MIDAS REX MR8 MR8-AS15 (UDI 00763000441081)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8-AS15 MR8 15CM STRAIGHT 2.4MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00763000441081), received between Mar 4, 2026 and Sep 28, 2026 — classified as 18 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event6
- Overheating of Device6
- Naturally Worn4
- Corroded1
- Difficult to Insert1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026No Apparent Adverse Event
Medtronic received information regarding a preventive maintenance, and no alleged product deficiencies were reported. At the time of report there was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910729 ↗MalfunctionSep 16, 2026Naturally Worn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50788615 ↗MalfunctionJul 3, 2026Difficult to Insert
IT WAS REPORTED THAT, PRIOR TO AN OPERATION, THE ATTACHMENT HAD DIFFICULTY INSERTING. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING WAS MISALINGED DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25700920 ↗MalfunctionJun 25, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25621576 ↗MalfunctionJun 24, 2026Overheating of Device
IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25606523 ↗MalfunctionJun 24, 2026Naturally Worn
IT WAS REPORTED THAT THE DEVICE NOT WORKING PROPERLY. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DAMAGE BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604453 ↗MalfunctionJun 24, 2026Overheating of Device
IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604023 ↗MalfunctionJun 24, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DAMAGE BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604005 ↗MalfunctionJun 24, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25603697 ↗MalfunctionJun 23, 2026Overheating of Device
IT WAS REPORTED THAT THE DEVICE WITH THE REPORT OF TEMPERATURE INCREASE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25591861 ↗MalfunctionJun 22, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DAMAGED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25573168 ↗MalfunctionJun 19, 2026Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT HAS BEARING ISSUE PRIOR TO THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25558732 ↗MalfunctionJun 19, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25556551 ↗MalfunctionJun 18, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DAMAGE BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25542977 ↗MalfunctionJun 17, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF NO REPORTED ISSUE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS BROKEN WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25534848 ↗MalfunctionJun 9, 2026Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT HAD BEARING PROBLEM. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444250 ↗MalfunctionApr 28, 2026Overheating of Device
IT WAS REPORTED THAT THE PRODUCTS WERE OVERHEATING DURING THE PROCEDURE. THERE WAS NO PATIENT IMPACT OR CASE DELAY NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25010866 ↗MalfunctionMar 4, 2026Corroded
IT WAS REPORTED THAT THE DEVICE WITH THE REPORT OF CORROSION. IT WAS REPORTED THAT THERE IS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT MISALIGNED DISTAL BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24510723 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page