MIDAS REX MR8 MR8-AS09 (UDI 00763000441043)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8-AS09 MR8 9CM STRAIGHT 3.2MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Midas Rex powered surgical attachment is a drill accessory. Many of the 124 reports describe attachments that overheat, wear out, or have bearings that come loose, and most say no patient was affected. The reports include no deaths and no injuries. Reports rose to 119 in the last 12 months from 3 in the previous 12. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 128
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 128 adverse event reports that list this exact device (DI 00763000441043), received between Sep 5, 2024 and Sep 29, 2026 — classified as 128 malfunctions.
The most frequently reported specific device problem is “Naturally Worn” (35 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Naturally Worn35
- Overheating of Device25
- Detachment of Device or Device Component21
- No Apparent Adverse Event20
- Difficult to Insert13
- Corroded6
- Vibration6
- Noise, Audible4
Patient problems
- No Clinical Signs, Symptoms or Conditions125
- Insufficient Information3
Recent adverse event reports
The 25 most recent of 128 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Difficult to Insert
It was reported that before operation, the tool could not be inserted into the attachment. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927484 ↗MalfunctionSep 15, 2026Naturally Worn
It was reported that the attachment bearings are broken. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50767587 ↗MalfunctionSep 15, 2026Naturally Worn
It was reported that bearing looseness in attachment. It was reported that there was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50767405 ↗MalfunctionSep 11, 2026Loose or Intermittent Connection
It was reported that after operation, the bearings were loose in the device. There was no patient involvement and patient impact in this event. Additional information received on evaluation that bearings were detached.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50726673 ↗MalfunctionSep 10, 2026No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50707398 ↗MalfunctionSep 9, 2026Loose or Intermittent Connection
It was reported that bearing in attachment has come loose. There was no patient involvement. Additional information received stating loose and damaged bearing was found during evaluation made it reportable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50683716 ↗MalfunctionAug 28, 2026Difficult to Remove; Naturally Worn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50492673 ↗MalfunctionAug 28, 2026No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50492570 ↗MalfunctionAug 21, 2026Naturally Worn
It was reported that the bearing was damaged. There was no parent involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50361081 ↗MalfunctionJul 2, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. AT THE TIME OF THE REPORT, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25689344 ↗MalfunctionJun 30, 2026Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT WOULD NOT WORK MAY BE BEARING ISSUE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25662637 ↗MalfunctionJun 29, 2026Difficult to Insert
IT WAS REPORTED THAT PREOPERATIVELY TOOL WAS DIFFICULT TO BE INSERTED IN THE DEVICE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25649242 ↗MalfunctionJun 26, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DAMAGE BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25633773 ↗MalfunctionJun 26, 2026Overheating of Device
IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25632931 ↗MalfunctionJun 26, 2026Overheating of Device
IT WAS REPORTED THAT DEVICE GETS HOT PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25632619 ↗MalfunctionJun 25, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. .IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT LOOSE BEARING AT TUBE DISTAL WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25626437 ↗MalfunctionJun 25, 2026Overheating of Device
IT WAS REPORTED THAT THE DEVICE EXPERIENCED OVERHEATING. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25622470 ↗MalfunctionJun 25, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25620001 ↗MalfunctionJun 25, 2026Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT BEARINGS IS FAILED. IT WAS REPORTED THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25619903 ↗MalfunctionJun 25, 2026Naturally Worn
IT WAS REPORTED THAT DEVICE HAS A DAMAGED INTERNAL BEARING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT LOOSE BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25619441 ↗MalfunctionJun 25, 2026Naturally Worn
IT WAS REPORTED THAT THE DEVICE BEARING WAS FAILED DURING THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS FALL OUT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25617189 ↗MalfunctionJun 24, 2026Corroded; Naturally Worn
IT WAS REPORTED THAT AFTER OPERATION THE BEARING WAS CORRODED AND FRACTURED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25605848 ↗MalfunctionJun 24, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PROPERLY. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED ON EVALUATION THAT BEARINGS WERE DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25605326 ↗MalfunctionJun 24, 2026Naturally Worn
IT WAS REPORTED THAT ATTACHMENT HAD A BEARING PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25605230 ↗MalfunctionJun 24, 2026Overheating of Device
IT WAS REPORTED THAT DURING TESTING, THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604837 ↗
FDA clearance and classification
Cleared through 510(k) K183515 on May 12, 2019 (decision: Substantially Equivalent). Submitted by Medtronic Powered Surgical Solutions as “MR8 Drill System, Midas Rex MR8 ClearView Tools”. The FDA review took 145 days. FDA record ↗
The FDA regulates product code HBE (drills, burrs, trephines & accessories (simple, powered)) under 21 CFR 882.4310 as a Class II device. 21 CFR 882.4310 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the MIDAS REX MR8 MR8-AS09
Has the MIDAS REX MR8 MR8-AS09 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000441043), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the MIDAS REX MR8 MR8-AS09?
The FDA's MAUDE database lists 128 adverse event reports for this device, received between Sep 5, 2024 and Sep 29, 2026. In those reports, the most common specific device problems are “Naturally Worn” (35), “Overheating of Device” (25) and “Detachment of Device or Device Component” (21). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MIDAS REX MR8 MR8-AS09?
None of the reports are classified as deaths, 0 as injuries and 128 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the MIDAS REX MR8 MR8-AS09 FDA approved?
It was cleared through 510(k) K183515 on May 12, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the MIDAS REX MR8 MR8-AS09, and is it still sold?
The FDA's device database lists Medtronic Powered Surgical Solutions as the company responsible for it (brand MIDAS REX MR8, model MR8-AS09). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,624 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)930 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)414 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)221 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)186 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)184 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)182 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)164 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record