NGV608286H, NGV608455H, NGV608460H, NGV608514H, NGV608558H, NGV608592H, NGV608612H, NGV608633H, NGV608641H, NGV612012H, NGV612069H, NGV612152H, NGV605198H, NGV605243H, NGV605520H, NGV605525H, NGV605527H, NGV605801H, NGV605829H, NGV607345H, NGV607544H, NGV607860H, NGV608059H; GTIN 00643169999961, Serial Numbers: NGV604346H, NGV604541H, NGV609697H, NGV609869H, NGV609904H, NGV600899H, NGV603534H, NGV609861H, NGV601209H, NGV604339H, NGV605492H, NGV609705H, NGV603539H, NGV609857H, NGV600865H, NGV605497H, NGV609670H, NGV609892H, NGV612368H, NGV613580H, NGV600823H, NGV600867H, NGV600902H, NGV601065H, NGV601066H, NGV601151H, NGV601214H, NGV605438H, NGV605496H, NGV609702H, NGV609856H, NGV609858H, NGV609882H, NGV609886H, NGV609887H, NGV609923H, NGV613161H, NGV609695H, NGV603536H, NGV609913H, NGV609675H, NGV600928H, NGV612058H, NGV601292H, NGV606776H, NGV606934H, NGV607826H, NGV606916H, NGV612053H, NGV600905H, NGV608922H, NGV613153H, NGV552613H, NGV600929H, NGV600930H, NGV606168H, NGV611022H, NGV612055H, NGV612350H, NGV612352H, NGV613118H, NGV613135H, NGV601297H, NGV608468H, NGV608538H, NGV607365H, NGV612785H, NGV613134H, NGV601285H, NGV608887H, NGV612117H, NGV604320H, NGV606714H, NGV611021H, NGV606178H, NGV606744H, NGV601296H, NGV605465H, NGV605504H, NGV612351H, NGV612811H, NGV605470H, NGV605469H, NGV612046H, NGV612065H, NGV612116H, NGV606715H, NGV607510H, NGV611835H, NGV601313H, NGV552936H, NGV601078H, NGV601271H, NGV601272H, NGV601308H, NGV605460H, NGV606710H, NGV606747H, NGV607547H, NGV608447H, NGV608885H, NGV609151H, NGV610032H, NGV611884H, NGV612097H, NGV612361H, NGV601287H, NGV608654H, NGV612355H, NGV605456H, NGV601275H, NGV604335H, NGV611846H, NGV613036H, NGV606877H, NGV609626H, NGV601280H, NGV608591H, NGV606910H, NGV606936H, NGV600895H, NGV601273H, NGV606750H, NGV606940H, NGV607986H, NGV608669H, NGV612359H, NGV600901H, NGV601211H, NGV606732H, NGV607612H, NGV608931H, NGV613151H, NGV613156H, NGV601315H, NGV605448H, NGV605498H, NGV605502H, NGV606147H, NGV606217H, NGV606218H, NGV606774H, NGV607689H, NGV607708H, NGV607792H, NGV609120H, NGV610484H, NGV611255H, NGV612142H, NGV612410H, NGV613052H, NGV613139H, NGV608660H, NGV601299H, NGV604323H, NGV606757H, NGV608981H, NGV608979H, NGV611015H, NGV611057H, NGV601070H, NGV600915H, NGV600924H, NGV600950H, NGV601058H, NGV601300H, NGV608462H, NGV611016H, NGV612805H, NGV601084H, NGV601319H, NGV607533H, NGV608632H, NGV610554H, NGV610596H, NGV612126H, NGV606791H, NGV606876H, NGV608602H, NGV608634H, NGV610877H, NGV606630H, NGV600913H, NGV606958H, NGV610904H, NGV612071H, NGV609570H, NGV606656H, NGV605500H, NGV606197H, NGV607844H, NGV612360H, NGV604337H, NGV604355H, NGV605608H, NGV601289H, NGV600907H, NGV606789H, NGV606644H, NGV606911H, NGV608904H, NGV610997H, NGV607424H, NGV601253H, NGV603033H, NGV611603H, NGV611665H, NGV612089H, NGV601123H, NGV601135H, NGV601136H, NGV601146H, NGV601175H, NGV601181H, NGV601199H, NGV601205H, NGV601234H, NGV601240H, NGV601241H, NGV601244H, NGV601249H, NGV601255H, NGV601256H, NGV601260H, NGV601262H, NGV601265H, NGV602616H, NGV602617H, NGV602618H, NGV603605H, NGV603608H, NGV603618H, NGV603621H, NGV603631H, NGV603636H, NGV603641H, NGV611585H, NGV611586H, NGV611600H, NGV611675H, NGV611696H, NGV601091H, NGV601118H, NGV602578H, NGV602579H, NGV603027H, NGV611604H, NGV602536H, NGV602540H, NGV603098H, NGV601138H, NGV602964H, NGV601239H, NGV602728H, NGV602737H, NGV602857H, NGV611555H, NGV611661H, NGV612156H, NGV600996H, NGV601001H, NGV601221H, NGV601223H, NGV601741H, NGV602606H, NGV603647H, NGV610827H, NGV611626H, NGV612139H, NGV601105H, NGV601012H, NGV601028H, NGV601149H, NGV601215H, NGV603088H, NGV603581H, NGV603585H, NGV611612H, NGV611613H, NGV611618H, NGV612321H, NGV612325H, NGV612328H, NGV600970H, NGV600980H, NGV602816H, NGV601168H, NGV602530H, NGV611658H, NGV612104H, NGV612149H, NGV612150H, NGV602594H, NGV602601H, NGV602604H, NGV602931H, NGV602961H, NGV603579H, NGV603644H, NGV611664H, NGV611549H, NGV611558H, NGV601226H, NGV601147H, NGV601224H, NGV601668H, NGV602523H, NGV602793H, NGV602802H, NGV602817H, NGV603021H, NGV603041H, NGV611608H, NGV611630H, NGV611631H, NGV611674H, NGV611508H, NGV603047H, NGV603592H, NGV601002H, NGV600845H, NGV600968H, NGV601227H, NGV601229H, NGV601243H, NGV601248H, NGV602557H, NGV602564H, NGV602565H, NGV602795H, NGV603060H, NGV603074H, NGV603078H, NGV603106H, NGV603113H, NGV603115H, NGV611561H, NGV611579H, NGV611580H, NGV612096H, NGV612133H, NGV612159H, NGV612161H, NGV612167H, NGV603036H, NGV601061H, NGV601250H, NGV611491H, NGV611643H, NGV601062H, NGV601068H, NGV601076H, NGV601230H, NGV601257H, NGV601680H, NGV601681H, NGV611635H, NGV611641H, NGV601031H, NGV612129H, NGV602797H, NGV602962H, NGV605433H, NGV608416H, NGV608417H, NGV611849H, NGV602967H, NGV604651H, NGV606501H, NGV606537H, NGV610839H, NGV610886H, NGV604767H, NGV612084H, NGV600788H, NGV610600H, NGV609005H, NGV605167H, NGV600780H, NGV610704H, NGV610715H, NGV612318H, NGV610695H, NGV612099H, NGV604760H, NGV605162H, NGV605178H, NGV605188H, NGV605415H, NGV605420H, NGV610492H, NGV610684H, NGV610706H, NGV610712H, NGV611081H, NGV612092H, NGV612348H, NGV612373H, NGV604496H, NGV604504H, NGV605194H, NGV605428H, NGV610459H, NGV610568H, NGV612335H, NGV600762H, NGV600782H, NGV600808H, NGV604015H, NGV604764H, NGV605160H, NGV610685H, NGV610694H, NGV610698H, NGV610906H, NGV612372H, NGV605176H, NGV611018H, NGV612340H, NGV611700H, NGV605169H, NGV600807H, NGV612087H, NGV612342H, NGV604759H, NGV605413H, NGV606839H, NGV606841H, NGV608986H, NGV610516H, NGV600793H, NGV605429H, NGV610900H, NGV612346H, NGV604835H, NGV611007H, NGV610678H, NGV612145H, NGV607293H, NGV607310H, NGV607318H, NGV607347H, NGV610908H, NGV611300H, NGV611351H, NGV611379H, NGV611392H, NGV611393H, NGV611398H, NGV611415H, NGV611420H, NGV611431H, NGV611436H, NGV602764H, NGV610838H, NGV613617H, NGV604266H, NGV609535H, NGV609940H, NGV610018H, NGV611688H, NGV609195H, NGV611985H, NGV607147H, NGV607536H, NGV605227H, NGV605236H, NGV606854H, NGV601191H, NGV605129H, NGV607747H, NGV605220H, NGV610437H, NGV610786H, NGV612786H, NGV612804H, NGV609655H, NGV609673H, NGV601169H, NGV601206H, NGV606847H, NGV610873H, NGV611826H, NGV611943H, NGV612406H, NGV612794H, NGV613647H, NGV601184H, NGV605127H, NGV607582H, NGV607604H, NGV610836H, NGV611945H, NGV607580H, NGV607599H, NGV611703H, NGV609855H, NGV610457H, NGV610718H, NGV610855H, NGV610870H, NGV611949H, NGV612397H, NGV605229H, NGV606842H, NGV607734H, NGV602768H, NGV604310H, NGV608026H, NGV611941H, NGV605425H, NGV605430H, NGV605508H, NGV606235H, NGV611901H, NGV607302H, NGV607758H, NGV607759H, NGV602960H, NGV603419H, NGV603424H, NGV605175H, NGV608368H, NGV608371H, NGV603175H, NGV603181H, NGV608515H, NGV608530H, NGV603466H, NGV606143H, NGV601160H, NGV601213H, NGV608533H, NGV602788H, NGV602771H, NGV612781H, NGV612244H, NGV600830H, NGV603150H, NGV612314H, NGV602227H, NGV612235H, NGV600816H, NGV600818H, NGV606844H, NGV612146H, NGV612409H, NGV552756H, NGV552941H, NGV553234H, NGV600858H, NGV605503H, NGV605821H, NGV606146H, NGV606186H, NGV608540H, NGV608570H, NGV611025H, NGV611030H, NGV605444H, NGV612407H, NGV600795H, NGV607629H, NGV606138H, NGV606126H, NGV606141H, NGV605848H, NGV601222H, NGV601252H, NGV609142H, NGV611113H, NGV611127H, NGV611133H, NGV611136H, NGV611161H, NGV611180H, NGV611320H, NGV606836H, NGV606845H, NGV607373H, NGV607399H, NGV606834H, NGV603558H, NGV604852H, NGV606640H, NGV606641H, NGV606642H, NGV606655H, NGV606795H, NGV606802H, NGV607456H, NGV603549H, NGV605202H, NGV606634H, NGV606638H, NGV606814H, NGV606816H, NGV606821H, NGV606825H, NGV611087H, NGV607606H, NGV607632H, NGV600896H, NGV611166H, NGV603655H, NGV608355H, NGV601155H, NGV609440H, NGV602763H, NGV604481H, NGV604514H, NGV611187H, NGV611973H, NGV611980H, NGV612163H, NGV608469H, NGV611178H, NGV612009H, NGV604507H, NGV603425H, NGV603431H, NGV603435H, NGV605132H, NGV609500H, NGV605140H, NGV609617H, NGV601269H, NGV609220H, NGV609559H, NGV609649H, NGV601322H, NGV601328H, NGV606054H, NGV609226H, NGV609599H, NGV605134H, NGV605153H, NGV605155H, NGV606587H, NGV606599H, NGV607500H, NGV607590H, NGV609639H, NGV610699H, NGV612072H, NGV613192H, NGV607507H, NGV606223H, NGV609213H, NGV601337H, NGV607639H, NGV609638H, NGV606574H, NGV609210H, NGV601345H, NGV606579H, NGV609227H, NGV605141H, NGV605148H, NGV601321H, NGV601332H, NGV601335H, NGV609216H, NGV609646H, NGV613211H, NGV609541H, NGV609651H, NGV612090H, NGV601349H, NGV613168H, NGV605130H, NGV606182H, NGV610670H, NGV601333H, NGV607628H, NGV610389H, NGV610887H, NGV611882H, NGV607640H, NGV609451H, NGV603545H, NGV603949H, NGV604301H, NGV605206H, NGV605207H, NGV608176H, NGV608409H, NGV610655H, NGV606164H, NGV606174H, NGV609448H, NGV606707H, NGV607021H, NGV603687H, NGV603734H, NGV606480H, NGV606763H, NGV606812H, NGV609193H, NGV600839H, NGV601018H, NGV601033H, NGV601074H, NGV601162H, NGV601197H, NGV602217H, NGV602987H, NGV603674H, NGV603679H, NGV603685H, NGV603694H, NGV603696H, NGV603711H, NGV603758H, NGV605437H, NGV606145H, NGV606368H, NGV606672H, NGV607659H, NGV608010H, NGV608180H, NGV608187H, NGV610262H, NGV610321H, NGV610333H, NGV610348H, NGV610668H, NGV610907H, NGV611883H, NGV611937H, NGV611939H, NGV601019H, NGV607367H, NGV608172H, NGV601207H, NGV603979H, NGV608177H, NGV612452H, NGV607542H, NGV603456H, NGV608506H, NGV601217H, NGV608212H, NGV602203H, NGV610633H, NGV606745H, NGV608535H, NGV610500H, NGV607391H, NGV604297H, NGV605208H, NGV606271H, NGV606505H, NGV607636H, NGV611851H, NGV608319H, NGV610649H, NGV610614H, NGV603955H, NGV606502H, NGV607444H, NGV605410H, NGV605889H, NGV606608H, NGV606861H, NGV611950H, NGV612765H, NGV611874H, NGV604275H, NGV605201H, NGV610585H, NGV610903H, NGV611822H, NGV603212H, NGV607495H, NGV607583H, NGV609161H, NGV602940H, NGV605605H, NGV608379H, NGV610398H, NGV601079H, NGV603527H, NGV608335H, NGV601011H, NGV606850H, NGV611825H, NGV610639H, NGV610665H, NGV601134H, NGV606479H, NGV611832H, NGV609184H, NGV610671H, NGV610024H, NGV606123H, NGV603526H, NGV606603H, NGV610623H, NGV610631H, NGV611448H, NGV606523H, NGV600814H, NGV601163H, NGV603895H, NGV610653H, NGV611345H, NGV604466H, NGV611196H, NGV604475H, NGV608485H, NGV611185H, NGV611193H, NGV611198H, NGV611200H, NGV611306H, NGV611982H, NGV611987H, NGV604452H, NGV611990H, NGV612002H, NGV604704H, NGV602929H, NGV601029H, NGV601187H, NGV601216H, NGV603219H, NGV605414H, NGV606467H, NGV606560H, NGV608429H, NGV611259H, NGV611854H, NGV611896H, NGV603946H, NGV601043H, NGV601185H, NGV611910H, NGV611918H, NGV603145H, NGV603370H, NGV608431H, NGV610592H, NGV610823H, NGV610830H, NGV611931H, NGV600841H, NGV601158H, NGV602994H, NGV603410H, NGV608387H, NGV610922H, NGV603760H, NGV603763H, NGV603772H, NGV603809H, NGV603821H, NGV608238H, NGV608239H, NGV608241H, NGV608255H, NGV611953H, NGV611964H, NGV611965H, NGV611969H, NGV611970H, NGV612424H, NGV603218H, NGV603403H, NGV603409H, NGV610834H, NGV611380H, NGV611860H, NGV609155H, NGV609175H, NGV605818H, NGV603443H, NGV611277H, NGV603327H, NGV603339H, NGV603343H, NGV606617H, NGV606622H, NGV602230H, NGV603375H, NGV603384H, NGV604288H, NGV605419H, NGV606567H, NGV611934H, NGV610602H, NGV601087H, NGV601110H, NGV601124H, NGV601130H, NGV605412H, NGV605600H, NGV605604H, NGV607648H, NGV608204H, NGV608210H, NGV608443H, NGV611332H, NGV611339H, NGV611352H, NGV611373H, NGV612486H, NGV612490H, NGV603875H, NGV603880H, NGV603994H, NGV608442H, NGV610634H, NGV606544H, NGV609196H, NGV611828H, NGV610547H, NGV610550H, NGV610572H, NGV610610H, NGV607752H, NGV603128H, NGV603143H, NGV607647H, NGV607655H, NGV612457H, NGV606569H, NGV606575H, NGV612493H, NGV612760H, NGV612761H, NGV605855H, NGV611299H, NGV603377H, NGV607714H, NGV607719H, NGV600781H, NGV603365H, NGV609192H, NGV611912H, NGV603856H, NGV603874H, NGV608045H, NGV608215H, NGV608221H, NGV608233H, NGV603155H, NGV603401H, NGV603445H, NGV603889H, NGV610493H, NGV612434H, NGV602231H, NGV602952H, NGV603009H, NGV603190H, NGV603198H, NGV603210H, NGV610654H, NGV611475H; GTIN 00643169999978, Serial Numbers: NGV601228H, NGV601232H, NGV603582H, NGV603584H, NGV604952H, NGV610028H, NGV610697H, NGV603426H, NGV604271H, NGV604274H, NGV607735H, NGV609780H, NGV609801H, NGV610840H, NGV612187H, NGV610041H, NGV606901H, NGV603544H, NGV603559H, NGV606892H, NGV608546H, NGV609091H, NGV611932H, NGV612759H, NGV612766H, NGV602923H, NGV603517H, NGV603602H, NGV606753H, NGV606856H, NGV607286H, NGV607488H, NGV607553H, NGV607554H, NGV608916H, NGV609035H, NGV609486H, NGV609609H, NGV609718H, NGV612137H, NGV613013H, NGV605376H, NGV608926H, NGV608933H, NGV610913H, NGV611093H, NGV611115H, NGV611130H, NGV607469H, NGV605091H, NGV607284H, NGV602722H, NGV602739H, NGV602740H, NGV602766H, NGV602785H, NGV602947H, NGV602972H, NGV603018H, NGV603603H, NGV603645H, NGV604277H, NGV604282H, NGV604299H, NGV604302H, NGV604303H, NGV604314H, NGV604319H, NGV604342H, NGV604350H, NGV604354H, NGV604361H, NGV604362H, NGV605217H, NGV605240H, NGV605242H, NGV605249H, NGV605258H, NGV605269H, NGV605294H, NGV605380H, NGV605392H, NGV605403H, NGV605405H, NGV605481H, NGV605529H, NGV605534H, NGV600811H, NGV600846H, NGV608401H, NGV608411H, NGV608422H, NGV608424H, NGV610651H, NGV601233H, NGV601982H, NGV601989H, NGV602000H, NGV602786H, NGV602792H, NGV603046H, NGV603054H, NGV603075H, NGV603109H, NGV605106H, NGV605354H, NGV605359H, NGV607600H, NGV607616H, NGV607645H, NGV607658H, NGV607660H, NGV607662H, NGV608007H, NGV601976H, NGV602041H, NGV602045H, NGV603344H, NGV607564H, NGV606879H, NGV600809H, NGV600976H, NGV606781H, NGV606801H, NGV606829H, NGV606830H, NGV606851H, NGV606957H, NGV606959H, NGV611834H, NGV611847H, NGV611871H, NGV601969H, NGV601971H, NGV603322H, NGV603345H, NGV603385H, NGV606798H, NGV606917H, NGV606926H, NGV607477H, NGV608254H, NGV608473H, NGV608972H, NGV608997H, NGV611548H, NGV613018H, NGV604502H, NGV606921H, NGV608466H, NGV612343H, NGV601236H, NGV602002H, NGV602052H, NGV602522H, NGV603394H, NGV603407H, NGV604461H, NGV604462H, NGV604995H, NGV605010H, NGV605049H, NGV605055H, NGV605078H, NGV607683H, NGV607684H, NGV607717H, NGV608502H, NGV608906H, NGV608909H, NGV609465H, NGV609572H, NGV611528H, NGV612066H, NGV610826H, NGV612354H, NGV612405H, NGV602035H, NGV609956H, NGV610017H; GTIN 00643169999985, Serial Numbers: NGV605748H, NGV601046H, NGV601219H, NGV601639H, NGV602049H, NGV602933H, NGV602997H, NGV603020H, NGV603969H, NGV603971H, NGV603989H, NGV604267H, NGV604294H, NGV604298H, NGV605136H, NGV605152H, NGV605163H, NGV605173H, NGV605190H, NGV605221H, NGV605246H, NGV605396H, NGV605409H, NGV605627H, NGV608397H, NGV608661H, NGV609543H, NGV609567H, NGV600904H, NGV600906H, NGV600912H, NGV603201H, NGV603204H, NGV603206H, NGV603844H, NGV603869H, NGV603901H, NGV604332H, NGV605212H, NGV605247H, NGV608426H, NGV608440H, NGV608465H, NGV608947H, NGV608983H, NGV608987H, NGV608989H, NGV609113H, NGV609571H, NGV609678H, NGV609704H, NGV609719H, NGV609845H, NGV610871H, NGV610891H, NGV610909H, NGV610939H, NGV611117H, NGV611139H, NGV611170H, NGV611172H, NGV611318H, NGV611331H, NGV600844H, NGV602053H, NGV602056H, NGV602537H, NGV602566H, NGV602729H, NGV602796H, NGV603378H, NGV603572H, NGV603589H, NGV603591H, NGV603609H, NGV605045H, NGV605048H, NGV605080H, NGV605082H, NGV605108H, NGV605255H, NGV605341H, NGV605344H, NGV605353H, NGV605358H, NGV605387H, NGV605393H, NGV605521H, NGV606946H, NGV606961H, NGV608402H, NGV608427H, NGV608453H, NGV611048H, NGV611105H, NGV612744H, NGV612776H, NGV612777H, NGV603691H, NGV603718H, NGV603730H, NGV603785H, NGV603787H, NGV603797H, NGV603804H, NGV603808H, NGV607690H, NGV608169H, NGV608174H, NGV608216H, NGV608229H, NGV608524H, NGV608630H, NGV608650H, NGV611984H, NGV611334H, NGV611377H, NGV611684H, NGV611966H, NGV603906H, NGV600843H, NGV600946H, NGV600997H, NGV603820H, NGV605119H, NGV605214H, NGV605250H, NGV605251H, NGV605583H, NGV606929H, NGV608405H, NGV608973H, NGV609053H, NGV609579H, NGV609645H, NGV609906H, NGV610861H, NGV610995H, NGV611027H, NGV611075H, NGV600800H, NGV600914H, NGV603826H, NGV603952H, NGV603953H, NGV605789H, NGV608412H, NGV610832H, NGV603829H, NGV605117H, NGV605549H, NGV605759H, NGV605788H, NGV608432H, NGV600840H, NGV605548H, NGV612141H, NGV612147H, NGV612720H; GTIN 00763000000004, Serial Numbers: NGV608627H, NGV608646H, NGV608656H, NGV602986H, NGV603034H, NGV603042H, NGV603045H, NGV602800H, NGV605905H, NGV605963H, NGV606278H, NGV606288H, NGV606464H, NGV606478H, NGV606497H, NGV606504H, NGV606511H, NGV606533H, NGV608193H, NGV608194H, NGV608199H, NGV608209H, NGV608227H, NGV608228H, NGV608230H, NGV608235H, NGV608243H, NGV608245H, NGV608246H, NGV608250H, NGV608267H, NGV608519H, NGV608585H, NGV608588H, NGV608589H, NGV608601H, NGV608605H, NGV608606H, NGV608621H, NGV607321H, NGV607818H, NGV607864H, NGV608110H, NGV608170H, NGV608173H, NGV608184H, NGV608185H, NGV605586H, NGV605596H, NGV606538H, NGV606539H, NGV606542H, NGV608629H; GTIN 00763000000011, Serial Numbers: NGV605385H, NGV611671H, NGV605349H, NGV602046H, NGV600882H, NGV601044H, NGV602008H, NGV602029H, NGV605386H, NGV610476H, NGV610920H, NGV612741H, NGV612745H, NGV612747H, NGV600820H, NGV603333H, NGV606887H, NGV607834H, NGV610872H, NGV600900H, NGV604290H, NGV605374H, NGV610292H, NGV611407H, NGV605377H, NGV602968H, NGV603001H, NGV606963H, NGV613196H; GTIN 00763000000028, Serial Numbers: NGV609212H, NGV609749H, NGV609703H, NGV609751H, NGV610835H, NGV611046H, NGV611055H, NGV611074H, NGV611077H; GTIN 00763000000035, Serial Numbers: NGV603866H, NGV608191H, NGV603878H, NGV608208H, NGV605514H, NGV605157H, NGV603854H, NGV603877H, NGV605185H, NGV605537H, NGV605726H; GTIN 00763000018740, Serial Numbers: NGV605780H, NGV605551H, NGV607790H, NGV607791H, NGV606119H, NGV606837H, NGV607594H, NGV611283H, NGV611527H, NGV611249H, NGV605536H, NGV607530H, NGV609568H, NGV606163H, NGV605554H, NGV605598H, NGV605567H, NGV605592H, NGV607319H, NGV605783H, NGV607631H; GTIN 00763000018764, Serial Numbers: NGV602998H; GTIN 00763000080747, Serial Numbers: NGV605546H; GTIN 00763000122751, Serial Numbers: NGV001638H, NGV652501H, NGV711766H, NGV716008H, NGV716201H, NGV709188H, NGV000014H, NGV701935H, NGV704160H, NGV704163H, NGV704164H, NGV702652H, NGV005107H, NGV702845H, NGV703755H, NGV704547H, NGV708075H, NGV712119H, NGV712929H, NGV714061H, NGV714999H, NGV716480H, NGV721673H, NGV722140H, NGV723363H, NGV001401H, NGV709727H, NGV714933H, NGV722186H, NGV000030H, NGV000958H, NGV001362H, NGV704778H, NGV705236H, NGV706791H, NGV712250H, NGV713155H, NGV713161H, NGV714548H, NGV714773H, NGV714934H, NGV714935H, NGV714936H, NGV714938H, NGV714939H, NGV710884H, NGV703805H, NGV706489H, NGV708276H, NGV709732H, NGV000760H, NGV710417H, NGV650532H, NGV650548H, NGV650887H, NGV652021H, NGV652497H, NGV709385H, NGV710714H, NGV714106H, NGV723040H, NGV001393H, NGV701844H, NGV704779H, NGV713085H, NGV700517H, NGV000815H, NGV002314H, NGV002546H, NGV650757H, NGV651317H, NGV651599H, NGV651770H, NGV702825H, NGV702958H, NGV704000H, NGV704742H, NGV705774H, NGV707906H, NGV707916H, NGV708021H, NGV712934H, NGV712942H, NGV713472H, NGV714223H, NGV715364H, NGV716195H, NGV716393H, NGV716468H, NGV723416H, NGV700011H, NGV000941H, NGV001366H, NGV001376H, NGV002629H, NGV002763H, NGV003621H, NGV003802H, NGV650703H, NGV651899H, NGV652458H, NGV701467H, NGV704594H, NGV705049H, NGV705055H, NGV707829H, NGV710619H, NGV710843H, NGV712443H, NGV712900H, NGV712941H, NGV713944H, NGV714740H, NGV714749H, NGV714751H, NGV715062H, NGV715063H, NGV715075H, NGV716493H, NGV717709H, NGV720162H, NGV722195H, NGV722197H, NGV722210H, NGV722250H, NGV722328H, NGV722787H, NGV722849H, NGV002172H, NGV710684H, NGV650534H, NGV700095H, NGV705072H, NGV706558H, NGV711362H, NGV713415H, NGV713553H, NGV713658H, NGV713942H, NGV714914H, NGV718112H, NGV720233H, NGV720291H, NGV000015H, NGV001365H, NGV002365H, NGV003369H, NGV003651H, NGV004713H, NGV005139H, NGV650005H, NGV650343H, NGV650546H, NGV650562H, NGV650575H, NGV650670H, NGV650672H, NGV650959H, NGV651106H, NGV651578H, NGV652023H, NGV652050H, NGV701473H, NGV702254H, NGV702255H, NGV702522H, NGV702769H, NGV702818H, NGV702821H, NGV702883H, NGV704147H, NGV704798H, NGV704816H, NGV705916H, NGV706523H, NGV706585H, NGV709163H, NGV710630H, NGV711668H, NGV713420H, NGV713566H, NGV713724H, NGV714776H, NGV714858H, NGV715022H, NGV716310H, NGV716324H, NGV720578H, NGV722168H, NGV722596H, NGV723387H, NGV716127H, NGV709398H, NGV716139H, NGV000016H, NGV001412H, NGV001606H, NGV002607H, NGV002873H, NGV002884H, NGV003633H, NGV651030H, NGV651951H, NGV652370H, NGV652560H, NGV700001H, NGV700532H, NGV703389H, NGV706002H, NGV709384H, NGV710801H, NGV711767H, NGV712422H, NGV714237H, NGV714963H, NGV715015H, NGV722376H, NGV707806H, NGV001647H, NGV003115H, NGV003373H, NGV003400H, NGV004693H, NGV700089H, NGV700805H, NGV702804H, NGV705003H, NGV705767H, NGV706153H, NGV706516H, NGV708073H, NGV708138H, NGV710708H, NGV710800H, NGV710968H, NGV712278H, NGV713089H, NGV713090H, NGV713118H, NGV713555H, NGV713941H, NGV714376H, NGV714499H, NGV714846H, NGV714915H, NGV715357H, NGV715995H, NGV716735H, NGV718668H, NGV719551H, NGV722494H, NGV722597H, NGV722620H, NGV003519H, NGV703705H, NGV706494H, NGV700183H, NGV707905H, NGV720150H, NGV723129H, NGV003272H, NGV712258H, NGV714482H, NGV719585H, NGV722248H, NGV650289H, NGV702238H, NGV702361H, NGV706134H, NGV714538H, NGV715060H, NGV715370H, NGV718173H, NGV001555H, NGV002352H, NGV002369H, NGV002426H, NGV002718H, NGV003510H, NGV004665H, NGV700528H, NGV700844H, NGV701874H, NGV701927H, NGV702244H, NGV702833H, NGV703543H, NGV703744H, NGV706007H, NGV706061H, NGV706064H, NGV706066H, NGV706147H, NGV706800H, NGV709366H, NGV710490H, NGV712088H, NGV712089H, NGV713109H, NGV713110H, NGV713412H, NGV713748H, NGV713750H, NGV713795H, NGV713797H, NGV713876H, NGV714375H, NGV714470H, NGV714746H, NGV714788H, NGV714789H, NGV714797H, NGV714891H, NGV715170H, NGV716143H, NGV716144H, NGV716392H, NGV720301H, NGV721451H, NGV721727H, NGV721730H, NGV722166H, NGV722175H, NGV722181H, NGV722224H, NGV722274H, NGV722355H, NGV000953H, NGV002368H, NGV706583H, NGV706821H, NGV707930H, NGV712161H, NGV713871H, NGV714786H, NGV720759H, NGV721073H, NGV650671H, NGV706142H, NGV710511H, NGV002270H, NGV003329H, NGV003331H, NGV004726H, NGV716215H, NGV720758H, NGV002744H, NGV002765H, NGV701769H, NGV709658H, NGV713482H, NGV002000H, NGV002761H, NGV650022H, NGV709336H, NGV714670H, NGV714842H, NGV700035H, NGV002010H, NGV002012H, NGV652588H, NGV652593H, NGV713490H, NGV714885H, NGV715380H, NGV716316H, NGV716494H, NGV716496H, NGV722208H, NGV703725H, NGV705054H, NGV709435H, NGV719596H, NGV719604H, NGV719608H, NGV650667H, NGV700410H, NGV703484H, NGV703737H, NGV704740H, NGV708278H, NGV713859H, NGV715021H, NGV720218H, NGV723117H, NGV723368H, NGV000804H, NGV001612H, NGV002060H, NGV705075H, NGV710603H, NGV710689H, NGV712169H, NGV712985H, NGV713789H, NGV714985H, NGV717459H, NGV721658H, NGV002784H, NGV650729H, NGV652048H, NGV704821H, NGV713798H, NGV001026H, NGV001216H, NGV001282H, NGV002174H, NGV002496H, NGV002895H, NGV002897H, NGV003323H, NGV003325H, NGV003817H, NGV005088H, NGV005140H, NGV005147H, NGV005170H, NGV005173H, NGV005176H, NGV650032H, NGV650041H, NGV650872H, NGV651602H, NGV651613H, NGV651775H, NGV651985H, NGV652229H, NGV700290H, NGV700293H, NGV700295H, NGV702772H, NGV702964H, NGV704041H, NGV704042H, NGV704571H, NGV704592H, NGV704741H, NGV705071H, NGV705315H, NGV707828H, NGV708132H, NGV708137H, NGV708387H, NGV708388H, NGV708417H, NGV708418H, NGV708420H, NGV709465H, NGV710454H, NGV710550H, NGV710585H, NGV710784H, NGV711365H, NGV711369H, NGV711370H, NGV712935H, NGV712947H, NGV713370H, NGV713782H, NGV713914H, NGV714716H, NGV714720H, NGV714812H, NGV715102H, NGV715136H, NGV715326H, NGV715692H, NGV715997H, NGV716409H, NGV716412H, NGV716421H, NGV716474H, NGV716479H, NGV716737H, NGV716741H, NGV716742H, NGV716808H, NGV718649H, NGV719309H, NGV719599H, NGV720048H, NGV720064H, NGV721562H, NGV721722H, NGV721723H, NGV722337H, NGV722338H, NGV723070H, NGV723385H, NGV001160H, NGV652300H, NGV703800H, NGV710562H, NGV722151H, NGV723120H, NGV712928H, NGV713406H, NGV713407H, NGV714529H, NGV000969H, NGV001604H, NGV001605H, NGV001811H, NGV002427H, NGV002766H, NGV700029H, NGV701500H, NGV705078H, NGV710576H, NGV711764H, NGV720774H, NGV723384H, NGV003216H, NGV702971H, NGV703509H, NGV703513H, NGV706776H, NGV707837H, NGV709448H, NGV710545H, NGV710557H, NGV710602H, NGV711772H, NGV713411H, NGV713760H, NGV714555H, NGV719047H, NGV719623H, NGV001281H, NGV002241H, NGV650142H, NGV702870H, NGV704768H, NGV705044H, NGV713907H, NGV714506H, NGV714674H, NGV715055H, NGV715337H, NGV710671H, NGV721653H, NGV650147H, NGV701667H, NGV712075H, NGV722819H, NGV002329H, NGV650950H, NGV705376H, NGV713226H, NGV713227H, NGV714036H, NGV715049H, NGV715053H, NGV715335H, NGV716171H, NGV719300H, NGV719482H, NGV723062H, NGV650194H, NGV701129H, NGV702251H, NGV706068H, NGV706127H, NGV711768H, NGV714655H, NGV714671H, NGV715118H, NGV722408H, NGV000950H, NGV701105H, NGV701441H, NGV703507H, NGV706047H, NGV714659H, NGV000959H, NGV700002H, NGV701448H, NGV704849H, NGV706318H, NGV714038H, NGV714562H, NGV716147H, NGV716485H, NGV721450H, NGV002004H, NGV002881H, NGV004677H, NGV005116H, NGV650150H, NGV650739H, NGV701937H, NGV703469H, NGV704803H, NGV704804H, NGV707737H, NGV707824H, NGV709343H, NGV709345H, NGV713385H, NGV714663H, NGV714665H, NGV715220H, NGV715923H, NGV716129H, NGV722310H, NGV722354H, NGV722538H, NGV722543H, NGV723065H, NGV715083H, NGV002272H, NGV002752H, NGV717766H, NGV713103H, NGV715366H, NGV710657H, NGV720240H, NGV720264H, NGV721454H, NGV722244H, NGV714919H, NGV716131H, NGV650111H, NGV650683H, NGV706073H, NGV713422H, NGV713433H, NGV714052H, NGV714059H, NGV714763H, NGV714764H, NGV715104H, NGV715105H, NGV715605H, NGV720419H, NGV000482H, NGV000827H, NGV002827H, NGV003019H, NGV004586H, NGV004702H, NGV005101H, NGV650868H, NGV650998H, NGV651435H, NGV652454H, NGV701398H, NGV702279H, NGV703515H, NGV703875H, NGV705727H, NGV705754H, NGV708390H, NGV709721H, NGV710558H, NGV710616H, NGV710853H, NGV710985H, NGV712884H, NGV712888H, NGV713065H, NGV713111H, NGV713147H, NGV713453H, NGV714457H, NGV714471H, NGV714475H, NGV714536H, NGV714677H, NGV714725H, NGV714730H, NGV714756H, NGV714784H, NGV714790H, NGV714814H, NGV715117H, NGV715215H, NGV717564H, NGV722286H, NGV722619H, NGV002840H, NGV650511H, NGV651598H, NGV652635H, NGV702781H, NGV702909H, NGV704137H, NGV709466H, NGV712171H, NGV712997H, NGV713458H, NGV713879H, NGV714037H, NGV716388H, NGV716506H, NGV718090H, NGV003032H, NGV652599H, NGV700938H, NGV701003H, NGV709390H, NGV709391H, NGV710568H, NGV710575H, NGV710874H, NGV712173H, NGV712920H, NGV713397H, NGV713400H, NGV713454H, NGV714917H, NGV714989H, NGV715338H, NGV716371H, NGV716461H, NGV720764H, NGV001306H, NGV001519H, NGV002351H, NGV003082H, NGV003634H, NGV650243H, NGV650475H, NGV650854H, NGV651601H, NGV700519H, NGV701088H, NGV701938H, NGV703468H, NGV703749H, NGV703872H, NGV706017H, NGV706126H, NGV706540H, NGV707807H, NGV708060H, NGV710872H, NGV712448H, NGV713189H, NGV713239H, NGV713324H, NGV714416H, NGV714772H, NGV715323H, NGV715324H, NGV716470H, NGV719401H, NGV719404H, NGV719433H, NGV719595H, NGV720143H, NGV720773H, NGV721734H, NGV722146H, NGV722482H, NGV722616H, NGV723108H, NGV723406H, NGV001544H, NGV700269H, NGV701770H, NGV715328H, NGV716190H, NGV003033H, NGV651594H, NGV708022H, NGV709197H, NGV713164H, NGV714400H, NGV715001H, NGV722369H, NGV708259H, NGV001533H, NGV002526H, NGV002874H, NGV003392H, NGV004270H, NGV004619H, NGV650270H, NGV651783H, NGV651954H, NGV651979H, NGV652017H, NGV702972H, NGV703463H, NGV703797H, NGV705738H, NGV706502H, NGV708275H, NGV710580H, NGV712112H, NGV713398H, NGV713756H, NGV715010H, NGV715890H, NGV718218H, NGV718222H, NGV722203H, NGV722610H, NGV004582H, NGV703307H, NGV703322H, NGV000002H, NGV001167H, NGV001490H, NGV002669H, NGV003396H, NGV003562H, NGV003747H, NGV004598H, NGV004905H, NGV650160H, NGV706019H, NGV708273H, NGV708274H, NGV709189H, NGV709726H, NGV710503H, NGV710556H, NGV710827H, NGV713852H, NGV714553H, NGV714739H, NGV715219H, NGV716165H, NGV716787H, NGV719450H, NGV720000H, NGV722514H, NGV722524H, NGV723405H, NGV720202H, NGV722813H, NGV723389H, NGV712903H, NGV000294H, NGV650341H, NGV651777H, NGV701426H, NGV707928H, NGV712991H, NGV714119H, NGV714515H, NGV715025H, NGV715322H, NGV718495H, NGV001199H, NGV002327H, NGV003086H, NGV005197H, NGV650174H, NGV651125H, NGV700016H, NGV700018H, NGV703741H, NGV706538H, NGV708054H, NGV710447H, NGV710690H, NGV716331H, NGV719338H, NGV722641H, NGV723123H, NGV002251H, NGV002478H, NGV700015H, NGV703850H, NGV705760H, NGV720791H, NGV722275H, NGV003335H, NGV003558H, NGV005090H, NGV708077H, NGV717714H, NGV713383H, NGV714968H, NGV715003H, NGV712280H, NGV720221H, NGV001322H, NGV001545H, NGV002013H, NGV002725H, NGV003327H, NGV004055H, NGV004920H, NGV651663H, NGV652272H, NGV703496H, NGV706568H, NGV710637H, NGV710808H, NGV710817H, NGV713252H, NGV714765H, NGV715087H, NGV716434H, NGV716790H, NGV718277H, NGV720765H, NGV720832H, NGV003250H, NGV003542H, NGV003631H, NGV004628H, NGV650519H, NGV700274H, NGV702349H, NGV703521H, NGV703751H, NGV707765H, NGV710734H, NGV712907H, NGV713413H, NGV719397H, NGV722390H, NGV700972H, NGV703544H, NGV713465H, NGV716141H, NGV718100H, NGV722152H, NGV723407H, NGV700854H, NGV700935H, NGV710572H, NGV710574H, NGV714981H, NGV715098H, NGV720102H, NGV720114H, NGV001163H, NGV001423H, NGV001486H, NGV001603H, NGV001660H, NGV002559H, NGV003101H, NGV003452H, NGV004575H, NGV650151H, NGV650195H, NGV650197H, NGV650756H, NGV700949H, NGV705232H, NGV706819H, NGV709400H, NGV710507H, NGV710573H, NGV713719H, NGV714713H, NGV715106H, NGV715113H, NGV715120H, NGV716149H, NGV716243H, NGV719587H, NGV720147H, NGV720837H, NGV715678H, NGV000936H, NGV001467H, NGV650997H, NGV651802H, NGV700008H, NGV700271H, NGV704806H, NGV707763H, NGV708035H, NGV710835H, NGV712181H, NGV713444H, NGV716133H, NGV716330H, NGV716367H, NGV719480H, NGV721505H, NGV721507H, NGV000749H, NGV000937H, NGV001389H, NGV001457H, NGV002654H, NGV002667H, NGV002668H, NGV003046H, NGV003414H, NGV650822H, NGV651551H, NGV652297H, NGV701269H, NGV701751H, NGV701807H, NGV702771H, NGV706150H, NGV709428H, NGV714683H, NGV714844H, NGV714987H, NGV715433H, NGV716341H, NGV716406H, NGV719614H, NGV707755H, NGV001639H, NGV701266H, NGV703748H, NGV710566H, NGV712923H, NGV001426H, NGV001427H, NGV650561H, NGV650753H, NGV650771H, NGV650859H, NGV650878H, NGV650988H, NGV700278H, NGV700376H, NGV700380H, NGV701390H, NGV703900H, NGV704177H, NGV704835H, NGV706009H, NGV708344H, NGV708409H, NGV709404H, NGV709724H, NGV712917H, NGV713726H, NGV714466H, NGV714745H, NGV715084H, NGV715095H, NGV715339H, NGV716723H, NGV717443H, NGV717444H, NGV719481H, NGV000023H, NGV000940H, NGV000956H, NGV001345H, NGV001403H, NGV001429H, NGV001466H, NGV001488H, NGV001508H, NGV001625H, NGV001627H, NGV001641H, NGV002044H, NGV002074H, NGV002299H, NGV002353H, NGV002431H, NGV002486H, NGV002602H, NGV002614H, NGV002875H, NGV003017H, NGV003038H, NGV003354H, NGV003407H, NGV003497H, NGV003539H, NGV003611H, NGV003624H, NGV004537H, NGV004667H, NGV005093H, NGV005129H, NGV005135H, NGV005223H, NGV650154H, NGV650254H, NGV650267H, NGV650442H, NGV650474H, NGV650493H, NGV650525H, NGV650550H, NGV650719H, NGV650746H, NGV650799H, NGV650802H, NGV650910H, NGV650914H, NGV651565H, NGV651644H, NGV651776H, NGV651940H, NGV651941H, NGV652035H, NGV652299H, NGV652407H, NGV652522H, NGV652527H, NGV700948H, NGV701373H, NGV701394H, NGV701416H, NGV701672H, NGV701752H, NGV701803H, NGV702347H, NGV702873H, NGV702874H, NGV702877H, NGV702882H, NGV702906H, NGV703083H, NGV703460H, NGV703461H, NGV703516H, NGV703528H, NGV703722H, NGV703729H, NGV703779H, NGV703860H, NGV704046H, NGV704047H, NGV704053H, NGV704055H, NGV704184H, NGV704242H, NGV704250H, NGV704544H, NGV704730H, NGV704734H, NGV704754H, NGV704765H, NGV705041H, NGV705993H, NGV705997H, NGV706484H, NGV706490H, NGV706491H, NGV706493H, NGV707748H, NGV707915H, NGV708067H, NGV708071H, NGV708143H, NGV708144H, NGV708279H, NGV708398H, NGV708424H, NGV709331H, NGV709361H, NGV709372H, NGV710468H, NGV710510H, NGV710594H, NGV710610H, NGV710695H, NGV710699H, NGV710705H, NGV710719H, NGV710720H, NGV710813H, NGV710815H, NGV710825H, NGV710885H, NGV711657H, NGV711676H, NGV712157H, NGV712179H, NGV712263H, NGV712269H, NGV712880H, NGV712882H, NGV712918H, NGV712926H, NGV712987H, NGV713058H, NGV713060H, NGV713086H, NGV713126H, NGV713130H, NGV713432H, NGV713438H, NGV713450H, NGV713474H, NGV713545H, NGV713746H, NGV713764H, NGV713850H, NGV714109H, NGV714240H, NGV714384H, NGV714540H, NGV714554H, NGV714644H, NGV714692H, NGV714695H, NGV714701H, NGV714712H, NGV714714H, NGV714757H, NGV714845H, NGV714876H, NGV714879H, NGV714904H, NGV714907H, NGV714928H, NGV714931H, NGV715023H, NGV715048H, NGV715074H, NGV715078H, NGV715080H, NGV715194H, NGV715195H, NGV715210H, NGV715279H, NGV715378H, NGV715450H, NGV715459H, NGV715461H, NGV715599H, NGV715600H, NGV715727H, NGV715769H, NGV716154H, NGV716309H, NGV716317H, NGV716325H, NGV716339H, NGV716366H, NGV716815H, NGV717436H, NGV717561H, NGV717786H, NGV718039H, NGV718067H, NGV718071H, NGV718091H, NGV718243H, NGV718263H, NGV718271H, NGV719165H, NGV719394H, NGV720277H, NGV720280H, NGV721735H, NGV722133H, NGV722144H, NGV722213H, NGV722282H, NGV722293H, NGV722365H, NGV722366H, NGV722547H, NGV722600H, NGV722662H, NGV722781H, NGV722823H, NGV723084H, NGV723087H, NGV723390H, NGV723391H, NGV723403H, NGV001547H, NGV650448H, NGV651662H, NGV651961H, NGV652024H, NGV700394H, NGV701929H, NGV703905H, NGV706543H, NGV706544H, NGV706545H, NGV706546H, NGV709328H, NGV709630H, NGV710614H, NGV712271H, NGV712284H, NGV713149H, NGV713214H, NGV713761H, NGV713911H, NGV714753H, NGV716724H, NGV716727H, NGV722206H, NGV700006H, NGV701221H, NGV701626H, NGV701738H, NGV703743H, NGV000899H, NGV000962H, NGV001359H, NGV001616H, NGV002038H, NGV002670H, NGV003097H, NGV004705H, NGV005226H, NGV005232H, NGV005249H, NGV700956H, NGV701134H, NGV701408H, NGV706787H, NGV707825H, NGV709196H, NGV709257H, NGV710628H, NGV712446H, NGV715147H, NGV716170H, NGV720237H, NGV000011H, NGV000928H, NGV000929H, NGV000932H, NGV001383H, NGV001618H, NGV002658H, NGV003016H, NGV702778H, NGV703489H, NGV706625H, NGV710724H, NGV710732H, NGV712257H, NGV713544H, NGV716363H, NGV717432H, NGV718653H, NGV708355H, NGV710607H, NGV710852H, NGV710889H, NGV719635H, NGV650295H, NGV722500H, NGV715190H, NGV001493H, NGV650682H, NGV651903H, NGV701748H, NGV703795H, NGV708397H, NGV709408H, NGV714752H, NGV714754H, NGV002381H, NGV002382H, NGV003513H, NGV650455H, NGV705317H, NGV708017H, NGV708058H, NGV710460H, NGV710531H, NGV710691H, NGV711675H, NGV712177H, NGV713194H, NGV713234H, NGV713242H, NGV713563H, NGV713868H, NGV714363H, NGV714637H, NGV716166H, NGV716503H, NGV718069H, NGV719367H, NGV719616H, NGV722135H, NGV722212H, NGV722363H, NGV723419H, NGV002757H, NGV002843H, NGV003100H, NGV003383H, NGV005118H, NGV650389H, NGV700995H, NGV701263H, NGV701389H, NGV702464H, NGV705796H, NGV708347H, NGV710542H, NGV710688H, NGV711782H, NGV713369H, NGV713395H, NGV713720H, NGV714035H, NGV714525H, NGV714948H, NGV715018H, NGV719393H, NGV719617H, NGV723071H, NGV000819H, NGV001785H, NGV650345H, NGV650504H, NGV651763H, NGV652418H, NGV700375H, NGV700848H, NGV701400H, NGV701758H, NGV702961H, NGV704045H, NGV705742H, NGV705984H, NGV706499H, NGV708147H, NGV708399H, NGV709368H, NGV710722H, NGV712106H, NGV713434H, NGV714519H, NGV715886H, NGV716337H, NGV716674H, NGV717447H, NGV717782H, NGV703788H, NGV001656H, NGV002656H, NGV002831H, NGV003244H, NGV003312H, NGV003395H, NGV003525H, NGV003845H, NGV004571H, NGV004617H, NGV004632H, NGV004903H, NGV004904H, NGV650279H, NGV650464H, NGV650520H, NGV650714H, NGV650762H, NGV650884H, NGV651321H, NGV652471H, NGV652480H, NGV700443H, NGV701032H, NGV701621H, NGV702356H, NGV702373H, NGV702374H, NGV702793H, NGV702849H, NGV703752H, NGV703753H, NGV703754H, NGV704836H, NGV706534H, NGV706792H, NGV707833H, NGV709407H, NGV709445H, NGV710415H, NGV710456H, NGV710563H, NGV710633H, NGV710635H, NGV711371H, NGV712415H, NGV713091H, NGV713225H, NGV714039H, NGV714354H, NGV714820H, NGV714833H, NGV716418H, NGV716713H, NGV718106H, NGV722362H, NGV722427H, NGV722457H, NGV722468H, NGV722469H, NGV722470H, NGV650500H, NGV001611H, NGV003447H, NGV702457H, NGV703786H, NGV706587H, NGV709401H, NGV709705H, NGV714865H, NGV718284H, NGV710789H, NGV650531H, NGV702957H, NGV710547H, NGV711750H, NGV714767H, NGV715431H, NGV650446H, NGV652294H, NGV701113H, NGV704007H, NGV709720H, NGV714513H, NGV714759H, NGV714848H, NGV000933H, NGV702834H, NGV712902H, NGV715143H, NGV717431H, NGV722137H, NGV002169H, NGV000753H, NGV004690H, NGV706552H, NGV708335H, NGV722825H, NGV650184H, NGV651994H, NGV001347H, NGV003012H, NGV005195H, NGV650287H, NGV650314H, NGV651584H, NGV651984H, NGV700810H, NGV701132H, NGV701850H, NGV702630H, NGV702631H, NGV703331H, NGV703391H, NGV703736H, NGV704735H, NGV705370H, NGV706129H, NGV706571H, NGV706573H, NGV707738H, NGV707746H, NGV708009H, NGV708046H, NGV708436H, NGV710786H, NGV710787H, NGV712233H, NGV712237H, NGV713728H, NGV713943H, NGV714723H, NGV714734H, NGV714770H, NGV716369H, NGV717558H, NGV718085H, NGV002278H, NGV000522H, NGV710596H, NGV711781H, NGV713168H, NGV713170H, NGV713172H, NGV714925H, NGV714926H, NGV716329H, NGV718064H, NGV720702H, NGV722276H, NGV003292H, NGV713812H, NGV714166H, NGV001666H, NGV707933H, NGV709399H, NGV710443H, NGV710570H, NGV712082H, NGV714800H, NGV714813H, NGV722189H, NGV002852H, NGV710721H, NGV712292H, NGV001471H, NGV001472H, NGV002604H, NGV002850H, NGV003247H, NGV004637H, NGV650157H, NGV650208H, NGV650496H, NGV650497H, NGV652295H, NGV700853H, NGV701765H, NGV704022H, NGV704136H, NGV706074H, NGV707913H, NGV709640H, NGV710459H, NGV710465H, NGV710834H, NGV711659H, NGV711663H, NGV714758H, NGV714930H, NGV714976H, NGV715004H, NGV715065H, NGV715081H, NGV715432H, NGV715675H, NGV716424H, NGV719365H, NGV719578H, NGV720036H, NGV702908H, NGV705781H, NGV706565H, NGV650552H, NGV000008H, NGV001633H, NGV003018H, NGV003864H, NGV650206H, NGV650304H, NGV650492H, NGV650555H, NGV651636H, NGV651976H, NGV701124H, NGV701507H, NGV702646H, NGV702832H, NGV704805H, NGV704809H, NGV705755H, NGV706076H, NGV706485H, NGV706802H, NGV707899H, NGV708141H, NGV708393H, NGV708430H, NGV710655H, NGV710718H, NGV710975H, NGV710976H, NGV711754H, NGV711777H, NGV712241H, NGV712295H, NGV712324H, NGV712960H, NGV713068H, NGV713752H, NGV714359H, NGV714761H, NGV714847H, NGV715108H, NGV715898H, NGV716161H, NGV717763H, NGV719364H, NGV719999H, NGV721547H, NGV722187H, NGV723135H, NGV000521H, NGV000938H, NGV712096H, NGV712252H, NGV716319H, NGV004653H, NGV650945H, NGV703467H, NGV710832H, NGV713745H, NGV713860H, NGV714009H, NGV715036H, NGV720781H, NGV003298H, NGV651426H, NGV651437H, NGV700164H, NGV704157H, NGV707927H, NGV710653H, NGV712285H, NGV712296H, NGV713211H, NGV714408H, NGV714520H, NGV000519H, NGV001144H, NGV001646H, NGV002775H, NGV700321H, NGV700971H, NGV701406H, NGV702840H, NGV703494H, NGV706492H, NGV710638H, NGV710683H, NGV710685H, NGV710856H, NGV713794H, NGV714460H, NGV717568H, NGV720002H, NGV720840H, NGV722233H, NGV000032H, NGV701133H, NGV715114H, NGV716390H, NGV003334H, NGV652432H, NGV708011H, NGV000799H, NGV002292H, NGV002641H, NGV002653H, NGV002760H, NGV003030H, NGV003528H, NGV004491H, NGV650436H, NGV701369H, NGV701665H, NGV703499H, NGV703740H, NGV703760H, NGV706059H, NGV708074H, NGV709427H, NGV709654H, NGV710524H, NGV710577H, NGV712273H, NGV712904H, NGV712989H, NGV713765H, NGV713870H, NGV714483H, NGV714688H, NGV714708H, NGV714711H, NGV714897H, NGV715088H, NGV716158H, NGV717765H, NGV718399H, NGV719216H, NGV719452H, NGV720234H, NGV720771H, NGV721509H, NGV722192H, NGV723113H, NGV000751H, NGV001455H, NGV002468H, NGV002617H, NGV701742H, NGV706579H, NGV002242H, NGV708423H, NGV709431H, NGV713105H, NGV000857H, NGV713468H, NGV650268H, NGV713452H, NGV003041H, NGV701842H, NGV714703H, NGV720768H, NGV722478H, NGV713791H, NGV714918H, NGV002175H, NGV000353H, NGV722357H, NGV723337H, NGV721642H, NGV700534H, NGV702976H, NGV712945H, NGV701604H, NGV001535H, NGV701474H, NGV706576H, NGV710636H, NGV716152H, NGV700103H, NGV700843H, NGV709936H, NGV710858H, NGV712079H, NGV712081H, NGV713784H, NGV714222H, NGV714960H, NGV718497H, NGV719597H, NGV719620H, NGV722243H, NGV708044H, NGV715683H, NGV718056H, NGV700366H, NGV000356H, NGV001636H, NGV650216H, NGV702783H, NGV703546H, NGV704190H, NGV710527H, NGV712374H, NGV719172H, NGV002855H, NGV704008H, NGV003607H, NGV710807H, NGV715185H, NGV703676H, NGV709180H, NGV714738H, NGV001437H, NGV709432H, NGV717753H, NGV652306H, NGV705989H, NGV713751H, NGV714984H, NGV716328H, NGV702863H, NGV714060H, NGV714715H, NGV650451H, NGV700166H, NGV700176H, NGV701813H, NGV714912H, NGV720296H, NGV709246H, NGV710605H, NGV713377H, NGV713880H, NGV714808H, NGV718623H, NGV723060H, NGV000925H, NGV004470H, NGV650491H, NGV651583H, NGV651754H, NGV651780H, NGV652567H, NGV700165H, NGV700842H, NGV701093H, NGV702864H, NGV704143H, NGV707762H, NGV714552H, NGV714640H, NGV714818H, NGV715320H, NGV715686H, NGV716096H, NGV716135H, NGV716433H, NGV716807H, NGV717756H, NGV718081H, NGV718648H, NGV720177H, NGV721740H, NGV722229H, NGV722341H, NGV722528H, NGV001172H, NGV002358H, NGV002633H, NGV701302H, NGV713152H, NGV713436H, NGV650763H, NGV702889H, NGV710492H, NGV713092H, NGV713729H, NGV716726H, NGV002577H, NGV650558H, NGV650559H, NGV700150H, NGV705246H, NGV002673H, NGV002830H, NGV002856H, NGV003745H, NGV651616H, NGV652071H, NGV701384H, NGV702277H, NGV704044H, NGV710469H, NGV713095H, NGV714551H, NGV715217H, NGV717540H, NGV650432H, NGV722236H, NGV713374H, NGV713396H, NGV715666H, NGV004868H, NGV701686H, NGV702779H, NGV001470H, NGV705435H, NGV714810H, NGV001622H, NGV706513H, NGV713082H, NGV712103H, NGV713447H, NGV714851H, NGV715979H, NGV715987H, NGV716155H, NGV000534H, NGV002029H, NGV002030H, NGV703490H, NGV706078H, NGV706081H, NGV707751H, NGV707814H, NGV710623H, NGV713204H, NGV713495H, NGV714108H, NGV715989H, NGV722167H, NGV722604H, NGV001416H, NGV705101H, NGV705746H, NGV712408H, NGV715111H, NGV716800H, NGV718803H, NGV718827H, NGV707820H, NGV708384H, NGV708385H, NGV710645H, NGV710649H, NGV710752H, NGV710757H, NGV714893H, NGV715244H, NGV720210H, NGV720213H, NGV000810H, NGV713419H, NGV714859H, NGV005007H, NGV651446H, NGV651448H, NGV651932H, NGV651945H, NGV651968H, NGV703395H, NGV703480H, NGV703511H, NGV709287H, NGV710521H, NGV710591H, NGV712274H, NGV713070H, NGV713364H, NGV713740H, NGV716802H, NGV713072H, NGV001623H, NGV702900H, NGV714747H, NGV714853H, NGV650514H, NGV652302H, NGV715020H, NGV716333H, NGV716472H, NGV722840H, NGV001150H, NGV003027H, NGV003028H, NGV003029H, NGV005017H, NGV651023H, NGV701740H, NGV709179H, NGV709181H, NGV710876H, NGV713478H, NGV713479H, NGV713484H, NGV714563H, NGV721641H, NGV722231H, NGV003321H, NGV650891H, NGV652049H, NGV701739H, NGV703283H, NGV703723H, NGV705059H, NGV715225H, NGV716126H, NGV719267H, NGV002470H, NGV700284H, NGV700628H, NGV706514H, NGV713953H, NGV714524H, NGV714696H, NGV720223H, NGV703709H, NGV709174H, NGV713207H, NGV705444H, NGV705992H, NGV706157H, NGV713264H, NGV713273H, NGV713274H, NGV722303H, NGV722305H, NGV722324H, NGV707718H, NGV710746H, NGV715199H, NGV719202H, NGV714941H, NGV002637H, NGV711763H, NGV719194H, NGV001391H, NGV004485H, NGV703885H, NGV001444H, NGV650366H, NGV700507H, NGV701487H, NGV704840H, NGV706071H, NGV710457H, NGV710599H, NGV710681H, NGV713926H, NGV714726H, NGV715223H, NGV715373H, NGV706512H, NGV710625H, NGV708268H, NGV003035H, NGV003840H, NGV650101H, NGV713887H, NGV714461H, NGV722270H, NGV650902H, NGV704181H, NGV705743H, NGV714699H, NGV716335H, NGV001168H, NGV002245H, NGV002249H, NGV002839H, NGV003524H, NGV003552H, NGV700261H, NGV700266H, NGV700268H, NGV707732H, NGV707734H, NGV707757H, NGV710880H, NGV713190H, NGV713192H, NGV713209H, NGV713219H, NGV714777H, NGV714961H, NGV714962H, NGV715235H, NGV715352H, NGV716799H, NGV717686H, NGV720787H, NGV720789H, NGV000778H, NGV000788H, NGV002829H, NGV707912H, NGV710678H, NGV710751H, NGV715068H, NGV720479H, NGV000765H, NGV709416H, NGV002665H, NGV650669H, NGV712121H, NGV713459H, NGV714115H, NGV715002H, NGV720166H, NGV722264H, NGV000935H, NGV001609H, NGV002334H, NGV650282H, NGV700954H, NGV701678H, NGV706085H, NGV706363H, NGV707753H, NGV710809H, NGV714990H, NGV714991H, NGV720215H, NGV003810H, NGV715951H, NGV716483H, NGV722349H, NGV651096H, NGV713457H, NGV719329H, NGV703527H, NGV715341H, NGV722240H, NGV002425H, NGV003580H, NGV003600H, NGV003601H, NGV003603H, NGV710472H, NGV710476H, NGV710604H, NGV710838H, NGV712090H, NGV712095H, NGV713904H, NGV714675H, NGV720316H, NGV722173H, NGV722234H, NGV002660H, NGV002661H, NGV700225H, NGV700706H, NGV705026H, NGV710601H, NGV713721H, NGV714766H, NGV714998H, NGV715353H, NGV001614H, NGV650013H, NGV700311H, NGV705311H, NGV714799H, NGV003500H, NGV652012H, NGV700160H, NGV702789H, NGV702829H, NGV705314H, NGV710582H, NGV713173H, NGV713564H, NGV713717H, NGV714112H, NGV714768H, NGV715330H, NGV715333H, NGV721537H, NGV705386H, NGV707839H, NGV710811H, NGV713180H, NGV714685H, NGV702884H, NGV705986H, NGV707768H, NGV709127H, NGV722805H, NGV651996H, NGV701695H, NGV703730H, NGV710850H, NGV714774H, NGV719053H, NGV722703H, NGV710700H, NGV720420H, NGV002866H, NGV002867H, NGV002868H, NGV002869H, NGV650061H, NGV650093H, NGV651634H, NGV651988H, NGV708280H, NGV708281H, NGV708282H, NGV708284H, NGV708285H, NGV722606H, NGV000811H, NGV702887H, NGV721536H, NGV701809H, NGV650131H, NGV650898H, NGV651952H, NGV651972H, NGV651978H, NGV651981H, NGV651992H, NGV651993H, NGV702249H, NGV702250H, NGV702266H, NGV702375H, NGV703712H, NGV703716H, NGV703717H, NGV703720H, NGV703721H, NGV709468H, NGV709540H, NGV709572H, NGV709606H, NGV709626H, NGV715228H, NGV715230H, NGV715232H, NGV715237H, NGV715239H, NGV718226H, NGV718229H, NGV718231H, NGV718233H, NGV718236H, NGV718239H, NGV002636H, NGV652013H, NGV703559H, NGV713373H, NGV001378H, NGV001511H, NGV003555H, NGV004600H, NGV706531H, NGV716187H, NGV705069H, NGV712442H, NGV000348H, NGV000351H, NGV001643H, NGV002887H, NGV002889H, NGV650888H, NGV701942H, NGV702004H, NGV705042H, NGV705377H, NGV705381H, NGV705382H, NGV705733H, NGV713035H, NGV713108H, NGV714113H, NGV714114H, NGV720822H, NGV721670H, NGV721672H, NGV652451H, NGV651779H, NGV651957H, NGV702247H, NGV703487H, NGV706517H, NGV002513H, NGV701948H, NGV000318H, NGV652022H, NGV709195H, NGV701939H, NGV703531H, NGV706050H, NGV706550H, NGV002320H, NGV002663H, NGV708415H, NGV722629H, NGV714111H, NGV002250H, NGV003103H, NGV704028H, NGV704032H, NGV704801H, NGV714909H, NGV001304H, NGV002842H, NGV702455H, NGV702980H, NGV712122H, NGV713799H, NGV714544H, NGV716159H, NGV000796H, NGV002343H, NGV002347H, NGV002555H, NGV002613H, NGV002618H, NGV002622H, NGV002624H, NGV002625H, NGV714549H, NGV714643H, NGV714702H, NGV722158H, NGV722162H, NGV003616H, NGV005084H, NGV650224H, NGV650409H, NGV652212H, NGV652264H, NGV652314H, NGV700228H, NGV700246H, NGV704834H, NGV705007H, NGV705983H, NGV707934H, NGV708272H, NGV709711H, NGV710494H, NGV710496H, NGV711255H, NGV711401H, NGV713121H, NGV713417H, NGV713418H, NGV714225H, NGV714226H, NGV714234H, NGV714826H, NGV714836H, NGV714840H, NGV715124H, NGV715130H, NGV715131H, NGV715135H, NGV715355H, NGV715358H, NGV717433H, NGV722622H, NGV722626H, NGV004673H, NGV703097H, NGV703733H, NGV704169H, NGV705040H, NGV706135H, NGV706137H, NGV706511H, NGV723729H, NGV705073H, NGV710470H, NGV004265H, NGV652598H, NGV713203H, NGV722272H, NGV004666H, NGV701679H, NGV701684H, NGV706022H, NGV707739H, NGV709394H, NGV711377H, NGV714996H, NGV718108H, NGV723348H, NGV000355H, NGV002664H, NGV003790H, NGV701840H, NGV703758H, NGV703782H, NGV704791H, NGV706131H, NGV710505H, NGV710704H, NGV711901H, NGV712996H, NGV713367H, NGV715041H, NGV723080H, NGV650490H, NGV651689H, NGV701126H, NGV716492H, NGV720287H, NGV723090H, NGV714969H, NGV003068H, NGV003072H, NGV003518H, NGV650007H, NGV703790H, NGV708354H, NGV713233H, NGV713235H, NGV713379H, NGV713402H, NGV713875H, NGV715576H, NGV719410H, NGV700300H, NGV709173H, NGV711644H, NGV712464H, NGV712467H, NGV716404H, NGV000735H, NGV721074H, NGV650707H, NGV651785H, NGV652019H, NGV652482H, NGV703486H, NGV703701H, NGV708291H, NGV708340H, NGV709135H, NGV714043H, NGV714044H, NGV714735H, NGV720294H, NGV707758H, NGV707761H, NGV708412H, NGV002221H, NGV002335H, NGV002837H, NGV708437H, NGV712140H, NGV714829H, NGV714830H, NGV716482H, NGV719283H, NGV702645H, NGV703995H, NGV703998H, NGV708070H, NGV709928H, NGV712076H, NGV713112H, NGV715040H, NGV716500H, NGV703526H, NGV650074H, NGV708050H, NGV710513H, NGV713445H, NGV000943H, NGV002255H, NGV002293H, NGV002774H, NGV650362H, NGV650510H, NGV650899H, NGV651707H, NGV652003H, NGV702871H, NGV706145H, NGV706504H, NGV706508H, NGV710477H, NGV712994H, NGV713736H, NGV714727H, NGV715218H, NGV715385H, NGV718082H, NGV719048H, NGV719049H, NGV720156H, NGV000790H, NGV001648H, NGV001650H, NGV001652H, NGV002769H, NGV003344H, NGV709916H, NGV709922H, NGV712101H, NGV714686H, NGV714691H, NGV716431H, NGV718046H, NGV720148H, NGV722287H, NGV003502H, NGV650297H, NGV650481H, NGV701379H, NGV723733H, NGV700933H, NGV701853H, NGV003620H, NGV651658H, NGV701478H, NGV703996H, NGV711664H, NGV714863H, NGV718099H, NGV721649H, NGV001439H, NGV702252H, NGV702302H, NGV702472H, NGV706533H, NGV703849H, NGV004344H, NGV004459H, NGV650553H, NGV652007H, NGV701301H, NGV703863H, NGV706563H, NGV706790H, NGV710736H, NGV711375H, NGV711389H, NGV715079H, NGV715574H, NGV715575H, NGV720678H, NGV722490H, NGV650482H, NGV651900H, NGV652009H, NGV710713H, NGV713562H, NGV713763H, NGV702272H, NGV702549H, NGV708126H, NGV710452H, NGV710879H, NGV716306H, NGV717542H, NGV712078H, NGV714744H, NGV651697H, NGV651702H, NGV651704H, NGV704610H, NGV707716H, NGV710799H, NGV710851H, NGV710886H, NGV712468H, NGV713015H, NGV713441H, NGV714706H, NGV715019H, NGV720047H, NGV720238H, NGV723324H, NGV708130H, NGV715350H, NGV715363H, NGV712277H, NGV651815H, NGV700960H, NGV703538H, NGV708129H, NGV716396H, NGV718101H, NGV722780H, NGV722836H, NGV000315H, NGV004638H, NGV708264H, NGV708431H, NGV712352H, NGV714895H, NGV715879H, NGV720185H, NGV720191H, NGV721742H, NGV003316H, NGV709185H, NGV700509H, NGV700987H, NGV711378H, NGV721712H, NGV722776H, NGV723402H, NGV703466H, NGV709930H, NGV710829H, NGV710830H, NGV651587H, NGV703877H, NGV710487H, NGV705050H, NGV709380H, NGV703785H, NGV709396H, NGV714494H, NGV714817H, NGV714822H, NGV723079H, NGV723082H, NGV723085H, NGV701502H, NGV718252H, NGV718308H, NGV002291H, NGV712435H, NGV717778H, NGV003098H, NGV003836H, NGV705039H, NGV706756H, NGV706782H, NGV706784H, NGV712107H, NGV712108H, NGV716793H, NGV720180H, NGV720183H, NGV004610H, NGV000812H, NGV004479H, NGV005085H, NGV702007H, NGV708365H, NGV713353H, NGV716394H, NGV718634H, NGV718637H, NGV704830H, NGV716407H, NGV000293H, NGV000308H, NGV001624H, NGV002076H, NGV002090H, NGV002101H, NGV002253H, NGV003375H, NGV703337H, NGV703382H, NGV706056H, NGV706564H, NGV710840H, NGV711778H, NGV711779H, NGV712232H, NGV713218H, NGV714785H, NGV714892H, NGV714955H, NGV715216H, NGV716473H, NGV716487H, NGV716491H, NGV723396H, NGV003410H, NGV650487H, NGV650489H, NGV712265H, NGV712266H, NGV718429H, NGV003384H, NGV712226H, NGV712979H, NGV003034H, NGV707736H, NGV708366H, NGV713011H, NGV718078H, NGV702949H, NGV704130H, NGV705374H, NGV713743H, NGV713848H, NGV714805H, NGV720819H, NGV721495H, NGV722214H, NGV702962H, NGV005105H, NGV005106H, NGV005113H, NGV650710H, NGV705063H, NGV706569H, NGV708029H, NGV708030H, NGV710870H, NGV710871H, NGV710989H, NGV716112H, NGV716372H, NGV716675H, NGV716678H, NGV002727H, NGV710768H, NGV713864H, NGV713927H, NGV714533H, NGV716387H, NGV002841H, NGV652176H, NGV705102H, NGV708056H, NGV709377H, NGV710738H, NGV713394H, NGV714032H, NGV651751H, NGV702973H, NGV718224H, NGV706487H, NGV650501H, NGV651016H, NGV709176H, NGV710565H, NGV710642H, NGV710644H, NGV714351H, NGV714944H, NGV708367H, NGV712306H, NGV001642H, NGV002318H, NGV002731H, NGV703504H, NGV707782H, NGV713568H, NGV713569H, NGV713570H, NGV713645H, NGV720288H, NGV002778H, NGV717009H, NGV703993H, NGV709365H, NGV710741H, NGV714539H, NGV001540H, NGV003812H, NGV708125H, NGV713744H, NGV713901H, NGV717760H, NGV719618H, NGV713768H, NGV716471H, NGV723362H, NGV701112H, NGV701490H, NGV701499H, NGV707821H, NGV710581H, NGV711783H, NGV711784H, NGV722850H, NGV652026H, NGV710740H, NGV711753H, NGV713831H, NGV716311H, NGV716818H, NGV715067H, NGV719289H, NGV723380H, NGV001789H, NGV651958H, NGV652352H, NGV652427H, NGV703529H, NGV707698H, NGV708145H, NGV714480H, NGV720295H, NGV722611H, NGV703906H, NGV710608H, NGV711381H, NGV714229H, NGV703534H, NGV710805H, NGV714512H, NGV714521H, NGV708361H, NGV718054H, NGV709419H, NGV710508H, NGV704186H, NGV704187H, NGV713004H, NGV713205H, NGV713488H, NGV714681H, NGV003052H, NGV709356H, NGV715577H, NGV000536H, NGV001385H, NGV001386H, NGV001388H, NGV002256H, NGV002257H, NGV002264H, NGV002768H, NGV002776H, NGV002777H, NGV003899H, NGV003926H, NGV003936H, NGV004266H, NGV004893H, NGV004900H, NGV651906H, NGV651925H, NGV700838H, NGV700841H, NGV701767H, NGV701833H, NGV701837H, NGV702843H, NGV702857H, NGV702865H, NGV703913H, NGV703986H, NGV703989H, NGV706132H, NGV706141H, NGV706152H, NGV707750H, NGV707759H, NGV707901H, NGV707914H, NGV710486H, NGV710659H, NGV710661H, NGV711755H, NGV711757H, NGV711759H, NGV712932H, NGV712933H, NGV713075H, NGV713078H, NGV713770H, NGV713775H, NGV713776H, NGV715012H, NGV715016H, NGV715148H, NGV715168H, NGV716172H, NGV716174H, NGV716175H, NGV717445H, NGV718193H, NGV718212H, NGV721512H, NGV721516H, NGV721517H, NGV721519H, NGV721531H, NGV711771H, NGV650059H, NGV723375H, NGV722262H, NGV003385H, NGV701374H, NGV702368H, NGV722301H, NGV709364H, NGV001632H, NGV713431H, NGV709706H, NGV712451H, NGV651995H, NGV709449H, NGV001182H, NGV001192H, NGV003314H, NGV710694H, NGV713174H, NGV723161H, NGV003060H, NGV707922H, NGV708343H, NGV709147H, NGV710428H, NGV714795H, NGV701159H, NGV713195H, NGV713201H, NGV723211H, NGV003391H, NGV706012H, NGV713088H, NGV000013H, NGV003357H, NGV003405H, NGV706496H, NGV710847H, NGV722267H, NGV723191H, NGV001159H, NGV706148H, NGV713464H, NGV723076H, NGV720775H, NGV713723H, NGV003529H, NGV003531H, NGV705290H, NGV709657H, NGV710512H, NGV705297H, NGV002501H, NGV650686H, NGV704786H, NGV704833H, NGV706658H, NGV709443H, NGV722219H, NGV707771H, NGV708338H, NGV710831H, NGV002386H, NGV002401H, NGV652045H, NGV652205H, NGV710837H, NGV717781H, NGV715167H, NGV723086H, NGV709363H, NGV722257H, NGV004073H, NGV700077H, NGV704179H, NGV706025H, NGV706030H, NGV706036H, NGV708047H, NGV709161H, NGV709166H, NGV710532H, NGV713403H, NGV718075H, NGV721659H, NGV722302H, NGV722484H, NGV722617H, NGV722618H, NGV001010H, NGV002348H, NGV003009H, NGV700965H, NGV706154H, NGV706486H, NGV707900H, NGV710982H, NGV712289H, NGV715348H, NGV716007H, NGV721563H, NGV000005H, NGV000736H, NGV651757H, NGV651895H, NGV651922H, NGV652174H, NGV706004H, NGV710502H, NGV710606H, NGV712153H, NGV714228H, NGV715006H, NGV716259H, NGV718695H, NGV719159H, NGV720705H, NGV720806H, NGV723367H, NGV720043H, NGV721556H, NGV713551H, NGV003279H, NGV003943H, NGV701854H, NGV706051H, NGV709387H, NGV710821H, NGV713216H, NGV713423H, NGV713741H, NGV714492H, NGV719375H, NGV720227H, NGV722835H, NGV709178H, NGV705296H, NGV002754H, NGV713462H, NGV714378H, NGV709164H, NGV709165H, NGV709172H, NGV701383H, NGV701125H, NGV701137H, NGV701410H, NGV703767H, NGV703770H, NGV703771H, NGV703772H, NGV703773H, NGV703775H, NGV703776H, NGV703778H, NGV703780H, NGV703783H, NGV002206H, NGV004859H, NGV652016H, NGV701004H, NGV706557H, NGV706559H, NGV710621H, NGV714728H, NGV714975H, NGV715565H, NGV718213H, NGV722370H, NGV716437H, NGV705006H, NGV708411H, NGV003791H, NGV703762H, NGV703784H, NGV703864H, NGV707691H, NGV711642H, NGV711643H, NGV712883H, NGV713425H, NGV713787H, NGV714868H, NGV714874H, NGV719168H, NGV719296H, NGV722815H, NGV723392H, NGV723404H, NGV701079H, NGV702458H, NGV705100H, NGV707749H, NGV709915H, NGV711760H, NGV712246H, NGV714232H, NGV716178H, NGV716185H, NGV002608H, NGV002620H, NGV711388H, NGV702242H, NGV709373H, NGV714379H, NGV650930H, NGV652371H, NGV652417H, NGV000934H, NGV706786H, NGV708382H, NGV710861H, NGV710866H, NGV710869H, NGV714743H, NGV715094H, NGV715099H, NGV715100H, NGV721715H, NGV723092H, NGV723100H, NGV714103H, NGV001413H, NGV004734H, NGV700514H, NGV001453H, NGV001454H, NGV005075H, NGV650732H, NGV652613H, NGV701272H, NGV706052H, NGV706055H, NGV712452H, NGV712456H, NGV712461H, NGV713254H, NGV713261H, NGV714804H, NGV715028H, NGV715121H, NGV715122H, NGV720119H, NGV720214H, NGV720216H, NGV700963H, NGV701931H, NGV714781H, NGV715164H, NGV716436H, NGV651944H, NGV721648H, NGV703761H, NGV708368H, NGV708380H, NGV709375H, NGV709454H, NGV709712H, NGV710672H, NGV714760H, NGV716373H, NGV716376H, NGV716385H, NGV716389H, NGV718111H, NGV718240H, NGV722705H, NGV722713H, NGV713362H, NGV714791H, NGV003508H, NGV001371H, NGV001447H, NGV716405H, NGV001372H, NGV700147H, NGV000927H, NGV702905H, NGV707754H, NGV710598H, NGV713183H, NGV713393H, NGV715656H, NGV719439H, NGV720024H, NGV714731H, NGV708358H, NGV702891H, NGV652018H, NGV001645H, NGV700504H, NGV702967H, NGV703326H, NGV707819H, NGV718061H, NGV705068H, NGV714832H, NGV003346H, NGV003787H, NGV701696H, NGV702246H, NGV703750H, NGV704166H, NGV704618H, NGV705070H, NGV709190H, NGV710423H, NGV713100H, NGV713429H, NGV714382H, NGV721453H, NGV722336H, NGV723374H, NGV651762H, NGV708277H, NGV714775H, NGV714233H, NGV708414H, NGV714959H, NGV715009H, NGV715573H, NGV722527H, NGV000730H, NGV002583H, NGV003625H, NGV004592H, NGV702956H, NGV705229H, NGV705757H, NGV706581H, NGV710554H, NGV710839H, NGV712992H, NGV713096H, NGV713923H, NGV713938H, NGV714365H, NGV715007H, NGV002783H, NGV709194H, NGV723104H, NGV723105H, NGV000012H, NGV001363H, NGV001395H, NGV001441H, NGV002269H, NGV002429H, NGV003326H, NGV651971H, NGV652467H, NGV652511H, NGV700084H, NGV700287H, NGV703336H, NGV705024H, NGV706672H, NGV708139H, NGV708379H, NGV708405H, NGV710551H, NGV710588H, NGV710589H, NGV710666H, NGV710790H, NGV710797H, NGV710819H, NGV712256H, NGV713171H, NGV713378H, NGV713392H, NGV713410H, NGV714047H, NGV714054H, NGV714971H, NGV714994H, NGV715096H, NGV715351H, NGV718417H, NGV719413H, NGV719626H, NGV720161H, NGV722340H, NGV722422H, NGV723371H, NGV002296H, NGV003036H, NGV713867H, NGV714828H, NGV722821H, NGV723379H, NGV004544H, NGV004919H, NGV652539H, NGV702469H, NGV708350H, NGV713010H, NGV713734H, NGV713735H, NGV723106H, NGV703711H, NGV000278H, NGV004481H, NGV651787H, NGV651974H, NGV703719H, NGV710414H, NGV710696H, NGV710697H, NGV710717H, NGV713489H, NGV713494H, NGV713722H, NGV713730H, NGV713739H, NGV714028H, NGV714030H, NGV714063H, NGV714497H, NGV714508H, NGV718160H, NGV718167H, NGV002035H, NGV003330H, NGV004608H, NGV701289H, NGV701290H, NGV701819H, NGV703514H, NGV710641H, NGV711775H, NGV713463H, NGV713919H, NGV714523H, NGV714855H, NGV715011H, NGV718083H, NGV701814H, NGV711751H, NGV714921H, NGV723415H, NGV000193H, NGV001432H, NGV003008H, NGV003770H, NGV003786H, NGV004565H, NGV005082H, NGV708356H, NGV708360H, NGV709715H, NGV710592H, NGV715325H, NGV716467H, NGV718619H, NGV719478H, NGV700367H, NGV700371H, NGV003557H, NGV651748H, NGV700145H, NGV702788H, NGV712889H, NGV713865H, NGV713872H, NGV714406H, NGV714443H, NGV723115H, NGV713486H, NGV722131H, NGV652025H, NGV714794H, NGV700149H, NGV704193H, NGV710522H, NGV713202H, NGV714894H, NGV721646H, NGV712144H, NGV000746H, NGV700073H, NGV703852H, NGV708146H, NGV714517H, NGV721062H, NGV723383H, NGV721082H, NGV004740H; GTIN 00763000122768, Serial Numbers: NGV004483H, NGV004538H, NGV004672H, NGV710664H, NGV710841H, NGV719230H, NGV722315H, NGV722599H, NGV710646H, NGV715381H, NGV714527H, NGV000923H, NGV001975H, NGV003273H, NGV004461H, NGV004664H, NGV714530H, NGV714648H, NGV714911H, NGV715132H, NGV718906H, NGV720057H, NGV723428H, NGV004633H, NGV004655H, NGV710810H, NGV720211H, NGV716653H, NGV723420H, NGV714662H, NGV718900H, NGV720706H, NGV722127H, NGV722132H, NGV722148H, NGV722153H, NGV003355H, NGV004543H, NGV004555H, NGV710824H, NGV711354H, NGV713169H, NGV713554H, NGV715276H, NGV716150H, NGV720058H, NGV720242H, NGV720268H, NGV000604H, NGV000614H, NGV000803H, NGV000833H, NGV000900H, NGV004464H, NGV004471H, NGV004531H, NGV004542H, NGV004554H, NGV004603H, NGV004660H, NGV710826H, NGV710873H, NGV713138H, NGV713175H, NGV714373H, NGV714522H, NGV714546H, NGV715445H, NGV715446H, NGV716134H, NGV716725H, NGV716809H, NGV717748H, NGV720235H, NGV720310H, NGV720786H, NGV719086H, NGV720272H, NGV000784H, NGV004532H, NGV722154H, NGV716661H, NGV004540H, NGV710882H, NGV715430H, NGV004550H, NGV004656H, NGV000638H, NGV650228H, NGV703474H, NGV708263H, NGV720841H, NGV721084H, NGV005168H, NGV703393H, NGV703394H, NGV708148H, NGV713939H, NGV714796H, NGV714838H, NGV715340H, NGV716137H, NGV718043H, NGV718047H, NGV722245H, NGV722247H, NGV722265H, NGV004661H, NGV650215H, NGV702963H, NGV712170H, NGV716130H, NGV717750H, NGV707923H, NGV005013H, NGV715379H, NGV705766H, NGV719454H, NGV701612H, NGV704013H, NGV707929H, NGV708042H, NGV708348H, NGV712176H, NGV713473H, NGV713773H, NGV713924H, NGV714041H, NGV714862H, NGV715327H, NGV717729H, NGV722268H, NGV650235H, NGV650263H, NGV700067H, NGV701818H, NGV703724H, NGV705770H, NGV707836H, NGV708341H, NGV713158H, NGV713456H, NGV713466H, NGV713471H, NGV713485H, NGV713881H, NGV715156H, NGV717937H, NGV722139H, NGV002787H, NGV650250H, NGV716797H, NGV717457H, NGV720445H, NGV700690H, NGV701811H, NGV702014H, NGV704158H, NGV705764H, NGV713934H, NGV714045H, NGV716395H, NGV650977H, NGV705782H, NGV708018H, NGV714742H, NGV715097H, NGV715157H, NGV717749H, NGV719406H, NGV719408H, NGV722104H, NGV722150H, NGV722227H, NGV650960H, NGV004669H, NGV701694H, NGV701763H, NGV707920H, NGV714967H, NGV714970H, NGV721502H, NGV721506H, NGV703498H, NGV002069H, NGV002888H, NGV704150H, NGV705737H, NGV708332H, NGV716138H, NGV718424H, NGV002222H, NGV000949H, NGV000970H, NGV703688H, NGV712135H, NGV712150H, NGV717752H, NGV707805H, NGV700054H, NGV701273H, NGV701480H, NGV703476H, NGV707799H, NGV712134H, NGV714689H, NGV003066H, NGV701261H, NGV707823H, NGV720112H, NGV703731H, NGV000606H, NGV000607H, NGV000632H, NGV000946H, NGV000948H, NGV000982H, NGV001373H, NGV002178H, NGV002213H, NGV002216H, NGV002861H, NGV003040H, NGV003045H, NGV003553H, NGV003554H, NGV004614H, NGV650018H, NGV650241H, NGV650949H, NGV651135H, NGV651137H, NGV651185H, NGV700079H, NGV700080H, NGV700082H, NGV703390H, NGV703505H, NGV703508H, NGV703518H, NGV703734H, NGV704132H, NGV704162H, NGV705784H, NGV706498H, NGV707796H, NGV707826H, NGV708020H, NGV708283H, NGV711656H, NGV712159H, NGV713125H, NGV713873H, NGV713882H, NGV714645H, NGV714748H, NGV714835H, NGV714837H, NGV714869H, NGV714973H, NGV714979H, NGV714993H, NGV715082H, NGV715090H, NGV715092H, NGV715103H, NGV715144H, NGV715214H, NGV715227H, NGV715367H, NGV715382H, NGV718080H, NGV718407H, NGV718409H, NGV720113H, NGV720220H, NGV722091H, NGV722102H, NGV722157H, NGV722170H, NGV722172H, NGV000619H, NGV002079H, NGV650226H, NGV700085H, NGV701609H, NGV702346H, NGV703475H, NGV703718H, NGV706505H, NGV708352H, NGV002982H, NGV707802H, NGV003039H, NGV000603H, NGV651188H, NGV701920H, NGV713166H, NGV000621H, NGV004612H, NGV712127H, NGV713067H, NGV713163H, NGV713178H, NGV714866H, NGV720825H, NGV701744H, NGV714048H, NGV715365H, NGV716375H, NGV004618H, NGV702350H, NGV003022H, NGV711647H, NGV713076H, NGV713930H, NGV004620H, NGV705064H, NGV705067H, NGV717960H, NGV719389H, NGV001542H, NGV002309H, NGV650256H, NGV701405H, NGV702966H, NGV705772H, NGV705779H, NGV712162H, NGV713492H, NGV714860H, NGV715151H, NGV715213H, NGV715318H, NGV715319H, NGV715331H, NGV715347H, NGV719392H, NGV000851H, NGV650972H, NGV703545H, NGV713157H, NGV000954H, NGV002832H, NGV002838H, NGV002851H, NGV005109H, NGV707931H, NGV715376H, NGV000944H, NGV001553H, NGV003020H, NGV003024H, NGV700064H, NGV701921H, NGV704161H, NGV707813H, NGV713154H, NGV714974H, NGV715154H, NGV715388H, NGV716125H, NGV716145H, NGV717758H, NGV719455H, NGV720770H, NGV651147H, NGV700701H, NGV000836H, NGV001370H, NGV720173H, NGV000613H, NGV703510H, NGV706028H, NGV708294H, NGV713187H, NGV713477H, NGV714980H, NGV715221H, NGV719459H, NGV701945H, NGV703523H, NGV704221H, NGV708038H, NGV714046H, NGV715149H, NGV716489H, NGV700056H, NGV701768H, NGV713182H, NGV714992H, NGV716785H, NGV721511H, NGV700036H, NGV701810H, NGV708290H, NGV713455H, NGV715356H, NGV000605H, NGV000617H, NGV000620H, NGV000839H, NGV001548H, NGV002188H, NGV002199H, NGV002217H, NGV003037H, NGV004606H, NGV004652H, NGV650245H, NGV651143H, NGV651145H, NGV652387H, NGV700078H, NGV700697H, NGV701947H, NGV703522H, NGV703540H, NGV703727H, NGV703798H, NGV704056H, NGV704152H, NGV704222H, NGV708008H, NGV708036H, NGV708063H, NGV708072H, NGV708076H, NGV708150H, NGV708342H, NGV712160H, NGV712166H, NGV712168H, NGV712302H, NGV712303H, NGV713878H, NGV714839H, NGV714843H, NGV714861H, NGV714997H, NGV715027H, NGV715198H, NGV715222H, NGV715240H, NGV715242H, NGV715329H, NGV715345H, NGV715349H, NGV715359H, NGV715377H, NGV716162H, NGV716486H, NGV716680H, NGV717725H, NGV717945H, NGV718096H, NGV718428H, NGV720059H, NGV723121H, NGV002891H, NGV715369H, NGV002989H, NGV003015H, NGV650009H, NGV650014H, NGV700061H, NGV708334H, NGV712131H, NGV713073H, NGV713774H, NGV714650H, NGV712163H, NGV651139H, NGV703548H, NGV703550H, NGV700076H, NGV703472H, NGV000618H, NGV000924H, NGV001541H, NGV002082H, NGV705741H, NGV710890H, NGV713405H, NGV714031H, NGV714825H, NGV714864H, NGV718079H, NGV715224H, NGV000631H, NGV002197H, NGV005070H, NGV650258H, NGV703396H, NGV703512H, NGV707817H, NGV707834H, NGV708013H, NGV713416H, NGV713928H, NGV714831H, NGV715161H, NGV715212H, NGV715387H, NGV716157H, NGV716686H, NGV717764H, NGV717768H, NGV717958H, NGV718437H, NGV723122H, NGV002198H, NGV002310H, NGV703542H, NGV720129H, NGV650015H, NGV650028H, NGV701831H, NGV704140H, NGV707728H, NGV707729H, NGV701816H, NGV704129H, NGV713421H, NGV002227H, NGV707780H, NGV004613H, NGV005077H, NGV651228H, NGV707801H, NGV714867H, NGV715066H, NGV715332H, NGV650257H, NGV652367H, NGV703503H, NGV703497H, NGV704154H, NGV708267H, NGV713487H, NGV717460H, NGV717767H, NGV718065H, NGV722271H, NGV650063H, NGV650236H, NGV651243H, NGV707800H, NGV712172H, NGV715336H, NGV002786H, NGV003031H, NGV650979H, NGV713193H, NGV720311H, NGV722281H, NGV002788H, NGV707798H, NGV708266H, NGV713184H, NGV713757H, NGV705083H, NGV712136H, NGV714798H, NGV714811H, NGV714819H, NGV714841H, NGV718021H, NGV718244H, NGV714870H, NGV715342H, NGV704708H, NGV710536H, NGV717449H, NGV717548H, NGV722364H, NGV705390H, NGV705438H, NGV714750H, NGV722221H, NGV703804H, NGV702805H, NGV702815H, NGV702819H, NGV703555H, NGV703704H, NGV705982H, NGV712244H, NGV716380H, NGV716384H, NGV716449H, NGV718251H, NGV718254H, NGV719431H, NGV722791H, NGV722804H, NGV716465H, NGV722176H, NGV713131H, NGV705456H, NGV701035H, NGV702817H, NGV703777H, NGV703703H, NGV702523H, NGV703678H, NGV716248H, NGV704024H, NGV703485H, NGV704145H, NGV722235H, NGV722238H, NGV716250H, NGV722205H, NGV703793H, NGV701027H, NGV702017H, NGV702240H, NGV703801H, NGV710499H, NGV713139H, NGV722147H, NGV722155H, NGV722156H, NGV722255H, NGV723088H, NGV704037H, NGV702516H, NGV702519H, NGV702524H, NGV703856H, NGV703862H, NGV703874H, NGV706144H, NGV706151H, NGV706495H, NGV706580H, NGV706582H, NGV702837H, NGV704226H, NGV702245H, NGV702526H, NGV702809H, NGV703843H, NGV708024H, NGV713098H, NGV713101H, NGV713122H, NGV714903H, NGV715458H, NGV716488H, NGV712956H, NGV702527H, NGV702806H, NGV715456H, NGV716252H, NGV724532H, NGV724539H, NGV701986H, NGV701998H, NGV702253H, NGV702304H, NGV702808H, NGV702810H, NGV702824H, NGV702835H, NGV702902H, NGV703695H, NGV704251H, NGV706488H, NGV708023H, NGV710484H, NGV710495H, NGV713127H, NGV713129H, NGV713143H, NGV713144H, NGV713162H, NGV714906H, NGV715449H, NGV715453H, NGV715455H, NGV715658H, NGV716245H, NGV716247H, NGV716386H, NGV716397H, NGV716399H, NGV716415H, NGV716484H, NGV717450H, NGV717541H, NGV718105H, NGV718249H, NGV719421H, NGV719440H, NGV722149H, NGV722179H, NGV722225H, NGV722246H, NGV722437H, NGV723091H, NGV723114H, NGV706759H, NGV701479H, NGV703502H, NGV703702H, NGV712967H, NGV713003H, NGV713061H, NGV703794H, NGV715610H, NGV722143H, NGV702547H, NGV706588H, NGV703891H, NGV703896H, NGV710586H, NGV717555H, NGV710571H, NGV710569H, NGV717556H, NGV717551H, NGV708027H, NGV709699H, NGV717557H, NGV722256H, NGV722342H, NGV722483H, NGV722217H, NGV718153H, NGV717451H, NGV722325H, NGV708039H, NGV717582H, NGV717733H, NGV702015H, NGV716663H, NGV717719H, NGV708438H, NGV718279H, NGV710567H, NGV717560H, NGV723075H, NGV717570H, NGV722260H, NGV718098H, NGV712961H, NGV708026H, NGV718159H, NGV722339H, NGV702838H, NGV718154H, NGV702012H, NGV702261H, NGV727061H, NGV727143H, NGV718165H, NGV723570H, NGV701162H, NGV701164H, NGV702836H, NGV702878H, NGV706867H, NGV710823H, NGV714709H, NGV719388H, NGV724336H, NGV729604H, NGV729606H, NGV729957H, NGV002662H, NGV713083H, NGV713087H, NGV713217H, NGV724523H, NGV002282H, NGV002338H, NGV005080H, NGV701482H, NGV703747H, NGV703759H, NGV705732H, NGV705734H, NGV705739H, NGV705744H, NGV706355H, NGV706547H, NGV711762H, NGV711765H, NGV715234H, NGV716426H, NGV718282H, NGV720779H, NGV721063H, NGV704043H, NGV704050H, NGV002715H, NGV002720H, NGV002762H, NGV003744H, NGV003746H, NGV003761H, NGV003788H, NGV003789H, NGV005083H, NGV707673H, NGV707726H, NGV707745H, NGV712178H, NGV712180H, NGV712238H, NGV712248H, NGV712914H, NGV712915H, NGV712924H, NGV712995H, NGV712998H, NGV713000H, NGV713002H, NGV714916H, NGV715360H, NGV729976H, NGV001600H, NGV002277H, NGV002377H, NGV002655H, NGV003808H, NGV005143H, NGV005227H, NGV701107H, NGV701175H, NGV703669H, NGV703765H, NGV703766H, NGV707776H, NGV708127H, NGV712091H, NGV712137H, NGV713228H, NGV715112H, NGV715137H, NGV715138H, NGV720239H, NGV720243H, NGV722184H, NGV722465H, NGV724558H, NGV729959H, NGV729962H, NGV001483H, NGV005137H, NGV705726H, NGV707717H, NGV713224H, NGV713230H, NGV713802H, NGV714690H, NGV714806H, NGV714940H, NGV715125H, NGV717713H, NGV718281H, NGV720231H, NGV721732H, NGV723591H, NGV727055H, NGV729951H, NGV000622H, NGV001294H, NGV002649H, NGV002651H, NGV003801H, NGV003806H, NGV003814H, NGV004909H, NGV005119H, NGV005124H, NGV700007H, NGV700009H, NGV700013H, NGV701346H, NGV701484H, NGV701498H, NGV701501H, NGV702828H, NGV703500H, NGV703693H, NGV703739H, NGV703745H, NGV704232H, NGV706155H, NGV706353H, NGV707710H, NGV707713H, NGV707777H, NGV708012H, NGV708364H, NGV708370H, NGV709151H, NGV709470H, NGV710828H, NGV710875H, NGV710881H, NGV712147H, NGV713062H, NGV713215H, NGV714660H, NGV714664H, NGV714937H, NGV715153H, NGV715229H, NGV715233H, NGV716196H, NGV716236H, NGV716365H, NGV716420H, NGV716422H, NGV716481H, NGV716798H, NGV719468H, NGV719474H, NGV720225H, NGV722304H, NGV724333H, NGV726922H, NGV727015H, NGV729860H, NGV002340H, NGV002350H, NGV701166H, NGV702893H, NGV702896H, NGV702897H, NGV702899H, NGV707789H, NGV707793H, NGV707795H, NGV712149H, NGV713231H, NGV713496H, NGV713497H, NGV713498H, NGV713680H, NGV713766H, NGV713777H.NGV713778H, NGV713788H, NGV720155H, NGV722485H, NGV722486H, NGV723465H, NGV723472H, NGV723630H, NGV724531H, NGV723094H, NGV724471H, NGV000645H, NGV000647H, NGV000685H, NGV000686H, NGV000754H, NGV000759H, NGV000762H, NGV001208H, NGV005126H, NGV708118H, NGV708122H, NGV708123H, NGV708124H, NGV710804H, NGV710822H, NGV711752H, NGV713759H, NGV713863H, NGV715110H, NGV715231H, NGV715372H, NGV716184H, NGV716439H, NGV716475H, NGV716810H, NGV717736H, NGV718408H, NGV718414H, NGV720157H, NGV721733H, NGV722242H, NGV723098H, NGV723099H, NGV724331H, NGV724515H, NGV724536H, NGV726792H, NGV729946H, NGV729973H, NGV730033H, NGV716466H, NGV719378H, NGV719382H, NGV720142H, NGV729972H, NGV701182H, NGV702364H, NGV703897H, NGV703904H, NGV710877H, NGV714717H, NGV715139H, NGV718405H, NGV724254H, NGV729944H, NGV000641H, NGV001165H, NGV001288H, NGV002370H, NGV002659H, NGV005087H, NGV005098H, NGV005102H, NGV005115H, NGV005117H, NGV703768H, NGV704005H, NGV704133H, NGV705730H, NGV706869H, NGV706874H, NGV706876H, NGV706877H, NGV706881H, NGV708377H, NGV708428H, NGV711756H, NGV711761H, NGV712093H, NGV714815H, NGV715142H, NGV715165H, NGV715236H, NGV716180H, NGV716454H, NGV717458H, NGV717461H, NGV717784H, NGV721743H, NGV723096H, NGV724552H, NGV727020H, NGV729578H, NGV729586H, NGV702365H, NGV707764H, NGV707766H, NGV712148H, NGV715173H, NGV716464H, NGV719588H, NGV721075H, NGV002714H, NGV712139H, NGV714658H, NGV707788H, NGV003869H, NGV003954H, NGV705467H, NGV707925H, NGV713080H, NGV714945H, NGV714956H, NGV716445H, NGV717739H, NGV717740H, NGV718435H, NGV723607H, NGV724530H, NGV724543H, NGV727053H, NGV727054H, NGV001484H, NGV002504H, NGV002666H, NGV005229H, NGV702372H, NGV702892H, NGV703988H, NGV704009H, NGV704040H, NGV705745H, NGV707752H, NGV709463H, NGV712146H, NGV713151H, NGV713793H, NGV713804H, NGV713805H, NGV715116H, NGV715179H, NGV715186H, NGV716368H, NGV717708H, NGV717716H, NGV717780H, NGV720170H, NGV721524H, NGV723625H, NGV723672H, NGV724496H, NGV724518H, NGV724541H, NGV727032H, NGV000625H, NGV005134H, NGV702879H, NGV707774H, NGV710803H, NGV713742H, NGV715141H, NGV716417H, NGV720204H, NGV723485H, NGV000771H, NGV703558H, NGV709440H, NGV709442H, NGV709444H, NGV709452H, NGV722480H, NGV002387H, NGV002494H, NGV709458H, NGV711758H, NGV720245H, NGV729869H, NGV000630H, NGV001487H, NGV707756H, NGV715444H, NGV000616H, NGV001169H, NGV001475H, NGV001476H, NGV700003H, NGV700005H, NGV703997H, NGV706562H, NGV708007H, NGV708014H, NGV708015H, NGV708429H, NGV708435H, NGV709129H, NGV714964H, NGV715158H, NGV720784H, NGV723634H, NGV723751H, NGV723759H, NGV723934H, NGV724464H, NGV724465H, NGV714661H, NGV000767H, NGV701085H, NGV703686H, NGV706555H, NGV707707H, NGV707778H, NGV707779H, NGV707783H, NGV707784H, NGV708395H, NGV713079H, NGV713153H, NGV713212H, NGV713792H, NGV713862H, NGV717846H, NGV000623H, NGV005144H, NGV005169H, NGV709459H, NGV721080H, NGV721499H, NGV721503H, NGV721521H, NGV729871H, NGV001492H, NGV002373H, NGV708375H, NGV708376H, NGV708378H, NGV718420H, NGV720206H, NGV729601H, NGV000624H, NGV701083H, NGV701489H, NGV713063H, NGV713077H, NGV720778H, NGV729593H, NGV707767H, NGV713737H, NGV723095H, NGV729948H, NGV702381H, NGV702459H, NGV707712H, NGV718402H, NGV718403H, NGV724335H, NGV724579H, NGV001164H, NGV001239H, NGV701157H, NGV703738H, NGV707705H, NGV707787H, NGV708135H, NGV708142H, NGV712138H, NGV713222H, NGV716478H, NGV717742H, NGV717743H, NGV718164H, NGV722306H, NGV722477H, NGV724289H, NGV724325H, NGV724485H, NGV724513H, NGV727992H, NGV729574H, NGV729960H, NGV729967H, NGV729975H, NGV700178H, NGV000298H, NGV652466H, NGV702273H, NGV709171H, NGV709192H, NGV710529H, NGV715077H, NGV716168H, NGV716327H, NGV004894H, NGV706522H, NGV706526H, NGV715073H, NGV719205H, NGV722278H, NGV722279H, NGV702270H, NGV004465H, NGV712287H, NGV719219H, NGV712890H, NGV650806H, NGV714924H, NGV002303H, NGV711366H, NGV716672H, NGV650793H, NGV703479H, NGV705985H, NGV722782H, NGV706011H, NGV708261H, NGV708403H, NGV714887H, NGV719294H, NGV720262H, NGV002312H, NGV708019H, NGV713355H, NGV714008H, NGV001361H, NGV650564H, NGV650875H, NGV651020H, NGV705308H, NGV705396H, NGV706049H, NGV706156H, NGV710987H, NGV712253H, NGV713099H, NGV713181H, NGV713354H, NGV713946H, NGV716379H, NGV716490H, NGV716665H, NGV720250H, NGV650557H, NGV706248H, NGV716497H, NGV708416H, NGV650551H, NGV706072H, NGV708033H, NGV708400H, NGV708401H, NGV710994H, NGV712126H, NGV715070H, NGV715071H, NGV723067H, NGV716383H, NGV703478H, NGV706185H, NGV713094H, NGV715072H, NGV705099H, NGV713280H, NGV713281H, NGV716374H, NGV721525H, NGV717434H, NGV713283H, NGV706256H, NGV721729H, NGV706044H, NGV650865H, NGV716391H, NGV706000H, NGV702541H, NGV716495H, NGV650895H, NGV651033H, NGV706046H, NGV711360H, NGV713106H, NGV713718H, NGV709467H, NGV719632H, NGV722623H, NGV713738H, NGV002300H, NGV002506H, NGV002516H, NGV650995H, NGV706070H, NGV706200H, NGV708078H, NGV712251H, NGV713185H, NGV713731H, NGV650545H, NGV713945H, NGV714669H, NGV710990H, NGV716671H, NGV710506H, NGV716381H, NGV002316H, NGV650768H, NGV650984H, NGV720299H, NGV711361H, NGV716189H, NGV715093H, NGV713725H, NGV000301H, NGV001374H, NGV001381H, NGV001382H, NGV001394H, NGV001399H, NGV001402H, NGV001406H, NGV002171H, NGV002183H, NGV002186H, NGV004478H, NGV004487H, NGV004488H, NGV004489H, NGV004536H, NGV004539H, NGV004557H, NGV004563H, NGV004605H, NGV651008H, NGV702466H, NGV702467H, NGV702826H, NGV702827H, NGV709338H, NGV709339H, NGV709341H, NGV709348H, NGV709349H, NGV709351H, NGV709354H, NGV709355H, NGV709357H, NGV715986H, NGV716090H, NGV716092H, NGV720313H, NGV721528H, NGV721551H, NGV721554H, NGV721652H, NGV721655H, NGV721661H, NGV721663H, NGV721674H, NGV721721H, NGV708034H, NGV709358H, NGV710526H, NGV710609H, NGV710632H, NGV715133H, NGV715442H, NGV001658H, NGV004691H, NGV702950H, NGV704139H, NGV712175H, NGV712281H, NGV712282H, NGV714367H, NGV714910H, NGV714952H, NGV715160H, NGV716413H, NGV718086H, NGV718087H, NGV720241H, NGV720699H, NGV722646H, NGV723102H, NGV700019H, NGV702366H, NGV004599H, NGV702918H, NGV703115H, NGV703292H, NGV703301H, NGV706763H, NGV706765H, NGV706779H, NGV706825H, NGV710842H, NGV710844H, NGV710848H, NGV710857H, NGV710862H, NGV710863H, NGV710865H, NGV713426H, NGV713428H, NGV713430H, NGV713435H, NGV713437H, NGV713439H, NGV713440H, NGV001155H, NGV001485H, NGV001514H, NGV001530H, NGV001538H, NGV650536H, NGV701506H, NGV004595H, NGV650837H, NGV650855H, NGV702471H, NGV712930H, NGV715321H, NGV717442H, NGV723118H, NGV723119H, NGV712936H, NGV722251H, NGV650537H, NGV713213H, NGV650538H, NGV702379H, NGV713220H, NGV715254H, NGV717430H, NGV000597H, NGV650539H, NGV650540H, NGV703746H, NGV705392H, NGV712316H, NGV700202H, NGV722295H, NGV722607H, NGV717115H, NGV718431H, NGV722299H, NGV702376H, NGV720192H, NGV000602H, NGV702378H, NGV706038H, NGV000601H, NGV706033H, NGV000598H, NGV000600H, NGV702377H, NGV706537H, NGV712931H, NGV712937H, NGV712940H, NGV000599H, NGV715140H, NGV722316H, NGV722331H, NGV712311H, NGV706532H, NGV003825H, NGV003834H, NGV710597H, NGV710983H, NGV712290H, NGV714236H, NGV714500H, NGV715343H, NGV715344H, NGV722453H, NGV722467H, NGV723145H, NGV004647H, NGV701495H, NGV000626H, NGV001496H, NGV001862H, NGV002083H, NGV003492H, NGV701492H, NGV710549H, NGV710552H, NGV710553H, NGV710555H, NGV710559H, NGV713066H, NGV713120H, NGV713236H, NGV713268H, NGV713550H, NGV715163H, NGV723147H, NGV723443H, NGV701493H, NGV712305H, NGV000627H, NGV000628H, NGV000629H, NGV001501H, NGV001649H, NGV001669H, NGV003380H, NGV714682H, NGV715126H, NGV001178H, NGV001469H, NGV001491H, NGV003413H, NGV701365H, NGV701494H, NGV701754H, NGV713561H, NGV713854H, NGV714668H, NGV721741H, NGV001176H, NGV001461H, NGV001661H, NGV001714H, NGV001719H, NGV003359H, NGV701491H, NGV715389H, NGV716176H, NGV722311H, NGV705394H, NGV706577H, NGV715162H, NGV701002H, NGV702910H, NGV707792H, NGV708404H, NGV713851H, NGV714667H, NGV003533H, NGV702901H, NGV713856H, NGV720769H, NGV721736H, NGV001670H, NGV003412H, NGV003520H, NGV703713H, NGV706158H, NGV709362H, NGV709388H, NGV723448H, NGV722471H, NGV723130H, NGV722266H, NGV002640H, NGV002642H, NGV002643H, NGV002645H, NGV002648H, NGV002764H, NGV002767H, NGV002770H, NGV002773H, NGV003106H, NGV003208H, NGV003252H, NGV003254H, NGV003256H, NGV003257H, NGV003262H, NGV003263H, NGV003264H, NGV003268H, NGV003269H, NGV706539H, NGV706541H, NGV706882H, NGV706884H, NGV706887H, NGV706890H, NGV706891H, NGV706893H, NGV706896H, NGV706897H, NGV706900H, NGV706901H, NGV706902H, NGV706911H, NGV706915H, NGV706916H, NGV706917H, NGV706921H, NGV706926H, NGV706929H, NGV706931H, NGV706933H, NGV706934H, NGV002179H, NGV002552H, NGV650507H, NGV706014H, NGV709329H, NGV714216H, NGV001162H, NGV701805H, NGV703386H, NGV705020H, NGV705048H, NGV706865H, NGV714105H, NGV716400H, NGV717723H, NGV720691H, NGV720694H, NGV709359H, NGV722141H, NGV002201H, NGV650530H, NGV700621H, NGV703457H, NGV703458H, NGV708065H, NGV708357H, NGV710698H, NGV710711H, NGV714238H, NGV714239H, NGV714556H, NGV714557H, NGV716099H, NGV716169H, NGV722326H, NGV722424H, NGV722479H, NGV722540H, NGV004584H, NGV004896H, NGV700221H, NGV700526H, NGV704844H, NGV706529H, NGV706586H, NGV706766H, NGV706783H, NGV709170H, NGV709177H, NGV710523H, NGV710525H, NGV710792H, NGV713278H, NGV714545H, NGV714547H, NGV714550H, NGV716401H, NGV716402H, NGV716403H, NGV716410H, NGV709186H, NGV703699H, NGV700944H, NGV705980H, NGV705981H, NGV720260H, NGV705759H, NGV709439H, NGV716414H, NGV700087H, NGV702629H, NGV702644H, NGV713371H, NGV710560H, NGV710595H, NGV710564H, NGV712988H, NGV713335H, NGV701372H, NGV713005H, NGV713327H, NGV002022H, NGV713347H, NGV713008H, NGV713874H, NGV701367H, NGV713877H, NGV715375H, NGV700698H, NGV702005H, NGV712229H, NGV713071H, NGV716124H, NGV000639H, NGV000774H, NGV651138H, NGV701755H, NGV701757H, NGV701759H, NGV707719H, NGV707720H, NGV707722H, NGV707723H, NGV708040H, NGV708041H, NGV714652H, NGV714653H, NGV714657H, NGV715172H, NGV715371H, NGV719411H, NGV719448H, NGV002075H, NGV650062H, NGV708031H, NGV722285H, NGV701756H, NGV707727H, NGV712317H, NGV712459H, NGV713069H, NGV715171H, NGV715206H, NGV715368H, NGV000787H, NGV721513H, NGV712318H, NGV002574H, NGV002580H, NGV002597H, NGV716239H, NGV717685H, NGV717691H, NGV714495H, NGV714214H, NGV714507H, NGV714215H, NGV720788H, NGV000608H, NGV000609H, NGV000611H, NGV001340H, NGV001408H, NGV001410H, NGV001411H, NGV001417H, NGV001419H, NGV001424H, NGV001442H, NGV001448H, NGV001449H, NGV001450H, NGV001451H, NGV001452H, NGV001456H, NGV001458H, NGV001473H, NGV001474H, NGV001477H, NGV001478H, NGV001482H, NGV001550H, NGV001631H, NGV001654H, NGV002014H, NGV002016H, NGV002025H, NGV002026H, NGV002027H, NGV002265H, NGV002271H, NGV002273H, NGV002286H, NGV004696H, NGV004724H, NGV004733H, NGV004741H, NGV004845H, NGV650512H, NGV650513H, NGV650517H, NGV650522H, NGV650892H, NGV650893H, NGV650942H, NGV651042H, NGV651060H, NGV651061H, NGV651068H, NGV651110H, NGV651121H, NGV651133H, NGV651134H, NGV713499H, NGV713747H, NGV713783H, NGV713861H, NGV714107H, NGV715014H, NGV716427H, NGV716428H, NGV716430H, NGV716432H, NGV700020H, NGV700026H, NGV700027H, NGV700031H, NGV700032H, NGV700033H, NGV713191H, NGV713196H, NGV713208H, NGV713210H, NGV713323H, NGV000612H, NGV001296H, NGV001311H, NGV001317H, NGV001601H, NGV001607H, NGV001608H, NGV001610H, NGV001617H, NGV001662H, NGV001664H, NGV001667H, NGV001709H, NGV001743H, NGV001786H, NGV001787H, NGV001793H, NGV001798H, NGV001814H, NGV651893H, NGV651894H, NGV704548H, NGV704563H, NGV704574H, NGV704575H, NGV704602H, NGV704831H, NGV713858H, NGV714029H, NGV714349H, NGV714353H, NGV704252H, NGV704828H, NGV713014H, NGV713019H, NGV713020H, NGV713021H, NGV713064H, NGV713097H, NGV714676H, NGV714678H, NGV714680H, NGV714684H, NGV704762H, NGV708406H, NGV708408H, NGV710534H, NGV710535H, NGV710537H, NGV710538H, NGV710540H, NGV710541H, NGV710543H, NGV001829H, NGV704027H, NGV704029H, NGV003495H, NGV713133H, NGV713136H, NGV704246H, NGV001462H, NGV001463H, NGV001620H, NGV001640H, NGV001944H, NGV001960H, NGV002003H, NGV002005H, NGV002006H, NGV003289H, NGV003432H, NGV650280H, NGV650286H, NGV650291H, NGV650294H, NGV650299H, NGV650317H, NGV650318H, NGV650348H.NGV650356H, NGV650373H, NGV650397H, NGV650478H, NGV650486H, NGV650752H, NGV650758H, NGV702876H, NGV702881H, NGV702885H, NGV702886H, NGV702888H, NGV703400H, NGV703404H, NGV704763H, NGV704808H, NGV704810H, NGV704813H, NGV704818H, NGV704819H, NGV704825H, NGV704829H, NGV713142H, NGV713145H, NGV713146H, NGV713148H, NGV713179H, NGV713538H, NGV713803H, NGV714802H, NGV714849H, NGV714852H, NGV650417H, NGV714850H, NGV704033H, NGV714707H, NGV703433H, NGV703434H, NGV714741H, NGV701750H, NGV701753H, NGV712953H, NGV722838H, NGV723049H, NGV002584H, NGV002585H, NGV004917H, NGV004918H, NGV700157H, NGV701774H, NGV702257H, NGV703483H, NGV703488H, NGV703495H, NGV705751H, NGV705771H, NGV706128H, NGV706521H, NGV710489H, NGV710491H, NGV712300H, NGV712972H, NGV713001H, NGV713365H, NGV713376H, NGV714510H, NGV714511H, NGV714514H, NGV714516H, NGV716164H, NGV716167H, NGV719173H, NGV719184H, NGV004636H, NGV004624H, NGV004625H, NGV004646H, NGV004630H, NGV001815H, NGV004631H, NGV723063H, NGV000027H, NGV004269H, NGV004902H, NGV700239H, NGV710509H, NGV710514H, NGV710731H, NGV710737H, NGV712878H, NGV712925H, NGV715057H, NGV717727H, NGV717728H, NGV651585H, NGV700167H, NGV700832H, NGV705004H, NGV715386H, NGV724559H, NGV702326H, NGV000307H, NGV709253H, NGV004342H, NGV710533H, NGV004417H, NGV004576H, NGV004890H, NGV650543H, NGV651564H, NGV700151H, NGV701997H, NGV702355H, NGV705057H, NGV706584H, NGV707697H, NGV710516H, NGV712120H, NGV712898H, NGV712949H, NGV716322H, NGV720868H, NGV002591H, NGV004588H, NGV650001H, NGV709128H, NGV707743H, NGV724571H, NGV000033H, NGV001357H, NGV001358H, NGV004463H, NGV650084H, NGV651051H, NGV700640H, NGV710464H, NGV712895H, NGV715059H, NGV716732H, NGV720772H, NGV000010H, NGV000022H, NGV000031H, NGV000281H, NGV000352H, NGV000532H, NGV001146H, NGV001148H, NGV001151H, NGV002205H, NGV002550H, NGV002551H, NGV002611H, NGV003521H, NGV003523H, NGV004340H, NGV650502H, NGV650503H, NGV650505H, NGV650563H, NGV651556H, NGV651902H, NGV652565H, NGV700616H, NGV701794H, NGV702259H, NGV702367H, NGV703317H, NGV705038H, NGV705842H, NGV705991H, NGV706130H, NGV706503H, NGV706506H, NGV707744H, NGV707786H, NGV707797H, NGV707816H, NGV709335H, NGV709369H, NGV710500H, NGV710504H, NGV710687H, NGV710693H, NGV712123H, NGV712901H, NGV712906H, NGV713359H, NGV713361H, NGV713449H, NGV714100H, NGV714211H, NGV714217H, NGV714484H, NGV714541H, NGV714950H, NGV715044H, NGV715064H, NGV715089H, NGV715091H, NGV715634H, NGV716121H, NGV716308H, NGV716313H, NGV717730H, NGV717732H, NGV718895H, NGV719016H, NGV719077H, NGV720022H, NGV721498H, NGV721550H, NGV722142H, NGV722258H, NGV722277H, NGV722323H, NGV723124H, NGV651965H, NGV700826H, NGV705060H, NGV707704H, NGV710634H, NGV714102H, NGV715522H, NGV716238H, NGV719170H, NGV708383H, NGV005010H, NGV707822H, NGV712910H, NGV714101H, NGV714474H, NGV716323H, NGV715031H, NGV004587H, NGV005095H, NGV703447H, NGV705058H, NGV706140H, NGV712944H, NGV714220H, NGV717720H, NGV720019H, NGV714477H, NGV710478H, NGV720051H, NGV723081H, NGV001161H, NGV710627H, NGV712921H, NGV715047H, NGV715109H, NGV715275H, NGV717437H, NGV720034H, NGV700830H, NGV709149H, NGV710481H, NGV712322H, NGV712950H, NGV715058H, NGV720151H, NGV705045H, NGV714231H, NGV714509H, NGV650091H, NGV700163H, NGV703385H, NGV708427H, NGV709256H, NGV714498H, NGV722614H, NGV723056H, NGV712913H, NGV714221H, NGV715043H, NGV701773H, NGV004324H, NGV703454H, NGV705023H, NGV710791H, NGV714227H, NGV709191H, NGV709386H, NGV712894H, NGV714531H, NGV004322H, NGV000464H, NGV004276H, NGV004278H, NGV004402H, NGV004892H, NGV700231H, NGV700609H, NGV706016H, NGV707725H, NGV708292H, NGV709156H, NGV709168H, NGV000311H, NGV651955H, NGV714218H, NGV715050H, NGV721527H, NGV722603H, NGV722612H, NGV000350H, NGV720428H, NGV714396H, NGV716274H, NGV722237H, NGV723068H, NGV723128H, NGV002576H, NGV000279H, NGV000513H, NGV004343H, NGV707921H, NGV708061H, NGV710471H, NGV711658H, NGV720139H, NGV000194H, NGV000312H, NGV702653H, NGV705005H, NGV708601H, NGV714369H, NGV716173H, NGV721518H, NGV709360H, NGV714792H, NGV004915H, NGV705062H, NGV703310H, NGV707665H, NGV714769H, NGV002209H, NGV002295H, NGV705061H, NGV004331H, NGV650066H, NGV708394H, NGV709183H, NGV710788H, NGV713358H, NGV714779H, NGV714949H, NGV716370H, NGV719031H, NGV720839H, NGV705046H, NGV709141H, NGV712320H, NGV715511H, NGV712899H, NGV722239H, NGV723125H, NGV004593H, NGV705056H, NGV651272H, NGV709389H, NGV723126H, NGV709453H, NGV004316H, NGV004888H, NGV701472H, NGV702359H, NGV709382H, NGV712943H, NGV712946H, NGV714219H, NGV714230H, NGV714392H, NGV714646H, NGV714647H, NGV716103H, NGV719206H, NGV719996H, NGV003574H, NGV003584H, NGV004241H, NGV004898H, NGV650275H, NGV650506H, NGV650508H, NGV701771H, NGV702627H, NGV702895H, NGV705725H, NGV707702H, NGV710673H, NGV711902H, NGV713446H, NGV713448H, NGV714778H, NGV714943H, NGV715101H, NGV714370H, NGV716332H, NGV715474H, NGV004268H, NGV712887H, NGV651581H, NGV716113H, NGV722129H, NGV722309H, NGV723110H, NGV717724H, NGV704135H, NGV702773H, NGV706570H, NGV706575H, NGV710584H, NGV710703H, NGV713654H, NGV713667H, NGV713883H, NGV713921H, NGV714923H, NGV715572H, NGV715582H, NGV718415H, NGV720276H, NGV722161H, NGV715107H, NGV718893H, NGV004585H, NGV705998H, NGV700520H, NGV000520H, NGV701005H, NGV701012H, NGV708045H, NGV710497H, NGV714900H, NGV003511H, NGV003512H, NGV651114H, NGV702468H, NGV704780H, NGV704783H, NGV706124H, NGV709455H, NGV712912H, NGV714543H, NGV716160H, NGV721452H, NGV721455H, NGV721500H, NGV721508H, NGV721510H, NGV721552H, NGV003599H, NGV004923H, NGV702816H, NGV000515H, NGV709393H, NGV709397H, NGV710498H, NGV713391H, NGV715030H, NGV717011H, NGV722249H, NGV001158H, NGV002586H, NGV003823H, NGV651967H, NGV700218H, NGV702363H, NGV702460H, NGV702461H, NGV705848H, NGV708049H, NGV708059H, NGV708391H, NGV708392H, NGV709330H, NGV709333H, NGV717692H, NGV717696H, NGV721725H, NGV722164H, NGV003602H, NGV003613H, NGV003614H, NGV003622H, NGV003623H, NGV003627H, NGV707835H, NGV707838H, NGV707903H, NGV707909H, NGV707910H, NGV707911H, NGV711907H, NGV712262H, NGV712264H, NGV712268H, NGV712276H, NGV712279H, NGV714559H, NGV714560H, NGV714564H, NGV714638H, NGV714639H, NGV714641H, NGV000276H, NGV004346H, NGV004475H, NGV702369H, NGV704025H, NGV706138H, NGV708010H, NGV710679H, NGV713399H, NGV713424H, NGV722288H, NGV650533H, NGV713386H, NGV720431H, NGV722169H, NGV707700H, NGV708025H, NGV714493H, NGV715115H, NGV717722H, NGV717783H, NGV720285H, NGV720293H, NGV703450H, NGV707708H, NGV707932H, NGV714110H, NGV715034H, NGV716340H, NGV000349H, NGV001145H, NGV002489H, NGV002595H, NGV703481H, NGV705791H, NGV706040H, NGV706136H, NGV712919H, NGV713012H, NGV710626H, NGV722383H, NGV000007H, NGV000021H, NGV000524H, NGV000527H, NGV001889H, NGV001901H, NGV002598H, NGV002601H, NGV002603H, NGV002605H, NGV002606H, NGV002612H, NGV002616H, NGV002619H, NGV650265H, NGV650272H, NGV651574H, NGV652441H, NGV652456H, NGV652457H, NGV652459H, NGV652468H, NGV652470H, NGV652472H, NGV652490H, NGV652494H, NGV700388H, NGV700389H, NGV700503H, NGV700511H, NGV700512H, NGV702462H, NGV702463H, NGV705946H, NGV706058H, NGV706062H, NGV706063H, NGV709342H, NGV709346H, NGV709347H, NGV709374H, NGV709376H, NGV709378H, NGV709379H, NGV709381H, NGV710483H, NGV710515H, NGV710712H, NGV710716H, NGV710725H, NGV710726H, NGV712124H, NGV712897H, NGV712905H, NGV712908H, NGV712909H, NGV712911H, NGV714518H, NGV714526H, NGV714534H, NGV714537H, NGV716667H, NGV716669H, NGV717789H, NGV718089H, NGV718107H, NGV720408H, NGV720416H, NGV724573H, NGV727941H, NGV727959H, NGV004318H, NGV651586H, NGV703448H, NGV707818H, NGV710517H, NGV710518H, NGV714946H, NGV718163H, NGV720232H, NGV003571H, NGV003592H, NGV003596H, NGV004594H, NGV652440H, NGV702542H, NGV706443H, NGV714503H, NGV715051H, NGV706026H, NGV706034H, NGV001149H, NGV722291H, NGV722292H, NGV004569H, NGV004899H, NGV650526H, NGV703379H, NGV703451H, NGV706123H, NGV706133H, NGV708402H, NGV713223H, NGV714535H, NGV714888H, NGV716177H, NGV722130H, NGV722138H, NGV714117H, NGV000009H, NGV000029H, NGV000037H, NGV003281H, NGV003366H, NGV650008H, NGV708116H, NGV708117H, NGV714496H, NGV714642H, NGV714881H, NGV714898H, NGV714902H, NGV718084H, NGV651058H, NGV703449H, NGV722103H, NGV651091H, NGV715069H, NGV000038H, NGV712879H, NGV002234H, NGV002284H, NGV004084H, NGV004349H, NGV004390H, NGV004396H, NGV004897H, NGV651275H, NGV651333H, NGV702362H, NGV703459H, NGV703470H, NGV703471H, NGV704788H, NGV704793H, NGV705787H, NGV705835H, NGV718873H, NGV718891H, NGV720040H, NGV001152H, NGV002210H, NGV002220H, NGV002582H, NGV003839H, NGV003843H, NGV003847H, NGV004263H, NGV004589H, NGV004591H, NGV004601H, NGV004621H, NGV004623H, NGV703328H, NGV703360H, NGV707688H, NGV707692H, NGV707696H, NGV707701H, NGV707706H, NGV707711H, NGV716670H, NGV001157H, NGV001902H, NGV002098H, NGV002374H, NGV002444H, NGV002557H, NGV004275H, NGV650535H, NGV650547H, NGV651915H, NGV702820H, NGV704866H, NGV711660H, NGV712117H, NGV713277H, NGV717704H, NGV717707H, NGV002554H, NGV002630H, NGV002635H, NGV002638H, NGV002639H, NGV652453H, NGV710656H, NGV710670H, NGV714213H, NGV714485H, NGV716733H, NGV717439H, NGV717440H, NGV719327H, NGV003509H, NGV701702H, NGV701790H, NGV707803H, NGV707808H, NGV707809H, NGV707832H, NGV715085H, NGV650789H, NGV719277H, NGV717726H, NGV708422H, NGV708426H; GTIN 00763000122775, Serial Numbers: NGV716502H, NGV717565H, NGV723411H, NGV002028H, NGV002071H, NGV002093H, NGV002279H, NGV002280H, NGV002283H, NGV002285H, NGV002287H, NGV002305H, NGV002308H, NGV002311H, NGV002317H, NGV701393H, NGV701503H, NGV701741H, NGV701746H, NGV701749H, NGV701949H, NGV702241H, NGV709184H, NGV709193H, NGV709371H, NGV709383H, NGV709392H, NGV709436H, NGV709464H, NGV710425H, NGV713807H, NGV713936H, NGV713937H, NGV714649H, NGV714673H, NGV714719H, NGV714896H, NGV715441H, NGV715443H, NGV715819H, NGV716318H, NGV721097H, NGV721526H, NGV001143H, NGV001154H, NGV001156H, NGV001171H, NGV001480H, NGV002631H, NGV002632H, NGV002647H, NGV002650H, NGV002652H, NGV651708H, NGV651744H, NGV652305H, NGV652538H, NGV652548H, NGV652597H, NGV702782H, NGV702785H, NGV702799H, NGV702801H, NGV704054H, NGV704141H, NGV704151H, NGV704156H, NGV707733H, NGV707741H, NGV707747H, NGV707760H, NGV707769H, NGV708419H, NGV708421H, NGV710654H, NGV710660H, NGV710662H, NGV710665H, NGV710667H, NGV712247H, NGV712249H, NGV712299H, NGV712881H, NGV712886H, NGV713401H, NGV713733H, NGV713755H, NGV713758H, NGV713772H, NGV713779H, NGV714729H, NGV714733H, NGV714908H, NGV714982H, NGV714986H, NGV717718H, NGV717738H, NGV718225H, NGV718314H, NGV718318H, NGV718412H, NGV718433H, NGV718516H, NGV650401H, NGV700985H, NGV702276H, NGV702281H, NGV702283H, NGV702291H, NGV702625H, NGV702776H, NGV702780H, NGV702981H, NGV706143H, NGV706149H, NGV706530H, NGV706560H, NGV706574H, NGV707830H, NGV707918H, NGV708133H, NGV708134H, NGV708149H, NGV708260H, NGV708262H, NGV713114H, NGV713132H, NGV713134H, NGV713135H, NGV713137H, NGV713408H, NGV713949H, NGV713950H, NGV713955H, NGV714528H, NGV714687H, NGV714698H, NGV714710H, NGV714809H, NGV714821H, NGV714857H, NGV714873H, NGV714875H, NGV714899H, NGV714922H, NGV714932H, NGV714942H, NGV714947H, NGV715243H, NGV715346H, NGV715354H, NGV715362H, NGV715539H, NGV715602H, NGV716192H, NGV716282H, NGV716812H, NGV717144H, NGV717150H, NGV717153H, NGV717216H, NGV717224H, NGV717228H, NGV717237H, NGV719464H, NGV719466H, NGV719467H, NGV719473H, NGV719479H, NGV719637H, NGV720176H, NGV720186H, NGV720188H, NGV720190H, NGV700024H, NGV700028H, NGV700030H, NGV700069H, NGV700094H, NGV700120H, NGV700137H, NGV712130H, NGV712141H, NGV712142H, NGV712143H, NGV714983H, NGV714988H, NGV715042H, NGV715045H, NGV715052H, NGV720244H, NGV720249H, NGV720290H, NGV720303H, NGV720309H, NGV716098H, NGV723370H, NGV651827H, NGV700801H, NGV712218H, NGV712893H, NGV712957H, NGV715977H, NGV716456H, NGV722518H, NGV000097H, NGV001971H, NGV705747H, NGV705979H, NGV707703H, NGV715437H, NGV715702H, NGV716668H, NGV721497H, NGV721739H, NGV712896H, NGV713262H, NGV721522H, NGV708131H, NGV713933H, NGV000178H, NGV002228H, NGV004458H, NGV650102H, NGV650260H, NGV652258H, NGV700684H, NGV712230H, NGV713404H, NGV716097H, NGV716411H, NGV718871H, NGV718901H, NGV719075H, NGV719379H, NGV720329H, NGV720687H, NGV720762H, NGV721530H, NGV003013H, NGV003619H, NGV003629H, NGV003805H, NGV003928H, NGV650300H, NGV650524H, NGV650541H, NGV651793H, NGV651834H, NGV700021H, NGV700023H, NGV700702H, NGV703806H, NGV705108H, NGV705723H, NGV706013H, NGV706053H, NGV706125H, NGV707831H, NGV708069H, NGV708333H, NGV708359H, NGV710590H, NGV711673H, NGV712240H, NGV712242H, NGV713409H, NGV713732H, NGV714700H, NGV714954H, NGV715134H, NGV716419H, NGV716443H, NGV718594H, NGV720834H, NGV721061H, NGV700362H, NGV003556H, NGV003615H, NGV703735H, NGV703774H, NGV703781H, NGV703796H, NGV703802H, NGV703803H, NGV723399H, NGV706589H, NGV002211H, NGV709175H, NGV710466H, NGV710648H, NGV712927H, NGV714504H, NGV717762H, NGV717770H, NGV720352H, NGV001602H, NGV003534H, NGV003536H, NGV003635H, NGV651788H, NGV652213H, NGV652247H, NGV700946H, NGV700979H, NGV702256H, NGV702274H, NGV702954H, NGV707919H, NGV709145H, NGV712220H, NGV712891H, NGV713074H, NGV713081H, NGV713140H, NGV713221H, NGV713272H, NGV713932H, NGV714722H, NGV715460H, NGV716120H, NGV716358H, NGV717462H, NGV717553H, NGV717554H, NGV720154H, NGV722652H, NGV717717H, NGV721645H, NGV000961H, NGV001613H, NGV652298H, NGV652628H, NGV703299H, NGV704241H, NGV707674H, NGV707772H, NGV708052H, NGV708396H, NGV708432H, NGV709332H, NGV712892H, NGV715054H, NGV718221H, NGV719030H, NGV003014H, NGV003612H, NGV700369H, NGV707724H, NGV709167H, NGV712145H, NGV714666H, NGV716651H, NGV720418H, NGV712224H, NGV713013H, NGV714886H, NGV715211H, NGV716326H, NGV716334H, NGV000112H, NGV003549H, NGV723069H, NGV723139H, NGV650301H, NGV002011H, NGV002626H, NGV003499H, NGV003652H, NGV003878H, NGV650126H, NGV700186H, NGV700387H, NGV700391H, NGV708055H, NGV708345H, NGV710730H, NGV710798H, NGV710802H, NGV710806H, NGV710864H, NGV710878H, NGV710887H, NGV711367H, NGV711652H, NGV711661H, NGV712239H, NGV720144H, NGV720314H, NGV720790H, NGV722488H, NGV004264H, NGV710675H, NGV720205H, NGV001653H, NGV003083H, NGV652654H, NGV702267H, NGV710814H, NGV710818H, NGV710845H, NGV716106H, NGV717772H, NGV718613H, NGV720816H, NGV650017H, NGV703462H, NGV713785H, NGV715238H, NGV715241H, NGV719453H, NGV001629H, NGV002627H, NGV002628H, NGV700381H, NGV700471H, NGV700505H, NGV700522H, NGV702983H, NGV707663H, NGV708062H, NGV708068H, NGV710796H, NGV000968H, NGV000971H, NGV001020H, NGV001716H, NGV001735H, NGV002181H, NGV002469H, NGV002519H, NGV004462H, NGV004467H, NGV004473H, NGV650298H, NGV700014H, NGV701163H, NGV701176H, NGV703305H, NGV703309H, NGV707735H, NGV707740H, NGV707773H, NGV713018H, NGV713159H, NGV713249H, NGV713259H, NGV714651H, NGV714654H, NGV718685H, NGV001668H, NGV002019H, NGV702959H, NGV702969H, NGV717242H, NGV001963H, NGV003054H, NGV003606H, NGV005128H, NGV005136H, NGV005221H, NGV650020H, NGV650037H, NGV650209H, NGV650221H, NGV650321H, NGV650350H, NGV650370H, NGV650388H, NGV651660H, NGV651667H, NGV651670H, NGV651682H, NGV652653H, NGV700530H, NGV700535H, NGV700699H, NGV703327H, NGV703465H, NGV704233H, NGV707926H, NGV708028H, NGV708270H, NGV708293H, NGV708349H, NGV710795H, NGV710868H, NGV713084H, NGV713102H, NGV714542H, NGV716684H, NGV717441H, NGV717684H, NGV717700H, NGV720141H, NGV720153H, NGV720424H, NGV720683H, NGV721745H, NGV704036H, NGV705105H, NGV714958H, NGV714965H, NGV720165H, NGV721657H, NGV722345H, NGV712225H, NGV717435H, NGV710583H, NGV712922H, NGV713059H, NGV001637H, NGV001651H, NGV002023H, NGV004493H, NGV004650H, NGV706075H, NGV706083H, NGV723061H, NGV723083H, NGV723378H; GTIN 00763000122782, Serial Numbers: NGV702456H, NGV700017H, NGV700022H, NGV700096H, NGV702907H, NGV710706H, NGV713571H, NGV722360H; GTIN 00763000122799, Serial Numbers: NGV714816H, NGV650544H, NGV651127H, NGV651131H, NGV704021H, NGV712094H, NGV713128H, NGV713381H, NGV713853H, NGV714116H, NGV714391H, NGV719424H, NGV720041H, NGV723111H, NGV651347H, NGV651680H, NGV711363H, NGV713558H, NGV715255H, NGV716377H, NGV716657H, NGV721566H, NGV003662H, NGV713229H, NGV714393H, NGV716140H, NGV718813H, NGV000302H, NGV000304H, NGV000310H, NGV001494H, NGV002847H, NGV002853H, NGV650944H, NGV651250H, NGV700034H, NGV700041H, NGV703452H, NGV704031H, NGV704144H, NGV704757H, NGV708374H, NGV709405H, NGV712235H, NGV713200H, NGV713537H, NGV713547H, NGV714479H, NGV714532H, NGV715542H, NGV716378H, NGV716452H, NGV718088H, NGV718214H, NGV721546H, NGV650521H, NGV650611H, NGV650711H, NGV702965H, NGV703763H, NGV703764H, NGV722621H, NGV706079H, NGV706086H, NGV721665H, NGV000740H, NGV000741H, NGV001460H, NGV708371H, NGV708372H, NGV713540H, NGV713546H, NGV713668H, NGV002835H, NGV004644H, NGV702948H, NGV712100H, NGV713543H, NGV713767H, NGV718881H, NGV705735H, NGV713282H, NGV715024H, NGV721090H, NGV710647H, NGV719234H, NGV715226H, NGV715427H, NGV718095H, NGV652361H, NGV651062H, NGV651105H, NGV651111H, NGV651115H, NGV713663H, NGV719407H, NGV707815H, NGV708369H, NGV003585H, NGV004482H, NGV650288H, NGV650509H, NGV650912H, NGV651103H, NGV652301H, NGV700043H, NGV703787H, NGV704149H, NGV713271H, NGV714502H, NGV714721H, NGV715029H, NGV715159H, NGV715267H, NGV718195H, NGV719434H, NGV722307H, NGV713855H, NGV720852H, NGV721671H, NGV722289H, NGV002857H, NGV004692H, NGV650516H, NGV714118H, NGV714490H, NGV715005H, NGV720828H, NGV722321H, NGV715311H, NGV002779H, NGV002833H, NGV003579H, NGV004578H, NGV004590H, NGV652646H, NGV710833H, NGV712365H, NGV713382H, NGV714050H, NGV714827H, NGV715032H, NGV722367H, NGV722557H, NGV722613H; GTIN 00763000122805, Serial Numbers: NGV003240H, NGV003501H, NGV003503H, NGV706020H, NGV707699H; GTIN 00763000122812, Serial Numbers: NGV712165H, NGV712167H, NGV716261H, NGV717231H, NGV718092H, NGV718093H, NGV718425H, NGV718434H, NGV721666H, NGV722128H, NGV722134H, NGV722163H, NGV722177H, NGV722209H, NGV722223H, NGV706035H, NGV706193H, NGV706331H, NGV712413H, NGV713165H, NGV708057H, NGV710749H, NGV710793H, NGV711379H, NGV711382H, NGV711383H, NGV710860H, NGV722347H, NGV718899H, NGV720818H, NGV710859H, NGV710867H, NGV703399H, NGV703853H, NGV721535H, NGV722145H, NGV722160H, NGV703756H, NGV721664H, NGV703851H, NGV651910H, NGV704812H, NGV004486H, NGV652461H, NGV700619H, NGV004908H, NGV701745H, NGV005094H, NGV005099H, NGV651998H, NGV705441H, NGV706031H, NGV714212H, NGV714224H, NGV714558H, NGV005110H, NGV701082H, NGV703517H, NGV703870H, NGV704578H, NGV704748H, NGV704814H, NGV704823H, NGV705035H, NGV705385H, NGV705395H, NGV705442H, NGV709713H, NGV710453H, NGV710462H, NGV710474H, NGV710715H, NGV710794H, NGV710812H, NGV710854H, NGV711386H, NGV712414H, NGV714235H, NGV714704H, NGV717149H, NGV707810H, NGV004477H, NGV005121H, NGV710479H, NGV710816H, NGV713007H, NGV713753H, NGV004474H, NGV651997H, NGV703868H, NGV004469H, NGV005068H, NGV005133H, NGV700025H, NGV703524H, NGV703530H, NGV703866H, NGV704775H, NGV713389H, NGV713390H, NGV652433H, NGV713117H, NGV714929H, NGV005091H, NGV651072H, NGV651860H, NGV700083H, NGV702937H, NGV705043H, NGV705107H, NGV709707H, NGV714705H, NGV703520H, NGV712156H, NGV712158H, NGV704824H, NGV700123H, NGV700983H, NGV704827H, NGV707709H, NGV713384H, NGV703303H, NGV713177H, NGV717223H, NGV713483H, NGV004912H, NGV652438H, NGV004916H, NGV705367H, NGV710778H, NGV703403H, NGV704600H, NGV704820H, NGV710624H, NGV710629H, NGV710701H, NGV712986H, NGV707811H, NGV707917H, NGV718875H, NGV703857H, NGV710493H, NGV710781H, NGV724564H, NGV000299H, NGV650211H, NGV650222H, NGV652572H, NGV652600H, NGV710528H, NGV720178H, NGV724570H, NGV700506H, NGV712444H, NGV650051H, NGV722352H, NGV700219H, NGV709914H, NGV710680H, NGV720709H, NGV712999H, NGV652307H, NGV000509H, NGV000967H, NGV002898H, NGV003782H, NGV652485H, NGV652503H, NGV703501H, NGV703895H, NGV000960H, NGV700242H, NGV700710H, NGV707902H, NGV708337H, NGV709728H, NGV724520H, NGV708339H, NGV717735H, NGV650016H, NGV700523H, NGV713754H, NGV713559H, NGV718406H, NGV719068H, NGV720302H, NGV700521H, NGV704815H, NGV708439H, NGV713123H, NGV713188H, NGV652569H, NGV714374H, NGV719274H, NGV714360H, NGV650012H, NGV708413H, NGV713548H, NGV650247H, NGV710618H, NGV650056H, NGV000306H, NGV002845H, NGV713167H, NGV713107H, NGV713866H, NGV712418H, NGV714377H, NGV717721H, NGV717731H, NGV720308H, NGV650283H, NGV703732H, NGV721065H, NGV724538H, NGV724574H, NGV700513H, NGV720297H, NGV713009H, NGV713762H, NGV714357H, NGV650290H, NGV707804H, NGV650237H, NGV713017H, NGV713093H, NGV724525H, NGV002893H, NGV700501H, NGV708346H, NGV721515H, NGV720430H; GTIN 00763000122829, Serial Numbers: NGV000773H, NGV000942H, NGV002717H, NGV003759H, NGV003765H, NGV003780H, NGV003783H, NGV003799H, NGV003821H, NGV003833H, NGV004271H, NGV004272H, NGV004649H, NGV004911H, NGV004913H, NGV705748H, NGV705752H, NGV705756H, NGV705789H, NGV705836H, NGV706015H, NGV706817H, NGV706827H, NGV706935H, NGV707521H, NGV709601H, NGV709722H, NGV710480H, NGV710539H, NGV710546H, NGV710600H, NGV710611H, NGV710676H, NGV710686H, NGV710692H, NGV710723H, NGV710727H, NGV710728H, NGV710729H, NGV710739H, NGV710742H, NGV710743H, NGV710744H, NGV710747H, NGV710891H, NGV713786H, NGV713790H, NGV713800H, NGV713801H, NGV713806H, NGV713808H, NGV713809H, NGV713849H, NGV713915H, NGV713917H, NGV713948H, NGV713951H, NGV715086H, NGV715361H, NGV715374H, NGV715436H, NGV715447H, NGV716440H, NGV716447H, NGV720744H, NGV720812H, NGV721087H, NGV721408H, NGV721429H, NGV721456H, NGV721514H, NGV721724H, NGV721744H, NGV722159H, NGV722259H, NGV722261H, NGV722263H, NGV722332H, NGV722595H, NGV723066H, NGV723109H, NGV723112H, NGV003909H, NGV004070H, NGV004071H, NGV004568H, NGV005123H, NGV701021H, NGV704153H, NGV723297H, NGV723314H, NGV723580H, NGV723584H, NGV003221H, NGV003235H, NGV003238H, NGV003241H, NGV003858H, NGV003886H, NGV003895H, NGV706166H, NGV706167H, NGV706500H, NGV706535H, NGV706542H, NGV706630H, NGV706757H, NGV706795H, NGV706799H, NGV706815H, NGV710612H, NGV710617H, NGV710620H, NGV710631H, NGV710640H, NGV710650H, NGV710651H, NGV710652H, NGV710668H, NGV710674H, NGV710855H, NGV712114H, NGV712125H, NGV712128H, NGV712151H, NGV712460H, NGV712462H, NGV712469H, NGV712471H, NGV712474H, NGV713360H, NGV713780H, NGV713781H, NGV713827H, NGV713920H, NGV713922H, NGV713929H, NGV713935H, NGV713940H, NGV713947H, NGV714051H, NGV714355H, NGV714371H, NGV714372H, NGV719356H, NGV719475H, NGV720138H, NGV721034H, NGV721060H, NGV721066H, NGV721070H, NGV721077H, NGV721523H, NGV721549H, NGV721643H, NGV721651H, NGV721719H, NGV721720H, NGV723127H, NGV723285H, NGV650438H, NGV712298H, NGV712304H, NGV712458H, NGV723078H, NGV723089H; GTIN 00763000122836, Serial Numbers: NGV001153H, NGV001170H, NGV001481H, NGV001396H, NGV001398H, NGV718171H, NGV705724H, NGV710888H, NGV713539H, NGV713241H, NGV713237H, NGV713267H; GTIN 00763000122850, Serial Numbers: NGV650217H, NGV650218H, NGV714978H, NGV724863H, NGV650230H, NGV712228H, NGV713481H, NGV650810H, NGV724276H, NGV724282H, NGV724346H, NGV700010H, NGV713115H, NGV724345H, NGV724526H, NGV724535H, NGV701026H, NGV650238H, NGV650202H, NGV714104H, NGV650617H, NGV704146H, NGV710643H, NGV713769H, NGV716132H, NGV717710H, NGV724334H, NGV724340H, NGV650233H, NGV713771H, NGV002859H, NGV712948H, NGV718178H, NGV000292H, NGV650252H, NGV707812H, NGV724846H, NGV715440H, NGV715638H, NGV724310H, NGV650214H, NGV650223H, NGV704138H, NGV714787H, NGV724837H, NGV650240H, NGV650229H, NGV001981H, NGV002790H, NGV712132H, NGV713556H, NGV000473H, NGV000490H, NGV700970H, NGV713016H, NGV724841H, NGV714771H, NGV715448H, NGV724548H, NGV724327H; GTIN 00763000421823, Serial Numbers: NGV724316H, NGV735395H, NGV735400H, NGV739998H, NGV732002H, NGV735376H, NGV727435H, NGV733098H, NGV734860H, NGV739588H, NGV724566H, NGV724576H, NGV724577H, NGV724578H, NGV730635H, NGV731955H, NGV731969H, NGV734753H, NGV740852H, NGV740853H, NGV740870H, NGV730631H, NGV730643H, NGV728258H, NGV728926H, NGV730629H, NGV732096H, NGV734110H, NGV734755H, NGV735061H, NGV737570H, NGV737595H, NGV740505H, NGV723987H, NGV725069H, NGV725118H, NGV727433H, NGV727966H, NGV730022H, NGV734754H, NGV737450H, NGV732750H, NGV723968H, NGV727132H, NGV729300H, NGV729473H, NGV733893H, NGV735693H, NGV735695H, NGV723431H, NGV727074H, NGV729640H, NGV729659H, NGV730684H, NGV730686H, NGV730687H, NGV730688H, NGV732061H, NGV732065H, NGV734133H, NGV734944H, NGV735944H, NGV735973H, NGV735977H, NGV731259H, NGV732037H, NGV726330H, NGV727258H, NGV729552H, NGV729940H, NGV729969H, NGV730160H, NGV730617H, NGV731426H, NGV731942H, NGV731944H, NGV731971H, NGV733082H, NGV733437H, NGV735378H, NGV735753H, NGV736655H, NGV737097H, NGV738287H, NGV738648H, NGV738653H, NGV739426H, NGV739877H, NGV740840H, NGV723433H, NGV724016H, NGV724848H, NGV724854H, NGV726354H, NGV727369H, NGV727421H, NGV728193H, NGV728822H, NGV729511H, NGV730016H, NGV730628H, NGV730663H, NGV731214H, NGV731501H, NGV731510H, NGV732081H, NGV733873H, NGV734493H, NGV734650H, NGV735409H, NGV735574H, NGV735696H, NGV735986H, NGV736485H, NGV737176H, NGV737682H, NGV737686H, NGV737694H, NGV737696H, NGV737742H, NGV740455H, NGV740744H, NGV727949H, NGV724694H, NGV725644H, NGV727424H, NGV727952H, NGV731264H, NGV731993H, NGV732490H, NGV733424H, NGV735741H, NGV739481H, NGV739676H, NGV740100H, NGV740740H, NGV723427H, NGV723471H, NGV727999H, NGV728371H, NGV731226H, NGV732560H, NGV734815H, NGV735716H, NGV736482H, NGV737102H, NGV737467H, NGV737771H, NGV738415H, NGV739318H, NGV740753H, NGV735770H, NGV727083H, NGV731448H, NGV734275H, NGV734408H, NGV735756H, NGV725670H, NGV736738H, NGV738895H, NGV738971H, NGV739315H, NGV740847H, NGV725013H, NGV725024H, NGV725151H, NGV725154H, NGV727951H, NGV729545H, NGV731959H, NGV731963H, NGV732641H, NGV732657H, NGV734089H, NGV734094H, NGV735401H, NGV735405H, NGV737126H, NGV737163H, NGV737621H, NGV737636H, NGV737641H, NGV738372H, NGV738571H, NGV738574H, NGV738849H, NGV739320H, NGV739398H, NGV739412H, NGV739418H, NGV739461H, NGV739486H, NGV739553H, NGV739577H, NGV740692H, NGV740695H, NGV736490H, NGV738514H, NGV738577H, NGV734756H, NGV737620H, NGV738920H, NGV725541H, NGV731400H, NGV732099H, NGV725540H, NGV732000H, NGV737461H, NGV730196H, NGV740009H, NGV740064H, NGV725347H, NGV732121H, NGV732123H, NGV738390H, NGV740096H, NGV740015H, NGV724763H, NGV727082H, NGV728993H, NGV730021H, NGV730404H, NGV734594H, NGV735575H, NGV735742H, NGV740156H, NGV728191H, NGV732059H, NGV734190H, NGV734872H, NGV737639H, NGV723435H, NGV723676H, NGV724973H, NGV725275H, NGV725672H, NGV725693H, NGV725839H, NGV725857H, NGV726912H, NGV727066H, NGV727120H, NGV727127H, NGV727954H, NGV727956H, NGV728043H, NGV728142H, NGV728144H, NGV728190H, NGV728249H, NGV728301H, NGV729155H, NGV730406H, NGV730409H, NGV730705H, NGV730706H, NGV730737H, NGV731395H, NGV731487H, NGV731953H, NGV732039H, NGV732198H, NGV732206H, NGV732675H, NGV733104H, NGV733957H, NGV733978H, NGV734002H, NGV734037H, NGV734042H, NGV734187H, NGV734391H, NGV734392H, NGV734394H, NGV734485H, NGV734572H, NGV734919H, NGV734931H, NGV736558H, NGV736969H, NGV737088H, NGV737181H, NGV737184H, NGV738314H, NGV738613H, NGV738833H, NGV738872H, NGV739013H, NGV739014H, NGV739016H, NGV739567H, NGV739568H, NGV740360H, NGV740732H, NGV724199H, NGV728097H, NGV728760H, NGV737087H, NGV737022H, NGV724074H, NGV724081H, NGV725778H, NGV728040H, NGV732751H, NGV736574H, NGV741164H, NGV723674H, NGV724948H, NGV729150H, NGV731962H, NGV732007H, NGV732366H, NGV732367H, NGV732636H, NGV733968H, NGV734409H, NGV734827H, NGV734829H, NGV737679H, NGV739319H, NGV724782H, NGV724866H, NGV725256H, NGV728368H, NGV730015H, NGV731195H, NGV732082H, NGV732085H, NGV732358H, NGV732690H, NGV735755H, NGV735847H, NGV736710H, NGV740001H, NGV725726H, NGV734373H, NGV739552H, NGV735667H, NGV725231H, NGV732746H, NGV734019H, NGV734711H, NGV734854H, NGV723644H, NGV724306H, NGV728151H, NGV733519H, NGV733525H, NGV737451H, NGV739883H, NGV732763H, NGV734603H, NGV740731H, NGV725833H, NGV727144H, NGV732102H, NGV732106H, NGV733376H, NGV734395H, NGV735757H, NGV739670H, NGV741165H, NGV724083H, NGV724093H, NGV732342H, NGV732589H, NGV734279H, NGV734280H, NGV734822H, NGV737192H, NGV737193H, NGV739990H, NGV739992H, NGV734360H, NGV739839H, NGV739871H, NGV723486H, NGV724106H, NGV725241H, NGV727408H, NGV728755H, NGV730742H, NGV732702H, NGV733542H, NGV733952H, NGV734423H, NGV735385H, NGV736564H, NGV737118H, NGV740637H, NGV724227H, NGV724814H, NGV726938H, NGV727147H, NGV728208H, NGV734057H, NGV734267H, NGV737743H, NGV738994H, NGV739422H, NGV723965H, NGV724505H, NGV725848H, NGV725926H, NGV727353H, NGV730604H, NGV730623H, NGV732682H, NGV733391H, NGV734517H, NGV735797H, NGV737744H, NGV738426H, NGV738591H, NGV723933H, NGV724307H, NGV724352H, NGV724357H, NGV724529H, NGV728166H, NGV728977H, NGV734079H, NGV736717H, NGV739044H, NGV740011H, NGV725058H, NGV730714H, NGV732725H, NGV734196H, NGV735060H, NGV735764H, NGV723948H, NGV723962H, NGV724533H, NGV724765H, NGV727231H, NGV728262H, NGV728807H, NGV729130H, NGV730647H, NGV731461H, NGV732013H, NGV732341H, NGV732343H, NGV732369H, NGV732483H, NGV733402H, NGV734106H, NGV734496H, NGV735381H, NGV735660H, NGV737165H, NGV737169H, NGV737483H, NGV738320H, NGV739019H, NGV739400H, NGV725897H, NGV730670H, NGV734509H, NGV738379H, NGV740814H, NGV724169H, NGV729608H, NGV730031H, NGV732113H, NGV739041H, NGV737577H, NGV737626H, NGV738603H, NGV724137H, NGV727176H, NGV727777H, NGV728099H, NGV728263H, NGV728291H, NGV732705H, NGV733414H, NGV733545H, NGV734585H, NGV734780H, NGV735348H, NGV735390H, NGV735668H, NGV736607H, NGV737191H, NGV738609H, NGV739548H, NGV723483H, NGV724805H, NGV724813H, NGV724965H, NGV728814H, NGV730027H, NGV732689H, NGV732708H, NGV733974H, NGV735699H, NGV738546H, NGV738602H, NGV739025H, NGV723571H, NGV723577H, NGV723977H, NGV732668H, NGV734758H, NGV734759H, NGV738525H, NGV738531H, NGV738645H, NGV739396H, NGV723429H, NGV724129H, NGV724463H, NGV727315H, NGV727317H, NGV728185H, NGV728247H, NGV728274H, NGV730096H, NGV732684H, NGV733408H, NGV735585H, NGV736580H, NGV739421H, NGV739569H, NGV740074H, NGV724096H, NGV724098H, NGV724553H, NGV725826H, NGV725919H, NGV730367H, NGV730637H, NGV732097H, NGV732757H, NGV735382H, NGV735383H, NGV738646H, NGV738647H, NGV739402H, NGV739404H, NGV739916H, NGV740253H, NGV740254H, NGV727428H, NGV731974H, NGV732196H, NGV732693H, NGV735644H, NGV738663H, NGV740056H, NGV727776H, NGV728803H, NGV730024H, NGV732191H, NGV733400H, NGV735356H, NGV736575H, NGV738380H, NGV739413H, NGV727310H, NGV733888H, NGV736601H, NGV739050H, NGV739428H, NGV725554H, NGV734951H, NGV740464H, NGV723445H, NGV723452H, NGV723886H, NGV724120H, NGV725841H, NGV731301H, NGV732769H, NGV734058H, NGV730624H, NGV732756H, NGV733434H, NGV734263H, NGV734490H, NGV736671H, NGV737645H, NGV739406H, NGV723646H, NGV725845H, NGV730342H, NGV735577H, NGV735598H, NGV724554H, NGV724956H, NGV727425H, NGV727441H, NGV727442H, NGV731231H, NGV734516H, NGV734521H, NGV734645H, NGV736652H, NGV736670H, NGV737173H, NGV737646H, NGV738594H, NGV738598H, NGV739032H, NGV740742H, NGV723437H, NGV724501H, NGV727056H, NGV728205H, NGV729602H, NGV730616H, NGV730724H, NGV732100H, NGV732640H, NGV732737H, NGV734804H, NGV737098H, NGV739017H, NGV739858H, NGV723478H, NGV727071H, NGV727443H, NGV728044H, NGV733513H, NGV734465H, NGV735666H, NGV736653H, NGV738926H, NGV728297H, NGV732016H, NGV732109H, NGV734805H, NGV736646H, NGV739322H, NGV739572H, NGV740829H, NGV729006H, NGV723458H, NGV723460H, NGV724134H, NGV724262H, NGV727239H, NGV728376H, NGV729509H, NGV729592H, NGV732172H, NGV732346H, NGV732683H, NGV734503H, NGV735838H, NGV736684H, NGV737482H, NGV738599H, NGV739030H, NGV740206H, NGV740230H, NGV740812H, NGV724060H, NGV724101H, NGV724231H, NGV724359H, NGV724472H, NGV724478H, NGV724511H, NGV724557H, NGV724563H, NGV724771H, NGV724773H, NGV724785H, NGV724799H, NGV724824H, NGV724826H, NGV724847H, NGV725235H, NGV725297H, NGV725519H, NGV725597H, NGV725673H, NGV726748H, NGV726759H, NGV727275H, NGV727341H, NGV727342H, NGV727969H, NGV727994H, NGV728153H, NGV728169H, NGV728200H, NGV728203H, NGV728359H, NGV728363H, NGV729542H, NGV729553H, NGV729559H, NGV729582H, NGV729611H, NGV729630H, NGV729882H, NGV730020H, NGV730064H, NGV730384H, NGV730603H, NGV730702H, NGV730738H, NGV731252H, NGV731256H, NGV731258H, NGV731262H, NGV731268H, NGV731528H, NGV731939H, NGV732135H, NGV732136H, NGV732153H, NGV732332H, NGV732574H, NGV732659H, NGV732660H, NGV732666H, NGV732681H, NGV732685H, NGV732714H, NGV732718H, NGV732723H, NGV732768H, NGV732771H, NGV733395H, NGV733425H, NGV733543H, NGV733544H, NGV733954H, NGV734115H, NGV734116H, NGV734126H, NGV734246H, NGV734410H, NGV734422H, NGV734494H, NGV734495H, NGV734576H, NGV734580H, NGV734752H, NGV734757H, NGV734769H, NGV734830H, NGV734843H, NGV734858H, NGV734859H, NGV735364H, NGV735367H, NGV735369H, NGV735384H, NGV735569H, NGV735571H, NGV735652H, NGV735758H, NGV735765H, NGV735779H, NGV736431H, NGV736682H, NGV736713H, NGV737094H, NGV737096H, NGV737651H, NGV737657H, NGV737658H, NGV737660H, NGV738381H, NGV738665H, NGV739037H, NGV739039H, NGV739302H, NGV739399H, NGV739464H, NGV739473H, NGV739485H, NGV739555H, NGV739689H, NGV739774H, NGV739984H, NGV740006H, NGV740070H, NGV740082H, NGV740087H, NGV740153H, NGV740166H, NGV740456H, NGV740743H, NGV740864H, NGV740873H, NGV723728H, NGV724794H, NGV727238H, NGV728150H, NGV728157H, NGV729504H, NGV730638H, NGV731230H, NGV731520H, NGV733527H, NGV736115H, NGV736494H, NGV736495H, NGV739423H, NGV739432H, NGV739439H, NGV739945H, NGV740003H, NGV724575H, NGV730699H, NGV732053H, NGV732662H, NGV736132H, NGV724102H, NGV725632H, NGV725784H, NGV725844H, NGV727844H, NGV729506H, NGV729587H, NGV732338H, NGV733517H, NGV733518H, NGV735583H, NGV739321H, NGV739401H, NGV739562H, NGV739563H, NGV740126H, NGV729441H, NGV731205H, NGV738641H, NGV739310H, NGV739864H, NGV740739H, NGV724778H, NGV724780H, NGV725602H, NGV727308H, NGV728080H, NGV728949H, NGV728969H, NGV730613H, NGV732676H, NGV735089H, NGV737111H, NGV738852H, NGV739298H, NGV739411H, NGV727991H, NGV724326H, NGV726919H, NGV727344H, NGV728973H, NGV730025H, NGV730606H, NGV731940H, NGV734276H, NGV734471H, NGV734659H, NGV735241H, NGV736118H, NGV737670H, NGV739043H, NGV739055H, NGV739578H, NGV739909H, NGV739987H, NGV740497H, NGV727576H, NGV728063H, NGV729877H, NGV731267H, NGV732711H, NGV734489H, NGV735372H, NGV739045H, NGV725237H, NGV728957H, NGV729131H, NGV730345H, NGV730680H, NGV732762H, NGV733401H, NGV734177H, NGV735363H, NGV735763H, NGV736599H, NGV737172H, NGV738385H, NGV738660H, NGV739405H, NGV739570H, NGV739571H, NGV739903H, NGV740768H, NGV740830H, NGV724951H, NGV724952H, NGV725605H, NGV727440H, NGV728033H, NGV728065H, NGV728282H, NGV728970H, NGV730689H, NGV731213H, NGV732044H, NGV732585H, NGV733407H, NGV733413H, NGV734088H, NGV734770H, NGV736109H, NGV738341H, NGV739305H, NGV739307H, NGV739564H, NGV725011H, NGV728377H, NGV728752H, NGV732163H, NGV724006H, NGV724801H, NGV726745H, NGV730151H, NGV731208H, NGV736136H, NGV739420H, NGV739918H, NGV740757H, NGV727363H, NGV727960H, NGV736113H, NGV724116H, NGV724821H, NGV725196H, NGV727409H, NGV728162H, NGV728241H, NGV732108H, NGV734583H, NGV734866H, NGV735649H, NGV739047H, NGV739407H, NGV739906H, NGV724512H, NGV725248H, NGV727919H, NGV728147H, NGV732036H, NGV734378H, NGV734867H, NGV737674H, NGV738543H, NGV738560H, NGV739469H, NGV724226H, NGV724953H, NGV724969H, NGV727064H, NGV727821H, NGV729627H, NGV732161H, NGV733435H, NGV740499H, NGV724284H, NGV727149H, NGV727154H, NGV730364H, NGV732586H, NGV732669H, NGV732691H, NGV736478H, NGV723590H, NGV724014H, NGV724087H, NGV725009H, NGV730602H, NGV730619H, NGV731228H, NGV731598H, NGV733418H, NGV734127H, NGV735671H, NGV736721H, NGV736724H, NGV737190H, NGV737655H, NGV737662H, NGV739904H, NGV740004H, NGV740704H, NGV740878H, NGV734772H, NGV739309H, NGV739445H, NGV724275H, NGV724354H, NGV725927H, NGV729483H, NGV730050H, NGV730349H, NGV730633H, NGV731255H, NGV732740H, NGV733530H, NGV735388H, NGV735783H, NGV735816H, NGV737090H, NGV737122H, NGV737123H, NGV739409H, NGV739579H, NGV740005H, NGV740860H, NGV740877H, NGV725120H, NGV727337H, NGV734248H, NGV735389H, NGV735788H, NGV723980H, NGV724844H, NGV725000H, NGV730131H, NGV732151H, NGV723482H, NGV724004H, NGV727028H, NGV727172H, NGV734587H, NGV734592H, NGV736117H, NGV737115H, NGV739554H, NGV739867H, NGV739869H, NGV727279H, NGV727851H, NGV730132H, NGV731306H, NGV731937H, NGV735394H, NGV737099H, NGV739419H, NGV723442H, NGV724125H, NGV724230H, NGV727427H, NGV730351H, NGV730615H, NGV732066H, NGV740738H, NGV737462H, NGV724166H, NGV733540H, NGV740139H, NGV740076H, NGV731249H, NGV735256H, NGV732091H, NGV724025H, NGV724580H, NGV724849H, NGV730112H, NGV723989H, NGV728818H, NGV730400H, NGV734402H, NGV738844H, NGV723988H, NGV729513H, NGV731184H, NGV731186H, NGV735417H, NGV736567H, NGV738897H, NGV732078H, NGV733894H, NGV736489H, NGV737770H, NGV739058H, NGV730139H, NGV734185H, NGV739898H, NGV736460H, NGV725704H, NGV725724H, NGV729934H, NGV738655H, NGV738667H, NGV739995H, NGV729529H, NGV732160H, NGV730658H, NGV724715H, NGV725213H, NGV735655H, NGV736620H, NGV724514H, NGV724835H, NGV725258H, NGV725584H, NGV725589H, NGV725769H, NGV725777H, NGV726339H, NGV727412H, NGV727415H, NGV730042H, NGV730645H, NGV734272H, NGV734424H, NGV736590H, NGV736727H, NGV738397H, NGV738506H, NGV740073H, NGV724278H, NGV724462H, NGV724696H, NGV724702H, NGV724704H, NGV724709H, NGV724711H, NGV732359H, NGV732561H, NGV732573H, NGV732578H, NGV732638H, NGV735265H, NGV735267H, NGV735326H, NGV735328H, NGV735341H, NGV724088H, NGV728802H, NGV739574H, NGV740470H, NGV725779H, NGV730163H, NGV735648H, NGV735670H, NGV741398H, NGV725150H, NGV725153H, NGV728055H, NGV729007H, NGV738989H, NGV739476H, NGV740845H, NGV735665H, NGV739863H, NGV740128H, NGV740511H, NGV732048H, NGV735664H, NGV739663H, NGV735386H, NGV724859H, NGV728173H, NGV735255H, NGV736128H, NGV736749H, NGV739994H, NGV732699H, NGV734182H, NGV735658H, NGV735745H, NGV735771H, NGV730138H, NGV733112H, NGV733389H, NGV734175H, NGV737164H, NGV724047H, NGV724050H, NGV727191H, NGV727242H, NGV727245H, NGV727248H, NGV728829H, NGV728830H, NGV729866H, NGV731218H, NGV731236H, NGV731277H, NGV731279H, NGV735775H, NGV735776H, NGV735778H, NGV737593H, NGV737715H, NGV737717H, NGV739541H, NGV739542H, NGV739544H, NGV739545H, NGV739546H, NGV736612H, NGV734169H, NGV724043H, NGV728152H, NGV728155H, NGV723775H, NGV725292H, NGV725565H, NGV726363H, NGV727345H, NGV728186H, NGV729618H, NGV731945H, NGV733411H, NGV733412H, NGV735088H, NGV735774H, NGV739313H, NGV739584H, NGV740114H, NGV740483H, NGV740489H, NGV740714H, NGV725637H, NGV728261H, NGV733537H, NGV733538H, NGV730614H, NGV724029H, NGV727938H, NGV739308H, NGV738846H, NGV739312H, NGV739878H, NGV740122H, NGV723587H, NGV727430H, NGV730646H, NGV739482H, NGV737476H, NGV737766H, NGV725338H, NGV727990H, NGV728168H, NGV724080H, NGV724286H, NGV724297H, NGV727336H, NGV728347H, NGV728362H, NGV740516H, NGV730402H, NGV735732H, NGV731196H, NGV731199H, NGV724344H, NGV727352H, NGV728175H, NGV728182H, NGV730601H, NGV734026H, NGV734129H, NGV736430H, NGV736696H, NGV737773H, NGV731220H, NGV732330H, NGV738583H, NGV722625H, NGV724264H, NGV732031H, NGV732058H, NGV734231H, NGV736493H, NGV739885H, NGV740473H, NGV724153H, NGV728045H, NGV730677H, NGV734033H, NGV734240H, NGV735588H, NGV736663H, NGV738391H, NGV740129H, NGV740131H, NGV725852H, NGV725856H, NGV729571H, NGV729577H, NGV730143H, NGV731244H, NGV731481H, NGV732063H, NGV732064H, NGV737470H, NGV738836H, NGV724252H, NGV724766H, NGV724977H, NGV728001H, NGV730639H, NGV734740H, NGV736700H, NGV739933H, NGV725594H, NGV729598H, NGV731943H, NGV734647H, NGV726460H, NGV733375H, NGV725277H, NGV731516H, NGV728669H, NGV729532H, NGV729536H, NGV729540H, NGV729605H, NGV730361H, NGV733887H, NGV736690H, NGV737739H, NGV737741H, NGV740501H, NGV740509H, NGV740769H, NGV728950H, NGV729557H, NGV735972H, NGV734472H, NGV725204H, NGV726221H, NGV727070H, NGV732648H, NGV736591H, NGV738524H, NGV740176H, NGV724600H, NGV729563H, NGV731424H, NGV739543H, NGV723908H, NGV725569H, NGV729899H, NGV734109H, NGV734278H, NGV732098H, NGV732148H, NGV732150H, NGV734511H, NGV734514H, NGV736439H, NGV736444H, NGV738563H, NGV738640H, NGV740454H, NGV728197H, NGV740763H, NGV727943H, NGV733416H, NGV734260H, NGV737712H, NGV726920H, NGV724091H, NGV740479H, NGV723979H, NGV724298H, NGV724303H, NGV727754H, NGV727761H, NGV728266H, NGV729142H, NGV732752H, NGV733111H, NGV733370H, NGV734737H, NGV734743H, NGV740192H, NGV740706H, NGV740709H, NGV741393H, NGV724991H, NGV728048H, NGV739052H, NGV724078H, NGV724079H, NGV731182H, NGV741325H, NGV728978H, NGV731180H, NGV739848H, NGV730053H, NGV730057H, NGV730113H, NGV730120H, NGV730125H, NGV732754H, NGV732760H, NGV737512H, NGV737558H, NGV739059H, NGV739061H, NGV739065H, NGV739067H, NGV741400H, NGV741403H, NGV741412H, NGV741414H, NGV741420H, NGV741428H, NGV741435H, NGV741436H, NGV741439H, NGV741442H, NGV727305H, NGV728201H, NGV735410H, NGV735796H, NGV740200H, NGV741410H, NGV730019H, NGV734810H, NGV738893H, NGV739430H, NGV740766H, NGV740771H, NGV725652H, NGV734917H, NGV739680H, NGV730041H, NGV724604H, NGV724607H, NGV724608H, NGV724836H, NGV727997H, NGV728736H, NGV730156H, NGV731934H, NGV731986H, NGV737186H, NGV737207H, NGV737211H, NGV737561H, NGV737562H, NGV737604H, NGV739449H, NGV739849H, NGV739852H, NGV728751H, NGV732075H, NGV732177H, NGV730071H, NGV724044H, NGV725674H, NGV738490H, NGV723441H, NGV736576H, NGV731932H, NGV727361H, NGV727368H, NGV727984H, NGV728165H, NGV732650H, NGV736639H, NGV737456H, NGV738358H, NGV739311H, NGV723596H, NGV729620H, NGV729625H, NGV732158H, NGV735375H, NGV739874H, NGV727326H, NGV728811H, NGV732190H, NGV732735H, NGV734461H, NGV739902H, NGV740010H, NGV740832H, NGV724036H, NGV725549H, NGV725663H, NGV725665H, NGV727241H, NGV727416H, NGV728158H, NGV729556H, NGV729681H, NGV729682H, NGV729709H, NGV732073H, NGV732086H, NGV733378H, NGV733381H, NGV733398H, NGV735590H, NGV735591H, NGV735592H, NGV739502H, NGV739557H, NGV739559H, NGV739560H, NGV739876H, NGV739889H, NGV739919H, NGV739921H, NGV739929H, NGV740474H, NGV740475H, NGV740854H, NGV740857H, NGV734802H, NGV736582H, NGV739583H, NGV724517H, NGV727747H, NGV729003H, NGV740478H, NGV740720H, NGV740772H, NGV723973H, NGV738674H, NGV740255H, NGV728094H, NGV730626H, NGV736597H, NGV738904H, NGV740765H, NGV728161H, NGV730715H, NGV732729H, NGV733436H, NGV737116H, NGV739480H, NGV732068H, NGV728759H, NGV732025H, NGV734761H, NGV728812H, NGV732087H, NGV734112H, NGV734741H, NGV724061H, NGV724343H, NGV728284H, NGV734173H, NGV727252H, NGV727431H, NGV727962H, NGV727972H, NGV731269H, NGV739587H, NGV740773H, NGV740836H, NGV741406H, NGV730133H, NGV730703H, NGV728159H, NGV728799H, NGV730111H, NGV731285H, NGV731436H, NGV724356H, NGV725240H, NGV727769H, NGV733903H, NGV735380H, NGV740077H, NGV740767H, NGV724003H, NGV725065H, NGV725862H, NGV726904H, NGV728732H, NGV729579H, NGV732590H, NGV733410H, NGV724270H, NGV735750H, NGV739892H, NGV724337H, NGV730037H, NGV730664H, NGV732563H, NGV732649H, NGV735244H, NGV735249H, NGV723978H, NGV724263H, NGV724995H, NGV727299H, NGV728176H, NGV730108H, NGV736602H, NGV738909H, NGV740520H, NGV740825H, NGV723912H, NGV726755H, NGV731221H, NGV740521H, NGV725070H, NGV734801H, NGV741388H, NGV739444H, NGV724071H, NGV735748H, NGV739443H, NGV739057H, NGV733963H, NGV727073H, NGV728269H, NGV732094H, NGV734488H, NGV739056H, NGV725005H, NGV732331H, NGV734738H, NGV736588H, NGV724957H, NGV728952H, NGV729566H, NGV730665H, NGV735736H, NGV738364H, NGV728124H, NGV727280H, NGV727414H, NGV732593H, NGV736605H, NGV730164H, NGV734327H, NGV732664H, NGV732686H, NGV733390H, NGV733397H, NGV732697H, NGV734048H, NGV735578H, NGV739585H, NGV739861H, NGV739862H, NGV739982H, NGV740688H, NGV724869H, NGV724944H, NGV729203H, NGV734101H, NGV734102H, NGV734105H, NGV737745H, NGV737747H, NGV724982H, NGV734775H, NGV739986H, NGV732088H, NGV724042H, NGV735659H, NGV730126H, NGV731957H, NGV727413H, NGV734389H, NGV736105H, NGV736552H, NGV736557H, NGV738842H, NGV738924H, NGV738925H, NGV739297H, NGV739997H, NGV740666H, NGV732736H, NGV737091H, NGV734063H, NGV734077H, NGV737074H, NGV738922H, NGV738972H, NGV731177H, NGV740810H, NGV726737H, NGV737454H, NGV725072H, NGV725137H, NGV733419H, NGV733421H, NGV733422H, NGV739290H, NGV739291H, NGV739292H, NGV739296H, NGV725876H, NGV725880H, NGV729531H, NGV731248H, NGV732645H, NGV729868H, NGV730149H, NGV730179H, NGV732079H, NGV732559H, NGV733900H, NGV734242H, NGV735752H, NGV738923H, NGV740471H, NGV741192H, NGV741323H, NGV741411H, NGV739575H, NGV723974H, NGV732134H, NGV733514H, NGV739900H, NGV739913H, NGV724172H, NGV724603H, NGV724768H, NGV725249H, NGV730659H, NGV731455H, NGV732368H, NGV733415H, NGV733971H, NGV734792H, NGV734874H, NGV739901H, NGV727970H, NGV730076H, NGV730080H, NGV731413H, NGV731418H, NGV732564H, NGV732698H, NGV734778H, NGV734779H, NGV736481H, NGV737105H, NGV739547H, NGV739549H, NGV741383H, NGV729507H, NGV732009H, NGV736635H, NGV727306H, NGV728246H, NGV734589H, NGV730618H, NGV732092H, NGV724145H, NGV731274H, NGV731276H, NGV731280H, NGV727768H, NGV729865H, NGV730051H, NGV730667H, NGV732337H, NGV734765H, NGV735661H, NGV735663H, NGV740823H, NGV727197H, NGV727343H, NGV727378H, NGV734798H, NGV734799H, NGV734816H, NGV739593H, NGV739660H, NGV729147H, NGV728951H, NGV725060H, NGV728156H, NGV730620H, NGV732335H, NGV734791H, NGV724205H, NGV725658H, NGV734377H, NGV736695H, NGV724612H, NGV735656H, NGV740106H, NGV740463H, NGV738312H, NGV740506H, NGV728139H, NGV730161H, NGV733532H, NGV732187H, NGV734942H, NGV740717H, NGV736127H, NGV730625H, NGV736589H, NGV736658H, NGV725781H, NGV725782H, NGV728815H, NGV728817H, NGV733431H, NGV734325H, NGV739448H, NGV739457H, NGV741392H, NGV724139H, NGV725849H, NGV729010H, NGV729127H, NGV734863H, NGV734885H, NGV737108H, NGV737110H, NGV738916H, NGV740813H, NGV739429H, NGV732582H, NGV733970H, NGV723650H, NGV732713H, NGV723466H, NGV723984H, NGV740820H, NGV739910H, NGV724328H, NGV724330H, NGV724583H, NGV725882H, NGV725891H, NGV727250H, NGV728019H, NGV728032H, NGV728204H, NGV728206H, NGV729595H, NGV731462H, NGV731466H, NGV731469H, NGV732224H, NGV732225H, NGV732235H, NGV734785H, NGV734786H, NGV734789H, NGV735769H, NGV736499H, NGV736551H, NGV737609H, NGV737624H, NGV738393H, NGV738467H, NGV738473H, NGV739323H, NGV739410H, NGV739434H, NGV739438H, NGV736116H, NGV738392H, NGV726691H, NGV724027H, NGV725206H, NGV726696H, NGV735069H, NGV723735H, NGV728183H, NGV730720H, NGV732077H, NGV734776H, NGV728481H, NGV734773H, NGV737196H, NGV738851H, NGV734597H, NGV723969H, NGV724011H, NGV739603H, NGV727347H, NGV728092H, NGV729614H, NGV730640H, NGV735242H, NGV737602H, NGV737616H, NGV739896H, NGV730385H, NGV730672H, NGV724593H, NGV724595H, NGV727964H, NGV732168H, NGV735075H, NGV737188H, NGV737189H, NGV739303H, NGV730157H, NGV740485H, NGV724180H, NGV731179H, NGV732773H, NGV735565H, NGV735773H, NGV739907H, NGV739996H, NGV736750H, NGV736781H, NGV736783H, NGV727338H, NGV727981H, NGV730608H, NGV734808H, NGV728813H, NGV737614H, NGV728947H, NGV740495H, NGV730358H, NGV727065H, NGV727161H, NGV727374H, NGV727965H, NGV730671H, NGV731227H, NGV732095H, NGV735747H, NGV736436H, NGV740012H, NGV740079H, NGV740504H, NGV733529H, NGV730158H, NGV727772H, NGV734800H, NGV737627H, NGV739581H, NGV724581H, NGV727232H, NGV724975H, NGV728015H, NGV732183H, NGV732186H, NGV739873H, NGV739886H, NGV740013H, NGV740014H, NGV727230H, NGV728174H, NGV728943H, NGV730084H, NGV730673H, NGV732124H, NGV732567H, NGV733373H, NGV734194H, NGV729632H, NGV731470H, NGV734739H, NGV734814H, NGV739431H, NGV739983H, NGV730086H, NGV737735H, NGV737737H, NGV740828H, NGV726918H, NGV738899H, NGV735567H, NGV732716H, NGV740494H, NGV737457H, NGV728299H, NGV734491H, NGV738378H, NGV734510H, NGV741166H, NGV723430H, NGV728170H, NGV734497H, NGV735640H, NGV738520H, NGV724845H, NGV728635H, NGV738376H, NGV739474H, NGV740469H, NGV740648H, NGV741408H, NGV735754H, NGV728054H, NGV736697H, NGV737723H, NGV740248H, NGV730405H, NGV732254H, NGV735391H, NGV735636H, NGV723983H, NGV724197H, NGV740730H, NGV732089H, NGV732213H, NGV739915H, NGV740338H, NGV723426H, NGV724152H, NGV727775H, NGV732894H, NGV738995H, NGV738997H, NGV739004H, NGV739006H, NGV724519H, NGV723904H, NGV723966H, NGV728081H, NGV728089H, NGV728143H, NGV728748H, NGV728820H, NGV728823H, NGV728826H, NGV728913H, NGV732046H, NGV732114H, NGV735710H, NGV738821H, NGV738828H, NGV738830H, NGV730352H, NGV736577H, NGV728209H, NGV723976H, NGV724319H, NGV724320H, NGV728149H, NGV730061H, NGV733512H, NGV735085H, NGV739466H, NGV741387H, NGV724309H, NGV740256H, NGV724008H, NGV727186H, NGV739437H, NGV740634H, NGV740595H, NGV723936H, NGV727307H, NGV728211H, NGV733886H, NGV739416H, NGV725247H, NGV728207H, NGV739905H, NGV727228H, NGV730708H, NGV737174H, NGV731273H, NGV727081H, NGV731253H, NGV733423H, NGV733426H, NGV733433H, NGV737055H, NGV739659H, NGV731946H, NGV732679H, NGV734586H, NGV735985H, NGV738505H, NGV738581H, NGV739500H, NGV741385H, NGV728164H, NGV734251H, NGV736111H, NGV723882H, NGV724545H, NGV724546H, NGV724855H, NGV725642H, NGV725909H, NGV725912H, NGV725913H, NGV727423H, NGV727439H, NGV728593H, NGV729292H, NGV731985H, NGV732008H, NGV732093H, NGV732694H, NGV733405H, NGV734022H, NGV734234H, NGV734270H, NGV734379H, NGV734384H, NGV734532H, NGV735647H, NGV735654H, NGV737538H, NGV737545H, NGV739010H, NGV739550H, NGV739551H, NGV739937H, NGV739942H, NGV740002H, NGV740008H, NGV724059H, NGV724210H, NGV724292H, NGV725033H, NGV727311H, NGV731201H, NGV740250H, NGV740518H, NGV730612H, NGV731404H, NGV731408H, NGV723907H, NGV723993H, NGV723994H, NGV723996H, NGV723999H, NGV734473H, NGV734476H, NGV734480H, NGV734482H, NGV734484H, NGV728179H, NGV729590H, NGV729637H, NGV740755H, NGV740727H, NGV733409H, NGV728958H, NGV731271H, NGV731281H, NGV739540H, NGV740078H, NGV741386H, NGV731203H, NGV732060H, NGV724300H, NGV725057H, NGV725668H, NGV727314H, NGV732733H, NGV733879H, NGV736659H, NGV739415H, NGV739566H, NGV739665H, NGV740308H, NGV727989H, NGV730056H, NGV725520H, NGV729287H, NGV731251H, NGV735749H, NGV735589H; GTIN 00763000421830, Serial Numbers: NGV728016H, NGV725837H, NGV730725H, NGV731260H, NGV735834H, NGV740747H, NGV723740H, NGV724198H, NGV724201H, NGV724232H, NGV724238H, NGV724249H, NGV727983H, NGV727995H, NGV728003H, NGV728034H, NGV728831H, NGV729770H, NGV729774H, NGV729935H, NGV736406H, NGV736427H, NGV738578H, NGV723589H, NGV727077H, NGV730709H, NGV730713H, NGV735711H, NGV735759H, NGV730697H, NGV724184H, NGV727921H, NGV730054H, NGV734056H, NGV736466H, NGV736641H, NGV733429H, NGV738419H, NGV738590H, NGV740707H, NGV724193H, NGV724241H, NGV725831H, NGV727955H, NGV728195H, NGV730065H, NGV730066H, NGV730712H, NGV730719H, NGV730732H, NGV731287H, NGV731405H, NGV731954H, NGV732120H, NGV734054H, NGV734530H, NGV735708H, NGV736405H, NGV738398H, NGV738580H, NGV738586H, NGV738587H, NGV738589H, NGV740172H, NGV740496H, NGV724351H, NGV730731H, NGV733428H, NGV736455H, NGV724127H, NGV724338H, NGV726441H, NGV727134H, NGV727898H, NGV730726H, NGV731178H, NGV733393H, NGV735715H, NGV735723H, NGV737887H, NGV738610H, NGV724055H, NGV727021H, NGV727187H, NGV727243H, NGV727869H, NGV729543H, NGV729759H, NGV733427H, NGV734400H, NGV738413H, NGV736403H, NGV728036H, NGV728083H, NGV734470H, NGV736461H, NGV724186H, NGV730058H, NGV732041H, NGV733521H, NGV734016H, NGV735734H, NGV724122H, NGV727857H, NGV730728H, NGV737898H, NGV740725H, NGV727068H, NGV731286H, NGV735712H, NGV736468H, NGV738357H, NGV728140H, NGV731398H, NGV734015H, NGV724128H, NGV724223H, NGV727125H, NGV727146H, NGV727234H, NGV727244H, NGV727256H, NGV727912H, NGV729521H, NGV729535H, NGV729757H, NGV729762H, NGV730704H, NGV730707H, NGV730721H, NGV731243H, NGV731266H, NGV733392H, NGV734050H, NGV734053H, NGV734364H, NGV734393H, NGV735728H, NGV735729H, NGV736581H, NGV736609H, NGV736616H, NGV737606H, NGV737640H, NGV737886H, NGV737889H, NGV738608H, NGV738611H, NGV740728H, NGV730723H, NGV736467H, NGV736624H, NGV726926H, NGV734405H, NGV736578H, NGV740370H, NGV725835H, NGV725836H, NGV737883H, NGV736648H, NGV727174H, NGV724177H, NGV728071H, NGV734271H, NGV734273H, NGV736463H, NGV729938H, NGV736484H, NGV740719H, NGV732003H, NGV733975H, NGV738576H, NGV724117H, NGV735843H, NGV724243H, NGV731391H, NGV733438H, NGV730059H, NGV730069H, NGV731317H, NGV731530H, NGV734479H, NGV736585H, NGV738593H, NGV740512H, NGV740721H, NGV726913H, NGV730695H, NGV736456H, NGV724208H, NGV724239H, NGV727047H, NGV727260H, NGV729927H, NGV729933H, NGV736587H, NGV739938H, NGV739939H, NGV739940H, NGV740215H, NGV736593H, NGV736604H, NGV739503H, NGV736525H, NGV724136H, NGV725736H, NGV726722H, NGV729301H, NGV729758H, NGV730722H, NGV733396H, NGV734274H, NGV738409H, NGV738600H, NGV739931H, NGV740729H, NGV725688H, NGV728093H, NGV734268H, NGV734526H, NGV734573H, NGV726709H, NGV727044H, NGV728819H, NGV733976H, NGV735851H, NGV738402H, NGV739687H, NGV739690H, NGV739881H, NGV723456H, NGV724222H, NGV726356H, NGV726361H, NGV726688H, NGV727017H, NGV727019H, NGV727076H, NGV727116H, NGV727181H, NGV727350H, NGV727915H, NGV728116H, NGV728192H, NGV728260H, NGV730045H, NGV731299H, NGV731307H, NGV732010H, NGV733386H, NGV734011H, NGV734017H, NGV734266H, NGV734477H, NGV735823H, NGV736475H, NGV738394H, NGV738417H, NGV738573H, NGV738579H, NGV738582H, NGV738595H, NGV738673H, NGV740500H, NGV740722H, NGV740751H, NGV724216H, NGV724250H, NGV724358H, NGV730729H, NGV734641H, NGV734642H, NGV740488H, NGV734023H, NGV736404H, NGV724178H, NGV724245H, NGV730043H, NGV731589H, NGV732125H, NGV733993H, NGV734035H, NGV734040H, NGV735731H, NGV738421H, NGV739558H, NGV724195H, NGV727259H, NGV728021H, NGV724175H, NGV727895H, NGV738607H, NGV723462H, NGV724219H, NGV724248H, NGV726338H, NGV727263H, NGV730073H, NGV724329H, NGV727257H, NGV729937H, NGV736457H, NGV736464H, NGV740278H, NGV740487H, NGV729285H, NGV730694H, NGV740372H, NGV740468H, NGV730698H, NGV740746H, NGV728025H, NGV729920H, NGV729924H, NGV730727H, NGV735831H, NGV728008H, NGV734390H, NGV724229H, NGV734469H, NGV736454H, NGV724114H, NGV728907H, NGV739924H, NGV739934H, NGV740764H, NGV724069H, NGV724082H, NGV724244H, NGV730078H, NGV738666H, NGV734107H, NGV738639H, NGV725824H, NGV727767H, NGV736412H, NGV736383H, NGV736387H, NGV736389H, NGV727778H, NGV728131H, NGV728196H, NGV728348H, NGV728980H, NGV736421H, NGV728270H, NGV728369H, NGV728004H, NGV728006H, NGV728267H, NGV729634H, NGV736396H, NGV728373H, NGV727764H, NGV728365H, NGV728366H, NGV727757H, NGV727758H, NGV727763H, NGV727944H, NGV728010H, NGV728067H, NGV728068H, NGV728117H, NGV728254H, NGV728264H, NGV728941H, NGV728984H, NGV728988H, NGV728989H, NGV735942H, NGV736286H, NGV736399H, NGV736415H, NGV728009H, NGV728245H, NGV728268H, NGV736408H, NGV736426H, NGV736410H, NGV727749H, NGV736424H, NGV728017H, NGV728356H, NGV736325H, NGV727756H, NGV728024H, NGV728360H, NGV728255H, NGV729564H, NGV728123H, NGV727744H, NGV736394H, NGV728991H, NGV736397H, NGV736375H, NGV727752H, NGV736373H, NGV728013H, NGV736378H, NGV728031H, NGV728199H, NGV728930H, NGV728922H, NGV727755H, NGV739487H, NGV738361H, NGV739484H, NGV732587H, NGV732633H, NGV732634H, NGV732637H, NGV732356H, NGV732667H, NGV732722H, NGV732724H, NGV738305H, NGV736732H, NGV732707H, NGV734777H, NGV734840H, NGV734841H, NGV732361H, NGV732717H, NGV732720H, NGV732731H, NGV732732H, NGV732891H, NGV733522H, NGV733878H, NGV734950H, NGV732580H, NGV732653H, NGV734918H, NGV738370H, NGV739605H, NGV732328H, NGV732340H, NGV732592H, NGV732635H, NGV732639H, NGV732703H, NGV732738H, NGV734742H, NGV734748H, NGV734766H, NGV734845H, NGV736680H, NGV737607H, NGV738584H, NGV739462H, NGV739467H, NGV739607H, NGV739608H, NGV739686H, NGV732654H, NGV732658H, NGV734870H, NGV734873H, NGV734877H, NGV734882H, NGV734896H, NGV734910H, NGV739674H, NGV732631H, NGV732764H, NGV734852H, NGV739478H, NGV739616H, NGV739620H, NGV739666H, NGV740519H, NGV740758H, NGV732566H, NGV734916H, NGV740246H, NGV732103H, NGV732329H, NGV732570H, NGV732709H, NGV732727H, NGV733533H, NGV734029H, NGV734204H, NGV740240H, NGV740244H, NGV740770H, NGV732741H, NGV733523H, NGV739662H, NGV732033H, NGV732696H, NGV734043H, NGV732018H, NGV732583H, NGV732643H, NGV732706H, NGV732728H, NGV734656H, NGV739468H, NGV739617H, NGV740335H, NGV734467H, NGV734064H, NGV734750H, NGV738304H, NGV739678H, NGV732884H, NGV732885H, NGV734922H, NGV734935H, NGV734940H, NGV739454H, NGV739465H, NGV740486H, NGV739615H, NGV733528H, NGV734065H, NGV734082H, NGV734844H, NGV737167H, NGV740761H, NGV732334H, NGV732588H, NGV733524H, NGV734749H, NGV740491H, NGV732688H, NGV733541H, NGV733875H, NGV732748H, NGV739611H, NGV734025H, NGV740819H, NGV732581H, NGV732644H, NGV732687H, NGV732734H, NGV733531H, NGV733877H, NGV734207H, NGV734464H, NGV739459H, NGV739463H, NGV739671H, NGV739677H, NGV740332H, NGV724295H, NGV729705H, NGV729707H, NGV729710H, NGV731409H, NGV740368H, NGV723997H, NGV727229H, NGV727438H, NGV728042H, NGV729872H, NGV730656H, NGV731406H, NGV736428H, NGV740369H, NGV728146H, NGV727616H, NGV729699H, NGV727884H, NGV728189H, NGV728062H, NGV734655H, NGV729696H, NGV729149H, NGV729243H, NGV729916H, NGV730026H, NGV732142H, NGV728265H, NGV729922H, NGV730145H, NGV729753H, NGV731222H, NGV727610H, NGV728091H, NGV728101H, NGV729145H, NGV731188H, NGV728072H, NGV729755H, NGV727880H, NGV728187H, NGV728075H, NGV728078H, NGV728181H, NGV729063H, NGV728372H, NGV728967H, NGV730017H, NGV728968H, NGV729911H, NGV731265H, NGV728073H, NGV732035H, NGV731192H, NGV732141H, NGV730036H, NGV727633H, NGV728171H, NGV727872H, NGV729921H, NGV724815H, NGV724816H, NGV725829H, NGV725842H, NGV725846H, NGV725847H, NGV725850H, NGV725851H, NGV725853H, NGV725854H, NGV725855H, NGV725873H, NGV725889H, NGV725903H, NGV725906H, NGV725915H, NGV725917H, NGV725918H, NGV725921H, NGV725929H, NGV726016H, NGV729546H, NGV729558H, NGV729560H, NGV729643H, NGV729644H, NGV729646H, NGV729652H, NGV729654H, NGV729655H, NGV729672H, NGV729674H, NGV729677H, NGV729680H, NGV729687H, NGV729688H, NGV729690H, NGV729691H, NGV729692H, NGV729694H, NGV729701H, NGV729702H, NGV734483H, NGV734486H, NGV734487H, NGV734492H, NGV734499H, NGV734505H, NGV734508H, NGV734512H, NGV734513H, NGV734519H, NGV739600H, NGV739601H, NGV723044H, NGV723970H, NGV724202H, NGV724582H, NGV724710H, NGV724718H, NGV728251H, NGV734084H, NGV734658H, NGV734661H, NGV734662H, NGV734837H, NGV735718H, NGV729612H, NGV734249H, NGV736573H, NGV740851H, NGV727973H, NGV728194H, NGV731949H, NGV740700H, NGV740811H, NGV740694H, NGV741597H, NGV735935H, NGV740868H, NGV741594H, NGV741429H, NGV741448H, NGV735948H, NGV736126H, NGV736133H, NGV736137H, NGV740855H, NGV740859H, NGV741409H, NGV741739H, NGV741893H, NGV741170H, NGV741430H, NGV741432H, NGV741738H, NGV723905H, NGV732138H, NGV733902H, NGV735637H, NGV735662H, NGV736053H, NGV739490H, NGV727963H, NGV732646H, NGV735957H, NGV732004H, NGV727961H, NGV733897H, NGV723982H, NGV724001H, NGV729875H, NGV733947H, NGV735978H, NGV737555H, NGV740274H, NGV740280H, NGV740281H, NGV723991H, NGV735979H, NGV736122H, NGV740283H, NGV727939H, NGV723975H, NGV732721H, NGV723923H, NGV733944H, NGV734824H, NGV727968H, NGV735952H, NGV728037H, NGV732674H, NGV729765H, NGV729766H, NGV734736H, NGV734895H, NGV739488H, NGV724867H, NGV729897H, NGV731526H, NGV724164H, NGV724260H, NGV724596H, NGV731407H, NGV731479H, NGV740295H, NGV740304H, NGV740348H, NGV740354H, NGV731414H, NGV731419H, NGV731529H, NGV724156H, NGV731480H, NGV731508H, NGV739932H, NGV740292H, NGV740303H, NGV740350H, NGV724601H, NGV731432H, NGV737468H, NGV724157H, NGV724206H, NGV731428H, NGV731494H, NGV740822H, NGV724035H, NGV724146H, NGV724207H, NGV731477H, NGV731492H, NGV737575H, NGV740293H, NGV740353H, NGV740371H, NGV731483H, NGV729901H, NGV731421H, NGV731440H, NGV731513H, NGV731527H, NGV737469H, NGV724155H, NGV724916H, NGV729885H, NGV731602H, NGV731493H, NGV729711H, NGV729769H, NGV740229H, NGV724105H, NGV728948H, NGV730681H, NGV733090H, NGV733371H, NGV737515H, NGV729914H, NGV739435H, NGV739440H, NGV739441H, NGV739446H, NGV724769H, NGV724783H, NGV724798H, NGV724860H, NGV731947H, NGV732758H, NGV733969H, NGV733972H, NGV736104H, NGV736106H, NGV736110H, NGV736123H, NGV737629H, NGV737665H, NGV737684H, NGV737689H, NGV738634H, NGV738635H, NGV727946H, NGV734406H, NGV732744H, NGV734418H, NGV734651H, NGV727987H, NGV735719H, NGV734407H, NGV734648H, NGV734649H, NGV732765H, NGV724274H, NGV728000H, NGV734745H, NGV739841H, NGV739891H, NGV739895H, NGV739925H, NGV739927H, NGV740872H, NGV741342H, NGV741424H, NGV741425H, NGV741444H, NGV741445H, NGV741568H, NGV741574H, NGV741584H, NGV741592H, NGV736416H, NGV736418H, NGV724341H, NGV727971H, NGV728810H, NGV738339H, NGV738572H, NGV723606H, NGV729290H, NGV736425H, NGV724350H, NGV738346H, NGV728731H, NGV728805H, NGV738625H, NGV735564H, NGV735566H, NGV735568H, NGV735572H, NGV735599H, NGV735642H, NGV735650H, NGV740513H, NGV726203H, NGV741354H, NGV734092H, NGV734111H, NGV724537H, NGV724544H, NGV724551H, NGV724555H, NGV724560H, NGV724561H, NGV728992H, NGV734838H, NGV734839H, NGV736594H, NGV736595H, NGV736603H, NGV737688H, NGV725263H, NGV725603H, NGV725643H, NGV727967H, NGV728020H, NGV728022H, NGV729537H, NGV729561H, NGV731581H, NGV728981H, NGV729151H, NGV729609H, NGV729779H, NGV729780H, NGV729782H, NGV729852H, NGV730636H, NGV730676H, NGV730678H, NGV733951H, NGV733964H, NGV733965H, NGV733967H, NGV737799H, NGV738569H, NGV729837H, NGV729918H, NGV728915H, NGV725570H, NGV725574H, NGV725617H, NGV729786H, NGV729832H, NGV730674H, NGV731533H, NGV732101H, NGV733880H, NGV733882H, NGV733945H, NGV733948H, NGV733950H, NGV734091H, NGV734097H, NGV734104H, NGV735669H, NGV735726H, NGV735746H, NGV735751H, NGV736401H, NGV736402H, NGV729861H, NGV739501H, NGV723998H, NGV725649H, NGV725657H, NGV725669H, NGV735586H, NGV735737H, NGV737817H, NGV737836H, NGV737857H, NGV727340H, NGV727868H, NGV741449H, NGV727265H, NGV727316H, NGV727327H, NGV727333H, NGV727357H, NGV727974H, NGV734062H, NGV734072H, NGV734073H, NGV734080H, NGV734130H, NGV740193H, NGV729909H, NGV729913H, NGV740826H, NGV740180H, NGV727046H, NGV734059H, NGV734081H, NGV740164H, NGV741451H, NGV734131H, NGV727364H, NGV727887H, NGV727171H, NGV735684H, NGV736443H, NGV728066H, NGV729881H, NGV730654H, NGV735686H, NGV740273H, NGV723895H, NGV724692H, NGV726342H, NGV732652H, NGV735674H, NGV735698H, NGV735740H, NGV735850H, NGV740834H, NGV740849H, NGV724267H, NGV724830H, NGV732117H, NGV733420H, NGV724020H, NGV728064H, NGV730655H, NGV730735H, NGV733959H, NGV736625H, NGV738228H, NGV738232H, NGV738656H, NGV739930H, NGV740842H, NGV727038H, NGV730652H, NGV729004H, NGV727420H, NGV728041H, NGV728049H, NGV737460H, NGV739496H, NGV723918H, NGV724144H, NGV724154H, NGV724572H, NGV724818H, NGV725130H, NGV725208H, NGV725268H, NGV727128H, NGV727137H, NGV727418H, NGV727419H, NGV727436H, NGV727896H, NGV727897H, NGV727977H, NGV728051H, NGV729090H, NGV729502H, NGV729884H, NGV730127H, NGV730657H, NGV730685H, NGV730733H, NGV730734H, NGV731278H, NGV731392H, NGV731410H, NGV732030H, NGV732350H, NGV732352H, NGV732353H, NGV732357H, NGV733368H, NGV733369H, NGV733960H, NGV734463H, NGV734474H, NGV735597H, NGV735694H, NGV735738H, NGV736453H, NGV736491H, NGV736498H, NGV736548H, NGV736549H, NGV736561H, NGV738299H, NGV739492H, NGV739879H, NGV740221H, NGV740285H, NGV740821H, NGV731472H, NGV727976H, NGV729524H, NGV731435H, NGV726571H, NGV730736H, NGV736471H, NGV728050H, NGV729196H, NGV730739H, NGV738294H, NGV740507H, NGV736583H, NGV729874H, NGV725587H, NGV729633H, NGV732024H, NGV734213H, NGV737455H, NGV737477H, NGV727986H, NGV728052H, NGV732348H, NGV724034H, NGV724355H, NGV724838H, NGV729600H, NGV732116H, NGV732354H, NGV733367H, NGV736569H, NGV729905H, NGV734239H, NGV736483H, NGV739491H, NGV724142H, NGV728057H, NGV732127H, NGV740225H, NGV727151H, NGV725827H, NGV726347H, NGV734024H, NGV734235H, NGV734264H, NGV736458H, NGV736542H, NGV736544H, NGV737447H, NGV737472H, NGV738592H, NGV739922H, NGV723208H, NGV737452H, NGV724806H, NGV724857H, NGV739436H, NGV727164H, NGV730660H, NGV730682H, NGV730140H, NGV724842H, NGV732327H, NGV724012H, NGV732028H, NGV732670H, NGV726176H, NGV731450H, NGV724486H, NGV725596H, NGV725865H, NGV726008H, NGV729888H, NGV729889H, NGV730692H, NGV736492H, NGV736592H, NGV726182H, NGV738234H, NGV729900H, NGV734180H, NGV734188H, NGV734466H, NGV735690H, NGV736477H, NGV739493H, NGV740881H, NGV731442H, NGV724013H, NGV726156H, NGV734201H, NGV736479H, NGV737458H, NGV739442H, NGV741433H, NGV727975H, NGV732665H, NGV724701H, NGV730675H, NGV730691H, NGV740508H, NGV740879H, NGV728059H, NGV728060H, NGV724840H, NGV734233H, NGV726017H, NGV729603H, NGV732128H, NGV740198H, NGV740490H, NGV740827H, NGV740875H, NGV724031H, NGV724339H, NGV729157H, NGV736488H, NGV738628H, NGV728285H, NGV726018H, NGV740279H, NGV723232H, NGV725604H, NGV732349H, NGV741902H, NGV725825H, NGV726014H, NGV726261H, NGV726697H, NGV727012H, NGV729202H, NGV729207H, NGV729892H, NGV729894H, NGV731500H, NGV732026H, NGV732131H, NGV732892H, NGV732893H, NGV732998H, NGV733961H, NGV734176H, NGV736469H, NGV736472H, NGV736546H, NGV737685H, NGV738288H, NGV738289H, NGV740218H, NGV740219H, NGV741381H, NGV741401H, NGV727142H, NGV727367H, NGV728026H, NGV728028H, NGV728038H, NGV728039H, NGV728292H, NGV724831H, NGV726289H, NGV727079H, NGV727432H, NGV729930H, NGV732047H, NGV732119H, NGV736474H, NGV738303H, NGV739453H, NGV741438H, NGV733962H, NGV739495H, NGV740861H, NGV728501H, NGV729001H, NGV724823H, NGV735643H, NGV740741H, NGV740824H, NGV740750H, NGV740756H, NGV733002H, NGV737170H, NGV737179H, NGV725840H, NGV727411H, NGV727888H, NGV728061H, NGV729638H, NGV731950H, NGV732110H, NGV732126H, NGV734462H, NGV734653H, NGV734654H, NGV735968H, NGV735971H, NGV735975H, NGV730668H, NGV735725H, NGV736645H, NGV738672H, NGV724010H, NGV724812H, NGV728809H, NGV728816H, NGV728834H, NGV729700H, NGV730642H, NGV730650H, NGV731938H, NGV731951H, NGV732014H, NGV738238H, NGV726026H, NGV726035H, NGV726036H, NGV726165H, NGV726171H, NGV726178H, NGV727136H, NGV727158H, NGV732040H, NGV733884H, NGV733885H, NGV733890H, NGV733895H, NGV734811H, NGV734813H, NGV729876H, NGV730683H, NGV730690H, NGV731401H, NGV733377H, NGV734652H, NGV725931H, NGV732565H, NGV730661H, NGV730679H, NGV739451H, NGV731961H, NGV724046H, NGV727130H, NGV727910H, NGV731973H, NGV731978H, NGV731994H, NGV732034H, NGV733372H, NGV733379H, NGV739494H, NGV739497H, NGV739498H, NGV739499H, NGV728202H, NGV732339H, NGV732360H, NGV727264H, NGV730627H, NGV732140H, NGV724022H, NGV724024H, NGV724028H, NGV727276H, NGV727339H, NGV727359H, NGV727365H, NGV727422H, NGV727958H, NGV727979H, NGV728253H, NGV729572H, NGV729588H, NGV729589H, NGV729597H, NGV729599H, NGV734643H, NGV734644H, NGV734646H, NGV739592H, NGV730666H, NGV734481H, NGV725956H, NGV725958H, NGV730079H, NGV730081H, NGV730082H, NGV732712H, NGV738249H, NGV725514H, NGV726023H, NGV726024H, NGV726251H, NGV727313H, NGV728210H, NGV734256H, NGV734468H, NGV725599H, NGV725607H, NGV738669H, NGV725552H, NGV724587H, NGV734823H, NGV732742H, NGV725272H, NGV726334H, NGV728257H, NGV728757H, NGV729923H, NGV732107H, NGV737576H, NGV737587H, NGV737625H, NGV724834H, NGV725030H, NGV725083H, NGV725085H, NGV729539H, NGV729547H, NGV729550H, NGV734258H, NGV734259H, NGV735981H, NGV736114H, NGV736119H, NGV736121H, NGV736480H, NGV738658H, NGV738659H, NGV724811H, NGV725884H, NGV727284H, NGV728250H, NGV729915H, NGV732118H, NGV719181H, NGV728944H, NGV726335H, NGV728910H, NGV733380H, NGV733383H, NGV733385H, NGV734782H, NGV734795H, NGV732042H, NGV732067H, NGV732074H, NGV736712H, NGV736715H, NGV736716H, NGV736741H, NGV736743H, NGV736760H, NGV736824H, NGV736856H, NGV737038H, NGV737089H, NGV737103H, NGV737107H, NGV737117H, NGV737119H, NGV737121H, NGV729908H, NGV730085H, NGV733546H; GTIN 00763000421847, Serial Numbers: NGV730040H, NGV731956H, NGV731965H, NGV731968H, NGV736392H, NGV738990H, NGV729581H, NGV729584H, NGV729628H, NGV732122H, NGV732129H, NGV732143H, NGV735355H, NGV735371H, NGV735373H, NGV735735H, NGV735743H, NGV738643H, NGV740069H, NGV740083H, NGV723910H, NGV724808H, NGV725149H, NGV729205H, NGV729209H, NGV729288H, NGV730035H, NGV730135H, NGV732569H, NGV732591H, NGV734203H, NGV735093H, NGV735777H, NGV736708H, NGV736714H, NGV738239H, NGV738293H, NGV738363H, NGV738366H, NGV739425H, NGV739586H, NGV740247H, NGV740252H, NGV740718H, NGV740723H, NGV724045H, NGV724072H, NGV724130H, NGV724131H, NGV724839H, NGV725023H, NGV725039H, NGV725067H, NGV732978H, NGV733108H, NGV733515H, NGV733520H, NGV733526H, NGV733539H, NGV734195H, NGV734821H, NGV734856H, NGV734862H, NGV734879H, NGV734921H, NGV736720H, NGV736734H, NGV736735H, NGV736739H, NGV736759H, NGV737546H, NGV738570H, NGV738596H, NGV738597H, NGV738642H, NGV738983H, NGV740321H, NGV740452H, NGV740482H, NGV740631H, NGV740640H, NGV740657H, NGV740702H, NGV727274H, NGV727760H, NGV727773H, NGV727871H, NGV727878H, NGV727882H, NGV727926H, NGV727978H, NGV727985H, NGV727998H, NGV728005H, NGV728007H, NGV730077H, NGV730083H, NGV731931H, NGV731964H, NGV731991H, NGV731997H, NGV732005H, NGV732011H, NGV732027H, NGV732032H.NGV732105H, NGV732139H, NGV733883H, NGV733899H, NGV733955H, NGV733956H, NGV733958H, NGV733973H, NGV734018H, NGV734028H, NGV734039H, NGV734051H, NGV734174H, NGV734181H, NGV734186H, NGV734227H, NGV736665H, NGV740264H, NGV740270H, NGV740287H, NGV740306H, NGV740307H, NGV740309H, NGV740310H, NGV740311H, NGV740314H, NGV740316H, NGV740318H, NGV740712H, NGV740726H, NGV724077H, NGV724104H, NGV724111H, NGV724161H, NGV724165H, NGV724185H, NGV724191H, NGV732365H, NGV732575H, NGV732584H, NGV732632H, NGV732647H, NGV732692H, NGV732700H, NGV732759H, NGV732761H, NGV739590H, NGV739597H, NGV739598H, NGV739602H, NGV740715H, NGV740716H, NGV730178H, NGV734324H, NGV735761H, NGV735766H, NGV734744H, NGV734751H, NGV735709H, NGV735721H, NGV735724H, NGV735727H, NGV735733H, NGV740133H, NGV740135H, NGV740142H, NGV740149H, NGV740155H, NGV740340H, NGV740352H, NGV740733H, NGV740735H, NGV729195H, NGV732029H, NGV732083H, NGV732111H, NGV738466H, NGV738562H, NGV724113H, NGV724290H, NGV724321H, NGV729961H, NGV730047H, NGV735772H, NGV729197H, NGV724041H, NGV727269H, NGV729302H, NGV731998H, NGV738568H, NGV727262H, NGV724312H, NGV725074H, NGV727271H, NGV730030H, NGV731981H, NGV740460H, NGV723971H, NGV723981H, NGV729510H, NGV732115H, NGV732181H, NGV738538H, NGV738557H, NGV740007H, NGV740017H, NGV740104H, NGV740257H, NGV740262H, NGV739070H, NGV739069H, NGV739071H, NGV735351H, NGV735349H, NGV739005H, NGV739020H, NGV729315H, NGV729318H, NGV731284H, NGV731298H, NGV729887H, NGV729891H, NGV729903H, NGV729939H, NGV729942H, NGV734221H, NGV736681H, NGV729631H, NGV722455H, NGV724119H, NGV724194H, NGV724342H, NGV725627H, NGV729291H, NGV729575H, NGV730048H, NGV730074H, NGV731987H, NGV735094H, NGV735413H, NGV735848H, NGV738373H, NGV738644H, NGV738650H, NGV738654H, NGV739040H, NGV739048H, NGV740242H, NGV729262H, NGV723700H, NGV724037H, NGV725633H, NGV728145H, NGV730124H, NGV732022H, NGV732577H, NGV734774H, NGV734926H, NGV735076H, NGV738464H, NGV727272H, NGV735086H, NGV740453H, NGV735713H, NGV735744H, NGV735641H, NGV731992H, NGV732012H, NGV732019H, NGV732021H, NGV729253H, NGV739403H, NGV739408H, NGV740000H, NGV729255H, NGV739060H, NGV739062H, NGV739063H, NGV739064H, NGV740737H, NGV739424H, NGV727173H, NGV727246H, NGV727247H, NGV727251H, NGV727253H, NGV727254H, NGV731247H, NGV731254H, NGV731257H, NGV731272H, NGV734936H, NGV736686H, NGV739036H, NGV728141H, NGV729198H, NGV729210H, NGV730018H, NGV733953H, NGV735399H, NGV735762H, NGV736409H, NGV738300H, NGV738317H, NGV738367H, NGV740216H, NGV740711H, NGV732043H, NGV732056H, NGV738227H, NGV738231H, NGV723897H, NGV724949H, NGV725008H, NGV725129H, NGV727270H, NGV729265H, NGV735396H, NGV735832H, NGV736579H, NGV736719H, NGV738483H, NGV738495H, NGV738979H, NGV740127H, NGV724214H, NGV730100H, NGV723727H, NGV725132H, NGV730194H, NGV732363H, NGV732364H, NGV738973H, NGV724123H, NGV724160H, NGV724255H, NGV724299H, NGV724748H, NGV724772H, NGV724803H, NGV729271H, NGV732568H, NGV732701H, NGV732743H, NGV733905H, NGV734502H, NGV735760H, NGV736417H, NGV736673H, NGV740317H, NGV740705H, NGV724054H, NGV729289H, NGV732156H, NGV735406H, NGV735768H, NGV738465H, NGV740481H, NGV731318H, NGV732145H, NGV736669H, NGV728163H, NGV730049H, NGV730060H, NGV736420H, NGV740710H, NGV738664H, NGV725143H, NGV728012H, NGV728184H, NGV738996H, NGV739888H, NGV740630H, NGV735418H, NGV738368H, NGV739021H, NGV739023H, NGV739026H, NGV739049H, NGV739053H, NGV739054H, NGV740320H, NGV738837H, NGV740243H, NGV740503H, NGV737615H, NGV738217H, NGV735584H, NGV735638H, NGV738826H, NGV739908H, NGV739911H, NGV739920H, NGV739926H, NGV740059H, NGV740089H, NGV740322H, NGV740326H, NGV740346H, NGV740465H, NGV740477H, NGV740498H, NGV740329H, NGV740510H, NGV739673H, NGV739684H, NGV735657H, NGV736675H, NGV739038H, NGV734928H, NGV740653H, NGV735352H, NGV735397H, NGV735804H, NGV735815H, NGV736701H, NGV739875H, NGV739999H, NGV740341H, NGV736842H, NGV737212H; GTIN 00763000421854, Serial Numbers: NGV725251H, NGV725257H, NGV725648H, NGV725650H, NGV725666H, NGV729928H, NGV732680H, NGV732132H, NGV734416H, NGV729929H; GTIN 00763000421861, Serial Numbers: NGV729479H, NGV734846H, NGV735365H, NGV739306H, NGV736656H, NGV724084H, NGV730641H, NGV730644H, NGV733898H, NGV735387H, NGV740458H, NGV740461H, NGV723652H, NGV724021H, NGV730741H, NGV723771H, NGV723986H, NGV729565H, NGV730648H, NGV730649H, NGV734076H, NGV734090H, NGV734100H, NGV734803H, NGV734849H, NGV734851H, NGV734868H, NGV735374H, NGV736683H, NGV736689H, NGV736772H, NGV739294H, NGV739295H, NGV739884H, NGV739917H, NGV739923H, NGV740071H, NGV740152H, NGV730034H, NGV730039H, NGV724246H, NGV729527H, NGV740835H, NGV734404H, NGV734787H, NGV735595H, NGV729549H, NGV729568H, NGV730599H, NGV732146H, NGV735582H, NGV739314H, NGV730063H, NGV731972H, NGV731990H, NGV734571H, NGV729499H, NGV730730H, NGV734124H, NGV724032H, NGV733901H, NGV733977H, NGV734269H, NGV724039H, NGV729492H, NGV729515H, NGV734401H, NGV735377H, NGV739288H, NGV739880H, NGV740816H, NGV740837H, NGV724049H, NGV724076H, NGV730740H, NGV732051H, NGV732137H, NGV734790H, NGV736654H, NGV736677H, NGV736775H, NGV739299H, NGV739853H, NGV740462H, NGV740472H, NGV731979H, NGV734331H, NGV730055H, NGV732069H, NGV734161H, NGV735579H, NGV735581H, NGV739397H, NGV731967H, NGV736667H, NGV739293H, NGV739847H, NGV740476H; GTIN 00763000421878, Serial Numbers: NGV727782H, NGV727873H, NGV727879H, NGV727883H; GTIN 00763000421885, Serial Numbers: NGV740261H, NGV740263H, NGV740266H, NGV740267H, NGV740269H, NGV740339H, NGV732753H, NGV732747H, NGV734071H, NGV734078H, NGV729204H, NGV729862H, NGV734055H, NGV734066H, NGV734068H, NGV734069H, NGV740169H, NGV729585H, NGV731935H, NGV729576H, NGV734806H, NGV735307H, NGV740207H, NGV740214H, NGV740220H, NGV728252H, NGV740205H, NGV740231H, NGV740234H, NGV740237H, NGV727920H, NGV727923H, NGV728167H, NGV728242H, NGV728243H, NGV727859H, NGV727870H, NGV734842H, NGV724007H, NGV740075H, NGV740080H, NGV740085H, NGV740167H, NGV732594H, NGV740016H, NGV740148H, NGV727184H, NGV725646H, NGV736694H, NGV726446H, NGV727141H, NGV727160H, NGV727165H, NGV727182H, NGV727240H, NGV727273H, NGV729932H, NGV729936H, NGV732739H, NGV734012H, NGV734120H, NGV734797H, NGV734820H, NGV736134H, NGV736135H, NGV736723H, NGV739845H, NGV740108H, NGV740132H, NGV740186H, NGV740239H, NGV740245H, NGV740259H, NGV733990H, NGV740101H, NGV740103H, NGV725243H, NGV726168H, NGV726336H, NGV736685H, NGV740057H, NGV740137H, NGV740195H, NGV740260H, NGV732045H, NGV728095H, NGV732695H, NGV736138H, NGV736661H, NGV740054H, NGV740061H, NGV740130H, NGV740197H, NGV734796H, NGV725651H, NGV726344H, NGV735357H, NGV736139H, NGV736707H, NGV739893H, NGV732642H, NGV740184H, NGV727286H, NGV735342H, NGV733992H, NGV736745H, NGV732671H, NGV740181H, NGV736728H, NGV739850H, NGV739882H, NGV724224H, NGV734098H, NGV724215H, NGV729570H, NGV729583H, NGV729626H, NGV729917H, NGV739943H, NGV740058H, NGV725238H, NGV723892H, NGV729580H, NGV729591H, NGV729641H, NGV729931H, NGV734117H, NGV736733H, NGV739941H, NGV740062H, NGV740063H, NGV740258H, NGV734123H, NGV740066H, NGV731936H, NGV728154H, NGV728350H, NGV728745H, NGV728361H, NGV729864H, NGV728354H, NGV728756H, NGV728259H, NGV728806H, NGV728750H, NGV729642H, NGV729664H, NGV729947H, NGV730029H, NGV734075H, NGV734083H, NGV734093H, NGV734096H, NGV740067H, NGV740068H, NGV740227H; GTIN 00763000421892, Serial Numbers: NGV740173H, NGV730717H, NGV723582H, NGV723616H, NGV723643H, NGV723738H, NGV723985H, NGV724002H, NGV724026H, NGV724067H, NGV724075H, NGV724085H, NGV724151H, NGV724171H, NGV724200H, NGV724218H, NGV730023H, NGV730044H, NGV734052H, NGV734061H, NGV734113H, NGV734215H, NGV734223H, NGV734261H, NGV734507H, NGV734515H, NGV734518H, NGV734525H, NGV734527H, NGV734575H, NGV734747H, NGV734762H, NGV734763H, NGV734767H, NGV734783H, NGV734793H, NGV735786H, NGV735801H, NGV735814H, NGV735817H, NGV735818H, NGV735822H, NGV735830H, NGV735849H, NGV736596H, NGV736598H, NGV736600H, NGV736644H, NGV736687H, NGV736691H, NGV736737H, NGV740084H, NGV740086H, NGV740088H, NGV740134H, NGV740140H, NGV740141H, NGV740151H, NGV740154H, NGV740159H, NGV740175H, NGV740185H, NGV740191H, NGV740194H, NGV740208H, NGV740224H, NGV740228H, NGV740233H, NGV740328H, NGV740331H, NGV740334H, NGV740367H, NGV740459H, NGV740492H, NGV740502H, NGV740659H, NGV740708H, NGV740713H, NGV740724H, NGV740748H, NGV740762H, NGV740145H, NGV740147H, NGV724483H, NGV724190H, NGV724204H, NGV724234H, NGV724236H, NGV724237H, NGV724353H, NGV724605H, NGV724609H, NGV724613H, NGV724683H, NGV724693H, NGV730062H, NGV730068H, NGV730070H, NGV731952H, NGV731960H, NGV732038H, NGV732774H, NGV733906H, NGV733946H, NGV733966H, NGV736688H, NGV736692H, NGV736693H, NGV736731H, NGV736747H, NGV736751H, NGV736766H, NGV736834H, NGV737453H, NGV737474H, NGV737479H, NGV740143H, NGV730046H, NGV730141H, NGV730142H, NGV730148H, NGV730187H, NGV730621H, NGV734122H, NGV739855H, NGV739887H, NGV739936H; GTIN 00763000421908, Serial Numbers: NGV725954H, NGV726228H, NGV727354H, NGV730716H, NGV734857H, NGV725557H, NGV731983H, NGV725871H, NGV726161H, NGV729890H, NGV732020H, NGV740065H, NGV732076H, NGV725922H, NGV725660H, NGV728177H, NGV729133H, NGV729893H, NGV732017H, NGV732062H, NGV732080H, NGV732347H, NGV732767H, NGV734831H, NGV740136H, NGV740158H, NGV725260H, NGV725344H, NGV725352H, NGV725354H, NGV725566H, NGV725620H, NGV725690H, NGV725860H, NGV725885H, NGV726155H, NGV727211H, NGV727349H, NGV727360H, NGV728180H, NGV728188H, NGV729879H, NGV729886H, NGV730365H, NGV730392H, NGV732015H, NGV732055H, NGV732070H, NGV732133H, NGV732336H, NGV732351H, NGV734901H, NGV740144H, NGV725546H, NGV725677H, NGV725681H, NGV729314H, NGV730696H, NGV732052H, NGV726349H, NGV730067H, NGV729907H, NGV729919H, NGV729926H, NGV731308H, NGV731933H, NGV731941H, NGV731958H, NGV732049H, NGV732054H, NGV732057H, NGV729309H, NGV729750H, NGV731970H, NGV731977H, NGV734414H, NGV734817H, NGV740355H, NGV741143H, NGV728014H, NGV728023H, NGV729154H, NGV734254H, NGV729555H, NGV741384H, NGV741158H, NGV734350H, NGV734398H, NGV741395H, NGV732772H; GTIN 00763000421922, Serial Numbers: NGV734262H, NGV734338H, NGV741340H, NGV729299H, NGV734245H, NGV734247H, NGV734341H, NGV734411H, NGV737449H, NGV737471H, NGV737478H, NGV737560H, NGV739798H, NGV740324H, NGV740325H, NGV734184H, NGV740342H, NGV729544H, NGV737840H, NGV737865H, NGV737866H, NGV739683H, NGV739692H, NGV739799H, NGV737466H, NGV737875H, NGV740815H, NGV741417H, NGV741593H, NGV727902H, NGV728011H, NGV734252H, NGV737860H, NGV739595H, NGV740357H, NGV740841H, NGV739589H, NGV739594H, NGV740165H, NGV740313H, NGV741330H, NGV734255H, NGV736553H, NGV739685H, NGV739688H, NGV740222H, NGV741328H, NGV741334H, NGV741413H, NGV741416H, NGV741567H, NGV741576H, NGV741585H, NGV741590H, NGV734217H, NGV739621H, NGV739672H, NGV740336H, NGV727980H, NGV727982H, NGV729596H, NGV737586H, NGV737839H, NGV737863H, NGV739693H, NGV739775H, NGV740330H, NGV740333H, NGV740337H, NGV741440H, NGV740210H, NGV741434H, NGV740297H, NGV727905H, NGV728002H, NGV729870H, NGV738649H, NGV740179H, NGV740300H, NGV740301H, NGV741319H, NGV741320H, NGV741324H, NGV741390H, NGV741418H, NGV741419H, NGV741450H, NGV734399H, NGV739679H, NGV734046H, NGV734220H, NGV740356H, NGV741185H, NGV729863H, NGV729971H, NGV734200H, NGV738671H, NGV739596H, NGV739604H, NGV740275H, NGV740296H, NGV739591H, NGV740162H, NGV737876H, NGV740235H, NGV729970H, NGV734214H, NGV740319H, NGV740282H, NGV740286H, NGV734277H, NGV740249H, NGV741427H, NGV741443H, NGV736560H, NGV737565H, NGV740268H, NGV740294H, NGV741362H, NGV740315H, NGV739661H, NGV739609H, NGV739610H, NGV739614H, NGV740277H, NGV740376H, NGV740178H, NGV740199H, NGV740302H, NGV740817H, NGV727903H, NGV739612H, NGV734219H, NGV737556H, NGV734189H, NGV740174H, NGV740345H, NGV739682H, NGV740265H, NGV739618H, NGV739619H, NGV741446H, NGV740312H, NGV739667H, NGV739613H, NGV727901H, NGV729295H, NGV734396H, NGV734403H, NGV739675H, NGV739777H, NGV739783H, NGV740226H, NGV740276H, NGV741396H, NGV741402H, NGV741407H, NGV740323H, NGV741421H; GTIN 00763000422615, Serial Numbers: NGV728427H, NGV728428H, NGV728601H, NGV728603H, NGV728604H, NGV728605H, NGV728606H, NGV728607H, NGV728609H.NGV728610H, NGV728611H, NGV728613H, NGV728614H, NGV728615H, NGV728616H, NGV728617H, NGV729247H, NGV729248H, NGV729249H, NGV729250H, NGV729251H, NGV729252H, NGV729254H, NGV729256H, NGV729257H, NGV729258H, NGV729260H, NGV729261H, NGV729263H, NGV729264H, NGV729266H, NGV729267H, NGV729268H, NGV729269H, NGV729675H, NGV729676H, NGV729678H, NGV729679H, NGV729683H, NGV729684H, NGV729685H, NGV729686H, NGV729689H, NGV730366H, NGV730368H, NGV730369H, NGV730370H, NGV730371H, NGV730372H, NGV730373H, NGV730374H, NGV730375H, NGV730376H, NGV730377H, NGV730378H, NGV730379H, NGV730380H, NGV730382H, NGV731901H, NGV731902H, NGV731903H, NGV731904H, NGV731905H, NGV731906H, NGV731907H, NGV731908H, NGV731909H, NGV731911H, NGV731914H, NGV731915H, NGV731916H, NGV731917H, NGV731918H, NGV731919H, NGV731920H, NGV731923H, NGV731924H, NGV731925H, NGV731926H, NGV731928H, NGV731929H, NGV731930H, NGV731984H, NGV732084H, NGV732090H, NGV732144H, NGV733263H, NGV733264H, NGV733265H, NGV733266H, NGV733267H, NGV733268H, NGV733269H, NGV733590H, NGV733939H, NGV733941H, NGV733942H, NGV733979H, NGV733980H, NGV733981H, NGV733982H, NGV733984H, NGV733985H, NGV733986H, NGV733987H, NGV734198H, NGV734199H, NGV734915H, NGV734929H, NGV734938H, NGV734941H, NGV734943H, NGV734945H, NGV734946H, NGV734960H, NGV734961H, NGV734964H, NGV734965H, NGV734966H, NGV734967H, NGV734995H, NGV734997H, NGV734998H, NGV734999H, NGV736018H, NGV736019H, NGV736020H, NGV736021H, NGV736022H, NGV736368H, NGV736370H, NGV736371H, NGV736374H, NGV736376H, NGV736377H, NGV736702H, NGV736703H, NGV736704H, NGV736705H, NGV736706H, NGV736709H, NGV736844H, NGV736845H, NGV736846H, NGV736847H, NGV736848H, NGV736849H, NGV736850H, NGV736851H, NGV736852H, NGV736853H, NGV736854H, NGV736855H, NGV736857H, NGV736858H, NGV736859H, NGV736860H, NGV737826H, NGV737827H, NGV737828H, NGV737829H, NGV737830H, NGV737831H, NGV737832H, NGV737833H, NGV737834H, NGV737835H, NGV737837H, NGV737838H, NGV737841H, NGV737842H, NGV737843H, NGV737844H, NGV737845H, NGV737846H, NGV737847H, NGV737849H, NGV737850H, NGV737851H, NGV737852H, NGV737853H, NGV737855H, NGV737856H, NGV737858H, NGV737862H, NGV739387H, NGV739388H, NGV739389H, NGV739390H, NGV739391H, NGV739392H, NGV739393H, NGV739394H, NGV739395H, NGV739471H, NGV739504H, NGV739505H, NGV739506H, NGV739507H, NGV739508H, NGV739509H, NGV739510H, NGV739511H, NGV739512H, NGV739513H, NGV739514H, NGV739515H, NGV739516H, NGV739517H, NGV739518H, NGV739519H, NGV739520H, NGV739521H, NGV739522H, NGV739523H, NGV739524H, NGV739525H, NGV739526H, NGV739527H, NGV739528H, NGV739529H, NGV739530H, NGV739531H, NGV739532H, NGV739533H, NGV739534H, NGV739536H, NGV739580H, NGV740574H, NGV740575H, NGV740576H, NGV740577H, NGV740578H, NGV740579H, NGV740580H, NGV740581H, NGV740582H, NGV740583H, NGV740935H, NGV740936H, NGV740937H, NGV740938H, NGV740939H, NGV740940H, NGV740941H, NGV740942H, NGV740943H, NGV741314H, NGV741315H, NGV741317H, NGV741318H, NGV741321H, NGV741322H, NGV741353H, NGV741355H, NGV741356H, NGV741357H, NGV741358H, NGV741359H, NGV741360H, NGV741361H, NGV741363H, NGV741364H, NGV741365H, NGV741366H, NGV741367H, NGV741368H, NGV741369H, NGV741370H, NGV741371H, NGV741372H, NGV741373H, NGV741374H, NGV741375H, NGV741376H, NGV741378H, NGV741379H, NGV741380H, NGV741389H, NGV741391H, NGV741397H, NGV741404H, NGV741415H, NGV741431H, NGV741437H, NGV741441H, NGV741452H, NGV740944H, NGV739964H, NGV741765H, NGV741766H, NGV730089H, NGV730343H, NGV737086H, NGV730951H, NGV732390H, NGV732484H, NGV733167H, NGV735900H, NGV736334H, NGV731580H, NGV735616H, NGV735861H, NGV729115H, NGV733216H, NGV727822H, NGV730889H, NGV727817H, NGV728796H, NGV729345H, NGV733271H, NGV735881H, NGV736223H, NGV728921H, NGV729653H, NGV729719H, NGV729720H, NGV729721H, NGV730482H, NGV730483H, NGV732305H, NGV732306H, NGV732307H, NGV736239H, NGV737139H, NGV728742H, NGV732553H, NGV732936H, NGV728838H, NGV730329H, NGV730519H, NGV730754H, NGV730831H, NGV732481H, NGV726907H, NGV729844H, NGV730451H, NGV730795H, NGV730802H, NGV730803H, NGV731570H, NGV732429H, NGV730927H, NGV728513H, NGV736341H, NGV728640H, NGV730273H, NGV732169H, NGV737229H, NGV731870H, NGV734454H, NGV729330H, NGV732241H, NGV727195H, NGV729339H, NGV729412H, NGV736198H, NGV736809H, NGV728460H, NGV728699H, NGV729321H, NGV731699H, NGV732482H, NGV732798H, NGV732912H, NGV733151H, NGV733651H, NGV733681H, NGV733755H, NGV735347H, NGV736254H, NGV736505H, NGV736822H, NGV736825H, NGV736888H, NGV736889H, NGV736926H, NGV736978H, NGV737019H, NGV729058H, NGV732595H, NGV732827H, NGV735477H, NGV735908H, NGV729399H, NGV732281H, NGV728555H, NGV728680H, NGV730115H, NGV730746H, NGV731879H, NGV728836H, NGV730413H, NGV730459H, NGV732846H, NGV733417H, NGV734349H, NGV736318H, NGV736528H, NGV736529H, NGV737203H, NGV728649H, NGV729526H, NGV732935H, NGV733750H, NGV736298H, NGV735450H, NGV728779H, NGV733074H, NGV734871H, NGV729355H, NGV733048H, NGV735717H, NGV728692H, NGV730752H, NGV731875H, NGV733650H, NGV734335H, NGV734449H, NGV734542H, NGV730772H, NGV732243H, NGV733768H, NGV732402H, NGV733586H, NGV733871H, NGV728711H, NGV727820H, NGV728229H, NGV732218H, NGV735298H, NGV735420H, NGV736873H, NGV728423H, NGV730295H, NGV732406H, NGV736817H, NGV733081H, NGV732262H, NGV731750H, NGV732538H, NGV736623H, NGV729016H, NGV733222H, NGV735096H, NGV730274H, NGV730328H, NGV730333H, NGV730426H, NGV731329H, NGV732835H, NGV735068H, NGV735550H, NGV736496H, NGV727945H, NGV728851H, NGV729159H, NGV730750H, NGV730973H, NGV731333H, NGV731446H, NGV731545H, NGV732315H, NGV732600H, NGV733283H, NGV733284H, NGV735062H, NGV735073H, NGV730431H, NGV728664H, NGV729029H, NGV729622H, NGV734828H, NGV736793H, NGV728880H, NGV736630H, NGV735931H, NGV736165H, NGV737221H, NGV728311H, NGV728557H, NGV732393H, NGV732558H, NGV735169H, NGV735925H, NGV728357H, NGV729858H, NGV731181H, NGV736250H, NGV734337H, NGV729471H, NGV730243H, NGV730952H, NGV730978H, NGV731412H, NGV732164H, NGV732453H, NGV733598H, NGV735412H, NGV735939H, NGV736196H, NGV736197H, NGV727208H, NGV728558H, NGV728729H, NGV730506H, NGV732475H, NGV734443H, NGV734625H, NGV735926H, NGV735927H, NGV736333H, NGV736784H, NGV729743H, NGV729744H, NGV730276H, NGV730277H, NGV730278H, NGV730279H, NGV731067H, NGV731069H, NGV731070H, NGV731071H, NGV734292H, NGV734293H, NGV734640H, NGV735946H, NGV735947H, NGV737067H, NGV737068H, NGV730347H, NGV732474H, NGV727396H, NGV730974H, NGV735157H, NGV736048H, NGV729457H, NGV729812H, NGV729813H, NGV729814H, NGV729815H, NGV729816H, NGV731463H, NGV731464H, NGV731628H, NGV731744H, NGV733272H, NGV728572H, NGV729125H, NGV729740H, NGV729741H, NGV730287H, NGV730288H, NGV732864H, NGV733927H, NGV729541H, NGV730827H, NGV731850H, NGV732516H, NGV733626H, NGV734031H, NGV736698H, NGV730200H, NGV732439H, NGV732440H, NGV732441H, NGV732449H, NGV732630H, NGV736870H, NGV729980H, NGV731747H, NGV732554H, NGV733814H, NGV734419H, NGV734621H, NGV735965H, NGV736972H, NGV737077H, NGV729402H, NGV729496H, NGV733115H, NGV731396H, NGV733032H, NGV734956H, NGV730945H, NGV735335H, NGV736046H, NGV736208H, NGV736350H, NGV736352H, NGV736565H, NGV727170H, NGV728566H, NGV728667H, NGV731503H, NGV733571H, NGV734300H, NGV735203H, NGV736101H, NGV737240H, NGV729139H, NGV730598H, NGV732229H, NGV729319H, NGV733721H, NGV730346H, NGV731109H, NGV727156H, NGV728705H, NGV729386H, NGV729624H, NGV730284H, NGV730591H, NGV731079H, NGV732514H, NGV732609H, NGV733076H, NGV733839H, NGV734913H, NGV736222H, NGV736234H, NGV736617H, NGV736957H, NGV736963H, NGV728792H, NGV730967H, NGV729415H, NGV729422H, NGV735497H, NGV735913H, NGV726700H, NGV726715H, NGV728324H, NGV728326H, NGV728453H, NGV729035H, NGV729103H, NGV729104H, NGV729105H, NGV729106H, NGV732269H, NGV733157H, NGV734368H, NGV734412H, NGV734555H, NGV735379H, NGV735451H, NGV735844H, NGV737026H, NGV728435H, NGV730900H, NGV731691H, NGV731692H, NGV734717H, NGV736765H, NGV731293H, NGV733199H, NGV727287H, NGV730745H, NGV732986H, NGV728730H, NGV728738H, NGV728739H, NGV729014H, NGV729797H, NGV729799H, NGV729800H, NGV729801H, NGV729802H, NGV730957H, NGV731029H, NGV732989H, NGV732990H, NGV733179H, NGV733180H, NGV733295H, NGV728352H, NGV728895H, NGV728896H, NGV729456H, NGV729514H, NGV729790H, NGV729983H, NGV729984H, NGV730447H, NGV730554H, NGV731160H, NGV731612H, NGV732876H, NGV732950H, NGV733043H, NGV733352H, NGV733359H, NGV734003H, NGV734306H, NGV734308H, NGV734582H, NGV734584H, NGV735007H, NGV735008H, NGV735458H, NGV737141H, NGV727823H, NGV728434H, NGV729046H, NGV729134H, NGV729276H, NGV732231H, NGV732288H, NGV732503H, NGV734606H, NGV735175H, NGV735176H, NGV735178H, NGV735179H, NGV730970H, NGV731611H, NGV734728H, NGV728456H, NGV728500H, NGV730771H, NGV732395H, NGV729341H, NGV730899H, NGV732480H, NGV735074H, NGV730539H, NGV730844H, NGV731667H, NGV733566H, NGV735032H, NGV736003H, NGV730520H, NGV731802H, NGV730813H, NGV732916H, NGV733690H, NGV734554H, NGV735492H, NGV735837H, NGV735998H, NGV727025H, NGV727092H, NGV728100H, NGV728338H, NGV728340H, NGV728341H, NGV728716H, NGV729072H, NGV730401H, NGV730477H, NGV730565H, NGV730890H, NGV730891H, NGV731290H, NGV731291H, NGV731444H, NGV731486H, NGV731536H, NGV731652H, NGV731653H, NGV732249H, NGV732960H, NGV733061H, NGV733099H, NGV733119H, NGV733148H, NGV733154H, NGV733177H, NGV733181H, NGV733304H, NGV733784H, NGV734147H, NGV734334H, NGV735781H, NGV735785H, NGV735929H, NGV736268H, NGV736506H, NGV736950H, NGV730395H, NGV735246H, NGV731152H, NGV736190H, NGV728782H, NGV727842H, NGV728393H, NGV729020H, NGV729851H, NGV731017H, NGV732870H, NGV733818H, NGV731033H, NGV734725H, NGV728892H, NGV730556H, NGV731149H, NGV731368H, NGV733153H, NGV735874H, NGV730972H, NGV728508H, NGV730097H, NGV736767H, NGV730448H, NGV733613H, NGV729439H, NGV735958H, NGV735982H, NGV736219H, NGV736252H, NGV736297H, NGV736300H, NGV729822H, NGV730429H, NGV731018H, NGV732324H, NGV732325H, NGV732874H, NGV732875H, NGV733007H, NGV733770H, NGV734539H, NGV727571H, NGV728835H, NGV728843H, NGV728844H, NGV729152H, NGV729317H, NGV730116H, NGV730341H, NGV730920H, NGV731746H, NGV731847H, NGV732159H, NGV732211H, NGV732274H, NGV732460H, NGV732618H, NGV732969H, NGV733275H, NGV733794H, NGV733797H, NGV733913H, NGV736066H, NGV736206H, NGV736773H, NGV736886H, NGV736887H, NGV736938H, NGV737218H, NGV737252H, NGV728888H, NGV736193H, NGV733811H, NGV729226H, NGV731283H, NGV732255H, NGV730212H, NGV731427H, NGV732293H, NGV732465H, NGV732624H, NGV733925H, NGV734313H, NGV735201H, NGV737072H, NGV727829H, NGV728315H, NGV729880H, NGV730239H, NGV730240H, NGV731447H, NGV731564H, NGV731900H, NGV733868H, NGV734172H, NGV735461H, NGV736885H, NGV736973H, NGV729334H, NGV731002H, NGV731585H, NGV734312H, NGV734356H, NGV737208H, NGV731345H, NGV732980H, NGV728279H, NGV728309H, NGV728335H, NGV728421H, NGV728713H, NGV728828H, NGV729409H, NGV729763H, NGV729764H, NGV729767H, NGV729768H, NGV729772H, NGV729773H, NGV730118H, NGV731066H, NGV731367H, NGV731397H, NGV731499H, NGV731515H, NGV731682H, NGV731743H, NGV731803H, NGV731873H, NGV732304H, NGV732494H, NGV732495H, NGV732496H, NGV732613H, NGV732833H, NGV732971H, NGV733006H, NGV733277H, NGV733289H, NGV733290H, NGV733323H, NGV733646H, NGV734212H, NGV734317H, NGV734319H, NGV734320H, NGV734425H, NGV734437H, NGV734444H, NGV734566H, NGV734630H, NGV734713H, NGV734716H, NGV734881H, NGV735029H, NGV735030H, NGV735257H, NGV735314H, NGV735315H, NGV735316H, NGV735465H, NGV735517H, NGV735518H, NGV735519H, NGV735546H, NGV736264H, NGV736273H, NGV736274H, NGV736357H, NGV736718H, NGV736725H, NGV727928H, NGV729726H, NGV730087H, NGV731006H, NGV735126H, NGV736365H, NGV737237H, NGV737238H, NGV728404H, NGV728876H, NGV734605H, NGV737140H, NGV737152H, NGV737199H, NGV728128H, NGV730235H, NGV731242H, NGV731320H, NGV732391H, NGV732427H, NGV733309H, NGV733319H, NGV735215H, NGV735453H, NGV736451H, NGV737154H, NGV731712H, NGV732845H, NGV728336H, NGV728994H, NGV729998H, NGV735826H, NGV728783H, NGV729229H, NGV729361H, NGV729118H, NGV733934H, NGV728668H, NGV731772H, NGV735867H, NGV728478H, NGV728547H, NGV728548H, NGV728549H, NGV728878H, NGV729062H, NGV730233H, NGV731443H, NGV731693H, NGV731694H, NGV731695H, NGV733155H, NGV733156H, NGV733558H, NGV733748H, NGV733850H.NGV733851H, NGV733869H, NGV735556H, NGV735767H, NGV735780H, NGV736034H, NGV736628H, NGV736937H, NGV737259H, NGV731877H, NGV735183H, NGV736011H, NGV736531H, NGV736761H, NGV730587H, NGV736541H, NGV736543H, NGV736041H, NGV736632H, NGV728464H, NGV729089H, NGV730903H, NGV730919H, NGV731148H, NGV732462H, NGV733139H, NGV733215H, NGV736999H, NGV728497H, NGV732531H, NGV732955H, NGV729189H, NGV729349H, NGV730845H, NGV730980H, NGV731482H, NGV733330H, NGV733789H, NGV734160H, NGV734229H, NGV734367H, NGV734914H, NGV735362H, NGV735479H, NGV728518H, NGV727549H, NGV728778H, NGV730873H, NGV736388H, NGV728317H, NGV731165H, NGV731347H, NGV727835H, NGV727837H, NGV727838H, NGV727839H, NGV727840H, NGV728657H, NGV728658H, NGV728660H, NGV729005H, NGV729008H, NGV729009H, NGV729011H, NGV729012H, NGV729362H, NGV729363H, NGV729364H, NGV729829H, NGV729830H, NGV729831H, NGV729833H, NGV729834H, NGV729835H, NGV729836H, NGV729838H, NGV729839H, NGV729840H, NGV731107H, NGV731108H, NGV731679H, NGV731759H, NGV731798H, NGV731832H, NGV731833H, NGV732799H, NGV732800H, NGV732801H, NGV732802H, NGV732803H, NGV732805H, NGV732806H, NGV732994H, NGV734287H, NGV734386H, NGV735036H, NGV735037H, NGV735301H, NGV735302H, NGV735558H, NGV735559H, NGV731488H, NGV732377H, NGV732383H, NGV734924H, NGV735498H, NGV730578H, NGV727825H, NGV728110H, NGV729051H, NGV729518H, NGV729519H, NGV729795H, NGV729796H, NGV730168H, NGV730306H, NGV730574H, NGV730822H, NGV730823H, NGV730896H, NGV731015H, NGV731360H, NGV731565H, NGV731660H, NGV731666H, NGV731827H, NGV733012H, NGV733143H, NGV733332H, NGV733732H, NGV734639H, NGV735097H, NGV735483H, NGV735854H, NGV735855H, NGV736248H, NGV736649H, NGV736650H, NGV731698H, NGV728305H, NGV728374H, NGV728543H, NGV728733H, NGV728934H, NGV730259H, NGV730568H, NGV730991H, NGV731056H, NGV731187H, NGV731867H, NGV732382H, NGV732411H, NGV732610H, NGV732866H, NGV732867H, NGV732868H, NGV733741H, NGV733762H, NGV733909H, NGV734143H, NGV734144H, NGV735152H, NGV735697H, NGV735833H, NGV736107H, NGV736112H, NGV736814H, NGV736832H, NGV736981H, NGV737060H, NGV729400H, NGV730480H, NGV731857H, NGV732242H, NGV732292H, NGV732442H, NGV732979H, NGV733000H, NGV733553H, NGV733695H, NGV734709H, NGV735528H, NGV736023H, NGV736062H, NGV729185H, NGV729737H, NGV730137H, NGV730357H, NGV731603H, NGV731649H, NGV732443H, NGV732444H, NGV733210H, NGV733256H, NGV733257H, NGV733356H, NGV733563H, NGV733591H, NGV734721H, NGV734722H, NGV734986H, NGV736440H, NGV729440H, NGV731606H, NGV731704H, NGV734677H, NGV736986H, NGV728395H, NGV729405H, NGV730514H, NGV730569H, NGV730867H, NGV731809H, NGV733054H, NGV733607H, NGV733822H, NGV730785H, NGV727163H, NGV728461H, NGV728979H, NGV728982H, NGV733066H, NGV733567H, NGV733761H, NGV735122H, NGV735124H, NGV735125H, NGV735819H, NGV735820H, NGV735824H, NGV736229H, NGV728400H, NGV729218H, NGV730757H, NGV735416H, NGV735502H, NGV731793H, NGV735587H, NGV731596H, NGV731362H, NGV731423H, NGV735190H, NGV728871H, NGV729336H, NGV729451H, NGV734615H, NGV735158H, NGV729136H, NGV728999H, NGV730892H, NGV732216H, NGV731797H, NGV737137H, NGV728443H, NGV730144H, NGV730515H, NGV731375H, NGV731839H, NGV733830H, NGV730528H, NGV731683H, NGV733191H, NGV734618H, NGV735154H, NGV735217H, NGV735402H, NGV735424H, NGV736075H, NGV728698H, NGV736084H, NGV736263H, NGV730908H, NGV728525H, NGV734433H, NGV732194H, NGV733639H, NGV735360H, NGV733726H, NGV735439H, NGV729123H, NGV729124H, NGV730415H, NGV731211H, NGV731212H, NGV732930H, NGV733713H, NGV735119H, NGV730525H, NGV733190H, NGV735909H, NGV732180H, NGV733611H, NGV736061H, NGV733346H, NGV734438H, NGV729036H, NGV729086H, NGV729665H, NGV731796H, NGV737230H, NGV729213H, NGV730961H, NGV734612H, NGV735226H, NGV734900H, NGV728704H, NGV730006H, NGV731800H, NGV735054H, NGV727180H, NGV729141H, NGV734947H, NGV733322H, NGV735536H, NGV729214H, NGV729215H, NGV727035H, NGV728302H, NGV729856H, NGV729985H, NGV730175H, NGV730199H, NGV731019H, NGV731120H, NGV731176H, NGV731215H, NGV732528H, NGV734459H, NGV730180H, NGV728953H, NGV730850H, NGV730861H, NGV732312H, NGV733258H, NGV736054H, NGV736077H, NGV736078H, NGV730389H, NGV732272H, NGV732550H, NGV733253H, NGV735480H, NGV736880H, NGV730966H, NGV731363H, NGV732837H, NGV733642H, NGV733727H, NGV734984H, NGV735918H, NGV736217H, NGV737016H, NGV728902H, NGV731632H, NGV731633H, NGV732184H, NGV732253H, NGV732505H, NGV734286H, NGV737255H, NGV729828H, NGV733641H, NGV734560H, NGV736092H, NGV737195H, NGV736231H, NGV728559H, NGV729217H, NGV729610H, NGV730510H, NGV731100H, NGV731117H, NGV731656H, NGV733928H, NGV734976H, NGV735221H, NGV730460H, NGV731009H, NGV731715H, NGV731795H, NGV732917H, NGV732975H, NGV733027H, NGV733044H, NGV733045H, NGV733173H, NGV733307H, NGV733743H, NGV734456H, NGV734602H, NGV735296H, NGV735466H, NGV735467H, NGV736311H, NGV736805H, NGV727826H, NGV727827H, NGV728433H, NGV730387H, NGV732397H, NGV732398H, NGV729857H, NGV730523H, NGV733025H, NGV734138H, NGV736184H, NGV736358H, NGV736411H, NGV736413H, NGV737083H, NGV736289H, NGV727321H, NGV727929H, NGV727930H, NGV727931H, NGV728138H, NGV728237H, NGV728238H, NGV728239H, NGV728240H, NGV728271H, NGV728272H, NGV728273H, NGV728275H, NGV728276H, NGV728277H, NGV732821H, NGV732822H, NGV733661H, NGV733662H, NGV733663H, NGV733664H, NGV733666H, NGV733667H, NGV735560H, NGV735561H, NGV735563H, NGV735570H, NGV735993H, NGV729013H, NGV728553H, NGV728990H, NGV729367H, NGV729444H, NGV729650H, NGV729669H, NGV729708H, NGV730052H, NGV730579H, NGV730807H, NGV730887H, NGV730916H, NGV731302H, NGV731310H, NGV731576H, NGV731713H, NGV732209H, NGV732409H, NGV733018H, NGV733229H, NGV733238H, NGV733535H, NGV733564H, NGV733595H, NGV733623H, NGV733809H, NGV734598H, NGV734668H, NGV734680H, NGV734695H, NGV734864H, NGV735066H, NGV735236H, NGV735836H, NGV735839H, NGV735954H, NGV735990H, NGV736385H, NGV736386H, NGV736967H, NGV736968H, NGV737070H, NGV737227H, NGV729043H, NGV730428H, NGV730444H, NGV730590H, NGV731842H, NGV731843H, NGV732264H, NGV732534H, NGV733172H, NGV733201H, NGV733249H, NGV734604H, NGV735012H, NGV736179H, NGV728749H, NGV733753H, NGV730092H, NGV728232H, NGV728233H, NGV728234H, NGV728430H, NGV728689H, NGV731124H, NGV735956H, NGV729966H, NGV729945H, NGV730205H, NGV730907H, NGV731592H, NGV731593H, NGV732973H, NGV732974H, NGV734288H, NGV734290H, NGV737032H, NGV727527H, NGV729335H, NGV729337H, NGV731648H, NGV732219H, NGV732951H, NGV734322H, NGV736140H, NGV730796H, NGV729084H, NGV731599H, NGV732256H, NGV732267H, NGV736058H, NGV728903H, NGV730421H, NGV731388H, NGV734714H, NGV736293H, NGV728867H, NGV728898H, NGV729192H, NGV729375H, NGV730423H, NGV731709H, NGV731837H, NGV732419H, NGV733001H, NGV733280H, NGV733921H, NGV733930H, NGV733931H, NGV734137H, NGV734954H, NGV735620H, NGV736220H, NGV736360H, NGV736364H, NGV729788H, NGV737053H, NGV731162H, NGV731163H, NGV728623H, NGV730318H, NGV731054H, NGV732958H, NGV735924H, NGV737200H, NGV736094H, NGV735153H, NGV729783H, NGV728134H, NGV728135H, NGV728390H, NGV728391H, NGV728392H, NGV729855H, NGV730197H, NGV730457H, NGV730458H, NGV731055H, NGV731190H, NGV731193H, NGV731297H, NGV731369H, NGV731370H, NGV732433H, NGV732598H, NGV732599H, NGV732796H, NGV732952H, NGV732953H, NGV733192H, NGV733285H, NGV733678H, NGV735156H, NGV735524H, NGV735955H, NGV736068H, NGV736069H, NGV736275H, NGV737151H, NGV737247H, NGV737248H, NGV734552H, NGV736828H, NGV736881H, NGV735018H, NGV729002H, NGV730893H, NGV733867H, NGV728442H, NGV728712H, NGV728758H, NGV729778H, NGV729781H, NGV730246H, NGV731788H, NGV731789H, NGV733312H, NGV736621H, NGV734347H, NGV728837H, NGV734197H, NGV735883H, NGV730500H, NGV733051H, NGV733162H, NGV733239H, NGV733783H, NGV735966H, NGV736154H, NGV736615H, NGV737264H, NGV728120H, NGV733597H, NGV733866H, NGV735021H, NGV736976H, NGV727818H, NGV728132H, NGV728718H, NGV730184H, NGV731490H, NGV732279H, NGV732282H, NGV732375H, NGV733200H, NGV733920H, NGV733999H, NGV735209H, NGV735269H, NGV736004H, NGV736183H, NGV736191H, NGV736335H, NGV737217H, NGV736025H, NGV734139H, NGV737204H, NGV731438H, NGV730437H, NGV733130H, NGV731012H, NGV731782H, NGV733152H, NGV730320H, NGV731799H, NGV733303H, NGV733305H, NGV733604H, NGV733698H, NGV734847H, NGV735038H, NGV735223H, NGV735791H, NGV735813H, NGV737158H, NGV727093H, NGV728445H, NGV731512H, NGV728408H, NGV730582H, NGV730760H, NGV736362H, NGV728511H, NGV733339H, NGV728121H, NGV728824H, NGV728825H, NGV735219H, NGV728079H, NGV728230H, NGV728520H, NGV729049H, NGV729146H, NGV729148H, NGV729162H, NGV730153H, NGV730340H, NGV730588H, NGV731014H, NGV731341H, NGV731786H, NGV731818H, NGV732220H, NGV732232H, NGV732258H, NGV732299H, NGV732818H, NGV733011H, NGV733056H, NGV733131H, NGV733225H, NGV733250H, NGV733357H, NGV733643H, NGV733760H, NGV733864H, NGV734163H, NGV734835H, NGV734902H, NGV735149H, NGV735260H, NGV735343H, NGV735869H, NGV735992H, NGV736308H, NGV736366H, NGV736438H, NGV736678H, NGV736928H, NGV728471H, NGV729326H, NGV729371H, NGV730442H, NGV730491H, NGV730492H, NGV730561H, NGV731687H, NGV732195H, NGV732526H, NGV732779H, NGV733202H, NGV733848H, NGV735936H, NGV735969H, NGV736085H, NGV728763H, NGV728765H, NGV729883H, NGV729958H, NGV730193H, NGV730195H, NGV731166H, NGV731681H, NGV731706H, NGV731707H, NGV733588H, NGV736036H, NGV736944H, NGV737250H, NGV730782H, NGV731876H, NGV727362H, NGV727377H, NGV729219H, NGV729635H, NGV732956H, NGV733710H, NGV734291H, NGV736867H, NGV736868H, NGV727832H, NGV728367H, NGV728439H, NGV728527H, NGV736331H, NGV736769H, NGV736031H, NGV736838H, NGV731636H, NGV733016H, NGV733037H, NGV733606H, NGV735345H, NGV727401H, NGV728226H, NGV730004H, NGV730129H, NGV730190H, NGV733849H, NGV735232H, NGV735333H, NGV736532H, NGV737039H, NGV732949H, NGV733738H, NGV737028H, NGV734968H, NGV731740H, NGV736384H, NGV736770H, NGV728850H, NGV729429H, NGV729433H, NGV729435H, NGV730575H, NGV730576H, NGV731197H, NGV732522H, NGV732809H, NGV732929H, NGV733005H, NGV733084H, NGV734898H, NGV735019H, NGV728879H, NGV730435H, NGV732830H, NGV732831H, NGV732856H, NGV734724H, NGV728133H, NGV728472H, NGV728740H, NGV729057H, NGV731374H, NGV733123H, NGV730394H, NGV730493H, NGV730538H, NGV732326H, NGV733343H, NGV736884H, NGV732228H, NGV733654H, NGV736524H, NGV730390H, NGV730344H, NGV733237H, NGV733776H, NGV729522H, NGV729523H, NGV736307H, NGV728426H, NGV728562H, NGV728589H, NGV728637H, NGV728847H, NGV729174H, NGV730311H, NGV730573H, NGV734442H, NGV735120H, NGV736433H, NGV736450H, NGV728938H, NGV729784H, NGV730774H, NGV730775H, NGV731031H, NGV731523H, NGV731524H, NGV731525H, NGV735133H, NGV735134H, NGV735452H, NGV736348H, NGV729015H, NGV734434H, NGV734601H, NGV737059H, NGV730417H, NGV731065H, NGV732730H, NGV732910H, NGV733733H, NGV734027H, NGV734861H, NGV736103H, NGV736782H, NGV735860H, NGV736936H, NGV728220H, NGV728869H, NGV729408H, NGV732879H, NGV733235H, NGV733608H, NGV733686H, NGV733827H, NGV733862H, NGV733863H, NGV734826H, NGV735576H, NGV728312H, NGV728720H, NGV730997H, NGV731097H, NGV732620H, NGV732621H, NGV732622H, NGV732623H, NGV735330H, NGV735622H, NGV735878H, NGV735879H, NGV735880H, NGV736249H, NGV736980H, NGV737214H, NGV728753H, NGV729476H, NGV730282H, NGV730529H, NGV731337H, NGV731352H, NGV731807H, NGV732784H, NGV732945H, NGV732999H, NGV733826H, NGV733911H, NGV734387H, NGV735987H, NGV729087H, NGV729088H, NGV729739H, NGV729745H, NGV731630H, NGV731663H, NGV732861H, NGV732862H, NGV733059H, NGV733310H, NGV733804H, NGV734949H, NGV735318H, NGV735319H, NGV736163H, NGV728514H, NGV728515H, NGV729092H, NGV729173H, NGV729333H, NGV729459H, NGV731090H, NGV731670H, NGV731822H, NGV732555H, NGV733195H, NGV733742H, NGV735916H, NGV736095H, NGV736130H, NGV736989H, NGV737080H, NGV737206H, NGV737235H, NGV728485H, NGV728985H, NGV728986H, NGV730213H, NGV730214H, NGV730462H, NGV730463H, NGV730465H, NGV730793H, NGV731172H, NGV731173H, NGV732176H, NGV732178H, NGV732179H, NGV732421H, NGV733163H, NGV733231H, NGV733270H, NGV733712H, NGV734067H, NGV734070H, NGV734085H, NGV735248H, NGV735250H, NGV736510H, NGV736512H, NGV737061H, NGV737062H, NGV732201H, NGV732285H, NGV735866H, NGV728416H, NGV728707H, NGV728795H, NGV733122H, NGV733345H, NGV735003H, NGV737222H, NGV729898H, NGV730999H, NGV733092H, NGV733093H.NGV733160H, NGV733587H, NGV728642H, NGV731076H, NGV732625H, NGV733705H, NGV733932H, NGV735191H, NGV733685H, NGV730327H, NGV735606H, NGV731638H, NGV726660H, NGV728726H, NGV728920H, NGV729168H, NGV730937H, NGV730938H, NGV735234H, NGV736634H, NGV736826H, NGV735448H, NGV731377H, NGV731478H, NGV735320H, NGV735425H, NGV728877H, NGV727324H, NGV727779H, NGV728552H, NGV729310H, NGV730986H, NGV731688H, NGV731689H, NGV731690H, NGV732287H, NGV732836H, NGV734340H, NGV734342H, NGV735150H, NGV736266H, NGV736267H, NGV736610H, NGV736611H, NGV736965H, NGV737063H, NGV737064H, NGV737065H, NGV737066H, NGV729747H, NGV730201H, NGV730355H, NGV731383H, NGV731574H, NGV731815H, NGV732832H, NGV733224H, NGV734302H, NGV736332H, NGV736435H, NGV737001H, NGV728914H, NGV729078H, NGV729503H, NGV729693H, NGV729695H, NGV733853H, NGV733854H, NGV733855H, NGV735289H, NGV735293H, NGV735294H, NGV735469H, NGV735470H, NGV735471H, NGV735472H, NGV727848H, NGV731677H, NGV728564H, NGV728639H, NGV729059H, NGV729236H, NGV729354H, NGV729413H, NGV729878H, NGV730088H, NGV730326H, NGV731169H, NGV731170H, NGV731202H, NGV731522H, NGV731794H, NGV732149H, NGV732174H, NGV732175H, NGV732226H, NGV732227H, NGV732276H, NGV732551H, NGV732995H, NGV732996H, NGV732997H, NGV733227H, NGV733561H, NGV734145H, NGV734884H, NGV734990H, NGV735058H, NGV735067H, NGV735129H, NGV735235H, NGV735793H, NGV735934H, NGV736124H, NGV736241H, NGV736242H, NGV736429H, NGV736459H, NGV737029H, NGV737095H, NGV734049H, NGV736007H, NGV728463H, NGV728496H, NGV728563H, NGV728721H, NGV729187H, NGV729389H, NGV730114H, NGV730486H, NGV730537H, NGV731312H, NGV731313H, NGV731402H, NGV731431H, NGV731760H, NGV732277H, NGV732502H, NGV732556H, NGV732797H, NGV732872H, NGV733861H, NGV734607H, NGV735464H, NGV735798H, NGV737054H, NGV729270H, NGV730011H, NGV730888H, NGV732459H, NGV733142H, NGV734980H, NGV728743H, NGV728987H, NGV729760H, NGV731752H, NGV731754H, NGV733622H, NGV733842H, NGV728387H, NGV728477H, NGV729754H, NGV729981H, NGV729982H, NGV731379H, NGV731458H, NGV731758H, NGV731836H, NGV736255H, NGV729826H, NGV729827H, NGV730503H, NGV730505H, NGV730564H, NGV730969H, NGV731240H, NGV731471H, NGV732378H, NGV734563H, NGV735501H, NGV736309H, NGV728573H, NGV728574H, NGV728575H, NGV728576H, NGV728577H, NGV728578H, NGV728579H, NGV728580H, NGV728581H, NGV728582H, NGV728583H, NGV728584H, NGV728585H, NGV728586H, NGV728587H, NGV729241H, NGV729242H, NGV729244H, NGV729245H, NGV729294H, NGV729296H, NGV729297H, NGV729298H, NGV729303H, NGV729304H, NGV729305H, NGV729306H, NGV730555H, NGV730597H, NGV731036H, NGV734014H, NGV734887H, NGV734888H, NGV734889H, NGV734891H, NGV734892H, NGV735138H, NGV735139H, NGV735140H, NGV735141H, NGV736096H, NGV736097H, NGV736141H, NGV736142H, NGV736143H, NGV736144H, NGV736145H, NGV736516H, NGV736517H, NGV736518H, NGV736519H, NGV736520H, NGV736521H, NGV736571H, NGV736584H, NGV736606H, NGV732204H, NGV732205H, NGV732781H, NGV732962H, NGV733718H, NGV734635H, NGV735039H, NGV735063H, NGV735227H, NGV735468H, NGV735890H, NGV728875H, NGV731763H, NGV734125H, NGV734339H, NGV736214H, NGV732549H, NGV728466H, NGV729443H, NGV729464H, NGV729465H, NGV729466H, NGV729467H, NGV729468H, NGV729469H, NGV731756H, NGV732543H, NGV732544H, NGV732545H, NGV732546H, NGV732547H, NGV734315H, NGV735432H, NGV735433H, NGV735434H, NGV735435H, NGV735436H, NGV736379H, NGV729038H, NGV729182H, NGV732485H, NGV734149H, NGV735259H, NGV735868H, NGV728422H, NGV727745H, NGV727748H, NGV727751H, NGV727753H, NGV727759H, NGV727762H, NGV727765H, NGV727766H, NGV727770H, NGV732415H, NGV732416H, NGV732417H, NGV732418H, NGV732239H, NGV732518H, NGV732628H, NGV732629H, NGV733601H, NGV736076H, NGV737153H, NGV728410H, NGV729332H, NGV734893H, NGV735180H, NGV735186H, NGV735917H, NGV736613H, NGV730009H, NGV731049H, NGV735055H, NGV729394H, NGV729395H, NGV731537H, NGV731538H, NGV728222H, NGV729227H, NGV730780H, NGV730835H, NGV731497H, NGV731600H, NGV731625H, NGV731710H, NGV731764H, NGV732268H, NGV733779H, NGV734140H, NGV734344H, NGV735280H, NGV735901H, NGV735902H, NGV736149H, NGV737092H, NGV729194H, NGV732407H, NGV733047H, NGV735288H, NGV736748H, NGV735487H, NGV730811H, NGV731825H, NGV730790H, NGV734010H, NGV736863H, NGV737138H, NGV728686H, NGV730169H, NGV732423H, NGV735865H, NGV736230H, NGV731131H, NGV731132H, NGV731137H, NGV733682H, NGV733683H, NGV734698H, NGV734699H, NGV734701H, NGV734702H, NGV734703H, NGV734704H, NGV735700H, NGV736865H, NGV737216H, NGV728118H, NGV728804H, NGV731346H, NGV732203H, NGV733593H, NGV735368H, NGV735539H, NGV736504H, NGV735991H, NGV736037H, NGV730360H, NGV730787H, NGV730471H, NGV730472H, NGV730473H, NGV734202H, NGV736072H, NGV736073H, NGV736074H, NGV729738H, NGV735540H, NGV735904H, NGV731540H, NGV736930H, NGV736982H, NGV737254H, NGV728399H, NGV729042H, NGV729325H, NGV729825H, NGV729848H, NGV730002H, NGV730433H, NGV734682H, NGV729482H, NGV729485H, NGV729487H, NGV732498H, NGV730924H, NGV731052H, NGV730122H, NGV730236H, NGV728313H, NGV729442H, NGV729792H, NGV729992H, NGV729993H, NGV729995H, NGV730150H, NGV730188H, NGV730397H, NGV730446H, NGV730474H, NGV731004H, NGV731232H, NGV731672H, NGV731739H, NGV732422H, NGV732548H, NGV732931H, NGV732932H, NGV732933H, NGV732984H, NGV733058H, NGV733146H, NGV733297H, NGV733715H, NGV733716H, NGV733766H, NGV733774H, NGV733846H, NGV734963H, NGV735127H, NGV735346H, NGV735621H, NGV736626H, NGV737128H, NGV727546H, NGV733720H, NGV734693H, NGV729201H, NGV733584H, NGV735189H, NGV735974H, NGV736970H, NGV737178H, NGV728516H, NGV729246H, NGV729843H, NGV731657H, NGV732210H, NGV732607H, NGV733069H, NGV733070H, NGV733708H, NGV736939H, NGV736940H, NGV737198H, NGV728346H, NGV728918H, NGV729025H, NGV729161H, NGV730455H, NGV730456H, NGV731209H, NGV731210H, NGV731678H, NGV732233H, NGV732388H, NGV733194H, NGV733311H, NGV733337H, NGV733825H, NGV735366H, NGV729895H, NGV730165H, NGV730913H, NGV732420H, NGV734875H, NGV727816H, NGV728631H, NGV730419H, NGV730420H, NGV730755H, NGV730794H, NGV731146H, NGV731389H, NGV731745H, NGV731791H, NGV731792H, NGV733723H, NGV733724H, NGV735025H, NGV735028H, NGV736261H, NGV736292H, NGV736301H, NGV737020H, NGV737244H, NGV730857H, NGV737205H, NGV729661H, NGV732265H, NGV734304H, NGV735594H, NGV728107H, NGV728405H, NGV728891H, NGV729039H, NGV729040H, NGV729163H, NGV731498H, NGV731546H, NGV731685H, NGV732476H, NGV735174H, NGV736192H, NGV736195H, NGV736961H, NGV737084H, NGV732627H, NGV731701H, NGV732961H, NGV735784H, NGV732283H, NGV728491H, NGV728526H, NGV728761H, NGV731382H, NGV731631H, NGV733636H, NGV736614H, NGV736964H, NGV730443H, NGV732532H, NGV733694H, NGV736945H, NGV732448H, NGV734295H, NGV736147H, NGV730988H, NGV736811H, NGV731037H, NGV732948H, NGV736150H, NGV728394H, NGV728444H, NGV728862H, NGV730219H, NGV730232H, NGV731058H, NGV731059H, NGV731060H, NGV731093H, NGV731441H, NGV731584H, NGV731605H, NGV731817H, NGV732509H, NGV732510H, NGV732511H, NGV733734H, NGV733736H, NGV734309H, NGV734439H, NGV734440H, NGV734441H, NGV734613H, NGV736537H, NGV736786H, NGV731028H, NGV736819H, NGV728109H, NGV734723H, NGV730998H, NGV734543H, NGV728332H, NGV730091H, NGV732399H, NGV726489H, NGV728624H, NGV730909H, NGV733318H, NGV733633H, NGV735011H, NGV736162H, NGV736337H, NGV732859H, NGV732860H, NGV732982H, NGV728780H, NGV730580H, NGV731544H, NGV731771H, NGV731882H, NGV733254H, NGV734114H, NGV734328H, NGV735001H, NGV734970H, NGV728638H, NGV732428H, NGV732838H, NGV733242H, NGV727382H, NGV727398H, NGV728565H, NGV728936H, NGV731808H, NGV732379H, NGV732710H, NGV734191H, NGV734421H, NGV735064H, NGV735222H, NGV736088H, NGV736089H, NGV736090H, NGV732403H, NGV737109H, NGV729095H, NGV730542H, NGV730939H, NGV731647H, NGV735411H, NGV735440H, NGV729065H, NGV729066H, NGV731138H, NGV731145H, NGV732221H, NGV732222H, NGV733116H, NGV733117H, NGV735530H, NGV736559H, NGV736562H, NGV736563H, NGV729120H, NGV734561H, NGV732259H, NGV732469H, NGV733186H, NGV733995H, NGV735272H, NGV735601H, NGV736315H, NGV737210H, NGV726663H, NGV736305H, NGV729193H, NGV729438H, NGV736211H, NGV736306H, NGV729421H, NGV732869H, NGV727934H, NGV728567H, NGV728568H, NGV728569H, NGV728570H, NGV729698H, NGV729703H, NGV729704H, NGV729706H, NGV729896H, NGV730848H, NGV730849H, NGV731074H, NGV731094H, NGV731095H, NGV731096H, NGV731161H, NGV731327H, NGV731869H, NGV732389H, NGV732914H, NGV733821H, NGV734669H, NGV734670H, NGV734880H, NGV734886H, NGV735404H, NGV735999H, NGV736000H, NGV736771H, NGV737018H, NGV737069H, NGV728734H, NGV728735H, NGV729484H, NGV729497H, NGV729501H, NGV730289H, NGV731371H, NGV731372H, NGV731373H, NGV732851H, NGV732852H, NGV732853H, NGV733614H, NGV733615H, NGV733616H, NGV736277H, NGV736280H, NGV736281H, NGV736282H, NGV731430H, NGV731583H, NGV729237H, NGV730895H, NGV732291H, NGV732819H, NGV733138H, NGV735284H, NGV735285H, NGV729391H, NGV730812H, NGV731613H, NGV732466H, NGV734920H, NGV734948H, NGV735478H, NGV735489H, NGV735490H, NGV736815H, NGV731509H, NGV728528H, NGV730593H, NGV732552H, NGV736313H, NGV728318H, NGV728560H, NGV729990H, NGV730768H, NGV731640H, NGV731641H, NGV732400H, NGV732824H, NGV732850H, NGV733124H, NGV733126H, NGV733845H, NGV735959H, NGV736185H, NGV736342H, NGV731614H, NGV728591H, NGV728592H, NGV729954H, NGV730007H, NGV730220H, NGV731834H, NGV732263H, NGV732500H, NGV732777H, NGV732778H, NGV732883H, NGV732904H, NGV733166H, NGV733816H, NGV733929H, NGV734729H, NGV735604H, NGV735923H, NGV735941H, NGV736016H, NGV736045H, NGV733679H, NGV734536H, NGV729850H, NGV730010H, NGV734297H, NGV730806H, NGV733569H, NGV730789H, NGV730914H, NGV735358H, NGV736319H, NGV736530H, NGV730315H, NGV730955H, NGV732616H, NGV733565H, NGV734150H, NGV729817H, NGV733703H, NGV733704H, NGV735441H, NGV728725H, NGV728727H, NGV730489H, NGV731322H, NGV733135H, NGV733136H, NGV735304H, NGV735896H, NGV737021H, NGV728457H, NGV730476H, NGV731475H, NGV731761H, NGV735310H, NGV736523H, NGV727437H, NGV728599H, NGV729989H, NGV731642H, NGV732825H, NGV732251H, NGV735231H, NGV728429H, NGV729728H, NGV729729H, NGV729730H, NGV729731H, NGV732376H, NGV732530H, NGV735415H, NGV735510H, NGV735511H, NGV728418H, NGV729994H, NGV733714H, NGV733773H, NGV735142H, NGV735143H, NGV728571H, NGV728717H, NGV728845H, NGV732215H, NGV732217H, NGV732410H, NGV733127H, NGV733129H, NGV733638H, NGV733790H, NGV733907H, NGV734735H, NGV730761H, NGV733072H, NGV734413H, NGV727579H, NGV735482H, NGV736742H, NGV729516H, NGV730751H, NGV734346H, NGV736338H, NGV736012H, NGV729432H, NGV732858H, NGV733198H, NGV728594H, NGV728127H, NGV728236H, NGV728793H, NGV729950H, NGV733050H, NGV733178H, NGV735943H, NGV736956H, NGV728777H, NGV729528H, NGV729530H, NGV730104H, NGV730105H, NGV729231H, NGV729234H, NGV729358H, NGV729359H, NGV731156H, NGV731159H, NGV731378H, NGV732943H, NGV732944H, NGV733149H, NGV734420H, NGV735035H, NGV735295H, NGV735626H, NGV735807H, NGV736087H, NGV736452H, NGV736570H, NGV736808H, NGV736959H, NGV728316H, NGV730763H, NGV730764H, NGV730870H, NGV732431H, NGV733226H, NGV733924H, NGV729847H, NGV731616H, NGV731680H, NGV733432H, NGV734818H, NGV736540H, NGV732863H, NGV728227H, NGV731476H, NGV729099H, NGV731639H, NGV736527H, NGV736797H, NGV729404H, NGV731013H, NGV731348H, NGV737057H, NGV734689H, NGV733302H, NGV729393H, NGV730833H, NGV730834H, NGV732604H, NGV732785H, NGV729129H, NGV730337H, NGV731485H, NGV735071H, NGV731048H, NGV733039H, NGV734667H, NGV735002H, NGV735485H, NGV735624H, NGV728493H, NGV729312H, NGV731582H, NGV734151H, NGV726760H, NGV728087H, NGV728480H, NGV728714H, NGV728741H, NGV729132H, NGV729996H, NGV730408H, NGV731674H, NGV731811H, NGV731816H, NGV732278H, NGV732854H, NGV733575H, NGV733631H, NGV733806H, NGV734637H, NGV735092H, NGV735202H, NGV735609H, NGV735937H, NGV736533H, NGV736535H, NGV736536H, NGV733042H, NGV728839H, NGV729538H, NGV732471H, NGV732899H, NGV734636H, NGV734671H, NGV734697H, NGV737079H, NGV728106H, NGV730305H, NGV731092H, NGV735825H, NGV736941H, NGV729398H, NGV732826H, NGV727933H, NGV729397H, NGV729449H, NGV729953H, NGV730410H, NGV730411H, NGV731150H, NGV731445H, NGV731495H, NGV731535H, NGV731844H, NGV732197H, NGV732499H, NGV733355H, NGV733785H, NGV733786H, NGV733787H, NGV734141H, NGV734142H, NGV735462H, NGV735557H, NGV736813H, NGV730203H.NGV731098H, NGV731129H, NGV731394H, NGV733079H, NGV733245H, NGV735332H, NGV729110H, NGV729824H, NGV732492H, NGV735163H, NGV728364H, NGV729712H, NGV729818H, NGV735121H, NGV726943H, NGV730910H, NGV730990H, NGV735022H, NGV728407H, NGV730454H, NGV732834H, NGV734397H, NGV735216H, NGV736204H, NGV728554H, NGV731838H, NGV732311H, NGV732517H, NGV735508H, NGV736079H, NGV737013H, NGV730102H, NGV729384H, NGV731123H, NGV732271H, NGV735237H, NGV736216H, NGV730098H, NGV732185H, NGV734688H, NGV728827H, NGV735430H, NGV735393H, NGV729017H, NGV729018H, NGV730571H, NGV731417H, NGV732373H, NGV733692H, NGV733693H, NGV737130H, NGV731474H, NGV735687H, NGV727537H, NGV729055H, NGV730586H, NGV730921H, NGV730984H, NGV731064H, NGV731664H, NGV731665H, NGV731810H, NGV732165H, NGV733687H, NGV734294H, NGV735829H, NGV736640H, NGV728308H, NGV736501H, NGV731437H, NGV730481H, NGV730928H, NGV731700H, NGV733014H, NGV733313H, NGV734359H, NGV735782H, NGV736507H, NGV736509H, NGV728685H, NGV727850H, NGV727924H, NGV727927H, NGV729636H, NGV729645H, NGV729647H, NGV734579H, NGV734581H, NGV734622H, NGV734623H, NGV727828H, NGV728939H, NGV729222H, NGV729673H, NGV730388H, NGV730468H, NGV731115H, NGV731321H, NGV732540H, NGV732541H, NGV732542H, NGV736245H, NGV736246H, NGV736247H, NGV736990H, NGV736991H, NGV728129H, NGV728467H, NGV729392H, NGV730322H, NGV731561H, NGV731609H, NGV732521H, NGV733281H, NGV733282H, NGV734952H, NGV735305H, NGV736322H, NGV736324H, NGV736380H, NGV736555H, NGV736762H, NGV728432H, NGV733833H, NGV732493H, NGV733038H, NGV736818H, NGV730502H, NGV733029H, NGV728378H, NGV728784H, NGV731270H, NGV731338H, NGV733605H, NGV733348H, NGV733063H, NGV733064H, NGV733065H, NGV732972H, NGV734568H, NGV732615H, NGV728451H, NGV733700H, NGV735015H, NGV730513H, NGV728544H, NGV729107H, NGV729108H, NGV729109H, NGV733120H, NGV733596H, NGV733810H, NGV735070H, NGV736161H, NGV737037H, NGV737228H, NGV732557H, NGV736934H, NGV729648H, NGV731711H, NGV729365H, NGV730307H, NGV731737H, NGV736049H, NGV736432H, NGV728655H, NGV729842H, NGV731871H, NGV732157H, NGV733647H, NGV736008H, NGV728899H, NGV734450H, NGV736205H, NGV726252H, NGV728420H, NGV728446H, NGV728882H, NGV733659H, NGV735600H, NGV736984H, NGV728448H, NGV730182H, NGV734314H, NGV730154H, NGV730155H, NGV730159H, NGV730490H, NGV730783H, NGV730784H, NGV734906H, NGV735919H, NGV736355H, NGV736927H, NGV737201H, NGV737202H, NGV732405H, NGV728894H, NGV730014H, NGV731841H, NGV733009H, NGV733799H, NGV728386H, NGV728870H, NGV728873H, NGV729027H, NGV729135H, NGV729952H, NGV730786H, NGV731385H, NGV732425H, NGV732486H, NGV732487H, NGV733087H, NGV733360H, NGV733652H, NGV733831H, NGV733918H, NGV733935H, NGV734135H, NGV735787H, NGV735914H, NGV736071H, NGV736839H, NGV736943H, NGV737017H, NGV737234H, NGV731696H, NGV736099H, NGV733648H, NGV728334H, NGV728675H, NGV730466H, NGV730959H, NGV730983H, NGV731874H, NGV732434H, NGV733188H, NGV733189H, NGV733214H, NGV735306H, NGV735308H, NGV735457H, NGV728375H, NGV728966H, NGV730119H, NGV733801H, NGV734329H, NGV735214H, NGV736237H, NGV736243H, NGV736258H, NGV736284H, NGV727293H, NGV726680H, NGV730836H, NGV731089H, NGV732322H, NGV732323H, NGV736240H, NGV736618H, NGV736791H, NGV736792H, NGV731621H, NGV728868H, NGV730198H, NGV731105H, NGV731121H, NGV736070H, NGV736361H, NGV737124H, NGV728682H, NGV728684H, NGV728900H, NGV730467H, NGV730851H, NGV730852H, NGV730859H, NGV731619H, NGV731675H, NGV731676H, NGV732236H, NGV733599H, NGV733602H, NGV734994H, NGV736050H, NGV736091H, NGV736966H, NGV731846H, NGV733834H, NGV735828H, NGV735853H, NGV728343H, NGV730162H, NGV731263H, NGV732954H, NGV733637H, NGV735414H, NGV736883H, NGV729122H, NGV733730H, NGV735198H, NGV736295H, NGV736627H, NGV728437H, NGV733347H, NGV737002H, NGV728762H, NGV734730H, NGV735155H, NGV736056H, NGV731627H, NGV731806H, NGV734629H, NGV733689H, NGV729275H, NGV733294H, NGV734281H, NGV731753H, NGV733185H, NGV733823H, NGV732888H, NGV736768H, NGV736958H, NGV727215H, NGV731022H, NGV735617H, NGV735618H, NGV736740H, NGV728388H, NGV728389H, NGV730313H, NGV731835H, NGV734923H, NGV736962H, NGV729821H, NGV730947H, NGV737246H, NGV728849H, NGV729717H, NGV730949H, NGV731384H, NGV732467H, NGV732468H, NGV734593H, NGV735057H, NGV735291H, NGV736233H, NGV736287H, NGV736354H, NGV728122H, NGV730323H, NGV730324H, NGV732199H, NGV732200H, NGV733326H, NGV735299H, NGV736200H, NGV734004H, NGV734005H, NGV736796H, NGV728703H, NGV728709H, NGV727099H, NGV727830H, NGV727831H, NGV730293H, NGV731590H, NGV732880H, NGV733688H, NGV735168H, NGV736534H, NGV737197H, NGV729322H, NGV730260H, NGV731080H, NGV731376H, NGV731821H, NGV734537H, NGV735795H, NGV736042H, NGV736538H, NGV730319H, NGV732907H, NGV734812H, NGV731452H, NGV734991H, NGV733158H, NGV732202H, NGV728349H, NGV728874H, NGV728901H, NGV729853H, NGV730297H, NGV730485H, NGV730557H, NGV730871H, NGV731010H, NGV731650H, NGV732529H, NGV733133H, NGV733618H, NGV733737H, NGV733758H, NGV733759H, NGV734086H, NGV734095H, NGV734103H, NGV734166H, NGV734224H, NGV734230H, NGV734446H, NGV734457H, NGV734460H, NGV734534H, NGV734546H, NGV734551H, NGV734676H, NGV736044H, NGV736086H, NGV736180H, NGV736213H, NGV736473H, NGV737242H, NGV737243H, NGV728383H, NGV728424H, NGV731316H, NGV733922H, NGV731207H, NGV732946H, NGV728498H, NGV730302H, NGV731624H, NGV728298H, NGV728300H, NGV730012H, NGV731671H, NGV733144H, NGV733286H, NGV733287H, NGV733288H, NGV734383H, NGV734385H, NGV736514H, NGV728523H, NGV728524H, NGV728700H, NGV729048H, NGV729342H, NGV729488H, NGV729489H, NGV729490H, NGV730498H, NGV730566H, NGV730981H, NGV730982H, NGV731189H, NGV731233H, NGV731234H, NGV732294H, NGV732295H, NGV732297H, NGV732814H, NGV732815H, NGV732816H, NGV732817H, NGV733196H, NGV733728H, NGV733729H, NGV734128H, NGV734132H, NGV735431H, NGV735625H, NGV735789H, NGV735875H, NGV735930H, NGV735989H, NGV736344H, NGV736755H, NGV736756H, NGV736975H, NGV737012H, NGV730854H, NGV736010H, NGV736158H, NGV728507H, NGV729376H, NGV729406H, NGV729085H, NGV731456H, NGV735281H, NGV729533H, NGV729534H, NGV731749H, NGV732537H, NGV735845H, NGV736971H, NGV733711H, NGV728431H, NGV730202H, NGV730841H, NGV734715H, NGV735437H, NGV727393H, NGV727746H, NGV728328H, NGV728329H, NGV728330H, NGV728331H, NGV729221H, NGV730962H, NGV730963H, NGV730964H, NGV731465H, NGV731467H, NGV732380H, NGV732381H, NGV733341H, NGV733342H, NGV734426H, NGV734427H, NGV734959H, NGV734969H, NGV734977H, NGV734978H, NGV734981H, NGV736441H, NGV736442H, NGV736446H, NGV736447H, NGV737006H, NGV737007H, NGV737009H, NGV728084H, NGV728085H, NGV728678H, NGV728679H, NGV730308H, NGV730309H, NGV734591H, NGV735218H, NGV735961H, NGV729733H, NGV733031H, NGV735252H, NGV728961H, NGV730283H, NGV732871H, NGV736102H, NGV728691H, NGV731626H, NGV736057H, NGV729789H, NGV730106H, NGV733572H, NGV729311H, NGV729377H, NGV727794H, NGV732506H, NGV731502H, NGV733077H, NGV733233H, NGV734570H, NGV730440H, NGV733274H, NGV733329H, NGV733534H, NGV736328H, NGV736746H, NGV731393H, NGV732913H, NGV734451H, NGV734927H, NGV728278H, NGV728280H, NGV728701H, NGV730570H, NGV732266H, NGV732432H, NGV732886H, NGV733193H, NGV733325H, NGV733335H, NGV735359H, NGV735514H, NGV736336H, NGV736547H, NGV736758H, NGV736977H, NGV736985H, NGV737125H, NGV727581H, NGV728306H, NGV728380H, NGV728672H, NGV729128H, NGV729327H, NGV729607H, NGV729742H, NGV730531H, NGV731399H, NGV733278H, NGV733751H, NGV735442H, NGV735456H, NGV735486H, NGV735619H, NGV734681H, NGV728681H, NGV736017H, NGV728590H, NGV729437H, NGV729955H, NGV729956H, NGV733046H, NGV734436H, NGV735792H, NGV736882H, NGV728595H, NGV730814H, NGV731315H, NGV729474H, NGV728645H, NGV729670H, NGV730439H, NGV730577H, NGV731588H, NGV732189H, NGV733248H, NGV733619H, NGV736285H, NGV730452H, NGV730930H, NGV729033H, NGV730338H, NGV730946H, NGV731032H, NGV731294H, NGV728747H, NGV730917H, NGV733197H, NGV736820H, NGV736821H, NGV728411H, NGV730166H, NGV730210H, NGV730298H, NGV730470H, NGV730801H, NGV730839H, NGV730985H, NGV731114H, NGV732424H, NGV732426H, NGV732602H, NGV732603H, NGV733211H, NGV734417H, NGV734431H, NGV735159H, NGV735207H, NGV735311H, NGV735317H, NGV736080H, NGV736081H, NGV736228H, NGV736276H, NGV736777H, NGV737075H, NGV729176H, NGV729177H, NGV729178H, NGV729179H, NGV732240H, NGV732301H, NGV733919H, NGV735553H, NGV735554H, NGV735555H, NGV730918H, NGV736572H, NGV733028H, NGV730191H, NGV732865H, NGV733209H, NGV734616H, NGV728223H, NGV728224H, NGV729978H, NGV733813H, NGV735027H, NGV727200H, NGV728474H, NGV728909H, NGV731053H, NGV735960H, NGV736343H, NGV730497H, NGV731714H, NGV734696H, NGV735297H, NGV736014H, NGV736637H, NGV736194H, NGV728484H, NGV730979H, NGV736550H, NGV733725H, NGV728618H, NGV735173H, NGV729649H, NGV731601H, NGV733279H, NGV733771H, NGV735484H, NGV736929H, NGV728358H, NGV728632H, NGV728633H, NGV731078H, NGV732970H, NGV733132H, NGV733812H, NGV734388H, NGV735527H, NGV735953H, NGV729346H, NGV731781H, NGV733856H, NGV733857H, NGV735877H, NGV734989H, NGV729385H, NGV728111H, NGV728114H, NGV728881H, NGV728887H, NGV729453H, NGV730960H, NGV731116H, NGV732611H, NGV732786H, NGV732903H, NGV732992H, NGV733040H, NGV734305H, NGV734719H, NGV736027H, NGV736201H, NGV736202H, NGV736931H, NGV736932H, NGV736933H, NGV727013H, NGV728551H, NGV730971H, NGV735033H, NGV728788H, NGV728790H, NGV730965H, NGV731460H, NGV732244H, NGV732246H, NGV732247H, NGV733168H, NGV734673H, NGV734674H, NGV734675H, NGV735516H, NGV732515H, NGV729216H, NGV729785H, NGV732257H, NGV732284H, NGV732719H, NGV730386H, NGV735607H, NGV730121H, NGV733640H, NGV728794H, NGV729282H, NGV731151H, NGV731489H, NGV734238H, NGV729382H, NGV729660H, NGV730317H, NGV735282H, NGV735014H, NGV736827H, NGV730393H, NGV733057H, NGV730223H, NGV732520H, NGV736795H, NGV728889H, NGV735213H, NGV731566H, NGV732214H, NGV729751H, NGV729752H, NGV731773H, NGV736326H, NGV733697H, NGV735322H, NGV735493H, NGV736251H, NGV737225H, NGV735263H, NGV736166H, NGV734146H, NGV736872H, NGV732882H, NGV732937H, NGV736225H, NGV728056H, NGV733363H, NGV730383H, NGV735938H, NGV728212H, NGV728643H, NGV731034H, NGV731534H, NGV733585H, NGV734678H, NGV734679H, NGV734710H, NGV734957H, NGV736156H, NGV731112H, NGV730522H, NGV734672H, NGV733696H, NGV734614H, NGV728468H, NGV731519H, NGV732261H, NGV734855H, NGV735889H, NGV736840H, NGV732412H, NGV733912H, NGV735894H, NGV736757H, NGV735611H, NGV728775H, NGV733019H, NGV736879H, NGV737257H, NGV734429H, NGV727556H, NGV729396H, NGV735340H, NGV737131H, NGV728626H, NGV728627H, NGV728628H, NGV728630H, NGV731083H, NGV731085H, NGV731086H, NGV731087H, NGV731765H, NGV731768H, NGV732303H, NGV735225H, NGV733668H, NGV735172H, NGV728728H, NGV729475H, NGV730441H, NGV733260H, NGV735224H, NGV735044H, NGV733752H, NGV735707H, NGV736131H, NGV731770H, NGV731304H, NGV732959H, NGV735463H, NGV734634H, NGV732394H, NGV734170H, NGV729313H, NGV729353H, NGV729390H, NGV730749H, NGV731548H, NGV731654H, NGV733624H, NGV734726H, NGV735047H, NGV735171H, NGV735211H, NGV736974H, NGV728912H, NGV729116H, NGV729117H, NGV730874H, NGV732497H, NGV732501H, NGV737245H, NGV728521H, NGV731118H, NGV733351H, NGV729277H, NGV732842H, NGV732844H, NGV732967H, NGV736067H, NGV736381H, NGV731686H, NGV733298H, NGV733782H, NGV734836H, NGV729032H, NGV730262H, NGV730264H, NGV731102H, NGV731103H, NGV731104H, NGV731106H, NGV733175H, NGV733176H, NGV727935H, NGV727936H, NGV727940H, NGV727942H, NGV728883H, NGV728884H, NGV728886H, NGV729153H, NGV729156H, NGV729723H, NGV729724H, NGV729727H, NGV729999H, NGV730432H, NGV730560H, NGV730756H, NGV730905H, NGV731858H, NGV731859H, NGV731860H, NGV731861H, NGV731862H, NGV731863H, NGV731864H, NGV731865H, NGV731866H, NGV733217H, NGV733219H, NGV733221H, NGV733574H, NGV733672H, NGV734243H, NGV734244H, NGV734282H, NGV734283H, NGV734620H, NGV734899H, NGV734907H, NGV734908H, NGV734909H, NGV734911H, NGV734912H, NGV735165H, NGV735167H, NGV735270H, NGV727947H, NGV727950H, NGV727957H, NGV727988H, NGV728746H, NGV729143H, NGV729144H, NGV731353H, NGV731354H, NGV731355H, NGV731356H, NGV731359H, NGV732316H, NGV732317H, NGV732318H, NGV732319H, NGV734374H, NGV734375H, NGV734376H, NGV735077H, NGV735078H, NGV735079H, NGV735080H, NGV735081H, NGV735083H, NGV735084H, NGV732508H, NGV735251H, NGV730237H.NGV733621H, NGV734333H, NGV736502H, NGV737135H, NGV729629H, NGV735608H, NGV733749H, NGV728113H, NGV728971H, NGV728972H, NGV728974H, NGV728975H, NGV728976H, NGV729191H, NGV729548H, NGV729551H, NGV730225H, NGV730226H, NGV730227H, NGV730228H, NGV730230H, NGV730824H, NGV730826H, NGV730828H, NGV731183H, NGV731848H, NGV731849H, NGV731851H, NGV731852H, NGV731853H, NGV731854H, NGV731856H, NGV733075H, NGV733137H, NGV733204H, NGV733205H, NGV733206H, NGV733207H, NGV733208H, NGV733562H, NGV733625H, NGV733628H, NGV734032H, NGV734036H, NGV734038H, NGV734041H, NGV734045H, NGV734353H, NGV735049H, NGV735050H, NGV735051H, NGV735052H, NGV735053H, NGV735631H, NGV735633H, NGV735634H, NGV735635H, NGV735651H, NGV735672H, NGV735673H, NGV735676H, NGV735678H, NGV735679H, NGV735680H, NGV735681H, NGV735682H, NGV729348H, NGV733262H, NGV733406H, NGV733803H, NGV736526H, NGV728438H, NGV731063H, NGV727824H, NGV728619H, NGV728620H, NGV728621H, NGV729722H, NGV729845H, NGV732384H, NGV732385H, NGV735188H, NGV736799H, NGV737082H, NGV733557H, NGV728754H, NGV729849H, NGV730356H, NGV736028H, NGV729024H, NGV730336H, NGV730517H, NGV731332H, NGV733232H, NGV733338H, NGV736785H, NGV736789H, NGV730944H, NGV729794H, NGV729979H, NGV730275H, NGV732536H, NGV734001H, NGV735886H, NGV735910H, NGV735964H, NGV735967H, NGV736347H, NGV736622H, NGV736841H, NGV728937H, NGV730776H, NGV730777H, NGV735031H, NGV736633H, NGV736875H, NGV728458H, NGV731618H, NGV732188H, NGV736391H, NGV737260H, NGV729508H, NGV734336H, NGV728954H, NGV730545H, NGV731608H, NGV733635H, NGV734218H, NGV734222H, NGV734330H, NGV736257H, NGV736794H, NGV737005H, NGV730509H, NGV730563H, NGV731457H, NGV733086H, NGV733775H, NGV732396H, NGV736005H, NGV733350H, NGV730842H, NGV730911H, NGV733549H, NGV734167H, NGV734428H, NGV735200H, NGV736476H, NGV736953H, NGV733796H, NGV728904H, NGV735337H, NGV728710H, NGV729734H, NGV730362H, NGV733013H, NGV730285H, NGV732898H, NGV735421H, NGV735422H, NGV729340H, NGV733843H, NGV734550H, NGV729715H, NGV735240H, NGV735243H, NGV728911H, NGV731113H, NGV734833H, NGV731429H, NGV729166H, NGV731433H, NGV736221H, NGV737133H, NGV737134H, NGV735475H, NGV729431H, NGV730332H, NGV737236H, NGV733815H, NGV732155H, NGV726686H, NGV730562H, NGV730818H, NGV733404H, NGV735842H, NGV728764H, NGV728786H, NGV728787H, NGV729206H, NGV729846H, NGV731615H, NGV731635H, NGV732783H, NGV733159H, NGV734369H, NGV734819H, NGV728307H, NGV729175H, NGV730810H, NGV730940H, NGV730941H, NGV731343H, NGV733603H, NGV735905H, NGV736942H, NGV731820H, NGV735627H, NGV736316H, NGV736807H, NGV737262H, NGV737263H, NGV728861H, NGV732488H, NGV735026H, NGV735997H, NGV736676H, NGV729977H, NGV731314H, NGV734371H, NGV730552H, NGV730099H, NGV730348H, NGV734982H, NGV729077H, NGV728115H, NGV731091H, NGV731468H, NGV731702H, NGV736522H, NGV730989H, NGV733255H, NGV734631H, NGV729068H, NGV732477H, NGV733030H, NGV735137H, NGV732909H, NGV728866H, NGV729240H, NGV730242H, NGV730294H, NGV730797H, NGV731573H, NGV733023H, NGV733121H, NGV733141H, NGV733241H, NGV735537H, NGV736182H, NGV736631H, NGV736643H, NGV736763H, NGV736829H, NGV737233H, NGV730109H, NGV733259H, NGV734993H, NGV737015H, NGV731119H, NGV729338H, NGV730846H, NGV730847H, NGV731366H, NGV731390H, NGV732881H, NGV733568H, NGV734289H, NGV734567H, NGV737011H, NGV729082H, NGV729083H, NGV730869H, NGV732430H, NGV733020H, NGV733021H, NGV733511H, NGV733589H, NGV734599H, NGV735005H, NGV735193H, NGV736265H, NGV733820H, NGV730434H, NGV734707H, NGV733570H, NGV729126H, NGV729278H, NGV736314H, NGV733767H, NGV731415H, NGV731669H, NGV729378H, NGV731261H, NGV728863H, NGV730866H, NGV731597H, NGV733071H, NGV737231H, NGV732849H, NGV733183H, NGV732479H, NGV737156H, NGV728412H, NGV731826H, NGV731812H, NGV733838H, NGV734357H, NGV734734H, NGV735520H, NGV735521H, NGV735522H, NGV730915H, NGV731684H, NGV732313H, NGV728688H, NGV730331H, NGV733052H, NGV733240H, NGV736877H, NGV737239H, NGV730778H, NGV733261H, NGV732963H, NGV732981H, NGV735303H, NGV736866H, NGV729381H, NGV734553H, NGV735980H, NGV735162H, NGV729820H, NGV731051H, NGV733915H, NGV735873H, NGV729172H, NGV728425H, NGV728470H, NGV728505H, NGV728997H, NGV728998H, NGV729000H, NGV729224H, NGV729787H, NGV729963H, NGV730291H, NGV730312H, NGV730314H, NGV730449H, NGV731125H, NGV731130H, NGV731830H, NGV733169H, NGV733170H, NGV733362H, NGV733671H, NGV733744H, NGV734685H, NGV734897H, NGV734955H, NGV735247H, NGV735419H, NGV736001H, NGV736497H, NGV730445H, NGV731194H, NGV733327H, NGV733719H, NGV733754H, NGV736290H, NGV729505H, NGV728622H, NGV730292H, NGV731607H, NGV733701H, NGV736227H, NGV736259H, NGV736302H, NGV735024H, NGV732726H, NGV731697H, NGV735392H, NGV730567H, NGV731824H, NGV733073H, NGV733763H, NGV734453H, NGV735605H, NGV735704H, NGV728415H, NGV728697H, NGV728769H, NGV732298H, NGV733847H, NGV735010H, NGV726352H, NGV728355H, NGV729423H, NGV729424H, NGV729425H, NGV729427H, NGV731021H, NGV732308H, NGV732309H, NGV735443H, NGV735444H, NGV729491H, NGV731311H, NGV727233H, NGV727841H, NGV736181H, NGV728136H, NGV728137H, NGV729718H, NGV731439H, NGV731473H, NGV731594H, NGV732902H, NGV732947H, NGV735090H, NGV735290H, NGV735891H, NGV736486H, NGV728676H, NGV729235H, NGV729656H, NGV730596H, NGV733676H, NGV733933H, NGV734972H, NGV734973H, NGV735147H, NGV736776H, NGV737033H, NGV737241H, NGV728653H, NGV730290H, NGV732905H, NGV734712H, NGV734925H, NGV729212H, NGV731623H, NGV732260H, NGV733594H, NGV733644H, NGV733691H, NGV734975H, NGV736082H, NGV736299H, NGV730215H, NGV730868H, NGV733551H, NGV733746H, NGV731738H, NGV734578H, NGV737004H, NGV728381H, NGV728382H, NGV729343H, NGV729410H, NGV729987H, NGV729988H, NGV730008H, NGV730350H, NGV730507H, NGV730508H, NGV731147H, NGV731334H, NGV732911H, NGV733780H, NGV733781H, NGV728479H, NGV735181H, NGV735194H, NGV728447H, NGV731705H, NGV733795H, NGV729028H, NGV729323H, NGV729667H, NGV729668H, NGV729748H, NGV731504H, NGV732617H, NGV732991H, NGV733805H, NGV734225H, NGV734226H, NGV734348H, NGV735045H, NGV728917H, NGV731324H, NGV732820H, NGV731748H, NGV728724H, NGV730231H, NGV730748H, NGV731175H, NGV732504H, NGV735630H, NGV728108H, NGV728965H, NGV729052H, NGV734708H, NGV735000H, NGV736803H, NGV728785H, NGV728797H, NGV728798H, NGV729070H, NGV729093H, NGV729111H, NGV729803H, NGV729804H, NGV729805H, NGV729806H, NGV730176H, NGV730585H, NGV730744H, NGV731303H, NGV731449H, NGV731542H, NGV731813H, NGV732789H, NGV732790H, NGV732791H, NGV732792H, NGV732813H, NGV733321H, NGV734232H, NGV734610H, NGV735238H, NGV735344H, NGV735531H, NGV735532H, NGV735533H, NGV735534H, NGV736988H, NGV737024H, NGV737025H, NGV731751H, NGV736511H, NGV730865H, NGV728217H, NGV728417H, NGV728612H, NGV729064H, NGV730094H, NGV730183H, NGV730521H, NGV730901H, NGV731736H, NGV731829H, NGV734617H, NGV736861H, NGV734971H, NGV735541H, NGV735542H, NGV735544H, NGV734848H, NGV728776H, NGV729736H, NGV736503H, NGV735995H, NGV736955H, NGV727834H, NGV730270H, NGV730271H, NGV732455H, NGV735610H, NGV730773H, NGV735962H, NGV729188H, NGV727819H, NGV728532H, NGV731708H, NGV734694H, NGV730487H, NGV731319H, NGV730211H, NGV733764H, NGV728661H, NGV728662H, NGV730931H, NGV732182H, NGV733234H, NGV736038H, NGV736153H, NGV736218H, NGV736619H, NGV736711H, NGV736780H, NGV737253H, NGV735692H, NGV729320H, NGV735852H, NGV728321H, NGV734332H, NGV736979H, NGV732968H, NGV733182H, NGV728848H, NGV731840H, NGV732250H, NGV734540H, NGV736304H, NGV727833H, NGV735016H, NGV730902H, NGV731637H, NGV733049H, NGV733841H, NGV736556H, NGV728963H, NGV728219H, NGV728476H, NGV728517H, NGV729793H, NGV730152H, NGV730398H, NGV730416H, NGV731073H, NGV731216H, NGV731224H, NGV732472H, NGV732857H, NGV732988H, NGV735144H, NGV737129H, NGV727849H, NGV728766H, NGV736026H, NGV735970H, NGV736951H, NGV728519H, NGV735928H, NGV727297H, NGV727847H, NGV730693H, NGV730838H, NGV732392H, NGV731325H, NGV734435H, NGV729079H, NGV729080H, NGV729366H, NGV733299H, NGV734205H, NGV734665H, NGV734666H, NGV735884H, NGV735912H, NGV733610H, NGV734345H, NGV736288H, NGV736545H, NGV737081H, NGV729379H, NGV731872H, NGV733354H, NGV729034H, NGV730412H, NGV734632H, NGV736024H, NGV728283H, NGV728286H, NGV728287H, NGV728288H, NGV728289H, NGV728290H, NGV728293H, NGV728294H, NGV728529H, NGV728530H, NGV728531H, NGV728533H, NGV728534H, NGV728535H, NGV728536H, NGV728537H, NGV728538H, NGV728539H, NGV728540H, NGV728541H, NGV730177H, NGV730244H, NGV730245H, NGV730247H, NGV730249H, NGV730250H, NGV730252H, NGV730253H, NGV730254H, NGV730255H, NGV730256H, NGV730257H, NGV730258H, NGV731038H, NGV731040H, NGV731041H, NGV731042H, NGV731043H, NGV731045H, NGV731046H, NGV731328H, NGV731549H, NGV731550H, NGV731552H, NGV731553H, NGV731554H, NGV731557H, NGV731558H, NGV731559H, NGV731560H, NGV731562H, NGV731716H, NGV731717H, NGV731718H, NGV731720H, NGV731721H, NGV731722H, NGV731723H, NGV731724H, NGV731725H, NGV731728H, NGV731729H, NGV731730H, NGV731731H, NGV731732H, NGV731735H, NGV731762H, NGV731886H, NGV731887H, NGV731888H, NGV731889H, NGV731891H, NGV731892H, NGV731893H, NGV731894H, NGV731895H, NGV731896H, NGV731897H, NGV731898H, NGV732300H, NGV733808H, NGV734153H, NGV734154H, NGV734155H, NGV734156H, NGV734157H, NGV734158H, NGV734159H, NGV734164H, NGV734241H, NGV734876H, NGV735196H, NGV735264H, NGV735266H, NGV735552H, NGV735799H, NGV735800H, NGV735802H, NGV735803H, NGV735805H, NGV735806H, NGV735808H, NGV735809H, NGV735810H, NGV735811H, NGV735841H, NGV736167H, NGV736168H, NGV736169H, NGV736172H, NGV736173H, NGV736174H, NGV736175H, NGV736176H, NGV736177H, NGV736539H, NGV736651H, NGV736657H, NGV736660H, NGV736662H, NGV736664H, NGV736666H, NGV736668H, NGV736672H, NGV736674H, NGV737041H, NGV737042H, NGV737043H, NGV737044H, NGV737045H, NGV737046H, NGV737047H, NGV737048H, NGV737049H, NGV737050H, NGV733022H, NGV729165H, NGV730897H, NGV732873H, NGV728088H, NGV728767H, NGV728890H, NGV730339H, NGV730518H, NGV731563H, NGV731757H, NGV732275H, NGV733096H, NGV733896H, NGV734381H, NGV735628H, NGV735863H, NGV736029H, NGV736207H, NGV728791H, NGV729091H, NGV730001H, NGV730524H, NGV732605H, NGV734832H, NGV736039H, NGV728465H, NGV736790H, NGV737136H, NGV733630H, NGV736272H, NGV728556H, NGV730101H, NGV736238H, NGV730819H, NGV732614H, NGV735300H, NGV728105H, NGV731801H, NGV733015H, NGV735876H, NGV736864H, NGV732704H, NGV728509H, NGV730422H, NGV730479H, NGV732457H, NGV733828H, NGV735702H, NGV735899H, NGV729069H, NGV731521H, NGV735703H, NGV729307H, NGV729446H, NGV729495H, NGV729512H, NGV730484H, NGV731305H, NGV731339H, NGV731340H, NGV731342H, NGV731591H, NGV732408H, NGV732491H, NGV733358H, NGV733556H, NGV733677H, NGV733832H, NGV735135H, NGV735160H, NGV735286H, NGV736009H, NGV736100H, NGV736159H, NGV736871H, NGV736946H, NGV731000H, NGV734370H, NGV735906H, NGV729328H, NGV729101H, NGV734211H, NGV736878H, NGV734633H, NGV729167H, NGV730189H, NGV730532H, NGV731296H, NGV728708H, NGV729138H, NGV730856H, NGV730934H, NGV731020H, NGV729461H, NGV728770H, NGV728771H, NGV728773H, NGV728774H, NGV730265H, NGV730266H, NGV730267H, NGV732450H, NGV732451H, NGV732452H, NGV734556H, NGV734557H, NGV734559H, NGV735041H, NGV735042H, NGV735043H, NGV736638H, NGV733757H, NGV730192H, NGV729470H, NGV732454H, NGV728402H, NGV730594H, NGV728473H, NGV729477H, NGV730583H, NGV730792H, NGV730923H, NGV732152H, NGV732154H, NGV737031H, NGV735983H, NGV733024H, NGV728351H, NGV735048H, NGV732162H, NGV734992H, NGV732314H, NGV732437H, NGV737076H, NGV728385H, NGV729184H, NGV728353H, NGV730117H, NGV733091H, NGV728396H, NGV728398H, NGV728821H, NGV728872H, NGV730800H, NGV730935H, NGV731295H, NGV731434H, NGV731507H, NGV732829H, NGV733320H, NGV733874H, NGV734303H, NGV734619H, NGV735206H, NGV736083H, NGV728600H, NGV735460H, NGV736513H, NGV730093H, NGV728504H, NGV728673H, NGV728231H, NGV732273H, NGV729662H, NGV730781H, NGV732234H, NGV733745H, NGV733788H, NGV735933H, NGV736954H, NGV729114H, NGV730134H, NGV730238H, NGV731547H, NGV731823H, NGV732270H, NGV733134H, NGV735258H, NGV736699H, NGV729160H, NGV730858H, NGV728469H, NGV728487H, NGV732906H, NGV733859H, NGV732619H, NGV731828H, NGV732776H, NGV731881H, NGV731884H, NGV732596H, NGV732597H, NGV735454H, NGV735455H, NGV735988H, NGV736203H, NGV729663H, NGV730816H, NGV732456H, NGV735447H.NGV727235H, NGV727237H, NGV727249H, NGV727255H, NGV727577H, NGV728337H, NGV728370H, NGV728441H, NGV728452H, NGV728542H, NGV728634H, NGV728652H, NGV728670H, NGV728702H, NGV728781H, NGV728853H, NGV728854H, NGV728855H, NGV728856H, NGV728857H, NGV728858H, NGV728859H, NGV728860H, NGV728919H, NGV728923H, NGV728924H, NGV728927H, NGV728928H, NGV728929H, NGV728932H, NGV728933H, NGV728945H, NGV728959H, NGV728960H, NGV729041H, NGV729140H, NGV729388H, NGV729841H, NGV730170H, NGV730171H, NGV730172H, NGV730299H, NGV730300H, NGV730301H, NGV730424H, NGV730427H, NGV730546H, NGV730547H, NGV730548H, NGV730549H, NGV730550H, NGV730559H, NGV730875H, NGV730876H, NGV730877H, NGV730878H, NGV730879H, NGV730880H, NGV730881H, NGV730882H, NGV730883H, NGV730886H, NGV731030H, NGV731568H, NGV731569H, NGV731571H, NGV731774H, NGV731775H, NGV731776H, NGV731777H, NGV731779H, NGV731780H, NGV731814H, NGV732208H, NGV732290H, NGV732370H, NGV732371H, NGV732374H, NGV732435H, NGV732436H, NGV732839H, NGV732840H, NGV732841H, NGV732938H, NGV732939H, NGV732940H, NGV732941H, NGV733067H, NGV733068H, NGV733101H, NGV733102H, NGV733105H, NGV733107H, NGV733109H, NGV733110H, NGV733114H, NGV733291H, NGV733292H, NGV733293H, NGV733317H, NGV733655H, NGV733656H, NGV733657H, NGV733658H, NGV734148H, NGV734361H, NGV734362H, NGV734363H, NGV734365H, NGV734366H, NGV734430H, NGV734445H, NGV734447H, NGV734624H, NGV734684H, NGV734686H, NGV734690H, NGV734691H, NGV734930H, NGV734932H, NGV734933H, NGV734934H, NGV734939H, NGV735170H, NGV735239H, NGV735261H, NGV735324H, NGV735325H, NGV735331H, NGV735509H, NGV735523H, NGV735548H, NGV735821H, NGV736186H, NGV736187H, NGV736291H, NGV736320H, NGV736349H, NGV736353H, NGV736437H, NGV736730H, NGV736960H, NGV736992H, NGV736993H, NGV736994H, NGV736995H, NGV736996H, NGV736997H, NGV736998H, NGV737000H, NGV737003H, NGV737040H, NGV737052H, NGV737056H, NGV737220H, NGV737266H, NGV727843H, NGV727845H, NGV727846H, NGV727996H, NGV731335H, NGV731336H, NGV733910H, NGV735312H, NGV736260H, NGV736500H, NGV730821H, NGV737073H, NGV728940H, NGV728942H, NGV729047H, NGV734118H, NGV736210H, NGV736244H, NGV730167H, NGV728597H, NGV729096H, NGV733629H, NGV733791H, NGV734608H, NGV735505H, NGV729200H, NGV733552H, NGV736949H, NGV730110H, NGV730853H, NGV735949H, NGV735950H, NGV733184H, NGV728384H, NGV733316H, NGV726668H, NGV730335H, NGV730788H, NGV735006H, NGV730186H, NGV730321H, NGV731517H, NGV733747H, NGV733807H, NGV728126H, NGV730181H, NGV730208H, NGV730209H, NGV730325H, NGV731111H, NGV731617H, NGV735338H, NGV728221H, NGV735279H, NGV729119H, NGV728342H, NGV733228H, NGV731126H, NGV735013H, NGV735204H, NGV735268H, NGV728916H, NGV729808H, NGV730234H, NGV730535H, NGV730995H, NGV731330H, NGV731331H, NGV733699H, NGV735895H, NGV728323H, NGV734979H, NGV729617H, NGV730488H, NGV734009H, NGV735613H, NGV736329H, NGV726885H, NGV734298H, NGV735132H, NGV737030H, NGV729798H, NGV730862H, NGV736449H, NGV730438H, NGV731110H, NGV733792H, NGV735615H, NGV735449H, NGV729419H, NGV729943H, NGV733041H, NGV733645H, NGV733673H, NGV733674H, NGV733675H, NGV733722H, NGV733739H, NGV736831H, NGV733251H, NGV733252H, NGV734380H, NGV736294H, NGV734733H, NGV729180H, NGV730572H, NGV731044H, NGV731845H, NGV732147H, NGV732170H, NGV732212H, NGV732957H, NGV733769H, NGV733914H, NGV735187H, NGV735423H, NGV736015H, NGV736339H, NGV736346H, NGV736874H, NGV727023H, NGV734285H, NGV735262H, NGV729225H, NGV735161H, NGV727625H, NGV728522H, NGV729725H, NGV733220H, NGV736178H, NGV731200H, NGV733554H, NGV736188H, NGV736189H, NGV730494H, NGV732166H, NGV731506H, NGV737132H, NGV730453H, NGV736629H, NGV729573H, NGV730000H, NGV733680H, NGV734687H, NGV735230H, NGV735846H, NGV735963H, NGV736063H, NGV735438H, NGV729030H, NGV729372H, NGV729373H, NGV729873H, NGV730943H, NGV733165H, NGV734548H, NGV734564H, NGV735059H, NGV735500H, NGV735503H, NGV735504H, NGV735506H, NGV729735H, NGV730976H, NGV731011H, NGV737256H, NGV732173H, NGV734171H, NGV731164H, NGV728864H, NGV731062H, NGV730530H, NGV728414H, NGV729671H, NGV733247H, NGV734351H, NGV735350H, NGV736035H, NGV728216H, NGV735921H, NGV730173H, NGV732895H, NGV728842H, NGV729344H, NGV729447H, NGV730216H, NGV730334H, NGV730558H, NGV731326H, NGV731496H, NGV733246H, NGV733361H, NGV733824H, NGV733916H, NGV735148H, NGV735323H, NGV735473H, NGV736262H, NGV730933H, NGV733103H, NGV733308H, NGV733315H, NGV735897H, NGV735920H, NGV735922H, NGV731610H, NGV735535H, NGV729569H, NGV730450H, NGV731643H, NGV728397H, NGV728648H, NGV731578H, NGV735932H, NGV736120H, NGV736212H, NGV732977H, NGV730855H, NGV735184H, NGV736752H, NGV737161H, NGV729022H, NGV729158H, NGV735496H, NGV733670H, NGV733620H, NGV728436H, NGV728719H, NGV728964H, NGV731035H, NGV733097H, NGV734311H, NGV734692H, NGV735283H, NGV735512H, NGV737155H, NGV731005H, NGV731556H, NGV735459H, NGV736160H, NGV733649H, NGV730269H, NGV732855H, NGV732964H, NGV733055H, NGV733865H, NGV735095H, NGV728693H, NGV728694H, NGV729232H, NGV729233H, NGV729283H, NGV729284H, NGV729286H, NGV729293H, NGV730992H, NGV730993H, NGV730994H, NGV731133H, NGV731134H, NGV731135H, NGV731142H, NGV731143H, NGV731144H, NGV732463H, NGV732464H, NGV732919H, NGV732920H, NGV732921H, NGV732923H, NGV732924H, NGV732925H, NGV732926H, NGV732927H, NGV733576H, NGV733577H, NGV733578H, NGV733579H, NGV733580H, NGV733581H, NGV733582H, NGV733583H, NGV733798H, NGV733835H, NGV733836H, NGV733837H, NGV735426H, NGV735427H, NGV735428H, NGV735429H, NGV737142H, NGV737144H, NGV737145H, NGV737146H, NGV737147H, NGV737148H, NGV737149H, NGV737150H, NGV730762H, NGV728069H, NGV734541H, NGV730221H, NGV728459H, NGV728483H, NGV728506H, NGV728588H, NGV729238H, NGV729239H, NGV729403H, NGV730534H, NGV730581H, NGV730798H, NGV730840H, NGV731300H, NGV731575H, NGV732445H, NGV732446H, NGV732447H, NGV734683H, NGV735274H, NGV735275H, NGV735276H, NGV735278H, NGV728512H, NGV728345H, NGV734905H, NGV733118H, NGV729100H, NGV728646H, NGV729411H, NGV729809H, NGV729810H, NGV736801H, NGV737113H, NGV728406H, NGV729324H, NGV730414H, NGV731351H, NGV731380H, NGV735996H, NGV729360H, NGV735911H, NGV731309H, NGV734316H, NGV729480H, NGV736148H, NGV728545H, NGV728550H, NGV730747H, NGV735507H, NGV730261H, NGV730830H, NGV730926H, NGV732310H, NGV732606H, NGV732878H, NGV735185H, NGV735253H, NGV730820H, NGV734448H, NGV729273H, NGV729112H, NGV728996H, NGV730469H, NGV731454H, NGV732787H, NGV732901H, NGV733036H, NGV732513H, NGV736835H, NGV727771H, NGV731595H, NGV735128H, NGV730551H, NGV730817H, NGV730832H, NGV732167H, NGV732810H, NGV733399H, NGV735840H, NGV731072H, NGV731658H, NGV732795H, NGV733026H, NGV733707H, NGV735885H, NGV736312H, NGV736636H, NGV726942H, NGV727564H, NGV730103H, NGV731662H, NGV735145H, NGV728218H, NGV730906H, NGV735688H, NGV735689H, NGV735474H, NGV728486H, NGV733017H, NGV731878H, NGV728852H, NGV731185H, NGV733627H, NGV729308H, NGV731484H, NGV736002H, NGV728651H, NGV728671H, NGV730147H, NGV731453H, NGV732976H, NGV728098H, NGV729073H, NGV730403H, NGV731416H, NGV734574H, NGV733817H, NGV730541H, NGV729430H, NGV732601H, NGV735339H, NGV735526H, NGV737027H, NGV729520H, NGV736816H, NGV728695H, NGV728696H, NGV728885H, NGV729274H, NGV729594H, NGV730495H, NGV730843H, NGV730863H, NGV732171H, NGV734299H, NGV735614H, NGV731567H, NGV731206H, NGV732890H, NGV732900H, NGV730217H, NGV730478H, NGV733793H, NGV735164H, NGV735685H, NGV736060H, NGV736152H, NGV736987H, NGV730003H, NGV729026H, NGV729054H, NGV730929H, NGV736744H, NGV728494H, NGV728598H, NGV730206H, NGV731579H, NGV728449H, NGV729316H, NGV731007H, NGV731008H, NGV733145H, NGV737194H, NGV728495H, NGV729716H, NGV731505H, NGV734538H, NGV728905H, NGV729279H, NGV736787H, NGV737226H, NGV730592H, NGV728677H, NGV732223H, NGV728096H, NGV729807H, NGV733870H, NGV737085H, NGV734326H, NGV726795H, NGV730204H, NGV730303H, NGV730953H, NGV732812H, NGV733083H, NGV733364H, NGV729991H, NGV730363H, NGV730766H, NGV730769H, NGV730770H, NGV732401H, NGV733844H, NGV735273H, NGV734296H, NGV728663H, NGV732896H, NGV728656H, NGV729823H, NGV730013H, NGV730872H, NGV732535H, NGV732877H, NGV733053H, NGV733088H, NGV733140H, NGV733536H, NGV733548H, NGV734087H, NGV734284H, NGV736032H, NGV736462H, NGV728103H, NGV731651H, NGV733717H, NGV729272H, NGV732966H, NGV733094H, NGV735915H, NGV736869H, NGV733334H, NGV741141H, NGV741819H, NGV741504H, NGV741988H, NGV740528H, NGV740390H, NGV737583H, NGV738143H, NGV738144H, NGV738145H, NGV738116H, NGV738681H, NGV738772H, NGV741700H, NGV737700H, NGV739099H, NGV737424H, NGV737897H, NGV738163H, NGV738164H, NGV738349H, NGV737301H, NGV738286H, NGV738886H, NGV739339H, NGV739340H, NGV739713H, NGV739815H, NGV741225H, NGV740784H, NGV739089H, NGV740527H, NGV737668H, NGV738218H, NGV738235H, NGV738297H, NGV738539H, NGV738763H, NGV738795H, NGV741184H, NGV741186H, NGV741187H, NGV741754H, NGV741774H, NGV738497H, NGV741253H, NGV737302H, NGV738412H, NGV738703H, NGV740788H, NGV741587H, NGV738805H, NGV739698H, NGV741016H, NGV741017H, NGV741018H, NGV738189H, NGV739756H, NGV741641H, NGV740405H, NGV739273H, NGV739204H, NGV739277H, NGV741336H, NGV741337H, NGV741775H, NGV739152H, NGV737286H, NGV738197H, NGV737306H, NGV739779H, NGV739769H, NGV741828H, NGV741596H, NGV739194H, NGV737541H, NGV738302H, NGV739743H, NGV740606H, NGV740607H, NGV740780H, NGV741851H, NGV737300H, NGV738343H, NGV739003H, NGV739737H, NGV739738H, NGV740591H, NGV738501H, NGV738502H, NGV738542H, NGV738544H, NGV739963H, NGV740980H, NGV740998H, NGV740999H, NGV741000H, NGV741051H, NGV741052H, NGV741203H, NGV737278H, NGV737337H, NGV737790H, NGV737793H, NGV738156H, NGV738975H, NGV741350H, NGV737423H, NGV741541H, NGV740550H, NGV740551H, NGV740809H, NGV740818H, NGV740831H, NGV738523H, NGV739626H, NGV740613H, NGV741138H, NGV737419H, NGV737802H, NGV738171H, NGV740404H, NGV740663H, NGV741607H, NGV738329H, NGV738219H, NGV738891H, NGV738158H, NGV741182H, NGV739109H, NGV740703H, NGV740979H, NGV741637H, NGV737379H, NGV737705H, NGV738676H, NGV738714H, NGV738716H, NGV738717H, NGV738755H, NGV738773H, NGV738810H, NGV738823H, NGV740412H, NGV740605H, NGV741181H, NGV739160H, NGV741564H, NGV737320H, NGV737381H, NGV737421H, NGV737427H, NGV737991H, NGV738338H, NGV739635H, NGV739788H, NGV741834H, NGV741835H, NGV741836H, NGV740444H, NGV738028H, NGV738814H, NGV740523H, NGV739134H, NGV739329H, NGV739330H, NGV739701H, NGV740019H, NGV740545H, NGV740636H, NGV740638H, NGV737285H, NGV737287H, NGV737323H, NGV738318H, NGV738319H, NGV738321H, NGV738400H, NGV740379H, NGV740658H, NGV740660H, NGV740951H, NGV740991H, NGV741022H, NGV741023H, NGV741024H, NGV741778H, NGV740564H, NGV738422H, NGV738424H, NGV738425H, NGV738428H, NGV739370H, NGV738410H, NGV738728H, NGV738930H, NGV740986H, NGV737927H, NGV740590H, NGV737363H, NGV737517H, NGV737901H, NGV738998H, NGV740912H, NGV740926H, NGV737491H, NGV737756H, NGV738015H, NGV738024H, NGV738027H, NGV738097H, NGV738781H, NGV738915H, NGV739249H, NGV739251H, NGV740416H, NGV741523H, NGV739155H, NGV741563H, NGV737376H, NGV737819H, NGV738076H, NGV739623H, NGV740931H, NGV741688H, NGV741808H, NGV737314H, NGV740423H, NGV740639H, NGV740413H, NGV737444H, NGV739648H, NGV738149H, NGV738880H, NGV737312H, NGV737336H, NGV737377H, NGV739259H, NGV739260H, NGV739343H, NGV740541H, NGV740552H, NGV740592H, NGV737321H, NGV737721H, NGV739193H, NGV739632H, NGV741582H, NGV741781H, NGV741730H, NGV738960H, NGV740565H, NGV737293H, NGV737652H, NGV738223H, NGV741036H, NGV741076H, NGV737325H, NGV737327H, NGV738000H, NGV738001H, NGV738146H, NGV738147H, NGV738429H, NGV738730H, NGV738746H, NGV738786H, NGV738787H, NGV738788H, NGV738789H, NGV738790H, NGV738905H, NGV739086H, NGV739216H, NGV739741H, NGV740628H, NGV740654H, NGV740655H, NGV740697H, NGV740893H, NGV740894H, NGV740895H, NGV741009H, NGV741010H, NGV741011H, NGV741144H, NGV741145H, NGV741335H, NGV741552H, NGV741625H, NGV741626H, NGV741691H, NGV741122H, NGV738408H, NGV737362H, NGV738887H, NGV740994H, NGV741008H, NGV741026H, NGV741027H, NGV738399H, NGV739752H, NGV741602H, NGV737435H, NGV737500H, NGV739133H, NGV741278H, NGV738396H, NGV740617H, NGV741520H, NGV741673H, NGV741498H, NGV737909H, NGV738311H, NGV738744H, NGV739344H, NGV739537H, NGV739708H, NGV739794H, NGV740544H, NGV740803H, NGV741114H.NGV737298H, NGV740024H, NGV740025H, NGV741619H, NGV741662H, NGV741891H, NGV738835H, NGV738843H, NGV739157H, NGV741312H, NGV741806H, NGV737569H, NGV738702H, NGV738817H, NGV741251H, NGV741464H, NGV739078H, NGV741901H, NGV738913H, NGV739819H, NGV740526H, NGV740681H, NGV740682H, NGV739207H, NGV740536H, NGV737746H, NGV741254H, NGV738215H, NGV738670H, NGV738675H, NGV739965H, NGV740109H, NGV740571H, NGV740572H, NGV740886H, NGV740887H, NGV741123H, NGV741124H, NGV741125H, NGV741126H, NGV741127H, NGV741326H, NGV737356H, NGV739178H, NGV738282H, NGV738283H, NGV738750H, NGV738753H, NGV739012H, NGV739015H, NGV739205H, NGV739746H, NGV739747H, NGV739824H, NGV739825H, NGV740626H, NGV741214H, NGV741216H, NGV741667H, NGV737294H, NGV737365H, NGV737564H, NGV737983H, NGV738102H, NGV740593H, NGV740948H, NGV741207H, NGV741208H, NGV738355H, NGV738618H, NGV740530H, NGV741899H, NGV741581H, NGV741586H, NGV741769H, NGV738155H, NGV739958H, NGV739353H, NGV741516H, NGV739332H, NGV737859H, NGV738492H, NGV738966H, NGV739081H, NGV739694H, NGV740734H, NGV740752H, NGV741507H, NGV739771H, NGV739772H, NGV740418H, NGV740419H, NGV740915H, NGV740917H, NGV740949H, NGV741907H, NGV741858H, NGV737372H, NGV738999H, NGV740563H, NGV741732H, NGV737885H, NGV738254H, NGV738255H, NGV741495H, NGV741862H, NGV738131H, NGV738622H, NGV738624H, NGV739136H, NGV738059H, NGV738476H, NGV738802H, NGV741829H, NGV737821H, NGV738348H, NGV738977H, NGV739267H, NGV740033H, NGV741331H, NGV738210H, NGV739072H, NGV741698H, NGV738963H, NGV741570H, NGV739961H, NGV738917H, NGV738063H, NGV737437H, NGV737683H, NGV740629H, NGV740671H, NGV738477H, NGV738533H, NGV738537H, NGV738860H, NGV738046H, NGV741242H, NGV741761H, NGV741762H, NGV739148H, NGV739350H, NGV740960H, NGV741660H, NGV737632H, NGV737633H, NGV741872H, NGV737964H, NGV737310H, NGV737353H, NGV737501H, NGV737680H, NGV737905H, NGV739125H, NGV739175H, NGV741235H, NGV741236H, NGV741237H, NGV737881H, NGV738071H, NGV741244H, NGV741776H, NGV738092H, NGV739214H, NGV739724H, NGV739725H, NGV740094H, NGV741255H, NGV737516H, NGV740983H, NGV740806H, NGV737664H, NGV739105H, NGV739106H, NGV737369H, NGV737810H, NGV738689H, NGV741040H, NGV741852H, NGV738211H, NGV739832H, NGV741191H, NGV741460H, NGV741461H, NGV741082H, NGV741204H, NGV739954H, NGV737340H, NGV738123H, NGV740678H, NGV738248H, NGV737282H, NGV741241H, NGV741304H, NGV739163H, NGV739782H, NGV741545H, NGV737403H, NGV737404H, NGV737405H, NGV737406H, NGV737408H, NGV737409H, NGV737410H, NGV737481H, NGV737751H, NGV737753H, NGV737754H, NGV737755H, NGV739001H, NGV739975H, NGV740449H, NGV740667H, NGV740901H, NGV740902H, NGV740903H, NGV741030H, NGV741032H, NGV741035H, NGV741735H, NGV739073H, NGV737422H, NGV737761H, NGV738005H, NGV738008H, NGV738337H, NGV738540H, NGV738541H, NGV738858H, NGV739033H, NGV739143H, NGV739805H, NGV739809H, NGV740616H, NGV741065H, NGV741066H, NGV741103H, NGV741183H, NGV741562H, NGV741627H, NGV741694H, NGV741695H, NGV741849H, NGV741850H, NGV741856H, NGV737559H, NGV737988H, NGV738192H, NGV737769H, NGV741259H, NGV741260H, NGV737798H, NGV739142H, NGV740601H, NGV739144H, NGV740596H, NGV740599H, NGV741311H, NGV738165H, NGV739127H, NGV739153H, NGV740383H, NGV740792H, NGV741599H, NGV737732H, NGV738047H, NGV738048H, NGV738791H, NGV739102H, NGV739372H, NGV739976H, NGV740547H, NGV740624H, NGV741193H, NGV741243H, NGV741302H, NGV741816H, NGV741911H, NGV738090H, NGV740403H, NGV738677H, NGV737962H, NGV741554H, NGV741860H, NGV741892H, NGV739091H, NGV739092H, NGV737332H, NGV737389H, NGV737612H, NGV737731H, NGV737925H, NGV737930H, NGV737932H, NGV738078H, NGV738263H, NGV738264H, NGV738313H, NGV738315H, NGV738726H, NGV738770H, NGV739029H, NGV739118H, NGV739138H, NGV739208H, NGV739759H, NGV739784H, NGV740097H, NGV740406H, NGV740407H, NGV741210H, NGV741211H, NGV741496H, NGV741535H, NGV741646H, NGV741764H, NGV741818H, NGV741848H, NGV739328H, NGV738785H, NGV740115H, NGV740691H, NGV740693H, NGV740961H, NGV741569H, NGV737492H, NGV739333H, NGV739757H, NGV740598H, NGV741512H, NGV741880H, NGV739962H, NGV737288H, NGV737692H, NGV737693H, NGV737726H, NGV738936H, NGV738962H, NGV740433H, NGV741352H, NGV741497H, NGV741857H, NGV739335H, NGV739336H, NGV738890H, NGV741701H, NGV741613H, NGV741638H, NGV737788H, NGV737979H, NGV741684H, NGV741991H, NGV740400H, NGV737276H, NGV737322H, NGV738782H, NGV737375H, NGV737540H, NGV739791H, NGV737967H, NGV737972H, NGV738157H, NGV738170H, NGV738230H, NGV738478H, NGV739169H, NGV739191H, NGV737794H, NGV737892H, NGV738306H, NGV738679H, NGV738764H, NGV740799H, NGV737797H, NGV737966H, NGV741643H, NGV741717H, NGV738333H, NGV739385H, NGV738073H, NGV740970H, NGV737368H, NGV737502H, NGV737613H, NGV737795H, NGV739083H, NGV740890H, NGV740967H, NGV739215H, NGV738683H, NGV738685H, NGV737488H, NGV741539H, NGV739695H, NGV741346H, NGV737975H, NGV738185H, NGV738731H, NGV739802H, NGV739803H, NGV741531H, NGV738902H, NGV740896H, NGV740898H, NGV737916H, NGV738982H, NGV739758H, NGV739947H, NGV741826H, NGV739077H, NGV739183H, NGV739640H, NGV741272H, NGV741565H, NGV741529H, NGV737415H, NGV737804H, NGV737900H, NGV738178H, NGV738221H, NGV738222H, NGV737749H, NGV737758H, NGV737760H, NGV737912H, NGV737985H, NGV737986H, NGV737987H, NGV738077H, NGV739262H, NGV739356H, NGV741189H, NGV741199H, NGV737436H, NGV737584H, NGV737796H, NGV738181H, NGV738183H, NGV738353H, NGV738803H, NGV739631H, NGV740434H, NGV740555H, NGV741286H, NGV741485H, NGV737494H, NGV740984H, NGV741525H, NGV737275H, NGV737305H, NGV737329H, NGV738186H, NGV738266H, NGV738395H, NGV738405H, NGV738701H, NGV739790H, NGV740049H, NGV740426H, NGV740642H, NGV741348H, NGV741692H, NGV741693H, NGV739115H, NGV739116H, NGV739117H, NGV739269H, NGV740921H, NGV740922H, NGV740928H, NGV740934H, NGV737695H, NGV738954H, NGV738976H, NGV741674H, NGV737385H, NGV741733H, NGV741846H, NGV738308H, NGV737291H, NGV737535H, NGV737537H, NGV738194H, NGV738242H, NGV739719H, NGV740696H, NGV741277H, NGV741281H, NGV741890H, NGV737441H, NGV741028H, NGV738631H, NGV739173H, NGV739283H, NGV740439H, NGV740440H, NGV741101H, NGV741658H, NGV737297H, NGV737936H, NGV737938H, NGV737941H, NGV738758H, NGV738832H, NGV739113H, NGV739668H, NGV739681H, NGV739801H, NGV740900H, NGV741020H, NGV741021H, NGV739087H, NGV740789H, NGV741468H, NGV741696H, NGV737634H, NGV737643H, NGV738162H, NGV738190H, NGV738948H, NGV739110H, NGV739120H, NGV739276H, NGV739733H, NGV739826H, NGV740112H, NGV740385H, NGV740573H, NGV740645H, NGV740950H, NGV740969H, NGV741198H, NGV741200H, NGV741202H, NGV737292H, NGV737364H, NGV737371H, NGV737971H, NGV738022H, NGV738873H, NGV739009H, NGV740448H, NGV740467H, NGV741726H, NGV741727H, NGV737996H, NGV737997H, NGV737999H, NGV738004H, NGV738006H, NGV738007H, NGV737296H, NGV737895H, NGV737896H, NGV739748H, NGV738250H, NGV738251H, NGV740042H, NGV740542H, NGV741755H, NGV739383H, NGV737923H, NGV737526H, NGV737527H, NGV737528H, NGV738807H, NGV740856H, NGV740862H, NGV741001H, NGV741002H, NGV741003H, NGV741004H, NGV741005H, NGV741006H, NGV741007H, NGV739158H, NGV740393H, NGV737822H, NGV738362H, NGV738365H, NGV738369H, NGV738374H, NGV738375H, NGV738377H, NGV741057H, NGV737268H, NGV737272H, NGV740438H, NGV738011H, NGV738012H, NGV738013H, NGV741861H, NGV738491H, NGV740093H, NGV737508H, NGV738686H, NGV739150H, NGV739121H, NGV739122H, NGV741098H, NGV741099H, NGV738804H, NGV738885H, NGV740396H, NGV738330H, NGV738882H, NGV741882H, NGV739762H, NGV741679H, NGV737872H, NGV739369H, NGV741217H, NGV738203H, NGV738204H, NGV738205H, NGV738632H, NGV738633H, NGV739337H, NGV738627H, NGV740644H, NGV741078H, NGV737942H, NGV740909H, NGV740910H, NGV740957H, NGV740958H, NGV741013H, NGV741014H, NGV741062H, NGV740635H, NGV741157H, NGV741159H, NGV739146H, NGV738110H, NGV738516H, NGV738517H, NGV738519H, NGV738521H, NGV738522H, NGV737653H, NGV738856H, NGV741573H, NGV737347H, NGV737417H, NGV738199H, NGV738201H, NGV738236H, NGV738237H, NGV738532H, NGV738968H, NGV738969H, NGV741128H, NGV741220H, NGV741487H, NGV741488H, NGV741767H, NGV741993H, NGV737525H, NGV741801H, NGV741802H, NGV737430H, NGV737711H, NGV739386H, NGV739709H, NGV740982H, NGV741176H, NGV741459H, NGV738139H, NGV738322H, NGV739793H, NGV740425H, NGV740546H, NGV740793H, NGV738154H, NGV737530H, NGV738272H, NGV741273H, NGV737373H, NGV737748H, NGV738928H, NGV738951H, NGV740098H, NGV740795H, NGV741514H, NGV741553H, NGV737935H, NGV739380H, NGV738970H, NGV741853H, NGV741205H, NGV740647H, NGV741537H, NGV738558H, NGV738132H, NGV741338H, NGV741341H, NGV741506H, NGV741885H, NGV739274H, NGV738095H, NGV737702H, NGV737704H, NGV737709H, NGV738030H, NGV738527H, NGV738528H, NGV739362H, NGV739363H, NGV739364H, NGV739365H, NGV739366H, NGV740447H, NGV740540H, NGV741779H, NGV738684H, NGV740838H, NGV737390H, NGV740968H, NGV741795H, NGV739949H, NGV737511H, NGV738096H, NGV737346H, NGV737475H, NGV738688H, NGV740092H, NGV740382H, NGV741197H, NGV741718H, NGV741881H, NGV741886H, NGV741887H, NGV740603H, NGV737281H, NGV741571H, NGV741895H, NGV737722H, NGV738175H, NGV738176H, NGV738177H, NGV738918H, NGV738919H, NGV741313H, NGV737867H, NGV737434H, NGV738888H, NGV738889H, NGV738869H, NGV740997H, NGV738687H, NGV739034H, NGV740028H, NGV741546H, NGV741737H, NGV739104H, NGV741528H, NGV737666H, NGV738310H, NGV738762H, NGV739352H, NGV741663H, NGV740123H, NGV737445H, NGV737446H, NGV737707H, NGV737708H, NGV737734H, NGV737738H, NGV739084H, NGV739085H, NGV739198H, NGV739232H, NGV739233H, NGV739279H, NGV739280H, NGV739281H, NGV739282H, NGV740524H, NGV740525H, NGV741068H, NGV741069H, NGV741070H, NGV741071H, NGV741219H, NGV741843H, NGV741845H, NGV740963H, NGV740965H, NGV741782H, NGV738224H, NGV738479H, NGV738480H, NGV738481H, NGV740116H, NGV741067H, NGV741274H, NGV741275H, NGV741508H, NGV738992H, NGV739797H, NGV741479H, NGV740048H, NGV738743H, NGV738853H, NGV741572H, NGV741577H, NGV741620H, NGV738691H, NGV738326H, NGV737352H, NGV737499H, NGV737563H, NGV737924H, NGV738138H, NGV738861H, NGV738862H, NGV738863H, NGV739141H, NGV739177H, NGV739179H, NGV739180H, NGV739371H, NGV741294H, NGV741745H, NGV738617H, NGV737547H, NGV738070H, NGV738818H, NGV741454H, NGV741706H, NGV737330H, NGV739744H, NGV741019H, NGV738718H, NGV738719H, NGV740559H, NGV741096H, NGV741759H, NGV739068H, NGV739082H, NGV739837H, NGV741699H, NGV740397H, NGV737763H, NGV737764H, NGV737825H, NGV737944H, NGV739243H, NGV740032H, NGV741457H, NGV741752H, NGV738551H, NGV740378H, NGV740619H, NGV740884H, NGV741540H, NGV739096H, NGV739726H, NGV741728H, NGV741666H, NGV741530H, NGV738678H, NGV740566H, NGV741526H, NGV739235H, NGV740899H, NGV737520H, NGV738208H, NGV740389H, NGV740587H, NGV741747H, NGV737497H, NGV739257H, NGV739258H, NGV740973H, NGV740976H, NGV741106H, NGV737384H, NGV737816H, NGV737818H, NGV737952H, NGV738765H, NGV740421H, NGV738420H, NGV737963H, NGV738857H, NGV740932H, NGV738243H, NGV738245H, NGV738247H, NGV739345H, NGV739346H, NGV737361H, NGV738945H, NGV739114H, NGV737440H, NGV738273H, NGV740424H, NGV741509H, NGV741740H, NGV741757H, NGV737439H, NGV741224H, NGV737358H, NGV737438H, NGV737899H, NGV737910H, NGV737945H, NGV737973H, NGV738276H, NGV738747H, NGV738907H, NGV738988H, NGV739161H, NGV739735H, NGV739736H, NGV739960H, NGV740650H, NGV740800H, NGV741465H, NGV741466H, NGV741534H, NGV741645H, NGV741682H, NGV741683H, NGV741807H, NGV741912H, NGV737380H, NGV740673H, NGV740641H, NGV741349H, NGV737334H, NGV737550H, NGV738496H, NGV739244H, NGV739245H, NGV740026H, NGV740102H, NGV740683H, NGV740776H, NGV740778H, NGV741041H, NGV741042H, NGV741045H, NGV741046H, NGV741074H, NGV741075H, NGV741399H, NGV739162H, NGV740794H, NGV738166H, NGV738868H, NGV738967H, NGV738014H, NGV739271H, NGV739285H, NGV739286H, NGV739988H, NGV740120H, NGV740669H, NGV741171H, NGV741283H, NGV741678H, NGV741817H, NGV737864H, NGV740885H, NGV741785H, NGV741589H, NGV737929H, NGV737947H, NGV738987H, NGV739131H, NGV739538H, NGV739622H, NGV739734H, NGV741132H, NGV741786H, NGV737981H, NGV739239H, NGV741542H, NGV738709H, NGV738710H, NGV739977H, NGV739981H, NGV740558H, NGV740906H, NGV741228H, NGV741229H, NGV741233H, NGV741238H, NGV741239H, NGV740387H, NGV738064H, NGV738065H, NGV737315H, NGV737982H, NGV737995H, NGV738069H, NGV739076H, NGV739792H.NGV740556H, NGV741058H, NGV741218H, NGV741664H, NGV738087H, NGV738336H, NGV738344H, NGV741044H, NGV737904H, NGV739341H, NGV738298H, NGV739742H, NGV741455H, NGV737426H, NGV740604H, NGV740972H, NGV741090H, NGV741091H, NGV739804H, NGV740651H, NGV738733H, NGV737585H, NGV741690H, NGV737495H, NGV741743H, NGV740585H, NGV741551H, NGV737490H, NGV737915H, NGV738049H, NGV741909H, NGV741910H, NGV737386H, NGV737387H, NGV737388H, NGV737697H, NGV738202H, NGV738207H, NGV739714H, NGV739715H, NGV739716H, NGV739717H, NGV739718H, NGV739828H, NGV739829H, NGV740918H, NGV741920H, NGV741921H, NGV741964H, NGV741971H, NGV740533H, NGV737572H, NGV738252H, NGV739189H, NGV740796H, NGV740988H, NGV738019H, NGV738107H, NGV738108H, NGV738284H, NGV738536H, NGV739810H, NGV739654H, NGV739657H, NGV739702H, NGV738792H, NGV739156H, NGV740118H, NGV737631H, NGV737890H, NGV738072H, NGV738091H, NGV740384H, NGV741245H, NGV741303H, NGV741777H, NGV737663H, NGV740125H, NGV741709H, NGV741710H, NGV737919H, NGV737920H, NGV740543H, NGV741246H, NGV741247H, NGV737714H, NGV738793H, NGV738794H, NGV739209H, NGV739767H, NGV740535H, NGV740791H, NGV741633H, NGV741647H, NGV740409H, NGV741129H, NGV741351H, NGV741492H, NGV741787H, NGV739100H, NGV739101H, NGV741603H, NGV739822H, NGV738099H, NGV738100H, NGV740956H, NGV741212H, NGV741213H, NGV738621H, NGV740589H, NGV741489H, NGV739382H, NGV739739H, NGV741547H, NGV741548H, NGV737416H, NGV738332H, NGV737727H, NGV739139H, NGV741201H, NGV737428H, NGV739368H, NGV741844H, NGV737319H, NGV738354H, NGV738745H, NGV739818H, NGV741630H, NGV738953H, NGV739147H, NGV741798H, NGV738124H, NGV741729H, NGV741628H, NGV739154H, NGV739634H, NGV741629H, NGV739833H, NGV737382H, NGV737815H, NGV737913H, NGV738526H, NGV738693H, NGV738883H, NGV738950H, NGV740035H, NGV740050H, NGV740053H, NGV740091H, NGV740399H, NGV740971H, NGV741549H, NGV741557H, NGV741559H, NGV741689H, NGV741809H, NGV737914H, NGV741307H, NGV740964H, NGV738937H, NGV741668H, NGV738811H, NGV738812H, NGV739373H, NGV740557H, NGV740904H, NGV740905H, NGV741089H, NGV737317H, NGV737318H, NGV737398H, NGV737718H, NGV737720H, NGV738127H, NGV738128H, NGV738340H, NGV738815H, NGV738816H, NGV738824H, NGV739192H, NGV739237H, NGV739238H, NGV739242H, NGV739246H, NGV739357H, NGV739358H, NGV739359H, NGV739638H, NGV740105H, NGV740414H, NGV740614H, NGV740615H, NGV740689H, NGV740690H, NGV740952H, NGV741054H, NGV741055H, NGV741517H, NGV741518H, NGV741521H, NGV741522H, NGV739703H, NGV739705H, NGV739706H, NGV739707H, NGV740993H, NGV738884H, NGV740386H, NGV740876H, NGV737801H, NGV738307H, NGV738738H, NGV739256H, NGV741344H, NGV737805H, NGV737806H, NGV737807H, NGV738809H, NGV739324H, NGV739325H, NGV739836H, NGV740391H, NGV740392H, NGV740883H, NGV741670H, NGV741671H, NGV741672H, NGV737510H, NGV738074H, NGV737667H, NGV741486H, NGV737811H, NGV737812H, NGV740586H, NGV741814H, NGV741624H, NGV740787H, NGV737509H, NGV741768H, NGV737304H, NGV738209H, NGV739740H, NGV741789H, NGV741790H, NGV741791H, NGV737383H, NGV739721H, NGV740408H, NGV740549H, NGV741072H, NGV741073H, NGV741240H, NGV741306H, NGV741462H, NGV737513H, NGV738017H, NGV738771H, NGV741854H, NGV737969H, NGV738756H, NGV741252H, NGV738009H, NGV741644H, NGV741792H, NGV737395H, NGV741669H, NGV737917H, NGV741703H, NGV738167H, NGV738168H, NGV737249H, NGV738018H, NGV738126H, NGV738134H, NGV738135H, NGV738261H, NGV738327H, NGV738328H, NGV738403H, NGV738720H, NGV738847H, NGV738881H, NGV738946H, NGV738956H, NGV740597H, NGV741598H, NGV741617H, NGV741618H, NGV739800H, NGV737484H, NGV737522H, NGV737861H, NGV738115H, NGV737534H, NGV738739H, NGV738708H, NGV740110H, NGV740664H, NGV737921H, NGV741533H, NGV738180H, NGV741578H, NGV741190H, NGV740532H, NGV737429H, NGV739184H, NGV738898H, NGV741708H, NGV740562H, NGV741305H, NGV737328H, NGV741261H, NGV737411H, NGV737412H, NGV738021H, NGV739381H, NGV740652H, NGV740843H, NGV741724H, NGV737507H, NGV738010H, NGV738406H, NGV739079H, NGV739950H, NGV741744H, NGV738345H, NGV740977H, NGV740978H, NGV741595H, NGV738056H, NGV738094H, NGV738112H, NGV739187H, NGV740569H, NGV740107H, NGV738104H, NGV738089H, NGV739355H, NGV740052H, NGV741258H, NGV737539H, NGV738331H, NGV738334H, NGV738418H, NGV738713H, NGV738878H, NGV741188H, NGV739272H, NGV738191H, NGV741631H, NGV738957H, NGV741847H, NGV737493H, NGV739710H, NGV741100H, NGV737354H, NGV738057H, NGV738748H, NGV739168H, NGV737506H, NGV737519H, NGV737581H, NGV740420H, NGV738768H, NGV738769H, NGV740466H, NGV741285H, NGV737529H, NGV738182H, NGV738184H, NGV739624H, NGV740668H, NGV741772H, NGV741749H, NGV741716H, NGV738839H, NGV740537H, NGV740539H, NGV738270H, NGV740612H, NGV738741H, NGV738734H, NGV741604H, NGV738947H, NGV737949H, NGV738414H, NGV738416H, NGV740807H, NGV740888H, NGV740889H, NGV740891H, NGV740892H, NGV740989H, NGV741038H, NGV737496H, NGV741458H, NGV739755H, NGV740923H, NGV740924H, NGV741591H, NGV737284H, NGV737671H, NGV737675H, NGV737676H, NGV738118H, NGV738119H, NGV738120H, NGV738121H, NGV738122H, NGV738559H, NGV738561H, NGV738564H, NGV738565H, NGV738566H, NGV738567H, NGV738604H, NGV738612H, NGV738614H, NGV738615H, NGV738619H, NGV738944H, NGV739823H, NGV740919H, NGV740920H, NGV741160H, NGV741161H, NGV741162H, NGV741163H, NGV741168H, NGV741750H, NGV741751H, NGV741897H, NGV737780H, NGV737782H, NGV737783H, NGV737784H, NGV737785H, NGV737787H, NGV741149H, NGV741150H, NGV741151H, NGV741153H, NGV741169H, NGV741172H, NGV741469H, NGV741471H, NGV741474H, NGV740618H, NGV737402H, NGV737706H, NGV737814H, NGV738081H, NGV738082H, NGV738084H, NGV738085H, NGV738086H, NGV739375H, NGV739376H, NGV739377H, NGV739378H, NGV739379H, NGV741130H, NGV741131H, NGV741155H, NGV741156H, NGV741173H, NGV741175H, NGV741177H, NGV741178H, NGV741179H, NGV741685H, NGV741686H, NGV741687H, NGV738845H, NGV740034H, NGV741877H, NGV741878H, NGV737309H, NGV737591H, NGV738141H, NGV738552H, NGV738553H, NGV738554H, NGV741793H, NGV741794H, NGV739750H, NGV741612H, NGV741810H, NGV740381H, NGV739132H, NGV737551H, NGV738760H, NGV738761H, NGV741615H, NGV738106H, NGV738212H, NGV738535H, NGV738735H, NGV738737H, NGV740790H, NGV741288H, NGV737906H, NGV739200H, NGV740945H, NGV740677H, NGV739129H, NGV741234H, NGV737504H, NGV738694H, NGV739137H, NGV739261H, NGV739754H, NGV738274H, NGV739360H, NGV737605H, NGV738117H, NGV739338H, NGV738949H, NGV741249H, NGV738482H, NGV738484H, NGV738485H, NGV740846H, NGV740848H, NGV740850H, NGV740863H, NGV740865H, NGV740866H, NGV741047H, NGV741048H, NGV741049H, NGV741050H, NGV741116H, NGV741117H, NGV741118H, NGV741119H, NGV741120H, NGV741121H, NGV741748H, NGV739361H, NGV738682H, NGV738929H, NGV737344H, NGV741859H, NGV737485H, NGV738759H, NGV738075H, NGV737990H, NGV738213H, NGV738216H, NGV737505H, NGV737869H, NGV738766H, NGV740621H, NGV741291H, NGV741292H, NGV741917H, NGV737316H, NGV737588H, NGV737590H, NGV739240H, NGV739247H, NGV739248H, NGV737261H, NGV738040H, NGV738041H, NGV738043H, NGV738044H, NGV740030H, NGV740774H, NGV740775H, NGV741209H, NGV737350H, NGV737656H, NGV738268H, NGV741310H, NGV741875H, NGV737803H, NGV739749H, NGV739834H, NGV739835H, NGV741876H, NGV737277H, NGV741896H, NGV741499H, NGV739697H, NGV737597H, NGV738038H, NGV739195H, NGV739197H, NGV739781H, NGV740554H, NGV740990H, NGV740992H, NGV741015H, NGV737443H, NGV737891H, NGV738225H, NGV738241H, NGV738721H, NGV739270H, NGV741500H, NGV738136H, NGV738692H, NGV739827H, NGV740925H, NGV741543H, NGV741866H, NGV737338H, NGV738767H, NGV740422H, NGV740798H, NGV739002H, NGV737524H, NGV738961H, NGV739535H, NGV739098H, NGV741813H, NGV741888H, NGV740401H, NGV741871H, NGV738512H, NGV738513H, NGV738515H, NGV739287H, NGV741276H, NGV741513H, NGV741640H, NGV741113H, NGV738140H, NGV740600H, NGV741284H, NGV741347H, NGV741712H, NGV737976H, NGV738130H, NGV737348H, NGV738068H, NGV738244H, NGV739149H, NGV739201H, NGV739206H, NGV739213H, NGV741490H, NGV741494H, NGV741821H, NGV741822H, NGV741837H, NGV741838H, NGV739367H, NGV740779H, NGV738952H, NGV737928H, NGV738941H, NGV739628H, NGV741753H, NGV737295H, NGV739075H, NGV737933H, NGV738050H, NGV738052H, NGV738055H, NGV740111H, NGV738003H, NGV739103H, NGV739972H, NGV741609H, NGV741681H, NGV741725H, NGV737948H, NGV738808H, NGV739130H, NGV739347H, NGV739773H, NGV739776H, NGV739959H, NGV740046H, NGV740785H, NGV740801H, NGV741295H, NGV741550H, NGV740594H, NGV737710H, NGV737931H, NGV738271H, NGV740632H, NGV741226H, NGV741527H, NGV738080H, NGV740782H, NGV739124H, NGV741639H, NGV739786H, NGV738291H, NGV738292H, NGV738296H, NGV738638H, NGV739107H, NGV739108H, NGV741025H, NGV738705H, NGV738706H, NGV741606H, NGV737299H, NGV737349H, NGV738401H, NGV740037H, NGV741771H, NGV737342H, NGV737661H, NGV738088H, NGV738256H, NGV738257H, NGV738258H, NGV738259H, NGV738260H, NGV738722H, NGV738723H, NGV738724H, NGV738806H, NGV739978H, NGV740043H, NGV740241H, NGV741256H, NGV741916H, NGV740665H, NGV737413H, NGV737414H, NGV737922H, NGV738188H, NGV738493H, NGV740567H, NGV740568H, NGV741758H, NGV737553H, NGV737874H, NGV738101H, NGV741104H, NGV740882H, NGV741874H, NGV737918H, NGV740625H, NGV741815H, NGV737289H, NGV737980H, NGV737994H, NGV738061H, NGV738964H, NGV738965H, NGV740560H, NGV741855H, NGV739647H, NGV740047H, NGV738265H, NGV738359H, NGV741134H, NGV738045H, NGV737879H, NGV738093H, NGV738959H, NGV739263H, NGV740627H, NGV741833H, NGV737331H, NGV739024H, NGV741215H, NGV737341H, NGV738016H, NGV740662H, NGV737775H, NGV738749H, NGV737311H, NGV741715H, NGV738382H, NGV738387H, NGV738389H, NGV738461H, NGV738462H, NGV738468H, NGV738469H, NGV738470H, NGV738472H, NGV738695H, NGV738696H, NGV738697H, NGV738698H, NGV738699H, NGV738700H, NGV738796H, NGV738797H, NGV738798H, NGV738799H, NGV738800H, NGV738801H, NGV738864H, NGV738865H, NGV738866H, NGV738867H, NGV739217H, NGV739218H, NGV739220H, NGV739221H, NGV739222H, NGV739223H, NGV739763H, NGV739764H, NGV739765H, NGV740441H, NGV740442H, NGV740443H, NGV740588H, NGV740623H, NGV740869H, NGV740871H, NGV740874H, NGV741083H, NGV741084H, NGV741085H, NGV741086H, NGV741088H, NGV741650H, NGV741651H, NGV741652H, NGV741653H, NGV741654H, NGV741655H, NGV741721H, NGV741722H, NGV741723H, NGV741867H, NGV741868H, NGV741870H, NGV737307H, NGV738150H, NGV738151H, NGV739649H, NGV741309H, NGV741555H, NGV739268H, NGV739723H, NGV740656H, NGV737554H, NGV741918H, NGV740661H, NGV741491H, NGV741463H, NGV741467H, NGV738927H, NGV737759H, NGV737808H, NGV737894H, NGV738153H, NGV738894H, NGV739967H, NGV741736H, NGV738939H, NGV739627H, NGV741195H, NGV741287H, NGV741800H, NGV737355H, NGV738172H, NGV738173H, NGV738323H, NGV739653H, NGV741333H, NGV741648H, NGV741889H, NGV737273H, NGV737274H, NGV741884H, NGV737351H, NGV738547H, NGV738549H, NGV740029H, NGV737360H, NGV738503H, NGV738504H, NGV738507H, NGV738508H, NGV738509H, NGV738510H, NGV738511H, NGV738896H, NGV740570H, NGV737690H, NGV739164H, NGV737392H, NGV739630H, NGV740802H, NGV738233H, NGV737542H, NGV737543H, NGV739785H, NGV740781H, NGV741680H, NGV737514H, NGV737974H, NGV738114H, NGV739796H, NGV739811H, NGV739812H, NGV741713H, NGV741714H, NGV739969H, NGV737425H, NGV737733H, NGV739090H, NGV739151H, NGV739696H, NGV741475H, NGV737600H, NGV738933H, NGV737432H, NGV740388H, NGV738498H, NGV738499H, NGV738500H, NGV739831H, NGV740040H, NGV741290H, NGV741770H, NGV737335H, NGV738932H, NGV734627H, NGV737544H, NGV737566H, NGV737567H, NGV737592H, NGV737608H, NGV737628H, NGV737824H, NGV738032H, NGV738034H, NGV738036H, NGV738037H, NGV738269H, NGV738277H, NGV738279H, NGV738280H, NGV738281H, NGV738712H, NGV738774H, NGV738776H, NGV738777H, NGV738779H, NGV738874H, NGV738875H, NGV738876H, NGV738877H, NGV738993H, NGV739140H, NGV739165H, NGV739166H, NGV739167H, NGV739188H, NGV739190H, NGV739211H, NGV739252H, NGV739253H, NGV739254H, NGV739348H, NGV739349H, NGV739384H, NGV739727H, NGV739728H, NGV739729H, NGV739730H, NGV739731H, NGV739732H, NGV739751H, NGV739768H, NGV739952H, NGV739955H, NGV739956H, NGV739957H, NGV740022H, NGV740044H, NGV740428H, NGV740429H, NGV740430H, NGV740431H, NGV740432H, NGV740608H, NGV740609H, NGV740610H, NGV740611H, NGV740698H, NGV740699H, NGV740797H, NGV741250H, NGV741297H, NGV741298H, NGV741299H, NGV741300H, NGV741301H, NGV741481H, NGV741482H, NGV741483H, NGV741484H.NGV741632H, NGV741634H, NGV741635H, NGV741636H, NGV741707H, NGV741839H, NGV741840H, NGV741841H, NGV741842H, NGV738352H, NGV738912H, NGV738921H, NGV737267H, NGV738125H, NGV738704H, NGV741108H, NGV741820H, NGV740045H, NGV741257H, NGV738152H, NGV738931H, NGV737598H, NGV737599H, NGV738981H, NGV737574H, NGV738986H, NGV739284H, NGV740119H, NGV741676H, NGV741677H, NGV738985H, NGV739181H, NGV739760H, NGV737907H, NGV740117H, NGV740124H, NGV741900H, NGV737580H, NGV739637H, NGV740620H, NGV741480H, NGV741830H, NGV738187H, NGV739629H, NGV741111H, NGV741112H, NGV739135H, NGV739780H, NGV737993H, NGV737699H, NGV737279H, NGV737992H, NGV738555H, NGV740649H, NGV741248H, NGV737701H, NGV741136H, NGV739170H, NGV739813H, NGV738133H, NGV738324H, NGV741147H, NGV737391H, NGV738109H, NGV738411H, NGV739230H, NGV740038H, NGV738911H, NGV739224H, NGV739948H, NGV740680H, NGV741583H, NGV739814H, NGV737691H, NGV739028H, NGV739094H, NGV740036H, NGV740633H, NGV741501H, NGV741503H, NGV741536H, NGV740415H, NGV737397H, NGV739225H, NGV739226H, NGV739228H, NGV739231H, NGV739234H, NGV738160H, NGV737280H, NGV737820H, NGV739699H, NGV740529H, NGV740531H, NGV738196H, NGV738729H, NGV738754H, NGV739641H, NGV739645H, NGV740947H, NGV741477H, NGV741478H, NGV741580H, NGV741832H, NGV738991H, NGV740402H, NGV738848H, NGV739112H, NGV740867H, NGV738253H, NGV739210H, NGV741773H, NGV737908H, NGV741610H, NGV741519H, NGV741524H, NGV740534H, NGV738098H, NGV738267H, NGV737442H, NGV739374H, NGV741883H, NGV741538H, NGV737548H, NGV737549H, NGV737768H, NGV739711H, NGV739712H, NGV741456H, NGV738908H, NGV740410H, NGV741760H, NGV737269H, NGV737270H, NGV737271H, NGV737958H, NGV737961H, NGV739806H, NGV739807H, NGV739808H, NGV741221H, NGV741222H, NGV741223H, NGV741230H, NGV741231H, NGV741232H, NGV741262H, NGV741263H, NGV741265H, NGV741266H, NGV741267H, NGV741268H, NGV741269H, NGV741270H, NGV741271H, NGV741863H, NGV741865H, NGV737313H, NGV738486H, NGV738487H, NGV738488H, NGV738489H, NGV741614H, NGV741903H, NGV741904H, NGV741905H, NGV741906H, NGV739229H, NGV740643H, NGV741296H, NGV738335H, NGV741742H, NGV738002H, NGV739074H, NGV739264H, NGV739266H, NGV741077H, NGV739093H, NGV740395H, NGV738910H, NGV741282H, NGV737573H, NGV737579H, NGV737957H, NGV738020H, NGV739722H, NGV739770H, NGV741196H, NGV741608H, NGV741879H, NGV739278H, NGV739636H, NGV740394H, NGV741102H, NGV737893H, NGV738105H, NGV740121H, NGV740844H, NGV737800H, NGV737903H, NGV741139H, NGV741142H, NGV741558H, NGV741788H, NGV737394H, NGV741493H, NGV741763H, NGV737308H, NGV737290H, NGV737978H, NGV738025H, NGV738029H, NGV738031H, NGV738819H, NGV739212H, NGV741092H, NGV741093H, NGV741094H, NGV741095H, NGV741422H, NGV737884H, NGV738240H, NGV740981H, NGV741140H, NGV738879H, NGV737533H, NGV741621H, NGV741622H, NGV741623H, NGV737367H, NGV737678H, NGV738161H, NGV738356H, NGV739820H, NGV741135H, NGV741137H, NGV741803H, NGV741804H, NGV737503H, NGV739097H, NGV741657H, NGV737977H, NGV738680H, NGV740113H, NGV741566H, NGV741423H, NGV738193H, NGV741476H, NGV738871H, NGV738148H, NGV738942H, NGV738943H, NGV740553H, NGV741056H, NGV738142H, NGV740099H, NGV739095H, NGV741394H, NGV741825H, NGV741661H, NGV738938H, NGV741345H, NGV741649H, NGV738545H, NGV738113H, NGV738474H, NGV738742H, NGV741741H, NGV741502H, NGV739126H, NGV738784H, NGV737283H, NGV737582H, NGV739753H, NGV740041H, NGV740051H, NGV737487H, NGV738026H, NGV741913H, NGV741914H, NGV736983H, NGV728893H, NGV730316H, NGV732793H, NGV732387H, NGV741656H, NGV738039H, NGV734664H, NGV734705H, NGV734706H, NGV738169H, NGV733273H, NGV737789H, NGV737791H, NGV739008H, NGV728325H, NGV730330H, NGV732539H, NGV735177H, NGV738427H, NGV730805H, NGV735827H, NGV740975H, NGV730815H, NGV741780H, NGV733301H, NGV733349H, NGV734178H, NGV738309H, NGV740966H, NGV740786H, NGV737326H, NGV741146H, NGV735182H, NGV736778H, NGV733331H, NGV735691H, NGV735370H, NGV741898H, NGV739088H, NGV732985H, NGV740907H, NGV740995H, NGV741510H, NGV735495H, NGV735945H, NGV737823H, NGV727591H, NGV729137H, NGV736269H, NGV741719H, NGV741811H, NGV730281H, NGV737051H, NGV729455H, NGV734432H, NGV740548H, NGV740018H, NGV735254H, NGV731122H, NGV735882H, NGV736155H, NGV738955H, NGV740021H, NGV741037H, NGV735573H, NGV732525H, NGV740930H, NGV741560H, NGV735309H, NGV736935H, NGV737984H, NGV736434H, NGV740985H, NGV729666H, NGV735192H, NGV735407H, NGV740427H, NGV729401H, NGV732286H, NGV731629H, NGV737078H, NGV734021H, NGV736608H, NGV740858H, NGV736812H, NGV732413H, NGV732414H, NGV737058H, NGV729356H, NGV737219H, NGV730095H, NGV740913H, NGV731136H, NGV735872H, NGV736215H, NGV740908H, NGV740959H, NGV740962H, NGV731543H, NGV735903H, NGV741059H, NGV741060H, NGV741063H, NGV741064H, NGV741109H, NGV741110H, NGV734962H, NGV734119H, NGV728489H, NGV740522H, NGV741711H, NGV727566H, NGV737725H, NGV735529H, NGV737433H, NGV734562H, NGV736800H, NGV730753H, NGV735403H, NGV739199H, NGV734595H, NGV735488H, NGV728561H, NGV741227H, NGV728768H, NGV729094H, NGV735515H, NGV730268H, NGV731885H, NGV741105H, NGV729208H, NGV736798H, NGV740933H, NGV740027H, NGV737034H, NGV737036H, NGV741505H, NGV727562H, NGV740446H, NGV735481H, NGV732993H, NGV733010H, NGV735136H, NGV731323H, NGV727925H, NGV738275H, NGV731586H, NGV739203H, NGV734165H, NGV735857H, NGV735858H, NGV735859H, NGV734301H, NGV737965H, NGV734894H, NGV736232H, NGV739830H, NGV730837H, NGV738934H, NGV741133H, NGV728840H, NGV741039H, NGV738179H, NGV733852H, NGV734458H, NGV730499H, NGV731386H, NGV734134H, NGV735812H, NGV738813H, NGV740955H, NGV741053H, NGV737008H, NGV734310H, NGV729387H, NGV730804H, NGV738351H, NGV741327H, NGV734638H, NGV739111H, NGV729370H, NGV737010H, NGV728625H, NGV731082H, NGV731084H, NGV731766H, NGV731767H, NGV731769H, NGV740538H, NGV730263H, NGV739821H, NGV741167H, NGV728296H, NGV728744H, NGV731357H, NGV734372H, NGV735082H, NGV737786H, NGV740031H, NGV741148H, NGV741152H, NGV741470H, NGV741472H, NGV741473H, NGV735653H, NGV735675H, NGV735677H, NGV728281H, NGV733923H, NGV740946H, NGV732987H, NGV736400H, NGV741012H, NGV733819H, NGV735856H, NGV732828H, NGV734628H, NGV731819H, NGV733706H, NGV737970H, NGV737523H, NGV731604H, NGV737813H, NGV728070H, NGV735445H, NGV733772H, NGV729749H, NGV728409H, NGV734452H, NGV738494H, NGV730501H, NGV733078H, NGV729220H, NGV733164H, NGV734382H, NGV735336H, NGV736171H, NGV741087H, NGV729186H, NGV739816H, NGV728666H, NGV728104H, NGV732458H, NGV731047H, NGV731153H, NGV734193H, NGV734355H, NGV738548H, NGV738550H, NGV737359H, NGV741665H, NGV737370H, NGV733276H, NGV730280H, NGV740914H, NGV741206H, NGV737366H, NGV738195H, NGV737937H, NGV735976H, NGV737956H, NGV741264H, NGV741864H, NGV736802H, NGV735551H, NGV733033H, NGV733034H, NGV733035H, NGV739745H, NGV741079H, NGV741080H, NGV741081H, NGV740805H, NGV733333H, NGV736125H, NGV737873H, NGV741812H, NGV728462H, NGV728906H, NGV729061H, NGV736033H, NGV733113H, NGV734321H; GTIN 00763000604202, Serial Numbers: NGV737536H, NGV737571H, NGV737774H, NGV737777H, NGV737463H, NGV737594H, NGV737772H, NGV737464H, NGV737601H, NGV737776H, NGV740290H, NGV737703H, NGV737740H, NGV740291H, NGV740298H, NGV740299H, NGV737728H, NGV740163H, NGV737650H, NGV737654H, NGV737659H, NGV737673H, NGV737677H, NGV737698H, NGV740289H, NGV740305H, NGV740347H, NGV740361H, NGV740362H, NGV740365H, NGV740366H, NGV737713H, NGV737114H, NGV737729H, NGV737448H, NGV737681H, NGV737750H, NGV740363H, NGV740373H, NGV736722H, NGV736679H, NGV740202H, NGV740377H, NGV737160H, NGV740212H, NGV740349H, NGV740271H, NGV740359H, NGV740364H, NGV740177H, NGV737100H, NGV740170H, NGV737182H, NGV740351H, NGV736736H, NGV740171H, NGV740211H, NGV740201H, NGV736729H, NGV737159H, NGV737185H, NGV737215H, NGV740272H, NGV740343H, NGV737093H, NGV737177H, NGV737224H, NGV737162H, NGV740344H, NGV737175H, NGV740168H, NGV740189H, NGV737183H, NGV737112H, NGV740213H, NGV740358H, NGV737642H, NGV737618H, NGV737610H, NGV737644H, NGV737622H, NGV737623H, NGV737638H, NGV737619H, NGV737617H, NGV737630H, NGV737779H, NGV737781H, NGV737778H, NGV737767H, NGV737757H, NGV737648H; GTIN 00763000634100, Serial Numbers: NGV742298H, NGV742299H, NGV742300H, NGV742301H, NGV742302H, NGV742303H, NGV742305H, NGV742306H, NGV742308H, NGV742309H, NGV742310H, NGV742311H, NGV742313H, NGV742315H, NGV742317H, NGV742319H, NGV742320H, NGV742321H, NGV742322H, NGV742323H, NGV742324H, NGV742325H, NGV742326H, NGV742327H, NGV742328H, NGV742330H, NGV742331H, NGV742332H, NGV742333H, NGV742334H, NGV742336H, NGV742337H, NGV742338H, NGV742339H, NGV742340H, NGV742342H, NGV742343H, NGV742344H, NGV742345H, NGV742347H, NGV742349H, NGV742350H, NGV742351H, NGV742352H, NGV742353H, NGV742354H, NGV742355H, NGV742356H, NGV742357H, NGV742358H, NGV742359H, NGV742360H, NGV742361H, NGV742362H, NGV742769H, NGV742770H, NGV742771H, NGV742774H, NGV742775H, NGV742776H, NGV742778H, NGV742780H, NGV742781H, NGV742782H, NGV742783H, NGV742785H, NGV742786H, NGV742789H, NGV743735H, NGV743736H, NGV743737H, NGV743738H, NGV743739H, NGV743740H, NGV743741H, NGV743742H, NGV743743H, NGV743744H, NGV743747H, NGV743748H, NGV743749H, NGV743750H, NGV743751H, NGV743752H, NGV743753H, NGV743754H, NGV743755H, NGV743756H, NGV743757H, NGV743758H, NGV743760H, NGV743761H, NGV743763H, NGV743765H, NGV743767H, NGV743773H, NGV743774H, NGV743775H, NGV743776H, NGV743777H, NGV743778H, NGV743779H, NGV743781H, NGV743783H, NGV743786H, NGV743788H, NGV743789H, NGV743791H, NGV744702H, NGV744703H, NGV744705H, NGV744706H, NGV744707H, NGV744708H, NGV744709H, NGV744710H, NGV744711H, NGV744712H, NGV744713H, NGV744714H, NGV744715H, NGV744716H, NGV744717H, NGV744718H, NGV744719H, NGV744720H, NGV744721H, NGV744722H, NGV744723H, NGV744724H, NGV744725H, NGV744726H, NGV744727H, NGV744729H, NGV744730H, NGV744732H, NGV744733H, NGV744734H, NGV744941H, NGV744948H, NGV744949H, NGV744951H, NGV744952H, NGV744953H, NGV744954H, NGV744955H, NGV744957H, NGV744958H, NGV744959H, NGV744960H, NGV744961H, NGV744962H, NGV744963H, NGV744964H, NGV744965H, NGV744966H, NGV745330H, NGV745331H, NGV745332H, NGV745334H, NGV745335H, NGV745336H, NGV745337H, NGV745338H, NGV745339H, NGV745342H, NGV745343H, NGV745344H, NGV745345H, NGV745347H, NGV745348H, NGV745349H, NGV745352H, NGV745353H, NGV745354H, NGV745355H, NGV745361H, NGV745364H, NGV745887H, NGV745888H, NGV745889H, NGV745890H, NGV745891H, NGV745892H, NGV745893H, NGV745895H, NGV745896H, NGV745897H, NGV745898H, NGV745899H, NGV745900H, NGV745901H, NGV745902H, NGV745903H, NGV745904H, NGV745905H, NGV745906H, NGV745907H, NGV745912H, NGV747363H, NGV747364H, NGV747365H, NGV747366H, NGV747657H, NGV747658H, NGV747659H, NGV747660H, NGV747661H, NGV747662H, NGV747663H, NGV747664H, NGV747665H, NGV747666H, NGV747667H, NGV747668H, NGV747669H, NGV747670H, NGV747671H, NGV747672H, NGV747673H, NGV747674H, NGV747675H, NGV747676H, NGV747677H, NGV748116H, NGV748117H, NGV748118H, NGV748121H, NGV748122H, NGV748123H, NGV748124H, NGV748125H, NGV748127H, NGV748128H, NGV748129H, NGV748130H, NGV748131H, NGV748132H, NGV748133H, NGV748134H, NGV748135H, NGV748136H, NGV748137H, NGV748138H, NGV748139H, NGV748141H, NGV748142H, NGV748143H, NGV748144H, NGV748145H, NGV748146H, NGV748147H, NGV748148H, NGV748149H, NGV748150H, NGV748178H, NGV748451H, NGV748452H, NGV748453H, NGV748454H, NGV748456H, NGV748457H, NGV748458H, NGV748459H, NGV748460H, NGV748461H, NGV748462H, NGV748463H, NGV748464H, NGV748465H, NGV748466H, NGV748467H, NGV748468H, NGV748469H, NGV748470H, NGV748471H, NGV748472H, NGV748473H, NGV748474H, NGV748475H, NGV748476H, NGV748477H, NGV748478H, NGV748479H, NGV748480H, NGV748481H, NGV748537H, NGV748539H, NGV748540H, NGV748543H, NGV748547H, NGV748550H, NGV748552H, NGV748557H, NGV748559H, NGV748560H, NGV748562H, NGV748564H, NGV748566H, NGV748569H, NGV748570H, NGV748571H, NGV748573H, NGV748575H, NGV748576H, NGV748578H, NGV748580H, NGV748581H, NGV748582H, NGV748584H, NGV748587H, NGV748589H, NGV748590H, NGV748591H, NGV748594H, NGV748597H, NGV748598H, NGV748601H, NGV748604H, NGV748606H, NGV748607H, NGV748609H, NGV748610H, NGV748616H, NGV748617H, NGV748619H, NGV748625H, NGV748626H, NGV750875H, NGV750876H, NGV750877H, NGV750879H, NGV750880H, NGV750881H, NGV750882H, NGV750883H, NGV750884H, NGV750885H, NGV750888H, NGV750894H, NGV750895H, NGV750896H, NGV750897H, NGV750898H, NGV750899H, NGV750900H, NGV750901H, NGV750902H, NGV750903H, NGV750904H, NGV750905H, NGV750906H, NGV750907H, NGV750908H, NGV750909H, NGV750910H, NGV750911H, NGV750912H, NGV750913H, NGV750914H, NGV750915H, NGV750916H, NGV750917H, NGV750918H, NGV750919H, NGV750920H, NGV750922H, NGV750923H, NGV750924H, NGV750925H, NGV750926H, NGV750927H, NGV750928H, NGV750929H, NGV750930H, NGV750931H, NGV750932H, NGV750934H, NGV750935H, NGV750936H, NGV750937H, NGV750938H, NGV750939H, NGV750940H, NGV750941H, NGV750942H, NGV750943H, NGV750944H, NGV750945H, NGV750947H, NGV750948H, NGV750949H, NGV750950H, NGV750951H, NGV750952H, NGV750953H, NGV751222H, NGV751225H, NGV751227H, NGV751232H, NGV751234H, NGV751236H, NGV751237H, NGV751238H, NGV751239H, NGV751954H, NGV751955H, NGV751958H, NGV751962H, NGV751963H, NGV751964H, NGV751965H, NGV751966H, NGV751967H, NGV751968H, NGV751970H, NGV751974H, NGV751975H, NGV751976H, NGV751977H, NGV751979H, NGV751981H, NGV751982H, NGV751985H, NGV751990H, NGV751991H, NGV751994H, NGV751996H, NGV752001H, NGV752002H, NGV752076H, NGV752078H, NGV752087H, NGV752088H, NGV752090H, NGV752091H, NGV752092H, NGV752093H, NGV752094H, NGV752098H, NGV752104H, NGV752105H, NGV752108H, NGV752110H, NGV752111H, NGV752118H, NGV752120H, NGV752121H, NGV752122H, NGV752123H, NGV752124H, NGV752125H, NGV752126H, NGV752127H, NGV752128H, NGV752129H, NGV752130H, NGV752131H, NGV752132H, NGV752133H, NGV752134H, NGV752135H, NGV752136H, NGV752137H, NGV752139H, NGV752140H, NGV752141H, NGV752143H, NGV752144H, NGV752145H, NGV752146H, NGV752147H, NGV752191H, NGV752228H, NGV752231H, NGV752233H, NGV752602H, NGV752727H, NGV752728H, NGV752730H, NGV752731H, NGV752733H, NGV752734H, NGV752735H, NGV752736H, NGV752738H, NGV752739H, NGV752740H, NGV752743H, NGV752744H, NGV752745H, NGV752746H, NGV752747H, NGV752748H, NGV752750H, NGV752752H, NGV752791H, NGV752797H, NGV752807H, NGV752808H, NGV752812H, NGV752816H, NGV752819H, NGV752825H, NGV752826H, NGV752834H, NGV752839H, NGV752840H, NGV752845H, NGV752852H, NGV752857H, NGV752860H, NGV752861H, NGV752862H, NGV752865H, NGV752867H, NGV752868H, NGV752869H, NGV752871H, NGV752872H, NGV752873H, NGV752875H, NGV752898H, NGV752971H, NGV752972H, NGV752974H, NGV752977H, NGV752978H, NGV752979H, NGV752980H, NGV752981H, NGV752984H, NGV752985H, NGV752987H, NGV752989H, NGV752990H, NGV752991H, NGV752992H, NGV752993H, NGV752994H, NGV752995H, NGV752996H, NGV752997H, NGV752998H, NGV752999H, NGV753000H, NGV753001H, NGV753147H, NGV753159H, NGV753164H, NGV753189H, NGV753191H, NGV753204H, NGV753205H, NGV753206H, NGV753299H, NGV753310H, NGV753311H, NGV753312H, NGV753313H, NGV753314H, NGV753927H, NGV753928H, NGV753929H, NGV753930H, NGV753933H, NGV753935H, NGV753936H, NGV753937H, NGV753938H, NGV753939H, NGV753940H, NGV754015H, NGV754092H, NGV754093H, NGV754094H, NGV754095H, NGV754096H, NGV754097H, NGV754099H, NGV754100H, NGV754101H, NGV754102H, NGV754103H, NGV754104H, NGV754105H, NGV754107H, NGV754108H, NGV754109H, NGV754110H, NGV754111H, NGV754274H, NGV754276H, NGV754277H, NGV754281H, NGV754283H, NGV754284H, NGV754285H, NGV754287H, NGV754289H, NGV754291H, NGV754292H, NGV754294H, NGV754295H, NGV754296H, NGV754299H, NGV754332H, NGV754334H, NGV754335H, NGV754336H, NGV754337H, NGV754340H, NGV754343H, NGV754344H, NGV754348H, NGV754356H, NGV754364H, NGV754369H, NGV754371H, NGV754374H, NGV754375H, NGV754382H, NGV754386H, NGV754392H, NGV754410H, NGV754413H, NGV754419H, NGV754421H, NGV754423H, NGV754425H, NGV754428H, NGV754429H, NGV754430H, NGV754442H, NGV754443H, NGV754445H, NGV754446H, NGV754447H, NGV754448H, NGV754449H, NGV754450H, NGV754578H, NGV754579H, NGV754581H, NGV754582H, NGV754584H, NGV754585H, NGV754586H, NGV754587H, NGV754588H, NGV754625H, NGV754627H, NGV754629H, NGV754630H, NGV754631H, NGV754635H, NGV754636H, NGV754663H, NGV754665H, NGV754672H, NGV754675H, NGV754677H, NGV754678H, NGV754679H, NGV747362H, NGV742617H, NGV745300H, NGV751044H, NGV751043H, NGV746140H, NGV752694H, NGV753905H, NGV747844H, NGV753767H, NGV753391H, NGV750867H, NGV754783H, NGV748020H, NGV744781H, NGV742088H, NGV748838H, NGV749035H, NGV749470H, NGV746938H, NGV753087H, NGV742105H, NGV745070H, NGV746013H, NGV746672H, NGV746987H, NGV747542H, NGV748111H, NGV750407H, NGV751036H, NGV752044H, NGV752605H, NGV754189H, NGV754791H, NGV742253H, NGV743188H, NGV746168H, NGV746347H, NGV746348H, NGV750610H, NGV744567H, NGV749098H, NGV742044H, NGV744920H, NGV745834H, NGV745957H, NGV747221H, NGV747825H, NGV748923H, NGV753354H, NGV753355H, NGV755012H, NGV751086H, NGV742120H, NGV742839H, NGV744675H, NGV750311H, NGV751132H, NGV745427H, NGV743712H, NGV744172H, NGV744559H, NGV744560H, NGV748740H, NGV750024H, NGV750156H, NGV750716H, NGV747018H, NGV747933H, NGV753083H, NGV745872H, NGV743182H, NGV747020H, NGV747576H, NGV748629H, NGV749138H, NGV752434H, NGV743331H, NGV743944H, NGV746371H, NGV749195H, NGV750200H, NGV742194H, NGV741968H, NGV742080H, NGV742086H, NGV742233H, NGV743553H, NGV743617H, NGV743618H, NGV748436H, NGV748796H, NGV748797H, NGV749108H, NGV753772H, NGV754191H, NGV744842H, NGV745767H, NGV746312H, NGV747069H, NGV747344H, NGV747432H, NGV747821H, NGV747822H, NGV747864H, NGV748863H, NGV749363H, NGV749493H, NGV749658H, NGV749791H, NGV750282H, NGV750351H, NGV751489H, NGV752433H, NGV754247H, NGV742221H, NGV750004H, NGV752195H, NGV742807H, NGV753585H, NGV742841H, NGV746983H, NGV747635H, NGV749935H, NGV751598H, NGV752411H, NGV742198H, NGV743350H, NGV744053H, NGV744060H, NGV744483H, NGV745526H, NGV745547H, NGV745673H, NGV746432H, NGV746490H, NGV746527H, NGV746654H, NGV747215H, NGV747581H, NGV747919H, NGV749255H, NGV749488H, NGV749690H, NGV750368H, NGV751293H, NGV753519H, NGV753520H, NGV753541H, NGV753668H, NGV754021H, NGV754773H, NGV743552H, NGV744823H, NGV745649H, NGV745945H, NGV747271H, NGV747273H, NGV747766H, NGV749287H, NGV746783H, NGV743427H, NGV743428H, NGV749174H, NGV751647H, NGV746616H, NGV747829H, NGV749120H, NGV749539H, NGV749681H, NGV751176H, NGV748382H, NGV751052H, NGV742064H, NGV744161H, NGV743616H, NGV743447H, NGV743448H, NGV743513H, NGV743872H, NGV746867H, NGV746869H, NGV747932H, NGV748093H, NGV749244H, NGV749374H, NGV749650H, NGV751301H, NGV752467H, NGV743281H, NGV743603H, NGV744415H, NGV744581H, NGV744584H, NGV746649H, NGV747189H, NGV747805H, NGV748970H, NGV743426H, NGV753329H, NGV743421H, NGV745230H, NGV750317H, NGV752461H, NGV743278H, NGV743454H, NGV743652H, NGV743653H, NGV744314H, NGV745032H, NGV746004H, NGV746402H, NGV746403H, NGV746428H, NGV749626H, NGV750133H, NGV742252H, NGV746061H, NGV742070H, NGV745967H, NGV749139H, NGV750978H, NGV747258H, NGV748076H, NGV752755H, NGV753091H, NGV742273H, NGV742806H, NGV743397H, NGV743684H, NGV744809H, NGV747750H, NGV750011H, NGV750369H, NGV751885H, NGV742060H, NGV742666H, NGV743568H, NGV743569H, NGV743598H, NGV743976H, NGV744544H, NGV744545H, NGV744851H, NGV744939H, NGV746532H, NGV746986H, NGV747232H, NGV747415H, NGV748827H, NGV748829H, NGV749136H, NGV749423H, NGV749426H, NGV749640H, NGV750650H, NGV751840H, NGV752323H, NGV751242H, NGV742363H, NGV744997H, NGV744999H, NGV745024H, NGV745793H, NGV746948H, NGV746949H, NGV747700H, NGV747816H, NGV748485H, NGV748964H, NGV749685H, NGV751273H, NGV753657H, NGV744302H, NGV744305H, NGV744308H, NGV746668H, NGV747202H, NGV747203H, NGV750498H, NGV751765H, NGV751766H, NGV753846H, NGV753847H, NGV747371H, NGV748069H, NGV749273H, NGV751418H, NGV743221H, NGV744120H, NGV743432H, NGV743559H, NGV743561H, NGV743562H, NGV747973H, NGV748441H, NGV751069H, NGV752305H, NGV752306H, NGV743595H, NGV744149H, NGV747165H, NGV747166H, NGV750141H, NGV752150H, NGV752292H, NGV753064H, NGV753066H, NGV753076H, NGV753119H, NGV753120H, NGV743223H, NGV743247H, NGV743273H, NGV744072H, NGV747016H, NGV747733H, NGV747734H, NGV751110H, NGV751300H, NGV744913H, NGV747335H, NGV747991H, NGV748160H, NGV749955H, NGV744399H, NGV744461H, NGV744583H, NGV744586H, NGV745022H, NGV745693H, NGV746479H, NGV746648H, NGV746650H, NGV747188H, NGV747190H, NGV747806H, NGV748974H, NGV751064H, NGV751170H, NGV751773H, NGV753820H, NGV753822H, NGV754693H, NGV747985H, NGV748961H, NGV744121H, NGV745223H, NGV749905H, NGV750125H, NGV752591H, NGV743726H, NGV747568H, NGV748817H, NGV749433H, NGV751299H, NGV743807H, NGV743813H, NGV745036H, NGV749796H, NGV750064H, NGV751618H, NGV743268H, NGV742072H, NGV742220H, NGV742562H, NGV743228H, NGV743853H, NGV744753H, NGV747047H, NGV747590H, NGV747640H, NGV749010H, NGV749448H, NGV749793H, NGV751102H, NGV751488H, NGV751564H, NGV753952H, NGV754406H, NGV744924H, NGV744101H, NGV749122H, NGV751522H, NGV741977H, NGV742580H, NGV745102H, NGV746039H, NGV746049H, NGV746050H, NGV747297H, NGV747689H, NGV747993H, NGV748247H, NGV750095H, NGV750098H, NGV750602H, NGV751862H, NGV751863H, NGV751898H, NGV751585H, NGV744537H, NGV746308H, NGV745271H, NGV751581H, NGV741975H, NGV746705H, NGV747846H, NGV742104H, NGV742804H, NGV744011H, NGV744014H, NGV744017H, NGV744025H, NGV745291H, NGV745615H, NGV745687H, NGV745804H, NGV746319H, NGV748967H, NGV748968H, NGV750315H, NGV752363H, NGV753836H, NGV741965H, NGV741966H, NGV742215H, NGV742216H, NGV743435H, NGV743979H, NGV743980H, NGV743982H, NGV744173H, NGV744335H, NGV745141H, NGV746925H, NGV746926H, NGV746927H, NGV746928H, NGV747140H, NGV747994H, NGV748824H, NGV748891H, NGV750023H, NGV750028H, NGV750043H, NGV750281H, NGV752149H, NGV752916H, NGV753998H, NGV754207H, NGV744087H, NGV744088H, NGV743668H, NGV747800H, NGV750715H, NGV747706H, NGV745417H, NGV742605H, NGV746678H, NGV747090H, NGV747711H, NGV747906H, NGV745137H, NGV745145H, NGV746592H, NGV750736H, NGV750737H, NGV749213H, NGV749569H, NGV753587H, NGV754177H, NGV754377H, NGV754790H, NGV746863H, NGV747026H, NGV743932H, NGV747982H, NGV750510H, NGV742636H, NGV743387H, NGV744176H, NGV744571H, NGV744572H, NGV744837H, NGV745256H, NGV745270H, NGV745434H, NGV745771H, NGV745861H, NGV746457H, NGV746884H, NGV747036H, NGV747037H, NGV747292H, NGV747786H, NGV747886H, NGV747977H, NGV747978H, NGV748225H, NGV748437H, NGV749005H, NGV749301H, NGV749474H, NGV749475H, NGV749630H, NGV749644H, NGV749651H, NGV749972H, NGV750033H, NGV750034H, NGV750366H, NGV751582H, NGV751789H, NGV752300H, NGV752301H, NGV752302H, NGV752307H, NGV752484H, NGV752486H, NGV752487H, NGV753106H, NGV753107H, NGV753128H, NGV753714H, NGV753941H, NGV754063H, NGV754197H, NGV754475H, NGV755458H, NGV755459H, NGV745183H, NGV750866H, NGV742391H, NGV745319H, NGV747980H, NGV751077H, NGV751529H, NGV752200H, NGV742092H, NGV743390H, NGV745469H, NGV746712H, NGV747178H, NGV747179H, NGV747990H, NGV744521H, NGV749826H, NGV749845H, NGV743936H, NGV749489H, NGV752760H, NGV744940H, NGV744944H, NGV747997H, NGV747999H, NGV752271H, NGV742184H, NGV743687H, NGV745916H, NGV746930H, NGV744174H, NGV742160H, NGV744643H, NGV745327H, NGV745329H, NGV745985H, NGV746994H, NGV748335H, NGV751613H, NGV742227H, NGV742384H, NGV742385H, NGV743586H, NGV743625H, NGV743660H, NGV744129H, NGV744858H, NGV745357H, NGV745358H, NGV745720H, NGV746062H, NGV746126H, NGV746673H, NGV746674H, NGV746675H, NGV747343H, NGV749230H, NGV750344H, NGV750695H, NGV750832H, NGV751455H, NGV752153H, NGV742152H, NGV742284H, NGV742285H, NGV743322H, NGV743325H, NGV743716H, NGV743719H, NGV744747H, NGV744773H, NGV745197H, NGV746104H, NGV747349H, NGV747434H, NGV747479H, NGV747795H, NGV748028H, NGV749667H, NGV749891H, NGV750134H, NGV750799H, NGV751247H, NGV751561H, NGV751562H, NGV752341H, NGV752375H, NGV752661H, NGV743256H, NGV753056H, NGV753332H, NGV742281H, NGV744405H, NGV745764H, NGV743938H, NGV744499H, NGV747917H, NGV747924H, NGV748648H, NGV749071H, NGV749411H, NGV749414H, NGV753565H, NGV742101H, NGV742276H, NGV742278H, NGV743584H, NGV743585H, NGV743939H, NGV743958H, NGV744191H, NGV744193H, NGV745143H, NGV746625H, NGV747560H, NGV747561H, NGV748650H, NGV749920H, NGV751705H, NGV751782H, NGV752710H, NGV753699H, NGV754206H, NGV744796H, NGV744866H, NGV744139H, NGV745381H, NGV742246H, NGV743368H, NGV743373H, NGV743472H, NGV743638H, NGV743669H, NGV743708H, NGV744134H, NGV744135H, NGV744362H, NGV744406H, NGV744615H, NGV744650H, NGV744758H, NGV744777H, NGV744923H, NGV745184H, NGV745186H, NGV745247H, NGV745248H, NGV745700H, NGV745951H, NGV745975H, NGV746315H, NGV746316H, NGV746462H, NGV746653H, NGV747081H, NGV747417H, NGV747531H, NGV747823H, NGV747854H, NGV747896H, NGV747904H, NGV747967H, NGV748174H, NGV748199H, NGV748641H, NGV748642H, NGV748683H, NGV748835H, NGV748928H, NGV749079H, NGV749176H, NGV749288H, NGV749487H, NGV749643H, NGV749655H, NGV749673H, NGV749684H, NGV749924H, NGV749984H, NGV750065H, NGV750108H, NGV750218H, NGV750367H, NGV750372H, NGV750450H, NGV750453H, NGV750454H, NGV750692H, NGV750793H, NGV750963H, NGV751076H, NGV751128H, NGV751129H, NGV751171H, NGV751175H, NGV751818H, NGV752173H, NGV752174H, NGV752211H, NGV752213H, NGV752214H, NGV752215H, NGV752328H, NGV752680H, NGV752779H, NGV753219H, NGV753548H, NGV753613H, NGV753655H, NGV753818H, NGV753987H, NGV753988H, NGV754453H, NGV754608H, NGV754659H, NGV754660H, NGV754701H, NGV742468H, NGV744519H, NGV745376H, NGV747450H, NGV749008H, NGV749824H, NGV750622H, NGV752040H, NGV747325H, NGV747798H, NGV749146H, NGV748204H, NGV752419H, NGV754081H, NGV744663H, NGV744682H, NGV745190H, NGV745766H, NGV746956H, NGV747617H, NGV747690H, NGV749975H, NGV753132H, NGV743460H, NGV741970H, NGV744743H, NGV748002H, NGV750286H, NGV751159H, NGV744848H, NGV746473H, NGV747172H, NGV751628H, NGV751629H, NGV753126H, NGV754057H, NGV754199H, NGV754556H, NGV754857H, NGV755044H, NGV743349H, NGV744273H, NGV744510H, NGV744912H, NGV746167H, NGV746702H, NGV746988H, NGV747630H, NGV748189H, NGV748646H, NGV748647H, NGV748843H, NGV749113H, NGV750421H, NGV751220H, NGV751386H, NGV751674H, NGV752041H, NGV752218H, NGV752399H, NGV753671H, NGV754539H, NGV742460H, NGV747612H, NGV742076H, NGV742207H, NGV742501H, NGV745451H, NGV743449H, NGV744116H, NGV744752H, NGV745233H, NGV745701H, NGV748678H, NGV748907H, NGV751513H, NGV754037H, NGV754376H, NGV754713H, NGV754913H, NGV754914H, NGV743433H, NGV749942H, NGV751240H, NGV753481H, NGV754251H, NGV743326H, NGV745537H, NGV747008H, NGV747247H, NGV747772H, NGV747856H, NGV751849H, NGV752480H, NGV742819H, NGV744167H, NGV745433H, NGV746398H, NGV747194H, NGV748635H, NGV751261H, NGV751429H, NGV752439H, NGV752440H, NGV753959H, NGV753961H, NGV754269H, NGV742532H, NGV743399H, NGV743869H, NGV746719H, NGV749362H, NGV751148H, NGV753910H, NGV752338H, NGV754217H, NGV742075H, NGV742123H, NGV743906H, NGV743908H, NGV743909H, NGV743910H, NGV744552H, NGV744553H, NGV744554H, NGV744555H, NGV745800H, NGV745801H, NGV745802H, NGV746525H, NGV746548H, NGV746550H, NGV746551H, NGV746552H, NGV746890H, NGV746893H, NGV746894H, NGV747093H, NGV747097H, NGV747098H, NGV747099H, NGV747100H, NGV747101H, NGV747102H, NGV747105H, NGV747107H, NGV747108H, NGV750539H, NGV750542H, NGV750544H, NGV750548H, NGV750550H, NGV750551H, NGV750555H, NGV750556H, NGV750559H, NGV750564H, NGV750566H, NGV752362H, NGV744594H, NGV746954H, NGV751325H, NGV742752H, NGV743312H, NGV743663H, NGV744652H, NGV747618H, NGV748385H, NGV749267H, NGV750206H, NGV742377H, NGV742378H, NGV742509H, NGV742549H, NGV742632H, NGV742633H, NGV743205H, NGV743207H, NGV743953H, NGV744050H, NGV744548H, NGV744627H, NGV745196H, NGV745421H, NGV746604H, NGV746606H, NGV746684H, NGV746685H, NGV747656H, NGV748094H, NGV748497H, NGV750116H, NGV750229H, NGV750582H, NGV750583H, NGV751909H, NGV751910H, NGV751915H, NGV752477H, NGV753223H, NGV753383H, NGV753413H, NGV753419H, NGV753681H, NGV754083H, NGV754119H, NGV754240H, NGV754737H, NGV754738H, NGV745958H, NGV742459H, NGV743313H, NGV743951H, NGV744200H, NGV744400H, NGV744401H, NGV744402H, NGV747723H, NGV748935H, NGV749623H, NGV750261H, NGV750632H, NGV751700H, NGV745436H, NGV744260H, NGV744268H, NGV744479H, NGV746841H, NGV747779H, NGV749305H, NGV749951H, NGV749998H, NGV752167H, NGV752185H, NGV753142H, NGV753233H, NGV754064H, NGV754175H, NGV742156H, NGV742440H, NGV742441H, NGV742442H, NGV743270H, NGV743356H, NGV743363H, NGV743723H, NGV744454H, NGV745228H, NGV745727H, NGV746055H, NGV747198H, NGV747199H, NGV747392H, NGV748113H, NGV748439H, NGV748440H, NGV748685H, NGV748688H, NGV748830H, NGV749298H, NGV749804H, NGV750989H, NGV751139H, NGV751249H, NGV751265H, NGV751375H, NGV752420H, NGV752982H, NGV752269H, NGV743429H, NGV744598H, NGV746430H, NGV750533H, NGV750534H, NGV751476H, NGV746135H, NGV743703H, NGV746858H, NGV750858H, NGV744536H, NGV745908H, NGV748417H, NGV748908H, NGV750150H, NGV750973H, NGV751482H, NGV752847H, NGV754036H, NGV742563H, NGV742566H, NGV743424H, NGV743581H, NGV743588H, NGV743677H, NGV743679H, NGV746599H, NGV746601H, NGV746827H, NGV746895H, NGV749808H, NGV749809H, NGV752577H, NGV753605H, NGV754212H, NGV754213H, NGV754883H, NGV748818H, NGV749692H, NGV742085H, NGV743815H, NGV748182H, NGV742293H, NGV742379H, NGV743471H, NGV745685H, NGV751664H, NGV749129H, NGV753768H, NGV745813H, NGV749963H, NGV745745H, NGV747490H, NGV749550H, NGV751141H, NGV745047H, NGV745939H, NGV748196H, NGV750162H, NGV743222H, NGV744142H, NGV744144H, NGV742500H, NGV745885H, NGV746539H, NGV750209H, NGV748628H, NGV753145H, NGV742755H, NGV744803H, NGV744804H, NGV748000H, NGV741955H, NGV744684H, NGV744987H, NGV745886H, NGV747521H, NGV742280H, NGV747907H, NGV747033H, NGV752160H, NGV753080H, NGV751506H, NGV753090H, NGV752324H, NGV743798H, NGV744566H, NGV744808H, NGV751675H, NGV746859H, NGV749222H, NGV750859H, NGV754523H, NGV747044H, NGV747461H, NGV749190H, NGV750493H, NGV750495H, NGV742639H, NGV745227H, NGV743381H, NGV743718H, NGV745193H, NGV745664H, NGV745765H, NGV747347H, NGV747934H, NGV749469H, NGV749471H, NGV750220H, NGV750729H, NGV751666H, NGV751701H, NGV751702H, NGV753281H, NGV749585H, NGV744683H, NGV749354H, NGV742530H, NGV742531H, NGV747073H, NGV747609H, NGV749724H, NGV752698H, NGV742586H, NGV742542H, NGV743193H, NGV744328H, NGV744982H, NGV746626H, NGV748861H, NGV742644H, NGV743195H, NGV743352H, NGV743353H, NGV744569H, NGV744749H, NGV745642H, NGV746314H, NGV746449H, NGV746677H, NGV746707H, NGV747627H, NGV747648H, NGV747871H, NGV747935H, NGV748001H, NGV748862H, NGV750802H, NGV752156H, NGV752334H, NGV752336H, NGV753518H, NGV753898H, NGV753899H, NGV747636H, NGV743264H, NGV742066H, NGV742516H, NGV742518H, NGV742599H, NGV742600H, NGV744114H, NGV744600H, NGV746386H, NGV748976H, NGV748977H, NGV749885H, NGV752443H, NGV749848H, NGV749876H, NGV753954H, NGV753376H, NGV742370H, NGV744568H, NGV746874H, NGV752217H, NGV754492H, NGV743200H, NGV747588H, NGV743876H, NGV746847H, NGV743675H, NGV744360H, NGV744850H, NGV745794H, NGV746607H, NGV747869H, NGV748011H, NGV749364H, NGV749376H, NGV754462H, NGV742231H, NGV743305H, NGV743940H, NGV745198H, NGV745199H, NGV745577H, NGV745585H, NGV746051H, NGV746052H, NGV746450H, NGV747393H, NGV747394H, NGV747396H, NGV747397H, NGV749384H, NGV750657H, NGV750667H, NGV750678H, NGV750865H, NGV752418H, NGV744137H, NGV744155H, NGV744910H, NGV746939H, NGV750061H, NGV744748H, NGV745171H, NGV748009H, NGV749203H, NGV750284H, NGV752361H, NGV744323H, NGV744334H, NGV750038H, NGV750039H, NGV750669H, NGV743572H, NGV743573H, NGV751135H, NGV751136H, NGV748741H, NGV742153H, NGV742649H, NGV743310H, NGV743383H, NGV743654H, NGV743655H, NGV743885H, NGV743886H, NGV743887H, NGV743889H, NGV743890H, NGV743891H, NGV743892H, NGV743894H, NGV743895H, NGV743896H, NGV743897H, NGV743901H, NGV747304H, NGV747305H, NGV747306H, NGV747307H, NGV747308H, NGV748811H, NGV749709H, NGV749710H, NGV750301H, NGV750302H, NGV750305H, NGV750306H, NGV750309H, NGV751748H, NGV751749H, NGV752071H, NGV752072H, NGV752210H, NGV752702H, NGV752703H, NGV753213H, NGV753217H, NGV753361H, NGV753362H, NGV753456H, NGV753701H, NGV742520H, NGV747522H, NGV749591H, NGV742142H, NGV742143H, NGV742186H, NGV742211H, NGV742416H, NGV742417H, NGV743316H, NGV743794H, NGV743821H, NGV744356H, NGV744396H, NGV744398H, NGV744642H, NGV745068H, NGV745225H, NGV745770H, NGV745874H, NGV745875H, NGV745984H, NGV745986H, NGV746393H, NGV746825H, NGV747067H, NGV747121H, NGV747124H, NGV747131H, NGV747234H, NGV747610H, NGV747859H, NGV748077H, NGV748349H, NGV748367H, NGV749377H, NGV750191H, NGV750396H, NGV750964H, NGV751104H, NGV751320H, NGV751324H, NGV751411H, NGV751531H, NGV751549H, NGV751604H, NGV751882H, NGV751927H, NGV752585H, NGV752586H, NGV752587H, NGV752901H, NGV752908H, NGV753723H, NGV753737H, NGV753755H, NGV753995H, NGV754583H, NGV754691H, NGV754921H, NGV742418H, NGV743141H, NGV745934H, NGV745969H, NGV746448H, NGV752754H, NGV753207H, NGV754949H, NGV746831H, NGV742436H, NGV742443H, NGV753082H, NGV750721H, NGV750966H, NGV751041H, NGV741823H, NGV742558H, NGV743169H, NGV743275H, NGV743277H, NGV743720H, NGV744117H, NGV744475H, NGV744547H, NGV745214H, NGV745538H, NGV746351H, NGV746374H, NGV746897H, NGV749988H, NGV751351H, NGV749581H, NGV749180H, NGV747915H, NGV754619H, NGV744341H, NGV742166H, NGV744035H, NGV742059H, NGV743859H, NGV745075H, NGV745797H, NGV747645H, NGV747811H, NGV744143H, NGV745518H, NGV747796H, NGV749795H, NGV750640H, NGV754395H, NGV754398H, NGV742419H, NGV742420H, NGV743330H, NGV743734H, NGV744157H, NGV744669H, NGV745250H, NGV745418H, NGV745696H, NGV746003H, NGV746082H, NGV747845H, NGV748318H, NGV748509H, NGV748510H, NGV748518H, NGV748921H, NGV750345H, NGV751231H, NGV754521H, NGV749228H, NGV743393H, NGV744889H, NGV744891H, NGV748725H, NGV749149H, NGV749980H, NGV749981H, NGV750262H, NGV750263H, NGV751689H, NGV753352H, NGV754650H, NGV754829H, NGV754937H, NGV742081H, NGV743574H, NGV751534H, NGV742405H, NGV744315H, NGV744967H, NGV745213H, NGV748659H, NGV752180H, NGV752181H, NGV753232H, NGV741967H, NGV742237H, NGV742435H, NGV742498H, NGV743186H, NGV743354H, NGV743728H, NGV743903H, NGV743904H, NGV743986H, NGV744079H, NGV744207H, NGV744208H, NGV744783H, NGV744822H, NGV744825H, NGV744826H, NGV744985H, NGV745194H, NGV745699H, NGV746368H, NGV746369H, NGV746828H, NGV746829H, NGV746848H, NGV746902H, NGV746903H, NGV748941H, NGV748942H, NGV749103H, NGV749104H, NGV749721H, NGV750119H, NGV750124H, NGV750731H, NGV750732H, NGV750839H, NGV751112H, NGV751114H, NGV751115H, NGV751313H, NGV751314H, NGV751755H, NGV751756H, NGV751757H, NGV751805H, NGV752259H, NGV752260H, NGV752262H, NGV754786H, NGV742631H, NGV748373H, NGV744608H, NGV746138H, NGV747042H, NGV748450H, NGV742234H, NGV743283H, NGV744133H, NGV744926H, NGV745536H, NGV746418H, NGV746951H, NGV747938H, NGV749081H, NGV749397H, NGV749642H, NGV749665H, NGV751644H, NGV753438H, NGV742524H, NGV747368H, NGV744775H, NGV747715H, NGV743411H, NGV747574H, NGV754776H, NGV754894H, NGV746024H, NGV746876H, NGV747333H, NGV742180H, NGV742181H, NGV744033H, NGV744504H, NGV744589H, NGV745201H, NGV745915H, NGV746868H, NGV748101H, NGV748165H, NGV751093H, NGV751930H, NGV752056H, NGV752468H, NGV753370H, NGV754115H, NGV742041H, NGV742296H, NGV744123H, NGV744124H, NGV747824H, NGV747998H, NGV748095H, NGV748164H, NGV748924H, NGV749593H, NGV749683H, NGV750689H, NGV753953H, NGV754106H, NGV754331H, NGV742158H, NGV743163H, NGV743358H, NGV744603H, NGV744664H, NGV744751H, NGV745054H, NGV745763H, NGV745933H, NGV746012H, NGV747683H, NGV747759H, NGV747760H, NGV748228H, NGV749992H, NGV752402H, NGV754019H, NGV754184H, NGV743266H, NGV749454H, NGV752682H, NGV743857H, NGV751007H, NGV743265H, NGV743981H, NGV751935H, NGV744365H, NGV747144H, NGV747599H, NGV748380H, NGV747416H, NGV744477H, NGV745803H, NGV747437H, NGV748023H, NGV745772H, NGV745773H, NGV745774H, NGV746997H, NGV747975H, NGV747979H, NGV749670H, NGV749641H, NGV744165H, NGV742295H, NGV743192H, NGV746706H, NGV744887H, NGV750730H, NGV749163H, NGV743468H, NGV743469H, NGV743578H, NGV743579H, NGV744500H, NGV744501H, NGV746387H, NGV747585H, NGV747705H, NGV749664H, NGV749813H, NGV752421H, NGV752466H, NGV743451H, NGV743705H, NGV743858H, NGV744576H, NGV745251H, NGV745396H, NGV745539H, NGV746352H, NGV746620H, NGV746621H, NGV747850H, NGV747852H, NGV747882H, NGV749143H, NGV749249H, NGV749815H, NGV750046H, NGV750192H, NGV750193H, NGV750988H, NGV753101H, NGV753321H, NGV753922H, NGV754769H, NGV743601H, NGV743204H, NGV743276H, NGV746612H, NGV748987H, NGV751836H, NGV751837H, NGV754915H, NGV755442H, NGV743627H, NGV744933H, NGV746299H, NGV747613H, NGV748665H, NGV746870H, NGV742235H, NGV742544H, NGV744495H, NGV744543H, NGV743110H, NGV743351H, NGV744542H, NGV744755H, NGV745119H, NGV745123H, NGV745124H, NGV747272H, NGV747943H, NGV751350H, NGV754771H, NGV746534H, NGV747155H, NGV747619H, NGV750035H, NGV752046H, NGV752920H, NGV751340H, NGV746522H, NGV747132H, NGV748932H, NGV746727H, NGV742052H, NGV746309H, NGV746693H, NGV747996H, NGV749483H, NGV749497H, NGV749945H, NGV755100H, NGV742149H, NGV742150H, NGV743594H, NGV746521H, NGV746877H, NGV749428H, NGV751031H, NGV751616H, NGV753397H, NGV744674H, NGV747897H, NGV749089H, NGV749169H, NGV747707H, NGV742612H, NGV742613H, NGV743284H, NGV742113H, NGV742115H, NGV743218H, NGV744125H, NGV744332H, NGV744612H, NGV744972H, NGV745391H, NGV747034H, NGV747060H, NGV748175H, NGV749361H, NGV749582H, NGV749686H, NGV750197H, NGV753293H, NGV753606H, NGV743171H, NGV743942H, NGV744235H, NGV744236H, NGV744465H, NGV744466H, NGV744467H, NGV744468H, NGV747642H, NGV747643H, NGV747820H, NGV747849H, NGV749850H, NGV750145H, NGV750223H, NGV750226H, NGV750825H, NGV751008H, NGV751777H, NGV742522H, NGV743196H, NGV746397H, NGV746687H, NGV746688H, NGV753086H, NGV743420H, NGV746722H, NGV746872H, NGV749242H, NGV749256H, NGV751536H, NGV743315H, NGV743505H, NGV745359H, NGV745812H, NGV745998H, NGV745999H, NGV746005H, NGV746686H, NGV747867H, NGV748667H, NGV748670H, NGV749365H, NGV750327H, NGV751206H, NGV751567H, NGV751677H, NGV751951H, NGV752325H, NGV752326H, NGV754035H, NGV754548H, NGV755151H, NGV744641H, NGV744865H, NGV746331H, NGV746332H, NGV747330H, NGV748405H, NGV748493H, NGV748494H, NGV749185H, NGV749240H, NGV750143H, NGV750954H, NGV750955H, NGV751145H, NGV751588H, NGV751869H, NGV751872H, NGV752423H, NGV753345H, NGV753636H, NGV754558H, NGV754774H, NGV742145H, NGV742406H, NGV743514H, NGV743658H, NGV743711H, NGV743880H, NGV744148H, NGV744764H, NGV744774H, NGV744902H, NGV745257H, NGV745873H, NGV746303H, NGV746304H, NGV747052H, NGV747427H, NGV748448H, NGV749178H, NGV749817H, NGV749835H, NGV750153H, NGV750161H, NGV750401H, NGV751245H, NGV751793H, NGV751831H, NGV752061H, NGV752062H, NGV752340H, NGV752372H, NGV753078H, NGV754788H, NGV755388H, NGV742656H, NGV742810H, NGV743523H, NGV744502H, NGV744665H, NGV745840H, NGV745841H, NGV747300H, NGV747301H, NGV747451H, NGV749379H, NGV749561H, NGV749926H, NGV751317H, NGV752064H, NGV753621H, NGV754399H, NGV741928H, NGV743732H, NGV744645H, NGV744759H, NGV744927H, NGV744984H, NGV745043H, NGV745172H, NGV745525H, NGV745781H, NGV745829H, NGV746716H, NGV746982H, NGV747058H, NGV747065H, NGV747449H, NGV747595H, NGV747873H, NGV747874H, NGV749996H, NGV750128H, NGV750696H, NGV752859H, NGV754165H, NGV754514H, NGV755107H, NGV742210H, NGV742595H, NGV742596H, NGV743492H, NGV743535H, NGV744630H, NGV744631H, NGV745722H, NGV745725H, NGV745818H, NGV746772H, NGV746773H, NGV746774H, NGV747299H, NGV747570H, NGV748253H, NGV748257H, NGV748258H, NGV748259H, NGV749292H, NGV749535H, NGV749551H, NGV749556H, NGV749889H, NGV750003H, NGV750598H, NGV751322H, NGV751323H, NGV753455H, NGV746081H, NGV742374H, NGV742380H, NGV743382H, NGV743538H, NGV743733H, NGV744597H, NGV747378H, NGV747748H, NGV747866H, NGV747983H, NGV749461H, NGV749649H, NGV750045H, NGV751138H, NGV752659H, NGV753764H, NGV754226H, NGV754373H, NGV742476H, NGV742071H, NGV744410H, NGV745161H, NGV747512H, NGV750981H, NGV745449H, NGV746468H, NGV748003H, NGV745432H, NGV746630H, NGV746632H, NGV747000H, NGV747038H, NGV747193H, NGV748099H, NGV750969H, NGV751254H, NGV753374H, NGV753601H, NGV753602H, NGV753614H, NGV754267H, NGV754328H, NGV747753H, NGV746670H, NGV742102H, NGV742373H, NGV744094H, NGV744550H, NGV745982H, NGV745983H, NGV746446H, NGV746447H, NGV747313H, NGV747615H, NGV748645H, NGV749269H, NGV749270H, NGV750738H, NGV750739H, NGV750740H, NGV752464H, NGV743954H, NGV743955H, NGV743956H, NGV744084H, NGV746819H, NGV747180H, NGV747181H, NGV747183H, NGV747185H, NGV750733H, NGV751579H, NGV742177H, NGV742178H, NGV743480H, NGV743481H, NGV745237H, NGV745238H, NGV745240H, NGV745241H, NGV747389H, NGV747391H, NGV747728H, NGV747792H, NGV747793H, NGV747801H, NGV747802H, NGV749507H, NGV753457H, NGV753459H, NGV753696H, NGV753697H, NGV742201H, NGV742202H, NGV742204H, NGV743245H, NGV743966H, NGV743971H, NGV743983H, NGV744077H, NGV744205H, NGV744385H, NGV744404H, NGV744805H, NGV744806H, NGV745128H, NGV745131H, NGV745340H, NGV746025H, NGV746541H, NGV746547H, NGV747225H, NGV747302H, NGV747303H, NGV747549H, NGV748722H, NGV748822H, NGV749012H, NGV749359H, NGV749542H, NGV749567H, NGV749646H, NGV750144H, NGV750264H, NGV750385H, NGV753660H, NGV753667H, NGV754845H, NGV754918H, NGV755501H, NGV743826H, NGV744861H, NGV745778H, NGV745780H, NGV746467H, NGV747057H, NGV747836H, NGV747837H, NGV747838H, NGV749964H, NGV751053H, NGV751060H, NGV751061H, NGV752465H, NGV753072H, NGV753092H, NGV753093H, NGV753583H, NGV753982H, NGV753985H, NGV755148H, NGV742167H, NGV742168H, NGV742279H, NGV742392H, NGV742393H, NGV744373H, NGV744374H, NGV744376H, NGV744377H, NGV744379H, NGV744380H, NGV744381H, NGV744382H, NGV744384H, NGV744386H, NGV744387H, NGV744388H, NGV744389H, NGV744390H, NGV750235H, NGV750236H, NGV750238H, NGV750239H, NGV750241H, NGV750242H, NGV750244H, NGV750245H, NGV750249H, NGV750250H, NGV750252H, NGV750256H, NGV750257H, NGV742245H, NGV742453H, NGV743943H, NGV744770H, NGV745321H, NGV746155H, NGV746725H, NGV746726H, NGV747223H, NGV749480H, NGV750198H, NGV743407H, NGV743570H, NGV743577H, NGV743591H, NGV745210H, NGV745830H, NGV745831H, NGV746388H, NGV746389H, NGV747295H, NGV747408H, NGV747410H, NGV747637H, NGV747727H, NGV748799H, NGV748882H, NGV749622H, NGV749699H, NGV749700H, NGV751758H, NGV753850H, NGV753851H, NGV753852H, NGV742053H, NGV743490H, NGV743822H, NGV744038H, NGV745296H, NGV745297H, NGV748209H, NGV749829H, NGV751934H, NGV752268H, NGV753089H, NGV754042H, NGV754043H, NGV742750H, NGV743487H, NGV743509H, NGV743520H, NGV743666H, NGV743855H, NGV744130H, NGV745517H, NGV746385H, NGV747027H, NGV747814H, NGV747922H, NGV750272H, NGV750714H, NGV750819H, NGV750971H, NGV751661H, NGV752408H, NGV753441H, NGV753442H, NGV753504H, NGV753729H, NGV754959H, NGV755475H, NGV742487H, NGV743545H, NGV743546H, NGV743547H, NGV743548H, NGV743549H, NGV744102H, NGV744104H, NGV744197H, NGV744198H, NGV744628H, NGV744786H, NGV744844H, NGV745149H, NGV745151H, NGV745152H, NGV745160H, NGV746526H, NGV746529H, NGV746531H, NGV747383H, NGV747384H, NGV747386H, NGV747961H, NGV747968H, NGV748053H, NGV748696H, NGV748989H, NGV749223H, NGV749226H, NGV749386H, NGV749513H, NGV749629H, NGV749844H, NGV750070H, NGV750074H, NGV751670H, NGV751762H, NGV751763H, NGV752709H, NGV753716H, NGV753717H, NGV753721H, NGV753722H, NGV746804H, NGV746805H, NGV746806H, NGV752713H, NGV752714H, NGV742502H, NGV742503H, NGV745236H, NGV753624H, NGV754903H, NGV751884H, NGV748947H, NGV750365H, NGV751042H, NGV742809H, NGV744295H, NGV744296H, NGV744486H, NGV745820H, NGV745948H, NGV746466H, NGV747421H, NGV747423H, NGV747424H, NGV747425H, NGV747426H, NGV748415H, NGV749654H, NGV750682H, NGV750683H, NGV750684H, NGV750685H, NGV750686H, NGV752579H, NGV753338H, NGV753449H, NGV753637H, NGV753638H, NGV753639H, NGV753640H, NGV743167H, NGV743285H, NGV743883H, NGV743884H, NGV744534H, NGV744591H, NGV745167H, NGV745168H, NGV745263H, NGV747736H, NGV750419H, NGV751838H, NGV754722H, NGV754770H, NGV755024H, NGV743444H, NGV754931H, NGV745837H, NGV747294H, NGV744416H, NGV744417H, NGV743862H, NGV744394H, NGV744925H, NGV745552H, NGV746400H, NGV750985H, NGV751623H, NGV752588H, NGV742825H, NGV744288H, NGV744289H, NGV744292H, NGV744943H, NGV745169H, NGV746880H, NGV748630H, NGV749882H, NGV751595H, NGV754402H, NGV742421H, NGV745591H, NGV745827H, NGV750276H, NGV750348H, NGV751078H, NGV743984H, NGV749073H, NGV752192H, NGV747543H, NGV750323H, NGV750856H, NGV752049H, NGV752050H, NGV742029H, NGV742607H, NGV742611H, NGV742616H, NGV743860H, NGV744771H, NGV745016H, NGV745018H, NGV745261H, NGV746690H, NGV747401H, NGV747632H, NGV747646H, NGV751518H, NGV751535H, NGV751847H, NGV753825H, NGV753826H, NGV754201H, NGV742751H, NGV743602H, NGV750637H, NGV751739H, NGV753856H, NGV753857H, NGV742499H, NGV743246H, NGV743934H, NGV742618H, NGV744604H, NGV751984H, NGV743489H, NGV744614H, NGV749846H, NGV744041H, NGV744046H, NGV744047H, NGV744370H, NGV744372H, NGV746404H, NGV746405H, NGV746407H, NGV747149H, NGV747150H, NGV747153H, NGV747154H, NGV750473H, NGV750475H, NGV750477H, NGV750478H, NGV752762H, NGV752910H, NGV753733H, NGV753734H, NGV753736H, NGV754480H, NGV754483H, NGV754484H, NGV743199H, NGV747995H, NGV749289H, NGV750968H, NGV752187H, NGV752278H, NGV743406H, NGV743459H, NGV744802H, NGV744841H, NGV745816H, NGV745817H, NGV747296H, NGV748653H, NGV749558H, NGV753249H, NGV744670H, NGV746122H, NGV744196H, NGV746544H, NGV746821H, NGV747621H, NGV748075H, NGV751711H, NGV742397H, NGV742438H, NGV746732H, NGV752457H, NGV743318H, NGV745226H, NGV749635H, NGV752342H, NGV752343H, NGV744513H, NGV748910H, NGV744131H, NGV745425H, NGV746833H, NGV750469H, NGV751774H, NGV743301H, NGV743442H, NGV748506H, NGV742175H, NGV742176H, NGV742425H, NGV742426H, NGV742533H, NGV743455H, NGV743456H, NGV743868H, NGV744375H, NGV744772H, NGV744787H, NGV745322H, NGV745323H, NGV745783H, NGV745955H, NGV745956H, NGV746359H, NGV746664H, NGV746718H, NGV747054H, NGV747055H, NGV747056H, NGV747596H, NGV747839H, NGV747840H, NGV748005H, NGV748074H, NGV748502H, NGV748504H, NGV748505H, NGV748846H, NGV748847H, NGV749094H, NGV749820H, NGV750846H, NGV750871H, NGV750872H, NGV751147H, NGV751150H, NGV752058H, NGV752060H, NGV752410H, NGV752495H, NGV752909H, NGV752928H, NGV753372H, NGV754324H, NGV754456H, NGV755104H, NGV755106H, NGV744431H, NGV747246H, NGV742469H, NGV744610H, NGV744611H, NGV747376H, NGV749904H, NGV742283H, NGV743079H, NGV743434H, NGV746305H, NGV746306H, NGV750182H, NGV750184H, NGV742205H, NGV742206H, NGV743479H, NGV744838H, NGV745424H, NGV745567H, NGV745569H, NGV745942H, NGV748215H, NGV748498H, NGV748499H, NGV748997H, NGV749019H, NGV749184H, NGV749993H, NGV752236H, NGV752483H, NGV753117H, NGV753118H, NGV753658H, NGV753687H, NGV754545H, NGV745938H, NGV749009H, NGV750869H, NGV755612H, NGV742195H, NGV742478H, NGV743226H, NGV743516H, NGV743517H, NGV744563H, NGV744564H, NGV745428H, NGV745430H, NGV745520H, NGV746034H, NGV746038H, NGV746477H, NGV747197H, NGV747438H, NGV747515H, NGV747523H, NGV747524H, NGV747525H, NGV747527H, NGV747528H, NGV747529H, NGV747564H, NGV747703H, NGV749053H, NGV749057H, NGV749061H, NGV750509H, NGV751027H, NGV752761H, NGV752764H, NGV754014H, NGV754016H, NGV754017H, NGV754712H, NGV755059H, NGV742430H, NGV753131H, NGV742448H, NGV743817H, NGV747730H, NGV748188H, NGV748190H, NGV751266H, NGV752157H, NGV753136H, NGV742400H, NGV742401H, NGV742403H, NGV742404H, NGV743607H, NGV743833H, NGV743835H, NGV743847H, NGV744852H, NGV745878H, NGV747550H, NGV747712H, NGV747761H, NGV747762H, NGV747763H, NGV750792H, NGV751823H, NGV752364H, NGV753186H, NGV754018H, NGV743415H, NGV744112H, NGV742449H, NGV748072H, NGV748073H, NGV748413H, NGV751409H, NGV752475H, NGV752476H, NGV754703H, NGV743699H, NGV745365H, NGV744533H, NGV746382H, NGV744785H, NGV747681H, NGV749953H, NGV751827H, NGV752281H, NGV742051H, NGV742470H, NGV742471H, NGV742510H, NGV742619H, NGV742620H, NGV743554H, NGV743619H, NGV743620H, NGV744701H, NGV744793H, NGV745594H, NGV745612H, NGV745627H, NGV745705H, NGV745776H, NGV745882H, NGV745909H, NGV745910H, NGV745911H, NGV746032H, NGV746481H, NGV747326H, NGV747327H, NGV747329H, NGV747603H, NGV748643H, NGV749280H, NGV749797H, NGV749910H, NGV750386H, NGV751302H, NGV751514H, NGV751706H, NGV751888H, NGV753002H, NGV742238H, NGV744549H, NGV747354H, NGV753444H, NGV753491H, NGV743172H, NGV748660H, NGV750160H, NGV750207H, NGV750280H, NGV743376H, NGV751221H, NGV743589H, NGV744565H, NGV749800H, NGV744330H, NGV744970H, NGV745356H, NGV745786H, NGV747718H, NGV747885H, NGV748375H, NGV749161H, NGV751573H, NGV751850H, NGV752045H, NGV752197H, NGV752198H, NGV754463H, NGV742541H, NGV747419H, NGV750666H, NGV754561H, NGV754709H, NGV751672H, NGV745426H, NGV749725H, NGV749976H, NGV751344H, NGV743952H, NGV744092H, NGV744093H, NGV745828H, NGV746623H, NGV747725H, NGV747726H, NGV747893H, NGV747894H, NGV747895H, NGV748939H, NGV749217H, NGV751891H, NGV751892H, NGV751894H, NGV751896H, NGV754034H, NGV749662H, NGV747950H, NGV748112H, NGV742561H, NGV754000H, NGV745277H, NGV742036H, NGV742038H, NGV742039H, NGV742505H, NGV742506H, NGV742507H, NGV744522H, NGV744523H, NGV744524H, NGV746682H, NGV746683H, NGV746881H, NGV746882H, NGV748938H, NGV750617H, NGV751305H, NGV751724H, NGV753708H, NGV750679H, NGV745584H, NGV751133H, NGV744297H, NGV743338H, NGV744340H, NGV746639H, NGV749266H, NGV751029H, NGV751030H, NGV751900H, NGV752435H, NGV754515H, NGV745988H, NGV743402H, NGV744339H, NGV748168H, NGV749106H, NGV750639H, NGV751032H, NGV751262H, NGV752855H, NGV747022H, NGV742257H, NGV743211H, NGV743217H, NGV743334H, NGV743336H, NGV743987H, NGV743988H, NGV743989H, NGV743992H, NGV743993H, NGV744080H, NGV744081H, NGV744082H, NGV745589H, NGV746536H, NGV747388H, NGV747952H, NGV748806H, NGV749295H, NGV749296H, NGV749884H, NGV750233H, NGV750260H, NGV750852H, NGV751764H, NGV753289H, NGV753336H, NGV753350H, NGV753351H, NGV753832H, NGV742189H, NGV742190H, NGV743530H, NGV743531H, NGV743532H, NGV743533H, NGV746989H, NGV746991H, NGV746992H, NGV746993H, NGV750014H, NGV750018H, NGV750019H, NGV745867H, NGV751592H, NGV752581H, NGV744391H, NGV744392H, NGV744393H, NGV744452H, NGV747533H, NGV747580H, NGV747764H, NGV749803H, NGV750062H, NGV751621H, NGV753081H, NGV742213H, NGV744159H, NGV744160H, NGV745056H, NGV746800H, NGV746801H, NGV746803H, NGV751218H, NGV751219H, NGV752188H, NGV753403H, NGV747520H, NGV747860H, NGV742657H, NGV743466H, NGV743518H, NGV743641H, NGV746662H, NGV747356H, NGV744740H, NGV742268H, NGV742269H, NGV744757H, NGV744857H, NGV745964H, NGV746832H, NGV747012H, NGV747569H, NGV747953H, NGV748351H, NGV748354H, NGV750194H, NGV750980H, NGV751068H, NGV751100H, NGV751406H, NGV751407H, NGV751775H, NGV753323H, NGV753740H, NGV753745H, NGV753893H, NGV754775H, NGV747492H, NGV754493H, NGV742395H, NGV745785H, NGV747884H, NGV742111H, NGV742282H, NGV742803H, NGV744036H, NGV744180H, NGV744234H, NGV744304H, NGV744432H, NGV745595H, NGV746384H, NGV747028H, NGV747224H, NGV747519H, NGV747536H, NGV747911H, NGV747912H, NGV749457H, NGV749458H, NGV749659H, NGV749660H, NGV749994H, NGV749995H, NGV750268H, NGV750271H, NGV750568H, NGV750972H, NGV751291H, NGV751660H, NGV752407H, NGV753886H, NGV753888H, NGV753890H, NGV744742H, NGV750760H, NGV751854H, NGV752357H, NGV751503H, NGV753786H, NGV754457H, NGV742486H, NGV744471H, NGV743622H, NGV746492H, NGV746553H, NGV747332H, NGV747810H, NGV748309H, NGV749096H, NGV750994H, NGV751050H, NGV751057H, NGV751624H, NGV751673H, NGV752026H, NGV755153H, NGV743925H, NGV743926H, NGV746594H, NGV750222H, NGV751759H, NGV751760H, NGV754232H, NGV748387H, NGV746100H, NGV746343H, NGV747511H, NGV747986H, NGV754167H, NGV745303H, NGV747430H, NGV742098H, NGV742508H, NGV744989H, NGV746006H, NGV746423H, NGV746666H, NGV748634H, NGV749102H, NGV750373H, NGV751883H, NGV752533H, NGV752534H, NGV753785H, NGV753003H, NGV745795H, NGV748884H, NGV751233H, NGV751253H, NGV742266H, NGV743191H, NGV743250H, NGV743600H, NGV744578H, NGV744754H, NGV744756H, NGV744792H, NGV746791H, NGV747431H, NGV747653H, NGV748352H, NGV749492H, NGV749819H, NGV750535H, NGV752052H, NGV752053H, NGV753702H, NGV753703H, NGV753741H, NGV748501H, NGV744342H, NGV744344H, NGV746129H, NGV747210H, NGV747211H, NGV747214H, NGV750410H, NGV743389H, NGV743342H, NGV744031H, NGV744659H, NGV744760H, NGV744849H, NGV746456H, NGV747337H, NGV747547H, NGV748849H, NGV748904H, NGV751348H, NGV752360H, NGV753125H, NGV752286H, NGV748104H, NGV745710H, NGV742422H, NGV746313H, NGV746710H, NGV746943H, NGV747088H, NGV748427H, NGV747535H, NGV749691H, NGV751376H, NGV744579H, NGV752976H, NGV743590H, NGV742624H, NGV743558H, NGV747171H, NGV747809H, NGV748086H, NGV748389H, NGV749697H, NGV751046H, NGV753635H, NGV754758H, NGV743519H, NGV744169H, NGV744969H, NGV746633H, NGV746929H, NGV748262H, NGV749127H, NGV750195H, NGV750987H, NGV752429H, NGV753663H, NGV742127H, NGV742130H, NGV744470H, NGV744472H, NGV745409H, NGV746031H, NGV748638H, NGV748639H, NGV748640H, NGV749144H, NGV749145H, NGV749495H, NGV751034H, NGV751035H, NGV746697H, NGV744115H, NGV752208H, NGV744206H, NGV744244H, NGV745138H, NGV745140H, NGV745819H, NGV745824H, NGV746602H, NGV746605H, NGV746836H, NGV747161H, NGV747314H, NGV747315H, NGV747951H, NGV748666H, NGV748940H, NGV749063H, NGV749207H, NGV749482H, NGV749703H, NGV750227H, NGV750228H, NGV750258H, NGV750265H, NGV751785H, NGV751786H, NGV743692H, NGV746046H, NGV746660H, NGV748527H, NGV751505H, NGV746350H, NGV748526H, NGV752332H, NGV741959H, NGV746667H, NGV744905H, NGV747021H, NGV749862H, NGV749997H, NGV751834H, NGV743253H, NGV743803H, NGV744646H, NGV747548H, NGV749181H, NGV753322H, NGV748955H, NGV749619H, NGV750983H, NGV754912H, NGV746663H, NGV749087H, NGV738892H, NGV742658H, NGV748019H, NGV743563H, NGV745826H, NGV745848H, NGV749455H, NGV749603H, NGV750278H, NGV750796H, NGV752272H, NGV742197H, NGV744840H, NGV745959H, NGV747929H, NGV742630H, NGV745192H, NGV747462H, NGV753230H, NGV753365H, NGV742407H, NGV742744H, NGV743337H, NGV743339H, NGV743340H, NGV744828H, NGV744937H, NGV745647H, NGV745836H, NGV746640H, NGV747614H, NGV748384H, NGV749638H, NGV750607H, NGV751049H, NGV751901H, NGV753543H, NGV753674H, NGV753675H, NGV744049H, NGV747783H, NGV742161H, NGV742162H, NGV742164H, NGV742218H, NGV742483H, NGV742583H, NGV742663H, NGV742664H, NGV742745H, NGV743380H, NGV743401H, NGV743423H, NGV743526H, NGV744525H, NGV744526H, NGV744807H, NGV744818H, NGV744819H, NGV745281H, NGV745282H, NGV745283H, NGV745825H, NGV747357H, NGV747360H, NGV748398H, NGV748399H, NGV748400H, NGV748927H, NGV749107H, NGV749601H, NGV749840H, NGV749841H, NGV750012H, NGV750688H, NGV750862H, NGV751067H, NGV751070H, NGV751263H, NGV751264H, NGV751658H, NGV751663H, NGV752257H, NGV752846H, NGV753197H, NGV753201H, NGV753448H, NGV754686H, NGV744518H, NGV745598H, NGV745850H, NGV747537H, NGV749091H, NGV750967H, NGV752151H, NGV752463H, NGV742798H, NGV750165H, NGV744574H, NGV744621H, NGV744623H, NGV749631H, NGV747709H, NGV748024H, NGV741987H, NGV742414H, NGV742415H, NGV743181H, NGV743537H, NGV743539H, NGV743582H, NGV743593H, NGV743922H, NGV743923H, NGV743928H, NGV743929H, NGV743959H, NGV744640H, NGV746378H, NGV746379H, NGV746383H, NGV749018H, NGV749472H, NGV749864H, NGV744423H, NGV747433H, NGV747908H, NGV750136H, NGV742083H, NGV743504H, NGV746044H, NGV747679H, NGV748418H, NGV753955H, NGV742582H, NGV742585H, NGV742757H, NGV746896H, NGV746898H, NGV747013H, NGV747014H, NGV749437H, NGV749438H, NGV750189H, NGV752355H, NGV753137H, NGV754041H, NGV747082H, NGV750151H, NGV742479H, NGV742480H, NGV746461H, NGV748528H, NGV748909H, NGV751578H, NGV753242H, NGV742428H, NGV742429H, NGV742587H, NGV742589H, NGV742759H, NGV743672H, NGV743676H, NGV745265H, NGV746156H, NGV746984H, NGV746985H, NGV748193H, NGV749246H, NGV752576H, NGV745954H, NGV753676H, NGV742045H, NGV742046H, NGV748431H, NGV753659H, NGV751144H, NGV744950H, NGV746844H, NGV748091H, NGV748994H, NGV752784H, NGV744685H, NGV753522H, NGV742525H, NGV745737H, NGV746333H, NGV746335H, NGV746336H, NGV746941H, NGV747024H, NGV749112H, NGV749677H, NGV745181H, NGV746438H, NGV747641H, NGV751650H, NGV746865H, NGV742748H, NGV743422H, NGV750285H, NGV750350H, NGV751071H, NGV751662H, NGV754006H, NGV754542H, NGV747633H, NGV752436H, NGV746655H, NGV748016H, NGV742749H, NGV743511H, NGV743852H, NGV745384H, NGV747291H, NGV748014H, NGV748420H, NGV750346H, NGV751010H, NGV751048H, NGV752337H, NGV753108H, NGV753272H, NGV753279H, NGV753942H, NGV745385H, NGV755389H, NGV744201H, NGV744202H, NGV744203H, NGV744204H, NGV744248H, NGV746537H, NGV746538H, NGV746542H, NGV746543H, NGV746885H, NGV746886H, NGV746887H, NGV746889H, NGV747586H, NGV747587H, NGV748179H, NGV748180H, NGV750625H, NGV750626H, NGV750627H, NGV750628H, NGV753253H, NGV743725H, NGV743417H, NGV743912H, NGV743913H, NGV744075H, NGV746597H, NGV748711H, NGV748713H, NGV748714H, NGV748715H, NGV753469H, NGV743461H, NGV743690H, NGV743935H, NGV743977H, NGV743978H, NGV744253H, NGV744505H, NGV744877H, NGV745013H, NGV745027H, NGV745519H, NGV745768H, NGV745946H, NGV746287H, NGV746465H, NGV747743H, NGV747984H, NGV748088H, NGV748089H, NGV748090H, NGV748157H, NGV748166H, NGV748416H, NGV748906H, NGV749041H, NGV749042H, NGV749044H, NGV749605H, NGV750853H, NGV751288H, NGV752171H, NGV752374H, NGV754085H, NGV754730H, NGV742515H, NGV743947H, NGV752377H, NGV747987H, NGV743149H, NGV747819H, NGV750404H, NGV751515H, NGV751780H, NGV752431H, NGV753282H, NGV753545H, NGV753819H, NGV742567H, NGV743400H, NGV751877H, NGV751880H, NGV752266H, NGV753070H, NGV751073H, NGV751833H, NGV748401H, NGV749134H, NGV753600H, NGV753664H, NGV743706H, NGV744990H, NGV745644H, NGV746643H, NGV747002H, NGV750708H, NGV750710H, NGV750986H, NGV753757H, NGV744096H, NGV744097H, NGV746820H, NGV747114H, NGV754777H, NGV742450H, NGV743261H, NGV747652H, NGV747848H, NGV748231H, NGV745762H, NGV749416H, NGV753537H, NGV755703H, NGV749449H, NGV749801H, NGV754326H, NGV748433H, NGV742254H, NGV743201H, NGV743203H, NGV745650H, NGV745814H, NGV745815H, NGV747940H, NGV747941H, NGV748152H, NGV748496H, NGV748657H, NGV749001H, NGV750824H, NGV752296H, NGV752297H, NGV752927H, NGV746866H, NGV748374H, NGV748919H, NGV748988H, NGV742527H, NGV742528H, NGV742529H, NGV743197H, NGV743840H, NGV743842H, NGV745031H, NGV745235H, NGV746613H, NGV752264H, NGV752265H, NGV753055H, NGV742606H, NGV745262H, NGV746010H, NGV749086H, NGV743512H, NGV745990H, NGV746074H, NGV746341H, NGV746615H, NGV746638H, NGV746839H, NGV747458H, NGV748320H, NGV748322H, NGV748425H, NGV748426H, NGV748672H, NGV752448H, NGV753897H, NGV742399H, NGV745284H, NGV750201H, NGV750388H, NGV754112H, NGV742096H, NGV742551H, NGV742553H, NGV742808H, NGV743830H, NGV744655H, NGV746505H, NGV746862H, NGV747412H, NGV747538H, NGV749674H, NGV751204H, NGV753356H, NGV742628H, NGV753387H, NGV746165H, NGV750546H, NGV750554H, NGV751645H, NGV752316H, NGV754736H, NGV743306H, NGV743307H, NGV743308H, NGV743309H, NGV744007H, NGV744074H, NGV744095H, NGV744251H, NGV746559H, NGV746565H, NGV746792H, NGV746807H, NGV746808H, NGV747089H, NGV747167H, NGV747298H, NGV747534H, NGV747742H, NGV747813H, NGV748161H, NGV748239H, NGV748703H, NGV749028H, NGV749293H, NGV749294H, NGV749701H, NGV749708H, NGV750623H, NGV750756H, NGV751358H, NGV751474H, NGV751605H, NGV751630H, NGV751631H, NGV753344H, NGV753830H, NGV753831H, NGV754638H, NGV754639H, NGV755055H, NGV742768H, NGV745209H, NGV746391H, NGV746628H, NGV747233H, NGV749669H, NGV750157H, NGV742387H, NGV742388H, NGV744473H, NGV745617H, NGV745636H, NGV747353H, NGV747405H, NGV748424H, NGV748517H, NGV751316H, NGV751907H, NGV752321H, NGV753113H, NGV745541H, NGV748652H, NGV743398H, NGV744037H, NGV746720H, NGV746995H, NGV748008H, NGV751271H, NGV751699H, NGV743190H, NGV744750H, NGV745249H, NGV748524H, NGV748953H, NGV750594H, NGV750780H, NGV750984H, NGV751310H, NGV751945H, NGV752162H, NGV753141H, NGV754086H, NGV754797H, NGV742173H, NGV742800H, NGV743566H, NGV744113H, NGV744697H, NGV744699H, NGV744945H, NGV744946H, NGV745966H, NGV747455H, NGV747456H, NGV747457H, NGV749792H, NGV750442H, NGV750444H, NGV750447H, NGV752441H, NGV752442H, NGV754401H, NGV745182H, NGV745216H, NGV749020H, NGV749072H, NGV750745H, NGV753071H, NGV741984H, NGV742250H, NGV742764H, NGV745187H, NGV752404H, NGV752481H, NGV746360H, NGV743220H, NGV750273H, NGV742171H, NGV747918H, NGV750132H, NGV751988H, NGV742169H, NGV742199H, NGV744324H, NGV744325H, NGV745521H, NGV745544H, NGV745545H, NGV746434H, NGV746837H, NGV747582H, NGV749095H, NGV751056H, NGV751394H, NGV754098H, NGV755000H, NGV742124H, NGV742126H, NGV744185H, NGV744186H, NGV745289H, NGV745637H, NGV746451H, NGV746453H, NGV746610H, NGV746611H, NGV746641H, NGV744427H, NGV745796H, NGV747253H, NGV747256H, NGV749110H, NGV750868H, NGV745341H, NGV742297H, NGV744843H, NGV748879H, NGV748894H, NGV749443H, NGV749825H, NGV747780H, NGV749283H, NGV751414H, NGV753771H, NGV744199H, NGV746622H, NGV746624H, NGV744634H, NGV745266H, NGV744409H, NGV747372H, NGV747373H, NGV747374H, NGV747375H, NGV753793H, NGV753795H, NGV743727H, NGV744885H, NGV745528H, NGV749839H, NGV750213H, NGV750215H, NGV753123H, NGV753130H, NGV744767H, NGV747963H, NGV753958H, NGV742792H, NGV743210H, NGV743510H, NGV746816H, NGV747487H, NGV748203H, NGV750703H, NGV742766H, NGV743551H, NGV743628H, NGV747686H, NGV747827H, NGV749490H, NGV751498H, NGV752304H, NGV742514H, NGV744863H, NGV744624H, NGV744919H, NGV747830H, NGV754013H, NGV742484H, NGV742485H, NGV743164H, NGV743198H, NGV744592H, NGV744947H, NGV745081H, NGV746363H, NGV746366H, NGV747429H, NGV747489H, NGV747920H, NGV748163H, NGV748410H, NGV749261H, NGV749678H, NGV750374H, NGV750420H, NGV750691H, NGV750749H, NGV752489H, NGV754721H, NGV748198H, NGV752444H, NGV753320H, NGV751439H, NGV744810H, NGV745189H, NGV746717H, NGV748674H, NGV752206H, NGV753410H, NGV743231H, NGV743279H, NGV744316H, NGV747381H, NGV750029H, NGV744103H, NGV745735H, NGV748249H, NGV744846H, NGV749543H, NGV750747H, NGV755082H, NGV742090H, NGV742091H, NGV742116H, NGV742243H, NGV742244H, NGV742629H, NGV743483H, NGV743541H, NGV743606H, NGV743691H, NGV744580H, NGV744695H, NGV744968H, NGV745051H, NGV745170H, NGV745208H, NGV745718H, NGV746713H, NGV746776H, NGV747494H, NGV747638H, NGV747639H, NGV747790H, NGV747817H, NGV747923H, NGV747957H, NGV748926H, NGV749221H, NGV749577H, NGV750586H, NGV751163H, NGV751165H, NGV751563H, NGV752152H, NGV752190H, NGV752358H, NGV752474H, NGV753347H, NGV753525H, NGV754726H, NGV747001H, NGV751586H, NGV744938H, NGV747916H, NGV749023H, NGV743499H, NGV747514H, NGV751126H, NGV742432H, NGV744829H, NGV744831H, NGV746852H, NGV747551H, NGV748438H, NGV749571H, NGV752008H, NGV752235H, NGV747168H, NGV747169H, NGV747170H, NGV753535H, NGV754533H, NGV742461H, NGV742462H, NGV742463H, NGV742464H, NGV743693H, NGV743694H, NGV743698H, NGV743914H, NGV743915H, NGV743918H, NGV743998H, NGV744263H, NGV744366H, NGV744367H, NGV744368H, NGV744369H, NGV744480H, NGV744484H, NGV744765H, NGV744766H, NGV744833H, NGV744834H, NGV744835H, NGV745155H, NGV745158H, NGV745159H, NGV745880H, NGV746798H, NGV747174H, NGV747175H, NGV747319H, NGV747320H, NGV747488H, NGV747593H, NGV747887H, NGV748695H, NGV748950H, NGV744254H, NGV746849H, NGV744639H, NGV746779H, NGV748903H, NGV748975H, NGV743332H, NGV743282H, NGV744585H, NGV745688H, NGV750541H, NGV751063H, NGV751169H, NGV743280H, NGV743506H, NGV744812H, NGV745360H, NGV745431H, NGV742840H, NGV744111H, NGV745325H, NGV746147H, NGV747249H, NGV748995H, NGV749562H, NGV753993H, NGV742146H, NGV742147H, NGV742472H, NGV742474H, NGV742475H, NGV743346H, NGV743623H, NGV743624H, NGV744098H, NGV744099H, NGV744210H, NGV744252H, NGV744255H, NGV744741H, NGV746696H, NGV747606H, NGV747607H, NGV747608H, NGV748434H, NGV748435H, NGV749253H, NGV749306H, NGV749308H, NGV749353H, NGV750016H, NGV750099H, NGV750428H, NGV750429H, NGV750430H, NGV752177H, NGV752178H, NGV752179H, NGV753133H, NGV753134H, NGV753135H, NGV754025H, NGV747176H, NGV745298H, NGV746591H, NGV747218H, NGV747219H, NGV751392H, NGV751472H, NGV751926H, NGV752705H, NGV753451H, NGV753483H, NGV754323H, NGV743639H, NGV746157H, NGV746276H, NGV752176H, NGV752350H, NGV752354H, NGV742805H, NGV743170H, NGV744817H, NGV745962H, NGV751250H, NGV747910H, NGV755058H, NGV750746H, NGV752684H, NGV746789H, NGV753662H, NGV746883H, NGV747340H, NGV748171H, NGV742141H, NGV742451H, NGV741962H, NGV742560H, NGV745174H, NGV745179H, NGV750006H, NGV744638H, NGV747331H, NGV747399H, NGV749588H, NGV751290H, NGV752417H, NGV748500H, NGV750013H, NGV742647H, NGV744408H, NGV745976H, NGV746463H, NGV747540H, NGV747909H, NGV747928H, NGV747962H, NGV748929H, NGV748956H, NGV749153H, NGV749196H, NGV749611H, NGV751157H, NGV752677H, NGV753656H, NGV753753H, NGV753754H, NGV754306H, NGV742255H, NGV742256H, NGV744083H, NGV745113H, NGV746823H, NGV746824H, NGV750735H, NGV751716H, NGV754256H, NGV754388H, NGV754603H, NGV747352H, NGV748898H, NGV749133H, NGV749356H, NGV749450H, NGV744511H, NGV744879H, NGV742423H, NGV744788H, NGV745372H, NGV746145H, NGV750053H, NGV751162H, NGV754311H, NGV747493H, NGV744780H, NGV745588H, NGV747513H, NGV748514H, NGV745045H, NGV742389H, NGV745375H, NGV753280H, NGV745195H, NGV754707H, NGV743640H, NGV749632H, NGV751905H, NGV747064H, NGV749802H, NGV752919H, NGV745063H, NGV742286H, NGV743311H, NGV743372H, NGV743403H, NGV743414H, NGV743515H, NGV743521H, NGV743527H, NGV743882H, NGV743949H, NGV743950H, NGV745260H, NGV746132H, NGV746133H, NGV746291H, NGV747141H, NGV747143H, NGV747318H, NGV747737H, NGV747738H, NGV748158H, NGV748159H, NGV748704H, NGV749533H, NGV749534H, NGV750166H, NGV750167H, NGV750180H, NGV751037H, NGV751040H, NGV751542H, NGV751713H, NGV751714H, NGV751715H, NGV752707H, NGV752708H, NGV753095H, NGV753454H, NGV742196H, NGV742368H, NGV744146H, NGV744561H, NGV745053H, NGV745408H, NGV745799H, NGV748508H, NGV748682H, NGV748922H, NGV748957H, NGV749484H, NGV751632H, NGV751633H, NGV751791H, NGV752048H, NGV752574H, NGV752575H, NGV753079H, NGV753367H, NGV753420H, NGV754117H, NGV754118H, NGV754231H, NGV755095H, NGV744424H, NGV746944H, NGV748446H, NGV751949H, NGV754009H, NGV741973H, NGV748206H, NGV750288H, NGV745884H, NGV747030H, NGV746724H, NGV751371H, NGV744154H, NGV744455H, NGV746478H, NGV746647H, NGV747187H, NGV747554H, NGV749485H, NGV749892H, NGV750022H, NGV752492H, NGV753823H, NGV753965H, NGV742654H, NGV742655H, NGV744462H, NGV742660H, NGV748680H, NGV742760H, NGV743254H, NGV744573H, NGV744855H, NGV751065H, NGV751072H, NGV751074H, NGV754379H, NGV743503H, NGV743599H, NGV744237H, NGV753968H, NGV749636H, NGV750155H, NGV742665H, NGV747741H, NGV749368H, NGV752279H, NGV752075H, NGV745961H, NGV751874H, NGV742813H, NGV745219H, NGV747678H, NGV748404H, NGV749805H, NGV755008H, NGV754219H, NGV743271H, NGV743443H, NGV745111H, NGV745294H, NGV745597H, NGV747791H, NGV754479H, NGV744029H, NGV742761H, NGV750066H, NGV750069H, NGV742371H, NGV747050H, NGV747530H, NGV747545H, NGV749355H, NGV751565H, NGV752243H, NGV753229H, NGV753382H, NGV742110H, NGV742232H, NGV743174H, NGV743175H, NGV743412H, NGV743413H, NGV743580H, NGV745402H, NGV746658H, NGV751665H, NGV744974H, NGV742364H, NGV742366H, NGV743555H, NGV743556H, NGV743557H, NGV745006H, NGV746001H, NGV747377H, NGV749706H, NGV749929H, NGV750624H, NGV752462H, NGV752763H, NGV753516H, NGV753517H, NGV742535H, NGV748513H, NGV743596H, NGV745782H, NGV742411H, NGV747072H, NGV749548H, NGV744696H, NGV745811H, NGV743168H, NGV742526H, NGV743410H, NGV745083H, NGV750860H, NGV748397H, NGV753496H, NGV743320H, NGV744168H, NGV744930H, NGV745042H, NGV747191H, NGV747192H, NGV748442H, NGV748443H, NGV749105H, NGV742638H, NGV743274H, NGV745046H, NGV747685H, NGV751081H, NGV753709H, NGV744633H, NGV745163H, NGV745390H, NGV755574H, NGV743355H, NGV744259H, NGV748010H, NGV744539H, NGV751625H, NGV754051H, NGV749661H, NGV752155H, NGV744593H, NGV753088H, NGV746689H, NGV746048H, NGV747250H, NGV749115H, NGV751260H, NGV751881H, NGV753357H, NGV744607H, NGV745272H, NGV743467H, NGV746771H, NGV749812H, NGV749952H, NGV755111H, NGV749268H, NGV755642H, NGV742434H, NGV743235H, NGV743404H, NGV743805H, NGV746454H, NGV747452H, NGV747453H, NGV747454H, NGV750423H, NGV748966H, NGV749142H, NGV749617H, NGV750392H, NGV744110H, NGV744419H, NGV742446H, NGV744988H, NGV745821H, NGV751338H, NGV744348H, NGV746818H, NGV748192H, NGV742758H, NGV750870H, NGV745833H, NGV743176H, NGV746637H, NGV742554H, NGV742653H, NGV743300H, NGV743709H, NGV743863H, NGV744907H, NGV745416H, NGV747684H, NGV750188H, NGV753951H, NGV745913H, NGV742795H, NGV748379H, NGV743324H, NGV743657H, NGV745414H, NGV748282H, NGV749177H, NGV753084H, NGV753105H, NGV753364H, NGV742151H, NGV743656H, NGV743838H, NGV744737H, NGV744901H, NGV744904H, NGV745244H, NGV745415H, NGV747053H, NGV747350H, NGV747480H, NGV747776H, NGV747857H, NGV747858H, NGV747861H, NGV748449H, NGV748823H, NGV749427H, NGV749818H, NGV750032H, NGV752662H, NGV753077H, NGV753239H, NGV753526H, NGV753994H, NGV754225H, NGV742537H, NGV743183H, NGV744106H, NGV752282H, NGV752161H, NGV743632H, NGV747403H, NGV754530H, NGV744364H, NGV747145H, NGV751432H, NGV745173H, NGV745175H, NGV745176H, NGV745177H, NGV745178H, NGV753776H, NGV743946H, NGV748103H, NGV744859H, NGV748184H, NGV753439H, NGV753949H, NGV744150H, NGV747597H, NGV745437H, NGV747644H, NGV751776H, NGV751778H, NGV742270H, NGV752317H, NGV743576H, NGV753085H, NGV750536H, NGV753964H, NGV748022H, NGV743825H, NGV744538H, NGV745388H, NGV754305H, NGV744590H, NGV752249H, NGV743234H, NGV745028H, NGV743409H, NGV748911H, NGV749392H, NGV751121H, NGV752766H, NGV752767H, NGV753104H, NGV744166H, NGV744811H, NGV745857H, NGV751398H, NGV751861H, NGV744768H, NGV747049H, NGV749109H, NGV751859H, NGV745849H, NGV754389H, NGV750701H, NGV743919H, NGV743964H, NGV743965H, NGV748406H, NGV748407H, NGV748805H, NGV748899H, NGV749202H, NGV753710H, NGV753711H, NGV744599H, NGV748845H, NGV748070H, NGV747784H, NGV748210H, NGV747046H, NGV747958H, NGV752277H, NGV743439H, NGV744261H, NGV749200H, NGV741947H, NGV742545H, NGV744109H, NGV745682H, NGV751879H, NGV754807H, NGV754944H, NGV746790H, NGV744782H, NGV749179H, NGV751614H, NGV744214H, NGV743611H, NGV743612H, NGV744274H, NGV744276H, NGV744279H, NGV744280H, NGV744282H, NGV744283H, NGV744285H, NGV744287H, NGV746737H, NGV746738H, NGV746739H, NGV746740H, NGV746741H, NGV746748H, NGV746750H, NGV746751H, NGV746752H, NGV746753H, NGV746754H, NGV746757H, NGV746761H, NGV746762H, NGV746764H, NGV746765H, NGV746766H, NGV746767H, NGV751551H, NGV751552H, NGV751553H, NGV751554H, NGV752451H, NGV752452H, NGV742248H, NGV742249H, NGV744799H, NGV744800H, NGV744801H, NGV745920H, NGV745921H, NGV745923H, NGV745925H, NGV745926H, NGV745929H, NGV746743H, NGV746795H, NGV746796H, NGV746957H, NGV746958H, NGV746959H, NGV746961H, NGV746963H, NGV746967H, NGV746969H, NGV746970H, NGV746971H, NGV746973H, NGV746974H, NGV746976H, NGV746978H, NGV746979H, NGV746980H, NGV746981H, NGV747589H, NGV750176H, NGV750177H, NGV750178H, NGV748965H, NGV751855H, NGV747716H, NGV744411H, NGV748067H, NGV742155H, NGV747831H, NGV749494H, NGV741979H, NGV741981H, NGV742396H, NGV742439H, NGV742488H, NGV742489H, NGV742490H, NGV742491H, NGV742493H, NGV742494H, NGV742495H, NGV742496H, NGV742497H, NGV743224H, NGV743238H, NGV743239H, NGV743241H, NGV743243H, NGV743244H, NGV743646H, NGV743647H, NGV743648H, NGV743649H, NGV743873H, NGV743875H, NGV743877H, NGV743878H, NGV744692H, NGV745971H, NGV745972H, NGV745973H, NGV746730H, NGV746731H, NGV746734H, NGV746735H, NGV746736H, NGV746904H, NGV746906H, NGV746907H, NGV746908H, NGV746909H, NGV746910H, NGV746911H, NGV746912H, NGV746913H, NGV746914H, NGV746916H, NGV746917H, NGV746919H, NGV746920H, NGV746921H, NGV746922H, NGV746923H, NGV746924H, NGV747464H, NGV747465H, NGV747467H, NGV747468H, NGV747469H, NGV747470H, NGV747471H, NGV747472H, NGV747473H, NGV747868H, NGV747890H, NGV747925H, NGV748727H, NGV749711H, NGV749712H, NGV749716H, NGV749717H, NGV749719H, NGV749836H, NGV749867H, NGV749868H, NGV749870H, NGV749872H, NGV749873H, NGV749875H, NGV749878H, NGV749879H, NGV750481H, NGV750483H, NGV750485H, NGV750492H, NGV751274H, NGV751284H, NGV751693H, NGV751694H, NGV751695H, NGV751727H, NGV751731H, NGV751733H, NGV751735H, NGV751736H, NGV751737H, NGV752065H, NGV752066H, NGV752067H, NGV752068H, NGV752389H, NGV752456H, NGV752459H, NGV752460H, NGV753196H, NGV743227H, NGV743948H, NGV745267H, NGV745268H, NGV745788H, NGV746093H, NGV748218H, NGV748219H, NGV749092H, NGV749387H, NGV749831H, NGV751788H, NGV751876H, NGV744421H, NGV746822H, NGV749252H, NGV753453H, NGV744882H, NGV745789H, NGV751408H, NGV751590H, NGV742739H, NGV743478H, NGV744153H, NGV744178H, NGV744790H, NGV744821H, NGV745394H, NGV746300H, NGV746617H, NGV746799H, NGV747015H, NGV747815H, NGV748167H, NGV748169H, NGV750007H, NGV751858H, NGV753967H, NGV754381H, NGV746646H, NGV753966H, NGV743477H, NGV745542H, NGV746619H, NGV748515H, NGV749117H, NGV749989H, NGV751682H, NGV752251H, NGV752446H, NGV749271H, NGV742477H, NGV751241H, NGV743634H, NGV743636H, NGV744679H, NGV744681H, NGV744735H, NGV745025H, NGV745532H, NGV745534H, NGV746945H, NGV748445H, NGV749118H, NGV749119H, NGV751207H, NGV751213H, NGV751950H, NGV752438H, NGV754008H, NGV749093H, NGV749832H, NGV747017H, NGV747231H, NGV747257H, NGV754522H, NGV746709H, NGV742078H, NGV742079H, NGV742513H, NGV742625H, NGV742662H, NGV744789H, NGV747787H, NGV748516H, NGV749888H, NGV750742H, NGV751545H, NGV751816H, NGV752922H, NGV754001H, NGV749568H, NGV744744H, NGV744745H, NGV745004H, NGV747563H, NGV743396H, NGV743272H, NGV744078H, NGV744438H, NGV743142H, NGV743507H, NGV746125H, NGV746843H, NGV746873H, NGV747988H, NGV748156H, NGV748925H, NGV749574H, NGV749578H, NGV750694H, NGV751620H, NGV747227H, NGV748085H, NGV747402H, NGV742634H, NGV742659H, NGV749166H, NGV754725H, NGV742521H, NGV751363H, NGV744262H, NGV746856H, NGV748185H, NGV748186H, NGV752246H, NGV747654H, NGV747655H, NGV748026H, NGV748027H, NGV749985H, NGV749986H, NGV750293H, NGV750294H, NGV750340H, NGV750343H, NGV750432H, NGV750433H, NGV750435H, NGV751671H, NGV753340H, NGV753462H, NGV753464H, NGV753474H, NGV753476H, NGV749496H, NGV742604H, NGV743937H, NGV751642H, NGV753571H, NGV744635H, NGV745397H, NGV751873H, NGV742556H, NGV742557H, NGV744337H, NGV744338H, NGV747532H, NGV748172H, NGV748202H, NGV750054H, NGV749152H, NGV743637H, NGV746946H, NGV743564H, NGV743670H, NGV744929H, NGV743441H, NGV743650H, NGV748033H, NGV748057H, NGV749371H, NGV742828H, NGV743446H, NGV743474H, NGV744577H, NGV745379H, NGV745395H, NGV746533H, NGV747834H, NGV748396H, NGV748896H, NGV748982H, NGV748984H, NGV749648H, NGV751589H, NGV752043H, NGV752166H, NGV753782H, NGV744836H, NGV745881H, NGV746470H, NGV753511H, NGV750211H, NGV750361H, NGV742455H, NGV742456H, NGV743710H, NGV746932H, NGV746933H, NGV746934H, NGV746935H, NGV747309H, NGV747310H, NGV747312H, NGV750634H, NGV747604H, NGV751477H, NGV743604H, NGV747553H, NGV751771H, NGV751772H, NGV752494H, NGV742433H, NGV743608H, NGV744474H, NGV749849H, NGV744527H, NGV744528H, NGV747086H, NGV747475H, NGV747605H, NGV751810H, NGV751887H, NGV753984H, NGV746609H, NGV751931H, NGV743392H, NGV747959H, NGV751415H, NGV754455H, NGV744265H, NGV743327H, NGV745940H, NGV747379H, NGV747380H, NGV754194H, NGV745301H, NGV747771H, NGV750762H, NGV745944H, NGV747567H, NGV743395H, NGV746871H, NGV743430H, NGV743629H, NGV744001H, NGV744090H, NGV744091H, NGV744190H, NGV746535H, NGV746900H, NGV746901H, NGV747439H, NGV748068H, NGV748689H, NGV748710H, NGV748719H, NGV749038H, NGV749272H, NGV750326H, NGV750371H, NGV750399H, NGV750400H, NGV750990H, NGV751685H, NGV751781H, NGV753319H, NGV753412H, NGV753471H, NGV754244H, NGV742588H, NGV743131H, NGV745264H, NGV746281H, NGV746285H, NGV747969H, NGV747970H, NGV748195H, NGV749843H, NGV751015H, NGV751596H, NGV751597H, NGV752209H, NGV752412H, NGV753532H, NGV753533H, NGV753534H, NGV754537H, NGV743405H, NGV742791H, NGV751312H, NGV744266H, NGV747011H, NGV748017H, NGV748447H, NGV742796H, NGV743419H, NGV743450H, NGV743730H, NGV744596H, NGV745252H, NGV745864H, NGV745870H, NGV746699H, NGV746834H, NGV747835H, NGV748895H, NGV749047H, NGV749814H, NGV750956H, NGV752608H, NGV753100H, NGV753385H, NGV745044H, NGV745784H, NGV748402H, NGV749360H, NGV753396H, NGV750614H, NGV745635H, NGV752318H, NGV746782H, NGV744171H, NGV748736H, NGV750726H, NGV747254H, NGV747255H, NGV745387H, NGV750693H, NGV749407H, NGV750690H, NGV747701H, NGV751130H, NGV753878H, NGV747565H, NGV753326H, NGV746295H, NGV746723H, NGV747680H, NGV747735H, NGV746692H, NGV749498H, NGV747007H, NGV746063H, NGV747566H, NGV753821H, NGV747966H, NGV751446H, NGV742182H, NGV742651H, NGV742652H, NGV743870H, NGV744331H, NGV744587H, NGV744588H, NGV746629H, NGV747345H, NGV748177H, NGV749141H, NGV751090H, NGV751319H, NGV751547H, NGV751548H, NGV751583H, NGV752057H, NGV753381H, NGV743298H, NGV751441H, NGV748096H, NGV750863H, NGV742640H, NGV742641H, NGV743804H, NGV744231H, NGV747704H, NGV746701H, NGV752578H, NGV747775H, NGV753612H, NGV747035H, NGV747059H, NGV750608H, NGV747571H, NGV747883H, NGV749538H, NGV749906H, NGV751012H, NGV754688H, NGV742793H, NGV743597H, NGV744357H, NGV744358H, NGV745422H, NGV745423H, NGV745993H, NGV745994H, NGV745995H, NGV746181H, NGV747713H, NGV747714H, NGV747937H, NGV747974H, NGV749013H, NGV749015H, NGV749687H, NGV751116H, NGV751117H, NGV751251H, NGV751275H, NGV752270H, NGV752295H, NGV752403H, NGV752765H, NGV745628H, NGV747406H, NGV743225H, NGV753646H, NGV742382H, NGV743673H, NGV743328H, NGV744156H, NGV744535H, NGV746680H, NGV749381H, NGV749833H, NGV750140H, NGV743681H, NGV744034H, NGV744662H, NGV747398H, NGV748170H, NGV751648H, NGV746998H, NGV747204H, NGV747205H, NGV747483H, NGV747484H, NGV747485H, NGV747486H, NGV750641H, NGV750647H, NGV751014H, NGV753222H, NGV753789H, NGV753790H, NGV753791H, NGV744693H, NGV743689H, NGV744884H, NGV745952H, NGV747611H, NGV749842H, NGV749830H, NGV746084H, NGV744613H, NGV746101H, NGV746830H, NGV750283H, NGV746427H, NGV742602H, NGV745610H, NGV746148H, NGV747558H, NGV748883H, NGV742540H, NGV743209H, NGV744489H, NGV744490H, NGV744492H, NGV744494H, NGV744496H, NGV744498H, NGV750109H, NGV750456H, NGV750457H, NGV750458H, NGV750462H, NGV750464H, NGV750465H, NGV750466H, NGV750467H, NGV750472H, NGV750484H, NGV750488H, NGV753863H, NGV753867H, NGV751124H, NGV742049H, NGV745243H, NGV745530H, NGV749429H, NGV750874H, NGV753443H, NGV743502H, NGV744779H, NGV741942H, NGV744784H, NGV748392H, NGV748393H, NGV749126H, NGV751318H, NGV747518H, NGV750833H, NGV751478H, NGV742274H, NGV746069H, NGV750225H, NGV754011H, NGV743375H, NGV745366H, NGV745531H, NGV746501H, NGV749675H, NGV753343H, NGV746469H, NGV747173H, NGV748422H, NGV750364H, NGV744322H, NGV749838H, NGV750389H, NGV750391H, NGV742236H, NGV743543H, NGV743920H, NGV743930H, NGV743960H, NGV743961H, NGV743962H, NGV744187H, NGV744188H, NGV744189H, NGV749128H, NGV750979H, NGV751684H, NGV751687H, NGV752701H, NGV753824H, NGV747594H, NGV750363H, NGV750725H, NGV744395H, NGV748021H, NGV748972H, NGV751569H, NGV742492H, NGV744881H, NGV744915H, NGV745218H, NGV746669H, NGV746878H, NGV749150H, NGV749958H, NGV750008H, NGV752172H, NGV741954H, NGV742187H, NGV742188H, NGV742408H, NGV742409H, NGV743384H, NGV743385H, NGV743386H, NGV743452H, NGV743453H, NGV743540H, NGV744164H, NGV744433H, NGV744487H, NGV744605H, NGV744606H, NGV744666H, NGV744667H, NGV744668H, NGV745652H, NGV746644H, NGV746679H, NGV746729H, NGV747237H, NGV747418H, NGV747477H, NGV747478H, NGV749170H, NGV749172H, NGV749434H, NGV749645H, NGV750680H, NGV750681H, NGV750711H, NGV750712H, NGV750713H, NGV750807H, NGV750808H, NGV753894H, NGV753895H, NGV753896H, NGV744067H, NGV744068H, NGV744069H, NGV744070H, NGV746401H, NGV746846H, NGV747351H, NGV748105H, NGV748106H, NGV749131H, NGV749367H, NGV753989H, NGV746072H, NGV746342H, NGV747459H, NGV748321H, NGV748673H, NGV753683H, NGV747688H, NGV743314H, NGV745997H, NGV750335H, NGV754338H, NGV742050H, NGV743366H, NGV743367H, NGV743522H, NGV744170H, NGV744425H, NGV744426H, NGV744556H, NGV745393H, NGV745787H, NGV746020H, NGV746023H, NGV746634H, NGV747555H, NGV747936H, NGV749453H, NGV749786H, NGV749788H, NGV751315H, NGV751436H, NGV753724H, NGV754211H, NGV754485H, NGV742219H, NGV743216H, NGV743879H, NGV752320H, NGV750259H, NGV750439H, NGV744512H, NGV744649H, NGV746026H, NGV753103H, NGV745074H, NGV746715H, NGV747552H, NGV748018H, NGV748959H, NGV750126H, NGV750130H, NGV750146H, NGV750148H, NGV751088H, NGV751089H, NGV751707H, NGV751710H, NGV754160H, NGV745466H, NGV743189H, NGV750415H, NGV742590H, NGV742591H, NGV742598H, NGV743425H, NGV744071H, NGV746595H, NGV747147H, NGV748943H, NGV749943H, NGV750387H, NGV750620H, NGV750621H, NGV751038H, NGV752774H, NGV744349H, NGV746546H, NGV747341H, NGV749168H, NGV750613H, NGV750812H, NGV750848H, NGV751438H, NGV751761H, NGV752069H, NGV752699H, NGV753450H, NGV744762H, NGV750337H, NGV752148H, NGV745527H, NGV745011H, NGV745950H, NGV746704H, NGV749187H, NGV753395H, NGV744595H, NGV742159H, NGV742593H, NGV747942H, NGV747945H, NGV749628H, NGV747687H, NGV749823H, NGV749854H, NGV751416H, NGV742444H, NGV743524H, NGV744862H, NGV746781H, NGV747051H, NGV747584H, NGV749693H, NGV750214H, NGV742247H, NGV743360H, NGV742650H, NGV742802H, NGV744147H, NGV752373H, NGV754812H, NGV741926H, NGV741927H, NGV741929H, NGV744644H, NGV749211H, NGV743508H, NGV751308H, NGV742592H, NGV742645H, NGV743156H, NGV744637H, NGV744700H, NGV745935H, NGV746142H, NGV746676H, NGV746778H, NGV747463H, NGV747833H, NGV748071H, NGV751770H, NGV743704H, NGV744975H, NGV743146H, NGV748187H, NGV747891H, NGV752424H, NGV752482H, NGV745844H, NGV752163H, NGV744557H, NGV746425H, NGV743567H, NGV744601H, NGV747407H, NGV751486H, NGV743844H, NGV750111H, NGV748029H, NGV746652H, NGV742427H, NGV743347H, NGV746358H, NGV748220H, NGV748848H, NGV750841H, NGV751615H, NGV752059H, NGV745287H, NGV742107H, NGV749286H, NGV750210H, NGV751002H, NGV751118H, NGV751119H, NGV752331H, NGV752683H, NGV753292H, NGV754501H, NGV755031H, NGV747601H, NGV753114H, NGV743605H, NGV747804H, NGV749486H, NGV749890H, NGV751571H, NGV752490H, NGV754727H, NGV754734H, NGV751603H, NGV744893H, NGV744956H, NGV745523H, NGV746665H, NGV747781H, NGV748632H, NGV747602H, NGV744827H, NGV744934H, NGV747019H, NGV746380H, NGV748512H, NGV744558H, NGV746523H, NGV750205H, NGV750702H, NGV753907H, NGV742185H, NGV744107H, NGV745853H, NGV746608H, NGV747939H, NGV748857H, NGV749856H, NGV749861H, NGV750976H, NGV753144H, NGV741933H, NGV741934H, NGV741935H, NGV741989H, NGV741990H, NGV741992H, NGV741997H, NGV742002H, NGV742012H, NGV742015H, NGV742022H, NGV742258H, NGV742260H, NGV742261H, NGV742262H, NGV742263H, NGV742264H, NGV742265H, NGV742287H, NGV742288H, NGV742289H, NGV742290H, NGV742291H, NGV742292H, NGV743286H, NGV743287H, NGV743288H, NGV743289H, NGV743290H, NGV743291H, NGV743292H, NGV743293H, NGV743294H, NGV743295H, NGV743296H, NGV743297H, NGV745304H, NGV745305H, NGV745306H, NGV745308H, NGV745309H, NGV745310H, NGV745318H, NGV745470H, NGV745471H, NGV745510H, NGV745511H, NGV745512H, NGV745513H, NGV745514H, NGV745515H, NGV745516H, NGV746408H, NGV746409H, NGV746414H, NGV746415H, NGV746417H, NGV746809H, NGV746810H, NGV746813H, NGV746814H, NGV747851H, NGV747862H, NGV747863H, NGV747870H, NGV747877H, NGV747878H, NGV747879H, NGV747880H, NGV748486H, NGV748487H, NGV748488H, NGV748489H, NGV748490H, NGV748491H, NGV748492H, NGV749154H, NGV749155H, NGV749156H, NGV749158H, NGV749160H, NGV749463H, NGV749464H, NGV749465H, NGV749466H, NGV749467H, NGV749468H, NGV751004H, NGV751017H, NGV751019H, NGV751020H, NGV751021H, NGV751022H, NGV751023H, NGV751024H, NGV751025H, NGV751026H, NGV751451H, NGV751459H, NGV751462H, NGV751463H, NGV751468H, NGV751572H, NGV751717H, NGV751718H, NGV751719H, NGV751721H, NGV751722H, NGV751723H, NGV752598H, NGV752603H, NGV752604H, NGV744678H, NGV742741H, NGV750606H, NGV746636H, NGV747592H, NGV751243H, NGV742212H, NGV742697H, NGV743975H, NGV745621H, NGV746286H, NGV746864H, NGV746931H, NGV746955H, NGV747575H, NGV747947H, NGV748242H, NGV748246H, NGV749357H, NGV749358H, NGV749620H, NGV749621H, NGV750722H, NGV743177H, NGV744363H, NGV745655H, NGV745968H, NGV746777H, NGV747631H, NGV747739H, NGV747832H, NGV749612H, NGV750405H, NGV751083H, NGV751790H, NGV752207H, NGV753400H, NGV753943H, NGV755480H, NGV753947H, NGV742482H, NGV745157H, NGV748536H, NGV745643H, NGV742372H, NGV752164H, NGV745096H, NGV747573H, NGV745540H, NGV748409H, NGV752165H, NGV752339H, NGV743475H, NGV749657H, NGV744791H, NGV744820H, NGV743473H, NGV744651H, NGV745377H, NGV745378H, NGV749671H, NGV747355H, NGV748890H, NGV750727H, NGV754452H, NGV749076H, NGV747875H, NGV748841H, NGV746631H, NGV748656H, NGV743851H, NGV745879H, NGV746459H, NGV746681H, NGV747785H, NGV748181H, NGV751123H, NGV751342H, NGV751343H, NGV752220H, NGV752308H, NGV753510H, NGV754195H, NGV754520H, NGV741937H, NGV742424H, NGV745320H, NGV747370H, NGV747616H, NGV748100H, NGV749481H, NGV749544H, NGV750199H, NGV750743H, NGV751013H, NGV753251H, NGV742740H, NGV742822H, NGV745949H, NGV748914H, NGV750873H, NGV753209H, NGV743631H, NGV749173H, NGV742271H, NGV742398H, NGV742410H, NGV742797H, NGV744126H, NGV744127H, NGV744128H, NGV744151H, NGV744152H, NGV744319H, NGV744541H, NGV744794H, NGV744874H, NGV744875H, NGV745883H, NGV745943H, NGV746127H, NGV747207H, NGV747769H, NGV747892H, NGV747926H, NGV747927H, NGV747930H, NGV747931H, NGV748329H, NGV748355H, NGV748851H, NGV748852H, NGV749546H, NGV749566H, NGV749614H, NGV750063H, NGV750414H, NGV750609H, NGV750724H, NGV751051H, NGV751131H, NGV751846H, NGV751867H, NGV752469H, NGV752470H, NGV753666H, NGV753686H, NGV753781H, NGV753844H, NGV754065H, NGV754074H, NGV754114H, NGV754239H, NGV754496H, NGV754792H, NGV754793H, NGV746497H, NGV743374H, NGV743377H, NGV743610H, NGV744503H, NGV746134H, NGV752333H, NGV752535H, NGV753371H, NGV742214H, NGV748097H, NGV751134H, NGV745069H, NGV750393H, NGV747497H, NGV747498H, NGV747500H, NGV754166H, NGV746744H, NGV746745H, NGV746747H, NGV747158H, NGV747794H, NGV753316H, NGV745220H, NGV751544H, NGV751953H, NGV753221H, NGV754010H, NGV747842H, NGV743544H, NGV748738H, NGV751779H, NGV754900H, NGV744739H, NGV749265H, NGV747971H, NGV749263H, NGV742054H, NGV746298H, NGV747010H, NGV752782H, NGV744158H, NGV744860H, NGV745058H, NGV745060H, NGV745061H, NGV745620H, NGV749238H, NGV749241H, NGV751140H, NGV751998H, NGV752367H, NGV752773H, NGV753969H, NGV742200H, NGV742209H, NGV745918H, NGV746322H, NGV746325H, NGV746329H, NGV749855H, NGV750000H, NGV751268H, NGV753997H, NGV754216H, NGV744867H, NGV745222H, NGV749668H, NGV745822H, NGV751272H, NGV745914H, NGV746694H, NGV745200H, NGV747946H, NGV745274H, NGV742458H, NGV744108H, NGV744136H, NGV749189H, NGV745164H, NGV747767H, NGV743680H, NGV743682H, NGV743683H, NGV743686H, NGV743688H, NGV743696H, NGV744345H, NGV747238H, NGV747240H, NGV747241H, NGV747242H, NGV747243H, NGV753874H, NGV753875H, NGV742069H, NGV743933H, NGV744845H, NGV745524H, NGV746007H, NGV747003H, NGV747004H, NGV749130H, NGV749248H, NGV749444H, NGV749613H, NGV751106H, NGV753778H, NGV753779H, NGV754518H, NGV754843H, NGV755035H, NGV745419H, NGV742222H, NGV742223H, NGV742224H, NGV742225H, NGV742226H, NGV742228H, NGV742229H, NGV742230H, NGV742569H, NGV742572H, NGV742573H, NGV742574H, NGV742576H, NGV742577H, NGV742578H, NGV742579H, NGV742753H, NGV742754H, NGV742829H, NGV742832H, NGV742835H, NGV742836H, NGV743344H, NGV743865H, NGV743866H, NGV743867H, NGV744052H, NGV744065H, NGV744482H, NGV744506H, NGV744507H, NGV744508H, NGV744686H, NGV744688H, NGV744689H, NGV744690H, NGV744916H, NGV745065H, NGV745067H, NGV745202H, NGV745203H, NGV745204H, NGV745206H, NGV745403H, NGV745404H, NGV745405H, NGV745406H, NGV745667H, NGV745694H, NGV745805H, NGV745806H, NGV745807H, NGV745808H, NGV745809H, NGV745810H, NGV745977H, NGV745979H, NGV745980H, NGV745981H, NGV746065H, NGV746337H, NGV746338H, NGV746339H, NGV746340H, NGV746437H, NGV746784H, NGV746785H, NGV746786H, NGV746787H, NGV746788H, NGV747074H, NGV747075H, NGV747076H, NGV747077H, NGV747079H, NGV747080H, NGV747216H, NGV747435H, NGV747436H, NGV747441H, NGV747442H, NGV747443H, NGV747444H, NGV747445H, NGV747446H, NGV747447H, NGV747448H, NGV747577H, NGV747578H, NGV747691H, NGV747692H, NGV747693H, NGV747694H, NGV747696H, NGV747697H, NGV747698H, NGV747699H, NGV747797H, NGV747898H, NGV747899H, NGV747900H, NGV747901H, NGV747902H, NGV747903H, NGV748078H, NGV748079H, NGV748080H, NGV748081H, NGV748082H, NGV748083H, NGV748284H, NGV748287H, NGV748292H, NGV748302H, NGV748305H, NGV748430H, NGV748864H, NGV748870H, NGV748871H, NGV748875H, NGV748913H, NGV748915H, NGV748916H, NGV748917H, NGV748946H, NGV748980H, NGV749274H, NGV749275H, NGV749276H, NGV749373H, NGV749634H, NGV749679H, NGV749726H, NGV749728H, NGV749730H, NGV749785H, NGV749787H, NGV749790H, NGV749811H, NGV750080H, NGV750086H, NGV750088H, NGV750091H, NGV750352H, NGV750353H, NGV750354H, NGV750355H, NGV750356H, NGV750357H, NGV750358H, NGV750359H, NGV750360H, NGV750991H, NGV750993H, NGV750995H, NGV750996H, NGV750999H, NGV751000H, NGV751001H, NGV751137H, NGV751292H, NGV751294H, NGV751295H, NGV751297H, NGV751298H, NGV751321H, NGV751390H, NGV751400H, NGV751600H, NGV751923H, NGV752014H, NGV752016H, NGV752024H, NGV752038H, NGV752039H, NGV752158H, NGV752196H, NGV752254H, NGV752294H, NGV752313H, NGV752314H, NGV752315H, NGV752345H, NGV752347H, NGV752349H, NGV752496H, NGV752582H, NGV752583H, NGV752756H, NGV752903H, NGV752904H, NGV752905H, NGV753004H, NGV753042H, NGV753043H, NGV753046H, NGV753109H, NGV753110H, NGV753260H, NGV753261H, NGV753264H, NGV753527H, NGV753528H, NGV753529H, NGV753530H, NGV753531H, NGV753669H, NGV753672H, NGV753971H, NGV753972H, NGV753973H, NGV753977H, NGV753978H, NGV753979H, NGV753981H, NGV754012H, NGV754020H, NGV754022H, NGV754384H, NGV754385H, NGV754387H, NGV754540H, NGV754541H, NGV754543H, NGV754544H, NGV754696H, NGV754799H, NGV754800H, NGV755119H, NGV755121H, NGV755122H, NGV755123H, NGV755608H, NGV755610H, NGV741938H, NGV743321H, NGV744816H, NGV746149H, NGV746999H, NGV747163H, NGV748098H, NGV748390H, NGV748654H, NGV749415H, NGV749422H, NGV749990H, NGV751256H, NGV751259H, NGV751402H, NGV751403H, NGV751422H, NGV751423H, NGV751427H, NGV753377H, NGV753603H, NGV753615H, NGV744830H, NGV751244H, NGV751999H, NGV744140H, NGV744141H, NGV744145H, NGV745798H, NGV750418H, NGV751143H, NGV753678H, NGV745947H, NGV750982H, NGV741952H, NGV744437H, NGV744488H, NGV749171H, NGV749656H, NGV750709H, NGV751279H, NGV743319H, NGV748636H, NGV748992H, NGV751417H, NGV753327H, NGV755099H, NGV742172H, NGV743583H, NGV744179H, NGV743626H, NGV743731H, NGV745191H, NGV742452H, NGV745742H, NGV744439H, NGV744440H, NGV744444H, NGV744824H, NGV745165H, NGV745180H, NGV743609H, NGV743957H, NGV744625H, NGV744626H, NGV745917H, NGV746947H, NGV747591H, NGV749239H, NGV749491H, NGV750965H, NGV753617H, NGV746029H, NGV746721H, NGV754580H, NGV744918H, NGV747005H, NGV743299H, NGV744602H, NGV755013H, NGV750687H, NGV751190H, NGV744632H, NGV744900H, NGV746618H, NGV749175H, NGV752471H, NGV754881H, NGV753369H, NGV748107H, NGV742519H, NGV744520H, NGV745135H, NGV745930H, NGV745931H, NGV746940H, NGV751230H, NGV744122H, NGV748183H, NGV744100H, NGV744776H, NGV744847H, NGV745245H, NGV747420H, NGV747460H, NGV747482H, NGV748006H, NGV748155H, NGV750659H, NGV754551H, NGV754577H, NGV743715H, NGV745410H, NGV747040H, NGV747865H, NGV747964H, NGV750219H, NGV752592H, NGV753315H, NGV754023H, NGV742131H, NGV742133H, NGV742251H, NGV742294H, NGV743269H, NGV743333H, NGV743416H, NGV743528H, NGV743534H, NGV743714H, NGV743837H, NGV744359H, NGV744853H, NGV745188H, NGV745278H, NGV745535H, NGV745623H, NGV745877H, NGV747177H, NGV747288H, NGV747289H, NGV748217H, NGV749257H, NGV750406H, NGV750787H, NGV750791H, NGV751161H, NGV751352H, NGV751353H, NGV751354H, NGV751555H, NGV751609H, NGV751610H, NGV751825H, NGV752365H, NGV752447H, NGV753684H, NGV753762H, NGV742208H, NGV746474H, NGV748015H, NGV753509H, NGV752690H, NGV749941H, NGV750185H, NGV750186H, NGV753670H, NGV747747H, NGV749065H, NGV750822H, NGV742437H, NGV742614H, NGV747719H, NGV746714H, NGV747084H, NGV749297H, NGV747283H, NGV753268H, NGV742635H, NGV743437H, NGV743440H, NGV744481H, NGV746857H, NGV747777H, NGV748633H, NGV749304H, NGV749609H, NGV752168H, NGV754819H, NGV743665H, NGV743849H, NGV744181H, NGV744182H, NGV744183H, NGV744184H, NGV745639H, NGV747501H, NGV747502H, NGV747503H, NGV747504H, NGV747506H, NGV747507H, NGV747508H, NGV747510H, NGV748507H, NGV750266H, NGV750267H, NGV750269H, NGV750270H, NGV750274H, NGV750275H, NGV752853H, NGV753725H, NGV753730H, NGV753731H, NGV753732H, NGV745104H, NGV742648H, NGV743476H, NGV745706H, NGV747045H, NGV747085H, NGV749135H, NGV752688H, NGV754524H, NGV747043H, NGV743713H, NGV743864H, NGV746372H, NGV746645H, NGV753599H, NGV744509H, NGV746838H, NGV752753H, NGV746851H, NGV743388H, NGV744175H, NGV744540H, NGV745963H, NGV748012H, NGV748013H, NGV744546H, NGV746068H, NGV749682H, NGV746349H, NGV749029H, NGV744532H, NGV753513H, NGV742390H, NGV754210H, NGV742431H, NGV744832H, NGV745059H, NGV746850H, NGV742457H, NGV742842H, NGV745039H, NGV750718H, NGV750719H, NGV752405H, NGV747954H, NGV747955H, NGV747061H, NGV752783H, NGV747826H, NGV750031H, NGV752397H, NGV742445H, NGV743621H, NGV745389H, NGV746392H, NGV748087H, NGV749372H, NGV749633H, NGV751941H, NGV752757H, NGV754973H, NGV742467H, NGV743379H, NGV745832H, NGV746443H, NGV746445H, NGV746842H, NGV751607H, NGV751659H, NGV752256H, NGV752406H, NGV743968H, NGV743969H, NGV743972H, NGV743973H, NGV743974H, NGV744814H, NGV744815H, NGV745792H, NGV746965H, NGV747070H, NGV747071H, NGV747259H, NGV747262H, NGV747263H, NGV747265H, NGV747266H, NGV747267H, NGV747268H, NGV747649H, NGV747651H, NGV747751H, NGV747752H, NGV748229H, NGV748232H, NGV748234H, NGV748238H, NGV748886H, NGV748888H, NGV748889H, NGV749731H, NGV749732H, NGV749737H, NGV749746H, NGV749748H, NGV750886H, NGV750887H, NGV750889H, NGV750890H, NGV750891H, NGV750892H, NGV750893H, NGV752115H, NGV752116H, NGV752117H, NGV752649H, NGV753300H, NGV753302H, NGV753305H, NGV746635H, NGV744420H, NGV744422H, NGV745026H, NGV745533H, NGV747251H, NGV747252H, NGV748893H, NGV748931H, NGV749366H, NGV749391H, NGV750394H, NGV750395H, NGV751211H, NGV751216H, NGV751946H, NGV752437H, NGV753649H, NGV753651H, NGV753833H, NGV753919H, NGV745413H, NGV745420H, NGV749123H, NGV747195H, NGV746399H, NGV752182H, NGV745411H, NGV748381H, NGV745017H, NGV744267H, NGV744272H, NGV744854H, NGV745529H, NGV747539H, NGV749594H, NGV753394H, NGV749901H, NGV745052H, NGV746691H, NGV746373H, NGV747647H, NGV752526H, NGV742068H, NGV743678H, NGV743861H, NGV742642H, NGV744138H, NGV744163H, NGV744464H, NGV744977H, NGV745285H, NGV746053H, NGV748880H, NGV749088H, NGV749378H, NGV749821H, NGV752584H, NGV752907H, NGV753298H, NGV753598H, NGV743470H, NGV742504H, NGV745412H, NGV747598H, NGV749383H, NGV749479H, NGV751448H, NGV754945H, NGV749264H, NGV743267H, NGV747006H, NGV747339H, NGV754855H, NGV754927H, NGV742621H, NGV742622H, NGV744414H, NGV746377H, NGV746698H, NGV748853H, NGV751307H, NGV750002H, NGV743152H, NGV744333H, NGV745876H, NGV752054H, NGV752216H, NGV742481H, NGV743485H, NGV748153H, NGV749615H, NGV749688H, NGV748429H, NGV744896H, NGV742790H, NGV744132H, NGV745231H, NGV745232H, NGV747474H, NGV748036H, NGV750697H, NGV750698H, NGV751311H, NGV753390H, NGV742275H, NGV742277H, NGV742661H, NGV743302H, NGV743303H, NGV752416H, NGV754902H, NGV746861H, NGV753102H, NGV743486H, NGV744469H, NGV744491H, NGV746353H, NGV746394H, NGV746627H, NGV748482H, NGV748483H, NGV748484H, NGV748756H, NGV748760H, NGV748769H, NGV748785H, NGV743633H, NGV745711H, NGV749837H, NGV742394H, NGV742413H, NGV747634H, NGV748357H, NGV748850H, NGV749164H, NGV749279H, NGV745522H, NGV745651H, NGV746651H, NGV747196H, NGV747702H, NGV749565H, NGV750820H, NGV751652H, NGV752289H, NGV752290H, NGV753094H, NGV753540H, NGV753813H, NGV745871H, NGV745996H, NGV746189H, NGV749245H, NGV749794H, NGV751287H, NGV752114H, NGV751558H, NGV747092H, NGV747583H, NGV750706H, NGV752400H, NGV753921H, NGV749380H, NGV742958H, NGV753248H, NGV753386H, NGV744256H, NGV744476H, NGV744478H, NGV746854H, NGV749476H, NGV749477H, NGV749478H, NGV749608H, NGV749950H, NGV752184H, NGV752244H, NGV753523H, NGV742801H, NGV744609H, NGV752047H, NGV745254H, NGV744778H, NGV745775H, NGV747048H, NGV747623H, NGV748327H, NGV749607H, NGV749680H, NGV749694H, NGV751814H, NGV754027H, NGV754809H, NGV749070H, NGV744813H, NGV745326H, NGV746875H, NGV747855H, NGV744917H, NGV747062H, NGV747063H, NGV748252H, NGV743341H, NGV744032H, NGV744761H, NGV744892H, NGV746455H, NGV746711H, NGV746936H, NGV746952H, NGV746953H, NGV747881H, NGV747921H, NGV747948H, NGV747981H, NGV748191H, NGV749851H, NGV750291H, NGV750292H, NGV751750H, NGV742122H, NGV752899H, NGV754852H, NGV744763H, NGV745953H, NGV747041H, NGV752428H, NGV745543H, NGV752472H, NGV754723H, NGV744463H, NGV750075H, NGV754329H, NGV754400H, NGV754516H, NGV747235H, NGV747236H, NGV747956H, NGV748986H, NGV749369H, NGV749967H, NGV751575H, NGV754087H, NGV742794H, NGV745605H, NGV745608H, NGV749970H, NGV745153H, NGV744575H, NGV743651H, NGV749370H, NGV751059H, NGV744622H, NGV744856H, NGV745965H, NGV747039H, NGV748350H, NGV749182H, NGV749183H, NGV750106H, NGV750470H, NGV751097H, NGV754899H, NGV748936H, NGV743232H, NGV747740H, NGV753610H, NGV743391H, NGV743659H, NGV747342H, NGV749999H, NGV745435H, NGV748949H, NGV749696H, NGV750115H, NGV750861H, NGV751122H, NGV752221H, NGV752222H, NGV752310H, NGV753506H, NGV754198H, NGV754477H, NGV754555H, NGV742119H, NGV742241H, NGV743418H, NGV746937H, NGV747228H, NGV747755H, NGV747989H, NGV749672H, NGV750767H, NGV750814H, NGV750817H, NGV751619H, NGV752479H, NGV753227H, NGV742093H, NGV742242H, NGV744177H, NGV744878H, NGV746357H, NGV746775H, NGV749948H, NGV751413H, NGV753346H, NGV753653H, NGV753986H, NGV744931H, NGV746942H, NGV747208H, NGV747721H, NGV747818H, NGV749966H, NGV750788H, NGV749210H, NGV749382H, NGV753798H, NGV753801H, NGV753804H, NGV753810H, NGV742369H, NGV746840H, NGV748858H, NGV750202H, NGV754506H, NGV743258H, NGV744530H, NGV745936H, NGV749616H, NGV745851H, NGV747023H, NGV747025H, NGV747572H, NGV751285H, NGV751843H, NGV751844H, NGV753121H, NGV747009H, NGV747754H, NGV748432H, NGV753544H, NGV751530H, NGV753112H, NGV753116H, NGV751045H, NGV753392H, NGV742100H, NGV743431H, NGV745048H, NGV750630H, NGV752580H, NGV749285H, NGV753538H, NGV754003H, NGV754180H, NGV754674H, NGV746661H, NGV747091H, NGV748224H, NGV752359H, NGV752770H, NGV751520H, NGV755411H, NGV747541H, NGV753607H, NGV753176H, NGV752836H, NGV746817H, NGV746853H, NGV752432H, NGV752573H, NGV752879H, NGV753792H, NGV754828H, NGV754772H, NGV751802H, NGV754250H, NGV754116H, NGV749625H, NGV752225H, NGV753124H, NGV753956H, NGV749051H, NGV753025H, NGV750505H, NGV753848H, NGV754591H, NGV754592H, NGV754595H, NGV754741H, NGV754742H, NGV754749H, NGV754752H, NGV743674H, NGV745823H, NGV749078H, NGV753036H, NGV753037H, NGV753839H, NGV753840H, NGV753841H, NGV753842H, NGV753843H, NGV751511H, NGV750723H, NGV751355H, NGV752877H, NGV745224H, NGV747905H, NGV744019H, NGV750313H, NGV750316H, NGV753834H, NGV753835H, NGV753837H, NGV753838H, NGV744239H, NGV744243H, NGV745142H, NGV747135H, NGV747138H, NGV749520H, NGV749979H, NGV750041H, NGV751725H, NGV751806H, NGV752507H, NGV753342H, NGV754208H, NGV744085H, NGV744089H, NGV747275H, NGV747276H, NGV747277H, NGV747278H, NGV747281H, NGV747282H, NGV747286H, NGV750375H, NGV750376H, NGV750377H, NGV750381H, NGV750384H, NGV747206H, NGV749863H, NGV752379H, NGV743550H, NGV752663H, NGV752711H, NGV753034H, NGV753163H, NGV752189H, NGV748859H, NGV753738H, NGV753739H, NGV744194H, NGV746835H, NGV749881H, NGV753007H, NGV753849H, NGV754205H, NGV754759H, NGV754782H, NGV748173H, NGV752673H, NGV752675H, NGV752986H, NGV754611H, NGV754658H, NGV754697H, NGV754698H, NGV754761H, NGV755498H, NGV751686H, NGV751768H, NGV753495H, NGV754847H, NGV743905H, NGV743911H, NGV744238H, NGV746524H, NGV746891H, NGV750545H, NGV750553H, NGV750558H, NGV750560H, NGV753742H, NGV753743H, NGV753744H, NGV753746H, NGV753747H, NGV753748H, NGV753749H, NGV753750H, NGV753751H, NGV753752H, NGV742191H, NGV742192H, NGV742376H, NGV748651H, NGV754084H, NGV754237H, NGV754238H, NGV754528H, NGV754604H, NGV755671H, NGV755674H, NGV755676H, NGV755677H, NGV748084H, NGV750298H, NGV750631H, NGV752383H, NGV753502H, NGV754643H, NGV754644H, NGV754646H, NGV746568H, NGV746587H, NGV746590H, NGV752726H, NGV752841H, NGV753073H, NGV753074H, NGV753482H, NGV753492H, NGV754589H, NGV752835H, NGV746598H, NGV753358H, NGV755406H, NGV755593H, NGV755594H, NGV753368H, NGV753049H, NGV753500H, NGV753858H, NGV754642H, NGV747710H, NGV744073H, NGV744329H, NGV748918H, NGV752672H, NGV755450H, NGV752550H, NGV754935H, NGV744570H, NGV746826H, NGV753122H, NGV753283H, NGV753284H, NGV754739H, NGV743893H, NGV752384H, NGV752387H, NGV754762H, NGV754763H, NGV743317H, NGV744355H, NGV745574H, NGV746390H, NGV749116H, NGV753944H, NGV754513H, NGV754517H, NGV755108H, NGV751269H, NGV750440H, NGV752940H, NGV754615H, NGV755532H, NGV751691H, NGV752381H, NGV754680H, NGV754682H, NGV754936H, NGV755602H, NGV749723H, NGV754314H, NGV754315H, NGV754833H, NGV754834H, NGV754835H, NGV753152H, NGV753005H, NGV752536H, NGV752528H, NGV752725H, NGV753181H, NGV752422H, NGV752937H, NGV746064H, NGV754004H, NGV752833H, NGV743101H, NGV745076H, NGV751992H, NGV751995H, NGV743361H, NGV750836H, NGV751746H, NGV747133H, NGV747134H, NGV747146H, NGV752516H, NGV753673H, NGV753679H, NGV748855H, NGV748856H, NGV750403H, NGV752568H, NGV752970H, NGV752973H, NGV749536H, NGV749537H, NGV751556H, NGV752830H, NGV752831H, NGV752837H, NGV752519H, NGV748162H, NGV751690H, NGV753245H, NGV752193H, NGV754575H, NGV746671H, NGV754839H, NGV754841H, NGV754842H, NGV752838H, NGV753692H, NGV753693H, NGV753695H, NGV747390H, NGV749499H, NGV749505H, NGV749506H, NGV749508H, NGV753460H, NGV753463H, NGV753470H, NGV753691H, NGV753694H, NGV753698H, NGV743536H, NGV745992H, NGV746545H, NGV755464H, NGV755506H, NGV745146H, NGV745148H, NGV747746H, NGV750248H, NGV750253H, NGV751784H, NGV755615H, NGV755616H, NGV755685H, NGV746528H, NGV750611H, NGV753266H, NGV753267H, NGV753349H, NGV753458H, NGV753712H, NGV753713H, NGV753715H, NGV753718H, NGV753719H, NGV753720H, NGV752829H, NGV755116H, NGV755117H, NGV755118H, NGV749291H, NGV744294H, NGV755603H, NGV744418H, NGV750615H, NGV750616H, NGV744118H, NGV754471H, NGV743702H, NGV747083H, NGV749075H, NGV753097H, NGV753446H, NGV750320H, NGV752370H, NGV752371H, NGV753916H, NGV753917H, NGV753854H, NGV753855H, NGV742511H, NGV742512H, NGV743722H, NGV744209H, NGV744351H, NGV744352H, NGV744353H, NGV752780H, NGV754481H, NGV755595H, NGV755596H, NGV755597H, NGV755598H, NGV755599H, NGV743542H, NGV753246H, NGV754258H, NGV754961H, NGV754962H, NGV754963H, NGV745295H, NGV751740H, NGV744515H, NGV744516H, NGV744517H, NGV753625H, NGV753626H, NGV753628H, NGV753631H, NGV753634H, NGV743856H, NGV751703H, NGV755688H, NGV755689H, NGV751155H, NGV754890H, NGV755455H, NGV755637H, NGV748644H, NGV749911H, NGV752074H, NGV753051H, NGV753054H, NGV754849H, NGV744225H, NGV754519H, NGV752522H, NGV748937H, NGV753704H, NGV753705H, NGV753706H, NGV753707H, NGV744299H, NGV753212H, NGV753218H, NGV754464H, NGV753170H, NGV743335H, NGV743991H, NGV750036H, NGV752070H, NGV752944H, NGV752945H, NGV753761H, NGV754241H, NGV754242H, NGV754243H, NGV754316H, NGV754319H, NGV754320H, NGV754321H, NGV754391H, NGV754640H, NGV750020H, NGV753814H, NGV753815H, NGV753816H, NGV753817H, NGV753401H, NGV753405H, NGV753406H, NGV753472H, NGV743465H, NGV750629H, NGV753882H, NGV753883H, NGV753884H, NGV753885H, NGV753887H, NGV753889H, NGV753891H, NGV753379H, NGV754254H, NGV752356H, NGV746345H, NGV750441H, NGV754168H, NGV754760H, NGV746480H, NGV753040H, NGV744343H, NGV746131H, NGV747209H, NGV747213H, NGV750408H, NGV750409H, NGV750411H, NGV750412H, NGV753726H, NGV753727H, NGV753728H, NGV752952H, NGV742217H, NGV743587H, NGV752724H, NGV755582H, NGV752409H, NGV752413H, NGV752414H, NGV753859H, NGV754264H, NGV754846H, NGV739007H, NGV753224H, NGV753225H, NGV745280H, NGV755795H, NGV755796H, NGV755797H, NGV755801H, NGV755805H, NGV746657H, NGV753150H, NGV749627H, NGV749886H, NGV749887H, NGV750438H, NGV754606H, NGV746950H, NGV755027H, NGV749704H, NGV750047H, NGV750048H, NGV747546H, NGV752667H, NGV744247H, NGV753098H, NGV753250H, NGV753254H, NGV754764H, NGV754765H, NGV754766H, NGV754767H, NGV746596H, NGV753468H, NGV754654H, NGV754656H, NGV750855H, NGV755102H, NGV752957H, NGV752502H, NGV747481H, NGV754260H, NGV755465H, NGV752959H, NGV749934H, NGV751309H, NGV755726H, NGV755789H, NGV755790H, NGV752055H, NGV752548H, NGV754647H, NGV754649H, NGV746567H, NGV747853H, NGV748066H, NGV748716H, NGV749707H, NGV751767H, NGV752073H, NGV753828H, NGV753829H, NGV754209H, NGV754303H, NGV754600H, NGV754651H, NGV754652H, NGV754848H, NGV754870H, NGV754886H, NGV754887H, NGV754983H, NGV755113H, NGV755451H, NGV755622H, NGV748007H, NGV751792H, NGV754245H, NGV753445H, NGV752963H, NGV747579H, NGV752542H, NGV746370H, NGV747413H, NGV747414H, NGV750959H, NGV750961H, NGV749002H, NGV747724H, NGV750295H, NGV750296H, NGV751808H, NGV753812H, NGV751180H, NGV748311H, NGV748314H, NGV752388H, NGV754229H, NGV754233H, NGV751303H, NGV753353H, NGV751852H, NGV753536H, NGV754534H, NGV743916H, NGV743994H, NGV746517H, NGV746518H, NGV746519H, NGV746520H, NGV747322H, NGV747323H, NGV747324H, NGV747888H, NGV747889H, NGV748901H, NGV748902H, NGV749880H, NGV749977H, NGV750154H, NGV751158H, NGV752670H, NGV752671H, NGV753295H, NGV753296H, NGV753297H, NGV743178H, NGV747729H, NGV754024H, NGV755415H, NGV747949H, NGV753466H, NGV754322H, NGV755010H, NGV755011H, NGV749219H, NGV751304H, NGV754235H, NGV755040H, NGV753619H, NGV751280H, NGV748706H, NGV753096H, NGV753665H, NGV753853H, NGV755583H, NGV755584H, NGV745960H, NGV747960H, NGV753026H, NGV755145H, NGV747629H, NGV750704H, NGV742367H, NGV747491H, NGV749931H, NGV753514H, NGV753515H, NGV754803H, NGV754804H, NGV749624H, NGV753173H, NGV753963H, NGV745127H, NGV754753H, NGV754757H, NGV755640H, NGV750425H, NGV744327H, NGV747944H, NGV749114H, NGV749940H, NGV752571H, NGV752849H, NGV749852H, NGV749853H, NGV749857H, NGV749858H, NGV749859H, NGV753773H, NGV753774H, NGV753775H, NGV753777H, NGV753923H, NGV752913H, NGV755627H, NGV755628H, NGV755686H, NGV755687H, NGV746476H, NGV750413H, NGV744216H, NGV744217H, NGV744221H, NGV744224H, NGV749509H, NGV749511H, NGV749512H, NGV743613H, NGV743614H, NGV744277H, NGV746749H, NGV746755H, NGV746756H, NGV746758H, NGV746760H, NGV746768H, NGV752449H, NGV752450H, NGV752453H, NGV752454H, NGV744798H, NGV745919H, NGV745927H, NGV746742H, NGV746770H, NGV746793H, NGV746797H, NGV746960H, NGV746962H, NGV746964H, NGV746968H, NGV746975H, NGV746977H, NGV750168H, NGV750169H, NGV750172H, NGV750173H, NGV750174H, NGV750175H, NGV750179H, NGV753211H, NGV753234H, NGV753235H, NGV753236H, NGV753237H, NGV753240H, NGV754571H, NGV754572H, NGV754573H, NGV744618H, NGV743645H, NGV743874H, NGV746733H, NGV746905H, NGV747466H, NGV748731H, NGV749715H, NGV749865H, NGV749874H, NGV751289H, NGV751738H, NGV753013H, NGV753014H, NGV753015H, NGV753019H, NGV753065H, NGV753067H, NGV753069H, NGV753187H, NGV753192H, NGV753193H, NGV753194H, NGV753195H, NGV753198H, NGV753199H, NGV753202H, NGV753208H, NGV755014H, NGV755015H, NGV755016H, NGV755018H, NGV755019H, NGV755020H, NGV755084H, NGV747316H, NGV753099H, NGV753452H, NGV752958H, NGV752884H, NGV749124H, NGV749125H, NGV750436H, NGV750636H, NGV752558H, NGV752941H, NGV748752H, NGV749702H, NGV753700H, NGV753616H, NGV753339H, NGV753461H, NGV753465H, NGV753473H, NGV753477H, NGV752509H, NGV750633H, NGV752720H, NGV752721H, NGV752722H, NGV752723H, NGV742637H, NGV746058H, NGV746060H, NGV752741H, NGV752951H, NGV753012H, NGV754784H, NGV754823H, NGV752967H, NGV753153H, NGV755452H, NGV755454H, NGV753873H, NGV749243H, NGV749375H, NGV752514H, NGV755776H, NGV753214H, NGV753252H, NGV753275H, NGV753337H, NGV753787H, NGV753788H, NGV744497H, NGV750496H, NGV753860H, NGV753861H, NGV753862H, NGV753864H, NGV753865H, NGV753866H, NGV753868H, NGV753869H, NGV752947H, NGV747200H, NGV747201H, NGV752382H, NGV754253H, NGV754257H, NGV754259H, NGV742643H, NGV746642H, NGV747476H, NGV753682H, NGV753685H, NGV753689H, NGV753690H, NGV751804H, NGV752936H, NGV749944H, NGV750612H, NGV750618H, NGV752778H, NGV753623H, NGV747032H, NGV754917H, NGV741924H, NGV743717H, NGV750127H, NGV753677H, NGV749599H, NGV751434H, NGV755586H, NGV755587H, NGV752504H, NGV753154H, NGV745212H, NGV749157H, NGV751006H, NGV751016H, NGV752594H, NGV752595H, NGV752596H, NGV752597H, NGV752599H, NGV752600H, NGV752601H, NGV753588H, NGV753589H, NGV753590H, NGV753591H, NGV753592H, NGV753593H, NGV753594H, NGV753596H, NGV753597H, NGV752369H, NGV752518H, NGV744654H, NGV744864H, NGV752563H, NGV752832H, NGV752880H, NGV754030H, NGV754327H, NGV752298H, NGV752513H, NGV747495H, NGV747496H, NGV747499H, NGV746746H, NGV747157H, NGV747160H, NGV751741H, NGV751742H, NGV751743H, NGV751745H, NGV752390H, NGV752391H, NGV752393H, NGV752394H, NGV752396H, NGV752398H, NGV749799H, NGV749637H, NGV752063H, NGV753870H, NGV753871H, NGV753872H, NGV753876H, NGV753877H, NGV753879H, NGV753880H, NGV753881H, NGV753022H, NGV752500H, NGV752886H, NGV752892H, NGV752895H, NGV753177H, NGV753180H, NGV753980H, NGV743488H, NGV753165H, NGV753166H, NGV747230H, NGV753160H, NGV753168H, NGV744441H, NGV752556H, NGV753146H, NGV753900H, NGV753901H, NGV743721H, NGV748108H, NGV752543H, NGV754668H, NGV754740H, NGV747720H, NGV742138H, NGV746845H, NGV753161H, NGV748495H, NGV750728H, NGV750734H, NGV743664H, NGV747505H, NGV747509H, NGV749016H, NGV749902H, NGV753486H, NGV753735H, NGV755783H, NGV755785H, NGV751842H, NGV743967H, NGV745790H, NGV746966H, NGV747260H, NGV747264H, NGV747650H, NGV751326H, NGV751334H, NGV751336H, NGV752113H, NGV752651H, NGV752655H, NGV753301H, NGV753303H, NGV753304H, NGV753308H, NGV753974H, NGV753975H, NGV753976H, NGV754169H, NGV754170H, NGV754183H, NGV754341H, NGV754349H, NGV754363H, NGV754451H, NGV755004H, NGV746421H, NGV755780H, NGV744347H, NGV745560H, NGV753175H, NGV753171H, NGV746361H, NGV751752H, NGV751754H, NGV751809H, NGV752537H, NGV750656H, NGV753030H, NGV753033H, NGV747358H, NGV747367H, NGV747369H, NGV753759H, NGV753799H, NGV753800H, NGV753802H, NGV753803H, NGV753805H, NGV753806H, NGV753807H, NGV753808H, NGV753809H, NGV753811H, NGV743484H, NGV748991H, NGV750977H, NGV753384H, NGV753641H, NGV747913H, NGV752946H, NGV752567H; GTIN 00763000689476, Serial Numbers: NGV741986H, NGV742043H, NGV742724H, NGV742756H, NGV742765H, NGV742992H, NGV743007H, NGV743009H, NGV741972H, NGV741974H, NGV742067H, NGV742073H, NGV742082H, NGV742108H, NGV742712H, NGV746330H, NGV746420H, NGV746429H, NGV746433H, NGV746588H, NGV748319H, NGV748421H, NGV748538H, NGV749917H, NGV749969H, NGV750518H, NGV748771H, NGV750530H, NGV750644H, NGV750524H, NGV748786H, NGV749907H, NGV751374H, NGV748790H, NGV748793H, NGV745679H, NGV745743H, NGV746015H, NGV746016H, NGV748774H; GTIN 00763000689483, Serial Numbers: NGV750527H, NGV750567H, NGV743206H, NGV748844H, NGV752013H, NGV744211H, NGV746516H, NGV750312H, NGV750589H, NGV750597H, NGV750599H, NGV751798H, NGV751824H, NGV751936H, NGV746413H, NGV751932H, NGV752011H, NGV754596H, NGV754708H, NGV754729H, NGV743215H, NGV743259H, NGV743251H, NGV745443H, NGV742626H, NGV742704H, NGV742706H, NGV742707H, NGV742716H, NGV743304H, NGV746431H, NGV746575H, NGV748063H, NGV748256H, NGV749735H, NGV749749H, NGV749750H, NGV748661H, NGV743394H, NGV750303H, NGV746483H, NGV753169H, NGV748286H, NGV745465H, NGV745488H, NGV752965H, NGV753174H, NGV753162H, NGV749773H, NGV748279H, NGV743213H, NGV748250H, NGV749718H, NGV742855H, NGV742831H, NGV742944H, NGV744284H, NGV744293H, NGV742743H, NGV742874H, NGV748061H, NGV748065H, NGV748675H, NGV743014H, NGV744215H, NGV744219H, NGV746573H, NGV745478H, NGV742722H, NGV742772H, NGV745447H, NGV746435H, NGV746436H, NGV746581H, NGV748058H, NGV749756H, NGV750290H, NGV742863H, NGV744309H, NGV746504H, NGV748045H, NGV748872H, NGV749713H, NGV750571H, NGV742935H, NGV745497H, NGV746560H, NGV752964H, NGV752966H, NGV753149H, NGV753157H, NGV745498H, NGV745505H, NGV748280H, NGV748867H, NGV749738H, NGV752015H, NGV752950H; GTIN 00763000689490, Serial Numbers: NGV748428H, NGV748963H, NGV748973H, NGV748996H, NGV752285H, NGV753497H, NGV754200H, NGV754359H, NGV754505H, NGV748960H, NGV748383H, NGV752288H, NGV748340H, NGV753489H, NGV748394H, NGV749032H, NGV754500H; GTIN 00763000689506, Serial Numbers: NGV742678H, NGV742690H, NGV742720H, NGV746017H, NGV746036H, NGV746037H, NGV746070H, NGV746261H, NGV746311H, NGV746323H, NGV748304H, NGV748315H, NGV748874H, NGV749040H, NGV749052H, NGV749922H, NGV749991H, NGV750010H, NGV750040H, NGV750055H, NGV750067H, NGV750103H, NGV750655H, NGV752681H, NGV744005H, NGV744006H, NGV744016H, NGV745691H, NGV746076H, NGV746440H, NGV746441H, NGV748338H, NGV748339H, NGV748342H, NGV748348H, NGV748361H, NGV748363H, NGV748364H, NGV749909H, NGV746564H, NGV746569H, NGV746570H, NGV746572H, NGV746586H, NGV746892H, NGV747096H, NGV747110H, NGV747117H, NGV747125H, NGV747126H, NGV751191H, NGV751194H, NGV751195H, NGV751373H, NGV752027H, NGV754152H, NGV754193H, NGV754218H, NGV754227H, NGV754228H, NGV754268H, NGV754282H, NGV741930H, NGV741969H, NGV741983H, NGV742003H, NGV742004H, NGV742031H, NGV745717H, NGV745729H, NGV745732H, NGV745741H, NGV745757H, NGV745865H, NGV745869H, NGV746030H, NGV746033H, NGV745599H, NGV745606H, NGV749418H, NGV749456H, NGV754149H, NGV754951H, NGV749968H, NGV752499H, NGV746411H, NGV742543H, NGV742723H, NGV746416H, NGV748887H, NGV748999H, NGV750517H, NGV752267H, NGV752329H, NGV752335H, NGV752346H, NGV752544H, NGV753576H, NGV754144H, NGV742538H, NGV754150H, NGV741922H, NGV741985H, NGV741996H, NGV742568H, NGV742738H, NGV744003H, NGV745683H, NGV746412H, NGV746482H, NGV746502H, NGV748423H, NGV748522H, NGV748525H, NGV748877H, NGV749417H, NGV752224H, NGV752540H, NGV746018H, NGV749420H, NGV749896H, NGV745703H, NGV750117H, NGV751080H, NGV751127H, NGV742473H, NGV742536H, NGV742550H, NGV742674H, NGV743795H, NGV745654H, NGV745704H, NGV745721H, NGV746169H, NGV746334H, NGV749004H, NGV749006H, NGV750645H, NGV752229H, NGV752704H, NGV752712H, NGV754047H, NGV754345H, NGV745855H, NGV742547H, NGV750026H, NGV750508H, NGV752348H, NGV752715H, NGV753479H, NGV753480H, NGV746444H, NGV748378H, NGV749251H, NGV749258H, NGV742539H, NGV742559H, NGV742565H, NGV742702H, NGV742715H, NGV746424H, NGV748359H, NGV750021H, NGV750513H, NGV750842H, NGV754974H, NGV742736H, NGV745616H, NGV745697H, NGV746035H, NGV748372H, NGV748900H, NGV748934H, NGV749003H, NGV750017H, NGV750515H, NGV751142H, NGV752546H, NGV752561H, NGV754969H, NGV742548H, NGV747162H, NGV749247H, NGV748969H, NGV742019H, NGV750670H, NGV751085H; GTIN 00763000689513, Serial Numbers: NGV743219H, NGV748556H, NGV748567H, NGV749560H, NGV750758H, NGV750763H, NGV750775H, NGV751490H, NGV751747H, NGV748545H, NGV750587H, NGV751492H, NGV741946H, NGV750752H, NGV742028H, NGV743037H, NGV750563H, NGV750591H, NGV751729H, NGV743020H, NGV743072H, NGV743076H, NGV743087H, NGV751361H, NGV751480H, NGV741995H, NGV742900H, NGV742908H, NGV742911H, NGV742923H, NGV742927H, NGV742943H, NGV742962H, NGV742972H, NGV743043H, NGV743054H, NGV746327H, NGV746355H, NGV746486H, NGV746487H, NGV748403H, NGV749575H, NGV749580H, NGV751357H, NGV751485H, NGV751487H, NGV742999H, NGV743129H, NGV750829H, NGV751327H, NGV746321H, NGV742871H, NGV742897H, NGV746301H, NGV746307H, NGV750705H, NGV751502H, NGV751349H, NGV750562H, NGV743041H, NGV749774H, NGV746317H, NGV748563H, NGV743042H, NGV749602H, NGV748542H, NGV746296H, NGV742953H, NGV742946H, NGV748548H, NGV743212H, NGV742025H, NGV742027H, NGV742021H, NGV743248H, NGV745586H, NGV745751H, NGV748044H, NGV748054H, NGV748102H, NGV748275H, NGV748289H, NGV748377H, NGV748618H, NGV748623H, NGV748624H, NGV748631H, NGV748724H, NGV746041H, NGV742699H, NGV745978H, NGV751467H, NGV751469H, NGV742703H, NGV742737H, NGV745761H, NGV749877H, NGV742705H, NGV742725H, NGV746040H, NGV748328H, NGV750529H, NGV751365H, NGV751444H, NGV749547H, NGV748362H, NGV751405H, NGV749504H, NGV749914H, NGV743793H, NGV744300H, NGV742746H, NGV742849H, NGV744312H, NGV743111H, NGV744222H, NGV746580H, NGV741944H, NGV743785H, NGV743117H, NGV743780H, NGV744298H, NGV744310H, NGV748728H, NGV750526H, NGV751125H, NGV748037H, NGV744040H, NGV744027H, NGV744030H, NGV746584H, NGV746283H, NGV744213H, NGV746043H, NGV751627H, NGV743112H, NGV746513H, NGV751113H, NGV751111H, NGV744320H, NGV746047H, NGV746562H, NGV744227H, NGV744317H, NGV744435H, NGV744457H, NGV748588H, NGV742669H, NGV749921H, NGV749529H, NGV742675H, NGV751033H, NGV751587H, NGV743077H, NGV744429H, NGV745258H, NGV744232H, NGV751391H, NGV742668H, NGV744012H, NGV748042H, NGV750522H, NGV750525H, NGV750764H, NGV750769H, NGV751465H, NGV751681H, NGV751688H, NGV751692H, NGV745382H, NGV745565H, NGV744449H, NGV749925H, NGV749949H, NGV751528H, NGV744301H, NGV745363H, NGV743924H; GTIN 00763000689520, Serial Numbers: NGV751196H, NGV751198H, NGV751332H, NGV752885H, NGV743122H, NGV745546H, NGV748748H, NGV748751H, NGV749198H, NGV745503H, NGV749431H, NGV750341H, NGV750378H, NGV752106H, NGV745439H, NGV750347H, NGV745476H, NGV745492H, NGV748763H, NGV748765H, NGV748767H, NGV742767H, NGV743106H, NGV750333H, NGV745429H, NGV745489H, NGV745548H, NGV748758H, NGV748834H, NGV750338H, NGV751282H, NGV745557H, NGV742688H, NGV750324H, NGV748761H, NGV751330H, NGV752109H, NGV743109H, NGV743116H, NGV743148H, NGV745496H, NGV745500H, NGV745558H, NGV748586H, NGV748605H, NGV748747H, NGV748781H, NGV749254H, NGV750325H, NGV750336H, NGV750402H, NGV750426H, NGV750431H, NGV748770H, NGV749441H, NGV749442H, NGV749447H, NGV749199H, NGV748611H, NGV749459H, NGV745504H, NGV745448H, NGV751283H, NGV748819H, NGV745549H, NGV749278H, NGV751360H, NGV752101H, NGV742735H, NGV748753H, NGV745507H, NGV748592H, NGV748772H, NGV748836H, NGV748840H, NGV749460H, NGV750422H, NGV749435H, NGV748593H, NGV748599H, NGV751199H, NGV742875H, NGV745438H, NGV745475H, NGV745502H, NGV748757H, NGV748759H, NGV748803H, NGV748832H, NGV750349H, NGV750451H, NGV751331H, NGV752138H, NGV748596H, NGV748813H, NGV742681H, NGV745446H, NGV748608H, NGV749436H, NGV749439H, NGV750362H, NGV743115H, NGV743144H, NGV745495H, NGV748755H, NGV749262H, NGV750322H, NGV750434H, NGV750437H, NGV742693H, NGV748603H, NGV750379H, NGV745550H, NGV751286H, NGV743150H, NGV750445H, NGV743108H, NGV742691H, NGV742692H, NGV742773H, NGV743135H, NGV748787H, NGV748595H, NGV742679H, NGV742687H, NGV742689H, NGV748802H, NGV749281H, NGV749523H, NGV751812H, NGV745147H, NGV745134H, NGV745144H, NGV742006H, NGV743113H, NGV745130H, NGV746066H, NGV749080H, NGV746557H, NGV743003H, NGV745037H, NGV748064H, NGV748671H, NGV749502H, NGV750068H, NGV750118H, NGV745133H, NGV749530H, NGV749532H, NGV749742H, NGV751404H, NGV753060H, NGV753062H, NGV753068H, NGV744921H, NGV745866H, NGV749395H, NGV752081H, NGV748115H, NGV753618H, NGV749036H, NGV752378H, NGV752652H, NGV753570H, NGV753643H, NGV750138H, NGV750147H, NGV753560H, NGV753563H, NGV753644H, NGV754058H, NGV753273H, NGV754066H, NGV754050H, NGV752237H, NGV752376H, NGV752506H, NGV753552H, NGV748296H, NGV749501H, NGV753572H, NGV754053H, NGV754060H, NGV754061H, NGV754068H, NGV752380H, NGV753274H, NGV743140H, NGV754416H, NGV748568H, NGV742920H, NGV745551H, NGV748126H, NGV748530H, NGV749765H, NGV752666H, NGV742017H, NGV742671H, NGV742672H, NGV742907H, NGV742919H, NGV745009H, NGV745020H, NGV745038H, NGV746098H, NGV748055H, NGV753642H, NGV742087H, NGV742912H, NGV742988H, NGV743082H, NGV745091H, NGV747109H, NGV748039H, NGV748531H, NGV748876H, NGV749570H, NGV750078H, NGV750142H, NGV741994H, NGV743084H, NGV743154H, NGV743820H, NGV744980H, NGV745010H, NGV745033H, NGV745092H, NGV745641H, NGV745744H, NGV746019H, NGV746144H, NGV746150H, NGV746271H, NGV747111H, NGV747992H, NGV748869H, NGV748878H, NGV749194H, NGV749197H, NGV749767H, NGV749768H, NGV750538H, NGV750774H, NGV752781H, NGV742960H, NGV744981H, NGV745015H, NGV745021H, NGV745292H, NGV745640H, NGV746103H, NGV748038H, NGV748529H, NGV749062H, NGV753566H, NGV741963H, NGV742853H, NGV743017H, NGV744994H, NGV745398H, NGV745583H, NGV745618H, NGV746556H, NGV748701H, NGV748778H, NGV748794H, NGV749060H, NGV751185H, NGV744013H, NGV749284H, NGV742922H, NGV746179H, NGV742941H, NGV743155H, NGV745561H, NGV745638H, NGV748056H, NGV750110H, NGV750549H, NGV753228H, NGV753241H, NGV754069H, NGV741978H, NGV741982H, NGV742997H, NGV743004H, NGV743012H, NGV743085H, NGV743237H, NGV745000H, NGV745066H, NGV745736H, NGV746085H, NGV746109H, NGV747123H, NGV748201H, NGV749068H, NGV749510H, NGV749517H, NGV749521H, NGV750104H, NGV750497H, NGV750776H, NGV750777H, NGV751507H, NGV752096H, NGV752788H, NGV753238H, NGV749576H, NGV752918H, NGV748971H, NGV749021H, NGV750596H, NGV745090H, NGV744020H, NGV744066H, NGV745486H, NGV754159H, NGV741949H, NGV748535H, NGV742007H, NGV745399H, NGV742018H, NGV744009H, NGV746203H, NGV748700H, NGV749766H, NGV751184H, NGV742860H, NGV745029H, NGV746269H, NGV748052H, NGV748555H, NGV749516H, NGV742016H, NGV742873H, NGV742996H, NGV744015H, NGV744026H, NGV745553H, NGV745555H, NGV745614H, NGV745619H, NGV745622H, NGV745624H, NGV745625H, NGV745626H, NGV746139H, NGV746141H, NGV748216H, NGV748223H, NGV748273H, NGV748572H, NGV748574H, NGV748577H, NGV748833H, NGV749064H, NGV749067H, NGV750107H, NGV750170H, NGV750590H, NGV750593H, NGV751094H, NGV751800H, NGV751801H, NGV751943H, NGV752102H, NGV752194H, NGV752505H, NGV753573H, NGV745362H, NGV745392H, NGV745746H, NGV750491H, NGV752234H, NGV743088H, NGV745400H, NGV749563H, NGV750504H, NGV751410H, NGV741998H, NGV748783H, NGV744028H, NGV745293H, NGV745629H, NGV749764H, NGV750494H, NGV752508H, NGV742020H, NGV745632H, NGV745740H, NGV750159H, NGV745631H, NGV748197H, NGV748235H, NGV748312H, NGV748324H, NGV749572H, NGV750149H, NGV750158H, NGV751168H, NGV752842H, NGV753270H, NGV746217H, NGV748621H, NGV753645H, NGV750307H, NGV751602H, NGV751617H, NGV748655H, NGV749191H, NGV750463H, NGV750835H, NGV751608H, NGV751626H, NGV751634H, NGV751649H, NGV751533H, NGV751622H, NGV750826H, NGV742978H, NGV748691H, NGV742977H, NGV749776H, NGV742095H.NGV743249H, NGV750461H, NGV750781H, NGV750782H, NGV751521H, NGV746225H, NGV748637H, NGV751641H, NGV751646H, NGV745684H, NGV748730H, NGV742097H, NGV746238H, NGV750314H, NGV751606H, NGV745692H, NGV746236H, NGV750843H, NGV742713H, NGV743050H, NGV746216H, NGV746237H, NGV748693H, NGV748694H, NGV748697H, NGV750321H, NGV751612H, NGV749188H, NGV746220H, NGV746224H, NGV746226H, NGV746227H, NGV746228H, NGV746230H, NGV746231H, NGV746232H, NGV750786H, NGV750789H, NGV750790H, NGV742729H, NGV743105H, NGV746045H, NGV746075H, NGV746221H, NGV748687H, NGV748698H, NGV749192H, NGV750297H, NGV750319H, NGV750459H, NGV751524H, NGV751636H, NGV748739H, NGV748744H, NGV749866H, NGV749869H, NGV750785H, NGV750794H, NGV750838H, NGV751499H, NGV743055H, NGV743058H, NGV746198H, NGV748732H, NGV749781H, NGV749871H, NGV750289H, NGV750783H, NGV750784H, NGV750844H, NGV750845H, NGV751491H, NGV751495H, NGV751611H, NGV746212H, NGV746214H, NGV748664H, NGV748681H, NGV748737H, NGV748742H, NGV742135H, NGV742710H, NGV743021H, NGV743039H, NGV745671H, NGV746234H, NGV746243H, NGV748733H, NGV749779H, NGV749780H, NGV749782H, NGV749783H, NGV750795H, NGV745678H, NGV746235H, NGV751525H, NGV742968H, NGV743252H, NGV743260H, NGV748699H, NGV750299H, NGV742714H, NGV742721H, NGV743011H, NGV742089H, NGV742094H, NGV742726H, NGV742727H, NGV742984H, NGV743001H, NGV745670H, NGV746229H, NGV746233H, NGV748622H, NGV748745H, NGV750318H, NGV742099H, NGV742979H, NGV745672H, NGV743049H, NGV745689H, NGV749777H, NGV746239H, NGV749772H, NGV742728H, NGV743044H, NGV745702H, NGV749186H, NGV749775H, NGV750308H, NGV750310H, NGV750818H, NGV750850H, NGV742696H, NGV742732H, NGV745686H, NGV750830H, NGV750831H, NGV742698H, NGV742718H, NGV742719H, NGV742891H, NGV746240H, NGV746241H, NGV751527H, NGV742982H, NGV743046H, NGV750300H, NGV750304H, NGV746245H, NGV750813H, NGV750816H, NGV742695H, NGV742700H, NGV743030H, NGV743107H, NGV745677H, NGV745680H, NGV745698H, NGV750779H, NGV751538H, NGV751541H, NGV751580H, NGV751584H, NGV751594H, NGV754358H, NGV754362H, NGV750216H, NGV751084H, NGV751510H, NGV749393H, NGV751399H, NGV754441H, NGV750114H, NGV753547H, NGV754073H, NGV754601H, NGV749895H, NGV752674H, NGV750089H, NGV751920H, NGV748721H, NGV750479H, NGV752212H, NGV752510H, NGV753611H, NGV754158H, NGV751508H, NGV751509H, NGV753493H, NGV751396H, NGV753632H, NGV750552H, NGV745856H, NGV752923H, NGV748255H, NGV744991H, NGV744993H, NGV748370H, NGV753633H, NGV754614H, NGV752084H, NGV745634H, NGV749413H, NGV752804H, NGV753139H, NGV748236H, NGV749759H, NGV751906H, NGV752660H, NGV751164H, NGV753373H, NGV754350H, NGV742137H, NGV742570H, NGV753247H, NGV753388H, NGV753393H, NGV753475H, NGV753926H, NGV754415H, NGV751431H, NGV751933H, NGV749549H, NGV752931H, NGV753586H, NGV749055H, NGV750486H, NGV752230H, NGV753770H, NGV748720H, NGV751370H, NGV749747H, NGV742870H, NGV752552H, NGV752796H, NGV748395H, NGV749553H, NGV749827H, NGV750642H, NGV748565H, NGV752700H, NGV752557H, NGV748237H, NGV749058H, NGV754414H, NGV746096H, NGV754141H, NGV750771H, NGV750773H, NGV751442H, NGV753996H, NGV746088H, NGV753341H, NGV754444H, NGV749403H, NGV746579H, NGV749405H, NGV753554H, NGV753557H, NGV754355H, NGV747104H, NGV749396H, NGV745115H, NGV745582H, NGV747106H, NGV750677H, NGV751397H, NGV752646H, NGV745346H, NGV746094H, NGV749900H, NGV743080H, NGV745653H, NGV748376H, NGV748620H, NGV748762H, NGV748933H, NGV748945H, NGV750489H, NGV750605H, NGV750770H, NGV745483H, NGV746210H, NGV748208H, NGV748368H, NGV748707H, NGV750081H, NGV750221H, NGV750676H, NGV745370H, NGV746089H, NGV746090H, NGV746108H, NGV742033H, NGV743104H, NGV743139H, NGV744058H, NGV746248H, NGV748614H, NGV750090H, NGV752658H, NGV743137H, NGV744906H, NGV747129H, NGV747148H, NGV748046H, NGV748109H, NGV743143H, NGV746183H, NGV742926H, NGV743024H, NGV745987H, NGV746080H, NGV747095H, NGV748254H, NGV749137H, NGV749399H, NGV749760H, NGV750183H, NGV751099H, NGV751902H, NGV752089H, NGV745071H, NGV748712H, NGV751095H, NGV741956H, NGV742859H, NGV743796H, NGV746209H, NGV748386H, NGV748520H, NGV748718H, NGV750082H, NGV751098H, NGV752555H, NGV753129H, NGV751395H, NGV742851H, NGV742888H, NGV746091H, NGV746092H, NGV748268H, NGV748544H, NGV749074H, NGV751151H, NGV751202H, NGV752368H, NGV752656H, NGV752657H, NGV746054H, NGV752689H, NGV743062H, NGV743881H, NGV745129H, NGV747113H, NGV748062H, NGV751166H, NGV751182H, NGV753389H, NGV749604H, NGV750480H, NGV748663H, NGV749409H, NGV742880H, NGV743068H, NGV745661H, NGV749132H, NGV751167H, NGV751443H, NGV752792H, NGV753215H, NGV753402H, NGV743067H, NGV743160H, NGV743921H, NGV745989H, NGV748369H, NGV748709H, NGV749595H, NGV743073H, NGV743848H, NGV745373H, NGV746211H, NGV748519H, NGV750772H, NGV752095H, NGV743811H, NGV745105H, NGV748207H, NGV748677H, NGV749056H, NGV752202H, NGV752520H, NGV742074H, NGV742815H, NGV743090H, NGV743091H, NGV745862H, NGV748317H, NGV750600H, NGV750805H, NGV751091H, NGV751156H, NGV751369H, NGV742000H, NGV742005H, NGV742032H, NGV742035H, NGV742037H, NGV742936H, NGV742991H, NGV743035H, NGV743040H, NGV743064H, NGV743071H, NGV743166H, NGV745118H, NGV745580H, NGV745752H, NGV745894H, NGV746111H, NGV746112H, NGV746364H, NGV748120H, NGV748212H, NGV748388H, NGV748503H, NGV748521H, NGV748600H, NGV748705H, NGV748717H, NGV749007H, NGV749049H, NGV749099H, NGV749424H, NGV749514H, NGV749743H, NGV749744H, NGV750604H, NGV750649H, NGV750755H, NGV750759H, NGV751152H, NGV751908H, NGV751928H, NGV751929H, NGV752924H, NGV752926H, NGV753059H, NGV753428H, NGV742827H, NGV742928H, NGV742989H, NGV743002H, NGV748261H, NGV749059H, NGV749503H, NGV749541H, NGV751149H, NGV751366H, NGV751367H, NGV751368H, NGV752512H, NGV752541H, NGV753210H, NGV743132H, NGV743229H, NGV747152H, NGV749077H, NGV752107H, NGV752693H, NGV742047H, NGV748708H, NGV749121H, NGV749913H, NGV742673H, NGV742934H, NGV749400H, NGV742011H, NGV748558H, NGV743159H, NGV745386H, NGV745659H, NGV741943H, NGV742852H, NGV742881H, NGV742987H, NGV743052H, NGV743061H, NGV743100H, NGV743818H, NGV745107H, NGV745302H, NGV745581H, NGV746056H, NGV746095H, NGV748455H, NGV749090H, NGV749618H, NGV750084H, NGV750120H, NGV752083H, NGV742816H, NGV742931H, NGV744995H, NGV750092H, NGV741957H, NGV742976H, NGV743081H, NGV743099H, NGV743145H, NGV745080H, NGV745317H, NGV746021H, NGV749515H, NGV750087H, NGV750121H, NGV750754H, NGV752647H, NGV742001H, NGV742065H, NGV742077H, NGV742993H, NGV746077H, NGV748293H, NGV743018H, NGV743075H, NGV743078H, NGV745485H, NGV745088H, NGV749083H, NGV751925H, NGV745748H, NGV746067H, NGV750804H, NGV742879H, NGV748227H, NGV749398H, NGV746362H, NGV742010H, NGV742818H, NGV742932H, NGV743138H, NGV743771H, NGV746488H, NGV746507H, NGV748702H, NGV749540H, NGV750122H, NGV750766H, NGV751911H, NGV743092H, NGV743093H, NGV745350H, NGV746014H, NGV747094H, NGV748561H, NGV748602H, NGV748962H, NGV750123H, NGV750761H, NGV751062H, NGV752798H, NGV752801H, NGV753378H, NGV747103H, NGV751382H, NGV745613H, NGV742824H, NGV749525H, NGV743782H, NGV745633H, NGV745858H, NGV748233H, NGV746110H, NGV748692H, NGV743816H, NGV754048H, NGV742812H, NGV742833H, NGV742857H, NGV752553H, NGV753423H, NGV754070H, NGV754071H, NGV749519H, NGV744738H, NGV750072H, NGV752240H, NGV742903H, NGV744868H, NGV748345H, NGV748365H, NGV750765H, NGV745854H, NGV746282H, NGV748336H, NGV750806H, NGV749162H, NGV749592H, NGV754130H, NGV752897H, NGV752917H, NGV746219H, NGV753415H, NGV752085H, NGV752086H, NGV749584H, NGV749586H, NGV747115H, NGV753546H, NGV745121H, NGV753052H, NGV752034H, NGV749754H, NGV751445H, NGV748341H, NGV754133H, NGV750077H, NGV748412H, NGV750076H, NGV751215H, NGV753435H, NGV752786H, NGV753488H, NGV753550H, NGV753263H, NGV753404H, NGV752664H, NGV743120H, NGV742954H, NGV751214H, NGV752524H, NGV749299H, NGV749919H, NGV753932H, NGV743236H, NGV745150H, NGV745112H, NGV746273H, NGV742742H, NGV751277H, NGV752531H, NGV743157H, NGV750573H, NGV742817H, NGV742837H, NGV742862H, NGV742866H, NGV742867H, NGV745668H, NGV745669H, NGV745708H, NGV745713H, NGV745723H, NGV745724H, NGV745726H, NGV745730H, NGV745733H, NGV745734H, NGV746247H, NGV748251H, NGV752223H, NGV753507H, NGV742901H, NGV742947H, NGV748613H, NGV751449H, NGV745603H, NGV750254H, NGV742921H, NGV745756H, NGV743136H, NGV748308H, NGV749555H, NGV751153H, NGV751826H, NGV752258H, NGV745600H, NGV752525H, NGV742929H, NGV742964H, NGV744873H, NGV744876H, NGV744880H, NGV744886H, NGV744890H, NGV744895H, NGV744898H, NGV744899H, NGV746292H, NGV749932H, NGV749962H, NGV749983H, NGV750044H, NGV752501H, NGV752822H, NGV753151H, NGV742961H, NGV743770H, NGV742023H, NGV742026H, NGV742055H, NGV744044H, NGV744911H, NGV746509H, NGV748043H, NGV748346H, NGV750797H, NGV751377H, NGV751447H, NGV753652H, NGV753797H, NGV754333H, NGV742014H, NGV744872H, NGV744979H, NGV748612H, NGV745739H, NGV742762H, NGV742955H, NGV749151H, NGV749159H, NGV749587H, NGV752511H, NGV752569H, NGV753553H, NGV753620H, NGV742814H, NGV745234H, NGV745570H, NGV750584H, NGV746113H, NGV746115H, NGV746116H, NGV750579H, NGV750580H, NGV750581H, NGV745407H, NGV745645H, NGV750476H, NGV745863H, NGV748676H, NGV749140H, NGV743240H, NGV745106H, NGV745578H, NGV741931H, NGV745609H, NGV745611H, NGV749030H, NGV750512H, NGV751177H, NGV752247H, NGV752248H, NGV752250H, NGV752253H, NGV753407H, NGV753408H, NGV753409H, NGV753411H, NGV753414H, NGV742048H, NGV742058H, NGV748265H, NGV748285H, NGV748291H, NGV749167H, NGV749193H, NGV753494H, NGV746367H, NGV746489H, NGV750511H, NGV745859H, NGV748301H, NGV748411H, NGV749290H, NGV752498H, NGV753277H, NGV746582H, NGV748358H, NGV750652H, NGV745086H, NGV753226H, NGV753269H, NGV750487H, NGV747119H, NGV745576H, NGV748331H, NGV742963H, NGV743790H, NGV746123H, NGV746508H, NGV748551H, NGV750234H, NGV751364H, NGV751819H, NGV752280H, NGV752344H, NGV752776H, NGV753902H, NGV743053H, NGV743057H, NGV746208H, NGV752679H, NGV746124H, NGV746128H, NGV746130H, NGV746136H, NGV746137H, NGV745103H, NGV749034H, NGV749916H, NGV742994H, NGV745479H, NGV746117H, NGV749069H, NGV749755H, NGV745154H, NGV750243H, NGV754148H, NGV747116H, NGV748269H, NGV742957H, NGV745162H, NGV745369H, NGV745660H, NGV750094H, NGV751437H, NGV745753H, NGV745755H, NGV745758H, NGV745760H, NGV745860H, NGV746266H, NGV748240H, NGV748243H, NGV748245H, NGV749769H, NGV749770H, NGV749771H, NGV750543H, NGV750547H, NGV751412H, NGV751421H, NGV751424H, NGV752654H, NGV753348H, NGV753417H, NGV753418H, NGV753421H, NGV744010H, NGV750575H, NGV746105H, NGV746106H, NGV753050H, NGV745563H, NGV745564H, NGV745646H, NGV750800H, NGV752678H, NGV752858H, NGV748307H, NGV748615H, NGV751947H, NGV751948H, NGV746057H, NGV746571H, NGV751821H, NGV750576H, NGV742938H, NGV749589H, NGV750809H, NGV751828H, NGV751829H, NGV753048H, NGV745656H, NGV748276H, NGV749579H, NGV745657H, NGV745852H, NGV751205H, NGV745097H, NGV748034H, NGV751005H, NGV745125H, NGV745368H, NGV748263H, NGV748266H, NGV748299H, NGV749518H, NGV751255H, NGV753914H, NGV754353H, NGV745019H, NGV744004H, NGV746254H, NGV749084H, NGV746250H, NGV751924H, NGV752261H, NGV742939H, NGV748414H, NGV753285H, NGV753575H, NGV742965H, NGV750751H, NGV753931H, NGV746079H, NGV745008H, NGV745868H, NGV746277H, NGV746297H, NGV748353H, NGV751452H, NGV751453H, NGV753278H, NGV751362H, NGV753574H, NGV743831H, NGV743839H, NGV745255H, NGV745276H, NGV750778H, NGV751460H, NGV751921H, NGV752676H, NGV752790H, NGV745568H, NGV749751H, NGV749758H, NGV750240H, NGV750847H, NGV751047H, NGV751461H, NGV753549H, NGV745593H, NGV748532H, NGV746558H, NGV750112H, NGV753915H, NGV748272H, NGV753262H, NGV753433H, NGV743985H, NGV745100H, NGV751458H, NGV745034H, NGV752914H, NGV752275H, NGV752284H, NGV752242H, NGV743845H, NGV746246H, NGV749147H, NGV749148H, NGV750801H, NGV753028H, NGV753551H, NGV742861H, NGV743841H, NGV752241H, NGV748344H, NGV752669H, NGV754147H, NGV742966H, NGV746121H, NGV746205H, NGV746207H, NGV746491H, NGV746510H, NGV746511H, NGV748060H, NGV748274H, NGV748419H, NGV748444H, NGV749390H, NGV749590H, NGV750050H, NGV750051H, NGV750520H, NGV750521H, NGV750532H, NGV751210H, NGV751457H, NGV752572H, NGV745035H, NGV745064H, NGV745571H, NGV746249H, NGV746495H, NGV748040H, NGV748290H, NGV748310H, NGV748313H, NGV748325H, NGV748334H, NGV750519H, NGV751217H.NGV751456H, NGV746255H, NGV746265H, NGV746119H, NGV748213H, NGV750073H, NGV752645H, NGV742887H, NGV743850H, NGV745286H, NGV745290H, NGV746218H, NGV749903H, NGV753416H, NGV749714H, NGV749736H, NGV749739H, NGV752921H, NGV753290H, NGV753291H, NGV753328H, NGV752565H, NGV742864H, NGV745596H, NGV748295H, NGV749938H, NGV751813H, NGV753256H, NGV753422H, NGV742876H, NGV746344H, NGV751208H, NGV745681H, NGV749740H, NGV753143H, NGV753148H, NGV753288H, NGV752082H, NGV745573H, NGV745575H, NGV751470H, NGV753398H, NGV752559H, NGV742854H, NGV747120H, NGV742601H, NGV742615H, NGV742623H, NGV752789H, NGV750001H, NGV750052H, NGV750056H, NGV741925H, NGV741932H, NGV741939H, NGV751105H, NGV751107H, NGV751109H, NGV751146H, NGV748662H, NGV749401H, NGV746498H, NGV746099H, NGV746102H, NGV746484H, NGV746485H, NGV752758H, NGV752769H, NGV753577H, NGV753579H, NGV745007H, NGV750821H, NGV754038H, NGV749408H, NGV748288H, NGV743074H, NGV744002H, NGV754026H, NGV743165H, NGV744978H, NGV746257H, NGV751183H, NGV741951H, NGV750071H, NGV751101H, NGV751372H, NGV752843H, NGV743802H, NGV745630H, NGV748032H, NGV752844H, NGV749930H, NGV749956H, NGV748298H, NGV743005H, NGV745562H, NGV746215H, NGV748326H, NGV751103H, NGV750663H, NGV742677H, NGV742985H, NGV750514H, NGV753647H, NGV753648H, NGV749048H, NGV750100H, NGV750474H, NGV750668H, NGV743019H, NGV743768H, NGV749597H, NGV749598H, NGV751913H, NGV746107H, NGV749300H, NGV749527H, NGV748360H, NGV749741H, NGV753399H, NGV751803H, NGV753271H, NGV748553H, NGV747112H, NGV748356H, NGV748554H, NGV752668H, NGV743158H, NGV746071H, NGV745082H, NGV745094H, NGV745095H, NGV745122H, NGV746267H, NGV750164H, NGV751940H, NGV751961H, NGV753934H, NGV754126H, NGV742959H, NGV743060H, NGV749500H, NGV752521H, NGV752523H, NGV745126H, NGV751193H, NGV743162H, NGV750810H, NGV751054H, NGV751172H, NGV751173H, NGV753424H, NGV753436H, NGV754078H, NGV754124H, NGV742869H, NGV748211H, NGV748534H, NGV749024H, NGV745288H, NGV746514H, NGV746563H, NGV747137H, NGV748030H, NGV748050H, NGV748110H, NGV748690H, NGV749606H, NGV750009H, NGV750619H, NGV753243H, NGV746256H, NGV743089H, NGV744042H, NGV744055H, NGV745108H, NGV746022H, NGV749389H, NGV749027H, NGV754597H, NGV742682H, NGV742685H, NGV750329H, NGV750455H, NGV745579H, NGV745566H, NGV742686H, NGV752273H, NGV750101H, NGV748546H, NGV750131H, NGV748985H, NGV745665H, NGV743036H, NGV746577H, NGV745842H, NGV748041H, NGV742915H, NGV743917H, NGV746258H, NGV742009H, NGV743126H, NGV743127H, NGV743801H, NGV745374H, NGV743823H, NGV745062H, NGV743797H, NGV743828H, NGV745030H, NGV746260H, NGV749085H, NGV742909H, NGV749522H, NGV745087H, NGV746153H, NGV743031H, NGV743032H, NGV743063H, NGV743066H, NGV743800H, NGV743836H, NGV745587H, NGV745592H, NGV746264H, NGV748723H, NGV748881H, NGV749898H, NGV750753H, NGV751174H, NGV751188H, NGV751401H, NGV752263H, NGV752538H, NGV752911H, NGV753032H, NGV753038H, NGV744928H, NGV744935H, NGV744942H, NGV744973H, NGV744998H, NGV744045H, NGV749762H, NGV747122H, NGV742667H, NGV745089H, NGV745747H, NGV750501H, NGV743214H, NGV746263H, NGV748686H, NGV748839H, NGV746086H, NGV746259H, NGV743907H, NGV746252H, NGV746253H, NGV750854H, NGV750857H, NGV743118H, NGV744870H, NGV744871H, NGV751440H, NGV743102H, NGV751830H, NGV751835H, NGV752079H, NGV743854H, NGV749402H, NGV752539H, NGV742845H, NGV742930H, NGV746097H, NGV747130H, NGV747164H, NGV750803H, NGV751969H, NGV752570H, NGV753075H, NGV753127H, NGV753429H, NGV753430H, NGV741940H, NGV741941H, NGV741945H, NGV741950H, NGV748230H, NGV748523H, NGV749031H, NGV749050H, NGV750557H, NGV750561H, NGV750569H, NGV750570H, NGV750572H, NGV748294H, NGV751433H, NGV752497H, NGV751753H, NGV743034H, NGV743027H, NGV746078H, NGV751393H, NGV743006H, NGV743015H, NGV743022H, NGV745077H, NGV745078H, NGV745079H, NGV745084H, NGV746192H, NGV746196H, NGV746197H, NGV746199H, NGV746200H, NGV746201H, NGV746202H, NGV748059H, NGV748119H, NGV748244H, NGV748248H, NGV748260H, NGV751799H, NGV751839H, NGV751845H, NGV751853H, NGV751860H, NGV751864H, NGV754135H, NGV754136H, NGV754139H, NGV754301H, NGV754317H, NGV754412H, NGV754424H, NGV754433H, NGV744922H, NGV746059H, NGV750537H, NGV751154H, NGV742904H, NGV742913H, NGV742916H, NGV742925H, NGV746500H, NGV746503H, NGV750224H, NGV750232H, NGV752386H, NGV753629H, NGV749639H, NGV745093H, NGV750468H, NGV743997H, NGV752665H, NGV742877H, NGV751917H, NGV751918H, NGV752077H, NGV743059H, NGV751380H, NGV748281H, NGV750096H, NGV751903H, NGV751904H, NGV752080H, NGV742084H, NGV742933H, NGV752729H, NGV754365H; GTIN 00763000689537, Serial Numbers: NGV743096H, NGV743151H, NGV743161H, NGV748930H, NGV748668H, NGV745101H, NGV751817H, NGV749960H, NGV745662H, NGV744983H, NGV750672H, NGV743898H, NGV743931H, NGV744704H, NGV745001H, NGV745401H, NGV745658H, NGV745666H, NGV745759H, NGV749729H, NGV749897H, NGV751181H, NGV751212H, NGV751807H, NGV743808H, NGV743814H, NGV744694H, NGV744698H, NGV745311H, NGV745315H, NGV749822H, NGV749927H, NGV749928H, NGV751389H, NGV751815H, NGV744192H, NGV743070H, NGV745259H, NGV749894H, NGV750675H, NGV751179H, NGV744869H, NGV753555H, NGV746585H, NGV743086H, NGV744039H, NGV745002H, NGV743103H, NGV743123H, NGV743147H, NGV749260H, NGV749720H; GTIN 00763000689551, Serial Numbers: NGV746555H, NGV746561H, NGV746576H, NGV744021H, NGV744290H, NGV744303H, NGV744306H, NGV744307H, NGV744883H, NGV744888H, NGV744897H, NGV744903H, NGV744908H, NGV744914H, NGV745273H, NGV745307H, NGV745312H, NGV745675H, NGV745709H, NGV745712H, NGV745716H, NGV746554H; GTIN 00763000689568, Serial Numbers: NGV745139H, NGV745215H, NGV745005H, NGV745324H, NGV745493H, NGV745501H, NGV744529H, NGV744403H, NGV745023H, NGV744220H, NGV744233H, NGV745116H, NGV745205H, NGV745371H, NGV745494H, NGV742709H, NGV742711H, NGV742717H, NGV744246H, NGV744412H, NGV745098H, NGV745109H, NGV745110H, NGV745117H, NGV745120H, NGV745351H, NGV745480H, NGV742683H, NGV742694H, NGV742730H, NGV742731H, NGV744241H, NGV744245H, NGV744453H, NGV745012H, NGV745014H, NGV745040H, NGV745085H, NGV745477H, NGV745481H, NGV745484H, NGV745487H, NGV745490H, NGV745491H, NGV745506H, NGV745508H, NGV745509H, NGV744228H, NGV744986H, NGV745328H, NGV745313H, NGV745367H, NGV744226H, NGV744460H, NGV746083H, NGV746028H, NGV743806H, NGV743819H, NGV743799H, NGV743792H, NGV742914H, NGV742940H, NGV742967H, NGV743745H, NGV742986H, NGV743045H, NGV743772H, NGV743809H, NGV743013H, NGV742733H, NGV742896H, NGV742898H, NGV742899H, NGV742983H, NGV742924H, NGV743810H, NGV742980H, NGV742973H, NGV743023H, NGV742910H, NGV742981H, NGV743787H, NGV743827H, NGV742144H, NGV742154H, NGV742157H, NGV742163H, NGV742170H, NGV742174H, NGV742183H, NGV746187H, NGV746191H, NGV746194H, NGV746195H, NGV746204H, NGV746206H, NGV746213H, NGV749303H, NGV749307H, NGV750500H, NGV750503H, NGV750506H, NGV750507H, NGV750516H, NGV750523H, NGV750595H, NGV751897H, NGV751922H, NGV751939H, NGV751944H, NGV751957H, NGV751959H, NGV751971H, NGV751972H, NGV751978H, NGV745707H, NGV745719H, NGV745728H, NGV745738H, NGV745749H, NGV745750H, NGV745754H, NGV742893H, NGV742895H, NGV751989H, NGV751993H, NGV751997H, NGV752000H, NGV752097H, NGV752119H, NGV745442H, NGV745444H, NGV752327H, NGV752427H, NGV751501H, NGV742889H, NGV742892H, NGV744048H, NGV750079H, NGV750083H, NGV743121H, NGV743233H, NGV746458H, NGV750060H, NGV751481H, NGV751359H, NGV751494H, NGV743095H, NGV743114H, NGV749982H, NGV751345H, NGV751346H, NGV751471H, NGV751479H, NGV746471H, NGV743125H, NGV744063H, NGV750601H, NGV751333H, NGV751337H, NGV742990H, NGV746452H, NGV750037H, NGV751483H, NGV751500H, NGV743119H, NGV751381H, NGV742884H, NGV742890H, NGV745472H, NGV751871H, NGV751875H, NGV751886H, NGV745473H, NGV746442H, NGV748998H, NGV749014H, NGV749022H, NGV742865H, NGV743759H, NGV743764H, NGV743766H, NGV749432H, NGV742883H, NGV742139H, NGV749451H, NGV749025H, NGV749033H, NGV742670H, NGV743097H, NGV745445H, NGV749054H, NGV742734H, NGV746346H, NGV746439H, NGV748944H, NGV742747H, NGV745843H, NGV746161H, NGV748952H, NGV748993H, NGV742136H, NGV745099H, NGV748241H, NGV745847H, NGV746027H, NGV746146H, NGV746118H, NGV746284H, NGV746272H, NGV742106H, NGV742121H, NGV742555H, NGV745607H, NGV748277H, NGV743130H, NGV745452H, NGV746268H, NGV746152H, NGV749043H, NGV742129H, NGV749784H, NGV746275H, NGV745845H, NGV742948H, NGV746278H, NGV742564H, NGV746328H, NGV742134H, NGV749860H, NGV743128H, NGV745440H, NGV748267H, NGV746270H, NGV751895H, NGV745242H, NGV745835H, NGV745474H, NGV745602H, NGV745648H, NGV745676H, NGV742109H, NGV742114H, NGV742128H, NGV742546H, NGV746310H, NGV749302H, NGV750093H, NGV745383H, NGV748920H, NGV749037H, NGV751986H, NGV751987H, NGV746280H, NGV746289H, NGV746290H, NGV746293H, NGV751454H, NGV751464H, NGV742763H, NGV751466H, NGV751532H, NGV742847H, NGV751676H, NGV752226H, NGV746294H, NGV749410H, NGV749412H, NGV742856H, NGV742858H, NGV746042H, NGV746288H, NGV751983H, NGV745482H, NGV751866H, NGV752033H, NGV745441H, NGV749445H; GTIN 00763000689575, Serial Numbers: NGV743008H, NGV743026H, NGV743048H, NGV743028H, NGV743033H, NGV751678H, NGV742970H, NGV742971H, NGV742974H, NGV742975H, NGV742995H, NGV742998H, NGV743000H, NGV742811H, NGV742821H, NGV748788H, NGV748789H, NGV748792H, NGV748795H, NGV748798H, NGV748801H, NGV748804H, NGV748808H, NGV748810H, NGV748812H, NGV748837H, NGV749946H, NGV746494H, NGV746496H, NGV746506H, NGV746512H, NGV746515H, NGV746583H, NGV748047H, NGV748051H, NGV748114H, NGV748200H, NGV748264H, NGV748270H, NGV748271H, NGV748283H, NGV748297H, NGV748779H, NGV748821H, NGV748826H, NGV748831H, NGV749722H, NGV749954H, NGV750653H, NGV750654H, NGV750660H, NGV750661H, NGV750834H, NGV750851H, NGV751475H, NGV751683H, NGV751851H, NGV751865H, NGV754062H, NGV742823H, NGV754049H, NGV754056H
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall4 deaths1,964 reports
- SYNCHROMED®II 8637-40Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths1,478 reports
- SYNCHROMED®II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall1 death94 reports
Product lines
Other devices under the same FDA submission
44 devices cleared under P860004 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
- SynchroMed® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths1,920 reports
- SYNCHROMED™ III 8667-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 death1,937 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 deaths1,766 reports
- SYNCHROMED®II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall2 deaths1,646 reports
- SYNCHROMED™ III 8667-40Medtronic, Inc. · Pump, infusion, implanted, programmable2 deaths1,421 reports
- SYNCHROMED®II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall71 reports
- ASCENDA® 8780Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall33 reports
- ASCENDA® 8781Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall14 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93481).
- Z-0614-2024 · Open, ClassifiedMedtronic SynchroMed II, Model 8637-20, Programmable pump
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page