Medtronic, Inc. · CATHETER LS96360-025 BIO-MED VEN 25FR · Class 2

All models of Bio-Medicus Life Support (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000615581), received between Jun 2, 2023 and Aug 31, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem4
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Mechanical Problem

    PLI 10 and PLI 20: Medtronic received information that during use of a Bio-Medicus life support cannula, it got stuck in the valve case and customer experienced insertion/removal issues. The device was replaced to complete the procedure. Patient outcome and status were unknown.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533640 ↗
  • MalfunctionAug 31, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533635 ↗
  • MalfunctionJul 22, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT DURING THE INSERTION OF THE BIO-MEDICUS LIFE SUPPORT CATHETER AND INTRODUCER, AFTER THE PRODUCT WAS PLACED IN THE INTENDED POSITION, THE INTRODUCER COULD NOT BE WITHDRAWN, WHICH WAS UNUSUAL FOR THIS PRODUCT AND NOT PREVIOUSLY EXPERIENCED BY THE HOSPITAL. THE SURGICAL PROCEDURE WAS PERFORMED FOLLOWING THE USUAL PROTOCOL. NO PRE-USE TESTING WAS PERFORMED DUE TO THE EMERGENCY OF THE CASE. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THE ALTERNATIVE MODEL FUNCTIONED NORMALLY WITHOUT ANY COMPLICATIONS. IT WAS UNKNOWN IF THERE WAS ANY COMPLICATIONS AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22563742 ↗
  • MalfunctionNov 22, 2024Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, IT WAS REPORTED THAT UPON CANNULATING THE PATIENT VIA THE LEFT GROIN USING A CUT DOWN TECHNIQUE THE CLINICIAN COULD NOT REMOVE THE INTRODUCER FROM THE CATHETER BODY ( LS96360-025 MULTISTAGE DRAINAGE CATHETER). THE CLINICIAN COULD ONLY RETRACT THE INTRODUCER BY APPROXIMATELY 10CM. THE DECISION WAS MADE TO REPLACE THE CANNULA WITH ANOTHER LS96360-025 AND THE EXACT SAME ISSUE RESULTED. THE CLINICIAN THEN DECIDED TO CANNULATE USING A LIVANOVA RAP CANNULA AND THIS WAS SUCCESSFUL. THE PATIENTS VESSEL WAS DILATED USING THE LIVANOVA ACCESS KIT TO 24FR DILATION PRIOR TO CANNULATION. THE CANNULA BODY WAS WETTED OUT USING SALINE PRIOR TO INTRODUCING THE WHITE INTRODUCER (CUSTOMER PROTOCOL). THE CUSTOMER ASKED IF THE ELASTEON MATERIAL OF THE BIOMEDICUS LIFE SUPPORT CATHETER HAS ANY IMPACT ON FRICTION BETWEEN THE CATHETER BODY AND THE WHITE INTRODUCER MATERIAL. (PLEASE DETAIL THE INTRODUCER MATERIAL IN THE REPORT). THE CUSTOMER ASKED FOR A DETAILED CHECK ON ANY MANUFACTURING VARIANCE ON THE CANNULA BODY AND THE WHITE INTRODUCER. ARE THEY WITHIN SPECIFICATION AND WHAT IS THAT SPECIFICATION (WAS ONE AT THE TOP E

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20759175 ↗
  • MalfunctionJun 2, 2023Material Deformation; Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, THE CUSTOMER REPORTED THAT THE CONNECTOR WAS BROKEN. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17048781 ↗