Skip to content

MIDAS REX MR8 MR8-AVA15 (UDI 00763000441326)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8-AVA15 MR8 15CM VAR ANG 2.4MM · Class 2

Quick summary

This Midas Rex variable-angle drill attachment has 121 adverse event reports, with no deaths or injuries. Some reports describe the attachment coming apart or falling off, with detached bearings found on evaluation, and some describe noise or overheating. Reports rose from 6 in the previous year to 101 in the latest year. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
130
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 130 adverse event reports that list this exact device (DI 00763000441326), received between Dec 14, 2023 and Sep 29, 2026 — classified as 130 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 5 reportsSep 2024: 5 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 16 reportsApr 2026: 21 reportsMay 2026: 11 reportsJun 2026: 23 reportsJul 2026: 21 reportsAug 2026: 1 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event84
  • Detachment of Device or Device Component18
  • Overheating of Device13
  • Naturally Worn7
  • Vibration5
  • Difficult to Insert4
  • Loss of Power4
  • Difficult to Remove2

Patient problems

  • No Clinical Signs, Symptoms or Conditions116
  • Insufficient Information15

Recent adverse event reports

The 25 most recent of 130 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934077 ↗
  • MalfunctionSep 28, 2026Loss of Power

    It was reported that the device does not work. There was no patient impact. Additional information was received on evaluation stated the distal bearing was detached made this event reportable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912183 ↗
  • MalfunctionSep 27, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50910087 ↗
  • MalfunctionSep 23, 2026Overheating of Device

    It was reported that device gets overheated. At the time of report no patient was involved. Additional information received that bearings were worn and detached during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50864381 ↗
  • MalfunctionSep 16, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50784465 ↗
  • MalfunctionSep 15, 2026Naturally Worn

    It was reported that bearings of the device were bad. At the time of report patient involvement or impact was unknown. Additional information received stating that the bearings were misaligned.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771819 ↗
  • MalfunctionSep 15, 2026Overheating of Device

    It was reported that the attachment was overheating. There was no patient involved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769065 ↗
  • MalfunctionSep 13, 2026Insufficient Device Problem Information

    Patient problems: Insufficient Information

    MAUDE report 50747834 ↗
  • MalfunctionSep 10, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50708352 ↗
  • MalfunctionSep 9, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. At the time of report no patient impact was observed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50684035 ↗
  • MalfunctionSep 3, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50582471 ↗
  • MalfunctionAug 28, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50492512 ↗
  • MalfunctionJul 6, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25714095 ↗
  • MalfunctionJul 6, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: Insufficient Information

    MAUDE report 25713779 ↗
  • MalfunctionJul 5, 2026Noise, Audible

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF NOISE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25712212 ↗
  • MalfunctionJul 4, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE DRILL FALLS OFF, THE HOLE SEEMS TO BE WIDENED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 25710845 ↗
  • MalfunctionJul 3, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703937 ↗
  • MalfunctionJul 3, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703430 ↗
  • MalfunctionJul 3, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25702762 ↗
  • MalfunctionJul 3, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST WITH NO ALLEGED PRODUCT DEFICIENCY. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING A BEARING THAT WAS DETACHED WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25702460 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT HAD ISSUE/FAILURE THAT WAS NOT SPECIFIC, BUT THE CUSTOMER EXPRESSED DISSATISFACTION WITH THE PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25700497 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT HAS GENERAL DISSATISFACTION. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689724 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689187 ↗
  • MalfunctionJul 2, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT GETS HOT. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25688853 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25688781 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.