NA BB11U86R5Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB11U86R5 AGH CPB · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000740320), received between Jan 15, 2026 and Jan 15, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1027-2023 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Delivered as Unsterile Product1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT WHILE UNLOADING THE PACKS FROM THE CRATE, IT WAS OBSERVED THAT THE BOTTOM PACK HAD A SHOE SIZED HOLE IN THE TYVEK LID. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT NOTHING APPEARS TO BE MISSING FROM THE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24087251 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.
- Z-1895-2021 · Apr 29, 2021 · TerminatedMedtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.
- Z-1883-2021 · Apr 29, 2021 · TerminatedMedtronic HY10J00R12, Custom Pack HY10J00R12 ECC Small CA, Custom Perfusion System,
- Z-1878-2021 · Apr 29, 2021 · TerminatedMedtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
- Z-1887-2021 · Apr 29, 2021 · TerminatedMedtronic CB10R79R6, Custom Pack CB10R79R6 1/4X1/4 Loop, Custom Perfusion System, Sterile, Rx.
- Z-1875-2021 · Apr 29, 2021 · TerminatedMedtronic HY7M07R6, Custom Pack HY7M07R6 E 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1881-2021 · Apr 29, 2021 · TerminatedMedtronic CB4W67R20, Custom Pack CB4W67R20 Adult E, Custom Perfusion System, Sterile, Rx.
- Z-1889-2021 · Apr 29, 2021 · TerminatedMedtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
- Z-1871-2021 · Apr 29, 2021 · TerminatedMedtronic HY10Z48R2, Ctr., Custom Pack HY10Z48R2 Adult ECC 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1880-2021 · Apr 29, 2021 · TerminatedMedtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
- Z-1894-2021 · Apr 29, 2021 · TerminatedMedtronic BB9G78R8, Custom Pack BB9G78R8 ECC Supp, Custom Perfusion System, Sterile, Rx.
- Z-1891-2021 · Apr 29, 2021 · TerminatedMedtronic CB10Y89R10, Custom Pack CB10Y89R10 ECC 3/8 A/V, Custom Perfusion System, Sterile, Rx.
- Z-1892-2021 · Apr 29, 2021 · TerminatedMedtronic BB8M63R5, Custom Pack BB8M63R5 Adult ECC, Custom Perfusion system, Sterile, Rx.
- Z-1885-2021 · Apr 29, 2021 · TerminatedMedtronic HY10Y54R4, Custom Pack HY10Y54R4 ECC PED 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1879-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W63R1, Custom Pack CB10W63R1 HL 20 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1873-2021 · Apr 29, 2021 · TerminatedMedtronic BB10R94R6, Custom Pack BB10R94R6 1/4 ECCPK, Custom Perfusion System, Sterile, Rx.
…and 10 more.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.