PowerPICC SOLO2 Catheter with Sherlock Tip Location System (TLS) Stylet 9194108Recalls & reported problems
Bard Access Systems, Inc. · PowerPICC SOLO Catheter with Tip Location System (TLS) Stylet 4F Nurse Full Tray with Microintroducer (with Lidocain) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741028595), received between Jan 31, 2019 and Jun 9, 2022 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Malposition of Device” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 11, 2025Class IOpen, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Malposition of Device3
- Adverse Event Without Identified Device or Use Problem1
- Coagulation in Device or Device Ingredient1
- Fluid/Blood Leak1
Patient problems
- Foreign Body In Patient1
- Itching Sensation1
- Nerve Damage1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- Pain1
- Rash1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED A FRACTURED DEVICE. THE POWER PICC LINE WAS INSERTED INTO ONE OF OUR PATIENTS IN FACILITY. 4FR POWER PICC SOLO SINGLE LUMEN CATHETER. INSERTED INTO RIGHT BRACHIAL VEIN WITHOUT ISSUE. INSERTION DATE (B)(6) 2022 AT FACILITY. REMOVED ON (B)(6) 2022 DUE TO LEAK FROM SECURE-CATH AREA. FRACTURE PRESENT AT 12CM. LINE LENGTH: 45CM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14650644 ↗MalfunctionJan 4, 2021Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED VIA (B)(4): "PATIENT WITH POWERPICC SOLO (REFERENCE# 9194108. LOT# REDY2960) PLACED ON (B)(6) 2020 AND REMOVED ON (B)(6) 2020 ASKING FOR POWERPICC SOLO CATHETER MATERIAL CONTENT STATING SHE HAD A REACTION TO THE CATHETER. CALLER STATES HER DOCTOR REMOVED IT AND THERE WAS A 3" FIBRIN CLOT ON THE END OF THE CATHETER HER DOCTOR SAID MAY BE RELATED TO ALLERGY. CALLER STATES SHE HAS LYME'S DISEASE AND HAS INCREASED FIBRIN ACTIVITY. CALLER STATES SHE DOES NOT THINK THERE WAS A DEVICE PROBLEM BUT THAT SHE REACTED TO THE DEVICE MATERIAL. CALLER ALSO ASKING IN GENERAL FOR MATERIAL CONTENT THAT BARD/BD PORTS ARE MADE OF STATING SHE ASKED HER DOCTOR AND THEY SAID TO CALL MANUFACTURER." (B)(4) RESPONSE: "INFORMED CALLER POWERPICC SOLO IS A POLYURETHANE CATHETER WITH SILICONE VALVES. INFORMED THAT IN GENERAL, BARD/BD PORTS ARE TITANIUM OR PLASTIC WITH SILICONE SEPTUMS AND EITHER A POLYURETHANE CATHETER OR A SILICONE (GROSHONG) CATHETER. EXPLAINED THAT HER HEALTHCARE PROVIDER CAN ALSO CALL AND PROVIDED MY NAME PER REQUEST." ADDITIONAL INFORMATION RECEIVED 12/09/2020: "AS A PATIENT I DO NOT KNOW THE TYPE OF ATTACHMENT PRODUCT THAT WAS USED. I DO KNOW THAT WHEN CHANGED IT CLICKED I
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11110699 ↗MalfunctionJan 4, 2021Coagulation in Device or Device Ingredient
IT WAS REPORTED VIA MS&S: "PATIENT WITH POWERPICC SOLO (REFERENCE# (B)(4). LOT# REDY2960) PLACED ON (B)(6) 2020 AND REMOVED ON (B)(6) 2020 ASKING FOR POWERPICC SOLO CATHETER MATERIAL CONTENT STATING SHE HAD A REACTION TO THE CATHETER. CALLER STATES HER DOCTOR REMOVED IT AND THERE WAS A 3" FIBRIN CLOT ON THE END OF THE CATHETER HER DOCTOR SAID MAY BE RELATED TO ALLERGY. CALLER STATES SHE HAS LYME'S DISEASE AND HAS INCREASED FIBRIN ACTIVITY. CALLER STATES SHE DOES NOT THINK THERE WAS A DEVICE PROBLEM BUT THAT SHE REACTED TO THE DEVICE MATERIAL. CALLER ALSO ASKING IN GENERAL FOR MATERIAL CONTENT THAT BARD/BD PORTS ARE MADE OF STATING SHE ASKED HER DOCTOR AND THEY SAID TO CALL MANUFACTURER." MS&S RESPONSE: "INFORMED CALLER POWERPICC SOLO IS A POLYURETHANE CATHETER WITH SILICONE VALVES. INFORMED THAT IN GENERAL, BARD/BD PORTS ARE TITANIUM OR PLASTIC WITH SILICONE SEPTUMS AND EITHER A POLYURETHANE CATHETER OR A SILICONE (GROSHONG) CATHETER. EXPLAINED THAT HER HEALTHCARE PROVIDER CAN ALSO CALL AND PROVIDED MY NAME PER REQUEST." ADDITIONAL INFORMATION RECEIVED 12/09/2020: "AS A PATIENT I DO NOT KNOW THE TYPE OF ATTACHMENT PRODUCT THAT WAS USED. I DO KNOW THAT WHEN CHANGED IT /CLICKED IN PL
Patient problems: Itching Sensation; Rash
MAUDE report 11109363 ↗InjuryFeb 1, 2019Malposition of Device
IT WAS REPORTED PICC WAS PLACED BY PICC RN AT BEDSIDE, CHEST X-RAY OBTAINED, X-RAY DICTATED TIP OF CATHETER IN INNOMINATE VEIN. PATIENT BROUGHT TO INTERVENTIONAL RADIOLOGY FOR RIGHT SIDED PICC PLACEMENT DUE TO NERVE PAIN AT INSERTION SITE AND MALPOSITION OF PICC. THIS REPORT PERTAINS TO THE NERVE DAMAGE.
Patient problems: Nerve Damage
MAUDE report 8297918 ↗MalfunctionFeb 1, 2019Malposition of Device
IT WAS REPORTED PICC WAS PLACED BY PICC RN AT BEDSIDE, CHEST X-RAY OBTAINED, X-RAY DICTATED TIP OF CATHETER IN INNOMINANT VEIN. PATIENT BROUGHT TO INTERVENTIONAL RADIOLOGY FOR RIGHT SIDED PICC PLACEMENT DUE TO NERVE PAIN AT INSERTION SITE AND MALPOSITION OF PICC. THIS REPORT ADDRESS THE MALPOSITION OF THE PICC.
Patient problems: Foreign Body In Patient
MAUDE report 8297635 ↗InjuryJan 31, 2019Malposition of Device
IT WAS REPORTED PICC WAS PLACED BY PICC RN AT BEDSIDE, CHEST X-RAY OBTAINED, X-RAY DICTATED TIP OF CATHETER IN INNOMINATE VEIN. PATIENT BROUGHT TO INTERVENTIONAL RADIOLOGY FOR RIGHT SIDED PICC PLACEMENT DUE TO NERVE PAIN AT INSERTION SITE AND MALPOSITION OF PICC. SECOND CHEST X-RAY CONFIRMED FOREIGN OBJECT IN RIGHT VENTRICLE, MEASURING 6CM. CT SCAN PERFORMED IDENTIFYING AS WIRE STYLET. X-RAY ON ADMISSION IN ED CONFIRMED NO METAL OBJECT AT ADMISSION. PATIENT PLACED ON TELEMETRY INDICATING MULTIPLE RUNS OF PVC'S. PICC REMOVED, AND IR PHYSICIAN REMOVED WIRE. WIRE WAS SENT TO PATHOLOGY FOR REVIEW. PATIENT ON TELEMETRY WITH NO ILL EFFECTS. THIS REPORT ADDRESSES THE WIRE FRAGMENT.
Patient problems: Pain
MAUDE report 8295255 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record