Bio-Medicus Life Support™ LS96218-017Recalls & reported problems

Medtronic, Inc. · CATHETER LS96218-017 BIOMED ART JUG 17FR · Class 2

All models of Bio-Medicus Life Support (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000615420), received between Mar 30, 2023 and Sep 9, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Crack” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack7
  • Air/Gas in Device1
  • Fluid/Blood Leak1
  • Material Puncture/Hole1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2026Mechanical Problem

    Medtronic received information that during use of a Bio-Medicus Life Support Catheter, it was reported that there was difficulty separating the needle from the body of the cannula. No noticeable damage was observed on the cannula. The device was replaced to complete the case. There was no adverse patient effect associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50688924 ↗
  • MalfunctionOct 15, 2024Crack; Crack

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, IT WAS REPORTED THAT THERE WAS A CRACK AT THE SIDE HOLE OF THE CANNULA. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20449687 ↗
  • MalfunctionDec 6, 2023Crack; Crack

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, THE CUSTOMER REPORTED THAT THEY IDENTIFIED SMALL CRACKS IN THE COUPLING PIECE. THE CUSTOMER NOTICED THAT THE CRACKS APPEARED IMMEDIATELY WHEN INSERTING THE DEVICE. THEY APPEARED WHEN THE CUSTOMER PRESSED THE INTRODUCER INTO THE DEVICE. THERE WAS A PROBLEM WITH THE CANNULATION ACCORDING TO ROUTINE, HOWEVER, THE CUSTOMER NOTICED THAT THE CRACKS APPEARED IN CONNECTION WITH THE DILATOR BEING PUSHED IN. THE DEVICE WAS NOT EXPOSED TO HEAT OR COLD BEFORE INSERTION. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CRACKED PORTION IS AT THE TOP OF THE CONNECTING PART OF THE DEVICE. THE CANNULA WAS USED FOR AN EXTRA CORPOREAL MEMBRANE OXYGENATION (ECMO) PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18273342 ↗
  • MalfunctionNov 30, 2023Crack; Crack

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, THE CUSTOMER REPORTED THAT THERE WAS A SMALL CRACK IDENTIFIED IN THE COUPLING PIECE AND THEY SUSPECTED THERE WAS A CRACK AT THE TOP OF THE CONNECTOR. THE CUSTOMER STATED THERE WAS NO LEAK AND THE PATIENT WAS NOT HARMED. THE CRACK WAS DISCOVERED A DAY AFTER ADMISSION AND APPEARED LIKE A SMALL WHITE LINE. THE DURATION OF THE TREATMENT WITH THE DEFECTIVE DEVICE WAS THREE DAYS. THE CUSTOMER STATED THERE WAS NO PROBLEM INSERTING THIS CANNULA. THERE WAS NO HARM OR BLOOD LOSS RELATED TO THE CANNULA. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE AS THERE WAS NO LEAK OBSERVED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT, HOWEVER THE PATIENT IS NOW DECEASED AND THE CUSTOMER STATED THAT IT WAS NOT IN ANY WAY RELATED TO THE CANNULA. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE. THE DAMAGE WAS NOT IN THE LOCATION OF THE SUTURES. THE CANNULA WAS BEING USED FOR AN EXTRA CORPOREAL MEMBRANE OXYGENATION (ECMO) PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18236062 ↗
  • MalfunctionJun 12, 2023Crack; Crack

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, IT WAS REPORTED THE DEVICE HAD A DAMAGED CONNECTOR. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE WAS SEEN TO THE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17105472 ↗
  • MalfunctionMay 9, 2023Crack; Material Puncture/Hole; Crack; Material Puncture/Hole

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, THE CUSTOMER REPORTED THAT SMALL CRACKS WERE OBSERVED IN THE COUPLING PIECE. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WASN'T ANY LEAK FROM THE CANNULA, OR FROM THE CONNECTION BETWEEN A CANNULA AND A CONNECTOR. THERE WAS NO PATIENT BLOOD LOSS. NO TRANSFUSION WAS REQUIRED. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE. THERE WAS NO DAMAGE IN THE LOCATION OF THE SUTURES. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THIS DEVICE WAS BEING USED FOR AN ECMO PROCEDURE. THERE WAS NO INJURY RELATED TO THE CANNULA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16895119 ↗
  • MalfunctionApr 25, 2023Crack; Crack

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, THE CUSTOMER REPORTED THAT UPON OPENING THE PACKAGE A CRACK WAS OBSERVED IN THE COUPLING PIECE, THERE WAS NO LEAK. THE DEVICE WAS USED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE. THERE WERE NO SUTURES NEAR THE DAMAGED PART. JUST A SAFETY STRAP. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS USED FOR ECMO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16805643 ↗
  • MalfunctionMar 30, 2023Crack; Fluid/Blood Leak; Air/Gas in Device; Crack; Fluid/Blood Leak; Air/Gas in Device

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, IT WAS REPORTED THAT THE INSERTION OF THE CANNULA WAS EASY BUT THE LUER LOCK DID NOT LOOK 100% SAFE. IT WAS REPORTED THAT THERE WAS A POSSIBILITY OF SUCKING AIR INTO THE CIRCUIT. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE. THERE WAS NO DAMAGE IN THE LOCATION OF THE SUTURES. THE CUSTOMER STATED THAT AIR WAS OBSERVED IN THE CIRCUIT. IT WAS STATED THAT INFORMATION ON WHETHER THE CANNULA WAS USED FOR EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) OR CARDIOPULMONARY BYPASS (CPB) WAS UNKNOWN. IT WAS STATED THAT THE CONNECTION WAS NOT LEAKING, IT WAS THE LUERLOCK ON THE CANNULAE THAT SUCKED AIR. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS PLACED IN THE V. J. RIGHT WITHOUT ANY PROBLEMS. THE LUER LOCK WAS NOT TIGHT (LEAKY) AND AIR WAS DRAWN INTO THE CANNULA. IT WAS REPORTED THAT POTENTIALLY, THIS COULD HAVE LEAD TO AIR EMBOLISM AND BLEEDING (AS A RESULT OF EXCHANGE). IT WAS REPORTED THAT THE LEAK WAS LIKE THE CAP THAT CLOSED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16644573 ↗

Other Single lumen ECMO cannula devices

All Single lumen ECMO cannula devices →