Skip to content

MIDAS REX MR8 MR8-ASMC09 (UDI 00763000441173)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8-ASMC09 MR8 9CM METAL CUT 3.2MM · Class 2

Adverse event reports
17
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00763000441173), received between Apr 2, 2025 and Jul 4, 2026 — classified as 17 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 5 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 4 reportsJun 2026: 4 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device9
  • Detachment of Device or Device Component2
  • Difficult to Insert2
  • Difficult to Remove2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Displays Incorrect Message1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • Insufficient Information3

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 4, 2026Difficult to Remove

    IT WAS REPORTED THAT DURING POST OPERATION DRILL BIT STUCK. AT THE TIME OF REPORT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT SEIZED DISTAL BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25709728 ↗
  • MalfunctionJun 23, 2026Difficult to Remove

    IT WAS REPORTED THAT DURING PRE-OP THE DRILL BIT GOT STUCK . AT THE TIME OF REPORT NO PATIENT OR STAFF IMPACT IS BEEN OBSERVED . ADDITIONAL INFORMATION RECEIVED THAT BEARINGS WERE SEIZED DURING EVALUATION .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25590813 ↗
  • MalfunctionJun 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A DEVICE WAS SENT IN FOR SCHEDULED PREVENTIVE MAINTENANCE. THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25487228 ↗
  • MalfunctionJun 11, 2026Overheating of Device

    IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE, THE ATTACHMENT TEMPERATURE INCREASES. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25470115 ↗
  • MalfunctionJun 11, 2026Difficult to Insert

    IT WAS REPORTED THAT DURING SERVICING, THE ATTACHMENT WAS DIFFICULT TO MOUNT OR ATTACH. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25466943 ↗
  • MalfunctionMay 29, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEVICE BEARING WAS FELL OUT. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326693 ↗
  • MalfunctionMay 27, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT BEARING WAS DETACHED. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25300660 ↗
  • MalfunctionMay 22, 2026Difficult to Insert

    IT WAS REPORTED PRE-OPERATIVELY THAT IT WAS DIFFICULT TO INSERT THE TOOL/BLADE INTO THE ATTACHMENT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25256760 ↗
  • MalfunctionMay 22, 2026Overheating of Device

    IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE, A TEMPERATURE INCREASE WAS OBSERVED IN THE ATTACHMENT. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED ON EVALUATION STATING THAT DISTAL BEARING WAS MISALIGNED MADE THIS EVENT REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25252455 ↗
  • MalfunctionApr 27, 2026Overheating of Device

    IT WAS REPORTED THAT DEVICE TEMPERATURE INCREASED. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24994390 ↗
  • MalfunctionFeb 4, 2026Naturally Worn

    IT WAS REPORTED THAT THE DEVICE WITH THE REPORT OF BEARING FAILED. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24257097 ↗
  • MalfunctionNov 7, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE MOTOR WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23504343 ↗
  • MalfunctionNov 7, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WAS ALLEGED TO HAVE ISSUES OF OVERHEATING, RUNNING HOT, AND HEATING UP. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION STATES THAT THE ISSUE WAS IDENTIFIED BEFORE SERVICE AND THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23494594 ↗
  • MalfunctionNov 7, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 23494097 ↗
  • MalfunctionNov 7, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT OVERHEATING. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 23494092 ↗
  • MalfunctionNov 3, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 23448160 ↗
  • MalfunctionApr 2, 2025Overheating of Device; Device Displays Incorrect Message; Overheating of Device; Device Displays Incorrect Message

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING AND ERROR MESSAGE GIVEN. NO PATIENT IMPACT REPORTED. ON FOLLOW-UP IT WAS REPORTED THAT THERE WERE NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21750307 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.