Bardex® Lubri-Sil® I.C. All-Silicone Temp Sensing Foley Catheter 119312MRecalls & reported problems
C. R. Bard, Inc. · Bardex® Lubri-Sil® I.C. 400 Series Temperature Sensing All-Silicone 5 cc Foley Catheter · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741039898), received between Nov 30, 2018 and Dec 5, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Calcified” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Calcified1
- Fluid/Blood Leak1
- Inaccurate Flow Rate1
- Material Fragmentation1
- Material Rupture1
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 5, 2025Calcified; Calcified
IT WAS REPORTED THAT PATIENT WAS ON THE INCREASED INTRACRANIAL PRESSURE PROTOCOL WHICH INVOLVED BEEN ON A COOLING BLANKET WITH A CENTRAL TEMPERATURE MONITOR. THAT PATIENT TEMPERATURE READING WAS FROM A FOLEY CATHETER THAT WAS PLACED ON (B)(6). AT THE START OF THE 19L SHIFT THE TEMPERATURE READING WAS APPROPRIATE IN THE 36'S. AROUND 0000 ON (B)(6), THE PATIENT TEMPERATURE WAS ELEVATED TO 37.5 AND SCHEDULED TYLENOL WAS GIVEN. THE COOLING BLANKET WAS DECREASING TEMPERATURE APPROPRIATELY AND THEY PLACED ICE PACKS ON THE PATIENT TO TRY AND PREVENT FEVER. BY 0200 THE TEMPERATURE WAS READING 39.5 AND QUICKLY JUMPED TO 40 THEN AT 0300 WAS 41. AT 0234 THE PATIENT STARTED HAVING BRADYCARDIA TO THE 40'S. FELLOW WAS CALLED TO BEDSIDE AND EVALUATED. INCREASED INTRACRANIAL PRESSURE WAS BELOW 20, EVD WAS PATENT, PATIENT HAD APPROPRIATE BLOOD PRESSURE, AND APPROPRIATE PERFUSION. THEY GOT AM LABS EARLY TO ENSURE ELECTROLYTES WERE STABLE AND NO DROP IN HEMOGLOBIN HAD OCCURRED. THEY ALSO FLUSHED THE FOLEY TOWARDS THE BLADDER WITH 10CC OF SALINE TO ENSURE PATENCY BECAUSE IN REPORT THEY WERE TOLD THAT PATIENT HAD ISSUES WITH SEDIMENT BUILDING UP. THE FOLEY TEMP GAVE NO OBVIOUS SIGNS THAT THERE WAS AN I
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23727825 ↗MalfunctionDec 4, 2024No Display/Image; No Display/Image
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE FOLEY CATHETER TEMPERATURE SENSING COMPONENT WAS NOT SUCCESSFUL IN MONITORING PATIENTS' BODY TEMPERATURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20841681 ↗MalfunctionFeb 14, 2024Material Fragmentation; Material Rupture; Material Fragmentation; Material Rupture
IT WAS REPORTED THAT TEMPERATURE SENSING FOLEY CATHETER DEFLATED AND WHEN IT CAME OUT STAFF NOTICED THE ENTIRE BALLOON SECTION WAS MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18703197 ↗MalfunctionNov 30, 2018Inaccurate Flow Rate; Inaccurate Flow Rate; Fluid/Blood Leak
IT WAS REPORTED THAT THE CATHETER DID NOT DRAIN URINE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8118805 ↗
Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119314C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall3,889 reports
- Bardex® I.C. Foley Catheter 0165SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,994 reports
- Bardex® I.C. Foley Catheter 0165SI16C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,052 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES778 reports
- Bardex® I.C. Temperature Sensing Foley Catheter 129414MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES436 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Temp Sensing Foley Catheter 119316MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES231 reports
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119316C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES198 reports
- Bardex® I.C. Foley Catheter 0165SI18C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES181 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page