MIDAS REX MR8 MR8-AVS12 (UDI 00763000441388)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8-AVS12 MR8 12CM VAR STR 2.4MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000441388), received between Jul 9, 2025 and Sep 28, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Naturally Worn2
- Physical Resistance/Sticking2
- Difficult to Insert1
- Difficult to Remove1
- No Apparent Adverse Event1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Detachment of Device or Device Component; Physical Resistance/Sticking
It was reported that the attachment was locked/sleeve-sticking. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50912005 ↗MalfunctionSep 27, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910425 ↗MalfunctionSep 25, 2026Difficult to Remove
It was reported that the attachment was blocked. At the time of report, there was no patient or staff impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50893951 ↗MalfunctionSep 24, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. There was no patient i mpact as a result of the event. Additional information was received stating that bearing fallout at tube distal was found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50886192 ↗MalfunctionSep 24, 2026Overheating of Device
It was reported that during use, the attachment was overheating. It was reported that there was no patient impact. Additional information was received stating that detached bearings were found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885872 ↗MalfunctionJun 30, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE PROCEDURE THE BALL BEARING FAILURE AND FALLING OUT OF ATTACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT REPORTED. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25662593 ↗MalfunctionMay 21, 2026Naturally Worn
IT WAS REPORTED THAT, WHILE PREPARING FOR USE, THE BEARINGS WERE FOUND TO BE DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25238990 ↗MalfunctionApr 28, 2026Difficult to Insert; Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT HAS LOOSE BALL BEARINGS AND UNABLE TO SEAT THE BURR. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED ON EVALUATION CONFIRMED THE LOOSE BALL BEARING MADE THIS EVENT REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25010848 ↗MalfunctionJul 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF BEARING DETACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22564114 ↗MalfunctionJul 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF BEARING DETACHMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22454784 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page