Bardex® I.C. Foley Catheter 0167SI22Recalls & reported problems
C. R. Bard, Inc. · Bardex® I.C. 30cc/ml Balloon Foley Catheter · Class 2
All models of Bardex I.C. Foley Catheter (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00801741017315), received between Oct 3, 2019 and Dec 16, 2025 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove4
- Inaccurate Flow Rate2
- Adverse Event Without Identified Device or Use Problem1
- Burst Container or Vessel1
- Complete Blockage1
- Defective Device1
- Deflation Problem1
- Fitting Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- No Patient Involvement1
- Tissue Breakdown1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 16, 2025Material Protrusion/Extrusion; Material Protrusion/Extrusion
IT WAS REPORTED THAT THEY HAVE A DEFECTIVE FOLEY CATHETER, ONE OF THE 3 HOLES ON THE INSERTION TIP WAS RAISED, AND THE FLUID DID NOT FLOW OUT. THE WHOLE LOT WAS PULLED, SO THEY HAVE A TOTAL OF 8 TO BE RETURNED. (MATERIAL# 0167SI22, LOT# NGKR1988).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23822279 ↗MalfunctionDec 16, 2025Inaccurate Flow Rate; Inaccurate Flow Rate
IT WAS REPORTED THAT THEY HAVE A DEFECTIVE FOLEY CATHETER, ONE OF THE 3 HOLES ON THE INSERTION TIP WAS RAISED, AND THE FLUID DID NOT FLOW OUT. THE WHOLE LOT WAS PULLED, SO THEY HAVE A TOTAL OF 8 TO BE RETURNED. (MATERIAL# 0167SI22, LOT# NGKR1988).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23821935 ↗MalfunctionSep 16, 2025Difficult to Remove; Defective Device; Difficult to Remove; Defective Device
IT WAS REPORTED THAT CUSTOMER HAVING ISSUES WITH 22FR 3-WAY 30CC FOLEY CATHETERS. THERE HAVE BEEN MULTIPLE INSTANCES WHERE THE UROLOGIST HAS HAD TROUBLE WITH ASPIRATING FROM THE FOLEY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23062588 ↗MalfunctionMay 20, 2025Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE FOLEY CATHETER DID NOT ASPIRATE OR IRRIGATE. IT DID ONE BUT NOT THE OTHER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22053988 ↗MalfunctionMar 10, 2025Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE FOLEY CATHETER WAS BLOCKED AND WOULD NOT INFLATE OR FLUSH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21558743 ↗MalfunctionFeb 3, 2023Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE FOLEY BALLOON RUPTURED AT THE END OF A CASE. STAFF NOTICED THE FOLEY COMING OUT AND REALIZED THE BALLOON HAD RUPTURED. THE FOLEY WAS THEN REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16293984 ↗InjuryDec 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE SURVEY RESPONDENT WAS ASKED WHETHER THEY BELIEVE THE BENEFITS OF USING THE PRODUCT OUTWEIGH THE POTENTIAL RISKS IN RELATION TO THE BARDEX I.C. 3-WAY IRRIGATION FOLEY CATHETER AS PART OF A CUSTOMER SURVEY. THE RESPONDENT INDICATED THAT NO, THE BENEFITS OF USING THIS PRODUCT DO NOT OUTWEIGH THE POTENTIAL RISKS. WHEN ASKED TO ELABORATE, THE RESPONDENT STATED URETHRAL INJURY. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED FOR URETHRAL INJURY (0167SI16, 0167SI18, 0167SI20, 0167SI22, 0167SI24, 0167SI26).
Patient problems: Tissue Breakdown
MAUDE report 15982637 ↗MalfunctionJun 30, 2021Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT 3WAY FOLEY CATHETER WAS INSERTED AS PART OF THE PROCEDURE AND IT WAS NOT FLOWING. EVEN THE CLOT WAS FLUSHED AND ADJUSTED, IT CONTINUED TO NOT FLOW. ALSO STATED THAT THE BALLOON CUFF WOULD NOT DEFLATE AND CATHETER WAS REMOVED BY MANEUVERING AND REPLACED WITH SAME TYPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12093657 ↗MalfunctionJun 30, 2021Inaccurate Flow Rate; Inaccurate Flow Rate
IT WAS REPORTED THAT 3-WAY FOLEY CATHETER WAS INSERTED AS PART OF THE PROCEDURE AND IT WAS NOT FLOWING. EVEN THE CLOT WAS FLUSHED AND ADJUSTED, IT CONTINUED TO NOT FLOW. ALSO STATED THAT THE BALLOON CUFF WOULD NOT DEFLATE, AND CATHETER WAS REMOVED AND REPLACED WITH SAME TYPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12093573 ↗MalfunctionDec 29, 2020Fitting Problem; Fitting Problem
IT WAS REPORTED THAT 2 OF THE FOLEY WERE UNABLE TO HOOK UP THE SYRINGE PROPERLY. PER FOLLOW UP INFORMATION RECEIVED ON (B)(6) 2020, STATED THAT 2 CATHETER WAS OPENED AND THE SYRINGE WAS NOT ATTACHED TO THE BALLOON PORT ON THE FIELD THEN USED A DIFFERENT SIZE 3-WAY CATHETER ON PATIENT AND KEPT ALL THE 6 FOR TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 11092803 ↗MalfunctionOct 3, 2019Deflation Problem; Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE FOLEY WOULD NOT DEFLATE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9151555 ↗MalfunctionOct 3, 2019Fluid/Blood Leak; Leak/Splash; Gas/Air Leak; Leak/Splash
IT WAS REPORTED THAT THE BALLOON PORT WAS LEAKING THE FLUID PLACED IN IT CAUSING THE BALLOON TO DEFLATE AND DISLODGE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9151280 ↗
Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119314C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall3,889 reports
- Bardex® I.C. Foley Catheter 0165SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,994 reports
- Bardex® I.C. Foley Catheter 0165SI16C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,052 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES778 reports
- Bardex® I.C. Temperature Sensing Foley Catheter 129414MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES436 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Temp Sensing Foley Catheter 119316MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES231 reports
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119316C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES198 reports
- Bardex® I.C. Foley Catheter 0165SI18C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES181 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page