PowerPICC SOLO2 Catheter 3194108Recalls & reported problems
Bard Access Systems, Inc. · DOT PowerPICC, Solo, · Class 2 · Implant
All models of PowerPICC SOLO2 Catheter (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741027710), received between Aug 9, 2024 and Nov 18, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Class IOpen, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT ON (B)(6) 2024 I PLACED A POWERPICC SOLO CATHETER. ON (B)(6) 2024 I WAS CALLED TO THE INFUSION DEPARTMENT BECAUSE THE PICC LINE WAS LEAKING. I TOOK THE DRESSING OFF AND PULLED THE PICC AS I FLUSHED IT AND FOUND 2 CRACKS IN THE PICC ABOUT 5CM UP FROM THE HUB. THE PATIENT SAID HIS ARM FELT FULL AFTER HIS LAST INFUSION. THE PICC LINE WAS REMOVED AND I PLACED A 2.25 MIDLINE IN THE PATIENT SO HE COULD RECEIVE HIS TREATMENT. THE FOLLOWING DAY I PLACED ANOTHER PICC LINE IN THE OPPOSITE ARM. NO OTHER INFORMATION PROVIDED, AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20708902 ↗MalfunctionAug 9, 2024Material Frayed; Material Frayed
IT WAS REPORTED, THE STYLET BROKE WHEN THEY REMOVED IT FROM THE PATIENT DUE TO AN UNSUCCESSFUL PLACEMENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19954130 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page