Skip to content

Omni-Tract® 11275Recalls & reported problems

Integra Lifesciences Corporation · Omni-Tract® Speed-Tract®, 1in x 1in (2.5cm x 2.5cm) Coupler · Class 1

All models of Omni-Tract (447)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10381780271604), received between Oct 2, 2024 and Oct 7, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Corroded” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 5, 2026Class IIOpen, Classified

    The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

    Z-2714-2026 · Root cause: Labeling design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Corroded1
  • Problem with Sterilization1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 7, 2024

    NEW INTEGRA INSTRUMENTS ARRIVED WITH RUST. THESE ARE NEW INSTRUMENTS IN THE PACKAGE CURRENTLY UN-STERILE BUT INTENDED FOR STERILIZATION. UPON OPENING THE PACKAGE THERE WAS A LOT OF VISIBLE RUST PRESENT ON THE INSTRUMENTS. THIS IS THE SECOND ORDER OF INSTRUMENTS THAT HAVE ARRIVED IN THIS CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20384439 ↗
  • MalfunctionOct 2, 2024Corroded; Problem with Sterilization

    NEW INTEGRA INSTRUMENTS ARRIVED WITH RUST. THESE ARE NEW INSTRUMENTS IN THE PACKAGE CURRENTLY NON-STERILE BUT INTENDED FOR STERILIZATION. UPON OPENING THE PACKAGING THERE WAS A LOT OF VISIBLE RUST PRESENT ON THE INSTRUMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20350518 ↗

Other RETRACTOR devices

All RETRACTOR devices →

Sources and updates