Skip to content

MIDAS REX MR8 MR8-F2/7TA23 (UDI 00763000513412)Recalls & reported problems

Medtronic Powered Surgical Solutions · TOOL MR8-F2/7TA23 MR8 F2/7CM TAPER 2.3MM · Class 2

All models of MIDAS REX MR8 (137)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00763000513412), received between Dec 3, 2024 and Aug 27, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Difficult to Remove1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2026Difficult to Remove

    It was reported that the blade was stuck in the nose of the attachment. At the time of report, there was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50470576 ↗
  • MalfunctionMay 29, 2026Fracture

    IT WAS REPORTED THAT DURING CRANIOTOMY THE PART OF THE DRILL BIT THAT ATTACHES TO THE DRILL MOTOR'S COLLAR BROKE OFF AND WAS LOST. THIS CAUSED DELAY OF 20 TO 30 MINUTES IN THE PROCEDURE, SURGERY WAS COMPLETED WITH BACKUP PRODUCT. AT THE TIME OF THE REPORT, IMPACT TO THE PATIENT WAS UNKNOWN. ADDITIONAL INFORMATION RECEIVED STATING THAT THE THERE WAS NO IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25329329 ↗
  • MalfunctionDec 2, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A CRANIOTOMY PROCEDURE WITH A TAPERED TOOL AND ATTACHMENT, THE TOOL TIP BROKE OFF. ACCORDING TO SURGEON THERE IS NO SUSPICION OF TOOL MALFUNCTION. THE SURGEON ALSO STATED THAT TOOL TIP BROKE OFF AS A RESULT OF EXCESS POWER AND MANIPULATION DUE TO THE PATIENT´S BONE THICKNESS. NO PATIENT IMPACT REPORTED. THE PROCEDURE WAS COMPLETED WITH THE BACKUP PRODUCT. ON ADDITIONAL FOLLOW-UP, IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23693568 ↗
  • MalfunctionMay 29, 2025Fracture; Fracture

    REPAIR REQUEST INITIATED FOR DEVICE WITH NO REPORTED MALFUNCTION. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO DETACHED BEARING AND BROKEN TOOL.

    Patient problems: Insufficient Information

    MAUDE report 22103949 ↗
  • MalfunctionApr 30, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING CRANIOTOMY SURGERY WHEN USING THE TOOL THE HEAD IS BROKEN, THE TOOL TIP THAT WAS LEFT ON THE BONE BOARD IS REMOVED. NO PATIENT IMPACT REPORTED. IT WAS REPORTED THAT THERE WERE NO FRAGMENTS LEFT INSIDE THE PATIENT AND THERE WAS NO DELAY IN PROCEDURE AS  A RESULT OF THE EVENT. THE PROCEDURE WAS COMPLETED WITH THE BACKUP PRODUCT. NO ADDITIONAL INTERVENTION WAS PERFORMED OR PLANNED AS A RESULT OF THIS EVENT. ON FOLLOW UP, IT WAS CONFIRMED WITH THE HEALTH CARE PROFESSIONAL THAT THERE WAS NO PATIENT OR STAFF IMPACT  .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21933486 ↗
  • MalfunctionApr 8, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING THE CRANIOTOMY PROCEDURE, THE DRILL BIT OVERHEATED AT STOLLERY. NO PATIENT IMPACT REPORTED. ON FOLLOW-UP IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT, THERE WAS A SLIGHT DELAY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DRILL.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21794312 ↗
  • MalfunctionDec 3, 2024Fracture; Fracture

    ON FOLLOW UP IT WAS CONFIRMED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20834152 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page