Skip to content

MIDAS REX MR8 MR8-AVS10 (UDI 00763000441371)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8-AVS10 MR8 10CM VAR STR 2.4MM · Class 2

All models of MIDAS REX MR8 (137)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00763000441371), received between Aug 12, 2024 and Sep 29, 2026 — classified as 20 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 3 reportsJul 2026: 5 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event8
  • Detachment of Device or Device Component5
  • Overheating of Device4
  • Loss of Power2
  • Naturally Worn2
  • Physical Resistance/Sticking1
  • Vibration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions17
  • Insufficient Information3

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Detachment of Device or Device Component; Naturally Worn

    It was reported that the attachment with the report of bearings have fallen out, and the upper diameter of the attachment has become enlarged. It was reported that there was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50929827 ↗
  • MalfunctionSep 28, 2026No Apparent Adverse Event

    It was reported that the device was sent for preventive maintenance with no alleged issue. At the time of report, there was no patient impact. Additional information received upon evaluation stating that the bearing was misaligned.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917010 ↗
  • MalfunctionSep 15, 2026Loss of Power

    It was reported that during testing, the device's variable dial was not working/ device seized. There was no patient involvement or impact during testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50776714 ↗
  • MalfunctionSep 11, 2026Naturally Worn

    It was reported that the upper part of the attachment is widened and split, probably due to excessive drill bit movement. There is no patient impact reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50726271 ↗
  • MalfunctionJul 3, 2026Loss of Power

    IT WAS REPORTED THAT THE DEVICE SEIZED. NO PATIENT IMPACT WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25700860 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION

    Patient problems: Insufficient Information

    MAUDE report 25696343 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689529 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT PRIOR TO USE THE ATTACHMENT WAS NOT WORKING. AT THE TIME OF REPORT, THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689156 ↗
  • MalfunctionJul 1, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT HAD DISSATISFACTION. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION STATED THAT BEARING DETACHED FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25677874 ↗
  • MalfunctionJun 30, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING USE,THE ATTACHMENT INNER RING CAME OUT, IT WAS REPORTED THAT THERE WAS PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25672487 ↗
  • MalfunctionJun 27, 2026Overheating of Device

    IT WAS REPORTED THAT INTRA-OPERATIVELY, THE DEVICE WAS ALLEGED TO HAVE OVERHEATED. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25645005 ↗
  • MalfunctionJun 4, 2026Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WITH OVERHEATING AND BAD BEARING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389376 ↗
  • MalfunctionMay 12, 2026Detachment of Device or Device Component

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, IMPACT TO THE PATIENT WAS UNKNOWN. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DURING CLEANING IT WAS NOTED THAT THE BEARINGS FELL OUT ,THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25148299 ↗
  • MalfunctionMay 7, 2026Vibration

    IT WAS REPORTED THAT NON-STANDARD MOVEMENT IN THE DRILL PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25110052 ↗
  • MalfunctionApr 8, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT THE ATTACHMENT WITH DRILL COULD NOT BE SEPARATED FROM THE MOTOR. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED THAT THE BEARINGS WERE DETACHED DURING THE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24814518 ↗
  • MalfunctionApr 7, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF NO REPORTED ISSUE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24797597 ↗
  • MalfunctionMar 18, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. NO KNOWN PATIENT IMPACT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24629037 ↗
  • MalfunctionAug 7, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT IS BEING FOLLOWED UP FROM THE FACILITY.

    Patient problems: Insufficient Information

    MAUDE report 22735432 ↗
  • MalfunctionAug 25, 2024Detachment of Device or Device Component; No Apparent Adverse Event; Detachment of Device or Device Component; No Apparent Adverse Event

    REPAIR REQUEST INITIATED FOR DEVICE WITH NO REPORTED MALFUNCTION. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT ON EVALUATION DUE TO DETACHED BEARINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20060824 ↗
  • MalfunctionAug 12, 2024Detachment of Device or Device Component; No Apparent Adverse Event; Detachment of Device or Device Component; No Apparent Adverse Event

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF PROBLEM NOT REPORTED. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO DETACHED BEARING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19959073 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page