NA BB11U86R3Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB11U86R3 AGH CPB · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000726317), received between Nov 30, 2022 and Jan 10, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1027-2023 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Separation1
- Mechanical Problem1
- Misassembled During Installation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 10, 2023Fluid/Blood Leak; Mechanical Problem; Fluid/Blood Leak; Mechanical Problem
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THE VACUUM RELIEF VALVE (VRV) WAS STUCK OPEN, WHICH RESULTED IN 15 ML OF BLOOD LOSS DURING PRIMING. THE LEAK OCCURRED ON PAGE 11 LINE 10 OF THE CUSTOM TUBING PACK SPECIFICATION. USE OF DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT NO INFORMATION WAS OBTAINED ON EXACTLY WHERE ON THE CIRCUIT THE VALVE WAS STUCK OPEN. THERE WAS NO TRANSFUSION NEEDED AS A RESULT OF THIS LEAK BECAUSE IT WAS DISCOVERED PRIOR TO PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16131284 ↗MalfunctionNov 30, 2022Material Separation; Misassembled During Installation; Material Separation; Misassembled During Installation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE DURING SET-UP OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THE DRIP CHAMBER ON LINE 4 PAGE 6 OF THE SPECIFICATION WAS EASILY DISCONNECTED FROM THE TUBING WITH LITTLE PULLING PRESSURE. IT WAS CLEAR THAT THE BONDING WAS NOT SUFFICIENT TO HOLD THE TUBING. THE CUSTOMER USED THE OTHER DRIP CHAMBER FOR THE DURATION OF THE CASE. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECT OCCURRED. IT WAS STATED THAT NOTHING WAS DAMAGED IN OR OUT OF THE PACK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15885765 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.
- Z-1895-2021 · Apr 29, 2021 · TerminatedMedtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.
- Z-1875-2021 · Apr 29, 2021 · TerminatedMedtronic HY7M07R6, Custom Pack HY7M07R6 E 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1878-2021 · Apr 29, 2021 · TerminatedMedtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
- Z-1882-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W64R1, Custom Pack CB10W64R1 CH Access 3/8, Custom Perfusion System, Sterile, Rx.
- Z-1893-2021 · Apr 29, 2021 · TerminatedMedtronic BB7D76R13, Custom Pack BB7D76R13 CORR HL, Custom Perfusion System, Sterile, Rx.
- Z-1888-2021 · Apr 29, 2021 · TerminatedMedtronic CB10R55R5, Custom Pack CB10R55R5 3/8X3/8 Loop, Custom Perfusion System, Sterile, Rx.
- Z-1884-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W65R1, Custom Pack CB10W65R1 CH Access1/4, Custom Perfusion System, Sterile, Rx.
- Z-1874-2021 · Apr 29, 2021 · TerminatedMedtronic BB11J14R2, Custom Pack BB11J14R2 Bridge, Custom Perfusion System, Sterile, Rx.
- Z-1889-2021 · Apr 29, 2021 · TerminatedMedtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
- Z-1872-2021 · Apr 29, 2021 · TerminatedMedtronic HY8M63R4, Custom Pack HY8M63R4 Adult ECC, Custom Perfusion System, Sterile, Rx.
- Z-1876-2021 · Apr 29, 2021 · TerminatedMedtronic CB7J97R8, Custom Pack CB7J97R8 U14 E Sup, Custom Perfusion System, Sterile, Rx.
- Z-1892-2021 · Apr 29, 2021 · TerminatedMedtronic BB8M63R5, Custom Pack BB8M63R5 Adult ECC, Custom Perfusion system, Sterile, Rx.
- Z-1885-2021 · Apr 29, 2021 · TerminatedMedtronic HY10Y54R4, Custom Pack HY10Y54R4 ECC PED 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1887-2021 · Apr 29, 2021 · TerminatedMedtronic CB10R79R6, Custom Pack CB10R79R6 1/4X1/4 Loop, Custom Perfusion System, Sterile, Rx.
- Z-1891-2021 · Apr 29, 2021 · TerminatedMedtronic CB10Y89R10, Custom Pack CB10Y89R10 ECC 3/8 A/V, Custom Perfusion System, Sterile, Rx.
…and 10 more.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.