This Medtronic InterStim X is an implanted neurostimulator for urinary and bowel dysfunction. Reports describe energy not being delivered, unexpected electric shocks, and pain, and some mention the device moving out of place or infection. The records include 1,905 reports, 812 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,840
- Deaths reported
- 0
- Injuries reported
- 790
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,840 adverse event reports that list this exact device (DI 00763000892067), received between Sep 17, 2024 and Sep 30, 2026 — classified as 790 injuries, 1,050 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (323 reports).
7 other recalls cover devices cleared under the same FDA submission (P080025, P970004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem438
- Insufficient Device Problem Information323
- Failure to Deliver Energy242
- Migration or Expulsion of Device242
- Inappropriate/Inadequate Shock/Stimulation241
- Energy Output Problem223
- Therapy Delivered to Incorrect Body Area157
- Electromagnetic Compatibility Problem146
Patient problems
- Insufficient Information606
- No Clinical Signs, Symptoms or Conditions429
- Pain411
- Incontinence271
- Electric Shock240
- Discomfort239
- Urinary Frequency / Polyuria209
- Urinary Retention200
Recent adverse event reports
The 25 most recent of 1,840 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Overheating of Device
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for fecal incontinence, urinary urge incontinence, and urinary/bowel dysfunction. It was reported that the patient (pt) was having accidents every day and the area around the implant was warm to the touch since yesterday. Patient stated the implant helped before yesterday because they had not had any accidents at all, but yesterday the symptoms came back full force. The patient was redirected to their healthcare provider (HCP) to further address the issue. Patient mentioned they would have an appointment with wound care next Tuesday.
Patient problems: Burning Sensation; Incontinence
MAUDE report 50948696 ↗InjurySep 30, 2026Inappropriate/Inadequate Shock/Stimulation
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for unknown indications for use. It was reported that the reason for call was the patient (who was implanted yesterday) reported they were in pain. Patient further clarified, and stated they didn't feel it at first when they were at the medical facility but after the hour and a half drive home and as they were getting out of the car, it felt kind of \"weird, like a fishing line or something.\" The patient said it felt like rough skin and soap that kind of burned and not something they were used to. The patient said they used a bidet but that hadn't made it go away and at first they didn't feel the vibrations but they started to feel the sensation like one would get if they used a TENS unit and the pad was peeling off or not quite attached and \"you get a zap\" that's what it felt like near their vaginal area and it was hurting, like a \"deep deep ache\" and it was kind of intense. The patient said it was not external that the sensations were internal. They described a sensation like a \"wavy line\" in their system, that they could feel it vibrating like a \"straight sharp line, like
Patient problems: Electric Shock; Insufficient Information; Pain
MAUDE report 50948637 ↗InjurySep 30, 2026Electromagnetic Interference; Energy Output Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that the patient (pt) noticed after having an MRI an unusual feeling on their vaginal area. The pt mentioned they didn't know when the \"dancing sensation\" on their vaginal area started (asked event date unknown). The pt stated the MRI facility did not set their INS on MRI mode and the pt thought it might have caused the unwanted sensation on their vaginal area. The agent reviewed general therapy and device-use information about putting the INS on MRI Mode. The pt mentioned the unwanted sensation on their vaginal area were addressed already by their healthcare provider (HCP) and their INS therapy was adjusted by the HCP. The pt showed concern why the MRI facility were not trained enough to make sure their INS were set on MRI mode prior having the MRI. The pt would be requesting a manufacturer representative (rep) to be present next time that they will be having an MRI. The pt decided to call their HCP to make special arrangements as well when they have their next MRI.
Patient problems: Insufficient Information
MAUDE report 50948612 ↗MalfunctionSep 30, 2026Energy Output Problem; Insufficient Device Problem Information
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction It was reported that the reason for the call was the caller stated that for the past week or two they had been having issues with their handset and communicator not charging. The caller stated that it won't work, won't pick up, deactivate, or change the levels that they needed to change. The caller added that they were given a new cord, but that didn't resolve the issue. The caller did not have the equipment with them at the time of the call. The caller will call back with the equipment for further troubleshooting.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947434 ↗InjurySep 30, 2026Electromagnetic Interference; Failure to Deliver Energy
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for Urinary/bowel dysfunction. It was reported that ever since Wednesday when they had an MRI. They feel like they have more incontinence. Reviewed some therapy education information and redirected to their doctor. The patient (Pt) said they have not had any falls or traumas and yesterday had a swallowing test, 2 weeks ago they went to the ER due to problems with their heart, they had been on half a diuretic pill and went to one pill but when they were on one pill they are committed to being in the bathroom all the time so they had to go back to a half a pill. The patient (Pt) said they were on a diuretic before after another ER visit 2 months ago. The patient was redirected to their doctor.
Patient problems: Incontinence; Insufficient Information
MAUDE report 50947110 ↗MalfunctionSep 30, 2026Electromagnetic Interference; Energy Output Problem
Information was received from a patient (pt) who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that they noticed today that the communicator was not responding and they didn't have a way to turn off their therapy as they pass through airport security gates. The pt mentioned the communicator would show a blinking yellow light for 60 seconds then would turn off. The pt stated they tried charging the communicator, but it didn't show any indicator lights. The pt added their programmer takes time to charge but was still working and had 67% battery life. The agent reviewed general therapy and device-use information and agent sent repair request to replace the communicator. The agent sent expedited delivery for the replacement. The agent called pt back to gather additional event information, pt answered and provided information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943077 ↗InjurySep 30, 2026Migration or Expulsion of Device; Insufficient Device Problem Information
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that it helped their symptoms in the beginning but that did not last for more than a couple of weeks and now they are getting up 4-5 times a night or have to go every 45 minutes during the day. Pt said they met with the Medtronic rep (asked/unknown name/date) who put them on program C with cycling but when therapy came on it made their big toe curl and didn't help their symptoms, they turned therapy OFF the night before last and notice no difference in their symptoms when off or on. Pt said the doctor had the patient go for an xray and pt saw the wire looked a little crooked but the doctor didn't think it was an issue. Pt said they need to reset it again, the rep told them they would call in a couple of days but pt has waited several weeks but has not heard from the rep. Reviewed rep role, reviewed some therapy education, offered and assisted pt to use their equipment and pt was successful to change programs and increase confirming comfortable sensation. Pt to monitor the effect and continue working closely with their doctor. Offered and
Patient problems: No Clinical Signs, Symptoms or Conditions; Sleep Dysfunction; Urinary Frequency / Polyuria
MAUDE report 50942143 ↗InjurySep 30, 2026Insufficient Device Problem Information
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that they just had the INS revised on Monday. The agent reviewed general therapy and device-use information. The pt changed the therapy program and increased therapy settings to a comfortable level.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942108 ↗InjurySep 29, 2026Therapy Delivered to Incorrect Body Area; Insufficient Device Problem Information
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that loss of therapeutic benefit. The patient stated that they were very unsatisfied because they had zero symptom relief for six weeks and no one had called them. The patient reported that they had called the field representative three times but never received a call back. The patient confirmed manufacturer rep was notified of this issue a week to a week and a half after implant date via phone and text but never receive a response. The patient mentioned that only after the third call did, they receive a text message to come to the clinic today at noon, but they stated that they could not come. The patient also stated that they had not spoken with their doctor about the complication, but rather with the field representative, who stated that the pain was not related to the INS. The patient stated that nothing ever been checked since they received the implant. The patient they spoke to NP because their wound not heal was told to leave it alone and up until now the wound still wouldn't heal. The patient stated that they could feel nothing d
Patient problems: Ambulation or Postural Difficulties; Impaired Healing; Inappropriate Device Stimulation of Tissue; Incontinence; Insufficient Information; Pain
MAUDE report 50938019 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a patient via a manufacturer representative (rep) regarding a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that there was a suspected infection. Additional symptoms included redness and fluid discharge. The issue was ongoing. Antibiotics were required and were released to the healthcare provider (HCP) with prescribed antibiotics on 2026-Sep-03. The patient reported picking at the incision site previously.
Patient problems: Erythema; Fluid Discharge; Unspecified Infection
MAUDE report 50936874 ↗InjurySep 29, 2026Energy Output Problem; Electromagnetic Compatibility Problem
Information was received from a friend/family member regarding a patient (pt) who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that the pt had been having more issues with their therapy since after their surgery. Caller stated that probably like 2 weeks ago, the pt had been having more issues, but caller did not specify if this was regarding the pt having a return of symptoms. Caller was also unsure if the pt had their therapy turned off during the surgery and the caller wanted to check the if the pt's therapy was still on. Agent reviewed general device use and caller was able to connect to the pt's INS without any issues. Caller also confirmed that the pt's therapy was turned on. Agent then reviewed general therapy information and redirected the caller to the pt's healthcare provider (HCP) if they wanted to make sure there were no issues with the INS. Caller stated that they have an appointment with the pt's HCP in October.
Patient problems: Insufficient Information
MAUDE report 50935465 ↗InjurySep 29, 2026Therapy Delivered to Incorrect Body Area; Electromagnetic Compatibility Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that the patient reported stimulation in an unexpected area. The patient stated that about a week ago, when they turned their therapy back on after surgery, they felt the stimulation mostly on their tailbone. The patient stated that they did not change the program; however, they noticed that the pocket where the INS was implanted never used to bother them, and now when they were lying on their left side, it was quite uncomfortable. The agent did not ask about the circumstances that led to the reported issue. The agent reviewed considerations for therapy optimization, general programming guidance, and general use of external devices. The agent walked the patient through connecting to the INS. The patient elected to increase the amplitude. They were able to connect to the INS and increase the stimulation. During the call, the patient felt the stimulation on their tailbone instead of their bicycle seat area. The patient was redirected to their doctor if the issue persisted.
Patient problems: Discomfort; Inappropriate Device Stimulation of Tissue
MAUDE report 50935167 ↗MalfunctionSep 29, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932641 ↗InjurySep 29, 2026Inappropriate/Inadequate Shock/Stimulation; Electromagnetic Compatibility Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for Urinary/Bowel Dysfunction. It was reported that they felt shocking all night long on their right leg, where they have an implant on their right knee. The patient also went to the bathroom about 10-12 times last night. They said they were heavily medicated on Tuesday and yesterday they were taking pain pills but last night they could not sleep. They did not know if the shocking was coming from the device, but the INS was put on the same side as their knee. The patient was redirected to their healthcare provider (HCP) to further address the issue. The agent asked patient if they wanted to decrease the stimulation intensity of if they would prefer to turn the therapy off while they discussed the issue with their HCP but they said that they did not have the devices with them and disconnected the call. The patient called back and stated that they were having pain, also they got a knee replacement on their right side and was emitting an electric pain. They stated that they were having pain pills to see if that helps. They also mentioned that they were still having accidents and last ni
Patient problems: Electric Shock; Incontinence; Pain; Sleep Dysfunction; Urinary Frequency / Polyuria
MAUDE report 50927535 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and urge incontinence. It was reported that the therapy had never really helped their symptoms since they got the therapy but their \"residuals now\" were a lot more than they were at the beginning. Patient said over the course of time, it seemed that it got to where it was not helping at all. The patient said they worked with a nurse at their managing healthcare provider's (HCP's) office, and they had tried each program and none of the programs had helped and increasing higher didn't do anything for them either. The patient said the HCP had told them if none of the programs had helped them, they could always start over again with the programs so the patient had called for assistance in changing programs. Agent reviewed device description/function as well as programming and stimulation considerations and walked the patient through connecting to their settings and switching to a new program. Patient said \"that was so easy.\" Patient set to increase stimulation up and monitor their symptoms. The patient said they might wait a week before moving on to the
Patient problems: Urinary Retention
MAUDE report 50925158 ↗InjurySep 28, 2026Therapy Delivered to Incorrect Body Area; Patient Device Interaction Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary urge incontinence and urinary/bowel dysfunction. It was reported that the patient (pt) was trying to turn their stimulation down a notch or two because it feels like there's a TENS unit in their feet and butt cheeks so it's kind of uncomfortable to poop. Caller added that when they sit, it's like they have a tingling sensation at the bottom of their feet. Agent asked caller when they first noticed this and caller stated it was after the device was implanted, maybe like 2-3 days after the procedure. Reviewed therapy information and general programming guidance. Caller was able to successfully connect to their implant and decrease the stimulation. Patient would maintain stimulation level and will continue to track symptoms. Redirected to healthcare provider (HCP) if issue persisted.
Patient problems: Discomfort; Inappropriate Device Stimulation of Tissue
MAUDE report 50924361 ↗InjurySep 28, 2026Migration or Expulsion of Device; Insufficient Device Problem Information
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary dysfunction/sacral nerve stim and gastrointestinal/ pelvic floor. It was reported that they had to go get an MRI and they turned off the therapy and turned it back on and it was working great. Patient stated at night they would only have to wake up twice to go to the bathroom and they did not have any leaking. Patient reported in the last week they have been waking up 7-8 times and they had a leak again since about 4 days ago. Agent asked patient to connect with the external devices and patient stated the programmer was not charged. Patient was able to connect to the handset and confirmed the battery was depleted. The issue was not resolved through troubleshooting. The patient stated they will call back when the programmer is charged so they can check on the status of the therapy. Additional information was received from the patient. They reported previously reported information regarding their symptoms. Patient added that they were having to wake up at night 3-5 times again, and that if they drank a little water, they couldn't hold it until they get to the restroom. Pat
Patient problems: Micturition Urgency; Sleep Dysfunction; Urinary Frequency / Polyuria; Urinary Incontinence
MAUDE report 50924024 ↗MalfunctionSep 28, 2026Energy Output Problem; Electromagnetic Compatibility Problem; Therapeutic or Diagnostic Output Failure; Insufficient Device Problem Information
Additional information was received from the healthcare provider (HCP). The HCP indicated that the cause of not being able to turn off the INS in June was determined to be due to trying to turn the device off for an MRI. It was not determined the cause of the system operating at max settings, and it was unknown if this occurred due to EMI. The steps taken to resolve this was to contact a manufacturer representative (rep). The cause of the communicator not working was not determined. This issue was not resolved and no further action would be taken.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923068 ↗InjurySep 28, 2026Failure to Deliver Energy; Energy Output Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for fecal incontinence and urinary/bowel dysfunction. It was reported that they were having some pain and wondered if it was related to their device. The patient said the pain was on the left side of the labial area but noted it was getting better. The agent cold transferred patient to patient services (PS) PH queue due to long hold. The patient called back and mentioned previously reported information regarding the pain in their labial area on the lefthand side. The patient mentioned that the pain was a lot better than it was, but the pain hadn't gone completely, so they were wondering if it was caused by the INS. The patient stated that they couldn't turn their therapy off or lower their stimulation because if they did, they couldn't urinate so they were afraid to have accidents. The agent reviewed general therapy information and programming guidance. The patient then agreed that the best course of action was to follow-up with their healthcare provider (HCP) to further address the issue. The agent also reviewed role of manufacturing representative (rep) and provided patient with d
Patient problems: Incontinence; Pain
MAUDE report 50917627 ↗InjurySep 28, 2026Therapy Delivered to Incorrect Body Area; Insufficient Device Problem Information
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that the reason for the call was the patient said the device was not working very well. The patient said they kept upping the stimulation and it didn't get a proper response. The patient mentioned that they were talking to a representative and they kept increasing the stimulation to 1.3 on program 1 and it was still not helping. The agent reviewed with the patient that they may need to increase the stimulation further or try a different program to see if that different patternof stimulation would help. The patient also mentioned that their sciatica had gotten worse since implant, and they weren't sure if the device would cause their symptoms to exacerbate. The patient was redirected to their healthcare provider to further address the issue. On 2026-Sep-25 additional information was received from the patient. The caller stated the equipment was not working and the patient was getting the device not responding message and to try again. The agent walked the caller through feeling for the implant with their fingers and putting the communicat
Patient problems: Discomfort; Inappropriate Device Stimulation of Tissue; Insufficient Information; Pain
MAUDE report 50916439 ↗MalfunctionSep 28, 2026Electromagnetic Interference; Overheating of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915218 ↗InjurySep 28, 2026Failure to Deliver Energy; Energy Output Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that the reason for the call was the patient reported they screwed up the program and the intensity about 4 days ago. The patient stated the therapy was not helping at all. The patient stated that stimulation was not comfortable. The agent reviewed with the patient that the settings that they were using during the trial were not necessarily going to work for the permanent implant. The agent offered to assist the patient with turning down the level of stimulation to a comfortable level. The patient stated that they turned it down and now they didn't feel anything. The patient was redirected to their healthcare provider to further address the issue.
Patient problems: Discomfort
MAUDE report 50914862 ↗InjurySep 28, 2026Intermittent Continuity; Unintended Collision
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urge incontinence and urinary/bowel dysfunction. It was reported that since the beginning of this week noticed some tingling in their left toes while setting and watching tv in the evening around 8-9pm. The patient said it was intermittent and doesn't feel at any other time. The patient also mentioned that since Monday or Tuesday has started to feel a burning sensation in their vaginal area. The patient said that they hadn't made any adjustments to their settings and hadn't used the external devices in a long time, since February. When asked, patient said that did have a fall on August 1st, fell backwards. Reviewed general use of external devices and it was determined that patient had to plug in the handset. The patient did end up moving the adaptor to a separate outlet and then started to see a percentage and was able to power on the handset. With instruction, patient opened the therapy app, turned the communicator on and successfully connected to implant. Reviewed general programming guidance and patient decreased the setting until didn't feel any burning sensation. With instru
Patient problems: Burning Sensation; Paresthesia
MAUDE report 50914851 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a friend/family member regarding a patient who was implanted with an implantable neurostimulator (INS) for unknown indications for use. It was reported that the therapy has not helped the patient's symptoms at all since the INS was implanted yesterday, and their symptoms actually got worse. The patient said they were peeing all over themselves and had several accidents in their clothes. The patient also mentioned that they were in pain all night, but the caller did not think the pain was coming from the INS. The patient met with their managing healthcare provider (HCP) today to address their concerns. The HCP advised the patient to see their primary care doctor for the pain, and to contact their manufacturing representative (rep) to address the therapy issue. Evidently the HCP did not provide the caller with the rep's contact information, which was why the caller reached out to Pat ient Services. They were also frustrated because the patient was trained under anesthesia. The caller mentioned that the patient's representative assisted them last night. The caller stated that the amplitude was 1.5 mA when they connected to the INS last night, then they i
Patient problems: Insufficient Information; Pain; Urinary Frequency / Polyuria; Urinary Incontinence; Urinary Retention
MAUDE report 50914800 ↗InjurySep 28, 2026Failure to Deliver Energy; Electromagnetic Compatibility Problem
Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that they were in the hospital in November and they had to disconnect and turn therapy off, they did not know if it was turned back on but since then they have had really bad symptoms, last weekend they were very incontinent with their urine. The patient said they need to make an appointment with their doctor at the office on Wednesday and the doctor's office told them to contact manufacturer to coordinate the appointment. They offered to help patient sync with their equipment to check if therapy was off or on and offered to assist but patient said they have done that before with the manufacturing representative (rep). Reviewed the role of the rep and offered and sent email to the rep in the area. The patient was redirected to their healthcare provider to further address the issue. When agent asked the date for when their symptoms started and pt answered off and on since implant
Patient problems: Urinary Incontinence
MAUDE report 50914438 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P080025 on Mar 14, 2011. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Sacral Nerve Stimulation Therapy System”. The FDA review took 899 days. The FDA has since approved 350 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗
Approved through premarket approval (PMA) P970004 on Sep 29, 1997. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Therapy System For Urinary Control”. The FDA review took 242 days. The FDA has since approved 449 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗
The FDA regulates product code EZW (stimulator, electrical, implantable, for incontinence) under 21 CFR 876.5270 as a Class III device. 21 CFR 876.5270 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under P080025, P970004. One submission often covers many models, so check the product and lot details in each record.
- Z-1704-2025 · Apr 3, 2025 · Open, ClassifiedEnhanced Verify Evaluation Handset (CFN HH90130FA)
- Z-0372-2025 · Sep 18, 2024 · Open, ClassifiedMedtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
- Z-2365-2023 · Jul 6, 2023 · Open, ClassifiedMedtronic InterStim Ground Pad REF 041826
- Z-0998-2022 · Feb 28, 2022 · TerminatedMEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA,
- Z-1147-2015 · Jan 9, 2015 · TerminatedThe Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
- Z-0037-2008 · Jun 25, 2007 · TerminatedMedtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implanta
- Z-0040-2008 · Jun 25, 2007 · TerminatedMedtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim
Common questions about the InterStim™ X 97800
Has the InterStim™ X 97800 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000892067), based on the FDA recall file dated Nov 27, 2026. However, 7 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the InterStim™ X 97800?
The FDA's MAUDE database lists 1,840 adverse event reports for this device, received between Sep 17, 2024 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Deliver Energy” (242), “Migration or Expulsion of Device” (242) and “Inappropriate/Inadequate Shock/Stimulation” (241); the most common patient problems are “Pain” (411), “Incontinence” (271) and “Electric Shock” (240). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the InterStim™ X 97800?
None of the reports are classified as deaths, 790 as injuries and 1,050 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the InterStim™ X 97800 FDA approved?
It was approved through premarket approval (PMA) P970004 on Sep 29, 1997; its record lists 2 FDA submissions in total.
Is the InterStim™ X 97800 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the InterStim™ X 97800, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand InterStim™ X, model 97800). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stimulator, electrical, implantable, for incontinence devices
- INTERSTIM® 3058Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence6 deaths23,076 reports
- InterStim™ Micro SureScan™ 97810Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence2,762 reports
- Axonics 4101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence1 death818 reports
- INTERSTIM® 3058Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence818 reports
- Axonics 1101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence1 death399 reports
- Axonics 5101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence162 reports
- INTERSTIM® 3023Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1 death147 reports
- INTERSTIM® 3023Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence110 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record