Bardex® I.C. Foley Catheter 0166SI18Recalls & reported problems
C. R. Bard, Inc. · Bardex® I.C. 30cc/ml Balloon Foley Catheter · Class 2
All models of Bardex I.C. Foley Catheter (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00801741016851), received between Apr 8, 2019 and Jun 23, 2025 — classified as 2 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Off-Label Use” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Off-Label Use4
- Incorrect Measurement2
- Material Rupture2
- Partial Blockage2
- Burst Container or Vessel1
- Deflation Problem1
- Device Dislodged or Dislocated1
- Inadequacy of Device Shape and/or Size1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient3
- Foreign Body In Patient2
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT ON THE FIRST DAY OF PLACEMENT AFTER TOTAL CYSTECTOMY THE FOLEY CATHETER BALLOON BURST A WHILE LATER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22308646 ↗MalfunctionJun 17, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE BALLOON OF THE FOLEY CATHETER WAS RUPTURED AFTER 1ST DAY OF PLACEMENT OF A CATHETER IN THE PELVIS AFTER TOTAL CYSTECTOMY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22236024 ↗MalfunctionDec 13, 2024Partial Blockage; Partial Blockage
IT WAS REPORTED THAT FOLEY CATHETER GOT CLOGGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20925863 ↗MalfunctionMay 6, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE FOLEY CATHETER BALLOON WAS PLACED IN DETENTION ON (B)(6) 2024, AND THE CATHETER WAS REMOVED NATURALLY ON (B)(6) 2024 WHEN THE BALLOON RAN OUT OF WATER. PER FOLLOW UP INFORMATION RECEIVED VIA IBC ON 17APR2024, STATED THAT NO ABNORMALITIES WERE IDENTIFIED THROUGHOUT THE CATHETER. BAGS FROM OTHER MANUFACTURERS WERE DISCARDED. NO LEAFLET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19247734 ↗MalfunctionApr 21, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE FOLEY CATHETER BALLOON RUPTURED ABOUT 45 MINUTES AFTER INSERTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14166471 ↗InjuryApr 19, 2022Off-Label Use; Off-Label Use
IT WAS REPORTED THAT THE CUSTOMER HAS HAD 3 OF THE FOLEY CATHETERS BALLOONS RUPTURED WHICH THEY WERE USING FOR INDUCTION. STATED THAT THE DOCTOR INSERTED THE FOLEY CATHETER THROUGH THE CERVIX, AND THEY INFLATED THE BALLOON WITH APPROXIMATELY 60ML OF NORMAL SALINE, WHICH DID EXCEED THE RECOMMENDED AMOUNT OF 30ML, BUT THEY HAVE USED THEM FOR YEARS THIS WAY WITH HARDLY ANY PROBLEMS. CUSTOMER ALSO STATED THAT THE PATIENT LAST WEEK THAT HAD THE FOLEY BREAK WENT FOR A C-SECTION AND THE DOCTOR FOUND A PART OF THE BALLOON INSIDE THE UTERUS. IT IS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED. THE AFFECTED LOTS WERE NGFP3605, NGFT5738.
Patient problems: Foreign Body In Patient
MAUDE report 14140966 ↗MalfunctionApr 19, 2022Off-Label Use; Off-Label Use
IT WAS REPORTED THAT THE CUSTOMER HAS HAD 3 OF THE FOLEY CATHETERS BALLOONS RUPTURED WHICH THEY WERE USING FOR INDUCTION. STATED THAT THE DOCTOR INSERTED THE FOLEY CATHETER THROUGH THE CERVIX, AND THEY INFLATED THE BALLOON WITH APPROXIMATELY 60ML OF NORMAL SALINE, WHICH DID EXCEED THE RECOMMENDED AMOUNT OF 30ML, BUT THEY HAVE USED THEM FOR YEARS THIS WAY WITH HARDLY ANY PROBLEMS. CUSTOMER ALSO STATED THAT THE PATIENT LAST WEEK THAT HAD THE FOLEY BREAK WENT FOR A C-SECTION AND THE DOCTOR FOUND A PART OF THE BALLOON INSIDE THE UTERUS. IT IS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14140962 ↗MalfunctionApr 14, 2022Off-Label Use; Off-Label Use
IT WAS REPORTED THAT THE CUSTOMER HAS HAD 3 OF THE FOLEY CATHETERS BALLOONS RUPTURED WHICH THEY WERE USING FOR INDUCTION. STATED THAT THE DOCTOR INSERTED THE FOLEY CATHETER THROUGH THE CERVIX, AND THEY INFLATED THE BALLOON WITH APPROXIMATELY 60ML OF NORMAL SALINE, WHICH DID EXCEED THE RECOMMENDED AMOUNT OF 30ML, BUT THEY HAVE USED THEM FOR YEARS THIS WAY WITH HARDLY ANY PROBLEMS. CUSTOMER ALSO STATED THAT THE PATIENT LAST WEEK THAT HAD THE FOLEY BREAK WENT FOR A C-SECTION AND THE DOCTOR FOUND A PART OF THE BALLOON INSIDE THE UTERUS. IT IS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14112921 ↗InjuryApr 14, 2022Off-Label Use; Off-Label Use
IT WAS REPORTED THAT THE CUSTOMER HAS HAD 3 OF THE FOLEY CATHETERS BALLOONS RUPTURED WHICH THEY WERE USING FOR INDUCTION. STATED THAT THE DOCTOR INSERTED THE FOLEY CATHETER THROUGH THE CERVIX, AND THEY INFLATED THE BALLOON WITH APPROXIMATELY 60ML OF NORMAL SALINE, WHICH DID EXCEED THE RECOMMENDED AMOUNT OF 30ML, BUT THEY HAVE USED THEM FOR YEARS THIS WAY WITH HARDLY ANY PROBLEMS. CUSTOMER ALSO STATED THAT THE PATIENT LAST WEEK THAT HAD THE FOLEY BREAK WENT FOR A C-SECTION AND THE DOCTOR FOUND A PART OF THE BALLOON INSIDE THE UTERUS. IT IS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 14112920 ↗MalfunctionDec 23, 2019Deflation Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE BALLOON ON THE FOLEY CATHETER DEFLATED DURING USE AND CAUSED THE CATHETER TO FALL OUT OF THE PATIENT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9512705 ↗MalfunctionDec 23, 2019Partial Blockage; Partial Blockage
IT WAS REPORTED THAT THE FOLEY CATHETER GOT CLOGGED PREVENTING IT FROM DRAINING PROPERLY AT TIMES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9512423 ↗MalfunctionApr 9, 2019Incorrect Measurement; Material Deformation; Incorrect Measurement
IT WAS REPORTED THAT THE CATHETER APPEARED TO BE SHORTER THAN NORMAL.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8497732 ↗MalfunctionApr 8, 2019Incorrect Measurement; Inadequacy of Device Shape and/or Size; Incorrect Measurement
IT WAS REPORTED THAT THE FOLEY CATHETER WAS ABOUT 3 INCHES SHORTER THAN USUAL. THE CUSTOMER REPORTED THAT THE CATHETER DID NOT REACH THE SAME PLACE THE PREVIOUS ONES DID WHEN IT WAS INSERTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8492810 ↗
Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119314C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall3,889 reports
- Bardex® I.C. Foley Catheter 0165SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,994 reports
- Bardex® I.C. Foley Catheter 0165SI16C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,052 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES778 reports
- Bardex® I.C. Temperature Sensing Foley Catheter 129414MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES436 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Temp Sensing Foley Catheter 119316MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES231 reports
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119316C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES198 reports
- Bardex® I.C. Foley Catheter 0165SI18C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES181 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page