PowerPICC® Catheter 3174335Recalls & reported problems
Bard Access Systems, Inc. · PowerPICC Catheter 4F Single-Lumen Polyurethane Catheter Radiology Basic Tray (135cm Nitinol Guidewire) · Class 2 · Implant
All models of PowerPICC Catheter (33)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00801741027628), received between Jul 11, 2019 and Feb 27, 2026 — classified as 1 injury, 14 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Class IOpen, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak11
- Material Deformation3
- Deformation Due to Compressive Stress2
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Known Impact Or Consequence To Patient2
- Burning Sensation1
- Extravasation1
- Swelling/ Edema1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2026Fluid/Blood Leak
IT WAS REPORTED THAT PICC HAD A CRACK AND WAS LEAKING AROUND 2CM MARK INTERNALLY. PICC REMOVED AND A NEW ONE PLACED. CATHETER WAS SECURED WITH STATLOCK. PATIENT DIDN'T HAVE ANY SYMPTOMS RELATED TO THE LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24471653 ↗MalfunctionJun 9, 2025Material Deformation; Material Deformation
IT WAS REPORTED WE HAVE HAD INCREASED INCIDENCE OF TROUBLE WITH THE PEEL AWAY SHEATH IN THE PICC KIT. IT HAS BEEN BENDING VERY EASILY WHEN TRYING TO INTRODUCE OVER THE WIRE. ALSO LIPPING AT THE SHEATH SO IT'S NOT A SMOOTH TRANSITION FROM THE INTRODUCER TO THE PEEL AWAY. THE EXACT NUMBER OF OCCURRENCES WAS NOT PROVIDED BY THE COMPLAINANT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22170566 ↗MalfunctionMay 15, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED PATIENT PRESENTS WITH LEAKING OF LINE. WHEN FLUSHING, THE LINE IS LEAKING FROM UNDER THE SOLO VALVE LUER SITE. WHEN PULLED BACK, AIR IS ENTERING THE SYRINGE. PATIENT IS LAID BACK AND CURRENT PICC LINE IS REMOVED. PRESSURE HELD TO SITE UNTIL HOMEOSTASIS IS RECEIVED AND PATIENT LEFT IN THE SUPINE POSITION FOR VASCULAR ASSESSMENT OF THE ARM. ONCE REMOVED, LINE IS VISUALIZED TO HAVE SMALL HOLE APPROXIMATELY 5-6 CM UP LINE WHERE MEDICATION IS DRIPPING FROM, AT THE 6 OR 7 CM MARK. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22031929 ↗InjuryMay 9, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED; PATIENT WAS IN CLINIC RECEIVING CHEMOTHERAPY TREATMENT. ABOUT HALFWAY THROUGH TREATMENT, LEAKING WAS NOTED FROM THE PICC INSERTION SITE. AREA AROUND THE INSERTION SITE WAS ALSO NOTED TO BE RED/PURPLE. MORE PROXIMAL TO THE PATIENT, TOWARDS HER ARMPIT, AN AREA OF EDEMA WAS NOTED. PATIENT WAS ALSO EXPERIENCING A STINGING SENSATION. ORDERS WERE RECEIVED TO DISCONTINUE THE PICC LINE. ON INSPECTION, A CRACK WAS NOTICED IN THE LINE ITSELF. AS A RESULT, THE PATIENT EXPERIENCED AN EXTRAVASATION. PATIENT NOTED SOME BLEEDING FROM SITE WHEN ONSET. ONCOLOGY STAFF WERE ABLE TO ACCESS THE PICC LINE, WITHDRAW AND FLUSH PRIOR TO CHEMOTHERAPY ADMINISTRATION MEDICATION CONNECTED AND STARTED AS PER INFUSION GUIDELINES. PATIENT HAD PICC INSERTED ON (B)(6) 2025. LINE REMOVED AS PER PROCEDURE, NOTED CRACK 62 MM FROM HUB. REPLACEMENT OF PICC BEING ORGANIZED BY DOCTOR. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Burning Sensation; Extravasation; Swelling/ Edema
MAUDE report 21995289 ↗MalfunctionMay 7, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED "IR PLACED 4F SINGLE LUMEN PICC. CLINICIAN REPORTED LEAKING FROM THE END CAP. END CAP WAS REPLACED A FEW TIMES. A NEW SL PICC WAS PLACED. THE SAME THING OCCURRED. THEY THEN TRIED A DIFFERENT END CAP (NEEDLESS CONNECTOR) AND THIS RESOLVED THE ISSUE." THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14316472 ↗MalfunctionMay 7, 2022Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation
IT WAS REPORTED "IR PLACED 4F SINGLE LUMEN PICC. CLINICIAN REPORTED LEAKING FROM THE END CAP. END CAP WAS REPLACED A FEW TIMES. A NEW SL PICC WAS PLACED. THE SAME THING OCCURRED. THEY THEN TRIED A DIFFERENT END CAP (NEEDLESS CONNECTOR) AND THIS RESOLVED THE ISSUE." THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14316385 ↗MalfunctionFeb 25, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE PICC LEAKED WHERE THE LUER LOCK CONNECTED TO THE CAP. ONCE AGAIN, CAPS HAD BEEN CHANGED WITH SAME RESULT REPEATING. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13601631 ↗MalfunctionFeb 25, 2022Material Frayed; Material Frayed
IT WAS REPORTED "THE GUIDEWIRE GOT STUCK WITHIN THE NEEDLE, SO THE CLINICIAN WHO WAS PLACING THE LINE REMOVED THE WIRE AND NEEDLE COMPLETELY. THE WIRE THEN FRAYED AFTER IT WAS REMOVED FROM THE NEEDLE OUTSIDE OF THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13601065 ↗MalfunctionFeb 24, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE PICC LEAKED AT THE LUER LOCK CONNECTED TO THE CAP. ADDITIONAL INFO. RECEIVED 02/07/2022: "WE JUST EXCHANGED THE DEFECTIVE PICC ON THE PATIENT. THE LEAK WAS REPRODUCED MULTIPLE TIMES WHERE THE LUER LOCK CONNECTED TO THE CAP. ONCE AGAIN, CAPS HAD BEEN CHANGED WITH SAME RESULT REPEATING. ORIGINAL WAS PLACED (B)(6) 2022 AND EXCHANGED (B)(6) 2022. NO HARM OR INJURY WAS SUSTAINED BY THE PATIENT. "
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13596295 ↗MalfunctionFeb 11, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE FACILITY HAS "GONE THROUGH 3 CATHETERS AND ALL HAD THE TUBE LEAKING." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13518233 ↗MalfunctionFeb 11, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE FACILITY HAS "GONE THROUGH 3 CATHETERS AND ALL HAD THE TUBE LEAKING." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE THIRD DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13518232 ↗MalfunctionFeb 11, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE FACILITY HAS "GONE THROUGH 3 CATHETERS AND ALL HAD THE TUBE LEAKING." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13518229 ↗MalfunctionJan 25, 2022Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation
IT WAS REPORTED "ON 12/30/21, WE PLACED A 4FR SINGLE LUMEN PICC (REF (B)(4)) (LOT#:REFX0144) ON A PATIENT IN THE MORNING. BEFORE END OF DAY, WE WERE ASKED TO INSPECT THIS CATHETER FOR LEAKAGE. UPON INVESTIGATION, THE CATHETER WAS INDEED LEAKING AT THE LUER LOCK CONNECTION WITH THE POSITIVE PRESSURE CAP. THE CAP WAS CHANGED MULTIPLE TIMES, WITH SUBSEQUENT FLUSHING PRODUCING THE SAME RESULT OF LEAKAGE AT THE LUER LOCK CONNECTION. THE CATHETER WAS EXCHANGED THAT DAY FOR A NEW ONE, DIFFERENT LOT, WITH NO FURTHER ISSUES. NO HARM OR INJURY WAS SUSTAINED BY THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13350600 ↗MalfunctionJul 12, 2019Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE WIRE WITHIN THE PROTECTIVE PLASTIC LOOP PACKAGING WAS KINKED, NOT ALLOWING THE WIRE TO BE THREADED THROUGH THE NEEDLE. A SECOND KIT HAD TO BE OPENED. NO PATIENT HARM REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8784726 ↗MalfunctionJul 11, 2019Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE WIRE WITHIN THE PROTECTIVE PLASTIC LOOP PACKAGING WAS KINKED, NOT ALLOWING THE WIRE TO BE THREADED THROUGH THE NEEDLE. A SECOND KIT HAD TO BE OPENED. NO PATIENT HARM REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8782851 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page