Bio-Medicus Life Support™ LS96555-023Recalls & reported problems

Medtronic, Inc. · CATHETER LS96555-023 BIO-MED VEN 23FR · Class 2

All models of Bio-Medicus Life Support (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000615512), received between Jun 28, 2024 and Apr 25, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Crack” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack3
  • Air/Gas in Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2025Crack; Crack

    IT WAS REPORTED THAT DURING THE USE OF THE BIO-MEDICUS LIFE SUPPORT CANNULA, THE CUSTOMER OBSERVED CRAZING ON THE CONNECTION END OF THE CANNULA. THE CATHETER WAS STILL USED FOR VENO-VENOUS EXTRACORPOREAL MEMBRANE OXYGENATION (VV ECMO). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21906918 ↗
  • MalfunctionJan 16, 2025Crack; Crack

    IT WAS REPORTED THAT DURING AN EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) RETRIEVAL PROCEDURE, AFTER CANNULATING THE RIGHT FEMORAL VEIN, THE BIO-MEDICUS LIFE SUPPORT CANNULA'S CONNECTOR WAS FOUND TO BE CRACKED AND DEEMED UNSAFE TO USE. THE CLINICAL TEAM NOTED THAT THE CANNULA "DIDN'T CLICK LIKE THE OTHER ONE." IT WAS ALSO REPORTED THAT THE HAEMOSTASIS CAP WAS NOT USED DURING THE ASSEMBLY OF THE CANNULA. THE INVESTIGATION INTO POTENTIAL DAMAGE BY THE CLINICAL TEAM DURING INSERTION WAS INCONCLUSIVE, AND IT WAS SUGGESTED THAT THE TRANSLUCENT HAEMOSTASIS CAP MIGHT HAVE BEEN DISCARDED IN ERROR PRIOR TO ASSEMBLING THE INTRODUCER INTO THE CANNULA. THE CUSTOMER EXPRESSED CONCERN THAT THE TRANSLUCENT HAEMOSTASIS CAP'S COLOR MATCHED THE PACKAGING, POTENTIALLY LEADING TO OVERSIGHT. A NEW CANNULA WAS PLACED INTO THE PATIENT WITHOUT HARM. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21166158 ↗
  • MalfunctionJun 28, 2024Crack; Air/Gas in Device; Crack; Air/Gas in Device

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, IT WAS REPORTED THAT THERE WAS A CRACK IN THE CONNECTOR. THE DAMAGE WAS NOT IN THE LOCATION OF THE SUTURES. THE HEMOSTASIS CAP WAS USED WHEN THE INTRODUCER WAS INSERTED INTO THECANNULA BODY CONNECTOR. THE CUSTOMER INSERTED THE DEVICE BUT THEY CONTINUED TO HAVE AIR WHEN THEY WERE ATTEMPTING TO MAKE A WET CONNECTION. THEY DISCOVERED THE CRACK BEFORE GOING ON PUMP. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19631338 ↗