PowerPICC SOLO2 Catheter 3194135Recalls & reported problems
Bard Access Systems, Inc. · PowerPICC SOLO Catheter 4F Radiology Basic Tray (Single-Lumen) (135 cm) · Class 2 · Implant
All models of PowerPICC SOLO2 Catheter (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741027758), received between Apr 23, 2018 and Sep 29, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 11, 2025Class IOpen, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
- Failure to Infuse1
- Material Discolored1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2021Material Discolored; Material Discolored
IT WAS REPORTED VIA MS&S "WHAT COULD CAUSE THE POWERPICC LINE TO TURN BLACK? HAVE YOU HEARD OF A PICC LINE DOING THIS WITH AN ARSENIC TRIOXIDE INFUSION? POWERPICC LINE PLACED 6 MONTHS AGO. PATIENT RECEIVING ARSENIC TRIOXIDE THROUGH POWERPICC LINE. POWERPICC LINE TURNING BLACK FROM HUB TO WINGS. PATIENT ARRIVED AT THE FACILITY 1 MONTH AGO WITH DISCOLORED(BLACK) PICC. NURSE STATED THEY SPOKE WITH OTHER ONCOLOGY NURSES WHO HAVE INFUSED ARSENIC TRIOXIDE WITH POWER PICCS IN THE PAST WITH NO ISSUES. THE PHARMACIST ALSO CONTACTED THE DRUG MANUFACTURER AND STATED THERE ARE NO CONTRAINDICATIONS. PATIENT SCHEDULED TO HAVE POWERPICC EXCHANGED. FACILITY REQUESTING A PRODUCT RETURN KIT AS SOON AS POSSIBLE. NURSE STATES SHE WILL CHECK THE CHART FOR A REF # AND LOT# AND CALL BACK." ADDITIONAL INFO. RECEIVED 07/13/2021: WHAT TYPE OF CATHETER SECUREMENT WAS UTILIZED? DURING HER STAY AT FACILITY, WE UTILIZED STATLOCK. PATIENT ADMITTED (B)(6) 2021. DATE OF OCCURRENCE: FIRST DOCUMENTED (B)(6) 2021. PATIENT STATES IT WAS THIS COLOR FOR ¿MANY MONTHS¿. STAFF ALL THOUGHT IT WAS SUPPOSED TO LOOK THAT WAY UNTIL SEEN BY MYSELF ON (B)(6) 2021. WHAT THEY'RE BEING TREATED FOR: ACUTE PROMYELOCYTIC LEUKEMIA. CONC
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12550883 ↗MalfunctionAug 25, 2021Failure to Infuse; Failure to Infuse
IT WAS REPORTED "PATIENT RECEIVED 3 DOSES OF ALTEPLASE DURING HER STAY." MS&S RESPONSE "RESPONSE INFORMED WE DO NOT HAVE ANY TESTING WITH ARSENIC TRIOXIDE IN POWERPICC LINES. SUGGESTED SHE CONTACT THE DRUG MANUFACTURER FOR INFUSION RECOMMENDATIONS AND CATHETER COMPATIBILITY. WE RECOMMEND THE PICC LINE BE REPLACED WITH A NEW LINE THAT IS COMPATIBLE WITH THE DRUG MANUFACTURE GUIDELINES."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368772 ↗MalfunctionApr 23, 2018Break
IT WAS REPORTED THAT ON (B)(6) 2018 THE PATIENT EXPERIENCED A BROKEN PICC LINE. THE PICC WAS INSERTED ON (B)(6) 2018. THERE WAS NO REPORTED PATIENT INJURY. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7450640 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page