Skip to content

MIDAS REX MR8 MR8-AF01 (UDI 00763000440985)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8-AF01 MR8 FOOTED F1 B5 2.4MM · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
25
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 00763000440985), received between Dec 19, 2025 and Sep 30, 2026 — classified as 25 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 4 reportsMay 2026: 1 reportsJun 2026: 4 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 9 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event6
  • Overheating of Device5
  • Detachment of Device or Device Component4
  • Naturally Worn3
  • Corroded2
  • Material Deformation2
  • Material Twisted/Bent2
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request and no alleged product deficiencies were reported. There was no patient impact reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943837 ↗
  • MalfunctionSep 27, 2026Difficult to Insert

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50910396 ↗
  • MalfunctionSep 25, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. At the time of report no patient impact was observed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50894800 ↗
  • MalfunctionSep 24, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885886 ↗
  • MalfunctionSep 24, 2026Material Deformation

    It was reported that during testing, the markings on the attachment noted to be damaged and illegible. There was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50881786 ↗
  • MalfunctionSep 23, 2026Detachment of Device or Device Component

    It was reported that two small ring fell from the inside of attachment after use. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863715 ↗
  • MalfunctionSep 23, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. There was no patient impact. Additional information was received stating that the bearings were ruptured.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863435 ↗
  • MalfunctionSep 10, 2026Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50716444 ↗
  • MalfunctionSep 10, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50716263 ↗
  • MalfunctionAug 21, 2026Detachment of Device or Device Component; Naturally Worn

    It was reported that prior to use, the attachment had a bearing issue. There was no patient involved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50361120 ↗
  • MalfunctionAug 10, 2026Vibration

    It was reported that the attachment had vibration. It was reported there was no patient involvement. Additional information was received stating that detached bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50190041 ↗
  • MalfunctionJul 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SMALL PIECES FELL OFF THE PRODUCT. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25676500 ↗
  • MalfunctionJun 18, 2026Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT DURING PRE-OP THE ATTACHMENT HAD BEARING ISSUE AND FOOT SLIGHTLY BENDED. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25549067 ↗
  • MalfunctionJun 18, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THERE WAS NO ALLEGED PRODUCT DEFICIENCY. AT THE TIME OF REPORT, THE IMPACT TO THE PATIENT WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 25542569 ↗
  • MalfunctionJun 16, 2026Overheating of Device

    IT WAS REPORTED THAT THE DURING TESTING ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT IMPACT.ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25515227 ↗
  • MalfunctionJun 11, 2026Naturally Worn

    IT WAS REPORTED THAT DURING PRE-OPERATIVE PREPARATION FOR USE THE BEARINGS WERE RUNNING ROUGH. THERE WAS NO PATIENT IMPACT IN THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON EVALUATION THAT THE BEARINGS WERE MISALIGNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25466272 ↗
  • MalfunctionMay 18, 2026Corroded

    IT WAS REPORTED THAT DURING IN-SERVICE THE DEVICE'S BEARING HAD RUST OR CORROSION ON IT. THERE WAS NO PATIENT INVOLVEMENT DURING THE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25209118 ↗
  • MalfunctionApr 29, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE DEVICE WAS SENT IN WITH NO REPORTED ISSUE. NO ADDITIONAL INFORMATION WAS PROVIDED. THERE WERE ANALYSIS FINDINGS.

    Patient problems: Insufficient Information

    MAUDE report 25031555 ↗
  • MalfunctionApr 10, 2026Corroded

    IT WAS REPORTED THAT DURING IN-SERVICE THE RUST AND CORROSION WERE OBSERVED. AT THE TIME OF REPORT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24838187 ↗
  • MalfunctionApr 6, 2026Overheating of Device

    IT WAS REPORTED THAT DURING PRE OPERATION THE ATTACHMENT HAD TEMPERATURE INCREASE. THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT MISALIGNED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24796097 ↗
  • MalfunctionApr 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BEARINGS DROPPED FROM THE ATTACHMENT PRIOR TO USE. AT THE TIME OF REPORT, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24754491 ↗
  • MalfunctionMar 31, 2026Overheating of Device

    IT WAS REPORTED THAT PRE-OPERATIVELY THE ATTACHMENT HAD A TEMPERATURE INCREASE OR WAS RUNNING HOT. THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24745551 ↗
  • MalfunctionMar 31, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT HAD TEMPERATURE INCREASE PRE-OPERATIVELY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24745320 ↗
  • MalfunctionMar 31, 2026Overheating of Device

    IT WAS REPORTED THAT THERE WAS TEMPERATURE INCREASE PRE-OPERATIVELY. NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24745052 ↗
  • MalfunctionDec 19, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT FOOT BENT FOR THE ATTACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23847083 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates