Skip to content

Broviac 0601610Recalls & reported problems

Bard Access Systems, Inc. · 4.2F Broviac CV Catheters 0.7 mm Lumen (White Adapter) Repair Kit · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Broviac repair kit is used to fix long-term central venous catheters that run through the chest. Most of the 67 reports describe the tubing cracking or breaking, fluid or blood leaking out, and blood clotting inside the line, with 6 reports of patient injury and no deaths. One open recall, started in January 2023, covers the adhesive in the repair kits hardening or coagulating so they cannot be used.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
67
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI 00801741036385), received between Apr 8, 2016 and Jun 4, 2026 — classified as 6 injuries, 61 malfunctions.

The most frequently reported specific device problem is “Fracture” (14 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 4, 2023Open, Classified

    Due to the adhesive provided within the catheter repair kits maybe hardened/coagulated (unusable).

    Z-1051-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Fracture14
  • Coagulation in Device or Device Ingredient10
  • Fluid/Blood Leak10
  • Nonstandard Device10
  • Stretched10
  • Material Protrusion/Extrusion7
  • Break6
  • Loss of or Failure to Bond6

Patient problems

  • No Clinical Signs, Symptoms or Conditions35
  • No Consequences Or Impact To Patient18
  • No Known Impact Or Consequence To Patient5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Hernia2
  • Bacteremia1
  • Erythema1
  • Fever1

Recent adverse event reports

The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Material Discolored; Air/Gas in Device

    A PATIENT UNDERWENT CENTRAL VENOUS CATHETER PLACEMENT PROCEDURE USING HICKMAN/LEONARD/BROVIAC CENTRAL VENOUS CATHETER. AFTER SOME TIME, IT WAS NOTED THAT THE EXTERNAL PORTION OF THE CATHETER TUBING HAD TURNED PINK, AND AN AIR BUBBLE WAS OBSERVED WITHIN IT IN THE EXTERNAL PORTION THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25391246 ↗
  • MalfunctionDec 1, 2025Burst Container or Vessel; Fracture; Burst Container or Vessel; Fracture

    A PATIENT UNDERWENT A REPAIR KIT PLACEMENT PROCEDURE USING THE HICKMAN REPAIR KIT. ON (B)(6) 2025, THE CRACK WAS NOTED ON REPAIR LINE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23678597 ↗
  • MalfunctionAug 14, 2024Stretched; Material Protrusion/Extrusion; Stretched; Material Protrusion/Extrusion

    IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, WHEN POSITIVE PRESSURE WAS APPLIED, A PART OF THE CATHETER WAS ALLEGEDLY FOUND TO BE EXPANDED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19981655 ↗
  • MalfunctionAug 6, 2024Loss of or Failure to Bond; Device Dislodged or Dislocated; Loss of or Failure to Bond; Device Dislodged or Dislocated

    IT WAS REPORTED THAT SOMETIMES POST A CHRONIC CATHETER PLACEMENT, THE METAL STENT IN THE REPAIR KIT ALLEGEDLY MIGRATED. THE PROCEDURE WAS COMPLETED BY REPLACING WITH ANOTHER KIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917600 ↗
  • MalfunctionJan 31, 2024Loss of or Failure to Bond; Material Separation; Device Dislodged or Dislocated; Loss of or Failure to Bond; Material Separation; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A CHRONIC CATHETER REPAIR PROCEDURE, THE METAL PIECE ALLEGEDLY DID NOT STAY UNDER THE SLEEVE AFTER THE REPAIR. IT WAS FURTHER REPORTED THAT THE EVENT CAUSED AN ADDITIONAL REPAIR OF THE LINE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18611951 ↗
  • MalfunctionDec 13, 2023Fracture; Fracture

    IT WAS REPORTED IN AN ARTICLE IN THE JOURNAL OF THE ¿JAPANESE SOCIETY OF PEDIATRIC SURGEONS¿ TITLED "COMPLICATIONS AND MANAGEMENT OF CENTRAL VENOUS CATHETERS OF LONG-TERM TOTAL PARENTAL NUTRITION", THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE EXTERNAL PART OF THE CATHETER WAS ALLEGEDLY BROKEN. REPORTEDLY, THE BROKEN CATHETER WAS REPAIRED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18316669 ↗
  • InjuryDec 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED IN AN ARTICLE IN THE JOURNAL OF THE ¿JAPANESE SOCIETY OF PEDIATRIC SURGEONS¿ TITLED "COMPLICATIONS AND MANAGEMENT OF CENTRAL VENOUS CATHETERS OF LONG-TERM TOTAL PARENTAL NUTRITION", THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE PATIENT ALLEGEDLY DEVELOPED A CATHETER TUNNEL INFECTION DUE TO METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS BACTEREMIA. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Bacteremia

    MAUDE report 18310280 ↗
  • MalfunctionDec 12, 2023Fracture; Fracture

    IT WAS REPORTED IN AN ARTICLE IN THE JOURNAL OF THE ¿JAPANESE SOCIETY OF PEDIATRIC SURGEONS¿ TITLED "COMPLICATIONS AND MANAGEMENT OF CENTRAL VENOUS CATHETERS OF LONG-TERM TOTAL PARENTAL NUTRITION", THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE EXTERNAL PART OF THE CATHETER WAS ALLEGEDLY BROKEN. REPORTEDLY, THE BROKEN CATHETER WAS REPAIRED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18310140 ↗
  • MalfunctionSep 14, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT PRIOR TO A CHRONIC CATHETER PLACEMENT PROCEDURE, THE GLUE WAS ALLEGEDLY DRIED UP. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17743363 ↗
  • MalfunctionSep 8, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT PRIOR TO A REPAIR KIT PLACEMENT PROCEDURE, GLUE WAS ALLEGEDLY DRIED UP. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17709471 ↗
  • MalfunctionJul 5, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT PRIOR TO A CHRONIC CATHETER REPAIR PROCEDURE, THE GLUE WAS ALLEGEDLY HARD. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17259358 ↗
  • MalfunctionJun 5, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT DURING PREPARATION OF A CATHETER REPAIR PROCEDURE, THE GLUE HAD THE INTACT FOIL SEAL OVER THE END. IT WAS FURTHER REPORTED THAT THE GLUE WAS ROCK HARD WHEN PUNCTURED THE SEAL WITH THE CAP. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17058604 ↗
  • MalfunctionApr 6, 2023Fracture; Stretched; Material Protrusion/Extrusion; Fracture; Stretched; Material Protrusion/Extrusion

    IT WAS REPORTED THAT SOMETIME POST CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY STARTED OUT AS AN ANUERYSM AND EVENTUALLY TORN. IT WAS FURTHER REPORTED THAT THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16689904 ↗
  • MalfunctionMar 27, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT DURING PREPARATION OF A CATHETER REPAIR PROCEDURE, THE GLUE WAS ALLEGEDLY HARD. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16617193 ↗
  • MalfunctionMar 27, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT DURING PREPARATION OF A CATHETER REPAIR PROCEDURE, THE GLUE WAS ALLEGEDLY HARD. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16617142 ↗
  • MalfunctionFeb 16, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT DURING PREPARATION OF A CATHETER REPAIRING PROCEDURE, THE ADHESIVE ON THE DEVICE WAS ALLEGEDLY SOLID AND WOULDN'T COME OUT OF TUBE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16385692 ↗
  • MalfunctionFeb 8, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT DURING PREPARATION OF A CATHETER REPAIR PROCEDURE, THE GLUE WAS ALLEGEDLY HARDENED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16324166 ↗
  • MalfunctionFeb 7, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT DURING PREPARATION OF A CATHETER REPAIR PROCEDURE, THE GLUE WAS ALLEGEDLY HARDENED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16316288 ↗
  • MalfunctionJan 31, 2023Coagulation in Device or Device Ingredient; Nonstandard Device; Coagulation in Device or Device Ingredient; Nonstandard Device

    IT WAS REPORTED THAT DURING PREPARATION OF A CATHETER REPAIR PROCEDURE, THE GLUE WAS ALLEGEDLY HARDENED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16270156 ↗
  • MalfunctionJan 12, 2023Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT APPROXIMATELY FOURTEEN DAYS POST CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY DETACHED FROM THE PART OF THE ADHESIVE. IT WAS FURTHER REPORTED THAT THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16143673 ↗
  • MalfunctionApr 25, 2022Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    IT WAS REPORTED THAT DURING A CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY FRACTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193974 ↗
  • MalfunctionApr 21, 2022Stretched; Material Protrusion/Extrusion; Stretched; Material Protrusion/Extrusion

    IT WAS REPORTED THAT TWENTY TWO DAYS POST CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY BULGED WITH FLUSHING. THE PROCEDURE WAS COMPLETED BY PLACING REPAIR KIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14165818 ↗
  • MalfunctionNov 2, 2021Fracture; Fracture

    IT WAS REPORTED THAT POST CATHETER PLACEMENT PROCEDURE, THE CATHETER WAS ALLEGEDLY BROKEN. THERE WAS REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12734525 ↗
  • MalfunctionAug 31, 2021Burst Container or Vessel; Fracture; Burst Container or Vessel; Fracture

    IT WAS REPORTED THAT APPROXIMATELY SIX MONTHS POST CATHETER PLACEMENT, THE REPAIR KIT ALLEGEDLY DAMAGED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12390220 ↗
  • MalfunctionJun 16, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING THE CATHETER PLACEMENT, THE TWO REPAIR SHEATHS ALLEGEDLY CRACKED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12006262 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.