Omni-Tract® 4040Recalls & reported problems

Integra Lifesciences Corporation · Omni-Tract® SnapClamp™ Assembly · Class 1

All models of Omni-Tract (447)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10381780270638), received between Mar 5, 2025 and Feb 5, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 5, 2026Open, Classified

    The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

    Z-2714-2026 · Root cause: Labeling design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break2
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 5, 2026Break

    A FACILITY REPORTED THAT DURING A SURGICAL PROCEDURE, THE SNAP CLAMP ASSEMBLY (4040) EXHIBITED A DEFECT, WHICH WAS A TIGHTENING PROBLEM ON THE HOOP THAT RESULTED IN A 30-MINUTE DELAY IN SURGICAL TIME. ANOTHER SURGICAL INSTRUMENT WAS USED TO COMPLETE THE SURGERY. NO PATIENT INJURIES OR DEATHS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24267958 ↗
  • MalfunctionJul 1, 2025Mechanical Problem; Mechanical Problem

    A FACILITY ENGINEER REPORTED THAT THE CLAMPING FORCE OF THE SNAP CLAMP ASSEMBLY (4040) ON THE RETRACTOR ROD WAS "TOO LOW, THAT THE CLAMP YIELDS TO TISSUE PRESSURE." THERE WAS NO PATIENT INJURY; HOWEVER, A SURGICAL DELAY EXCEEDING 30 MINUTES WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22375287 ↗
  • MalfunctionMar 5, 2025Break; Break

    A FACILITY REPORTED THAT DURING A "HIGHLY COMPLEX DIGESTIVE SURGERY INVOLVING A PANCREATICODUODENECTOMY WITH VASCULAR RESECTION", THE SNAP CLAMP ASSEMBLY (4040) "SPONTANEOUSLY BROKE WHILE BEING MANIPULATED." THE BROKEN RETRACTOR DID NOT FALL INTO THE SURGICAL SITE, AND ALL BROKEN PARTS WERE SUCCESSFULLY RECOVERED. THERE WAS NO INJURY TO THE PATIENT; HOWEVER, A SURGICAL DELAY OF TWO HOURS WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21532619 ↗

Other RETRACTOR devices