Bio-Medicus Life Support™ LS96555-023Recalls & reported problems
Medtronic, Inc. · CATHETER LS96555-023 BIO-MED VEN 23FR · Class 2
All models of Bio-Medicus Life Support (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000615512), received between Jun 28, 2024 and Apr 25, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Crack” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack3
- Air/Gas in Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2025Crack; Crack
IT WAS REPORTED THAT DURING THE USE OF THE BIO-MEDICUS LIFE SUPPORT CANNULA, THE CUSTOMER OBSERVED CRAZING ON THE CONNECTION END OF THE CANNULA. THE CATHETER WAS STILL USED FOR VENO-VENOUS EXTRACORPOREAL MEMBRANE OXYGENATION (VV ECMO). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21906918 ↗MalfunctionJan 16, 2025Crack; Crack
IT WAS REPORTED THAT DURING AN EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) RETRIEVAL PROCEDURE, AFTER CANNULATING THE RIGHT FEMORAL VEIN, THE BIO-MEDICUS LIFE SUPPORT CANNULA'S CONNECTOR WAS FOUND TO BE CRACKED AND DEEMED UNSAFE TO USE. THE CLINICAL TEAM NOTED THAT THE CANNULA "DIDN'T CLICK LIKE THE OTHER ONE." IT WAS ALSO REPORTED THAT THE HAEMOSTASIS CAP WAS NOT USED DURING THE ASSEMBLY OF THE CANNULA. THE INVESTIGATION INTO POTENTIAL DAMAGE BY THE CLINICAL TEAM DURING INSERTION WAS INCONCLUSIVE, AND IT WAS SUGGESTED THAT THE TRANSLUCENT HAEMOSTASIS CAP MIGHT HAVE BEEN DISCARDED IN ERROR PRIOR TO ASSEMBLING THE INTRODUCER INTO THE CANNULA. THE CUSTOMER EXPRESSED CONCERN THAT THE TRANSLUCENT HAEMOSTASIS CAP'S COLOR MATCHED THE PACKAGING, POTENTIALLY LEADING TO OVERSIGHT. A NEW CANNULA WAS PLACED INTO THE PATIENT WITHOUT HARM. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 21166158 ↗MalfunctionJun 28, 2024Crack; Air/Gas in Device; Crack; Air/Gas in Device
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS LIFE SUPPORT CANNULA, IT WAS REPORTED THAT THERE WAS A CRACK IN THE CONNECTOR. THE DAMAGE WAS NOT IN THE LOCATION OF THE SUTURES. THE HEMOSTASIS CAP WAS USED WHEN THE INTRODUCER WAS INSERTED INTO THECANNULA BODY CONNECTOR. THE CUSTOMER INSERTED THE DEVICE BUT THEY CONTINUED TO HAVE AIR WHEN THEY WERE ATTEMPTING TO MAKE A WET CONNECTION. THEY DISCOVERED THE CRACK BEFORE GOING ON PUMP. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19631338 ↗
Other Single lumen ECMO cannula devices
- Bio-Medicus Life Support™ LS96218-017Medtronic, Inc. · Single lumen ECMO cannula8 reports
- Bio-Medicus Life Support™ LS96555-019Medtronic, Inc. · Single lumen ECMO cannula7 reports
- Bio-Medicus Life Support™ LS96360-025Medtronic, Inc. · Single lumen ECMO cannula5 reports
- Bio-Medicus Life Support™ LS96218-019Medtronic, Inc. · Single lumen ECMO cannula5 reports
- Bio-Medicus Life Support™ LS96218-015Medtronic, Inc. · Single lumen ECMO cannula4 reports
- Bio-Medicus Life Support™ LS96555-021Medtronic, Inc. · Single lumen ECMO cannula4 reports
- Bio-Medicus Life Support™ LS96360-023Medtronic, Inc. · Single lumen ECMO cannula3 reports
- Bio-Medicus Life Support™ LS96218-021Medtronic, Inc. · Single lumen ECMO cannula3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page