Bardex® I.C. Foley Catheter 0165SI22Recalls & reported problems
C. R. Bard, Inc. · Bardex® I.C. 5cc/mL Balloon Foley Catheter · Class 2
All models of Bardex I.C. Foley Catheter (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00801741016783), received between Aug 15, 2017 and Sep 14, 2026 — classified as 1 injury, 18 malfunctions.
The most frequently reported specific device problem is “Partial Blockage” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Partial Blockage3
- Decrease in Pressure2
- Deflation Problem2
- Device Contamination with Chemical or Other Material2
- Difficult to Remove2
- Fluid/Blood Leak2
- Incorrect Measurement2
- Material Split, Cut or Torn2
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- No Consequences Or Impact To Patient3
- Pain3
- Urinary Retention2
- Irritability1
- Irritation1
- No Known Impact Or Consequence To Patient1
- Urinary Tract Infection1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50764426 ↗MalfunctionJan 21, 2026Difficult to Remove
IT WAS REPORTED THAT DURING USE THE FOLEY CATHETER HAD DIFFICULTY IN DRAINING WATER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24138117 ↗MalfunctionMay 20, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE FOLEY CATHETER HAD RUST ON IT. THE PHOTOS OF THE SECOND CATHETER WERE FOUND TO BE DAMAGED. THE PACKAGING WAS ALL SEALED UP BEFORE OPENING FIRST LAYER, THEN OPENING BLUE COVER ON CATHETER. AS IF MENTIONED, WAS SECOND ONE IN A MATTER OF WEEKS AND HAD THROWN OUT FIRST ONE BUT KEPT THIS ONE TO SEND TO BARD. THE FLASH IS TURNED ON SO COULD SEE THE BROWN STUFF ALONG WITH SIZE OF HOLE. THAT'S CLEARLY BIGGER, THEY'RE STORED IN THE BEDROOM UNIT BESIDE MY BED. NO EXCESSIVE HEAT, CONDENSATION. THE SIZE OF HOLE WAS FIRST CLUE, HAD WITH FIRST FAULTY CATHETER AND THIS ONE EXACT SAME, THEN TURNED FLASH ON AND SAW HOW GROSS IT ACTUALLY WAS. THE PATIENT HAD THE TWO UNITS SINCE JANUARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22054002 ↗MalfunctionMay 20, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE FOLEY CATHETER HAD RUST ON IT. THE PHOTOS OF THE SECOND CATHETER WERE FOUND TO BE DAMAGED. THE PACKAGING WAS ALL SEALED UP BEFORE OPENING FIRST LAYER, THEN OPENING BLUE COVER ON CATHETER. AS IF MENTIONED, WAS SECOND ONE IN A MATTER OF WEEKS AND HAD THROWN OUT FIRST ONE BUT KEPT THIS ONE TO SEND TO BARD. THE FLASH IS TURNED ON SO COULD SEE THE BROWN STUFF ALONG WITH SIZE OF HOLE. THAT'S CLEARLY BIGGER, THEY'RE STORED IN THE BEDROOM UNIT BESIDE MY BED. NO EXCESSIVE HEAT, CONDENSATION. THE SIZE OF HOLE WAS FIRST CLUE, HAD WITH FIRST FAULTY CATHETER AND THIS ONE EXACT SAME, THEN TURNED FLASH ON AND SAW HOW GROSS IT ACTUALLY WAS. THE PATIENT HAD THE TWO UNITS SINCE JANUARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22053416 ↗MalfunctionMay 20, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THE FOLEY CATHETER HAD RUST ON IT. THE PHOTOS OF THE SECOND CATHETER WERE FOUND TO BE DAMAGED. THE PACKAGING WAS ALL SEALED UP BEFORE OPENING FIRST LAYER, THEN OPENING BLUE COVER ON CATHETER. AS IF MENTIONED, WAS SECOND ONE IN A MATTER OF WEEKS AND HAD THROWN OUT FIRST ONE BUT KEPT THIS ONE TO SEND TO BARD. THE FLASH IS TURNED ON SO COULD SEE THE BROWN STUFF ALONG WITH SIZE OF HOLE. THAT'S CLEARLY BIGGER, THEY'RE STORED IN THE BEDROOM UNIT BESIDE MY BED. NO EXCESSIVE HEAT, CONDENSATION. THE SIZE OF HOLE WAS FIRST CLUE, HAD WITH FIRST FAULTY CATHETER AND THIS ONE EXACT SAME, THEN TURNED FLASH ON AND SAW HOW GROSS IT ACTUALLY WAS. THE PATIENT HAD THE TWO UNITS SINCE JANUARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22053414 ↗MalfunctionMay 20, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THE FOLEY CATHETER HAD RUST ON IT. THE PHOTOS OF THE SECOND CATHETER WERE FOUND TO BE DAMAGED. THE PACKAGING WAS ALL SEALED UP BEFORE OPENING FIRST LAYER, THEN OPENING BLUE COVER ON CATHETER. AS IF MENTIONED, WAS SECOND ONE IN A MATTER OF WEEKS AND HAD THROWN OUT FIRST ONE BUT KEPT THIS ONE TO SEND TO BARD. THE FLASH IS TURNED ON SO COULD SEE THE BROWN STUFF ALONG WITH SIZE OF HOLE. THAT'S CLEARLY BIGGER, THEY'RE STORED IN THE BEDROOM UNIT BESIDE MY BED. NO EXCESSIVE HEAT, CONDENSATION. THE SIZE OF HOLE WAS FIRST CLUE, HAD WITH FIRST FAULTY CATHETER AND THIS ONE EXACT SAME, THEN TURNED FLASH ON AND SAW HOW GROSS IT ACTUALLY WAS. THE PATIENT HAD THE TWO UNITS SINCE JANUARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22052128 ↗MalfunctionJul 6, 2022Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE STERILIZED WATER LEAKED FROM THE INFLATION VALVE DURING PRETEST. THERE WAS NO BALLOON RUPTURE OR TEAR AND THE PATIENT DID NOT HAVE ANY URINE STONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14938306 ↗MalfunctionMar 17, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A CUT MARK FOUND IN THE SHAFT OF THE FOLEY CATHETER DURING PRETEST AND THE WATER LEAKED FROM THERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13796252 ↗MalfunctionJan 28, 2022Burst Container or Vessel; Material Split, Cut or Torn; Burst Container or Vessel; Material Split, Cut or Torn
IT WAS REPORTED THAT WATER LEAKED FROM THE FOLEY CATHETER BALLOON DURING PRETEST. IT WAS MENTIONED THAT A RUPTURE OR TEAR WAS FOUND ON THE FOLEY CATHETER BALLOON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13389209 ↗MalfunctionDec 12, 2021Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE FOLEY CATHETER WAS REMOVED SPONTANEOUSLY. WHEN THE BALLOON PART WAS CONFIRMED, THE DISTILLED WATER WAS DRAINED. IT WAS ALSO REPORTED THAT THERE WAS NO BALLOON RUPTURE OR TEAR. AS PER FOLLOW UP INFORMATION RECEIVED ON 02DEC2021, DAMAGE TO THE CATHETER WAS UNKNOWN AND THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12981763 ↗MalfunctionSep 21, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE FOLEY CATHETER HAD LEAKAGE ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12505821 ↗MalfunctionApr 6, 2021Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove
IT WAS REPORTED THAT THE FOLEY CATHETER DID NOT DEFLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11625146 ↗InjuryJan 2, 2021Biocompatibility; Biocompatibility
IT WAS REPORTED THAT THE PATIENT HAD RE OCCURRING URINARY TRACT INFECTIONS. PATIENT HAD PAIN AND IRRITATION WHEN INSERTING CATHETER. DOCTOR STATED IT WAS ALL FROM USAGE OF THE CATHETER. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED FOR URINARY TRACT INFECTIONS.
Patient problems: Irritability; Irritation; Pain; Urinary Tract Infection
MAUDE report 11103903 ↗MalfunctionDec 22, 2020Failure to Infuse; Failure to Infuse
IT WAS REPORTED THAT THE CATHETER WAS DIFFICULT TO INFLATE. IT WAS FURTHER REPORTED THAT INSTEAD OF THE BALLOON, THE SHAFT INFLATED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11055933 ↗MalfunctionOct 9, 2020Partial Blockage; Partial Blockage
IT WAS REPORTED THAT THE FOLEY CATHETERS GOT CLOGGED UP SOMETIMES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10658697 ↗MalfunctionJan 28, 2020Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THE FOLEY CATHETERS WERE NOT DRAINING CORRECTLY, WHICH CAUSED THE URINE TO BACK UP AND CAUSE THE PATIENT PAIN. NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: Pain
MAUDE report 9638767 ↗MalfunctionDec 1, 2018Partial Blockage; Partial Blockage; Deflation Problem; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT THERE WAS URINE LEAKAGE FROM THE CATHETER AND THE USER WAS UNABLE TO DEFLATE THE BALLOON WHEN TRYING TO REPLACE THE CATHETER. THE OPERATOR CUT THE CATHETER ACCORDING TO THE INSTRUCTIONS AND USED A GUIDEWIRE, HOWEVER THE USER WAS STILL UNABLE TO DEFLATE THE BALLOON. ON THE NEXT DAY, WHEN PUSHING THE DRAINAGE LUMEN OF THE CATHETER WITH A STYLET, THE OPERATOR FELT THAT THERE WAS CALCULI IN IT, HOWEVER THE SITUATION DID NOT IMPROVE. AFTER THAT, THE USER COULD REMOVE THE CATHETER AS THE BALLOON DEFLATED. THE USER CONFIRMED THERE WAS A HOLE ON THE REMOVED BALLOON, WHICH SEEMED TO HAVE BEEN PERFORATED WITH A GUIDEWIRE ON THE DAY BEFORE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8121935 ↗MalfunctionNov 13, 2017No Flow
IT WAS REPORTED THAT THE CATHETER, PLACED SUPRAPUBICALLY, WOULD NOT ALLOW URINE TO DRAIN. PER FOLLOW UP WITH THE COMPLAINANT SOME PAIN WAS EXPERIENCED, HOWEVER NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: Urinary Retention
MAUDE report 7024271 ↗MalfunctionAug 15, 2017Partial Blockage; No Flow
IT WAS REPORTED THAT THE SUPRAPUBIC CATHETER WAS BLOCKED AND DID NOT ALLOW URINE TO DRAIN. THE PATIENT EXPERIENCED PAIN WHICH LEAD HIM TO REMOVE THE CATHETER AND REPLACE IT WITH A NEW ONE. NO MEDICAL INTERVENTION REQUIRED.
Patient problems: Pain; Urinary Retention
MAUDE report 6796258 ↗
Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119314C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall3,889 reports
- Bardex® I.C. Foley Catheter 0165SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,994 reports
- Bardex® I.C. Foley Catheter 0165SI16C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,052 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES778 reports
- Bardex® I.C. Temperature Sensing Foley Catheter 129414MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES436 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Temp Sensing Foley Catheter 119316MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES231 reports
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119316C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES198 reports
- Bardex® I.C. Foley Catheter 0165SI18C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES181 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page