PowerPICC SOLO2 Catheter with Sherlock Tip Location System (TLS) Stylet 9194118Recalls & reported problems
Bard Access Systems, Inc. · PowerPICC SOLO2 Catheter with Tip Location System (TLS) Stylet 4F Basic Tray with Microintroducer (Single-Lumen) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00801741028618), received between Nov 17, 2015 and Mar 18, 2025 — classified as 4 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Material Frayed” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Class IOpen, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Frayed4
- Fluid/Blood Leak3
- Material Deformation2
- Break1
- Defective Component1
- Device Dislodged or Dislocated1
- Difficult to Advance1
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient3
- Chest Tightness/Pressure1
- Distress1
- Erythema1
- Foreign Body In Patient1
- Low Blood Pressure/ Hypotension1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2025Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation
IT WAS REPORTED THE PICC CATHETER BROKE BETWEEN 6-7 CM. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21635420 ↗MalfunctionNov 16, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED BY CUSTOMER THAT PT HAD THIS PICC INSERTED (B)(6) ,2020. ON (B)(6) ,2023 IT WAS NOTED THAT THIS PICC WAS MAL POSITIONED. NEW PICC PLACED WITH NO ISSUES. PICC REMOVED BY UNIT RN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18154943 ↗InjuryOct 19, 2023Break; Break
IT WAS REPORTED BY THE CUSTOMER THAT THE POWERPICC SOLO BROKE AT APPROX. THE 15CM MARK WHEN HCP ATTEMPTED TO REMOVE. PATIENT HAD PICC INSERTED IN 2020 IN THE LEFT BASILIC VEIN FOR CHEMOTHERAPY TREATMENTS, UNSURE OF ANY OTHER MEDICATIONS THAT WERE INFUSED. THE CUSTOMER STATED THE X-RAY SHOWED THE PICC IS FRACTURED IN 2 PLACES, FIRST BREAK AT THE AXILLIA. THE BROKEN SECTIONS REMAIN IN THE BRACHIOCEPHALIC VEIN. FURTHER INVESTIGATION IS REQUIRED BY RADIOLOGY TO ASSESS IF THE PIECES CAN BE REMOVED. PICC IN-SITU FOR APPROX. 4 YEARS, BROKEN SECTION OF PICC REMAINS IN THE PATIENT, UNCERTAIN OF REMOVAL. ADDITIONAL INFORMATION RECEIVED 17-OCT-2023: IT WAS REPORTED THE EVENT WAS NON-LIFE-THREATENING. THE NURSE ON SHIFT ATTEMPTED TO REMOVE THE CATHETER AND DISCOVERED ONLY 15 CM WAS REMOVED. UPON FURTHER X-RAY, IT WAS NOTED TWO PIECES OF THE CATHETER REMAINED INSIDE THE PATIENT. THERE WERE NO DELAYS IN TREATMENT BUT THE PATIENT WAS VERY UPSET ABOUT THE SITUATION. RADIOLOGY WAS ABLE TO REMOVE THE LARGER PORTION OF THE CATHETER BUT FELT THAT ONE SMALL FRAGMENT REMAINED. IT COULD NOT BE RETRIEVED. CXRS AND FLUOROSCOPY WERE USED TO ASSESS THE INTERNAL PORTION OF THE CATHETER. NO MEDICATIONS WERE IN
Patient problems: Distress; Foreign Body In Patient
MAUDE report 17969448 ↗MalfunctionAug 22, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT AFTER CLEAR FLUID NOTED IN DRUG WHEN PICC FLUSHED, APPROXIMATELY 7-8CM THERE IS A SMALL PINHOLE IN THE PICC LINE. THIS WOULD HAVE BEEN APPROXIMATELY 3-4 CM FROM THE INSERTION PIT/FIT UNDER THE SKIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17597536 ↗InjuryJun 23, 2021Material Frayed; Material Frayed
IT WAS REPORTED "PICC WAS INSERTED USING THE STYLET ON CXR, NOTED FOREIGN BODY AT THE SUBCLAVIAN... 3CM OF STYLET WIRE. PATIENT WILL BE TRANSFERRED TO INTERVENTIONAL RADIOLOGY FOR REMOVAL." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12049951 ↗MalfunctionApr 2, 2021Material Frayed; Material Frayed
IT WAS REPORTED THAT THE NEEDLE WENT INTO THE VEIN FIND AND WHEN THE GUIDEWIRE ADVANCED, THE PATIENT CLAIMED TO FEEL THAT. THE GUIDEWIRE SEEMED TO LOOP AND THE NURSE PULLED BACK ON THE GUIDEWIRE, ENCOUNTERING RESISTANCE WHEN PULLING BACK, BUT FINALLY ABLE TO SLOWLY PULL OUT THE GUIDEWIRE; HOWEVER, IT WAS VERY CLEAR THAT THE WIRE WAS FRAYED, UNRAVELLING, AND BENT AT THE TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11608847 ↗MalfunctionMar 19, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT ON (B)(6) 2019, NURSE ASSESSED THAT THERE WAS FAIR BLOOD RETURNS AND QUERIED LEAKING SALINE AND SOME BLOOD AT SITE WHEN PT WAS IN FOR PICC CARE AND BLOOD WORK AT ONCOLOGY. PICC NURSE ASSESSED SAME DAY AND FOUND NO LEAK AT THAT TIME. ON (B)(6) 2019 PT RECEIVED CHEMO TX PACLITAXEL AND CARBOPLATIN, AND PICC NURSE WAS NOTIFIED THE FOLLOWING WEEK THAT THE TX ROOM NURSE FOUND LEAKING FROM INSERTION SITE WITH NO BLOOD RETURNS AGAIN. PICC NURSE CONFIRMED LEAKING AT THE SITE. PICC WAS WITHDRAWN 2 CM. FLUSHED AND STILL HAD LEAKING. A PICC EXCHANGE WAS ATTEMPTED AND UNSUCCESSFUL DUE TO LINE CURLING IN THE AXILLA. LINE WAS REMOVED. NEW LINE INSERTION IN NEW SITE ON SAME DATE ((B)(6) 2019). ON REVIEW OF THE ORIGINAL LINE, AN HOLE WAS FOUND AT THE 20 CM MARK.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9856807 ↗MalfunctionFeb 28, 2020Material Frayed; Material Frayed
IT WAS REPORTED THAT AT THE FACILITY WHILE INSPECTING THE GUIDEWIRE, THE TIP BROKE OFF EASILY. THE DEVICE WAS NOT USED ON A PATIENT. IT WAS STATED THIS HAPPENED WITH TWO DEVICES WITH LOT NUMBER REDW3820. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9770159 ↗MalfunctionFeb 28, 2020Material Frayed; Material Frayed
IT WAS REPORTED THAT AT THE FACILITY WHILE INSPECTING THE GUIDEWIRE, THE TIP BROKE OFF EASILY. THE DEVICE WAS NOT USED ON A PATIENT. IT WAS STATED THIS HAPPENED WITH TWO DEVICES WITH LOT NUMBER REDW3820. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9770152 ↗MalfunctionMar 8, 2019Defective Component
IT WAS REPORTED THAT POWER PICC CONTINUES TO BLOCK OFF REPEATEDLY. THIS IS APPARENTLY A RECURRING EVENT. THE NUMBER OF OCCURRENCES WAS NOT PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8404556 ↗InjuryMar 2, 2018Insufficient Device Problem Information
IT WAS REPORTED THAT ON COMPLETION OF PICC INSERTION PATIENT BECAME FLUSHED, HYPOTENSIVE AND TACHYCARDIC. USER SUGGESTS LIKELY MAST CELL ACTIVATION REACTION. PATIENT APPARENTLY REQUIRED SUPPORTIVE MEASURES: 2L CRYSTALLOID, BENADRYL, STEROIDS. IT ALLEGEDLY TOOK CLOSE TO AN HOUR FOR PATIENT TO RECOVER AND CONSIDERATION WAS BEING MADE FOR AN ICU CONSULT.
Patient problems: Erythema; Low Blood Pressure/ Hypotension; Tachycardia
MAUDE report 7311521 ↗MalfunctionJan 17, 2018Failure to Advance
IT WAS REPORTED THAT THE PICC WAS INSERTED VIA LEFT ARM BASILICA VEIN, INITIAL DIFFICULTY IN ADVANCING THE CATHETER AS CATHETER WAS GOING TO THE RIGHT IJ, EVENTUALLY WAS ABLE TO DIRECT CATHETER TO DISTAL SVC, FIBRIN NOTED IN CATHETER LUMEN, FIBRIN WAS WITHDRAWN FROM SYRINGE AND SYRINGE FLUSHED WITH NORMAL SALINE, DRESSING COMPLETED. PATIENT SUPPOSEDLY STATED HAVING CHEST PRESSURE, VITAL SIGNS STABLE, NO SHORTNESS OF BREATH. NO APPARENT HARM TO THE PATIENT.
Patient problems: Chest Tightness/Pressure
MAUDE report 7201200 ↗MalfunctionOct 31, 2017Scratched Material
IT WAS REPORTED THAT NURSE NOTICED ROUGHNESS OF THE OUTER PORTION OF THE CATHETER ALONG THE CENTIMETER MARKS WHEN RUNNING HER FINGERS ALONG THE EDGE COMPARED TO OTHER CATHETERS. IT WAS STATED THE NURSE FELT AN INCREASE IN "BUMPINESS" RATHER THAN SMOOTHNESS. THE PROCEDURE WAS COMPLETED AND THE CATHETER LEFT IN SITU. THE CATHETER WAS LATER REMOVED. THIS REPORTED ADDRESSES CATHETER THREE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6990369 ↗MalfunctionOct 31, 2017Material Integrity Problem
IT WAS REPORTED THAT NURSE NOTICED ROUGHNESS OF THE OUTER PORTION OF THE CATHETER ALONG THE CENTIMETER MARKS WHEN RUNNING HER FINGERS ALONG THE EDGE COMPARED TO OTHER CATHETERS. IT WAS STATED THE NURSE FELT AN INCREASE IN "BUMPINESS" RATHER THAN SMOOTHNESS. THE PROCEDURE WAS COMPLETED AND THE CATHETER LEFT IN SITU. THE CATHETER WAS LATER REMOVED. THIS REPORTED ADDRESSES CATHETER TWO.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6990367 ↗MalfunctionOct 31, 2017Material Deformation
IT WAS REPORTED THAT NURSE NOTICED ROUGHNESS OF THE OUTER PORTION OF THE CATHETER ALONG THE CENTIMETER MARKS WHEN RUNNING HER FINGERS ALONG THE EDGE COMPARED TO OTHER CATHETERS. IT WAS STATED THE NURSE FELT AN INCREASE IN "BUMPINESS" RATHER THAN SMOOTHNESS. THE PROCEDURE WAS COMPLETED AND THE CATHETER LEFT IN SITU. THE CATHETER WAS LATER REMOVED. THIS REPORTED ADDRESSES CATHETER ONE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6990365 ↗MalfunctionMay 23, 2017Kinked; Occlusion Within Device
IT WAS REPORTED THAT THE PICC LINE WAS FOUND TO BE ALLEGEDLY KINKED IN LEFT ARM AND WAS REMOVED. ORIGINALLY INSERTED TO UPPER SVC, OCCLUSION NOT RESOLVED WITH TPA. CXR SHOWED LINE 10CM SHORTER BUT NO ADDITIONAL LENGTH ON SKIN. OBVIOUS KINK WHEN REMOVED. INFUSED CT PREVIOUS DAY. IT WAS NOTED THAT THIS HAS OCCURRED IN THE PAST NUMEROUS TIMES IN LATE 2015 AND EARLY 2016. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6585332 ↗InjuryJan 26, 2016Knotted; Difficult to Remove; Physical Resistance; Difficult to Advance
IT WAS REPORTED THAT THE CUSTOMER ACCESSED THE RIGHT BASILICA VEIN AND ATTEMPTED TO ADVANCE THE WIRE. THEY WERE ABLE TO ADVANCE IT 15CM THEN APPARENTLY MET RESISTANCE. THEY WERE APPARENTLY UNABLE TO REMOVE THE WIRE. A DERMATOTOMY WAS PERFORMED. THE INTRODUCER WAS INSERTED TO THE POINT OF RESISTANCE AND THEN THEY WERE ABLE TO REMOVE THE WIRE. THEY REPORTED A KNOT IN THE WIRE.
Patient problems: No Code Available
MAUDE report 5391700 ↗MalfunctionNov 17, 2015Malposition of Device
IT WAS REPORTED THAT PICC WAS ALLEGEDLY MALPOSITIONED CONTRALATERALLY DURING CT SCAN. - REPORTEDLY, SCOUT SCAN WAS DONE BEFORE CT SCAN AND CONFIRMED THAT PICC TIP WAS IN RIGHT LOCATION (CAVOATRIAL JUNCTION). - POWERPICC WAS USED FOR CT SCAN AND ALLEGEDLY, MALPOSITIONED DURING SCAN. - OVER COURSE OF NEXT 3 DAYS, PICC NURSES REPORTEDLY USED VARIOUS FLUSHING TECHNIQUES TO GET PICC BACK TO PROPER POSITION; ALLEGEDLY, GOT PICC TO GO BACK DOWN SVC. - REPORTEDLY, PICC NURSE CONFIRMED CT WON锠USE THE PICC FOR ANYMORE CT SCANS.
Patient problems: No Information
MAUDE report 5231072 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.