Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 73022SIRecalls & reported problems
C. R. Bard, Inc. · Bardex® Lubri-Sil® I.C. All-Silicone Coated 30 cc 3 way Foley Catheter · Class 2
All models of Bardex Lubri-Sil I.C. All-Silicone Foley Catheter (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741025518), received between Nov 11, 2022 and Mar 11, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Invagination” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Invagination1
- Material Puncture/Hole1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 11, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT TRANSURETHRAL RESECTION OF THE PROSTATE PROCEDURE WAS DONE AT THE END PLACED A BARD 22 FR 30CC/ML BALLOON LATEX FREE FOLEY CATHETER. AFTER THE CATHETER WAS ATTACHED TO CONTINUOUS BLADDER IRRIGATION, THE SURGEON NOTICED SALINE LEAKING FROM THE CATHETER NEAR THE TRIPLE LUMEN SITE. FOLEY CATHETER WAS NOTED TO HAVE A PUNCTURE IN IT CAUSING FLUID TO LEAK OUT. PHYSICIAN HAD TO QUICKLY REPLACE FOLEY CATHETER WITH A NEW ONE. CATHETER AND PACKAGING WERE PLACED IN BAG FOR RISK MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21574459 ↗MalfunctionNov 11, 2022Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT THE NURSE HAD DEFLATED THE BALLOON AND ATTEMPTED TO REMOVE FOLEY CATHETER SEVERAL TIMES WITHOUT SUCCESS. ABOUT AN INCH OF THE FOLEY WAS STUCK IN THE PENIS. IT WAS STATED THAT THE NURSE HAD CALLED FOR HELP AND GOT A SECOND OPINION BEFORE CALLING THE DOCTOR. IT WAS A NON-LATEX 22 FR FOLEY CATHETER THAT SOMETIMES HAD A RIDGE NEAR THE BALLOON. AFTER THE NURSE CAME TO THE PATIENT TO PULL IT, THEY FOUND THE BALLOON HAD DEFLATED AND THE PLASTIC HAD FOLDED OVER ON ITSELF INSTEAD OF LYING FLAT. STATED THAT CAUSED PAIN FOR THE PATIENT WHO HAD TO SIT WITH IT AT THE TIP OF THE PENIS FOR APPROXIMATELY 40 MINUTES WHILE STAFF TROUBLESHOOT THE PROBLEM. DOCTOR HAD TO TUG ON THE CATHETER FIRMLY UNTIL IT POPPED OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15780797 ↗MalfunctionNov 11, 2022Material Invagination; Material Invagination
IT WAS REPORTED THAT THE NURSE HAD DEFLATED THE BALLOON AND ATTEMPTED TO REMOVE FOLEY CATHETER SEVERAL TIMES WITHOUT SUCCESS. ABOUT AN INCH OF THE FOLEY WAS STUCK IN THE PENIS. IT WAS STATED THAT THE NURSE HAD CALLED FOR HELP AND GOT A SECOND OPINION BEFORE CALLING THE DOCTOR. IT WAS A NON-LATEX 22 FR FOLEY CATHETER THAT SOMETIMES HAD A RIDGE NEAR THE BALLOON. AFTER THE NURSE CAME TO THE PATIENT TO PULL IT, THEY FOUND THE BALLOON HAD DEFLATED AND THE PLASTIC HAD FOLDED OVER ON ITSELF INSTEAD OF LYING FLAT. ALSO STATED THAT CAUSED PAIN FOR THE PATIENT WHO HAD TO SIT WITH IT AT THE TIP OF THE PENIS FOR APPROXIMATELY 40 MINUTES WHILE STAFF TROUBLESHOOT THE PROBLEM. DOCTOR HAD TO TUG ON THE CATHETER FIRMLY UNTIL IT POPPED OUT.
Patient problems: Pain
MAUDE report 15779333 ↗
Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119314C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall3,889 reports
- Bardex® I.C. Foley Catheter 0165SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,994 reports
- Bardex® I.C. Foley Catheter 0165SI16C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,052 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES778 reports
- Bardex® I.C. Temperature Sensing Foley Catheter 129414MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES436 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Temp Sensing Foley Catheter 119316MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES231 reports
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119316C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES198 reports
- Bardex® I.C. Foley Catheter 0165SI18C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES181 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page