Catalog Number / Product Description recallZ-1051-2023 · Bard Peripheral Vascular Inc · initiated Jan 4, 2023
Class II Open, Classified All recalls posted in February 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to the adhesive provided within the catheter repair kits maybe hardened/coagulated (unusable).
FDA root cause category: Under Investigation by firm
What the company is doing
On 01/04/2023, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via FedEx or email to customers informing them that the firm has received complaints regarding hardened/coagulated adhesive associated with various Catheter Repair Kits. as a result of the hardened/coagulated adhesive, there is the potential to cause delays whilst an alternative repair kit, adhesive or replacement catheter is obtained, therefore potentially prolonging surgery or necessitating exchange. Customer are instructed to: 1. Discontinue use of product listed in Attachment A, and immediately dispose of all affected product remaining in their possession in accordance with their local facility s process. 2. Share this notification with all users within their facility network of the product to ensure they are also aware of this field action. 3. If customers purchased this product from a distributor, contact the distributor for further instructions and credit resolution. 4. Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not they have any of the product so that BD may acknowledge their receipt of this notification and subsequently process any applicable credit regarding impacted product. 5. Indicate on the response form the quantity of product identified at their facility and confirm that this product inventory was disposed of. For Distributors (in addition to the above): Identify all customers within their distribution network that purchased any affected product and provide a copy of the attached customer letter to all customers to advise them of this field action notification on BD s behalf. Contact their BD representative if they require assistance with this process or they may also contact the number below for further assistance. North American Regional Complaint Center Phone: 1-844-8BD- LIFE (1-844-823-5433) Say Recall when prompted M-F 8am - 5pm CT Email: [email protected]
Recall details
- Recall number
- Z-1051-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jan 4, 2023
- Posted by the FDA
- Feb 1, 2023
- Recalling firm
- Bard Peripheral Vascular Inc
- Quantity
- 7520 (Breakdown: US 3764; OUS 3243)
- Product code
- LJS
- FDA submissions
- K830406, K871998
- Distribution
- Worldwide Distribution: U.S.(nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; OUS (foreign) countries of: Australia, Belgium, Canada, Hong Kong and Japan. O.U.S.: Australia, Belgium, Canada, Hong Kong, and Japan.
Product description
Catalog Number / Product Description 0601600 2.7F Broviac CV Catheter Repair Kit 0601610 4.2 Fr Broviac CV Catheter Repair Kit 0601620 6.6F Broviac CV Catheter Repair Kit 0601630 9.6F Hickman CV Catheter Repair Kit 0601680 Hickman / Leonard CV Catheter Repair Kit (all except 13.5F) 0601690 Hickman / Leonard CV Catheter Repair Kit (all except 13.5F) 0601700 9F Hickman CV Catheter Repair Kit 0601710 12F Hickman CV Catheter Repair Kit 0601730 Hickman / Leonard CV Catheter Repair Kit (all except 13.5F) 0601740 12.5F Hickman CV Catheter Repair Kit 0601750 10F Leonard CV Catheter Repair Kit 0601760 7F Hickman CV Catheter Repair Kit 0601790 10F Hickman CV Catheter Repair Kit
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Catalog Numbers/Lot-Serial Numbers/UDI Code: 0601610 / REFX4541 / (01)00801741036385(17)230930(10)REFX4541 0601610 / REFX5636 / (01)00801741036385(17)240930(10)REFX5636 0601610 / REFY2820 / (01)00801741036385(17)230930(10)REFY2820 0601610 / REGP0198 / (01)00801741036385(17)241231(10)REGP0198 0601610 / REGP2720 / (01)00801741036385(17)250131(10)REGP2720 0601610 / REGQ1144 / (01)00801741036385(17)250228(10)REGQ1144 0601620 / REES0680 / (01)00801741074479(17)230430(10)REES0680 0601620 / REEU0726 / (01)00801741074479(17)230630(10)REEU0726 0601620 / REEU1536 / (01)00801741074479(17)230630(10)REEU1536 0601620 / REEU3300 / (01)00801741074479(17)230630(10)REEU3300 0601620 / REEV1906 / (01)00801741074479(17)230731(10)REEV1906 0601620 / REFX4543 / (01)00801741074479(17)230930(10)REFX4543 0601630 / REEQ0746 / (01)00801741036408(17)230228(10)REEQ0746 0601630 / REGN2382 / (01)00801741036408(17)241231(10)REGN2382 0601630 / REGQ1143 / (01)00801741036408(17)250228(10)REGQ1143 0601680 / REER3749 / (01)00801741036415(17)230331(10)REER3749 0601680 / REEU1626 / (01)00801741036415(17)230630(10)REEU1626 0601690 / REES0716 / (01)00801741036422(17)230131(10)REES0716 0601690 / REFX1172 / (01)00801741036422(17)230131(10)REFX1172 0601690 / REFY0874 / (01)00801741036422(17)230131(10)REFY0874 0601700 / REES2549 / (01)00801741036439(17)230430(10)REES2549 0601700 / REGQ1154 / (01)00801741036439(17)230731(10)REGQ1154 0601710 / REFY2852 / (01)00801741036446(17)231031(10)REFY2852 0601730 / REGQ1174 / (01)00801741036453(17)240229(10)REGQ1174 0601740 / REFZ0223 / (01)00801741036460(17)230930(10)REFZ0223 0601750 / REFY0744 / (01)00801741036477(17)230731(10)REFY0744 0601760 / REEU1534 / (01)00801741036484(17)230630(10)REEU1534 0601790 / REFW1651 / (01)00801741036514(17)231031(10)REFW1651 7741700 / REET0877 / (01)00801741036798(17)240131(10)REET0877 7741800 / REER0703 / (01)00801741036804(17)240531(10)REER0703 7741800 / REFW3156 / (01)00801741036804(17)250131(10)REFW3156 0601600 / REES06790 / (01)00801741036378(17)220831(10)REES0679 0601600 / REES2528 / (01)00801741036378(17)220831(10)REES2528 0601600 / REES3456 / (01)00801741036378(17)220831(10)REES3456 0601610 / REEQ1683 / (01)00801741036385(17)220831(10)REEQ1683 0601610 / REEQ4180 / (01)00801741036385(17)220831(10)REEQ4180 0601610 / REER3748 / (01)00801741036385(17)220831(10)REER3748 0601610 / REES0715 / (01)00801741036385(17)220831(10)REES0715 0601610 / REET3287 / (01)00801741036385(17)220831(10)REET3287 0601620 / REEQ2751 / (01)00801741074479(17)220831(10)REEQ2751 0601620 / REEQ4141 / (01)00801741074479(17)220831(10)REEQ4141 0601620 / REER3694 / (01)00801741074479(17)220831(10)REER3694 0601620 / REET3269 / (01)00801741074479(17)220831(10)REET3269 0601630 / REEU0729 / (01)00801741036408(17)220831(10)REEU0729 0601630 / REEU3297 / (01)00801741036408(17)220831(10)REEU3297 0601690 / REEU0727 / (01)00801741036422(17)220831(10)REEU0727 0601700 / REER3695 / (01)00801741036439(17)220831(10)REER3695 0601710 / REER3750 / (01)00801741036446(17)220831(10)REER3750 0601710 / REEU3314 / (01)00801741036446(17)220831(10)REEU3314 0601740 / REEU3298 / (01)00801741036460(17)220831(10)REEU3298 0601750 / REEU0730 / (01)00801741036477(17)220831(10)REEU0730 0601750 / REEU1608 / (01)00801741036477(17)220831(10)REEU1608 0601760 / REER3696 / (01)00801741036484(17)220831(10)REER3696 0601760 / REES3457 / (01)00801741036484(17)220831(10)REES3457 0601760 / REET2794 / (01)00801741036484(17)220831(10)REET2794 0601760 / REEU1184 / (01)00801741036484(17)220831(10)REEU1184
Devices named in this recall
15 devices in the FDA device registry whose identifier appears in the recall's code information.
- Broviac 0601610Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall67 reports
- Broviac 0601620Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall29 reports
- Hickman 0601680Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall12 reports
- Hickman 0601760Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall12 reports
- Hickman 0601630Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall8 reports
- Hickman 0601700Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall8 reports
- Hickman 0601690Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall7 reports
- Broviac 0601600Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall5 reports
- Hickman 0601790Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall2 reports
- Hickman 0601730Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall1 report
- Hickman 0601740Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall1 report
- Groshong 7742000Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall1 report
- Hickman 0601710Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall0 reports
- Leonard 0601750Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall0 reports
- Groshong 7741800Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page