Skip to content

MIDS REX MR8 MR8-10BA20DRecalls & reported problems

Medtronic Powered Surgical Solutions · TOOL MR8-10BA20D MR8 10CM BALL D 2MM · Class 2

All models of MIDS REX MR8 (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000487072), received between Oct 22, 2025 and Jan 21, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2026Fracture

    IT WAS REPORTED THAT DURING ENDOSCOPIC TRANSSPHENOIDAL RESECTION OF A PITUITARY TUMOR, THE BURR HEAD FRACTURED, RENDERING THE DEVICE UNUSABLE. IT WAS ALSO REPORTED THAT THERE IS A 5 MINUTES OF SURGICAL DELAY AND THE PROCEDURE COMPLETED WITH BACKUP PRODUCTS. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24127623 ↗
  • MalfunctionNov 18, 2025Fracture; No Apparent Adverse Event; Fracture; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING A POSTERIOR CERVICAL SPINE DECOMPRESSION AND FIXATION PROCEDURE, A MATCHHEAD TOOL FRACTURED WHILE CREATING A PILOT HOLE FOR A CERVICAL PEDICLE SCREW USING THE MOTOR. THE PATIENT¿S BONE WAS REPORTED TO BE SCLEROTIC, AND INCREASED FORCE WAS APPLIED AS THE SHAFT WAS THICKER THAN THE TIP BALL DIAMETER (2 MM). THE PROCEDURE WAS COMPLETED USING A SPARE PRODUCT. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23578604 ↗
  • MalfunctionOct 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A SPINAL SURGERY, THE TOOL FRACTURED WHILE IN USE. THE PROCEDURE WAS COMPLETED USING A BACKUP PRODUCT WITHOUT ANY REPORTED DELAY. THERE WAS NO REPORTED PATIENT INJURY OR IMPACT ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356342 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices