Omni-Tract® 4040Recalls & reported problems
Integra Lifesciences Corporation · Omni-Tract® SnapClamp™ Assembly · Class 1
All models of Omni-Tract (447)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10381780270638), received between Mar 5, 2025 and Feb 5, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jun 5, 2026Open, Classified
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Z-2714-2026 · Root cause: Labeling design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2026Break
A FACILITY REPORTED THAT DURING A SURGICAL PROCEDURE, THE SNAP CLAMP ASSEMBLY (4040) EXHIBITED A DEFECT, WHICH WAS A TIGHTENING PROBLEM ON THE HOOP THAT RESULTED IN A 30-MINUTE DELAY IN SURGICAL TIME. ANOTHER SURGICAL INSTRUMENT WAS USED TO COMPLETE THE SURGERY. NO PATIENT INJURIES OR DEATHS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24267958 ↗MalfunctionJul 1, 2025Mechanical Problem; Mechanical Problem
A FACILITY ENGINEER REPORTED THAT THE CLAMPING FORCE OF THE SNAP CLAMP ASSEMBLY (4040) ON THE RETRACTOR ROD WAS "TOO LOW, THAT THE CLAMP YIELDS TO TISSUE PRESSURE." THERE WAS NO PATIENT INJURY; HOWEVER, A SURGICAL DELAY EXCEEDING 30 MINUTES WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22375287 ↗MalfunctionMar 5, 2025Break; Break
A FACILITY REPORTED THAT DURING A "HIGHLY COMPLEX DIGESTIVE SURGERY INVOLVING A PANCREATICODUODENECTOMY WITH VASCULAR RESECTION", THE SNAP CLAMP ASSEMBLY (4040) "SPONTANEOUSLY BROKE WHILE BEING MANIPULATED." THE BROKEN RETRACTOR DID NOT FALL INTO THE SURGICAL SITE, AND ALL BROKEN PARTS WERE SUCCESSFULLY RECOVERED. THERE WAS NO INJURY TO THE PATIENT; HOWEVER, A SURGICAL DELAY OF TWO HOURS WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21532619 ↗
Other RETRACTOR devices
- Endo Paddle Retract 173046Covidien LP · RETRACTOR56 reports
- RadiaLux™ 50-101-1Medtronic Navigation, Inc. · RETRACTOR1 recall2 reports
- Omni-Tract® 11275Integra Lifesciences Corporation · RETRACTOR1 recall2 reports
- CHARNLEY 9620-04-000Depuy International Ltd · RETRACTOR1 recall2 reports
- NA 2176-10-000Depuy (Ireland) · RETRACTOR51 reports
- Fathom MR2-102050Seaspine Orthopedics Corporation · RETRACTOR1 recall1 report
- Omni-Tract® 3104RIntegra Lifesciences Corporation · RETRACTOR1 recall0 reports
- Omni-Tract® 3601Integra Lifesciences Corporation · RETRACTOR1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page