Hydro-Surg 0026870Recalls & reported problems

Davol Inc. · Hydro-Surg Plus Laparoscopic Irrigator with Nezhat-Dorsey SmokEvac Trumpet Valve, 33 cm Tip and Cojoined Suction and Irrigation Tubing · Class 2

All models of Hydro-Surg (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Davol Hydro-Surg laparoscopic irrigator is a surgical tool that flushes fluid and suctions during keyhole surgery. Most of its 209 reports describe fluid leaking from the device, often during setup, and some describe corrosion. The reports include 1 injury and no deaths. There is 1 open recall, saying the product cannot provide irrigation, which began in June 2022. No reports came in the last 12 months, versus 10 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
209
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 209 adverse event reports that list this exact device (DI 00801741000362), received between May 23, 2017 and May 22, 2025 — classified as 1 injury, 208 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (173 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 11 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 9 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 7, 2022Open, Classified

    The product is not able to provide irrigation.

    Z-1305-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak173
  • Corroded67
  • Detachment of Device or Device Component19
  • Device Contamination with Chemical or Other Material6
  • Overheating of Device5
  • Smoking5
  • Insufficient Device Problem Information4
  • Thermal Decomposition of Device4

Patient problems

  • No Clinical Signs, Symptoms or Conditions180
  • No Consequences Or Impact To Patient25
  • No Patient Involvement2
  • Laceration(s)1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 209 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, THREE HYDRO-SURG PLUS IRRIGATORS (DEVICE #1 , #2 & #3) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069861 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, FOUR HYDRO-SURG LAP IRRIGATORS (DEVICE #1, #2, #3 & #4) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069860 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, THREE HYDRO-SURG LAP IRRIGATORS (DEVICE #1, #2 & #3) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069844 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, FOUR HYDRO-SURG LAP IRRIGATORS (DEVICE #1, #2, #3 & #4) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069843 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, FOUR HYDRO-SURG LAP IRRIGATORS (DEVICE #1, #2, #3 & #4) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069842 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, THREE HYDRO-SURG LAP IRRIGATORS (DEVICE #1, #2 & #3) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069840 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, TWO HYDRO-SURG PLUS IRRIGATORS (DEVICE #1 & #2) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069838 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, TWO HYDRO-SURG PLUS IRRIGATORS (DEVICE #1 & #2) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069826 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING SET UP OF A PROCEDURE ON (B)(6) 2025, FOUR HYDRO-SURG LAP IRRIGATORS (DEVICE #1, #2, #3 & #4) WERE LEAKING IRRIGATION FLUID AFTER BEING PRIMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069819 ↗
  • MalfunctionJan 31, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING ROBOTIC CHOLECYSTECTOMY PROCEDURE ON (B)(6) 2025, HYDRO-SURG PLUS LAPAROSCOPIC IRRIGATOR WAS LEAKED PREOPERATIVELY. IT WAS REPORTED THAT THE LEAKING WAS NOTED "AT CONNECTION" AND A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21284594 ↗
  • MalfunctionSep 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, DURING A PROCEDURE, THE WHITE CAP (ND VALVE BODY PLUG) OF HYDRO-SURG PLUS IRRIGATOR POPPED OFF CAUSING THE HANDLE TO LEAK IRRIGATION FLUID. IT WAS REPORTED THAT THE OR STAFF TIGHTENED THE WHITE CAP PRIOR TO USE TO ENSURE IT WAS SECURED ONTO THE ND HANDLE. IT WAS ALSO REPORTED THAT ANOTHER IRRIGATOR WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20179705 ↗
  • MalfunctionAug 1, 2024Break; Break

    AS REPORTED, DURING LAPAROSCOPIC CHOLECYSTECTOMY, THE HYDRO-SURG IRRIGATOR BROKE AND MALFUNCTIONED DURING A CASE. IT WAS REPORTED THAT THE TWO DEVICES, THE PLASTIC AT THE END WHERE THE SUCTION TIP WAS CRACKED (MALE ADAPTER), MAKING IT IMPOSSIBLE TO GET THE PROPER SEAL AND SUCTION. THE SURGEON TRIED TURNING UP THE SUCTION FOR THE TWO THAT CRACKED BUT THE SUCTION TIP WOULD NOT STAY ATTACHED. IT WAS ALSO REPORTED THAT THE SURGEON USED ANOTHER NEW DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19882753 ↗
  • MalfunctionJul 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, DURING LAPAROSCOPIC HEMICOLECTOMY PROCEDURE ON (B)(6) 2024, THE PLASTIC WHITE CAP OF HYDRO-SURG IRRIGATOR FELL ONTO THE FLOOR WHEN HOOKING UP AT START OF CASE AND BECAME UNSTERILE. IT WAS REPORTED THAT TUBING DID NOT WORK, AND SUCTION COULD NOT BE MAINTAINED, IT WAS UNKNOWN WHETHER THE WHITE CAP WAS CHECKED AND SECURED PRIOR TO USE. THE PROCEDURE WAS COMPLETED USING ANOTHER HYDROSURG DEVICE OF THE SAME LOT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19646188 ↗
  • MalfunctionMar 7, 2024Corroded; Fluid/Blood Leak; Corroded; Fluid/Blood Leak

    AS REPORTED, DURING A PROCEDURE HYDRO-SURG HAD IRRIGATION FLUID LEAK. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18861283 ↗
  • MalfunctionFeb 26, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING AN UNKNOWN PROCEDURE HYDRO-SURG PLUS IRRIGATOR HAD A FLUID LEAK ISSUE. LEAKAGE ISSUE NOTED UPON HOOK UP OF THE IRRIGATION FLUID. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18784182 ↗
  • MalfunctionFeb 26, 2024Corroded; Fluid/Blood Leak; Corroded; Fluid/Blood Leak

    AS REPORTED, DURING A PROCEDURE, THE HYDRO-SURG LEAKED IRRIGATION FLUID IMMEDIATELY AFTER CONNECTING TO THE IRRIGATION FLUID. IT WAS REPORTED THAT NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE AND THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18784153 ↗
  • MalfunctionFeb 26, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING A PROCEDURE, THE HYDRO-SURG HAD IRRIGATION FLUID LEAK. THERE WAS NO PATIENT INJURY OR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18784152 ↗
  • MalfunctionFeb 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING A PROCEDURE, THE HYDRO-SURG LEAKED IRRIGATION FLUID WHERE THE TUBING AND BATTERY COMPARTMENT CONNECT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18713231 ↗
  • MalfunctionFeb 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING AN UNKNOWN PROCEDURE, THE HYDRO-SURG LEAKED IRRIGATION FLUID WHERE THE TUBING AND BATTERY COMPARTMENT CONNECT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18713230 ↗
  • MalfunctionFeb 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING AN UNKNOWN PROCEDURE, THE HYDRO-SURG LEAKED IRRIGATION FLUID WHERE THE TUBING AND BATTERY COMPARTMENT CONNECT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18713228 ↗
  • MalfunctionFeb 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING A PROCEDURE, THE HYDRO-SURG LEAKED IRRIGATION FLUID WHERE THE TUBING AND BATTERY COMPARTMENT CONNECT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18713091 ↗
  • MalfunctionFeb 5, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING AN UNKNOWN PROCEDURE HYDRO-SURG PLUS IRRIGATOR HAD A FLUID LEAK ISSUE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18640076 ↗
  • MalfunctionFeb 5, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING AN UNKNOWN PROCEDURE HYDRO-SURG PLUS IRRIGATOR HAD A FLUID LEAK ISSUE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18640075 ↗
  • MalfunctionFeb 5, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING A ROBOTIC-ASSISTED HYSTERECTOMY PROCEDURE HYDRO-SURG IRRIGATOR DEVICE WAS LEAKING IRRIGATION FLUID AT THE BASE OF THE PUMP ASSEMBLY IN THE BEGINNING OF A PROCEDURE AND THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18640072 ↗
  • MalfunctionFeb 5, 2024Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING AN UNKNOWN PROCEDURE HYDRO-SURG PLUS IRRIGATOR HAD A FLUID LEAK ISSUE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18640068 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →