Skip to content

RadiaLux™ 50-101-1 (UDI 00763000432805)Recalls & reported problems

Medtronic Navigation, Inc. · RETRACTOR 50-101-1 RADIALUX CE PINK · Class 1

All models of RadiaLux (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000432805), received between Oct 1, 2022 and Oct 1, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Dec 8, 2022Open, Classified

    Defect in sterile pouch seal, incomplete seal, of lighted Retractors, increases the risk of contamination, which can lead to post-operative infection.

    Z-0953-2023 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break2
  • Failure to Power Up2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 1, 2022Break; Failure to Power Up; Break; Failure to Power Up

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING TWO HANDPIECES. IT WAS REPORTED THAT THE SITE HAS TWO HANDPIECES THAT WOULD NOT POWER ON BECAUSE WHEN THE PROTECTIVE TAB WAS REMOVED, IT BROKE, PREVENTING PROPER CONTACT BETWEEN WITH THE BATTERY. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15525708 ↗
  • MalfunctionOct 1, 2022Break; Failure to Power Up; Break; Failure to Power Up

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING TWO HANDPIECES. IT WAS REPORTED THAT THE SITE HAS TWO HANDPIECES THAT WOULD NOT POWER ON BECAUSE WHEN THE PROTECTIVE TAB WAS REMOVED, IT BROKE, PREVENTING PROPER CONTACT BETWEEN WITH THE BATTERY. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15524517 ↗

Other RETRACTOR devices

All RETRACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page