Hickman 0601690Recalls & reported problems
Bard Access Systems, Inc. · Red Adapter Leg for Hickman and Leonard Multiple Lumen CV Catheters Repair Kit · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741036422), received between Jan 12, 2017 and Sep 15, 2022 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 4, 2023Open, Classified
Due to the adhesive provided within the catheter repair kits maybe hardened/coagulated (unusable).
Z-1051-2023 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Break5
- Fracture5
- Fluid/Blood Leak2
- Hole In Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT SOMETIME POST PORT PLACEMENT PROCEDURE, THE RED ADAPTER LEG ALLEGEDLY BROKEN AT THE SAME SPOT. IT WAS FURTHER REPORTED THAT THE LUMEN WAS REPAIRED FOR THE THIRD TIME WITH THE REPAIR KIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15425502 ↗MalfunctionSep 15, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT SOMETIME POST PORT PLACEMENT PROCEDURE, THE RED ADAPTER LEG ALLEGEDLY BROKEN AT THE SAME SPOT. IT WAS FURTHER REPORTED THAT THE LUMEN WAS REPAIRED FOR THE THIRD TIME WITH THE REPAIR KIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15425441 ↗MalfunctionSep 14, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT SOMETIME POST PORT PLACEMENT PROCEDURE, THE RED ADAPTER LEG ALLEGEDLY BROKEN AT THE SAME SPOT. IT WAS FURTHER REPORTED THAT THE LUMEN WAS REPAIRED FOR THE THIRD TIME WITH THE REPAIR KIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15416327 ↗MalfunctionSep 14, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT SOMETIME POST PORT PLACEMENT PROCEDURE, THE RED ADAPTER LEG ALLEGEDLY BROKEN AT THE SAME SPOT. IT WAS FURTHER REPORTED THAT THE LUMEN WAS REPAIRED FOR THE THIRD TIME WITH THE REPAIR KIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15416324 ↗MalfunctionAug 17, 2022Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT SOMETIME POST PORT PLACEMENT PROCEDURE, THE RED ADAPTER LEG ALLEGEDLY BROKEN AT THE SAME SPOT. IT WAS FURTHER REPORTED THAT THE LUMEN WAS REPAIRED FOR THE THIRD TIME WITH THE REPAIR KIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15238764 ↗MalfunctionMay 31, 2018Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE FACILITY HAS ALLEGED THE CATHETER HAS HOLES IN THE TUBING, SMALL SPLITS IN THE TUBING AND/OR SEPARATION AT THE REPAIR KIT. THE CATHETER WAS REPAIRED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 7558995 ↗MalfunctionJan 12, 2017Hole In Material
FACILITY CONTACT REPORTED THAT A DLHL 7FR LINE WAS PLACED ON (B)(6) 2016. ON (B)(6) 2016, A PINPOINT HOLE WAS REPORTABLY NOTED AT THE HUB OF THE RED LUMEN. LINE REPAIR REQUIRED. PATIENT WILL BE FOLLOWED ON TO SEE IF LINE INFECTION DETECTED 7 DAYS BEFORE OR AFTER LINE REPAIR OCCURRED. WILL NOTIFY BARD FIELD ASSURANCE IF LINE INFECTION FOUND.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6247542 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.