Hickman 0601700Recalls & reported problems
Bard Access Systems, Inc. · External Catheter Segment for Hickman 9F Round Dual-Lumen CV Catheters Repair Kit · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741036439), received between Feb 6, 2018 and Jul 10, 2023 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 4, 2023Open, Classified
Due to the adhesive provided within the catheter repair kits maybe hardened/coagulated (unusable).
Z-1051-2023 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak4
- Loss of or Failure to Bond2
- Migration2
- Device Dislodged or Dislocated1
- Device Operates Differently Than Expected1
- Material Protrusion/Extrusion1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2023Loss of or Failure to Bond; Material Separation; Device Dislodged or Dislocated; Material Protrusion/Extrusion; Loss of or Failure to Bond; Material Separation; Device Dislodged or Dislocated; Material Protrusion/Extrusion
IT WAS REPORTED THAT THREE YEARS, ONE MONTH, AND TWENTY-ONE DAYS POST CHRONIC CATHETER PLACEMENT, UPON PERFORMING THE CATHETER REPAIR, A METAL STENT SEGMENT WAS ALLEGEDLY FOUND TO BE PROTRUDING OUT OF THE CATHETER AT AN ANGLE. REPORTEDLY, THE CATHETER WAS REPAIRED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17285042 ↗InjurySep 29, 2022Migration; Migration
IT WAS REPORTED THAT DURING CATHETER PLACEMENT PROCEDURE, THE METAL SEGMENT OF THE REPAIR KIT ALLEGEDLY MIGRATED. CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15501671 ↗InjuryAug 8, 2022Loss of or Failure to Bond; Migration; Loss of or Failure to Bond; Migration
IT WAS REPORTED THAT POST CATHETER PLACEMENT PROCEDURE, THE METAL SEGMENT OF THE REPAIR KIT ALLEGEDLY DETACHED AND FLOATING IN PATIENT'S LINE. IT WAS FURTHER REPORTED THAT THE METAL SEGMENT ALLEGEDLY MIGRATED. REPORTEDLY DETACHED PART WAS REMOVED AND REPAIRED THE LINE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15181685 ↗MalfunctionMay 7, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT SOMETIME POST CHRONIC CATHETER PLACEMENT, WHEN REPAIR KIT WAS USED THE CATHETER ALLEGEDLY HAD A LEAK. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11788572 ↗MalfunctionApr 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT SOME TIME POST CATHETER PLACEMENT, THE CATHETER ALLEGEDLY HAD LEAKAGE BETWEEN THE TWO CATHETER LINES AT POINT OF BLUNT NEEDLE INSERTION. IT WAS REPORTED THAT PHYSICIAN DID THE REPAIR USING REPAIR KIT AFTER 10 DAYS THE CATHETER ALLEGEDLY LEAKED AGAIN SAME LOCATION OF SEALED JOINT WHILE FLUSHING ON 0.7 MM LINE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11714162 ↗MalfunctionMar 25, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING A CATHETER PLACEMENT, THE DEVICE ALLEGEDLY HAD LEAKAGE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11566867 ↗MalfunctionMar 25, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING A CATHETER PLACEMENT, THE DEVICE ALLEGEDLY HAD LEAKAGE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11566866 ↗MalfunctionFeb 6, 2018Device Operates Differently Than Expected
IT WAS REPORTED THAT THE NURSE FOUND THE METAL STENT PART TO BE MOVABLE AFTER PRIMING THE PRODUCT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7246315 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page