Skip to content

PowerPICC Catheter with Sherlock Tip Location System (TLS) Stylet 9174108DRecalls & reported problems

Bard Access Systems, Inc. · PowerPICC Catheter with Sherlock Tip Location System (TLS) Stylet 4F Maximal Barrier Tray · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741035555), received between Apr 7, 2017 and May 15, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Class IOpen, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Air Leak1
  • Connection Problem1
  • Difficult or Delayed Separation1
  • Leak/Splash1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2025Material Deformation; Difficult or Delayed Separation; Material Deformation; Difficult or Delayed Separation

    ADDITIONAL INFORMATION RECEIVED ON 21-MAY-2025: DURING THE PROCEDURE. THE NURSE HAD THREADED THE LINE. WHEN SHE WENT TO "PEEL" AWAY THE SHEATH THE ORANGE PORTION ON ONE SIDE SNAPPED OFF INTO HER HAND. THE NURSE WAS ABLE TO CAREFULLY REMOVE THE REST OF THE INTRODUCER FROM THE PATIENT. NO PORTION WAS RETAINED. NO HARM AND NO INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031931 ↗
  • MalfunctionMar 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER THAT "BLOOD RETURN NOTED ON FLOW SHEETS FOR MULTIPLE DAYS. NO TPA ADMINISTRATIONS OR POWER INJECTIONS NOTED IN CHART. PATIENT REQUIRED NO ADDITIONAL TREATMENT OR INTERVENTION ONCE LINE REMOVED. UPON REMOVAL, TEAR IN EXTERNAL PORTION OF THE LINE BETWEEN 35-40CM MARKINGS. THE TEAR SEEMS TO BE 5CM LONG."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16542239 ↗
  • MalfunctionNov 21, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT POWERPICC WAS FOUND LEAKING. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9354482 ↗
  • MalfunctionNov 20, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT POWERPICC WAS FOUND LEAKING. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9349486 ↗
  • MalfunctionApr 7, 2017Air Leak; Leak/Splash; Connection Problem

    PER SALES REP, THE PRIMARY NURSE AT THE FACILITY STATED THE PICC LINE WAS LEAKING AT THE HUB TUBING CONNECTION. IT WAS REPORTED THAT THE RN NOTED AIR IN THE LINE WHEN TRYING TO DRAW BLOOD. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6474185 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page