MIDAS REX MR8 MR8-AVA10 (UDI 00763000441265)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8-AVA10 MR8 10CM VAR ANG 2.4MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Midas Rex MR8 surgical drill appears in 91 adverse event reports, with no deaths and no injuries. Many reports describe overheating attachments, wobbling drill bits, and loss of power, with some noting detached bearings found during evaluation. Most reports list no patient impact. There were 77 reports in the past 12 months, up from 3 in the year before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 100
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 100 adverse event reports that list this exact device (DI 00763000441265), received between Jun 18, 2024 and Sep 28, 2026 — classified as 1 injury, 99 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event46
- Detachment of Device or Device Component12
- Difficult to Insert11
- Naturally Worn11
- Overheating of Device11
- Loss of Power6
- Vibration5
- Fluid/Blood Leak4
Patient problems
- No Clinical Signs, Symptoms or Conditions93
- Insufficient Information6
- Cerebrospinal Fluid Leakage1
- Dural Tear1
- Hydrocephalus1
- Hypovolemia1
Recent adverse event reports
The 25 most recent of 100 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Unintended System Motion
It was reported that, during a surgical procedure, decompression was performed under general anesthesia at L2/3 on the right, with c ontralateral over-the-top decompression on the left, as well as left-sided decompression at L4/5. During the L2/3 decompression, the drill unexpectedly became blocked and subsequently malfunctioned. At the same time, the patient made a sudden and forceful movement of the lower limb despite general anesthesia. Microscopic inspection revealed a focal dural opening adjacent to the right L3 nerve root with low-volume CSF leakage. An intraoperative nerve root injury was suspected. The decompression was carefully continued, andthe right L3 nerve root and contralateral spinal canal were adequately decompressed. The dural opening was treated locally with TachoSil. Hemostasis was achieved, and a Valsalva maneuver at the end of the repair showed no persistent CSF leakage. Additional information was received from the surgeon clarified that the instrument remained securely held, but the burr attachment suddenly thrust forward while operating at full speed. The event occurred within milliseconds without warning, and the exact mechanism could not beconfirmed retros
Patient problems: Cerebrospinal Fluid Leakage; Dural Tear; Hydrocephalus; Hypovolemia
MAUDE report 50916319 ↗MalfunctionSep 23, 2026No Apparent Adverse Event
It was reported that the customer had general dissatisfaction over the product. There was no patient impact. Additional information received as bearing detached found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50864364 ↗MalfunctionSep 16, 2026No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50787971 ↗MalfunctionSep 16, 2026No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50784393 ↗MalfunctionSep 15, 2026Detachment of Device or Device Component
It was reported that pre-op, ring is missing. At the time of report there was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50768385 ↗MalfunctionSep 11, 2026No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50740787 ↗MalfunctionSep 11, 2026Overheating of Device; Loss of Power
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50740525 ↗MalfunctionSep 11, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. At the time of report there was no patient impact. Additional information was received stating that detached bearings were found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50732690 ↗MalfunctionSep 11, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request and no alleged product deficiencies were reported. At the time of the report, there was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50729768 ↗MalfunctionSep 10, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request with no alleged product deficiency. At the time of the report, there was no known impact to the patient. Additional information was received stating that the it was loaner check up, there was no impact to patient and detached distal bearing was found during evaluation made it reportable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713747 ↗MalfunctionSep 1, 2026Difficult to Insert
It was reported that during intra-operative use of a drill, the drill could not be clicked into place and was described as difficult to insert. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50537073 ↗MalfunctionAug 28, 2026Use of Device Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50490803 ↗MalfunctionAug 27, 2026No Apparent Adverse Event
It was reported that the customer had general dissatisfaction over the product. There was no impact to the patient at the time of report. Additional information received as bearing detached found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50466523 ↗MalfunctionAug 26, 2026Vibration
It was reported that the device's drill bit was wobbling. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50460740 ↗MalfunctionAug 26, 2026No Apparent Adverse Event
It was reported that device was not working. At the time of report there was no patient impact. Additional information was received stating that detached bearings were found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50443935 ↗MalfunctionAug 20, 2026Difficult to Remove; Naturally Worn
It was reported that the tool was stuck inside and the bearings failed. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50341939 ↗MalfunctionAug 17, 2026Loss of Power
It was reported that prior to use, the attachment was frozen. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50277837 ↗MalfunctionJul 6, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST WITH NO ALLEGED PRODUCT DEFICIENCY. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING BEARING THAT WAS DETACHED WAS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25716148 ↗MalfunctionJul 6, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THERE WAS NO ALLEGED ISSUE AT THE TIME OF INTAKE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25714925 ↗MalfunctionJul 6, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25714778 ↗MalfunctionJul 6, 2026Difficult to Insert; Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT OVERHEAT AND DOES NOT HOLD THE TOOL. THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25714716 ↗MalfunctionJul 6, 2026Difficult to Insert
IT WAS REPORTED THAT THE ATTACHMENT WAS DEFECTIVE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25714079 ↗MalfunctionJul 5, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THERE WERE NO ALLEGED ISSUE WITH THE DEVICE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25712199 ↗MalfunctionJul 4, 2026Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT BEARINGS WERE BAD. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25709588 ↗MalfunctionJul 4, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST WITH NO ALLEGED PRODUCT DEFICIENCY. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING BEARING THAT WAS DETACHED WAS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25709460 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page