NA BB10B05R15Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB10B05R15 P N T 2-6 · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000513214), received between Jul 22, 2024 and Jul 22, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Obstruction of Flow” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1027-2023 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2024Obstruction of Flow; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT UPON PRIMING, THE CUSTOMER DISCOVERED THAT THE FLOW STOPPED WHEN THEY CLAMPED THE BRIDGE AND FORCED FLOW THROUGH THE PRE-BYPASS FILTER. THE OCCLUSION OCCURRED IN THE PRE BYPASS FILTER. THE FLOW ISSUE WAS NOT RELATED TO BLOOD ANTICOAGULATION AS IT OCCURRED IN PRE-BYPASS BEFORE THE BLOOD WAS INTRODUCED. THE DEVICE WAS REPLACED. THE CUSTOMER STATED THAT SUFFICIENT FLOW WAS ACHIEVED WITH THE REPLACEMENT DEVICE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19791489 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.
- Z-1895-2021 · Apr 29, 2021 · TerminatedMedtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.
- Z-1890-2021 · Apr 29, 2021 · TerminatedMedtronic HY10L28R3, Custom Pack HY10L28R3 1/4 Pack, Custom Perfusion System, Sterile, Rx.
- Z-1884-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W65R1, Custom Pack CB10W65R1 CH Access1/4, Custom Perfusion System, Sterile, Rx.
- Z-1878-2021 · Apr 29, 2021 · TerminatedMedtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
- Z-1885-2021 · Apr 29, 2021 · TerminatedMedtronic HY10Y54R4, Custom Pack HY10Y54R4 ECC PED 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1880-2021 · Apr 29, 2021 · TerminatedMedtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
- Z-1888-2021 · Apr 29, 2021 · TerminatedMedtronic CB10R55R5, Custom Pack CB10R55R5 3/8X3/8 Loop, Custom Perfusion System, Sterile, Rx.
- Z-1873-2021 · Apr 29, 2021 · TerminatedMedtronic BB10R94R6, Custom Pack BB10R94R6 1/4 ECCPK, Custom Perfusion System, Sterile, Rx.
- Z-1889-2021 · Apr 29, 2021 · TerminatedMedtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
- Z-1883-2021 · Apr 29, 2021 · TerminatedMedtronic HY10J00R12, Custom Pack HY10J00R12 ECC Small CA, Custom Perfusion System,
- Z-1876-2021 · Apr 29, 2021 · TerminatedMedtronic CB7J97R8, Custom Pack CB7J97R8 U14 E Sup, Custom Perfusion System, Sterile, Rx.
- Z-1891-2021 · Apr 29, 2021 · TerminatedMedtronic CB10Y89R10, Custom Pack CB10Y89R10 ECC 3/8 A/V, Custom Perfusion System, Sterile, Rx.
- Z-1882-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W64R1, Custom Pack CB10W64R1 CH Access 3/8, Custom Perfusion System, Sterile, Rx.
- Z-1886-2021 · Apr 29, 2021 · TerminatedMedtronic HY10D90R2, Custom Pack HY10D90R2 3/8 ACC, Custom Perfusion System, Sterile, Rx.
- Z-1892-2021 · Apr 29, 2021 · TerminatedMedtronic BB8M63R5, Custom Pack BB8M63R5 Adult ECC, Custom Perfusion system, Sterile, Rx.
…and 10 more.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page