MIDAS REX MR8 MR8-ASMC09 (UDI 00763000441173)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8-ASMC09 MR8 9CM METAL CUT 3.2MM · Class 2
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00763000441173), received between Apr 2, 2025 and Jul 4, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device9
- Detachment of Device or Device Component2
- Difficult to Insert2
- Difficult to Remove2
- Adverse Event Without Identified Device or Use Problem1
- Device Displays Incorrect Message1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- Insufficient Information3
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 4, 2026Difficult to Remove
IT WAS REPORTED THAT DURING POST OPERATION DRILL BIT STUCK. AT THE TIME OF REPORT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT SEIZED DISTAL BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25709728 ↗MalfunctionJun 23, 2026Difficult to Remove
IT WAS REPORTED THAT DURING PRE-OP THE DRILL BIT GOT STUCK . AT THE TIME OF REPORT NO PATIENT OR STAFF IMPACT IS BEEN OBSERVED . ADDITIONAL INFORMATION RECEIVED THAT BEARINGS WERE SEIZED DURING EVALUATION .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25590813 ↗MalfunctionJun 12, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A DEVICE WAS SENT IN FOR SCHEDULED PREVENTIVE MAINTENANCE. THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25487228 ↗MalfunctionJun 11, 2026Overheating of Device
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE, THE ATTACHMENT TEMPERATURE INCREASES. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25470115 ↗MalfunctionJun 11, 2026Difficult to Insert
IT WAS REPORTED THAT DURING SERVICING, THE ATTACHMENT WAS DIFFICULT TO MOUNT OR ATTACH. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25466943 ↗MalfunctionMay 29, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE BEARING WAS FELL OUT. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25326693 ↗MalfunctionMay 27, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT BEARING WAS DETACHED. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25300660 ↗MalfunctionMay 22, 2026Difficult to Insert
IT WAS REPORTED PRE-OPERATIVELY THAT IT WAS DIFFICULT TO INSERT THE TOOL/BLADE INTO THE ATTACHMENT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25256760 ↗MalfunctionMay 22, 2026Overheating of Device
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE, A TEMPERATURE INCREASE WAS OBSERVED IN THE ATTACHMENT. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED ON EVALUATION STATING THAT DISTAL BEARING WAS MISALIGNED MADE THIS EVENT REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25252455 ↗MalfunctionApr 27, 2026Overheating of Device
IT WAS REPORTED THAT DEVICE TEMPERATURE INCREASED. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24994390 ↗MalfunctionFeb 4, 2026Naturally Worn
IT WAS REPORTED THAT THE DEVICE WITH THE REPORT OF BEARING FAILED. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257097 ↗MalfunctionNov 7, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE MOTOR WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23504343 ↗MalfunctionNov 7, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS ALLEGED TO HAVE ISSUES OF OVERHEATING, RUNNING HOT, AND HEATING UP. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION STATES THAT THE ISSUE WAS IDENTIFIED BEFORE SERVICE AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23494594 ↗MalfunctionNov 7, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT.
Patient problems: Insufficient Information
MAUDE report 23494097 ↗MalfunctionNov 7, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT OVERHEATING. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 23494092 ↗MalfunctionNov 3, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO KNOWN PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23448160 ↗MalfunctionApr 2, 2025Overheating of Device; Device Displays Incorrect Message; Overheating of Device; Device Displays Incorrect Message
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING AND ERROR MESSAGE GIVEN. NO PATIENT IMPACT REPORTED. ON FOLLOW-UP IT WAS REPORTED THAT THERE WERE NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21750307 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.